BIOBASE GROUP V900 Laminar Flow Cabinet User Manual

June 1, 2024
BIOBASE GROUP

BIOBASE-GROUP-logo

BIOBASE GROUP V900 Laminar Flow Cabinet

BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-product

Specifications

  • Model: BKCB-V900
  • Dimension (WDH): 9006301760mm
  • Work Area Dimension (WDH): 800610550mm
  • Air Flow Mode: Vertical Flow Type
  • Height of Work Table: 750mm
  • Front Window: 4.8mm HEPA
  • HEPA Specification: 71040069mm, 99.995% PFE for 0.3m particles
  • Noise: 65dB(A)
  • illumination: 300 lx
  • Vibration: 5mrms
  • Temperature Rise: 8
  • Display: LCD Liquid Crystal Display
  • Waterproof Socket: Yes
  • Power Consumption: Does not exceed 500W
  • Grounding Resistance: 0.1
  • Power Supply: 15W UV Lamp, 8W LED Lamp
  • Air Flow Velocity: 0.2-0.45m/s
  • Safe Height of Front Window Opening: Min: 200mm, Max: 350mm
  • Safety Feature: Alarm for front window opening height, UV lamp delayed 10s open for operator safety

Product Usage Instructions

Chapter 2 Installation

Caution: This Laminar Flow Cabinet needs to be installed by trained and qualified engineers.

  1. Remove all packaging components.
  2. Inspect the external surface for any damages.
  3. Carefully inventory accessories according to the packing list.
  4. Move the cabinet to a suitable installation location.

Frequently Asked Questions

  • Q: Can I install the Laminar Flow Cabinet by myself?
    • A: No, the cabinet needs to be installed by trained and qualified engineers to ensure proper setup and functionality.
  • Q: How do I unlock the front window for access?
    • A: Use a Phillips screwdriver to loosen the pan head screw on the counterweight plate to unlock the front window.

Preface

Thank you for using the Laminar Flow Cabinet produced by our company. In order for you to use this product properly and to prevent damage to persons, property and experimental products, please be sure to read and comply with the contents of the manual carefully. If you do not use the cabinet in the way specified by the manufacturer and disassemble the cabinet by yourself, the protection provided by the cabinet lock may be damaged. Only our after-sales service personnel and technicians authorized by us can install and repair this cabinet, otherwise it may cause electric shock or fire. Be sure to place the cabinet firmly on a solid and horizontal surface. If the ground is not firm or the cabinet placed in an inappropriate place, it will cause the cabinet to fall over and cause injury to people. Please use the special power supply indicated on the nameplate of this cabinet, otherwise it may cause fire or electric shock. If the use of voltage does not meet the requirements of the marked power supply, you need to add 3000W or more automatic voltage regulator that suitable for motor load. Please use a grounded power outlet to prevent electric shock. If the power outlet is not grounded, be sure to have a professional technician install the grounding wire. This cabinet is for indoor use only, do not use it in the open air. If Laminar Flow Cabinet is wet by rain, it may cause leakage or electric shock. Do not put the container with water on the Laminar Flow Cabinet. If the item falls, it may cause injury to people. And the run-off water can cause cabinet leakage or electric shock. Do not ground Laminar Flow Cabinet through gas pipes, power pipes, telephone lines or lightning rods. Such grounding may cause electric shock or other greater danger. If the Laminar Flow Cabinet is not operating properly, stop using it and unplug it from the power source. Operating the cabinet in an abnormal condition may cause other hazards such as cabinet leakage. When the Laminar Flow Cabinet is not used for a long time, unplug the power supply to prevent the power cord from causing electric shock, leakage and other dangers due to aging. Do not store acid, alkali and other corrosive, flammable, explosive or volatile hazardous materials in the Laminar Flow Cabinet, or use combustible spray near the Laminar Flow Cabinet, which may easily cause cabinet damage as well as casualties and property damage. If you need to move the cabinet after installation, you need to confirm with the manufacturer. Dispose the replaced old filter as biological waste. The top air outlet must not be blocked. Our company reserves the right to change the design and user manual of the products, when they changes, the user will not be notified

Product Introduction

Product Description

There are strict requirements for the cleanliness of the air in the operating area including laboratories, electronic technology, aerospace, precision equipments and other fields at present. And clean air technology is also required in biological fields such as medicine and health, biopharmaceuticals, food, medical science experiments, sterile microorganisms, etc. Laminar Flow Cabinet is an ideal local air purification equipment to improve the clean working environment. The cleanliness of the work area of this Laminar Flow Cabinet reaches ISO class 5, which can effectively improve the process conditions, ensure the accuracy of products and provide a sterile environment in the fields of scientific experiments and biological experiments.

Product Features

Control of front window

The front window is counterweight controlled, it can be stopped smoothly at any height position within the range, so that the front window can also be controlled to take out the test products or instruments in the operation area when the power is out.

Structure

  1. The cabinet adopts 1.2mm thick cold-rolled steel plate and electrostatic spraying on the surface, which enhances the structural strength and makes the whole cabinet more stable and heavy.
  2. The work surface is made of stainless steel, which is beautiful and corrosion-resistant.
  3. The bracket is made of metal with electrostatic spraying on the surface.
  4. Control panel adopt touch switch to make the cabinet appearance beautiful and easy to operate.

UV sterilization

This Laminar Flow Cabinet using UV-C type 254nm wavelength UV sterilization, which can not only kill the active cells of microorganisms but also kill strong heat-resistant budding spores and other mold spores, in addition, that phage and viruses can quickly break under UV rays.

Technical Parameters

Model BKCB- V900
Dimension(WDH) 9006301760mm

Work area dimension

(WDH)

| 800610550mm
Air flow mode| Vertical Flow Type
Height of work table| 750mm
Front window| 4.8mm
HEPA| HEPA without partitions, with 99.995% PFE for φ0.3μm particles.
Specification of HEPA| 71040069mm
Noise| ≤65dB(A)
Illumination| ≥300 lx
Vibration| ≤5μm(rms)
Temperature rise| ≤8℃
Display| LCD Liquid Crystal Display


Waterproof socket

| ****

The power of the cabinet used does not exceed 500W.

Grounding resistance| ≤0.1Ω
Power supply| AC220V±10%,50 Hz(standard)           AC110V±10%, 60Hz (Optional)
Working area| 304 Brushed stainless steel
Cabinet| Made of 1.2mm (1mm) cold-rolled steel plate and the surface is

electrostatically sprayed.

UV Lamp appointed

timing

| UV lamp delayed 10s open for the safety of the operator.
UV Lamp| 15W
LED lamp| 8W
Velocity of air flow| 0.2-0.45m/s
Safe height of front window

opening

| ****

200mm

Max height of front window

opening

| ****

350mm

Alarm for height of front window opening| ****

Yes

Safety| Colonies numbers≤0.5CFU/30min

Product Structure

BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig
\(1\)

Installation

Caution

This Laminar Flow Cabinet needs to be installed by our trained and qualified engineers.

  1. Remove all packaging components.
  2. Inspect the external surface of the host for scratches, deformations or foreign objects.
  3. Carefully inventory accessories and information according to the packing list in the manual.
  4. Move the entire cabinet to a position as close as possible to the final location and easy to install.

Caution

When handling the Laminar Flow Cabinet, it is strictly forbidden to place or disassemble it upside down.

Installation Environment

  • a. Indoor use.
  • b. Humidity range: ≤80%.
  • c. Power supply voltage:AC 220V±10%(standard) AC 110V±10%(Optional).
  • d. Transient overvoltage: Facility category (overvoltage category) Ⅱ.
  • e. Rated pollution level: Class 2.

Location Selection

  1. Laminar Flow Cabinet should be installed in the clean laboratory or indoor away from the dust source and vibration source, not at the channel. If the laboratory has windows, it should always be in the closed state. Laminar Flow Cabinet should not be placed in the circulation air entrance, for fear that the air blow over the former operation area or blow to the HEPA.
  2. If space allows, there should be 30cm space behind and around the cabinet for cleaning, if not, there should be minimum 8cm per side and 3.8cm space at the back for cleaning. The cabinet power outlet can be closed to facilitate the maintenance, and there is no need to move the cabinet to ensure electrical safety. Do not put the cabinet in the position where it is difficult to operate and disconnect the cabinet.
  3. The power socket of this device can be turned off for maintenance purposes, and there is no need to move the laboratory clean workbench to ensure electrical safety.
    Do not place the device in a position where it is difficult to operate and disconnect the power supply.

Installation Steps

  • a. After moving to the designated indoor location, remove the packing material and check its hardware according to the packing list to make sure there is no damage during transportation.

  • b. Assemble and place the baseBIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig \(2\)
    Screw out the M10×55 inner HXBTs in the cross brace and T brace, assemble the base tightly and firmly refer to the above figure.

  • c. Connect the host and base Refer to the diagram below to connect the host and base, using M10 × Tighten the 55 hexagonal cylindrical head bolt, flat washer 10, and spring washer 10 firmly from bottom to top through the base and side plate.BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig \(3\)

  • d. Unlock the front window
    Before pushing and pulling the front window, it is necessary to use a cross screwdriver to loosen the fixing screws on both sides of the counterweight plate on the back of the cabinet, so that the front window can be pushed and pulled smoothly, as shown in the following figure.BIOBASE-GROUP-V900
-Laminar-Flow-Cabinet-fig \(4\)

  • e. Whole cabinet placement
    The cabinet should be placed in a protected area of airflow to prevent the airflow from ventilation system, air conditioner, door, window and personnel movement from affecting its functionBIOBASE-GROUP-V900-Laminar-Flow-Cabinet-
fig \(5\)
    Caution

    • The cabinet should be installed in a wide and spacious space.
    • When moving the cabinet, it is necessary to cut off the power and adjust the casters before moving slowly.
    • The function of the silicone strip is to prevent damage to the front window glass during transportation. Remove them after placing cabinet.
  • f. Castor adjustmentBIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig \(6\)

Rotate the red part of the castors clockwise to lower the corresponding base legs and lower the height of the cabinet. Simultaneously lowering the four casters can move the cabinet position; Rotate the red part of the casters counterclockwise, raise the corresponding base support feet, and raise the height of the cabinet. Four casters can be adjusted simultaneously to fix the cabinet; Four casters can be adjusted simultaneously to keep the cabinet in a horizontal and stable state.

Check after Installation

After powering on, follow the normal operating procedures to check the following items:

Test item Normal condition
Normal power on-off Connect the power supply and power on the cabinet.
Fan operation Click the “Fan” button,fan works normally.

Front window up and

down movement

| The front window slides smoothly up and down.
LED lamp| Click “LED lamp” button, the LED lamp is on.

  • This product is placed in various departments or teaching laboratories in the laboratory;
  • The installation site should be away from dust sources and seismic sources;
  • Please make sure to plug the power plug into a socket with a ground wire and ensure that the grounding terminal is reliably grounded. Please use a power supply with a leakage protection device;
  • After installation, the interior and surrounding environment of the cabinet must be cleaned carefully using a vacuum cleaner or tools that do not produce fibers.
  • After installation and cleaning, an anemometer can be used to measure the average wind speed in the working area. When the average airflow velocity is less than 0.2m/s, the airflow velocity should be increased. On the contrary, when the average airflow velocity is greater than 0.5m/s, it should be reduced.
    Caution:
    If there is any problem please contact your local dealer for installation adjustment.

Operating Instruction

Introduction to the Main ScreenBIOBASE-GROUP-V900-Laminar-Flow-Cabinet-
fig \(7\)

  • Air Flow Animation: When the cabinet running, display the air flow direction, but if the fan is closed, the animation will stop.
  • UV Lamp: Control the switch of UV lamp.
  • Fan: Control the fan switch.
  • Settings: After clicking “Settings”, it enters the settings interface, which allows you to set the following contents of the cabinet: Timing switch, fan settings, equipment life, standby settings.
  • Fan Gear Adjustment: Real time adjustment of fan gear, with 10 gears divided from low to high, and 10 gears at maximum wind speed;
  • Fan Gear Display: Real time display of fan gear;
  • Time: Real-time display of the current date and time.
  • LED Lamp: Control the open and close of LED lamp.

Settings InterfaceBIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig
\(8\)

Click “Settings” to enter the settings interface, in which you can set contents as shown in the figure.

Timing switch

Click “Timing Switch” in the above figure to enter the timing switch interface, click “Time Input/Display Area” to input the value to be timed, and then you can switch the timing for the corresponding time.BIOBASE-GROUP-V900
-Laminar-Flow-Cabinet-fig \(9\)

Standby settings

The standby time is the time when the screen is darkened and the equipment enters the screen protection state. Click “Standby Settings” to enter the standby settings interface, click the number in the “Standby Time Display Area” to modify the standby time.BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig
\(10\)

Introduction to the usage process

  1. Before the experiment: Lower the glass door to the bottom, click the screen UV lamp button, turn on the UV lamp, and disinfect for more than half an hour.
  2. Manually lift the glass door to the appropriate height, click the fan button, turn on the fan, and after running for half an hour, the experiment can begin.
  3. After the experiment is completed, lower the glass door to the bottom, click the screen UV lamp button, turn on the UV lamp, and disinfect for more than half an hour.
  4. After use, unplug the power plug and place it in the air to avoid contact with moisture and water sources to prevent leakage.

Caution

  1. When disinfecting, people should leave the room to protect their eyes and skin from inadvertent exposure.
  2. The intensity of the UV lamp is tested regularly according to the manufacturer’s specifications, and it is recommended that it be tested once a quarter. If it fails to pass, replace it.

Maintenance

The operability and safety of this cabinet can only be guaranteed if there are enough competent personnel and units for inspection, maintenance and repair.

Comprehensive Maintenance Cycle

Maintenance should be performed every week, every month, every year, every 1000 working hours and every restart.

Maintenance and Repair Methods

Cleaning

Under normal circumstances, cleaning only requires a small amount of household or commercial dishwashing detergent, which will be dissolved in water and wiped directly to wipe off the surface of the cabinet dirt.

  1. When the cabinet is used for the first time or long time out of use again. You should first use a dry cleaning towel to remove the floating dust on the surface, after several cleanings by towel wipe, to confirm the dust-free then use medical alcohol spray on medical gauze to disinfect the operating area, laminar flow plate and work surface, to wipe the comprehensive. Before use, turn on the UV lamp to disinfect, after 30 minutes, turn off the UV lamp and press the fan key, the fan will run.
  2. When the Laminar Flow Cabinet is used continuously. After using the laboratory clean work table every day, first clear all the debris on the work surface, wipe it again with a dry and wet cleaning towel, and finally disinfect the operation area, laminar flow plate and work table surface with medical alcohol spray on medical gauze, wipe it comprehensively to prepare for use on the next day.

Caution:

You can only spray alcohol on the medical gauze with alcohol spray, do not spray alcohol on the laminar flow plate, because the laminar flow plate has a HEPA, which can not be subject to moisture.

Regular maintenance

  1. Daily or weekly cleaning in the process of use
    Use medical alcohol to disinfect and clean the cabinet. Disinfect and clean the operation panel with medical alcohol. Use flexible cleaning agent or glass-specific cleaning agent to clean the outer surface of the area and glass. Check each function of the cabinet according to the user manual.

  2. Monthly cleaning
    Surface cleaning. Check functions of the cabinet, disinfect of the interior of the cabinet. Record this maintenance process.

  3. Annual maintenance
    Check the firm station of steel wire ropes on the front window. Check LED lamp tubes. Conduct a thorough inspection of cabinet’s performance for the safety of it. The maintenance fee is paid by the user.

Caution

  • a. The power supply should be disconnected before performing routine maintenance.
  • b. As the statistics of operating time will directly affect the judgment of maintenance needs, we suggest that a detailed record of operating time should be prepared for reference and inquiry when using this cabinet.
  • c. The fan must be inspected and maintained on a regular basis.
  • d. Surface cleaning: In order to keep the cabinet clean, please clean it regularly (at least once a week is recommended), wipe it with a soft cloth soaked with flexible cleaner and then wrung out and wiped. Please do not spray any chemicals on the operation panel or other labels to prevent discoloration of the label film or unclear writing. Clean the outer surface of the cabinet and the glass with a flexible cleaner or a special cleaner for glass.

Handling of Common Problems

When there are doubts, suspicion of equipment failure in the process of use, you can first refer to the following content to troubleshoot. Before diagnosing the fault, please confirm whether the power supply is connected properly, whether the power cord is significantly damaged, whether the fuse is in good condition, and whether the power lock is in the open state.

Fault Position Judgment basis Solution
LED lamp lamp is not bright or abnormal This cabinet is equipped with an

interlock program, and it is normal to not turn on the ultraviolet lamp when it is turned on. Therefore, it is confirmed whether

the ultraviolet lamp is turned off.

Lamp tube| Check  if

damaged.

| the| lamp| tube| is| Replace lamp tube.
Circuit| Check for poor connections in the

wiring.

| Check circuit.
Control board| Excluding the possibility of other

damages.

| Replace control board.
UV lamp does not lamp up or abnormal| This cabinet is equipped with an interlock program, and it is normal to not turn on the ultraviolet light when it is turned on. Therefore, it is confirmed whether

the ultraviolet light is turned off.

Lamp holder| Check     that    the     connection between the lamp tube and the

lamp holder is secure.

| Replace lamp holder.
Lamp tube| Check  if

damaged.

| the| lamp| tube| is| Replace lamp tube.
Ballast| Check if the ballast is damaged.| Replace ballast.
Circuit| Check for poor connections in the

wiring.

| Check circuit.
Control board| Excluding the possibility of other

damages.

| Replace control board.
The fan does not work| This cabinet is equipped with an interlock program, and it is normal for the fan to not turn on when the UV light is turned on. Therefore, it is confirmed

whether the UV light is turned off.

Fan| Check if the fan is damaged.| If damaged,

fan.

| replace the

Circuit

| Check for poor connections in the

wiring.

|

If any, rewire.

Touch control sensitive

|

screen is       not

| Circuit| Check for poor connections in the

wiring.

| If any, rewire.
Touch screen| Excluding the possibility of other

damages.

| Replace the circuit board.
The cabinet is

power-off

|

Power supply

| Check whether the power supply

is connected properly.

| Reconnect

supply.

| the     power
 | Power cord| Check the power cord for obvious

damage.

| Replace the power cord.
---|---|---|---
Fuse| Check if the safety tube is in good

condition.

| If damaged, replace the

safety tube.

Transformer

| Check whether the transformer output is normal.| If      there      are any

abnormalities, replace the transformer.

Circuit board| Excluding the possibility of other

damages.

| Replace the circuit board.

Caution:

  • a. The operation of the above electrical components must be carried out by qualified electricians under safe conditions (cutting off the power supply). Other components are not allowed to be disassembled, otherwise the consequences will be borne by the user themselves;
  • b. When the equipment encounters a malfunction other than the above, and the operator cannot immediately eliminate it, please notify our company’s maintenance department immediately. For your safety, please do not repair the equipment on your own;
  • c. The maintenance work of this cabinet is only undertaken by trained and recognized technical personnel;
  • d. If you need to order parts, you can contact our technical service department. Please indicate the model and number of the laboratory cleaning workbench you purchased.

Simple Parts Replacement

Operation of all electrical components of the equipment must be carried out by a qualified electrician under safe conditions. When the equipment has malfunction and the operator can not immediately troubleshoot, please notify the maintenance personnel immediately. Please do not repair the equipment by yourself for your safety. If you need to order parts, you can find our technical service department, inform us of the model and number of the Laminar Flow Cabinet you have purchased.

Replace the Fuse

Socket fuse [Φ520(5A)] is located on the right side of Laminar Flow Cabinet. When you replace it, first turn off the power and unplug it, use a cross screwdriver to press and screw the fuse seat counterclockwise, take off the fuse in the fuse seat and replace it with a new fuse of same specification, then press and screw the fuse seat back clockwise; fire line fuse [Φ520(10A)] is also located in the right side of cabinet, use a screwdriver to remove the fuse and replace the fuse with a new one, then press it backBIOBASE-
GROUP-V900-Laminar-Flow-Cabinet-fig \(11\)

Replace the UV Lamp

The product is equipped with life of UV lamp (T8, 40W) for a cumulative working time of 600 hours. We recommend that you regularly test the UV intensity in order to achieve good disinfection effect. You can use the UV intensity test card to confirm whether you need to replace the UV lamp or not. When replacing, first disconnect the power supply, then remove the lamp by screwing it 90°, take out a new UV lamp of same specification, put it on the lamp holder and screw 90° in the opposite direction.

BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig \(12\) BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig \(13\)

Replace the LED Lamp

When LED lamp of the Laminar Flow Cabinet needs to be replaced, please disconnect the power supply. Then tilt the LED integrated bracket to remove, unplug the right side, replace the new LED bracket and then tilt the card into the slot.BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig
\(14\)

Caution:

  1. It is forbidden to wipe the lamp body with a wet cloth while cleaning in the powered state.
  2. Prohibit the replacement in the powered state.

Caution

Storage Conditions

The cabinet should be stored in the warehouse with relative humidity not more than 75%, temperature less than 40℃, good ventilation, no acid, alkali and other corrosive gases. The storage cycle should not exceed one year. If it stored for more than one year, the open-package inspection is needed and only passed the inspection can it enter the circulation field.

Transportation Conditions

The cabinet should be transported in full accordance with the requirements shown on the outer surface of the packing area. The user should check the integrity of the packing area carefully when receiving the cabinet with packing area sent by the logistics company. If the packing area has damage, extrusion and other phenomena, please refuse to sign, and contact with our company in time.

Caution:

Please contact with us in time when the cabinet that has been installed and used is relocated and moved again.

Caution

  1. Before connecting the AC power supply, you need to ensure that the voltage of the power supply is consistent with input voltage and it is stable, and ensure that the rated load of the power socket is not less than the requirements. The cabinet adopts grounding plug, which has the third leg that could only match with grounding type power socket, so it is a safety equipment. If the plug can not be inserted into the socket, you should ask an electrician to install a grounding type power socket. Be sure to confirm good grounding when using.
  2. During the using process of the cabinet, do not put soft, fine items (for example: soft tissue paper) on the countertop, avoid sucking them into the negative pressure duct and fan by the inlet, which can affect the operation of the cabinet.
  3. The maximum weight of items placed in the cabinet should not exceed 23kg/25×25cm2.
  4. Avoid vibration: Avoid the use of vibrating equipments (such as centrifuges, vortex oscillators, etc.) in the cabinet, because vibration will make the particulate matter accumulated on the filter membrane shake off, resulting in the reduction of cleanliness inside the operating area. At the same time, if there is former operating surface balance failure, it will also cause Laminar Flow Cabinet contaminate to the operator.
  5. Open flame is prohibited: The use of open flame is prohibited in the Laminar Flow Cabinet. The use of open flame will lead to flocculation of airflow in the operating area, and will damage the filter. During the process of experiments that requiring high temperature sterilization, it is highly recommended to use infrared sterilizer.
  6. HEPA has service life. With the extension of the use period, dust and bacteria accumulation in the filter will lead to the increase of pressure loss of HEPA. When the increase of the velocity can not meet the requirements, you must promptly contact our service department to replace the HEPA, otherwise it will affect the safety of the equipment. Replaced filter should be disposed according to the medical waste.
  7. The fan and its lower side steel plate is the static pressure area cover, these air ducts are strictly sealed when leave factory and needed to maintain their tightness. The operator should not loosen or remove the screws of these parts. If there is a special need, it must be handled by our service personnel.
  8. The storage period of the cabinet is one year. When it exceed the storage period, it must be opened and inspected once by the company’s technical personnel, only qualified can it be used.

Caution:

Solemnly declare: If the cabinet is not used in accordance with the methods prescribed by our company, it may damage the protection provided by the cabinet, and our company should not be responsible for the risks caused by the operation that not in accordance with the provisions!

Label Description

BIOBASE-GROUP-V900-Laminar-Flow-Cabinet-fig
\(15\)

Warranty Commitment

  • The warranty period of the purchased product and maintenance contents are subject to the sales contract.
  • In the warranty period of the cabinet, if the user improper use caused by failure or damage, our company does not assume warranty obligations.
  • Out of the warranty period, our company is also responsible for maintenance, but charged the corresponding maintenance fees.
  • The service life is 8 years, production date see product label.
  • Our company trains and approves the maintenance unit and maintenance personnel to provide drawings and some necessary technical data of the cabinet.

Appendix Wiring Schematic Diagram

BIOBASE-GROUP-V900-Laminar-Flow-
Cabinet-fig \(16\)

  1. I.Starting current: 10A
  2. Operating power: 200W
  3. Power installation

Power should be of reliable grounding. Dedicated circuit should he adopted for the power, and the wiring capacity should not less than 4m2 solid copper wire. A three-core socket that meets the regulations should be used, the polarity of the socket should be used to match the Laminar Flow Cabinet. The specifications of power cord and the connection power cord should meet the relevant standards.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals