SIEMENS B22348-05 Rapid COVID-19 Antigen Self Test Instruction Manual
- May 15, 2024
- SIEMENS
Table of Contents
- INTENDED USE
- EXPLANATION OF THE TEST
- MATERIALS PROVIDED
- MATERIALS REQUIRED BUT NOT PROVIDED
- QUALITY CONTROL
- TEST PROCEDURES
- STORAGE AND STABILITY
- WARNINGS & PRECAUTIONS
- LIMITATIONS
- PERFORMANCE CHARACTERISTICS
- CLINICAL EVALUATION
- TECHNICAL SUPPORT
- ORDERING AND CONTACT INFORMATION
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
CLINITEST®
RAPID COVID-19
ANTIGEN SELF-TEST
Health Care Provider Instructions for Use (IFU)
For Emergency Use Authorization (EUA) Only.
In vitro diagnostic use only.
INTENDED USE
The CLINITEST® Rapid COVID-19 Antigen Self-Test is a lateral flow immunoassay
device intended for the qualitative detection of nucleocapsid protein antigen
from the SARS-CoV-2 virus.
This test is authorized for non-prescription home use with self-collected
anterior nasal (nares) swab samples from individuals aged 14 or older or adult
collected anterior nasal (nares) swab samples from individuals aged 2 years or
older. This test is authorized for individuals with symptoms of COVID-19
within the first 7 days of symptom onset when tested at least twice over three
days with at least 48 hours between tests, and for individuals without
symptoms or other epidemiological reasons to suspect COVID-19, when tested at
least three times over five days with at least 48 hours between tests.
The CLINITEST® Rapid COVID-19 Antigen Self-Test does not differentiate between
SARS-CoV or SARSCoV-2.
Results are for the identification of SARS-CoV-2 nucleocapsid protein antigen,
which is generally detectable in anterior nasal (nares) samples during the
acute phase of infection. Positive results indicate the presence of viral
antigens, but clinical correlation with patient history and other diagnostic
information is necessary to determine infection status. Positive results do
not rule out bacterial infection or co-infection with other viruses and the
agent detected may not be the definitive cause of disease.
Individuals who test positive with the CLINITEST® Rapid COVID-19 Antigen Self-
Test should self-isolate and seek follow up care with their physician or
healthcare provider as additional testing may be necessary.
All negative results are presumptive and confirmation with a molecular assay,
if necessary for patient management, may be performed. Negative results do not
rule out SARS-CoV-2 infection and should not be used as the sole basis for
treatment or patient management decisions, including infection control
measures such as isolating from others and wearing masks.
Negative results should be considered in the context of an individual’s recent
exposures, history, and the presence of clinical signs and symptoms consistent
with COVID-19.
Individuals who test negative and continue to experience COVID-19 like
symptoms of fever, cough and/or shortness of breath may still have SARS-CoV-2
infection and should seek follow up care with their physician or healthcare
provider.
Individuals should provide all results obtained with this product to their
healthcare provider for public health reporting and to receive appropriate
medical care. All healthcare providers will report all test results they
receive from individuals who use the
authorized product to relevant public health authorities in accordance with
local, state, and federal requirements, using appropriate LOINC and SNOMED
codes, as defined by the Laboratory In Vitro Diagnostics (LIVD) Test Code
Mapping for SARSCoV-2
Tests provided by CDC.
The CLINITEST® Rapid COVID-19 Antigen Self-Test is intended for non-
prescription self-use and/or as applicable, for an adult lay user testing
another aged 2 years or older in a non-laboratory setting. The CLINITEST®
Rapid COVID-19 Antigen Self-Test
is only for in vitro diagnostic use under the Food and Drug Administration’s
Emergency Use Authorization. This product has not been FDA cleared or
approved.
EXPLANATION OF THE TEST
COVID-19 (short for ‘Coronavirus Disease 2019’) is a disease first recognized
in 2019 that is caused by a type of novel coronavirus called SARS-CoV-2. Due
to its rapid spread, the World Health Organization (WHO) recognized the
disease as a global pandemic on
March 11, 2020. Individuals infected with SARS-CoV-2 may have a range of
symptoms from asymptomatic infection to severe respiratory illness and even
death. The virus is spread primarily from person to person through respiratory
particles, even by individuals without symptoms.
The CLINITEST Rapid COVID-19 Antigen Self-Test is a rapid, qualitative
immunochromatographic assay for the determination of the presence of SARS-
CoV-2 antigens in direct human anterior nasal swab specimens. The test strip
is composed of the following
parts: sample pad, reagent pad, reaction membrane, and absorbing pad. The
reagent pad contains the colloidal-gold conjugated with the monoclonal
antibodies against the nucleocapsid protein of SARS-CoV-2; the reaction
membrane contains the secondary
antibodies for nucleocapsid protein of SARS-CoV-2. The whole strip is fixed
inside a plastic device. When the sample is added into the sample well,
conjugates dried in the reagent pad are dissolved and migrate along with the
sample. If SARS-CoV-2
nucleocapsid antigen is present in the sample, a complex forms between the
anti-SARS-2 conjugate and the virus will be captured by the specific anti-
SARS-CoV-2 monoclonal antibodies coated on the test line region (T).
MATERIALS PROVIDED
Contents of the 1 Test Kit:
1 Test Device
1 Sterile Swab
1 Extraction Tube with Buffer and Tip
1 Tube Holder
Instructions for Use
Contents of the 2 Test Kit:
2 Test Devices
2 Sterile Swabs
2 Extraction Tubes with Buffer and Tips
1 Tube Holder
Instructions for Use
Contents of the 4 Test Kit:
4 Test Devices
4 Sterile Swabs
4 Extraction Tubes with Buffer and Tips
1 Tube Holder
Instructions for Use
Contents of the 5 Test Kit:
5 Test Devices
5 Sterile Swabs
5 Extraction Tubes with Buffer and Tips
1 Tube Holder
Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED
- Timer
- Optional Recuro Health web app for reporting
- iPhone or Android phone for using the optional web app (Requires modern mobile web browser e.g. Chrome, Safari)
QUALITY CONTROL
Each CLINITEST Rapid COVID-19 Antigen Self-Test has a built-in internal
procedural control. The reddish-purple line appearing at the “C” position is
an internal procedural control. This procedural control line indicates that
sufficient flow has occurred, and
the functional integrity of the test device has been maintained. A distinct
visible pink/red Control line should always appear if the test has been
performed correctly. If the Control line does not appear, the test result is
invalid and a new test should be
performed. External run controls are not required to use the CLINITEST Rapid
COVID-19 Self-Test in a home setting.
TEST PROCEDURES
Step 1. PLACE TUBE IN TUBE HOLDER
Find tube holder shown on the back of the box.
Push tube through outlined hole. Step 2. OPEN TUBE
Remove the seal from the tube. Avoid spilling the liquid.
Make sure the tube is standing up straight. Step 3. OPEN SWAB
Open the swab pouch on the end opposite the swab tip by peeling back the pouch
cover. Hold the plastic stick end of the swab and remove from pouch. Do NOT
touch the swab end and only handle by the stick end. SWAB BOTH NOSTRILS
Carefully insert swab tip into one nostril about 1/2 to 3/4 of an inch deep.
Do not insert the swab any further if you feel any resistance. Rub the insides
of the nostril in a complete circle at least 5 times. Make sure that you are
rubbing the insides of the nostril. Do not simply roll the swab. Take
approximately 15 seconds to collect the specimen. Be sure to collect any nasal
drainage that may be present on the swab. Remove swab from the nostril and
repeat in your other nostril with the SAME swab. NOTE: If you are swabbing others, please wear a
face mask. With children, the maximum depth of insertion into the nostril may
be less than ¾ of an inch, and you may need to have a second person to hold
the child’s head while swabbing.
NOTE: Failure to swab properly may cause false negative results.
Step 4. PLACE SWAB IN TUBE
Remove the swab from your nostril. Immediately take the tube out of the tube
holder and insert swab tip into the liquid inside the tube. Mix vigorously by
rolling the swab tip at least 6 times on the bottom and sides of the tube.
PLACE TUBE IN TUBE HOLDER
Place the tube back into the tube holder. Keep the swab inside of the tube.
Start timer for 1 minute. NOTE: Do not remove swab before 1 minute has
elapsed. Early removal of the swab may cause false negative results.
REMOVE SWAB FROM TUBE
After 1 minute take the tube out of the tube holder.
As you remove the swab from the tube, squeeze swab tip several times from
outside of the tube. Try to release as much liquid from the swab as possible.
Dispose the swab in the trash. INSERT TIP
Take a tube tip from the kit and push it into the top of the tube.
Make sure there is a tight fit.
NOTE: Please ensure the tip is securely fitted before proceeding. Step
5. OPEN TEST DEVICE
Open the test device pouch by tearing the area circled below. Place the test
device on a flat surface.
ADD 4 DROPS
Hold the tube straight up and down above the test device and gently squeeze to
add 4 drops of solution into the sample well, labeled as “S” on the test
device. Adding more or less than 4 drops of solution into the sample well may
result in incorrect results. START TIMER
Start timer for 15 minutes.
Do not move the test device. Keep on a flat surface.
Step 6. READ TEST RESULT
After 15 minutes find result window, labeled as “C” (for Control) and “T” (for
Test) on the test device. It is important to read your test result at 15-20
minutes. False negative or false positive results can occur if test results
are read before 15 minutes or after 30 minutes.
In the section below are examples for positive, negative and invalid test
results. Used test materials should be thrown away as household waste.
INTERPRETATION OF THE RESULTS
Repeat testing is needed to improve test accuracy. Please follow the table below when interpreting test results.
Status on first day of Testing| First Result Day 1| Second
Result Day 3| Third Result Day 5| Interpretation
---|---|---|---|---
With Symptoms| Positive| N/A| N/A| Positive for
COVID-19
Negative| Positive| N/A| Positive for
COVID-19
Negative| Negative| N/A| Negative for
COVID-19
Without Symptoms| Positive| N/A| N/A| Positive for
COVID-19
Negative| Positive| N/A| Positive for
COVID-19
Negative| Negative| Positive| Positive for
COVID-19
Negative| Negative| Negative| Negative for
COVID-19
Results should be considered in the context of an individual’s recent exposures, history, and the presence of clinical signs and symptoms consistent with COVID-19.
COVID-19 POSITIVE
If the test device looks like the examples below, then protein from the virus
that causes COVID-19 was detected in the sample. The test is positive if there
are two pink/red lines present, one at the Control “C” line and one at the
Test “T” line.
Look very closely for line next to “T”. This line can be very faint. Any
visible pink/ red “T” line is a positive result when the “C” line is also
present. Repeat testing does not need to be performed if patients have a
positive result at any time.
IF THE TEST IS POSITIVE
A positive test result means that the virus that causes COVID-19 was detected
in the sample, and it is very likely the individual has COVID-19 and is
contagious. Please contact the patient’s doctor/primary care physician (if
applicable) and the local health authority immediately and instruct your
patient to adhere to the local guidelines regarding self-isolation.
There is a very small chance that this test can give a positive result that is
incorrect (a false positive). Positive results do not rule out bacterial
infection or co-infection with other viruses. The agent detected may not be
the definite cause of disease.
Individuals who test positive with the CLINITEST Rapid COVID-19 Antigen Self-
Test should self-isolate and seek follow up care with their physician or
healthcare provider as additional confirmatory testing with a molecular test
for positive results may also be necessary, if there is a low likelihood of
COVID-19, such as in individuals without known exposures to COVID-19 or
residing in communities with low prevalence of infection.
COVID-19 NEGATIVE
If the test device looks like the example below then protein from the virus
that causes COVID-19 was not detected. You will only see one line next to “C”
and there will not be any line visible next to”T”. IF THE TEST IS NEGATIVE
To increase the chance that the negative result for COVID-19 is accurate, you should:
- Test again in 48 hours if the individual has symptoms on the first day of testing.
- Test 2 more times at least 48 hours apart if the individual does not have symptoms on the first day of testing.
A negtive test result indicates that the virus that causes COVID-19 was not detected in the sample. A negative result does not rule out COVID-19. There is a higher chance of false negative results with antigen tests compared to laboratory-based tests such as PCR tests. If the test is negative but COVID-19-like symptoms, e.g., fever, cough, and/or shortness of breath continue, follow up testing for SARS-CoV-2 with a molecular test or testing for other respiratory disease should be considered. If applicable, seek follow up care with the primary health care provider. All negative results should be treated as presumptive and confirmation with a molecular assay may be necessary if there is a high likelihood of SARS-CoV-2 infection, such as in an individual with a close contact with COVID-19 or with suspected exposure to COVID-19 or in communities with high prevalence of infection. Negative results do not rule out SARS-CoV-2 infection and should not be used as the sole basis for treatment or patient management decisions, including infection control decisions.
INVALID
If the test device looks like the examples below then the test was not able to
give a result and you must repeat the test with a new swab, a new tube, and a
new test device. The test is INVALID if there is no line next to “C”.
IF THE TEST IS INVALID
If at 15 minutes the line next to the “C” does not appear, even if any shade
of pink/red “T” line appears, the test result is invalid. If the test result
is invalid, a new swab should be collected, and the test should be performed
again with a new tube and test device.
STORAGE AND STABILITY
- CLINITEST Rapid COVID-19 Antigen Self-Test should be stored between 2 to 30 °C (35.6 to 86 °F).
- Kit components in the CLINITEST Rapid COVID-19 Self-Test are stable until the expiration date printed on the label.
- The Test Device must remain in the sealed foil pouch until use.
- For the most current expiration dates of this test, please refer to: https://www.fda.gov/covid-tests.
WARNINGS & PRECAUTIONS
-
Read all instructions carefully before performing the test. Failure to follow the instructions may result in inaccurate test results.
-
In the USA, this product has not been FDA cleared or approved, but has been authorized by FDA under an Emergency Use Authorization.
-
This product has been authorized only for the detection of proteins from SARS-CoV-2, not for any other viruses or pathogens.
-
The emergency use of this product is only authorized for the duration of the declaration that circumstances exist justifying the authorization of emergency use of in vitro diagnostics for detection and/or diagnosis of COVID-19 under Section 564(b)(1)
of the Federal Food, Drug, and Cosmetic Act, 21 U.S.C. § 360bbb-3(b)(1), unless the declaration is terminated or authorization is revoked sooner. -
Serial testing should be performed in individuals with negative results at least twice over three days (with 48 hours between tests) for symptomatic individuals and three times over five days (with at least 48 hours between tests) for asymptomatic individuals. You may need to purchase additional tests to perform this serial (repeat) testing.
-
If you have had symptoms longer than 7 days you should consider testing at least three times over five days with at least 48 hours between tests.
-
The Test is intended to aid in the diagnosis of a current COVID-19 infection. Please consult a healthcare professional to discuss your results and if any additional testing is required.
-
Keep test kit and materials out of the reach of children and pets before and after use.
-
Wear a safety mask or other face-covering when collecting a specimen from a child or another individual.
-
This test is read visually. Users with impaired vision or color-impaired vision may not be able to read the test.
-
An anterior nasal swab sample can be self-collected by an individual age 14 years and older.
-
Children aged 2 to 13 years of age should be tested by an adult.
-
The control line may show up within a few minutes of starting the test. It may take up to 15 minutes for a test line to show up.
-
Do not use on anyone under two years of age.
-
Do not open the kit contents until ready for use. If the test cassette is open for an hour or longer, invalid test results may occur.
-
Do not use the test after the expiration date shown on the kit box.
-
Do not use if any of the test kit contents or packaging is damaged or open.
-
Test components are single-use. Do not re-use.
-
Make sure there is sufficient light when testing. For best results, read test in a well-lit area.
-
Do not use nasal sprays for at least 30 minutes before collecting a nasal sample.
-
Remove any piercings from the nose before starting the test.
-
Do not use on anyone who is prone to nosebleeds or has had facial injuries or head injuries/surgery in the past six months.
-
Inadequate or improper nasal swab sample collection may yield false negative test results.
-
Do not touch the swab tip (specimen collection area) when handling the swab.
-
Once opened, the test card should be used within 60 minutes.
-
The test is intended to be read at 15 minutes. If the test is read before 15 minutes or after 30 minutes, false negative or false positive results may occur. If more than 30 minutes has elapsed, do not interpret the test results and the test should be repeated with a new test cassette.
-
Do not ingest any kit components.
-
Avoid exposure of your skin, eyes, nose, or mouth to the solution in the tube.
-
The solution in the tube and the test device contains an ingredient that is hazardous to skin and eyes (see table below).
If contact with the body occurs, rinse with water. If irritation persists, seek medical advice. https://www.poisonhelp.org or 1-800-222-1222. Chemical Name/CAS| GHS Code for applicable Ingredient| Concentration (%)
---|---|---
Sodium Azide/26628-22-8| Acute Tox. 2 (Oral), H300 Acute Tox. 1 (Dermal), H310| 0.02% (device) and 0.05% (tube)
Triton/9002-93-1| Acute Tox. 4 (Oral), H302| 1.5% -
For more information on EUAs please visit: https://www.fda.gov/emergency-preparedness-and-response/mcm-legal-regulatory-and-policy-framework/emergencyuse-authorization
-
For the most up to date information on COVID-19, please visit: http://www.cdc.gov/COVID19
LIMITATIONS
-
The performance of this test was established based on the evaluation of a limited number of clinical specimens collected between November 2021 – February 2022. The clinical performance has not been established for all circulating variants but is anticipated to be reflective of the prevalent variants in circulation at the time and location of the clinical evaluation.
Performance at the time of testing may vary depending on the variants circulating, including newly emerging strains of SARS-CoV-2 and their prevalence, which change over time. -
Do not use this test for individuals under [2] of age. The swab included in the kit is designed for collection of samples from adults and additional safety measures are needed for safe collection in children under 14 years of age.
-
There is a higher chance of false negative results with antigen tests than withlaboratory-based molecular tests due to the sensitivity of the test technology. This means that there is a higher chance this test will give a false negative result in an individual with COVID-19 as compared to a molecular test, especially in samples with low viral load.
-
This device is only used for testing direct human anterior nasal swab specimens. Viral transport media (VTM) should not be used with this test.
-
The amount of antigen in a sample may decrease as the duration of illness increases.
-
A negative test result may occur if the level of antigen in a sample is below the detection limit of the test or if the sample was collected improperly.
-
The performance of the CLINITEST Rapid COVID-19 Antigen Self-Test was evaluated using the procedures provided in these Instructions for Use (IFU) only. Modifications to these procedures may alter the performance of the test.
-
This test detects both viable (live) and non-viable SARS-CoV-2. Test performance depends on the amount of viral antigen in the sample and may or may not correlate with viral culture results performed on the same sample.
-
Test results must be evaluated in conjunction with other clinical data available to the physician.
-
Failure to follow the Test Procedure may adversely affect test performance and/or invalidate the test result. Incorrect test results may occur if a specimen is incorrectly collected or handled.
-
If the patient continues to have symptoms of COVID-19, and both the patient’s first and second tests are negative, the patient may not have COVID-19, however additional follow-up may be needed. If the test is positive, then proteins from the virus that causes COVID-19 have been found in the sample and the individual likely has COVID-19.
-
Positive test results do not rule out co-infections with other pathogens.
-
Negative results do not rule out SARS-CoV-2 infection, particularly in those who have been in contact with the virus.
Follow-up testing with a molecular diagnostic should be considered to rule out infection in these individuals. -
Negative test results are not intended to rule in other non-SARS viral or bacterial infections.
-
Negative results should be treated as presumptive and confirmed with a molecular assay for clinical management, if necessary.
-
The performance of this test was established based on the evaluation of a limited number of clinical specimens collected between November 2021 to December 2021. The clinical performance has not been established in all circulating variants but is anticipated to be reflective of the prevalent variants in circulation at the time and location of the clinical evaluation.
Performance at the time of testing may vary depending on the variants circulating, including newly emerging strains of SARS-CoV-2 and their prevalence, which change over time.
PERFORMANCE CHARACTERISTICS
a. Analytical Sensitivity: Limit of Detection (LoD)
The Limit of Detection (LoD) of the CLINITEST Rapid COVID-19 Antigen Self-Test
was determined using serial dilutions of the gamma irradiated SARS-CoV-2 (USA-
WA1/2020). A 50-µL sample of gamma irradiated SARS-CoV-2 diluted in PNW was
pipetted onto the dry swab and allowed to absorb for at least 10 seconds. The
swab was then transferred to a pre-filled vial of buffer and mixed for a
minimum of swab six (6) times on the bottom and sides of the tube and remained
in the tube for 1 minute as described in the IFU. Following addition and
mixing of the PNW sample, four (4) drops were added to the sample well of each
device as described in the IFU. Test results were read visually at 15 minutes.
LoD confirmation testing was performed by testing twenty (20) replicates at
the preliminary (1X) LoD concentration. The confirmed LoD for the CLINITEST
Rapid COVID-19 Antigen Self-Test was 7.0 x 103 TCID 50/mL. Based upon the
testing procedure for this study the LoD of 7.0 x 103 TCID 50/mL equates to
350 TCID
50/Swab.
The performance of this test device in the detection of the Omicron variant of
SARS-CoV-2 was evaluated in a dilution series of heat-inactivated clinical
specimens which were positive for the Omicron variant. This testing was
conducted by the National Institutes of Health (NIH) as a component of the
Rapid Acceleration of Diagnostics (RADx®) initiative. Compared to an EUA
authorized RT-PCR method, the CLINITEST Rapid COVID-19 Antigen Self-Test
detected 100% of heat-inactivated Omicron samples at a Ct-value of 23.6 (n=2)
and 100% of live virus Omicron samples at a Ct-value of 21.6 (n=5). Omicron
dilutions at lower viral concentrations (Ct-values greater than 23.6 for heat-
inactivated virus and greater than 21.6 for live virus) were not detected by
the CLINITEST Rapid COVID-19 Antigen Self-Test in this study. Performance in
the detection of SARS-CoV-2 Omicron variant was also demonstrated compared to
the SARS-CoV-2 B.1.2 and Delta variants.
SARS-CoV-2 Variant Detection in Heat-Inactivated and Live Samples
CLINITEST COVID-19 Ag Test| B.1.2 (HI)| Delta (HI)| Omicron (HI)
25.3 (n=2)| 22.1 (n=2)| 23.6 (n=2)
B.1.2 (Live)| Delta (Live)| Omicron (Live)
26.8 (n=2) Lot 1| 25.6 (n=2) Lot 1| 21.6 (n=5)
26.8 (n=2) Lot 2| 24.1 (n=2) Lot 2
(HI: Heat-Inactivated)
| Heat-Inactivated Samples| Live Samples
---|---|---
Sample| Average N2 Ct (n=3)| Percent Positive (n=2)| Average N2 Ct (n=3)|
Percent Positive (n=5)
Omicron- Dilution 1| 21.2| 100%| 19.3| 100%
Omicron-Dilution 2| 22.4| 100%| 19.8| 100%
Omicron-Dilution 3| 23.6| 100%| 20.9| 100%
Omicron- Dilution 4| 24.6| 0%| 21.6| 100%
Omicron-Dilution 5| 25.8| 0%| 22.9| 0%
Omicron-Dilution 6| 27.0| 0%| 23.9| 0%
Omicron-Dilution 7| 28.1| 0%| 24.9| 0%
Omicron-Dilution 8| 29.5| 0%| 26.0| 0%
Omicron-Dilution 9| 30.5| 0%| 26.9| 0%
Omicron-Dilution 10| 31.7| 0%| 27.7| 0%
Omicron-Dilution 11| N/A| N/A| 28.8| 0%
b. High-dose hook effect
The CLINITEST Rapid COVID-19 Antigen Self-Test was tested up to 2.86 x 106
TCID 50/mL of gamma irradiated SARS-CoV-2 (USA-WA1/2020) and no high-dose hook
effect was observed.
c. Endogenous Interfering Substances
The CLINITEST Rapid COVID-19 Antigen Self-Test was evaluated for performance
in the presence of potentially interfering substances that might be present in
a respiratory specimen. Negative specimens were evaluated in triplicate to
confirm that the potentially interfering substances were not cross-reactive
with the test. Specimens containing 3x LoD SARS-CoV-2 (1x LoD is 7.0 x 103
TCID 50/mL) were also evaluated in the presence of interfering substances in
triplicate to confirm that SARS-CoV-2 could still be detected. Interfering
substances testing was performed using a panel of endogenous and exogenous
substances tested at concentrations recommended by the FDA. At the
concentrations tested none of the substances caused a false-positive test
result in unspiked samples or interfered with the detection of a true positive
test result in 3X LoD spiked samples.
Substance | Concentration |
---|---|
Whole blood | 4% |
Mucin (porcine stomach, type II) | 0.50% |
Chloraseptic (Menthol/Benzocaine) | 1.5 mg/mL |
Naso GEL (NeilMed) | 5% v/v |
Nasal Drops (Phenylephrine) | 15% v/v |
Nasal Spray (Oxymetazoline) | 15% v/v |
Nasal Spray (Cromolyn) | 15% v/v |
Zicam | 5% v/v |
Homeopathic (Alkalol) | 10% v/v |
Sore Throat Phenol Spray | 15% v/v |
Tobramycin | 4 μg/mL |
Substance | Concentration |
--- | --- |
Mupirocin | 10 mg/mL |
Tamiflu (Oseltamivir Phosphate) | 5 mg/mL |
Fluticasone Propionate | 5% v/v |
Body and Hand Lotion | 0.5% w/v |
Body Lotion with 1.2% dimethicone | 0.5% w/v |
Hand Lotion | 5% w/v |
Hand Sanitizer with Aloe, 69% ethyl alcohol | 5% w/v |
Hand Sanitizer Cream Lotion | 15% v/v |
Hand Sanitizer, 85% ethanol, fast drying | 15% v/v |
Hand soap liquid gel | 10% v/v |
d. Analytical Specificity: Cross-reactivity and Microbial interference
Cross-reactivity and interference studies were performed for related
pathogens, high prevalence disease agents, and normal or pathogenic flora that
are reasonably likely to be encountered in the clinical specimen of the nasal
cavity. Each organism and virus were tested in both the absence and presence
of inactivated SARS-CoV-2 (SARS-CoV-2 isolate USA-WA1/2020). All testing
samples were prepared in the negative nasal wash. No cross-reactivity was
observed for any of the organisms tested, except for SARS-coronavirus which
exhibited cross-reactivity when tested at 1.58 x 104 TCID 50/mL (Table 8). A
titration of SARS-CoV was performed to find the concentration at which cross
reactivity was no longer observed. Cross reactivity was no longer observed for
SARS-CoV at 1.58 x 102 TCID 50/mL. These results are not unexpected in that
the CLINITEST Rapid COVID-19 Antigen Self-Test targets the nucleocapsid
protein which is present on both the SARS-CoV and SARS-CoV-2 viruses.
ID| Organism| Concentration Tested for Cross
Reactivity| Concentration Tested for Microbial Interference
---|---|---|---
229E| Human coronavirus 229E| 1.43 × 105 TCID50 /mL| 1.43× 105 TCID50 /mL
OC43| Human coronavirus OC43| 1.43 × 105 TCID50 /mL| 1.43 × 105 TCID 50/mL
NL63| Human coronavirus NL63| 1.43 × 105 TCID50 /mL| 1.43 × 105 TCID50 /mL
SARS| SARS-coronavirus| 1.58 × 104 TCID /mL 50| N/A
SARS1:1000| SARS-Coronavirus| 1.58 × 102 TCID 50/mL| 1.58 × 102 TCID50 /mL
MERS| MERS-coronavirus| 1.43 × 105 TCID 50/mL| 1.43 × 105 TCID50 /mL
AV1| Adenovirus| 1.43 × 105 TCID 50/mL| 1.43 × 105 TCID50 /mL
hMPV| Human metapneumovirus 4 Type B2| 1.43 × 105 TCID 50/mL| 1.43 × 105
TCID50 /mL
P1| Parainfluenza virus 1| 3.60 × 105 TCID 50/mL| 3.60 × 105 TCID50 /mL
P2| Parainfluenza virus 2| 1.43 × 105 TCID50 /mL| 1.43 × 105 TCID 50/mL
P3| Parainfluenza virus 3| 1.43 × 105 TCID 50/mL| 1.43 × 105 TCID 50/mL
P4| Parainfluenza virus 4b| 1.43 × 105 TCID50 /mL| 1.43 × 105 TCID50 /mL
FluA| Influenza A| 1.43 × 105 CEID50 /mL| 1.43 × 105 CEID50 /mL
FluB| Influenza B| 1.43 × 105 TCID 50/mL| 1.43 × 105 TCID 50/mL
EV68| Enterovirus 68| 1.43 × 105 TCID 50/mL| 1.43 × 105 TCID 50/mL
RSV| Respiratory syncytial virus| 1.43 × 105 pfu/mL| 1.43 × 105 pfu/mL
RV| Rhinovirus| 1.43 × 105 TCID 50/mL| 1.43 × 105 TCID 50/mL
HI| Haemophilus influenzae| 1.00 ×106 cfu/mL| 1.00 ×106 cfu/mL
SPN| Streptococcus pneumonia| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
SPY| Streptococcus pyogenes| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
CA| Candida albicans| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
CA| Candida albicans (Kansas City Test)| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
BP| Bordetella pertussis| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
MP| Mycoplasma pneumonia| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
CP| Chlamydia pneumoniae| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
LP| Legionella pneumophila| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
MT| Mycobacterium tuberculosis| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
PC| Pneumocystis carinii| 1.25 × 106 nuclei/mL| 1.25 × 106 nuclei/mL
PJ| P. jiroveci-S. cerevisiae| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
SA| Staphylococcus aureus subsp. aureus| 1.00 × 106 cfu/mL| 1.00 × 106 cfu/mL
SE| Staphylococcus epidermidis| 4.66 × 105 cfu/mL| 2.33 × 105 cfu/mL
PNM| Pooled Negative Matrix| N/A| N/A
e. Flex Study
A robust use of CLINITEST Rapid COVID-19 Antigen Self-Test was demonstrated by
six (6) flex studies as follows;
- Non-level positioning of Test Device
- Varying the Extraction Solution volume
- Varying the swab rotation number
- Sample volume variability
- Result reading time variability
- Temperature and humidity
CLINICAL EVALUATION
a. Serial Performance
A prospective clinical study was conducted between January 2021 and May 2022
as a component of the Rapid Acceleration of Diagnostics (RADx) initiative from
the National Institutes of Health (NIH). A total of 7,361 individuals were
enrolled via a decentralized clinical study design, with a broad geographical
representation of the United States. Per inclusion criteria, all individuals
were asymptomatic upon enrollment in the study and at least 14 days prior to
it and did not have a SARS-CoV-2 infection in the three months prior to
enrollment. Participants were assigned to one of three EUA authorized SARS-
CoV-2 OTC rapid antigen tests to conduct serial testing (every 48 hours) for
15 days. If an antigen test was positive, the serial-antigen testing result is
considered positive.
At each rapid antigen testing time point, study subjects also collected a
nasal swab for comparator testing using a home collection kit (using a
15-minute normalization window between swabs). SARS-CoV-2 infection status was
determined by a composite comparator method on the day of the first antigen
test, using at least two highly sensitive EUA RT-PCRs. If results of the first
two molecular test were discordant a third highlysensitive EUA RT-PCR test was
performed, and the final test result was based upon the majority rule.
Study participants reported symptom status throughout the study using the
MyDataHelps app. Two-day serial antigen testing is defined as performing two
antigen tests 36 – 48 hours apart. Three-day serial antigen testing is defined
as performing three antigen tests over five days with at least 48 hours
between each test.
Out of the 7,361 participants enrolled in the study, 5,609 were eligible for
analysis. Among eligible participants, 154 tested positive for SARS-CoV-2
infection based on RTPCR, of which 97 (62%) were asymptomatic on the first day
of their infection, whereas 57 (39%) reported symptoms on the first day of
infection. Pre-symptomatic subjects were included in the positive percent
agreement (PPA) of asymptomatic individuals, if they were asymptomatic on the
first day of antigen testing, regardless of whether they developed symptoms at
any time after the first day of testing.Performance of the antigen test with
serial testing in individuals is described in the table below.
Data establishing PPA of COVID-19 antigen serial testing compared to the
molecular comparator single day testing throughout the course of infection
with serial testing. Data is from all antigen tests in study combined.
DAYS AFTER FIRST PCR POSITIVE TEST RESULT| ASYMPTOMATIC ON FIRST DAY
OF TESTING| SYMPTOMATIC ON FIRST DAY OF TESTING
---|---|---
Ag Positive / PCR Positive (Antigen Test Performance % PPA)
1 Test| 2 Tests| 3 Tests| 1 Test| 2 Tests| 3 Tests
0| 9/97
(9.3%)| 35/89
(39.3%)| 44/78
(56.4%)| 34/57
(59.6%)| 47/51
(92.2%)| 44/47
(93.6%)
2| 17/34
(50.0%)| 23/34
(67.6%)| 25/32
(78.1%)| 58/62
(93.5%)| 59/60
(98.3%)| 43/43
(100%)
4| 16/21
(76.2%)| 15/20
(75.0%)| 13/15
(86.7%)| 55/58
(94.8%)| 53/54
(98.1%)| 39/40
(97.5%)
6| 20/28
(71.4%)| 21/27
(77.8%)| 16/18
(88.9%)| 27/34
(79.4%)| 26/33
(78.8%)| 22/27
(81.5%)
8| 13/23
(56.5%)| 13/22
(59.1%)| 4/11
(36.4%)| 12/17
(70.6%)| 12/17
(70.6%)| 7/11
(63.6%)
10| 5/9
(55.6%)| 5/8
(62.5%)| | 4/9
(44.4%)| 3/7
(42.9%)|
1 Test = one (1) test performed on the noted days after first PCR positive
test result. Day 0 is the first day of documented infection with SARS-CoV-2.
2 Tests = two (2) tests performed an average of 48 hours apart. The first
test performed on the indicated day and the second test performed 48 hours
later.
3 Tests = three (3) tests performance an average of 48 hours apart. The first
test performed on the indicated day, the second test performed 48 hours
later, and a final test performed 48 hours after the second test.
b. Clinical Performance
A prospective study was completed at five (5) sites in the United States for
clinical validation of the CLINITEST Rapid COVID-19 Antigen Self-Test for the
detection of the SARS-CoV-2 in subject-collected anterior nasal (AN) swab
samples. The study evaluated the investigational test’s performance in
symptomatic individuals (those suspected of COVID-19). A total of 277
symptomatic subjects were enrolled and each were currently experiencing
symptoms associated with COVID-19, within 7 days of symptom onset. Each
enrolled subject either self-collected one sample from their anterior nasal
passages (from both nostrils), or had one sample collected from him/her by
another individual. Each subject then had a mid-turbinate sample (from both
nostrils) collected from him/her by one of the study personnel. Test results
from the CLINITEST Rapid COVID-19 Antigen Self-Test Test (candidate test) were
compared to highly sensitive molecular FDA EUA Authorized SARS-CoV-2 assays to
determine test performance. As shown, the positive percent agreement (PPA) is
85.4% and the negative percent agreement (NPA) is 99.3% with the 95%
confidence interval bounds of 78.4% to 90.9% for the PPA and 95.9% to 100% for
the NPA, respectively.
CLINITEST Rapid COVID-19 Antigen Self-Test (candidate) results vs.
Comparator results
| Comparator Positives| Comparator Negatives| Total
---|---|---|---
Candidate Positives| 117| 1| 118
Candidate Negatives| 20| 139| 159
Total| 137| 140| 277
Positive Percent Agreement (PPA) = (117/137) x 100% = 85.4% (95% CI = 78.4 to
90.9%)
Negative Percent Agreement (NPA) = (139/140) x 100% = 99.3% (95% CI = 96.1 to
100%)
Age and Gender Analysis
Subject Age| Female| Male| Comparator Positives| %
Positivity Rate (by Comparator)
---|---|---|---|---
<14 years of age| 16| 18| 7| 20.6
14-24 years of age| 22| 13| 20| 57.1
24-64 years of age| 102| 82| 94| 51.1
≥65 years of age| 13| 11| 16| 66.7
Total| 153| 124| 137| 49.5
Clinical Performance in Subjects on Different Symptomatic Days
Days of COVID-19 Symptoms| Number of Specimens Tested|
Investigational Positives| Comparator Positives
---|---|---|---
Day 0-1| 53| 22| 27
Day 2| 79| 31| 32
Day 3| 55| 27| 27
Day 4| 35| 14| 17
Day 5| 28| 10| 14
Day 6| 14| 7| 9
Day 7| 13| 7| 11
Total| 277| 118| 137
TECHNICAL SUPPORT
For questions, or to report a problem, please call Technical Support at
833-933-2340 (Available Hours: Mon. to
Fri.: 9 a.m. – 5 p.m. PST) or covidhometest-USA.dl@siemens-healthineers.com
.
Test system problems may also be reported to the FDA using the MedWatch
reporting system (phone: 1-800 FDA-1088; fax: 1-800 FDA-1078: or
http://www.fda.gov/medwatch).
ORDERING AND CONTACT INFORMATION
Siemens Healthineers
covidhometest-USA.dl@siemens-healthineers.com
http://www.clinitest.siemens-healthineers.com/us
You may see one or more of these symbols on the labelling/packaging of this product:
| Manufacturer| | Date of manufacture
---|---|---|---
| Contains sufficient for
| In vitro diagnostic medical device| | Use-by date
| Consult instructions for use| | Batch code
| Temperature limit| | Do not reuse
Revision Date: 2023-02,
B22348-05 Rev. E
References
- Siemens Healthineers | Corporate Home
- Coronavirus Disease 2019 (COVID-19) | CDC
- CLINITEST® Rapid COVID-19 Self-Test - Instructions for Use (USA)
- MedWatch: The FDA Safety Information and Adverse Event Reporting Program | FDA
- Emergency Preparedness and Response | FDA
- 800-222-1222" class="nturl" > Poison Control | Call Poison Help Centers Now 1-800-222-1222
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>