PVCHECKs EN 2.01 Multifunction PV Tester User Manual
- June 1, 2024
- PVCHECKs
Table of Contents
- EN 2.01 Multifunction PV Tester
- Specifications:
- Product Information:
- Operating Instructions:
- 1. Introduction
- 2. Measuring Efficiency PV Plants
- 3. Measuring Parameters on PV Plant
- 4. Quick Check on PV Modules and Strings (IVCK)
- Q: What are the warranty conditions for PVCHECKs?
- Q: How can I store efficiency measures with PVCHECKs?
EN 2.01 Multifunction PV Tester
“`html
Specifications:
- Product Name: PVCHECKs
- Release: EN 2.01 – 01/10/2019
- Compliance: IEC/EN61010-1
Product Information:
The PVCHECKs is a specialized instrument designed for measuring
efficiency in PV plants. It offers various functions to help users
accurately assess PV modules and strings.
Operating Instructions:
1. Introduction
The PVCHECKs instrument is used to measure efficiency in PV
plants. Before using, ensure compliance with safety measures
according to directive IEC/EN61010-1.
2. Measuring Efficiency PV Plants
To measure efficiency in PV plants using the remote unit
SOLAR-02, follow the instructions provided in section 6.1 of the
user manual.
3. Measuring Parameters on PV Plant
If measuring parameters on a PV plant without using SOLAR-02,
refer to section 6.2 for detailed instructions.
4. Quick Check on PV Modules and Strings (IVCK)
For a quick check on PV modules and strings using IVCK, follow
the steps outlined in section 6.3 of the manual.
FAQ:
Q: What are the warranty conditions for PVCHECKs?
A: The warranty conditions for PVCHECKs can be found in section
12.1 of the user manual.
Q: How can I store efficiency measures with PVCHECKs?
A: To save efficiency measures, refer to section 7.1 of the
manual for detailed instructions on storing data.
“`
PVC HEC Ks
User manual
© Copyright HT ITALIA 2019
Release EN 2.01 – 01/10/2019
© Copyright HT ITALIA 2019
Release EN 2.01 – 01/10/2019
PVCHECKs EN – 1
TABLE OF CONTENTS
PVCHECKs
EN – 2
PVCHECKs
1. PRECAUTIONS AND SAFETY MEASURES………………………………………………………..3 1.1. Preliminary
instructions…………………………………………………………………………………………5 1.2. During
use…………………………………………………………………………………………………………..6 1.3. After
use……………………………………………………………………………………………………………..6 1.4. Definition of measurement
(overvoltage) category ……………………………………………………6
2. GENERAL DESCRIPTION ………………………………………………………………………………….7 2.1. Introduction
…………………………………………………………………………………………………………7 2.2. Instrument
functions……………………………………………………………………………………………..7
3. PREPARATION FOR USE…………………………………………………………………………………..8 3.1. Initial checks
……………………………………………………………………………………………………….8 3.2. Instrument power
supply……………………………………………………………………………………….8 3.3. Storage
………………………………………………………………………………………………………………8
4. NOMENCLATURE ……………………………………………………………………………………………..9 4.1. Instrument
description…………………………………………………………………………………………..9 4.2. Keyboard description
………………………………………………………………………………………….10 4.3. Display
description……………………………………………………………………………………………..10 4.4. Initial screen
………………………………………………………………………………………………………11
5. GENERAL MENU …………………………………………………………………………………………….12 5.1. SET instrument
settings……………………………………………………………………………………13
5.1.1. General ………………………………………………………………………………………………………………………13 5.1.2.
Measuring unit …………………………………………………………………………………………………………….14 5.1.3. Date and
time………………………………………………………………………………………………………………14 5.1.4. Remote
unit/Pyranometer ……………………………………………………………………………………………..15 5.1.5. Irradiance
……………………………………………………………………………………………………………………16 5.1.6. DC Clamp
…………………………………………………………………………………………………………………..16 5.2. EFF Efficiency test
settings for PV installations……………………………………………………18 5.2.1. Instrument
settings……………………………………………………………………………………………………….18 5.2.2. System
parameters………………………………………………………………………………………………………19 5.2.3. Selection of the
compensation relationship of temperature effects ……………………………………..21 5.3. LOW
Settings for continuity test with 200mA …………………………………………………….23 5.3.1.
Instrument settings……………………………………………………………………………………………………….23 5.4. M
Settings for insulation measurement…………………………………………………………….25 5.4.1.
Instrument settings……………………………………………………………………………………………………….25 5.5. IVCK
Settings for IVCK quick test………………………………………………………………………27 5.5.1. Instrument
settings……………………………………………………………………………………………………….27 5.6. DB Database module
management……………………………………………………………………30 5.6.1. How to define a new PV
module…………………………………………………………………………………….32 5.6.2. How to modify an existing PV
module …………………………………………………………………………….35 5.6.3. How to delete an existing PV
module ……………………………………………………………………………..37
6. OPERATING INSTRUCTIONS …………………………………………………………………………..38 6.1. Measuring
efficiency PV plants by using remote unit SOLAR-02 ………………………………38 6.2.
Measuring parameters on PV plant without using SOLAR-02 …………………………………..42 6.3.
Quick check on PV modules and strings (IVCK)……………………………………………………..46
6.3.1. Foreword…………………………………………………………………………………………………………………….46 6.3.2. Carrying
out an IVCK quick test without measuring irradiance……………………………………………47 6.3.3.
Carrying out an IVCK quick test and measuring irradiance
………………………………………………..52 6.3.4. Reset
Averages……………………………………………………………………………………………………………58
6.3.4.1. Anomalous situations when performing IVCK tests
…………………………………………………………………..60
6.4. Measurement of insulation on PV modules/strings/fields (M)
…………………………………63 6.4.1. Foreword…………………………………………………………………………………………………………………….63
6.4.2. Measuring insulation FIELD mode……………………………………………………………………………….64
6.4.3. Measuring insulation TIMER mode………………………………………………………………………………66
6.4.4. Measuring insulation STRING mode ……………………………………………………………………………68
6.4.4.1. Anomalous situations ……………………………………………………………………………………………………………71
6.5. Measurement of continuity on PV modules/strings/fields (LOW)
…………………………….73 6.5.1. Foreword…………………………………………………………………………………………………………………….73
6.5.2. Calibration of measuring cables……………………………………………………………………………………..73
6.5.3. Measuring continuity …………………………………………………………………………………………………….75
6.5.3.1. Anomalous situations ……………………………………………………………………………………………………………77
6.6. List of display messages……………………………………………………………………………………..79
EN – 3
PVCHECKs
7. STORING DATA……………………………………………………………………………………………….81 7.1. Saving efficiency
measures …………………………………………………………………………………82 7.2. Saving IVCK, M and LOW
measured results …………………………………………………….82 7.3. Operations with results
……………………………………………………………………………………….85
7.3.1. Recalling on the display the PV efficiency
results……………………………………………………………..85 7.3.2. Recalling on the display the results
of IVCK, M and LOW measurements……………………….86
7.3.2.1. Accessing the data saved in the memory Numeric
display………………………………………………………88 7.3.3. Deleting data from the
memory………………………………………………………………………………………91
8. CONNECTING THE INSTRUMENT TO THE PC ………………………………………………….92 9. MAINTENANCE
……………………………………………………………………………………………….95
9.1. General information ……………………………………………………………………………………………95 9.2. Battery
replacement……………………………………………………………………………………………95 9.3. Cleaning the instrument
………………………………………………………………………………………95 9.4. End of life
………………………………………………………………………………………………………….95 10.TECHNICAL
SPECIFICATIONS…………………………………………………………………………96 10.1. Technical specifications
for PV installation efficiency ………………………………………………96 10.2. Technical
specifications of the IVCK function …………………………………………………………97 10.3. Technical
specifications for electrical safety …………………………………………………………..97 10.4.
Reference standards…………………………………………………………………………………………..98
10.4.1. General ………………………………………………………………………………………………………………………98 10.5. General
characteristics ……………………………………………………………………………………….98 10.6. Environmental
conditions for use………………………………………………………………………….99 10.7.
Accessories……………………………………………………………………………………………………….99
11.APPENDIX THEORETICAL OUTLINE ……………………………………………………………100 11.1. Efficiency
test on PV installations ……………………………………………………………………….100
12.SERVICE ………………………………………………………………………………………………………102 12.1. Warranty
conditions ………………………………………………………………………………………….102 12.2. Service
……………………………………………………………………………………………………………102
PRECAUTIONS AND SAFETY MEASURES
The instrument has been designed in compliance with directive IEC/EN61010-1
relevant to electronic measuring instruments. Before and while carrying out
measurements, observe the following indications and read all notes preceded by
the symbol with the utmost attention.
· Do not carry out any voltage or current measurement in humid environments ·
Do not carry out any measurements in case gas, explosive materials or
flammables are
present, or in dusty environments · Avoid contact with the circuit being
measured if no measurements are being carried out · Avoid contact with exposed
metal parts, with unused measuring probes, etc. · Do not carry out any
measurement in case you find anomalies in the instrument such
as deformation, breaks, absence of display on the screen, etc. · Pay special
attention when measuring voltages higher than 25V in special
environments and 50V in normal environments, since a risk of electrical shock
exists.
In this manual, and on the instrument, the following symbols are used:
CAUTION: observe the instructions given in this manual; an improper use could
damage the instrument or its components
High voltage danger: electrical shock hazard
EN – 4
Double insulation DC voltage or current
PVCHECKs
Connection to earth
1. PRELIMINARY INSTRUCTIONS · This instrument has been designed for use in
the environmental conditions specified in
§ 10.6. Do not use in different environmental conditions. · The instrument may
be used for measuring VOLTAGE and CURRENT in CAT III 300V
with a maximum voltage of 1000VDC between inputs. Do not use on systems
exceeding the limit values specified in § 10.1, § 10.2 and § 10.3 · We
recommend following the normal safety rules devised to protect the user
against dangerous currents and the instrument against incorrect use. · Only
the accessories provided together with the instrument will guarantee safety
standards. They must be in good conditions and replaced with identical models,
when necessary. · Check that the batteries are correctly inserted. · Before
connecting the measuring cables to the circuit being measured, check that the
desired function has been selected.
EN – 5
2. DURING USE Please carefully read the following recommendations and instructions:
PVCHECKs
CAUTION
· Failure to comply with the caution notes and/or instructions may damage
the instrument and/or its components or be a source of danger for the
operator.
· The symbol ”
” indicates a full charge level of the internal batteries.
When battery charge decreases to a minimum level, the symbol ”
”
appears on the display. In this case, stop testing and replace the
batteries according to the indications given in § 9.2 · The instrument is capable of keeping data stored even without
batteries.
3. AFTER USE When measurements are completed, turn off the instrument by
pressing and holding the ON/OFF key for some seconds. If the instrument is not
to be used for a long time, remove the batteries and follow the instructions
given in § 3.3
4. DEFINITION OF MEASUREMENT (OVERVOLTAGE) CATEGORY Standard “IEC/EN61010-1:
Safety requirements for electrical equipment for measurement, control and
laboratory use, Part 1: General requirements” defines what measurement
category, commonly called overvoltage category, is. § 6.7.4: Measured
circuits, reads:
Circuits are divided into the following measurement categories:
· Measurement category IV is for measurements performed at the source of the
lowvoltage installation Examples are electricity meters and measurements on
primary overcurrent protection devices and ripple control units.
· Measurement category III is for measurements performed on installations
inside buildings Examples are measurements on distribution boards, circuit
breakers, wiring, including cables, bus-bars, junction boxes, switches,
socket-outlets in the fixed installation, and equipment for industrial use and
some other equipment, for example, stationary motors with permanent connection
to fixed installation.
· Measurement category II is for measurements performed on circuits directly
connected to the low-voltage installation
Examples are measurements on household appliances, portable tools and similar
equipment. · Measurement category I is for measurements performed on circuits
not directly connected to MAINS
Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient
stresses are variable; for that reason, the standard requires that the
transient withstand capability of the equipment is made known to the user.
EN – 6
PVCHECKs
2. GENERAL DESCRIPTION
2.1. INTRODUCTION This instrument has been designed to carry quick checks
(IVCK) on photovoltaic modules/ strings in order to verify the parameters
declared by the manufacturer. In addition, this instrument measures
insulation/continuity on PV modules/strings/fields and evaluates the
efficiency of a PV field.
IVCK and insulation/continuity measurements may be performed in sequential
order IVCK Insulation Continuity, or be carried out separately in manual mode.
2.2. INSTRUMENT FUNCTIONS The instrument has the following features:
Continuity test of protective conductors (LOW) · Test with test current >
200mA in compliance with standard IEC/EN62446 · Manual calibration of
measuring cables.
Measurement of insulation resistance on PV modules/strings (M) · Test voltages
of 250V, 500V, 1000VDC in compliance with standard IEC/EN62446 · 3 measuring
modes: Field, Timer, String · Insulation test of unearthed metal masses
Efficiency evaluation of a PV installation in the short/medium term (EFF) ·
Measurement of DC voltage, DC current and DC power at PV modules/strings’
outputs · Measurement of irradiation [W/m2] by means of a reference cell
connected to the
optional remote unit SOLAR-02 · Measurement of module and environmental
temperature by means of probe connected
to the optional remote unit SOLAR-02 · Application of compensation
relationships to DC efficiency. · Immediate evaluation of DC efficiency
according to the limits set by the user · Parameter recording of a PV system
with 5s to 60min programmable IP
Quick checks (IVCK) in compliance with standard IEC/EN62446 · Measurement of
open circuit voltage Voc on PV modules/strings up to 1000VDC · Measurement of
short-circuit current Isc on PV modules/strings up to 15A · Measurement of
irradiation by means of an optional reference cell · Immediate evaluation
(OK/NO) of test results · Possible connection with the optional remote unit
SOLAR-02 · Internal customizable database for managing up to 30 PV modules ·
Display of results in OPC and STC conditions
The instrument is provided with backlit display, internal contrast adjustment
and a HELP key able to give a valid help to the operator while connecting the
instrument to the installation. An Auto Power OFF function, which can also be
deactivated, is available after approx. 5 minutes’ idleness.
EN – 7
PVCHECKs
3. PREPARATION FOR USE
3.1. INITIAL CHECKS Before shipping, the instrument has been checked from an
electric as well as mechanical point of view. All possible precautions have
been taken so that the instrument is delivered undamaged. However, we
recommend checking it to detect any damage possibly suffered during transport.
In case anomalies are found, immediately contact the Dealer.
We also recommend checking that the packaging contains all components indicated in § 10.7. In case of discrepancy, please contact the Dealer. In case the instrument should be returned, please follow the instructions given in § 12.
3.2. INSTRUMENT POWER SUPPLY The instrument is battery supplied. For battery type and life, see § 10.5.
The symbol ”
” indicates a full charge level of the internal batteries. When battery
charge decreases to a minimum level, the symbol ”
” appears on the display. In this
case, stop testing and replace the batteries according to the indications given in § 9.2.
The instrument is capable of keeping data stored even without batteries.
The instrument is provided with advanced algorithms to maximize the batteries’ life.
During instrument operation, a short pressing of the
key turns on the display’s
backlighting. In order to save battery efficiency, backlighting automatically turns off after
approx. 30 seconds.
A frequent use of backlighting reduces the batteries’ life.
3.3. STORAGE In order to guarantee precise measurement, after a long storage time under extreme environmental conditions, wait for the instrument to come back to normal condition (see § 10.6).
EN – 8
4. NOMENCLATURE
4.1. INSTRUMENT DESCRIPTION
PVCHECKs
CAPTION: 1. Inputs 2. Display 3. Connector for optical output/USB
port 4. Arrow keys/ENTER key 5. GO/STOP key 6. SAVE key 7. ON/OFF key 8. HELP
/ key 9. ESC/MENU key
Fig. 1: Description of the front part of the instrument
CAPTION: 1. Input for irradiation measuring probe 2. Input for auxiliary
temperature
measuring probe / DC current clamp (IVCK, EFF) 3. P, N inputs for DC voltage
(IVCK, EFF) / Insulation (M) measurement 4. E, C inputs for continuity tests
(LOW)
Fig. 2: Description of the upper part of the instrument CAPTION: 1. Connector
for opto-insulated/USB output cable connection
Fig. 3: Description of the instrument’s side
EN – 9
4.2. KEYBOARD DESCRIPTION The keyboard includes the following keys:
ON/OFF key to switch on/off the instrument
PVCHECKs
ESC/MENU key to exit the selected menu without confirming and to go back to
the main menu
keys to move the cursor inside the various screens in order to select
programming parameters ENTER key to confirm modifications, selected
programming parameters and to select the function to be accessed from the
menu.
GO/STOP key to start measurements
SAVE key to save the measured values
HELP key (long pressing) to access the on-line help, to view the possible
connections between the instrument and the installation.
key (simple pressing) to activate the display’s backlighting
4.3. DISPLAY DESCRIPTION
The display is a graphic module with a resolution of 128 x 15/05/12 15:34:26
128 dots. The display’s first line indicates the system’s date/
time and contains the battery charge indicator.
At the bottom of the display, the function of the ENTER key
and the active mode are displayed.
The symbol
indicates the presence of an active radio
connection with the remote unit SOLAR-02.
The flashing symbol
indicates that the instrument is
currently searching for a radio connection with the remote
unit SOLAR-02.
Selection
EN – 10
PVCHECKs
4.4. INITIAL SCREEN
When switching on the instrument, the initial screen appears for a few
seconds. It shows: · The instrument model (PVCHECKs) · The manufacturer’s name
· The presence of the enabled internal module for RF
connection · The serial number (SN:) of the instrument · The firmware version
(FW:) in the instrument’s memory · The date of the last instrument calibration
(Calibration
date:)
PVCHECKs
HT
RF SN: 15345678
FW: 1.14 Calibration date:
09/09/2015
After a few seconds, the instrument switches to the last function selected.
EN – 11
5. GENERAL MENU
PVCHECKs
EN – 12
PVCHECKs
Pressing the ESC/MENU key in any condition of the 15/05/12 15:34:26
instrument displays the general menu screen, in which the
instrument may be set, the saved measures can be displayed I V C K Te s t m o d . / s t r i n g s and the desired measuring function may be selected.
Use the cursor to select one of the options and confirm with L O W P E C o n t i n u i t y
ENTER to access the desired function.
test
M Insulation test
EFF Efficiency test
SET DB MEM PC
Settings Archive modules Recall Memory Trasf. data to PC
Press ENTER to sel.
MENU
5.1. SET INSTRUMENT SETTINGS
Position the cursor onto SET by using the arrow keys (,) 15/05/12 15:34:26
and confirm with ENTER. The display shows the screen
which lists the different settings of the instrument.
General
Measuring unit
Settings will instrument.
be
maintained
also
after
switching
off
the D a t e a n d T i m e
Remote Unit – Pyranometer
Irradiance
DC Clamp
Press ENTER to sel. SET
5.1.1. General
1. Position the cursor onto General by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER.
2. The display shows the screen which allows: Setting the instrument language Enabling/disabling the auto power off function
L a n g u a g e : E n g l i s h
Auto Power Off : NO
Contrast:
: 40
Adjusting display contrast
Key Beep
:NO
Enabling the acoustic signal when pressing a key.
3. Use the arrow keys (,) for option setting, and select
the desired option by means of the arrow keys ( , ). 4. Press the SAVE key to save the settings made; the
SAVE to store
message “Data saved” will be displayed for a few
SET
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
EN – 13
PVCHECKs
5.1.2. Measuring unit This section allows setting the measuring units of some parameters included in the database (DB) for the management of PV modules (see § 5.6) when measuring IVCK.
1. Position the cursor onto “Measuring Unit” by using the 15/05/12 15:34:26
arrow keys (,) and confirm with ENTER.
2. The display shows the screen which allows setting the A l p h a measuring units of the following parameters:
: m A / ° C
· Alpha possible selection: “%/°C” and “mA/°C” · Beta possible selection: “%/°C” and “mV/°C” · Gamma expressed in “%/°C” · Voc and Isc tolerance expressed in “%”
Beta Gamma To l e r a n c e
: %/°C : %/°C : %
3. Set the desired measuring units by means of the arrow
keys ( , ).
SAVE to store
4. Press the SAVE key to save the settings made; the
SET
message “Data saved” will be displayed for a few
seconds.
5. Press the ESC/MENU key to exit without saving and go
back to the previous screen.
5.1.3. Date and time
1. Position the cursor onto “Date and Time” by using the 15/05/12 15:34:26
arrow keys (,) and confirm with ENTER.
2. The display shows the screen which allows setting the Ye a r system’s date/time both in the European (EU) or in the
USA (US) format.
Month Day
3. Set the desired measuring units by means of the arrow H o u r s
keys ( , ).
Minutes Format
4. Press the SAVE key to save the settings made; the
: 2 0 1 2 : 05 : 15 : 09 : 53 : EU
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
SAVE to store
and go back to the previous screen.
SET
EN – 14
PVCHECKs
5.1.4. Remote unit/Pyranometer This section allows selecting the type of
remote unit to be used (if available) and setting the values of typical
parameters (Sensitivity and Alpha) of the reference irradiance cell supplied
with the instrument. The values of these parameters, which are printed on the
back label of the cell, depend on the type of PV module being tested.
1. Position the cursor onto Remote Unit-Pyranometer by 15/05/12 15:34:26
using the arrow keys (,) and confirm with ENTER.
2. The display shows the screen which allows selecting the R e m o t e U n i t use of the remote unit for EFF or IVCK tests. Possible E F F :
Y E S
options are: YES (use of SOLAR-02)
Remote Unit IVCK: NO
NO (no use of SOLAR-02) In case the remote unit S e n s . : 3 1 . 0 m V / k W / m 2 SOLAR-02 is not used for IVCK tests, it is necessary to A l p h a : 0 . 0 6 0 % / ° C set the values of Sensitivity (Sens.) and Alpha
parameter of the supplied irradiance cell on the
instrument.
3. Set the desired values by means of the keys ( , )
4. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few
SAVE to store
seconds. Press the ESC/MENU key to exit without saving
SET
and go back to the previous screen
EN – 15
PVCHECKs
5.1.5. Irradiance This section allows setting the minimum irradiance threshold both for IVCK measurement and for efficiency test on a PV installation.
1. Position the cursor onto “Irradiance” by using the arrow 15/05/12 15:34:26
keys (,) and confirm with ENTER.
2. The display shows the screen containing “Min Irr IVCK”, M i n I r r I V C K : 0 0 0 W / m 2 which allows setting the minimum irradiance threshold expressed in W/m2, used by the instrument as a reference M i n I r r E F F : 6 0 0 W / m 2
during IVCK measurement, and “Min Irr EFF”, which
allows setting the minimum irradiance threshold
expressed in W/m2, used by the instrument as a reference
while measuring the efficiency of a PV installation. Use the
arrow keys (,) to switch between the two items. 3. To set the minimum irradiance threshold, use the arrow
SAVE to store
keys ( , ). To obtain results with the accuracy
SET
guaranteed in this manual, we recommend following the
instructions given in § 10. The value can be set within the
following range: 0 ÷ 800 W/m2
4. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
CAUTION
The “0 W/m2” setting for parameter “Irr min IVCK” allows carrying out IVCK measurements without the following conditions being checked: Connection of reference cell to IRR input of the instrument Unsteady irradiation values Number of modules consistent with the measured open circuit voltage
5.1.6. DC Clamp This option allows setting a possible correction factor K for
the DC clamp, in order to improve current measurement. If present, the
correction factor is indicated on the rear label of the clamp itself,
indicated as:
K= X.xxx
In case no label is present, set k = 1.000
1. Position the cursor onto “DC Clamp” by using the arrow 15/05/12 15:34:26
keys (,) and confirm with ENTER.
2. The display shows the item “DC clamp K”, which allows D C c l a m p setting the corrective factor in the following range 0.950 to K
: 1 . 0 0 0
1.050 Set th d si d l b
f th
ke
EN – 16
to 1.050. Set the desired values by means of the arrow keys ( , ). 3. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
PVCHECKs
SAVE to store SET
EN – 17
PVCHECKs
5.2. EFF EFFICIENCY TEST SETTINGS FOR PV INSTALLATIONS The purpose of this measurement is to evaluate the DC efficiency of a photovoltaic installation, with the possibility of obtaining a positive or negative result of the check/ recording depending on a limit on the nDC parameter, freely set by the user. For this test, the use of the optional remote unit SOLAR-02 is necessary (see § 6.1)
5.2.1. Instrument settings
1. Position the cursor onto EFF by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
values of the output electrical parameters of the
Irr
photovoltaic generator.
Pnom
– – 3.500
Tc
– – –
Te
– – –
Pdc
0.0
Vdc
0.000
Idc
0.0
ndc
– – –
W/m2 kW °C °C kW V A
GO to Start
Selection
EFF
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: System parameters and Instrument
settings.
3. Use
the
arrow
keys
(,)
to
select
Irr
“Instrument
settings” and confirm with ENTER. The instrument P n o m
– – 3.500
shows the following screen:
Tc
– – –
Te
– – –
W/m2 kW °C °C
Pdc
0.0
kW
Vdc
0.000
V
Idc
0.0
A
ndc
– – –
System parameters
Instrument settings
Selection
EFF
4. By using the arrow keys ( , ) it is possible to set:
15/05/12 15:34:26
the integration period (IP) which can be used by the
instrument when testing the parameters of a PV I P
: 5 s
system. The following values can be selected: 5s,
10s, 30s, 60s, 120s, 300s, 600s, 900s, 1800s, F S D C c l a m p : 1 0 0 0 A
3600s.
EN – 18
The FS of the DC clamp used for measuring DC current with value selectable
between 1A ÷ 3000A
5. Press the SAVE key to save the settings made; the message “Data saved”
will be displayed for a few seconds. Press the ESC/MENU key to exit without
saving and go back to the previous screen.
PVCHECKs
SAVE to store EFF
5.2.2. System parameters
1. Position the cursor onto EFF by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
values of the output electrical parameters of the
Irr
photovoltaic generator.
Pnom
– – 3.500
Tc
– – –
Te
– – –
Pdc
0.0
Vdc
0.000
Idc
0.0
ndc
– – –
W/m2 kW °C °C kW V A
GO to Start
Selection
EFF
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: System parameters and Instrument
settings
3. Use
the
arrow
keys
(,)
to
select
Irr
“System
Parameters” and confirm with ENTER. The instrument P n o m
– – 3.500
shows the following screen:
Tc
– – –
Te
– – –
Pdc
0.0
Vdc
0.000
Idc
0.0
ndc
– – –
System parameters
EN – 19
W/m2 kW °C °C kW V A
PVCHECKs
Instrument settings
Selection
EFF
4. By using the arrow keys ( , ) it is possible to set:
15/05/12 15:34:26
Pmax maximum total rated power of PV installation
expressed in kW
Pmax:
Range coefficient of power variation with
temperature, characteristic parameter of PV modules G a m m a
(range: -1.00 ÷ -0.01%/C)
NOCT
NOCT rated operating temperature of the cell,
characteristic parameter of PV modules (typically in Te
: 3 . 5 0 0 k W
:
-0.45 %/ °C
: 45 °C
: 40 °C
the following range: 42 ÷ 48°C)
Tc
: 45 °C
Te, Tc setting of default values of environmental n D C L i m : : 0 . 8 5
and (Te) and PV module (Tc) temperatures. These C o r r. Ty p e . : T. E n v. values are taken into consideration by the instrument
only when no auxiliary probe is connected to the
remote unit SOLAR-02 (range: Te = 0°C ÷ 80°C; Tc =
0°C ÷ 100°C)
SAVE to Store
nDC Lim minimum limit of DC efficiency (default
EFF
value: 0.85 , range: 0.01 ÷ 1.15)
Corr. Type Setting of the compensation
relationship on the calculation of Pdc power and the
maximization of DC efficiency (see § 5.2.3).
EN – 20
PVCHECKs
5.2.3. Selection of the compensation relationship of temperature effects This
option allows selecting the relationship to be used to correct the
measurements performed according to the modules’ temperature as regards the
calculation of nDC efficiency. The following modes are available: – T.Mod.:
Correction factor Rfv2 related to PV module Temp. (Italian guideline
CEI-82-25) – T.Env: Correction factor Rfv2 related to environmental Temp.
(Italian guideline CEI-82-25) – nDC: nDC correction according to module
temperature
CAUTION
If it’s performed a test according to Italian guideline it is recommended to select “T.Env.” correction
Corr. Type
Temperature used (Tcel)
Calculation of nDC
R ef.
T.Mod.
Tcel = Tmoduli_Mis
T.Env.
Tcel
=
Tamb
(NOCT
–
20)×
Irr 800
1
(se Tcel 40°C)
Rfv2
=
1
–
(Tcel
–
×
100
(se Tcel > 40°C)
CE
hence
I
nDC =
Pdc
8225
Rfv2
×
Gp GSTC
×
Pn
nDC
Tcel = Tmoduli_Mis
( ) nDC
=
GSTC Gp
×
1 +
100
×
Tcel
–
25
×
Pdc Pn
– –
where:
Symb ol
Description
G p Irradiance measured on PV module surface
Measuri ng unit
[W/m2 ]
G STC
Standard Irradiance = 1000
[W/m2 ] [ ] P Rated Power = sum of all Pmax values of the section of the PV plant being kW n measured
Pdc DC power measured at the PV generator’s output
[kW ]
Rfv2 Correction factor depending on the Temperature of PV Cells (Tcel)
measured or
calculated according to the type of selected correction relationship
EN – 21
PVCHECKs
Absolute value of the thermal coefficient of Pmax of the PV modules included in the plant section being measured.
NOCT
(Normal Operating Cell Temperature) = Temperature the cells under reference conditions are brought to (800W/m2, 20°C, AM=1.5, air speed =1m/s).
[%/ °C] [%/°C]
For further details see § 11.1
EN – 22
PVCHECKs
5.3. LOW SETTINGS FOR CONTINUITY TEST WITH 200MA
The purpose of this measurement is to test the continuity of protective
conductors and equipotential ones (e.g. from rod to earth and connected
foreign earth) and earth rods of SPDs on PV installations. The test must be
carried out using a test current > 200mA according to the prescriptions of
IEC/EN62446 guideline
5.3.1. Instrument settings
1. Position the cursor onto LOW by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
following screen:
RPE max :
1
Rcal
: —
Rpe
= —
Itest
= – – – mA
Selection
LOW
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: Settings and Leads calibration
3. Use the arrow keys (,) to select “Settings” and
RPE max :
confirm with ENTER. The instrument shows the following
screen:
Rcal
:
1 —
Rpe
= —
Itest
= – – – mA
Leads calibration Settings Selection
LOW
4. Using the arrow keys ( , ) it is possible to set the 15/05/12 15:34:26
maximum limit value of Rpe resistance which the
instrument uses as a reference while measuring.
RPE max
: 0 1
5. Press the SAVE key to save the settings made; the message “Data saved” will be displayed for a few
EN – 23
seconds. Press the ESC/MENU key to exit without saving and go back to the previous screen.
PVCHECKs
SAVE to store
Selection
LOW
CAUTION
The settings saved for max RPE also affect the settings for the Continuity
test contained in the IVCK measurement (MENU IVCK)
EN – 24
PVCHECKs
5.4. M SETTINGS FOR INSULATION MEASUREMENT 5.4.1. Instrument settings
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
following screen:
I n s . Te s t : 1 0 0 0 V
Ri min
: 1.0
M
Mode
: Field
Vtest – – – V – – – V
Ri(+)
– – – M
Ri(-)
– – – M
Rp
– – – M
Selection
M
2. Press the ENTER key. The instrument shows the option 15/05/12 15:34:26
Settings:
3. Confirm with ENTER. The instrument shows the
following screen:
I n s . Te s t : 1 0 0 0 V
Ri min
: 1.0
M
Mode
: Field
Vtest
– – V — V
Ri(+)
– – –
M
Ri(-)
– – –
M
Rp Settings Selection
– – –
M
M
4. Use the arrow keys (,) for option setting, and select 15/05/12 15:34:26
the desired option by means of the arrow keys ( , ).
The following parameters can be set:
I n s . Te s t : 1 0 0 0 V
Ins. Test test voltage: 250, 500, 1000VDC
Mode operating modes: Field, Timer, String
Ri min
Ri min minimum limit value for insulation resistance
Test time maximum value of test time (only for M o d e
TIMER mode) with a maximum valEuNe o- f23500s in steps
: 1.00 M : TIMER
PVCHECKs
Test time maximum value of test time (only for
TIMER mode) with a maximum value of 300s in steps
of 1s.
Te s t t i m e : 1 0 s
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
SAVE to store
seconds. Press the ESC/MENU key to exit without
saving and go back to the previous screen.
M
CAUTION
The settings saved for test voltage also affect the settings for the
insulation measurement contained in the IVCK measurement (MENU IVCK)
EN – 26
PVCHECKs
5.5. IVCK SETTINGS FOR IVCK QUICK TEST The purpose of this measurement is to
verify the functionality of the connections and of the strings in a
photovoltaic field, according to the provisions of standard IEC/EN62446 by
measuring the open circuit voltage and the short-circuit current at operating
conditions and referred to STC (via the optional measurement of irradiation)
and providing immediate results regarding the measurement just carried out,
both in absolute terms and by comparison with the strings previously tested.
The test also allows carrying out (if selected) the continuity test and an
insulation measurement in a sequence.
5.5.1. Instrument settings
1. Position the cursor onto IVCK by using the arro 15/05/12 15:34:26 (,) and confirm with ENTER. The display shoM o d u l e : S U N P W R 3 1 8 following screen:
Vdc
: 0.0
V
Irr
:0
W/
m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V = – – – M
)
Rpe (Cal)
= —
Selection
IVCK
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: Settings, Reset Averages (see § M o d u l e
6.3.4) and Leads calibration (see § 6.5.2).
Vdc
3. Use the arrow keys (,) to select “Settings” and
confirm with ENTER. The instrument shows the following I r r
: SUNPWR318
: 0.0
V
:0
W/m2
screen:
Tc
: Auto
°C
Voc,Isc:
Ri(1000V) = – – – M
EN – 27
Rpe (Cal) = – – –
Leads calibration
Reset Averages
Settings
Selection
IVCK
PVCHECKs
4. Use the arrow keys (,) for option setting, and select 15/05/12 15:34:26
the desired option by means of the arrow keys ( , ). M o d u l e : S U N P W R 2 1 0
The following parameters can be set:
Mod.Nx Str : 01
Module type of module being tested
Te m p
Mod. N x Str. number of modules in the string.
Allowable values 1 ÷ 50
: Manual
25
°C
Temp temperature measuring method. Possible To l . V o c : 0 5 % ( + 4 % )
options:
To l . I s c : 0 5 % ( + 4 % )
“Auto” automatic measurement (recommended) I n s . Te s t : 1 0 0 0
V
carried out according to the measured value of the R i m i n : 1 . 0
M
modules’ Voc. “Manual” setting, made by the operator, of the known R P E Te s t : 2
temperature value of the module in the line below.
Aux temperature measurement with auxiliary probe PT1000. Tol. Voc (%) percentage value of the desired limit
SAVE to store IVCK
tolerance (set by the operator according to his/her
own requirements) for the measurement of Voc
carried out by the instrument. Allowable values: 0% ÷
25%. The value between brackets (4%) indicates the
reading error of the instrument in Voc measurement.
Tol. Isc (%) percentage value of the desired limit
tolerance (set by the operator according to his/her
own requirements) for the measurement of Isc carried
out by the instrument. Allowable values: 0% ÷ 25%.
The value between brackets (4%) indicates the
reading error of the instrument in Isc measurement.
Ins Test enabling/disabling of insulation
measurement and setting of test voltage. Possible
options: OFF, 250V, 500V, 1000V. With enabled
function, the line “Ri min” appears, which allows
setting the minimum limit threshold in the range 0.1 ÷
100M
RPE Test enabling/disabling of continuity test and
setting of the limit threshold value for measurement.
Possible options OFF, 1 ÷ 5 in steps of 1
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without
saving and go back to the previous screen.
CAUTION
The settings saved for test voltage for the insulation measurement carried out
within the IVCK function also affect the settings for the single measurement
(MENU M).
EN – 28
PVCHECKs EN – 29
PVCHECKs
5.6. DB DATABASE MODULE MANAGEMENT The meter allows managing up to a maximum
of 30 different types of PV modules, further to a DEFAULT module (not editable
and not erasable) which can be used as reference case when no piece of
information about the type of module being tested is available.
The parameters, referred to 1 module, which can be set are described below in
Table 1 together with their measuring range, resolution and validity
condition.
Symbol Nms
Description Number of modules per string
Range 1 ÷ 50
Resol. 1
Conditions
Pmax Maximum rated power of module
50 ÷ 4800W
1W
Pmax – Vmpp I mpp 0.01
Pmax
Voc
Open-circuit voltage
15.00 ÷ 99.99V 100.0 ÷ 320.0V
0.01V 0.1V
Vmpp
Voltage on maximum power point
15.00 ÷ 99.99V 100.0 ÷ 320.0V
0.01V 0.1V
Isc
Short-circuit current
0.5 ÷ 15.00A
0.01A
Impp Current on maximum power point
0.5 ÷ 15.00A
0.01A
Toll –
Negative tolerance for Pmax provided by the module manufacturer
0% ÷ 25.0% 0 ÷ 99W
0.1% 1
Toll +
Positive tolerance for Pmax provided by the module manufacturer
0 ÷ 25% 0 ÷ 99W
0.1% 1
Alpha
Isc temperature coefficient
-0.100÷0.100%/°C 0.001%/°C -15.00 ÷ 15.00mA/°C 0.01mA/°C
Beta
Voc temperature coefficient
-0.99 ÷ -0.01%/°C 0.01%/°C –0.999 ÷ -0.001V/°C 0.001V/°C
Gamma
Pmax temperature coefficient
-0.99 ÷ -0.01%/°C 0.01%/°C
NOCT Rated working temperature of cell
0 ÷ 100°C
1°C
Tech.
Effects due to module technology
STD (standard) CAP(capacitive
eff.)
Rs
Internal serial resistance
0.00 ÷ 10.00
0.01
Table 1: Typical parameters of PV modules
Voc Vmpp Voc Vmpp Isc Impp Isc Impp 100Tol-/Pnom< 25
100Tol+/Pnom< 25
100Alfa / Isc 0.1
100Beta/Voc 0.999
CAUTION
EN – 30
PVCHECKs The “Tech” item is referred to the choose of the technology of the
module
on test. Select the “STD” option for test on “STANDARD” PV modules, “CAP”
option for test on higher capacitive effects PV modules A wrong choice of the
type of technology can lead to a negative outcome of the final test.
EN – 31
5.6.1. How to define a new PV module
PVCHECKs
EN – 32
PVCHECKs
1. Position the cursor onto DB by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the M o d e l : D E F A U LT
screen which contains: · The type of PV module
· The values of the parameters associated to the module P m a x =
185
W
(see Table 1)
Voc
=
44.5
V
2. Use the arrow keys ( , ) to select the “DEFAULT” V m p p = 3 7 . 5 V
module type and confirm with ENTER.
Isc
=
5.40
A
Impp =
4.95
A
To l l – =
0
%
Selection
DB
3. Press the ENTER key, select the “New” command (which 15/05/12 15:34:26
allows defining a new module) and confirm again with M o d e l ENTER. Use the
arrow keys (,) to scroll the list of parameters.
Pmax
: D E F A U L T
=
185
W
Voc
=
44.5
V
Vmpp =
37.5
V
Isc
=
5.40
A
Impp =
4.95
A
To l l – =
0
%
New
Selection
DB
4. By using the internal virtual keyboard it is possible 15/05/12 15:34:26
define the name of the module (e.g. SUNPOWER 210) M o d e l : means of the arrow keys (,, , ). Press ENTER digit any character of the desired name. 5. Press the SAVE key to save the name of the new mod P m a x =
185
W
thus defined or the ESC/MENU key to exit without savi V o c =
44.5
V
KEYBOARD
SUNPOWER 210
ABCD E F G H I J K L M N OP
QRS T U VW XYZ – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE / ESC
EN – 33
PVCHECKs
6. Insert the value of each parameter (see Table 1) 15/05/12 15:34:26
according to the possible datasheet of the Manufacturer.). M o d e l Position the cursor onto the parameter to be defined by using the arrow keys (,) and set the value using the arrow keys ( , ). Keep the arrow keys ( , ) pressed P m a x
: SUNPWR 210 = 0 W
to carry out a quick setting of the values.
Voc =
0.0
V
7. Press the SAVE key to save the settings or the ESC/ V m p p =
0.0
V
MENU key to exit without saving.
Isc
=
0.00
A
Impp =
0.00
A
To l l – =
0
%
DB
CAUTION
· If the value of any parameter is unknown, press and hold the HELP key for a
few seconds to set the default values.
· Upon pressing the SAVE key, the instrument checks the conditions indicated
in Table 1
· and, if one or more of these conditions are not fulfilled, one of the error
messages listed in § 6.6 is shown on the display. The instrument does not save
the configuration set before any error is solved.
EN – 34
5.6.2. How to modify an existing PV module
PVCHECKs
EN – 35
PVCHECKs
1. Select the PV module to be modified from the internal 15/05/12 15:34:26
database by means of the arrow keys ( , ).
Model
2. Press the ENTER key and select the “Modify” command using the arrow key ().
3. Confirm selection with ENTER.
Pmax
S U N P W R 2 1 0
=
210
W
Voc
= 47.70
V
Vmpp = 40.00
V
Isc
=
5.75
A
New
Modify
Delete
Delete all
Selection
DB
4. By using the internal virtual keyboard it is possible 15/05/12 15:34:26
define again the name of the module or leave M o d e l : S U N P W R 2 1 0
unchanged by means of the arrow keys (,, ,
Press ENTER to digit any character of the desired name
5. Press the SAVE key to save the name of the new mod P m a x =
185
W
thus defined or to access the new parameter setting. V o c =
44.5
V
KEYBOARD
SUNPOWER 210
ABCD E F G H I J K L M N OP
QRS T U VW XYZ – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE / ESC
6. Change the value of the desired parameters by using the 15/05/12 15:34:26
arrow keys (,) and set the value using the arrow keys M o d e l : S U N P W R 2 1 0
( , ). Keep the arrow keys ( , ) pressed to carry out a quick setting of the values. If the value of any parameter is unknown, press and hold the HELP key for a few P m a x = 2 1 0 W
seconds to set the default values.
Voc
=
47.70
V
7. Press the SAVE key to save the settings made or the V m p p = 4 0 . 0 0 V
ESC/MENU key to exit without saving. In this case, the I s c
=
5.75
A
instrument shows the message “Data not saved”.
Impp =
5.25
A
To l l – =
5
%
DB
EN – 36
PVCHECKs
5.6.3. How to delete an existing PV module
1. Select the PV module from the internal database by 15/05/12 15:34:26
means of the arrow keys ( , ).
M o d e l : S U N P W R 2 1 0
2. Press the ENTER key and select the “Delete” command
by means of the arrow key () to delete the selected
module.
Pmax =
210
W
3. Press the ENTER key and select the “Delete all” V o c = 4 7 . 7 0 V
command by means of the arrow key () to delete all V m p p = 4 0 . 0 0 V
modules in the database.
Isc
=
5.75
A
4. Confirm the selection with ENTER or press ESC/MENU to
exit.
New
Modify
Delete
Delete all
Selection
DB
CAUTION
It is not possible to modify nor delete the “DEFAULT” PV module which is a
factory setting.
EN – 37
PVCHECKs
6. OPERATING INSTRUCTIONS
6.1. MEASURING EFFICIENCY PV PLANTS BY USING REMOTE UNIT SOLAR-02 For the sake
of simplicity, further in this section, the word “string” will be used,
although often the term “photovoltaic field” would be more correct. From the
point of view of the instrument, the management of a single string or of more
parallel strings (PV field) is identical. The instrument PVCHECKs (Master)
allows measuring the efficiency of PV installations in combination with the
optional remote unit SOLAR-02, which the irradiance and temperature probes are
connected to. This remote unit is capable of communicating with the Master
unit (to manage data synchronization and download) via a radiofrequency (RF)
connection, which is active up to a maximum distance of 1m between the units.
CAUTION
· The maximum voltage between inputs P and N is 1000VDC. Do not measure
voltages exceeding the limits given in this manual. Exceeding these limits
could result in electrical shocks to the user and damage to the instrument.
· In order to guarantee the operator’s safety, while making the connections,
disable the system being measured by means of the switches/breakers upstream
and downstream of the DC/AC converter (inverter).
Fig. 4: Connection of instrument for efficiency measurement on a PV
installation 1. Check and, if necessary, set the sensitivity of the reference
cell on SOLAR-02
consistently with the type of PV modules which will be measured (please refer
to the user manual of SOLAR-02). 2. We recommend making a preliminary
evaluation of the value of irradiation on the surface of the PV modules being
tested by means of unit SOLAR-02 (operating independently) and the reference
cell. 3. Switch on PVCHECKs, check and, if necessary, change the instrument
settings relevant to the type of remote unit, to the minimum irradiation
threshold, to the full scale
EN – 38
PVCHECKs
of the DC clamp, to the integration period and to the parameters of the system
being measured (see § 5.1.4, § 5.1.5, § 0, § 5.2.1 and § 5.2.2).
4. In order to guarantee the operator’s safety, disable the system being
measured by means of the switches/breakers upstream and downstream of the
DC/AC converter (inverter).
5. Bring PVCHECKs and SOLAR-02 nearer (maximum distance of 1m between them). All instruments must be switched on (see the User Manual of SOLAR-2 for further details).
6. On PVCHECKS, press the MENU key, select the function EFF and press ENTER; wait for the two units to start communicating with each other. This condition is highlighted by the contemporary presence of the following indicators:
Symbol Symbol
steady (not flashing) on the display of PVCHECKs steady (not flashing) on the display of SOLAR-02
7. Connect the inputs P and N respectively to the positive and negative output of the string, respecting the colours indicated in Fig. 4
8. Connect the output connector of the DC clamp to the IDC input.
CAUTION
BEFORE CONNECTING THE DC CLAMP TO THE CONDUCTORS Switch on the clamp, check
the LED indicating the status of the clamp’s internal batteries (if present),
select the correct range, press the ZERO key on the DC clamp and check on the
display of PVCHECKs the actual zeroing of the corresponding Idc value (values
up to 0.02A are acceptable).
9. Connect the DC current clamp to the positive output conductor of the
string, respecting the direction of the arrow found on the clamp itself as
indicated in Fig. 4
10. Position the clamp so that the jaw is not near the negative conductor.
11. The display shows the first screen which contains the 15/05/12 15:34:26
values of the output electrical parameters of the
module/string.
12. Before starting measuring, check for the presence of I r r
– – –
symbol ” ” steady on the display, which indicated a correct RF connection with the remote unit SOLAR-02. P n o m 3 . 5 0 0
Tc
45
Ta
30
Pdc
3.125
Vdc
389
Idc
8.01
ndc
– – –
EN – 39
W/m2 kW °C °C kW V A
PVCHECKs
Go to Start
Selection
EFF
13. Keeping the remote unit SOLAR-02 always close to th 15/05/12 15:34:26
main unit, press the GO/STOP key on PVCHECKS to
start testing. The message “Waiting start recording…
appears on the display of the main unit and “HOLD” I r r
– – –
shown on the display of SOLAR-02, together with a indication of the time remaining, expressed in second P n o m 3 . 5 0 0
until “00” is reached.
Tc
45
Ta
30
Pdc
3.125
Vdc
389
Idc
8.01
ndc
– – –
W/m2 kW °C °C kW V A
Waiting start recording…
Selection
EFF
14. Upon reaching “00” after pressing the GO/STOP key, th 15/05/12 15:35:00
test is started and the two units are synchronized wit
each other. In these conditions the message “Recording
running…” is displayed on the main unit an Irr
– – –
“Recording…” is displayed by SOLAR-02.
Pnom 3.500
Tc
45
Te
30
Pdc
3.125
Vdc
389
Idc
8.01
ndc
– – –
W/m2 kW °C °C kW V A
EN – 40
Recording running…
Selection
EFF
PVCHECKs
15. At any time it will be possible to analyze the curren1t5/05/12 15:35:00
recording status by pressing the MENU key. The followin
information will be shown: Starting date and time of recording The value set for the integration period (IP)
Start 15/05/12 15:30:00
Period:
5s
The number of periods elapsed from the beginning oIfP N u m b e r 6 1
the recording
Rec. Time
0d 1h
The remaining memory capacity for recording
Press the ESC key to exit the screen
Recording running…
16. Now it is possible to bring the unit SOLAR-02 near the PV strings to
measure irradiation and temperature by means of the relevant probes. When the
distance between the unit SOLAR-02 and PVCHECKs does not allow a RF
connection, on the display of SOLAR-02,
the symbol ” ” flashes for approx. 30s and then disappears, while PVCHECKs
continues searching for the connection for approx. 1 minute. 17. Position the
reference cell onto the surface of the PV modules. Please refer to the
relevant User Manual for a correct assembly. 18. Put the temperature sensor in
contact with the rear side of the module and fasten it with some tape; avoid
touching it with your fingers in order not to alter the measurement result.
19. Wait for a few seconds to allow the probes to reach a steady measure and
then connect the irradiation probe to input PYRA/CELL and the temperature
probe to input TEMP of unit SOLAR-02. 20. Wait for the message “READY” to
appear on the display of SOLAR-02 to indicate that the unit has detected the
data with solar irradiation > minimum threshold set (see § 5.1.5). 21. With
the message “READY” shown on the display, wait for approximately 1 minute in
order to take a certain number of samples. 22. Disconnect the irradiation and
temperature probes from unit SOLAR-02 and bring the unit near PVCHECKs (max
distance 1m).
23. The main unit PVCHECKs must be in EFF mode. If there is no flashing
symbol ” “, press key to activate the RF connection search again.
24. Press key on SOLAR-02 to activate the RF connection again. Consequently,
the main unit will show the message “active radio connection”.
25. To stop testing, press the GO/STOP key on the instrument and confirm with
ENTER that you want to stop recording.
26. The display of the unit SOLAR-02 will display the message “SEND” to
indicate that the data will be transferred to the main unit.
27. After the automatic data transfer phase, the instrument will 15/05/12 15:35:00
display:
Do not show any results if do not exist on the PV I r r
712
installation a “stable irradiance” condition more than
the minimum irradiance threshold
Pnom 3.500
Display the best performance values if during the Tc
45
recording, the Irradiance values reached the “stable” Te
30
diti
d it l
hi h th th mi i
EN – 41
W/m2 kW °C °C
PVCHECKs
condition and its values were higher than the minimum P d c 3 . 1 2 5 k W
irradiance threshold Unable to carry out the analysis if irradiance has never V d c
389
V
reached a steady value higher than the minimum I d c
8.01
A
threshold set, or if no value exists which is valid during the n d c
0.88
whole recording (nDC > 1.15).
Analys Results
Selection
EFF
Press SAVE to save the results obtained (see § 7.1) or ESC to exit the screen and go back to the initial screen.
6.2. MEASURING PARAMETERS ON PV PLANT WITHOUT USING SOLAR-02 The “PV
installation efficiency” test carried out without using the optional remote
unit SOLAR-02 only allows evaluating output electrical parameters of a string
or a photovoltaic field (quantities, Vdc, Idc and Pdc), which can be
periodically recorded with a
EN – 42
PVCHECKs programmable integration period (see § 5.2.1). In this mode, the
values of Irradiation, Te, Tc, the nDC efficiency value are not evaluated and
no result is provided by the instrument.
CAUTION
· The maximum voltage between inputs P and N is 1000VDC. Do not measure
voltages exceeding the limits given in this manual. Exceeding these limits
could result in electrical shocks to the user and damage to the instrument.
· In order to guarantee the operator’s safety, while making the connections,
disable the system being measured by means of the switches/breakers upstream
and downstream of the DC/AC converter (inverter).
Fig. 5: Instrument connection for measuring PV installation parameters without
SOLAR-02 1. Select the option “NO” as regards the type of remote unit to be
used during EFF test
(see § 5.1.4), set the full scale of the DC clamp (see § 5.2.1), a possible
correction factor of the DC clamp (see § 0), the integration period and the
rated power of the system (see § 5.2.1 and § 5.2.2). 2. In order to guarantee
the operator’s safety, disable the system being measured by means of the
switches/breakers upstream and downstream of the DC/AC converter (inverter) 3.
Connect the inputs P and N respectively to the positive and negative output of
the string, respecting the colours indicated in Fig. 5 4. Connect the output
connector of the DC clamp to the IDC input.
CAUTION
EN – 43
PVCHECKs
BEFORE CONNECTING THE DC CLAMP TO THE CONDUCTORS Switch on the clamp, check
the LED indicating the status of the clamp’s internal batteries (if present),
select the correct range, press the ZERO key on the DC clamp and check on the
display of PVCHECKs the actual zeroing of the corresponding Idc value (values
up to 0.02A are acceptable).
5. Connect the DC current clamp to the positive output conductor of the
string, respecting the direction of the arrow found on the clamp itself as
indicated in Fig. 5. Position the clamp so that the jaw is not near the
negative conductor.
6. The display shows the first screen which contains th 15/05/12 15:34:26
values of the output electrical parameters of the module
string.
Irr
– – –
Pnom 3.500
Tc
– – –
Te
– – –
Pdc
3.125
Vdc
389
Idc
8.01
ndc
– – –
W/m2 kW °C °C kW V A
Go to Start
Selection
EFF
7. Press GO/STOP key on PVCHECKs to start the test. Th 15/05/12 15:34:26
message “Waiting start recording…” appears on the
display of the instrument until “00” is reached.
Irr
– – –
Pnom 3.500
Tc
45
Te
30
Pdc
3.125
Vdc
389
Idc
8.01
ndc
– – –
W/m2 kW °C °C kW V A
Waiting start recording…
Selection
EFF
8. Upon reaching “00” after pressing the GO/STOP key, th 15/05/12 15:35:00
test is started. Unser these conditions, the messag
“Recording running…” appears on the instrument I r r
– – –
display
EN – 44
W/m2
instrument’s display.
PVCHECKs
Pnom 3.500
kW
Tc
45
°C
Te
30
°C
Pdc
3.125
kW
Vdc
389
V
Idc
8.01
A
ndc
– – –
Recording running…
Selection
EFF
9. At any time it will be possible to analyze the curren1t5/05/12 15:35:00
recording status by pressing the MENU key. The followin
information will be shown: Starting date and time of recording The value set for the integration period
Start : 14/02/00 17:18:00
Period:
5s
The number of periods elapsed from the beginning oIfP N u m b e r 6 1
the recording
Rec. Time 0d 1h
The remaining memory capacity for recording
Press the ESC key to exit the screen
Recording running…
10. To stop testing, press the GO/STOP key on the instrument and confirm with
ENTER that you want to stop recording.
11. Press SAVE to save the results obtained (see § 7.1) or ESC to exit the
screen and go back to the initial screen.
EN – 45
PVCHECKs
6.3. QUICK CHECK ON PV MODULES AND STRINGS (IVCK) 6.3.1. Foreword This
function carries out a series of quick tests on a PV module/string, measuring
in a sequence:
The Voc voltage and the short-circuit current Isc according to the
prescriptions of standard IEC/EN62446 with the possibility of measuring (by
using the relevant probes) also the irradiance and module temperature values.
Insulation resistance measurement (if enabled see § 5.5.1), only carried out
in STRING mode (see § 6.4.4) i.e. by automatically producing an internal short
circuit between the input terminals P and N and measuring between this short-
circuit point and the input terminal E.
Continuity test of protective conductors (if enabled see § 5.5.1) with 200mA
between the input terminals E and C of the instrument.
Irradiance measurement can be carried out in one of the following modes:
· Reference cell directly connected to PVCHECKs · Reference cell connected to
SOLAR-02 in RF connection with PVCHECKs
Irradiance measurements are always carried out in real time; it is therefore
impossible to start a “remote” recording of irradiance values through
SOLAR-02.
If the minimum irradiance threshold (see § 5.1.5) is: · = 0 the instrument
does not control the presence of the reference cell, irradiance
changes, number of modules and does not display error messages is case it is
impossible to calculate Voc and Isc values transposed under STC. This mode is
suitable to carry out a test session on a considerable number of strings as
quickly as possible. · > 0 (recommended >700W/m2) the instrument carries out
all tests provided for I-V test, manages all conditions and error messages for
I-V test (wrong num. of mod., temp. out of range, cell presence, min. irr.,
etc.) and calculated the values of Voc and Isc under STC. This mode is
recommended whenever accurate tests are to be carried out on modules/ strings
being examined.
In general, the result page will include:
· Description of the module used · Irradiance and temperature values (if
available) · Average values of Voc and Isc calculated as average of the
corresponding values under
OPC on the last 10 tests saved in the memory. If the number of tests is < 10,
the average is calculated on the number of available tests. The first test
will display dashes in the field “average values” as there are no previous
tests on which to calculate an average. · The values of Voc and Isc measured
under OPC as well as any partial result (only available if STC values are not
available) obtained by comparison with average values. · The values of Voc and
Isc calculated under STC (if available) and any partial result obtained by
comparing the values calculated under STC with the nominal ones (inserted in
DB modules). · The overall test outcome (OK(NO). The overall result will be
calculated based on the partial results obtained: Based on partial results
under STC (if these are available) Based on partial results under OPC (if STC
values are not available)
The instrument will not display any overall result if no partial result is
available.
EN – 46
6.3.2. Carrying out an IVCK quick test without measuring irradiance
PVCHECKs
EN – 47
PVCHECKs
CAUTION
· The maximum voltage between inputs P, N, E and C is 1000VDC. Do not measure
voltages exceeding the limits given in this manual
· Do not perform test on PV modules or strings connected to the DC/AC
converter
· The maximum current which can be measured by the instrument is 15A. Before
carrying out the measures of IVCK and Insulation in “STRING” mode, always
ensure that the instrument is connected to ONE STRING and not more strings in
parallel to avoid possible damage of it
1. Switch on the instrument by pressing the ON/OFF key. 2. Check that the
remote unit SOLAR-02 is not selected (see § 5.1.4 setting NO) 3. Check that
the value of minimum irradiance set in section “Irradiance” (see § 5.1.5) is
equal to 0.
2. Position the cursor onto IVCK by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the M o d u l e
screen to the side: The meaning of the parameters is the V d c following:
Module type of module being tested
Irr
: SUNPWR318 : 0.0 V : – – – W/m2
Vdc value of output voltage from module/string Tc
: —
measured in real time
Irr value of irradiation measured in real time
Voc,Isc:
Tc value of module temperature (see § 5.5.1)
Voc, Isc section with display of result OK/NO of R i ( 1 0 0 0 V ) Voc and Isc measurement
– – – M
Ri() the value in brackets may be NO/selected test voltage (see § 5.5.1). The value of Ri indicates the R p e ( C a l )
—
insulation resistance
Rpe() the value in brackets may be NO, Cal or
NoCal (see § 5.5.1). The value of Rpe indicates the
result of the continuity test.
Selection
IVCK
3. Press the ENTER key, select “Settings” and confirm again with ENTER. Set
the instrument as indicated in § 5.5.1
4. If necessary, press the ENTER key, select “Reset Averages” and confirm
again with ENTER. If necessary, carry out this operation as indicated in §
6.3.4
5. If necessary, press the ENTER key, select “Leads calibration” and confirm
again with ENTER. If necessary, carry out this operation as indicated in §
6.5.2
6. Connect the instrument to the module/string being tested and, if
necessary, to the main earth node of the system and to the earthed metal
masses as shown in Fig. 6. In particular, connect the negative output pole of
the module/string to terminal N and the positive output pole of the
module/string to terminal P.
EN – 48
CAPTION: E: Green cable
PVCHECKs
Fig. 6: Connection for IVCK test without irradiance measurement
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by
the instrument (see § 6.6) and, therefore, the test cannot be started. Check
and eliminate, if possible, the problem causing the error message before going
on with the test.
9. Press GO/STOP to start the test. In case no erro15/05/12 15:34:26
conditions occur, the instrument displays the messagM o d u l e
“Measuring…” and the measure of open-circuit voltag
Vdc
between the terminals P and N and of short-circuit curre
(for Isc values 15A).
Irr
: SUNPWR318
: 548.0 V
:0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Measuring…
Selection
IVCK
10.When Voc and Isc measurements are complete, th15/05/12 15:34:26
message “OK” is shown in case the result of the test M o d u l e positive
(measured values within the tolerances set o
Vdc
the instrument) 11. With insulation measurement selected, the instrume I r r
: SUNPWR318
: 548.0 V
:0
W/m2
goes on with the test keeping terminals P and N sho Tc
: Auto °C
circuited, and carrying out the test between this point an
terminal E for the time necessary to obtain a steadV o c , I s c :
OK
value.
12.The value of insulation resistance is shown in field “R and the message “OK” appears in case the result of thR i ( 1 0 0 0 V )
11 6
M OK
test is positive (measured value higher than the minimu
limit value set on the instrument).
Rpe (Cal)
—
EN – 49
Measuring…
Selection
IVCK
PVCHECKs
13.With continuity measurement selected, the instrume 15/05/12 15:34:26
goes on by opening the short-circuit and carrying out thM o d u l e test
between terminals E and C.
Vdc
14.The value of resistance in the continuity test is shown field “Rpe” and the
message “OK” appears in case thI r r
: SUNPWR318
: 548.0 V
:0
W/m2
result of the test is positive (measured value lower thaTc
: Auto °C
the maximum limit value set on the instrument).
15.The message “Outcome:OK” is finally shown by thV o c , I s c :
OK
instrument in case the result of all performed tests
positive.
Ri(1000V)
11 6
M OK
Rpe (Cal)
2.00 OK
Outcome: OK
IVCK
16.Press the arrow key to show the following page in15/05/12 15:34:26
which the values of parameters Voc and Isc areM o d u l e : shown. It shows:
SUNPWR210
The module being used
Irr
The average values of Voc and Isc under OPC
—
W/ m2
conditions
Tc ( A U TO )
— °C
The values of Voc and Isc measured under OPC andV o c A v g @ O P the relevant partial results obtained by comparison C
647 V
with the average values. In general:
IscAvg@OP C
5.43 A
Voc@OPC
647 V
O K
Isc@OPC
5.35 A
O K
Voc@STC
— V
Isc@STC
— A
Outcome: OK
IVCK
EN – 50
PVCHECKs
Esito Voc@OPC = OK
se
100 × VocAvg@OPC -Voc@OPC (Tol Voc + 4%)
VocAvg @ OPC
Esito Isc@OPC = OK
se
100 × IscAvg@OPC – Isc@OPC (Tol Isc + 4%)
IscAvg @ OPC
The overall value of the results:
o OK: if all results under OPC are OK, o NO if one of the results under OPC is
NO
17.Press the key to go back to the previous screen. 18.Press the SAVE key to
store the test result in the instrument’s memory (see § 7.2) or
the ESC/MENU key to exit the screen without saving and go back to the main
measuring screen.
CAUTION
The average values of Voc and Isc are displayed in the results’ page. These values include the average values of Voc and Isc under OPC conditions calculated as mean of the last 10 tests previously saved. If the user has carried out and saved a number of tests <10 or reset the average values (see § 6.3.4) the average value displayed during test N+1 will be calculated on the available N values.
EN – 51
6.3.3. Carrying out an IVCK quick test and measuring irradiance
PVCHECKs
EN – 52
PVCHECKs
CAUTION
· The maximum voltage between inputs P, N, E and C is 1000VDC. Do not measure
voltages exceeding the limits given in this manual.
· Do not perform test on PV modules or strings connected to the DC/AC
converter
· The maximum current which can be measured by the instrument is 15A. Before
carrying out the measures of IVCK and Insulation in “STRING” mode, always
ensure that the instrument is connected to ONE STRING and not more strings in
parallel to avoid possible damage of it
1. Switch on the instrument by pressing the ON/OFF key. 2. Irradiance
measurement is carried out in one of the following modes:
Measurement by reference cell directly connected to PVCHECKs Measurement by
reference cell connected to a connected SOLAR-02 unit RF with
PVCHECKs 3. Check that the settings of remote unit SOLAR-02 are consistent
with the type of
measurement to be carried out (see § 5.1.4). 4. Check the minimum irradiance
value set (see § 5.1.5).
5. Position the cursor onto IVCK by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the M o d u l e
screen to the side: The meaning of the parameters is the V d c following:
Module type of module being tested
Irr
: SUNPWR318
: 0.0 V
:0
W/m2
Vdc value of output voltage from module/string Tc
: Auto °C
measured in real time
Irr value of irradiation measured in real time
Voc,Isc:
Tc value of module temperature (see § 5.5.1)
Voc, Isc section with display of result OK/NO of R i ( 1 0 0 0 V ) Voc and Isc measurement
– – – M
Ri() the value in brackets may be NO/selected test voltage (see § 5.5.1). The value of Ri indicates the R p e ( C a l )
—
insulation resistance
Rpe() the value in brackets may be NO, Cal or
NoCal (see § 5.5.1). The value of Rpe indicates the
result of the continuity test.
Selection
IVCK
6. Press the ENTER key, select “Settings” and confirm again with ENTER. Set
the instrument as indicated in § 5.5.1
7. If necessary, press the ENTER key, select “Average Reset” and confirm
again with ENTER. Carry out this operation as indicated in § 6.3.4
8. If necessary, press the ENTER key, select “Cable calibration” and confirm
again with ENTER. Carry out this operation as indicated in § 6.5.2
9. Mount the rod onto the disc of the optional accessory M304 and keep it in
touch with the module’s surface. Make sure that the shadow cast by the rod
onto the disc falls within the “limit internal circle” on the disc itself (see
Fig. 7). If this is not the case, the incidence angle between the sun rays and
the module surface is too high
EN – 53
PVCHECKs and, therefore, the measurements carried out by the instrument CANNOT
be considered reliable. Repeat the same operations at another daytime. 10.
Fasten the bracket to the module by using the supplied set of screws and mount
the reference cell on it, if possible with output terminals facing downwards.
Rotate the cell until it rests on the tang found on the bracket so that the
cell is perfectly parallel to the surface of the module, then fix it with the
relevant screws provided.
Fig. 7: Positioning the inclinometer M304 11. Connect the output of the cell
corresponding to the type of module being tested to the
IRR. input of the instrument using the cable supplied with the cell, or to the
PYRA/ CELL input of the remote unit SOLAR-02, if used (see Fig. 8 and Fig. 9)
12. If used, connect the temperature sensor to the TEMP input of the
instrument and fasten it to the backside of the module under a cell using some
adhesive tape. As an alternative, connect the sensor to the TEMP input of the
remote unit SOLAR-02, if used (see Fig. 8 and Fig. 9) 13. Connect the
instrument to the module/string being tested and, if necessary, to the main
earth node of the system and to the earthed metal masses as shown in Fig. 8
and Fig. 9. In particular, connect the negative output pole of the
module/string to terminal N and the positive output pole of the module/string
to terminal P.
CAUTION
In case the remote unit SOLAR-02 is used for measuring irradiance, make sure
that the RF communication with the master unit PVCHECKs is always active
(symbol ” ” on and steady on the display).
EN – 54
PVCHECKs
CAPTION:
E: Green cable C: Blue cable P: Red cable N: Black cable
1. PV module/string 2. Main system earthing 3. Earthed metal structure in the
system 4. Reference cell for irradiance
measurement 5. Te m p e r a t u r e s e n s o r ( i f
required)
Fig. 8: Connection for IVCK test with direct irradiance measurement
CAPTION:
E: Green cable C: Blue cable P: Red cable N: Black cable
1. PV module/string 2. Main system earthing 3. Earthed metal structure in the
system 4. Reference cell for irradiance
measurement 5. Temperature sensor (if required) 6. Unità remota SOLAR-02Remote
unit SOLAR-02
Fig. 9: Connection for IVCK test with irradiance measurement through SOLAR-02
14.In the initial screen of IVCK mode, the following values 15/05/12 15:34:26
are displayed in real time: Module type of module being tested
Module
: SUNPWR318
Vdc value of output voltage from module/string V d c
: 548.0 V
Irr irradiance (from a direct measurement or
SOLAR-02 connected via RF)
Irr
: 856 W/m2
Tc modulo temperature (in MAN or AUX mode) Tc and the relevant measuring mode, or “- – -” in AUTO
: Auto °C
mode
In case, the symbol ” SOLAR-02
” of RF connection with V o c , I s c : EN – 55
SOLAR-02
PVCHECKs
Ri(1000V)
– – – M
Rpe (Cal)
—
Selezione
IVCK
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
15.Press GO/STOP to start the test. In case no error 15/05/12 conditions occur, the instrument displays the message 15:34:26
“Measuring…” and the measure of open-circuit voltage M o d u l e
between the terminals P and N and of short-circuit V d c
current (for Isc values 15A).
Irr
Tc
: SUNPWR318 : 548.0 V : 856 W/m2 : Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Measuring…
Selection
IVCK
16.When Voc and Isc measurements are complete, the 15/05/12 message “OK” is
shown in case the result of the test is 15:34:26 positive (measured values
within the tolerances set on M o d u l e
EN – 56
: SUNPWR318
PVCHECKs
the instrument)
Vdc
:
17.With insulation measurement selected, the instrument I r r
:
goes on with the test keeping terminals P and N short-
circuited, and carrying out the test between this point Tc
:
and terminal E for the time necessary to obtain a steady
value.
Voc,Isc:
18.The value of insulation resistance is shown in field “Ri”
and the message “OK” appears in case the result of the R i ( 1 0 0 0 V ) test is positive (measured value higher than the
minimum limit value set on the instrument).
Rpe (Cal)
548.0 V 856 W/m2 Auto °C
OK
116 M
OK
—
Measuring…
Selection
IVCK
19.With continuity measurement selected, the instrument 15/05/12 goes on by opening the short-circuit and carrying out the 15:34:26
test between terminals E and C.
Module
:
20.The value of resistance in the continuity test is shown in V d c
:
field “Rpe” and the message “OK” appears in case the I r r
:
result of the test is positive (measured value lower than
the maximum limit value set on the instrument).
Tc
:
21.The message “Outcome:OK” is finally shown by the
instrument in case the result of all performed tests is V o c , I s c :
positive.
Ri(1000V)
SUNPWR318 548.0 V 856 W/m2 Auto °C
OK
116 M
OK
Rpe (Cal)
2.00 OK
Outcome: OK
IVCK
22.Press the arrow key to show the following page in which the values of parameters Voc and Isc are shown. It shows: The module being used The value of irradiance The value of temperature of the module The average values of Voc and Isc under OPC The values of Voc and Isc measured under OPC The values of Voc and Isc calculated under STC and the relevant partial results obtained by comparison with the nominal values.
In general:
EN – 57
15/05/12 15:34:26
Module:
SUNPWR210
Irr
—
W/ m2
Tc ( A U TO )
57 °C
VocAvg@OP C
647 V
IscAvg@OP C
5.43 A
Voc@OPC
647 V
Isc@OPC
5.35 A
In general:
PVCHECKs
Voc@STC
787 V
O K
Isc@STC
5.72 A
O K
Outcome: OK
IVCK
( ) Esito Voc @ STC
= OK
se
100 ×
VocNom @ STC
–
Voc @
STC
Tol Voc + 4%
VocNom
@ STC
( ) Esito
Isc @ STC
=
OK
se
100 ×
IscNom @ STC
–
Isc @ STC
Tol Isc + 4%
IscNom
@ STC
The rated values of Voc and Isc are the values found in the instrument’s internal modules’ DB (see § 5.6).
The overall value of the results:
o OK: if all results under STC are OK, o NO if one of the results under STC is NO
23.Press the key to go back to the previous screen. 24.Press the SAVE key to
store the test result in the instrument’s memory (see § 7.2) or
the ESC/MENU key to exit the screen without saving and go back to the main
measuring screen.
CAUTION
The average values of Voc and Isc are displayed in the results’ page. These values include the average values of Voc and Isc under OPC conditions calculated as mean of the last 10 tests previously saved. If the user has carried out and saved a number of tests <10 or reset the average values (see § 6.3.4) the average value displayed during test N+1 will be calculated on the available N values.
6.3.4. Reset Averages If irradiance values are measured, the instrument provides a result by comparing the values measured with the average values calculated according to the previously saved measured. In this case, the average values calculated by the instrument are therefore particularly important. In case a new measurement campaign is started with significant variations in irradiance or temperature, it is advisable to set the average reference values to zero in order to make new calculations based on new measurements. To reset the average values, please keep to the following steps:
EN – 58
PVCHECKs
1. In IVCK mode, press the ENTER key, select “Reset 15/05/12 15:34:26
Averages” and confirm by pressing ENTER again to M o d u l e zero the average
values calculated until that moment.
Vdc
: SUNPWR318 : 0.0 V
Irr
:0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Leads calibration
Reset Averages
Settings
Selection
IVCK
The average values are automatically reset also modifying and then saving one
of the following parameters:
Type of PV module Number of modules per string
The average values are not reset if the operator changes operating mode and
then goes back to this mode.
EN – 59
6.3.4.1. Anomalous situations when performing IVCK tests
PVCHECKs
EN – 60
PVCHECKs
1. In case the instrument detects a voltage higher than 15/05/12 15:34:26
1000V at terminals P-N, P-E and N-E, it does not carry M o d u l e out the
test, gives out a long sound and displays the
Vdc
message “Vin > 1000”.
Irr
: SUNPWR318
: 0.0 V
:0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Vin > 1000
Selection
IVCK
2. In case the instrument detects a voltage lower than 15V 15/05/12 15:34:26
at terminals P and N, it does not carry out the test, gives M o d u l e out a
long sound and displays the message “Low
Vdc
voltage”.
Irr
: SUNPWR318
: 0.0 V
:0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Low voltage
Selection
IVCK
3. In case the instrument detects a voltage higher than > 5V 15/05/12 15:34:26
at terminals E and C, it does not carry out the test, gives M o d u l e out a
long sound and displays the message “Voltage >
Vdc
Lim”.
Irr
: SUNPWR318
: 0.0 V
:0
W/m2
Tc
: Auto °C
Voc,Isc:
EN – 61
Ri(1000V) Rpe (Cal)
PVCHECKs – – – M —
Voltage > Limit
Selection
IVCK
4. In case the instrument detects an Isc current higher than 15/05/12 15:34:26
15A, it does not carry out the test, gives out a long sound M o d u l e and
displays the message “Isc current too high”.
Vdc
: SUNPWR318 : 0.0 V
Irr
:0
W/m2
Tc
: Auto °C
Voc,Isc:
Ri(1000V)
– – – M
Rpe (Cal)
—
Isc current too high
Selection
IVCK
EN – 62
PVCHECKs 6.4. MEASUREMENT OF INSULATION ON PV MODULES/STRINGS/FIELDS (M)
6.4.1. Foreword The purpose of this function is to measure the insulation
resistance of the active conductors of a module, string, of a whole PV field
and of possible unearthed metal masses according to the prescriptions
IEC/EN62446 guideline. In general, the instrument measures insulation in the
following modes: FIELD mode used for measuring insulation resistance of a PV
field (photovoltaic
generator) consisting in one or more strings connected in parallel. The
instrument performs the measurement on the Positive and Negative poles of the
PV field. TIMER mode the instrument measures continuously (with a max duration
of 300s) on terminal “P” only, displaying the minimum resistance value
obtained at the end of the selected time. It may be used for measuring the
insulation resistance of more unearthed metal masses. STRING mode used for
measuring insulation only on single PV modules or strings, automatically
creating an internal short-circuit between the Positive and Negative poles,
with no need to use an external switch to short-circuit the positive and
negative terminals and measuring between this short-circuit point and the
installation earthing.
EN – 63
PVCHECKs
6.4.2. Measuring insulation FIELD mode
1. Position the cursor onto M by using the arrow keys15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
screen to the side:
I n s . Te s t : 1 0 0 0
V
2. Press the ENTER key, activate “Settings” and possibly change the desired parameters (see § 5.4). The followingR i m i n
: 1.0
M
parameters are shown on the display:
Mode
: Field
Ins. Test selected test voltage (250, 500 o
1000VDC)
Vtest – – – V – – –
Ri min minimum limit threshold for insulation
measurement
Ri(+)
– – –
Mode measuring mode: FIELD
Vtest real test voltages applied respectivelyR i ( – )
– – –
between the Positive pole and the Negative pole o
the field respect the earthing
Rp
– – –
Ri (+) insulation resistance measurement between
the Positive pole of the PV field and the earthing
Ri (-) insulation resistance measurement betweenSelection
M
the Negative pole of the PV field and the earthing
Rp final value of the measurement obtained by the
parallel of Ri(+) e Ri(-) values which is compared with15/05/12 15:34:26 the Ri min limit by the instrument
key access to the second page with the
measured values of VPN, VEP and VEN voltages V P N
– – –
V M M M
V
VEP
— V
VEN
— V
Selection
M
3. Connect the instrument to the PV field being tested and to the main earth
node of the system shown in Fig. 10. In particular, connect the negative
output pole of the PV field to terminal N and the positive output pole of the
PV field to terminal P.
CAPTION:
E: Green cable P: Red cable N: Black cable
1. Unearthed PV field 2. Main system earthing
EN – 64
PVCHECKs Fig. 10: Instrument connection for insulation measurement in FIELD
mode
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
4. Press GO/STOP to start the test. In case no erro 15/05/12 15:34:26
conditions occur, the instrument displays the message
“Measuring…” as shown in the screen to the side.
I n s . Te s t : 1 0 0 0
V
Ri min
: 1.0
M
Mode
: Field
Vtest 1043
Ri(+)
1057
V
– – –
M
Ri(-)
– – –
M
Rp
– – –
M
Measuring…
Selection
M
5. When measurement (with a fixed duration of 10s) is15/05/12 15:34:26
complete, the instrument provides the values Ri (+) and
Ri (-), respectively the insulation resistance of the Positive and Negative poles of the PV field being tested Te s t I s o : 1 0 0 0
If both results are higher than the minimum limit set, theR l i m
: 1.0
instrument displays the message “Outcome:OK”;M o d o
: Campo
otherwise, it displays the message “Outcome: NO” as
shown in the screen to the side.
Vtest 1043
6. Press the SAVE key to store the test result in the
1057
instrument’s memory (see § 7.2) or the ESC/MENU keyR i ( + ) to exit the screen without saving and go back to the main
100
measuring screen.
Ri (-)
100
V
M
V M M
Rp
69
M
Outcome: OK
M
EN – 65
PVCHECKs
6.4.3. Measuring insulation TIMER mode
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
screen to the side.
I n s . Te s t : 1 0 0 0 V
2. Press the ENTER key, activate “Settings” and possibly change the desired parameters (see § 5.4). The following R i m i n
: 1.0
M
parameters are shown on the display:
Mode
: Timer
Ins. Test selected test voltage (250, 500 or V t e s t
1000VDC)
Ri min minimum limit threshold for insulation R i ( + ) measurement
Mode measuring mode: TIMER
Vtest real applied test voltage
Ri (+) minimum insulation resistance between the Positive pole and the earthing detected during the Te s t t i m e :
whole duration of measurement.
Test time test duration range of 10 ÷ 300s
which
can
be
set
within
a
Selection
key access to the second page with the
measured values of VPN, VEP and VEN voltages
— V – – – M
10s M
3. Connect the instrument to the possible unearthed metal masses and to the main earth node of the system (see Fig. 11)
CAPTION:
E: Green cable P: Red cable N: Black cable
1. Unearthed PV module/string
2. M a i n s y s t e m earthing
3. Unearthed metal mass
Fig. 11: Instrument connection for insulation measurement in TIMER mode
CAUTION
EN – 66
PVCHECKs
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
4. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message “Measuring…” as shown in
the screen to the side.
I n s . Te s t : 1 0 0 0 V
Ri min
: 1.0
M
Mode
: Timer
Vtest
1020 V
Ri(+)min
– – – M
Te s t t i m e : 1 0 s
Measuring…
Selection
M
5. When measurement is complete, the instrument provides 15/05/12 15:34:26
the Ri(+)min value, i.e. the minimum value of insulation
resistance of the PV module/string (or of other metal
masses) being tested, continuously measured during the I n s . Te s t : 1 0 0 0 V
whole duration of measurement. If the result is higher
than the minimum limit set, the instrument displays the R i m i n
: 1.0
M
message “Outcome:OK”; otherwise, it displays the M o d e message “Outcome:NO” as shown in the screen to the
: Timer
side.
6. Press the SAVE key to store the test result in the instrument’s memory (see § 7.2) or the ESC/MENU key V t e s t
= 1020 V
to exit the screen without saving and go back to the main
measuring screen.
Ri(+)min = >20 M 0
EN – 67
PVCHECKs
Te s t t i m e : 2 0 0 s Outcome: OK M
6.4.4. Measuring insulation STRING mode
CAUTION
· The maximum current which can be measured by the instrument is
15A. Before carrying out the measures of Insulation in “STRING”
mode, always ensure that the instrument is connected to ONE
STRING and not more strings in parallel to avoid possible damage of
it
· ALWAYS disconnect the string on test from the inverter before
fo i th
t
1. Position the cursor onto M by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
screen to the side.
I n s . Te s t : 1 0 0 0 V
2. Press the ENTER key, activate “Settings” and possibly change the desired parameters (see § 5.4). The following R l m i n
: 1.0
M
parameters are shown on the display:
Mode
: String
Ins. Test selected test voltage (250, 500 or
1000VDC)
Vtest
Ri min minimum limit threshold for insulation
measurement
Rp
Mode measuring mode: STRING
Vtest real applied test voltage
Rp final value of the measurement obtained by the
parallel of Ri(+) e Ri(-) values which is compared with
the Ri min limit by the instrument
key access to the second page with the
— V – – – M
EN – 68
key access to the second page with the measured values of VPN, VEP and VEN voltages Selection
PVCHECKs
M
3. Connect the instrument to the PV module/string being tested and to the
main earth node of the system as shown in . In particular, connect the
negative output pole of the PV field to terminal N and the positive output
pole of the PV field to terminal P.
CAPTION:
E: Green cable P: Red cable N: Black cable
1. Unearthed PV module/string 2. Main system earthing
Fig. 12: Instrument connection for insulation measurement in STRING mode
CAUTION
Upon pressing the GO/STOP key, different error messages can be displayed by the instrument (see § 6.6) and, therefore, the test cannot be started. Check and eliminate, if possible, the problem causing the error message before going on with the test.
4. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message “Measuring…” as shown in
the screen to the side.
I n s . Te s t : 1 0 0 0 V
Ri min
: 1.0
M
Mode
: Stringa
Vtest
= 1020 V
EN – 69
Rp
= – – – M
PVCHECKs
Measuring…
Selection
M
5. When measurement is complete, the instrument provides 15/05/12 15:34:26
the Ri value, i.e. the minimum value of insulation
resistance of the of other objects being tested,
continuously measured during the whole duration of I n s . Te s t : 1 0 0 0 V
measurement. If the result is higher than the minimum
limit set, the instrument displays the message Ri min
: 1.0
M
“Outcome:OK”; otherwise, it displays the message M o d e “Outcome:NO” as shown in the screen to the side.
: Stringa
6. Press the SAVE key to store the test result in the
instrument’s memory (see § 7.2) or to exit the screen without saving and
the go
ESC/MENU key back to the main
V
t
e
s
t
= 1020 V
measuring screen.
Rp
= >10 M
0
Outcome: OK M
EN – 70
PVCHECKs
6.4.4.1. Anomalous situations
1. In any operating mode, in case the instrument detects a 15/05/12 15:34:26
voltage higher than 1000V at terminals P-N, P-E and N-
E, it does not carry out the test, gives out a long sound
I n s . Te s t : 1 0 0 0 V
and displays the message “Vin > 1000”.
Ri min
: 1.0
M
Mode
: String
Vtest
= — V
Rp
= – – – M
Vin > 1000
Selection
M
2. In STRING mode, in case the instrument detects an Isc 15/05/12 15:34:26
current higher than 15A, it does not carry out the test,
gives out a long sound and displays the message “Isc
I n s . Te s t : 1 0 0 0 V
current too high”.
Ri min
: 1.0
M
Mode
: Stringa
Vtest
= — V
Rp
= – – – M
Isc current too high
Selection
M
3. In STRING mode, in case the instrument detects a 15/05/12 15:34:26
current < 0.2A between terminals P and N, it does not
carry out the test, gives out a long sound and displays
I n s . Te s t : 1 0 0 0 V
the message “Current < Lim”.
Ri min
: 1.0
M
Mode
: String
Vtest
= — V
EN – 71
Rp
= – – – M
PVCHECKs
Current < Lim
Selection
M
4. In STRING mode, in case the instrument detects a 15/05/12 15:34:26
voltage < 15V between terminals P and N, it does not
carry out the test and displays the message “Low
voltage”.
I n s . Te s t : 1 0 0 0 V
Ri min
: 1.0
M
Mode
: String
Vtest
= — V
Rp
= – – – M
Low voltage
Selection
M
EN – 72
PVCHECKs
6.5. MEASUREMENT OF CONTINUITY ON PV MODULES/STRINGS/FIELDS (LOW)
6.5.1. Foreword The purpose of this measurement is to test the continuity of
protective conductors and equipotential ones (e.g. from rod to earth and
connected foreign earth) and earth rods of SPDs on PV installations. The test
must be carried out using a test current > 200mA according to the
prescriptions of IEC/EN62446 guideline
6.5.2. Calibration of measuring cables
1. Position the cursor onto LOW by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
screen to the side.
Rpe max :
Fig. 13: Compensation of the measuring cables’ resistance
Connect the measuring cables as shown in Fig. 13:
Rcal
:
1 —
Compensation of the measuring cables’ resistance
Rpe
= —
Itest
= – – – mA
Selection
LOW
Fig. 13: Compensation of the measuring cables’ resistance
2. Press the ENTER key. The instrument shows the 15/05/12 15:34:26
following options: Settings and Leads calibration
3. Use the arrow keys (,) to select “Leads calibration” R P E m a x : and confirm with ENTER.
Rcal
:
1 —
Rpe
= —
Itest
= – – – mA
EN – 73
PVCHECKs
Leads calibration Settings Selection
LOW
4. Press GO/STOP to start calibration. The message 15/05/12 15:34:26
“Measuring…” is shown on the display.
5. At the end of the compensation procedure, in case the measured value of resistance is lower than 5, the R P E m a x :
instrument gives out a double sound to signal the positive R c a l
:
1 —
result of the test. The following screen appears on the
display:
Rpe
= —
Itest
= – – – mA
Measuring…
Selection
LOW
6. The value of compensated resistance of the cables which 15/05/12 15:34:26
will be subtracted from all subsequent continuity
measures is shown next to “Rcal” and the message
“Calibration OK” appears on the display.
RPE max :
7. To delete the value of compensated resistance, carry out a R c a l
:
1.0 0.02
new compensation procedure with a resistance higher than
5 such as, for example, with open leads. The value in R p e Rcal is zeroed on the display.
= —
Itest
= – – – mA
Calibration OK
Selection
LOW
EN – 74
PVCHECKs
6.5.3. Measuring continuity
1. Position the cursor onto LOW by using the arrow keys 15/05/12 15:34:26
(,) and confirm with ENTER. The display shows the
screen to the side.
RPE max :
2. Press the ENTER key, activate “Settings” and possibly
change the desired parameters (see § 5.3.1). The R c a l
:
1.0 —
following parameters are shown on the display:
RPE max maximum threshold for continuity R p e
measurement, which can be selected in a range 1 ÷
5 in steps of 1
Itest
Rcal the value of resistance of the measuring
cables after calibrating them
Rpe the result of continuity measurement
Itest real test current
= — = – – – mA
Selection
LOW
3. Press the ENTER key, activate “Cable calibration” (see § 6.5.2) to carry
out the initial calibration of the measuring cables.
4. Connect the instrument to the PV module/string being tested and to the
main earth node of the system as shown in Fig. 14
CAPTION:
E: Green cable C: Blue cable
1. PV module/string 2. Main system earthing 3. Earthed metal structure in the
system
Fig. 14: Connection for continuity measurement on structures of the PV
installation
CAUTION
EN – 75
PVCHECKs
Upon pressing the GO/STOP key, different error messages can be displayed by
the instrument (see § 6.6) and, therefore, the test cannot be started. Check
and eliminate, if possible, the problem causing the error message before going
on with the test.
5. Press GO/STOP to start the test. In case no error 15/05/12 15:34:26
conditions occur, the instrument displays the message
“Measuring…” as shown in the screen to the side.
RPE max :
1.0
Rcal
: —
Rpe
= —
Itest
= – – – mA
Measuring…
Selection
LOW
6. At the end of measurement, the instrument provides the 15/05/12 15:34:26
value of resistance of the object being tested. If the result
is lower than the maximum limit set, the instrument
Rpe max :
displays the message “Outcome:OK”; otherwise, it
displays the message “Outcome:NO” as shown in the R c a l
:
1.0 —
screen to the side.
7. Press the SAVE key to store the test result in the R p e instrument’s memory (see § 7.2) or the ESC/MENU key
= 0.23
to exit the screen without saving and go back to the main
measuring screen.
Itest
= 210 mA
Outcome: OK
EN – 76
PVCHECKs
Selection
LOW
6.5.3.1. Anomalous situations
1. In case the instrument detects a voltage higher than 5V at 15/05/12 15:34:26
its terminals E and C, it does not carry out the test, gives
out a long sound and displays the message “Voltage >
Lim”.
RPE max :
Rcal
:
1 —
Rpe
= —
Itest
= – – – mA
Voltage > Lim
Selection
LOW
2. In case the instrument detects that the calibrated 15/05/12 15:34:26
resistance is higher that the measured resistance, the
instrument gives out a long sound and displays the
RPE max :
message: “Calibration not OK”
Rcal
:
1 —
Rpe
= —
Itest
= – – – mA
EN – 77
PVCHECKs
Calibration not OK
Selection
LOW
3. In case the instrument detects a resistance higher than 15/05/12 15:34:26
5 at its terminals, it gives out a long sound, zeroes the
compensated value and displays the message RPE max : “Calibration reset”.
Rcal
:
1 0.00
Rpe
= —
Itest
= – – – mA
Calibration reset
Selection
LOW
EN – 78
6.6. LIST OF DISPLAY MESSAGES
PVCHECKs
MESSAGE
DESCRIPTION
Function not available
The selected function/characteristic is not available
Data not saved
The instrument was not able to save the data
Wrong date
Set a correct system date
RADIO transmission error
The instrument does not communicate via RF with external units
SOLAR-02: Firmware incorrect
Inconsistent SOLAR-02 FW. Update firmware.
Firmware incorrect
Unsuitable instrument FW. Update firmware.
Error 4: contact service
Internal instrument error
Database full
The number of modules added to the internal DB is > 30
Module already added
The name of the module has already been added to the DB
Memory full
Instrument memory full upon pressing the GO key
Error: Vmpp >= Voc
Check module settings inside DB
Error: Impp >= Isc
Check module settings inside DB
Error: Vmpp * Impp >= Pmax
Check module settings inside DB
Error: Alpha too high
Check module settings inside DB
Error: Beta too high
Check module settings inside DB
Error: Gamma too high
Check module settings inside DB
Error: Toll too high
Check module settings inside DB
Wait data analysis
Download data from SOLAR-02 and wait for the result of PV efficiency test
Download error
Contact service department
Error while saving
Problems with the access to memory area
Remote unit undetected
The instrument does not detect any remote unit SOLAR-02
Unable to perform analysis Problems with the data downloaded from SOLAR-02. Check settings.
Data unavailable
Generic error. Repeat the test.
Negative voltage
Check the polarity of the instrument’s input terminals
Low voltage
Check the voltage between input terminals P and N
Vin > 1000
Voltage between input terminals > 1000V
Wrong no. of modules. Continue?
Set no. of modules inconsistent with measured Voc
Ref. cell temp. exceeding limits
Temperature measured by reference cell too high
Cell temp. undetected (ENTER/ ESC)
Measurement not carried out on module cell
Low battery
Low battery charge. Insert new batteries.
Please wait for cooling…
Instrument overheated. Wait before resuming the test.
Irradiance too low
Irradiance value lower than the minimum threshold set
NTC Error
Internal NTC efficiency jeopardized. Contact service department
EN – 79
PVCHECKs
Isc current too high
Measured Isc current > 15A
Current < Lim
Current between P and N lower than the minimum detectable value
EEPROM error: contact service
Internal instrument error
FRAM error: contact service
Internal instrument error
RTC Error: contact service
Internal instrument error
RADIO error: contact service
Internal instrument error
FLASH error: contact service
Internal instrument error
IO EXP error: contact service
Internal instrument error
Voltage > limit
Voltage between terminals E and C > 10V
Label already assigned
Change the numeric reference of the marker associated with the measure
Isc current < Lim
Isc current lower than the minimum detectable value. Contact service department
CAUTION: internal short-circuit
Contact service department
CAUTION: blown fuse
Contact service department
Reset calibration. Press ENTER.
Resistance value of input cables > 2
Calibration not OK
Calibrated resistance value > measured resistance
Error: Isc offset measurement
Internal instrument error
Rcal > measured R
Calibrated resistance value > measured resistance
CAUTION: AC voltage at P-N terminals
Presence of AC input voltage
Wait for condenser discharge
Wait for discharge of tested object after insulation
EN – 80
PVCHECKs
7. STORING DATA
The instrument allows saving max 999 measured values. The saved data can be
recalled at display and deleted at any moment, and can be associated to
reference numerical markers relevant to the installation name, the PV string
and the PV module (max 250 markers).
EN – 81
PVCHECKs
7.1. SAVING EFFICIENCY MEASURES
1. Press the SAVE key with a measured result shown on 15/05/12 15:34:26
display. The instrument shows the screen to the si
containing a virtual keyboard.
Use
the
arrow
keys
(,)
and
(
,
)
to
add
a
Irr
sh
description (max 13 chars) relevant to the test performeP n o m
712 3.500
3. Press the SAVE key again to confirm data saving or ESTCc
45
MENU to exit without saving.
Te
30
W/m2 kW °C °C
Pdc
3.125
kW
KEYBOARD
PV PLANT
A BCD E FGH I J K LMNOP
Q R S T U VW X Y Z – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE/ESC
7.2. SAVING IVCK, M AND LOW MEASURED RESULTS
EN – 82
PVCHECKs
1. Press the SAVE key with a measured result shown o 15/05/12 15:34:26
display. The instrument shows the screen to the M E A S U R
containing the following items:
E
:
The first memory location available (“MEASURE”) A r e a
:
The 1st marker level (ex: Area) The 2nd marker level (ex: Field)
Field
:
The 3rd marker level (ex: Inverter)
Inverter :
The field “Comment” in which the operator can incl a short description (max 13 char) for the system. C o m m e n t :
004 0 0 1
001 001
M o d i f i c a t i o n SAVE
2. Each marker level can be assigned to different labeInverter
labels are built-in and 5 labels can be customized). SeArea the desired marker level using the arrow keys ( ,
press the ENTER key to get available strings.
Building
Select one of the items available (,) and confirm with ENTER.
using
the
arrow
Field
Plant
Comment :
004 0 0 2
001 001
M o d i f i c a t i o n SAVE
4. Before performing a test, by mean Topview PC softwSolar Power
it’s possible to add custom strings into the list of maPVPlant level strings. In this case, the new values ma factory
selected as an alternative to the buil-in ones, as showH
T
the screen to the side.
instruments
HT ITALIA
EN – 83
004 0 0 2
001
PV installation Inverter Area Building Field Plant
PVCHECKs
001
M o d i f i c a t i o n SAVE
CAUTION
· Marker strings can be uploaded into the instrument by mean of the TopView
software through a PC connection (“Marker manager”).
· It is possible to add up to 5 custom strings for each marker, further to the
5 provided as default values.
· Deletion of custom strings can only occur through the TopView software. (The
buil-in strings cannot be eliminated.)
5. Use the arrow keys (,) and ( , ) with the virtu 15/05/12 15:34:26
keyboard to add a short description (max 13 chars) in thM E A S U R E :
“Comment” field. Press ENTER to digit any character H T I TA L I A :
the desired name.
StringT
:
004 0 0 2
001
Module1 :
001
Comment :
TA S T I E R A
T E S T I M P. H T
A B CD E FG H I J K LMNOP
Q R S T U VW X Y Z – + 0 1 2 3
4 5 6 7 8 9 SPACE
DEL
SAVE/ESC
6. Press the SAVE key again to confirm data saving or ESC/MENU to exit without saving.
EN – 84
7.3. OPERATIONS WITH RESULTS 7.3.1. Recalling on the display the PV efficiency results
PVCHECKs
EN – 85
PVCHECKs
1. Press the ESC/MENU key to go back to the main menu 15/05/12 15:34:26
select “MEM” and confirm with ENTER to access theMEM section where saved values are displayed. The screen to the side is shown by the instrument and contains the001
TYPE SMP 08/04/2012
saved tests.
002
REC 13/05/2012
2. Using the arrow keys (,) and the arrow key , select003 “Recall”, then “Efficiency” and confirm with ENTER to
*REC 14/05/2012
display the results of test only.
3. Using the arrow key it is possible to display the E f f i c i e n c y
following labels:
Re IVCK,M, LOW
TYPE indicates the type of data saved: “REC” for ac
check with a precise final result YES/NO, “*REC” whenO p e n
the instrument does not have temperature values recorded by
the irradiance SOLAR-02 and
“IaSnTdDe
l
e
t
e
for saving the instant conditions on the display.
Selection
MEM – EFF
DATE indicates the date and time in which the
datum has been saved in the instrument.
DESCRIPTION indicates the description provided by
the user when saving the datum.
4. Select the type of “SMP” datum, the item “Open” and
confirm with ENTER. The instrument shows the following
screen
5. Select the type of “REC” datum, the item “Open” and15/05/12 15:35:00
confirm with ENTER. The instrument shows the screen
containing the final results.
Irr
6. Selecting the “*REC” datum type, the item “View” and
confirming with ENTER, the message “Unable toP n o m
712 3.500
perform analysis” is displayed due to the irradiance Tc
45
temperature values recorded by SOLAR-02 missing Te
30
The partial values of this measurement are onlyP d c
visible after downloading the data onto the PC (see §
- using TopView.
Vdc
Idc
3.125 389 8.01
ndc
0.88
W/m2 kW °C °C kW V A
Analysis Results
Selection
EFF
7.3.2. Recalling on the display the results of IVCK, M and LOW measurements EN – 86
PVCHECKs
1. Press the ESC/MENU key to go back to the main menu 15/05/12 15:34:26
select “MEM” and confirm with ENTER to access theMEM DATE
TYPE
section where saved values are displayed. The screen to
the side is shown by the instrument and contains the001 08/04/2012 10:38 LOW
saved tests.
002 13/04/2012 12:15 M
2. Using the arrow keys (,) and the arrow key , select003 15/05/12 12:20 IVCK
“Recall”, then “IVCK” and confirm with ENTER to display
the results of the I-V curve only.
3. The field “DATE” indicates the date/time in which the
Efficiency
result of measurement has been saved, the field “TYPE”R e c I V C K , M , L O W
indicates the type of test carried out (LOW, M, IVCK).
Open
Delete
Selection
MEM- IVCK
4. Use the arrow key to go to the “Comments” label
15/06/12 15:34:26
5. The instrument will display the comment inserted by theMEM
operator during the saving procedure (see § 7.2) as
regards the system.
001
Comments PV PLANT 1
6. The symbol “” next to the measurement number indicates002
PV PLANT 2
that the instrument has performed the tests with recording
of irradiance and temperature values through the remote
unit, but these values have not been transferred or are no
available. For these measurements, the values
translated to STC will not be available. Press ESC/MENU to exit the screen and
go
back
to
theSelection
MEM- IVCK
main menu.
EN – 87
7.3.2.1. Accessing the data saved in the memory Numeric display
PVCHECKs
EN – 88
PVCHECKs
1. Select a line corresponding to a saved result and pre15/05/12 15:34:26
ENTER.
MEM
2. Select “Open” and press ENTER to open the sect
where measurement results are displayed as:
001
Comments PV PLANT 1
Numeric screens of measured parameters un 002*
PV PLANT 2
standard (STC) conditions and under opera
(OPC) conditions for IVCK tests
Numeric screens of measured parameters w measuring insulation (M) and continuity (LOW). O p e n
Re c a l l
Delete
Selection
MEM-IVCK
3. For IVCK test, there are the values of the followi 15/05/12 15:34:26
parameters: The module being used
Module:
SUNPWR210
The value of irradiance
Irr
The value of temperature of the module
903
W/ m2
The average values of Voc and Isc under OPCTc ( A U T O )
57 °C
conditions
VocAvg@OP
The values of Voc and Isc measured under OPCC
– – -V
conditions
IscAvg@OPC
The values of Voc and Isc calculated under STC and
the relevant partial results obtained by comparisoVno c @ O P C
– – -A 647V
with the nominal values.
Isc@OPC
5.35A
Voc@STC Isc@STC
787V
O K
5.72A
O K
Outcome: OK
Selection
IVCK
4. For M test in FIELD mode there are the values of 15/05/12 15:34:26
following parameters:
Set rated test voltage
Minimum threshold set on insulation measurementI n s . Te s t : 1 0 0 0
V
The type of mode·selected The real value of the applied test voltage
Ri min
: 1.0
M
The insulation value of positive pole Ri (+) The insulation value of positive pole Ri (-)
Mode
: Field
Vtest 1065
1064
V
EN – 89
Ri (+)
100
M
PVCHECKs
Ri (-)
100
M
Rp
72
M
Outcome: OK
Selection
M
5. For M test in TIMER mode there are the values15/05/12 15:34:26
the following parameters:
Set rated test voltage
Minimum threshold set on insulation measurementI n s . Te s t : 1 0 0 0
V
The type of selected mode The real value of the applied test voltage
Ri min
: 1.0
M
The Ri(+) minimum value of insulation resistancMeo d e of the PV module/string (or of other met
: Timer
masses) being tested, continuously measure
during the whole duration of measurement. The set measuring time
Vtest
1020 V
Ri(+)min
200 M
Test time:
10s
Oucome: OK
Selection
M
6. For M test in STRING mode there are the values of the 15/05/12 15:34:26
following parameters:
Set rated test voltage Minimum threshold set on insulation measurement The type of mode selected
I n s . Te s t Ri min
: 1000 : 1.0
V
M
The real value of the applied test voltage
Mode
: String
The Rp final value of the measurement obtained by
the parallel of Ri(+) e Ri(-) values which is V t e s t compared with the Ri min limit by the instrument
1020 V
EN – 90
PVCHECKs
Rp
200 M
Outcome: OK
Selection
M
7. For LOW test, there are the values of the following 15/05/12 15:34:26
parameters:
Limit threshold set for continuity measurement Value of calibration resistance of test cables The value of resistance of the object being tested
RPE max :
Rcal
:
1.0 —
The real value of the applied test current
Rpe
= 0.99
Itest
= 212. mA
Outcome: OK
Selection
LOW
7.3.3. Deleting data from the memory
1. Within the list of saved results, press ENTER to display 15/05/12 15:34:26
the submenus.
MEM
2. Select “Delete”, then press . The instrument allows
selecting the following items:
001
TYPE SMP 08/04/2012
Delete Last deletes the last saved test
002
REC 13/04/2012
Delete All deletes the whole memory content
3. Use the arrow keys (,) to select the desired option,
then press the ENTER key to confirm selection
4. Press ESC/MENU to exit the screen and go back to the R e c a l l main
menu.
Open Delete Last
Delete Delete All
Selection
MEM EFF
EN – 91
8. CONNECTING THE INSTRUMENT TO THE PC
PVCHECKs
EN – 92
PVCHECKs
CAUTION
· The connection between instrument and PC is realized by means of cable
C2006.
· In order to transfer the data onto a PC it is necessary to install
beforehand both the management software Topview and the drivers of cable C2006
on the PC itself.
· Before connecting, it is necessary to select the port to be used and the
correct baud rate (57600 bps) on the PC. To set these parameters, launch the
TopView software provided and refer to the programme’s on-line help.
· The selected port must not be engaged by other devices or applications, e.g.
a mouse, a modem, etc. Close any applications running using the Microsoft
Windows Task Manager function.
· The optical port emits invisible LED radiations. Do not directly observe
with optical instruments. Class 1M LED apparatus according to standard IEC/EN
60825-1.
To transfer data to the PC, follow this procedure:
1. Switch on the instrument by pressing the ON/OFF key. 2. Connect the
instrument to the PC via the provided optical/USB cable C2006. 3. Press the
ESC/MENU key to open the main menu. 4. Use the arrow keys (,) to select “PC”
to access data transfer mode and confirm
with ENTER.
15/05/12 15:34:26
IVCK Test mod./strings
LOW PE Continuity Te s t
M I n s u l a t i o n Te s t
E F F E ff i c i e n c y Te s t
SET DB MEM PC
Settings Archive modules Recall Memory Trasf. data to PC
Press Enter to sel.
MENU
5. The instrument shows the following screen:
15/05/12 15:34:26
EN – 93
PVCHECKs
Connecting to PC
MENU
6. Use TopView controls to activate data transfer (please refer to the on-
line help of the programme).
EN – 94
PVCHECKs
9. MAINTENANCE
9.1. GENERAL INFORMATION The instrument you purchased is a precision
instrument. While using and storing the instrument, carefully observe the
recommendations listed in this manual in order to prevent possible damage or
danger during use. Do not use the instrument in environments with high
humidity levels or high temperatures. Do not expose to direct sunlight. Always
switch off the instrument after use. In case the instrument is not to be used
for a long time, remove the batteries to avoid liquid leaks that could damage
the instrument’s internal circuits.
9.2. BATTERY REPLACEMENT When the low battery symbol ” ” appears on the LCD
display, or if during a test the instrument shows the message “low battery”,
it is necessary to replace the batteries.
CAUTION
Only expert and trained technicians should perform this operation. Before
carrying out this operation, make sure you have disconnected all cables from
the input terminals.
1. Switch off the instrument by pressing and holding the ON/OFF key. 2.
Remove the cables from the input leads. 3. Loosen the battery compartment
cover fastening screw and remove the cover. 4. Remove all the batteries from
the battery compartment and replace them only with new
batteries of the same type only (see § 10.5) making sure to respect the
indicated polarities. 5. Restore the battery compartment cover into place and
fasten it by means of the relevant screw. 6. Do not scatter old batteries into
the environment. Use the relevant containers for disposal.
9.3. CLEANING THE INSTRUMENT Use a soft and dry cloth to clean the instrument.
Never use wet cloths, solvents, water, etc.
9.4. END OF LIFE
CAUTION: this symbol indicates that the appliance, its accessories and the
internal batteries must be collected separately and correctly disposed of.
EN – 95
PVCHECKs
10.TECHNICAL SPECIFICATIONS
10.1. TECHNICAL SPECIFICATIONS FOR PV INSTALLATION EFFICIENCY Accuracy is
indicated as [%reading + (no. of digits) * resolution] at 23°C ± 5°C, <80%HR
DC voltage Range [V] 5.0 ÷ 199.9
200.0 ÷ 999.9
Resolution [V] 0.1 0.5
Accuracy ±(1.0%rdg + 2dgt)
DC current (by means of external clamp transducer)
Range [mV]
Resolution [mV]
Accuracy
-1100 ÷ -5 5 ÷ 1100
0.1
±(0.5%rdg + 0.6mV)
The value of current is ALWAYS displayed with a positive sign: the value of current transduced in a voltage lower than 5mV is zeroed.
FS DC clamp [A] 1< FS 10
10< FS 100 100< FS 1000
Resolution [A] 0.001 0.01 0.1
Minimum read value [A] 0.05 0.5 5
DC power (Vmis > 150V) Clamp FS [A]
Range [W]
Resolution [W]
1< FS 10
0.000k ÷ 9.999k
0.001k
10< FS 100
0.00k ÷ 99.99k
0.01k
100< FS 1000
0.0k ÷ 999.9k
Vmis = voltage at which power is measured; Imis = measured current
0.1k
Irradiance (with HT304N reference cell)
Range [mV]
Resolution [mV]
1 ÷ 40.0
0.02
Accuracy ±(1.5%rdg + 3dgt)
(Imis < 10%FS) ±(1.5%rdg)
(Imis 10%FS)
Accuracy ±(1.0%rdg + 0.1mV)
Temperature (with probe type PT300N)
Range [°C]
Resolution [°C]
-20.0 ÷ 100.0
0.1
Accuracy ±(1.0%rdg + 1°C)
EN – 96
PVCHECKs
10.2. TECHNICAL SPECIFICATIONS OF THE IVCK FUNCTION DC Voltage @ OPC
Range [V]
Resolution [V]
Accuracy
5.0 ÷ 199.9
200 ÷ 999
Minimum VPN voltage to start the test: 15V
DC Current @ OPC Range [A]
0.1 0.5
Resolution [A]
±(1.0%rdg + 2dgt) Accuracy
0.10 ÷ 15.00
0.01
±(1.0%rdg + 2dgt)
DC Voltage @ STC Range [V] 5.0 ÷ 199.9 200 ÷ 999
Resolution [V] 0.1 1
Accuracy ±(4.0%rdg + 2dgt)
DC Current @ STC Range [A] 0.10 ÷ 15.00
Resolution [A] 0.01
Accuracy ±(4.0%rdg + 2dgt)
Irradiance (with HT304N reference cell)
Range [mV]
Resolution [mV]
1 ÷ 40.0
0.02
Accuracy ±(1.0%rdg + 0.1mV)
Temperature (with probe type PT300N)
Range [°C]
Resolution [°C]
-20.0 ÷ 100.0
0.1
Accuracy ±(1.0%rdg + 1°C)
10.3. TECHNICAL SPECIFICATIONS FOR ELECTRICAL SAFETY Continuity of protective conductors (LOW)
Range []
Resolution []
Accuracy
0.00 ÷ 1.99
0.01
2.0 ÷ 19.9
0.1
±(2.0%rdg + 2dgt)
20 ÷ 199
1
Test current >200mA DC up to 5 (cables included), resolution 1mA, accuracy ±(5.0%reading + 5digits) Open-circuit voltage 4 < V0 < 10V
Insulation resistance (M) Mode TIMER
Test voltage [V]
Range [M] Resolution [M]
Accuracy
0.01 ÷ 1.99
0.01
250, 500, 1000
2.0 ÷ 19.9
0.1
±(5.0%rdg + 5dgt)
Open voltage Short circuit current Generated voltage
20 ÷ 199
1
<1.25 x nominal test voltage < 15mA (peak) for each test voltage resolution 1V, accuracy ±(5.0%reading + 5digits) @ Rmis> 0.5% FS
EN – 97
PVCHECKs
Rated current measured
1mA with 1k @ Vnom
Insulation resistance (M) Mode FIELD (*), STRING (**)
Test voltage [V]
Range [M] Resolution [M]
Accuracy (***)
0.1 ÷ 1.9
0.1
250, 500, 1000
2 ÷ 99
1
±(20.0%rdg + 5dgt)
(*) For FIELD mode
if VPN >1V the minimum voltage VEP and VEN for the calculation of Ri(+) and Ri(-) is 1V
(**) For STRING mode
minimum VPN voltage to start the test: 15V
Open voltage
<1.25 x nominal test voltage
Short circuit current
< 15mA (peak) for each test voltage
Generated voltage
resolution 1V, accuracy ±(5.0%reading + 5digits) @ Rmis> 0.5% FS
Rated current measured (***) For FIELD mode:
1mA with 1k @ Vnom
{ } max R+ , R- 100 { } min R+ , R-
add 5 dgts to the accuracy if
10.4. REFERENCE STANDARDS
10.4.1. General
Instrument safety:
IEC/EN61010-1
EMC:
IEC/EN61326-1
Safety of measuring accessories:
IEC/EN61010-031
Measurements:
IEC/EN62446 (IVCK, LOW, M)
Insulation:
double insulation
Pollution level:
2
Measurement category:
CAT III 300V to earth
Max 1000V between inputs P, N, E, C
10.5. GENERAL CHARACTERISTICS
Display and memory
Type of display:
LCD custom, 128×128 pxl, with backlighting
Saved data:
max 999
PC interface:
optical / USB
Characteristics of radio module
Frequency range: 2.4835GHz R&TTE category: Max transmission power: Max RF
connection distance: PV system efficiency Integration Period: Memory of
SOLAR-02:
Power supply Battery type: Low battery indication: Battery life: Automatic
power off:
Mechanical characteristics Dimensions (L x W x H): Weight (batteries
included): Mechanical protection:
2.400 ÷
Class 1 30µW 1m
5,10,30,60,120,300,600,900,1800,3600s approx. 1.5 hours (@ PI = 5s) approx. 8
days (@ PI = 600s)
6×1.5V alkaline type AA LR06 MN1500 the display shows symbol ” ” approx 120
hours (PV efficiency) after 5 minutes
235 x 165 x 75mm (9 x 6 x 3in) 1.2kg (2.5lv) IP40
EN – 98
PVCHECKs
10.6. ENVIRONMENTAL CONDITIONS FOR USE
Reference temperature:
23°C ± 5°C (73° ± 41°F)
Operating temperature:
0°C ÷ 40°C (32 ÷ 104°F)
Allowable relative humidity:
<80%RH
Storage temperature:
-10°C ÷ 60°C (14 ÷ 140°F)
Storage humidity:
<80%RH
Max operating altitude:
2000m
This instrument satisfies the requirements of Low Voltage Directive 2014/35/EU
(LVD) and of EMC Directive 2014/30/EU
This instrument satisfies the requirements of 2011/65/EU (RoHS) directive and
2012/19/EU (WEEE) directive
10.7. ACCESSORIES See the attached packing list
EN – 99
PVCHECKs
11.APPENDIX THEORETICAL OUTLINE
11.1. EFFICIENCY TEST ON PV INSTALLATIONS According to the requirements of the
laws in force, DC efficiency test on a PV installation depends on the type of
correction used to compensate the effects of the module’s temperature and on
the mathematical relationship used to calculate the nDC parameter (see §
5.2.3).
Cor r.
Tcel value
Tm Tcel = Measured module
od
temp. value
Tcel = Calculated module
Ten
temp. value
v Tcel = Tamb + (NOCT – 20)× Gp
800
Mathematical relationship for nDC calculation
Rfv2
=
11-
(Tcel
–
×
100
(se Tcel 40°C) (se Tcel > 40°C)
hence
nDC =
Pdc
Rfv2
×
Gp GSTC
×
Pn
Standard Result
Italian guideline
CEI 82-25
OK/NO
nD Tcel = Measured module temp.
C
value
( ) nDC
=
GSTC Gp
×
1 +
100
×
Tcel
–
25
×
Pdc Pn
– – –
– – –
where:
Symb ol
Description
G p Irradiance measured on PV module surface
Measuri ng unit
[W/m2 ]
G STC
Standard Irradiance = 1000
[W/m2 ] [ ] P Rated Power = sum of all Pmax values of the section of the PV plant being kW n measured
Pdc DC power measured at the PV generator’s output
[kW ]
Rfv2 Correction factor depending on the Temperature of PV Cells (Tcel)
measured or
calculated according to the type of selected correction relationship
[ ] Absolute value of the thermal coefficient of Pmax of the PV modules
included in the
plant section being measured.
%/°C
[ ] NOCT
(Normal Operating Cell Temperature) = Temperature the cells under reference conditions are brought to (800W/m2, 20°C, AM=1.5, air speed =1m/s).
%/°C
The previous relationships are valid if Irradiance > Min Irradiance (see the
User Manual of the MASTER instrument) and the irradiance values are “steady”,
i.e. for each sample taken, with IP 1min, the difference between the maximum
and minimum irradiance values measured must be < 20W/m2
The final OUTCOME can be:
EN – 100
PVCHECKs Non-displayable: if the obtained values are inconsistent (e.g. nDC
1.15) or if
irradiation has never reached a steady value > minimum threshold set. The maximum performance of the system
EN – 101
PVCHECKs
12.SERVICE
12.1. WARRANTY CONDITIONS This instrument is warranted against any material or
manufacturing defect, in compliance with the general sales conditions. During
the warranty period, defective parts may be replaced. However, the
manufacturer reserves the right to repair or replace the product. Should the
instrument be returned to the After-sales Service or to a Dealer, transport
will be at the Customer’s charge. However, shipment will be agreed in advance.
A report will always be enclosed to a shipment, stating the reasons for the
product’s return. Only use original packaging for shipment; any damage due to
the use of non-original packaging material will be charged to the Customer.
The manufacturer declines any responsibility for injury to people or damage to
property.
The warranty shall not apply in the following cases:
· Repair and/or replacement of accessories and battery (not covered by
warranty). · Repairs that may become necessary as a consequence of an
incorrect use of the
instrument or due to its use together with non-compatible appliances. ·
Repairs that may become necessary as a consequence of improper packaging. ·
Repairs which may become necessary as a consequence of interventions performed
by unauthorized personnel. · Modifications to the instrument performed without
the manufacturer’s explicit
authorization. · Use not provided for in the instrument’s specifications or in
the instruction manual.
The content of this manual cannot be reproduced in any form without the
manufacturer’s authorization.
Our products are patented and our trademarks are registered. The manufacturer
reserves the right to make changes in the specifications and prices if this is
due to improvements in technology
12.2. SERVICE If the instrument does not operate properly, before contacting
the After-sales Service, please check the conditions of battery and cables and
replace them, if necessary. Should the instrument still operate improperly,
check that the product is operated according to the instructions given in this
manual. Should the instrument be returned to the After-sales Service or to a
Dealer, transport will be at the Customer’s charge. However, shipment will be
agreed in advance. A report will always be enclosed to a shipment, stating the
reasons for the product’s return. Only use original packaging for shipment;
any damage due to the use of non-original packaging material will be charged
to the Customer.
EN – 102
PVCHECKs EN – 1
PVCHECKs EN – 2
Y A M U M 0 0 5 3 H
HT INSTRUMENTS SA
C/ Legalitat, 89 08024 Barcelona – ESP Tel.: +34 93 408 17 77, Fax: +34 93 408
36 30 eMail: info@htinstruments.com eMail: info@htinstruments.es
Web: www.htinstruments.es
SRL
Via della Boaria, 40 48018 Faenza (RA) – ITA
Tel: +39 0546 621002 Fax: +39 0546 621144
eMail: ht@htitalia.it
Web: www.ht-instruments.com HT INSTRUMENTS BRASIL
Rua Aguaçu, 171, bl. Ipê, sala 108 13098321 Campinas SP – BRA Tel: +55 19
3367.8775 Fax: +55 19 9979.11325
eMail: vendas@ht-instruments.com.br Web: www.ht-instruments.com.br HT
INSTRUMENTS USA LLC
HT ITALIA
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>