Laminar Flow Cabinet
User Manual
(Apply to Horizontal Flow Type BKCB-H1500/1800)
Preface
Thank you for using the Laminar Flow Cabinet produced by our company. In order
to use this product properly and to prevent damage to persons, property and
experimental products, please be sure to read and comply with the contents of
the manual carefully. If you do not use the device in the way specified by the
manufacturer and disassemble the it by yourself, the protection provided by
the cabinet lock may be damaged.
Only our after-sales service personnel and technicians authorized by us can
install and repair this equipment, otherwise it may cause electric shock or
fire.
Be sure to place the device firmly on a solid and horizontal surface. If the
ground is not firm or the device placed in an inappropriate place, it will
cause the device to fall over and cause injury to people.
Please use the special power supply indicated on the nameplate, otherwise it
may cause fire or electric shock. If the use of voltage does not meet the
requirements of the marked power supply, you need to add 3000W or more
automatic voltage regulator that suitable for motor load.
Please use a grounded power outlet to prevent electric shock. If the power
outlet is not grounded, be sure to have a professional technician install the
grounding wire. This device is for indoor use only, do not use it in the open
air. If Laminar Flow Cabinet is wet by rain, it may cause leakage or electric
shock.
Do not put the container with water on the Laminar Flow Cabinet. If the item
falls, it may cause injury to people. And the run-off water can cause device
leakage or electric shock.
Do not ground Laminar Flow Cabinet through gas pipes, power pipes, telephone
lines or lightning rods. Such grounding may cause electric shock or other
greater danger.
If the Laminar Flow Cabinet is not operating properly, stop using it and
unplug it from the power supply. Operating the equipment in an abnormal
condition may cause other hazards such as equipment leakage.
When the Laminar Flow Cabinet is not used for a long time, unplug the power
supply to prevent the power cord from causing electric shock, leakage and
other dangers due to aging.
Do not store acid, alkali and other corrosive, flammable, explosive or
volatile hazardous materials in the Laminar Flow Cabinet, or use combustible
spray near the Laminar Flow Cabinet, which may easily cause device damage as
well as casualties and property damage.
If you need to move the device after installation, you need to confirm with
the manufacturer.
Dispose the replaced old filter as biological waste.
The top air outlet must not be blocked.
Our company reserves the right to change the design and user manual of the
products, when changes, the user will not be notified.
Chapter 1 Product Introduction
1.1 Product Description
There are strict requirements for the cleanliness of the air in the work area
including laboratories, electronic technology, aerospace, precision equipments
and other fields at present. And clean air technology is also required in
biological fields such as medicine
and health, biopharmaceuticals, food, medical science experiments, sterile
microorganisms, etc. Laminar Flow Cabinet is an ideal local air purification
equipment to improve the clean working environment.
The cleanliness of the Laminar Flow Cabinet work area can reach ISO class 5,
which can effectively improve the process conditions, ensure the accuracy of
products and provide a sterile environment in the fields of scientific
experiments and biological experiments.
1.2 Product Features
1.2.1 Classification
Horizontal laminar flow, single-sided operation.
1.2.2 Structure
- The surface of the cabinet is electrostatic sprayed, and 1.2mm cold-rolled steel plate and welded structure are used to enhance the structural strength and stability of the cabinet.
- The work surface is made of stainless steel, which is beautiful and corrosion-resistant.
- The bracket is made of metal with electrostatic spraying on the surface.
- Control panel adopt touch switch to make the cabinet appearance beautiful and easy to operate.
1.2.3 UV sterilization
This Laminar Flow Cabinet using UV-C type 254nm wavelength UV sterilization,
which can not only kill the active cells of microorganisms, but also kill
strong heat-resistant budding spores and other mold spores, in addition that
phage and viruses can quickly break under UV rays.
1.3 Technical Parameters
Type | BKCB-H I 500 | BKCB-H1800 |
---|---|---|
Dimension (WDH) | 15008001730mm | 18008001930mm |
Work area size (WDH) | 1400500600mm | 17005001730mm |
Air flow type | Horizontal flow | |
Work surface height | 750mm | 750mm |
HEPA filter | HEPA without partitions, with 99.995% filtration efficiency for |
90.3pm particles.
HEPA specification| 142063069mm| I 1620 66569mm
Noise| 565d6(A)
Illumination| a.300 lx
Vibration| 55pm (rms)
Temperature rise| 58°C
Display| LCD (Liquid Crystal Display)
Splash-proof socket| The power of the equipment used does not exceed 500W.
Grounding resistance| <0.1S2
Power supply| AC220V±10%,50 Hz(standard) ACI 10V±10%, 60Hz (Optional)
Work area| 304 brushed stainless steel
Cabinet| The cabinet is made of I.2mm (I mm) thick cold-rolled steel plate and
the surface is electrostatically sprayed.
UV Lamp appointed timing| UV lamp delayed lOs open to protect the safety of
the operator.
UV Lamp| 30W| 40W
LED lamp| 16W| 16W
Airflow velocity| 0.2-0A5m/s| 0.2-0.45m/s
Safety| Colony forming unit 50.5CFU/30min
1.4 Product Structure
Figure 1
- Footmaster caster
- Base stand
- Side window
- Upper cabinet
- Button switch
- Power socket
- Control panel
- LED screen
- Waterproof socket
- LED lamp
- UV lamp
- Hook hanger
Caution
- The power of the cabinet used on the receptacle should not exceed 500W (rated voltage rated current 220V 2.3A).
- The waterproof receptacle as waterproof only when its front cover is lay down; the receptacle cannot be considered as waterproof when the front cover is opened.
Note: IP44 is the socket degree of protection. According to IEC 60529: Protected the inside of the socket against foreign solid objects of 1.0mmΦ and greater from contacting the internal parts; protected the socket against splashing water from all directions from causing damage.
Chapter 2 Installation
Caution
This Laminar Flow Cabinet needs to be installed by our trained and qualified
engineers.
- Remove all packaging components.
- Inspect the external surface of the host for scratches, deformations or foreign objects.
- Carefully inventory accessories and information according to the packing list in the manual.
- Move the entire cabinet to a position as close as possible to the final location and easy to install.
Caution
When handling the Laminar Flow Cabinet, it is strictly forbidden to place or
disassemble it upside down.
2.1 Installation Environment
a. Indoor use.
b. Humidity: ≤80%.
c. Power supply voltage: AC 220V±10%(standard) AC 110V±10%(Optional).
d. Transient overvoltage: Facility category (overvoltage category) Ⅱ.
e. Rated pollution level: Class 2.
2.2 Location Selection
Laminar Flow Cabinet should be installed in the clean laboratory or indoor
environment away from the dust source and vibration source, not at the
channel. If the laboratory has windows, it should always be in the closed
state. Laminar Flow Cabinet should not be placed in the circulation air
entrance, so that the air can blow over the former operation area or blow to
the HEPA.
If space allows, there should be 30cm space behind and around the Laminar Flow
Cabinet for cleaning, if not, there should be minimum 8cm per side and 3.8cm
space at the back.
The Laminar Flow Cabinet power outlet can be closed to facilitate the
maintenance, and there is no need to move the device to ensure electrical
safety.
The power outlet can be closed to facilitate the maintenance, and the safety
of electrical appliances can be ensured without moving. Do not place the
device in a position where it is difficult to disconnect the power supply.
2.3 Installation Steps
a. After moving to the designated indoor location, remove the packing
material, check the device and accessories according to the packing list to
make sure there is no damage during transportation.
b. Assemble and place the base stand
Screw out the M10×55 inner HXBTs in the cross brace and T brace, assemble the
base tightly and firmly refer to the above figure.
c. Connect the upper cabinet and stand base
According to the above diagram, use M10×55 inner HXBTs, flat washers Φ10,
spring washers Φ10 to fasten firmly through the base and side plate from the
bottom up.
d. Place the whole device
The Laminar Flow Cabinet should be placed in a protected area of airflow to
prevent the airflow from ventilation system, air conditioner, door, window and
personnel movement it.
Caution:
- The cabinet should be installed in a wide and more spacious space .
- When moving the cabinet, you need to cut off the power and adjust the casters first and then move slowly.
e. Caster adjustment
Rotate the red part of the caster clockwise, lower the corresponding base
legs, lower the height of the cabinet, then lower four casters at the same
time to move the cabinet position; counterclockwise rotate the red part of the
caster, raise the corresponding base
legs, raise the height of the cabinet, raise four casters at the same time to
fix the cabinet; four casters can be adjusted at the same time to make the
cabinet in the horizontal stable state.
2.4 Check after Installation
After powering up, check the following items according to the normal use
process.
Test item | Normal condition |
---|---|
Normal power on-off | Connect the power supply and power on. |
Fan operation | Click the “Fan” button,fan works normally. |
LED lamp | Click “LED lamp” button, the LED lamp is on. |
Waterproof socket | Click “Waterproof socket”,output power is AC |
220V±10%(standard) AC 110V±10%(Optional ) tested by multimeter.
➢ This product is placed in laboratory sections or teaching laboratories.
➢ The installation location should be far away from dust and vibration
sources.
➢ Be sure to plug the power supply into an outlet with a ground wire and
ensure that the ground terminal is reliably grounded, and use a power supply
with a leakage protection device.
➢ The interior and surroundings of the cabinet must be carefully cleaned after
installation using a vacuum cleaner or a tool that does not produce fibers.
➢ Test the average velocity of working area with an anemometer after
installation and cleaning. When the average airflow velocity is less than
0.2m/s, the airflow velocity should be increased. On the contrary, when the
average airflow velocity is greater than 0.5m/s, it should be reduced.
Caution:
If there is any problem please contact your local dealer for installation
adjustment.
Chapter 3 Operating Instruction
3.1 Main Screen Introduction
Temperature/Humidity: Real-time measurement and display of temperature
and humidity in the operation area.
Air Flow Animation : When the device is running, display the air flow
direction, when close the fan, the animation stops.
UV Lamp : Control the switch of UV lamp.
Note: UV lamp and LED lamp have a chain mechanism, only when the fan and
LED lamp are closed can UV lamp be opened.
Fan: Control the fan switch.
Power socket: Control the switch of waterproof socket in the operation
area;
Silence: Switch sound on and off. “×” is in front of the horn when the
sound is off.
Settings: After clicking “Settings”, it enters the settings interface,
which allows you to set the following contents: “Timing Switch”, “Fan
Settings”, “Equipment Life”, “Standby Settings”.
Pressure: Real-time display of the pressure difference value before and
after the filter.
Velocity: Real-time display of horizontal airflow velocity magnitude.
Time: Real-time display of the current date and time.
LED Lamp : Control the open and close of LED lamp.
Waterproof socket : Control the power on and off of waterproof socket.
3.2 Settings Interface
Click “Settings” to enter the settings interface, in which you can set
contents as shown in the figure.
3.2.1 Timing switch
Click “Timing Switch” in the above figure to enter the timing switch
interface, click “Time Input/Display Area” to input the value to be timed, and
then you can switch the timing for the corresponding time.
Caution:
In order to protect users, the UV lamp is turned on with a delay of 10 seconds
by default. If users have other delay needs, you can set the UV lamp to turn
on at a time on this interface. Before turning on the UV lamp, the user should
set the timer off on this interface. In order to achieve a good effect of air
and surface disinfection, the irradiation time of the UV lamp should be
≥30min.
3.2.2 Equipment life
Click “Equipment Life” to enter the equipment life display interface, which
can display the UV lamp uesed time, the equipment uesed time and filter uesed
time. If the UV lamp or filter is replaced, it is needed to reset the time.
Click “Time Display Area”, enter the password that which can be obtained from
the manufacturer.
Caution:
If the filter needs to be replaced, please contact our customer service center
or local agent.
3.2.3 Standby settings
The standby time is the time when the screen is darkened and enters the screen
protection state.
Click “Standby Settings” to enter the standby settings interface, click the
number in the “Standby Time Display Area” to modify the standby time.
3.3 Introduction to the Usage Process
- Before the experiment:Click the UV lamp, turn on the UV lamp, disinfect for more than half an hour.
- Open the fan, after running for half an hour can the experiment be started.
- After the experiment, click the UV lamp, turn on the UV lamp and disinfect for more than half an hour.
- After use, unplug the power supply and place it in suspension to avoid power contact with moisture and water to avoid leakage.
Caution:
- When using UV lamp for disinfection, people should leave the room to protect their eyes and skin from inadvertent exposure.
- The intensity of the UV lamp should be tested regularly according to the manufacturer’s specifications, and it is recommended that it be replaced once a quarter if it fails to pass.
Chapter 4 Maintenance
The operability and safety of this device can only be guaranteed if there are
enough competent personnel and units for inspection, maintenance and repair.
4.1 Comprehensive Maintenance Cycle
Maintenance should be performed weekly, monthly, annual, or every 1000 work
hours, as well as every restart.
4.2 Maintenance and Repair Methods
4.2.1 Cleaning
Under normal circumstances, cleaning only requires a small amount of household
or commercial dishwashing detergent, which can be dissolved in water and wiped
the dirt on the surface directly.
-
When used for the first time or use again after long time out of use
You should first use a dry cleaning towel to remove the surface of the floating dust, after wiping with a clean towel several times, after making sure that there is no dust, spray medical alcohol on the medical gauze to disinfect the work area, laminar flow plate and work surface, comprehensively. Before use, turn on the UV sterilization lamp to disinfect, 30 minutes later turn off the UV sterilization lamp and press the fan key, the fan will run. -
When used continuously
After using the device every day, first clear all the debris on the work surface, wipe it again with a dry and wet cleaning towel, and finally disinfect the work area, laminar flow plate and work surface with medical alcohol spray on medical gauze comprehensively and prepare for use on the next day.
Caution:
You can only spray alcohol on the medical gauze with alcohol spray, do not
spray alcohol on the laminar flow plate, because the laminar flow plate has a
HEPA, which can not be subject to moisture.
4.2.2 Regular maintenance
a. Daily or weekly maintenance
- Use medical alcohol to disinfect and clean the device.
- Disinfect and clean the control panel with medical alcohol.
- Use flexible cleaning agent or glass-specific cleaning agent to clean the outer surface of the device and glass.
- Check each function of the device according to the user manual.
b. Monthly maintenance
- Surface cleaning.
- Check functions of the device, disinfection of the interior of it.
- Record this maintenance.
c. Annual maintenance
- Check UV lamp and LED lamp.
- Conduct a thorough inspection of device’s performance for the safety of it. The maintenance fee is paid by the user.
- Record this maintenance.
Caution:
a. Before daily maintenance, cut off the power supply.
b. Since the statistics of operation time will directly affect the judgment of
maintenance needs, we suggest that a detailed record of operation time should
be prepared when using the device for reference and inquiry.
c. The fan must be inspected and maintained regularly.
d. Surface cleaning: To keep the cabinet clean, please clean it regularly (at
least once a week is recommended). When wiping, dip a soft cloth in soapy
water and then wring it out. Do not spray any chemical reagent on the
operation panel or other labels to prevent
discoloration or illegible label film. Clean the outer surface of the cabinet
and glass with a flexible cleaning agent or glass-specific cleaning agent for
glass.
Chapter 5 Handling of Common Problems
When there are doubts, suspicion of device failure in the process of use, you
can first refer to the following content to troubleshoot.
Before diagnosing a fault, check whether the power supply is properly
connected, the power cable is obviously damaged, the fuse is in good
condition, and the button switch is on.
Fault | Position | Judgement | Solution |
---|
LED lamp light is not bright or
abnormal| Lamp tube is damaged.| Check whether the lamp| If damaged, replace
it.
Circuit| Check whether the cable is in poor contact.| If yes, reconnect the
cables.
Control board| Rule out other damage.| Replace control board.
UV lamp does not light up or
abnormal| Lamp holder| Check that the lamp tube is firmly connected to the
lamp holder, or the lamp holder is damaged.| If not, tighten it.
If damaged, replace lamp holder.
Lamp tube is damaged.| Check whether the lamp i| If damaged, replace it.
Ballast| Check whether the ballast is damaged.| If damaged, replace it.
Circuit| Check whether the cable is in poor contact.| If yes, reconnect the
cables.
Control board| Rule out other damage.| Replace control board.
Fan failure,| Fan| Check whether the fan is damaged.| If damaged, replace it.
Circuit| Check whether the cable is in poor contact| If yes, reconnect the
cables.
The display is not bright| Circuit| Check whether the cable is in poor
contact.| If yes, reconnect the cables.
Screen| Check whether the screen is good.| If not, replace it.
Control board| Rule out other damage.| Replace control board.
Touch screen control is not sensitive| Control screen| Make sure the power
supply is connected and the fuse is intact.| If not, reconnect the cables or
replace the fuse.
Determine the key is damaged.| If yes, contact us.
Determine whether the connecting wiring is poor contact.| If yes, reconnect
the cables.
Replace the touch screen.|
Device power-off
| Power supply| Power supply is not connected well.| Reconnect.
Power cord| Whether the power cord has obvious damage.| If damaged, replace
it.
Fuse| Whether fuse is good.| If not, replace it.
| Transformer| Whether transformer output is normal.| If not, replace it.
Control board| Rule out other damage.| Replace circuit board.
Caution:
- The operation of the above electrical components must be carried out by a qualified electrician under safe conditions (cut off the power). And other parts are not allowed to disassemble, otherwise the consequences are borne by the user.
- If there is other failure that is not in above list, or the operator can not immediately solve, please immediately notify our company’s service department immediately, for your safety, please do not repair the device by yourself;
- The maintenance work of this device can only be undertaken by trained and approved technical personnel;
- If you need to order parts, please find our technical service department, please indicate the model and number of the device you purchased.
Chapter 6 Simple Parts Replacement
Operation of all electrical components of the device must be carried out by a
qualified electrician under safe conditions. When the device has malfunction
and the operator can not immediately troubleshoot, please notify the
maintenance personnel immediately.
Please do not repair the device by yourself for your safety.
If you need to order parts, you can find our technical service department,
inform us of the model and number of the device you have purchased.
6.1 Replace the Fuse
Socket fuse [Φ520(5A)] is located on the right side of cabinet. When you
replace it, first turn off the power and unplug it, use a cross screwdriver to
press and screw the fuse seat counterclockwise, take off the fuse in the fuse
holder and replace it with a new fuse
of same specification, then press and screw the fuse seat clockwise to replace
it; fire line fuse [Φ520(10A)] is also located on the right side, use a flat-
head screwdriver to remove the fuse holder and replace the fuse with a new
one, then press it back.
6.2 Replace the UV Lamp
The product is equipped with life of UV lamp (T8, 40W) for a cumulative
working time of 600 hours. We recommend that you regularly test the UV
intensity in order to achieve good disinfection effect. You can use the UV
intensity test card to confirm whether need to replace the UV lamp or not.
When replacing, first disconnect the power supply, then remove the lamp by
screwing it 90°, take out a new UV lamp of same specification, put it on the
lamp holder and screw 90° in the opposite direction.
6.3 Replace the LED Lamp
When LED lamp needs to be replaced, please disconnect the power supply. Then
tilt the LED integrated bracket to remove, unplug the right bracket connector,
replace the new LED light and install back the bracket connector, and tilt the
LED integrated bracket
into the LED bracket slot.
Caution:
- It is forbidden to wipe the lamp with a wet cloth while cleaning in the powered state.
- It is forbidden to do the replacement in the powered state.
Chapter 7 Precautions
7.1 Storage Conditions
The device should be stored in the warehouse with relative humidity not more
than 75%, temperature less than 40℃, good ventilation, no acid, alkali and
other corrosive gases. The storage cycle should not exceed one year. If it
stored for more than one year, the
open-package inspection is needed and only passed the inspection can it enter
the circulation field.
7.2 Transportation Conditions
The device should be transported in full accordance with the requirements
shown on the outer surface of the packing area. The user should check the
integrity of the packing area carefully when receiving the device with packing
area sent by the logistics company. If the packing area has damage, extrusion
and other phenomena, please refuse to sign, and contact with our company in
time.
Caution:
Please contact with us in time when the Laminar Flow Cabinet that has been
installed and used is relocated and moved again.
7.3 Cautions
-
Before connecting the AC power supply, please ensure that the voltage of the power supply is consistent with input voltage and it is stable, ensure that the rated load of the power socket is not less than the requirements. The cabinet adopts grounding plug, which has the third leg that could only match with grounding type power socket, so it is a safety device. If the plug can not be inserted into the socket, you should ask an electrician to install a grounding type power socket. Be sure to confirm good grounding when using.
-
During the using process of the device, do not put soft, fine items (For example: soft tissue paper) on the work surface, avoid sucking them into the negative pressure duct and fan by the inlet, which can affect the operation of the device.
-
The maximum weight of items placed in the cabinet should not exceed 23kg/25×25cm2.
-
Avoid vibration: Avoid the use of vibrating equipments (such as centrifuges, vortex oscillators, etc.) in the cabinet, because the vibration will cause the accumulation of particulate matter on the filter membrane to shake off, resulting in a reduction in the internal cleanliness of the work area. At the same time, if the balance fails on the front work surface, it will also cause the pollution the operator.
-
Open flame is prohibited: The use of open flame is prohibited in the cabinet. The use of open flame will lead to flocculation of airflow in the work area, and will damage the filter. During the process of experiments that requiring high temperature sterilization, it is highly recommended to use infrared sterilizer.
-
Service life of HEPA. With the extension of the use period, dust and bacteria accumulation in the filter will lead to the increase of HEPA pressure loss. When the increase of the velocity can not meet the requirements, you must promptly contact our service department to replace the HEPA, otherwise it will affect the safety of the device.
Replaced filter should be disposed according to the medical waste. -
The fan and its lower side steel plate is the plenum cover, these air ducts are strictly sealed when leave factory and needed to maintain their tightness. The operator should not loosen or remove the screws of these parts. If there is a special need, it must be handled by our service personnel.
Caution:
Solemnly declare: If the device is not used in accordance with the methods
prescribed by the company, it may damage the protection provided by the
device, and the company should not be responsible for the risks caused by the
operation that not in accordance with the provisions!
Chapter 8 Label Description
| 10A Fuse Label
---|---
| 5A Tubular Fuse for Socket Label
| 10A Tubular Fuse for Blower Label
| Grounding Label
| UV Lamp Warning Label
| Load Requirement Label
| Filter Upstream Label
Chapter 9 Warranty
➢ The warranty period of the purchased product and maintenance contents are
subject to the sales contract.
➢ In the warranty period of the device, if the user improper use caused by
failure or damage, the company does not assume warranty obligations.
➢ Out of the warranty period, our company is also responsible for maintenance,
but charged the corresponding maintenance fees.
➢ The sevice life is 8 years, production date please see product label.
Appendix Wiring Schematic Diagram
- Starting current: 10A
- Operating power: 2050W
- Power installation:
Power should be of reliable grounding.
Dedicated circuit should be adopted for the power, and the wiring capacity
should not less than 4m2 solid copper wire. A three-core socket that meets the
regulations should be used, the polarity of the socket should be used to match
the Laminar Flow Cabinet.
The specifications of power cord and the connection power cord should meet the
relevant standards.