myjealthchecked 10322821 Iron Deficiency Lateral Flow Test Instruction Manual
- June 1, 2024
- myjealthchecked
Table of Contents
- myjealthchecked 10322821 Iron Deficiency Lateral Flow Test
- Specifications
- INTENDED USE
- SUMMARY
- PRINCIPLE
- PRECAUTIONS
- INSTRUCTIONS
- READING THE RESULTS
- LIMITATIONS
- PERFORMANCE CHARACTERISTICS
- FAQ
- INDEX OF SYMBOLS
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
- 550, Yinhai Street,
myjealthchecked 10322821 Iron Deficiency Lateral Flow Test
Specifications
- Intended Use: Rapid test for qualitative detection of ferritin in human finger-prick blood
- For: Aid in the diagnosis of iron deficiency and anemia
- Usage: Self-testing in vitro diagnostic use only
INTENDED USE
The Iron Deficiency (Ferritin) Rapid Test is a rapid test for the qualitative detection of ferritin in human finger-prick blood to aid the diagnosis of iron deficiency and anaemia. For self-testing in vitro diagnostic use only.
SUMMARY
Anaemia due to iron depletion is common in children and adults of all ages, particularly in menstruating women (at least 20% suffer from iron deficiency). The main symptoms are extreme fatigue, difficulty concentrating, headache, pale skin, muscle and joint pain, weight gain, palpitations, sometimes associated with sleep disturbances, weakness, chest pain, fast heartbeat or shortness of breath. Iron deficiency anaemia occurs when your body does not have enough iron to produce haemoglobin. Haemoglobin is the part of red blood cells that gives blood its red colour and enables the red blood cells to carry oxygenated blood throughout your body. If you are not consuming enough iron (due to a lack of iron in your diet or, an inability to absorb iron), or if you’re losing too much iron (from blood loss, or pregnancy for example), your body cannot produce enough haemoglobin, and iron deficiency anaemia will eventually develop. Mild iron deficiency anaemia usually doesn’t cause complications. However, left untreated, iron deficiency anaemia can become severe and lead to health problems, including heart failure, problems during pregnancy and delayed growth and development in children. Low ferritin may also indicate hypothyroidism, vitamin C deficiency or coeliac disease. Low ferritin levels are seen in some patients with restless legs syndrome, not necessarily related to anaemia, but perhaps due to low iron stores short of anaemia.1,2
PRINCIPLE
The Iron Deficiency (Ferritin) Rapid Test is a qualitative, lateral flow immunoassay for the detection of human ferritin in human whole blood. The membrane is precoated with anti-ferritin polyclonal antibody on the test line region. The gold is pre-coated with anti-ferritin monoclonal antibody and Rabbit IgG. During testing, the specimen reacts with the particle coated with an anti-ferritin monoclonal antibody. The mixture migrates upward on the membrane chromatographically by capillary action to react with anti-ferritin polyclonal antibody on the membrane and generate a coloured line. The line in test line region (T) appears if the ferritin level exceeds the cut-off level of 30 ng/mL. If the ferritin concentration is less than 30 ng/mL, the test line does not appear. To serve as a procedural control, a coloured line will always appear in the control line region, indicating that the proper volume of specimen has been added and membrane wicking has occurred.
PRECAUTIONS
Please read all the information in this Instructions for use before performing the test
- For self-testing in vitro diagnostic use only
- Do not eat, drink or smoke in the area where the specimens or kits are handled
- Store in a dry place at 2-30°C (36-86°F), avoiding areas of excess moisture. If the foil packaging is damaged or has been opened, please do not use
- This test kit is intended to be used as a preliminary test only and repeatedly abnormal results should be discussed with a doctor or medical professional
- Follow the indicated time strictly
- Use the test only once. Do not dismantle and touch the test window of the test cassette
- The kit must not be frozen or used after the expiration date printed on the package
- Keep out of the reach of children
- The used test should be discarded in your general household waste
STORAGE AND STABILITY
Store as packaged at room temperature or refrigerated (2-30°C). The test is
stable up until the expiration date printed on the sealed pouch or label of
the closed canister. The test must remain in the sealed pouch or closed
canister until use. Do not freeze. Do not use after the expiration date.
MATERIALS PROVIDED
- Test cassette
- Capillary dropper
- Buffer
- Alcohol pad
- Lancet
- Plaster
- Instructions for use
MATERIALS REQUIRED BUT NOT PROVIDED
- Timer
PREPARATION
- Wash your hands with soap and rinse with clear warm water.
- Bring the pouch to room temperature before opening it.
- Open the pouch, and remove the dropper, buffer vial, lancet, alcohol pad and test cassette placing them on a clean, level surface.
- Perform the test within one hour with the best results obtained if the test is performed immediately after opening the foil pouch.
INSTRUCTIONS
- Carefully pull off and dispose of the cap on the lancet.
- Use the alcohol pad to clean the fingertip which will be pricked with the lancet. Allow to air dry.
- Press the orange end of the lancet down against the fingertip to prick the puncture site. The tip retracts automatically and safely after use.
- Without touching the puncture site, massage the hand towards the fingertip, keeping the hand pointing downwards to obtain the sample of blood.
- Without squeezing the capillary dropper bulb, place the end of the capillary tube in the blood until it reaches the fill line. Massage your finger again to obtain more blood if the blood does not reach the indicated line. Avoid air bubbles.
- Transfer the collected blood into the sample well (S) of the cassette, by squeezing the dropper bulb. Apply the spot plaster to puncture site if required.
- Unscrew the cap of the buffer vial and add 2 drops of buffer into the sample well (S) of the cassette.
- Start a timer and wait for the coloured line(s) to appear.|
- Read results at 5 minutes. Do not interpret the result after 10 minutes.
READING THE RESULTS
LIMITATIONS
- The Iron Deficiency (Ferritin) Rapid Test provides only a qualitative analytical result. A secondary analytical method must be used to obtain a confirmed result.
- It is possible that technical or procedural errors, as well as other interfering substances in the whole blood specimen, may cause erroneous results.
- As with all diagnostic tests, all results must be considered with other clinical information available to the physician.
PERFORMANCE CHARACTERISTICS
The specimen correlation used a specimen number (n) equal to 102 specimens, including 79 normal whole blood specimens and 23 abnormal whole blood specimens were confirmed by CLIA. The result demonstrated showed that the abnormal coincidence rate is 91.3%, the normal coincidence rate is 96.2% and the total coincidence rate is 95.1%..
Method **CLIA** | Total Results |
---|
Iron Deficiency (Ferritin) Rapid Test
| Results| Abnormal| Normal
Abnormal| 21| 3| 24
Normal| 2| 76| 78
Total Results| 23| 79| 102
- Abnormal coincidence rate= 21/(21+2)*100%=91.3%
- Normal coincidence rate=76/(3+76)*100%=96.2%
- Total coincidence rate=(21+76)/(21+3+2+76) *100% = 95.1%
Accuracy
- The Iron Deficiency (Ferritin) Rapid Test has been compared with a leading commercial
- Ferritin CLIA test. The correlation between these two systems is over 95.0%.
Analytical Sensitivity
- The Iron Deficiency (Ferritin) Rapid Test can detect levels of ferritin in human fingerstick blood as low as 30 ng/mL.
Intra-Assay
Within-run precision has been determined by using 10 replicates of three
specimens: 0 ng/mL, 30 ng/mL and 100 ng/mL specimens. The specimens were
correctly identified > 99% of the time.
Cross-Reactivity
An evaluation was performed to determine the cross-reactivity and
interferences of the Iron Deficiency (Ferritin) Rapid Test. There is no cross-
reactivity with HAMA, RF, Human serum albumin, human AFP, Ferric Chloride,
human transferrin and human hemoglobin.
Inter-Assay
Between-run precision has been determined by 10 independent assays on the
same 3 specimens: 0 ng/mL ferritin, 30 ng/mL ferritin, and 100 ng/mL ferritin
standard sample. Three different lots of the Iron Deficiency (Ferritin) Rapid
Test have been tested using these specimens. The specimens were correctly
identified >99% of the time.
BIBLIOGRAPHY
- Kryger MH, Otake K, Foerster J (March 2002). “Low body stores of iron and restless legs syndrome: a correctable cause of insomnia in adolescents and teenagers” .SleepMed.3(2): 127-32.
- Mizuno S, Mihara T, Miyaoka T, Inagaki T, Horiguchi J (14 March 2005). “CSF iron, ferritin and transferrin levels in restless legs syndrome”. J Sleep Resl: 43-7.
FAQ
How does the Iron Deficiency (Ferritin) Rapid Test work?
Ferritin is a protein and the primary form of iron stored inside cells. An abnormal result means that the ferritin concentration in blood is lower than 30 ng/mL and a possible iron deficiency.
2. When should the test be used?
The Iron Deficiency (Ferritin) Rapid Test can be performed in case of symptoms like paleness, feeling tired, headaches, faster heartbeat or shortness of breath during exercise; mainly, if female, when pregnant or in case of excessive bleeding during periods. The test can be performed anytime of the day, but must not be performed in cases of disease, acute inflammation, or spleen or liver injury. Abnormal results can be obtained even in case of no iron deficiency.
2 Can the recult he incorrent?
The results are accurate as long as the instructions are carefully followed. However, the result can be incorrect if the ferritin test gets wet before test is performed or if the quantity of blood dispensed in the sample well is not sufficient. The capillary dropper provided in the box ensures the collected blood volume is correct. Due to immuno- logical principles involved, there is still a chance of false results in rare cases. A consultation with the doctor is always recommended for such tests based on immunological principles.
4. What is the line that appears under the c (control) line?
When this line appears, it confirms that the test has performed correctly.
5. If i read the result after 10 minutes, will the result be reliable?
No. The result should be read at 5 minutes after adding the buffer. The result is not reliable after 10 minutes.
6. What do i have to do if the result is abnormal?
If the result is abnormal, it means that the ferritin level is lower than normal (30 ng/mL) and that you should consult a physician to discuss the test result and decide whether additional analysis should be performed.
7. What do i have to do if the result is normal?
If the result is normal, it means that the ferritin level is higher than 30 ng/mL and is within the normal range. However, if the symptoms persist, it is recommended to consult a physician.
INDEX OF SYMBOLS
- Hangzhou AliTest Biotech Co., Ltd.
-
550, Yinhai Street,
- Hangzhou Economic & Technological Development Area
- Hangzhou, 310018 P.R. China
Lancet:
- Promised Hangzhou Meditech Co., Ltd.
- No. 1388 Cangxing Street, Cangqian Community,
- Yuhang District, Hangzhou City, 311121 Zhejiang, China
OR
- Ningbo Medsun Medical Co., Ltd.
- No. 55 Jinxi Road, Zhenhai 315221 Ningbo
- People’s Republic of China
Alcohol pad
- Jiangsu Sunclean Medical Co., Ltd.
- No. 11 Fenghuang South Road,
- Hutang Town, Wujin District
- 213162 Changzhou City, Jiangsu Province, P.R. China
OR
- Ningbo Medsun Medical Co., Ltd.
- No. 55 Jinxi Road, Zhenhai 315221 Ningbo
- People’s Republic of China
Plaster
-
Nantong City YOJO Medical Products C, Ltd
-
Industrial Park, Sanxing Town, Haimen City,
-
Jiangsu Province 226112 China
-
MedNet EC-REP GmbH, Borkstrasse 10, 48163 Muenster, Germany
-
MT Promedt Consulting GmbH, Altenhofstr. 80 66386 St., Ingbert, Germany
-
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany
-
Medpath GmbH, Mies-van-der-Rohe Strasse 8, 80807 Munich, Germany
-
Shanghai International Holding Corp. GmbH (Europe) Eiffestrasse 80, 20537 Hamburg, Germany
-
SUNGO Europe B.V. Olympisch Stadion 24, 1076DE Amsterdam, Netherlands
Number: 14601325600
Revision date: 2023-02-03
Ferritin Rapid Test Cassette
DHF-0078
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>