TANEL WIP-15A Moisture Meter Paper User Manual
- June 1, 2024
- Tanel
Table of Contents
TANEL WIP-15A Moisture Meter Paper
Application
Moisture Meter for Paper WIP-15A is an electronic device for measuring
moisture contents of paper. It measures the dielectric constant (SIC) of
paper. The final reading is an average level of moisture in a layer of paper
50 mm thick.
Moisture Meter WIP-15A is a modified and improved version of previously
produced WIP-20A.
Technical data
Range | 2 % – 15 % |
---|---|
Thickness | |
of layer measured | approx. 50 mm |
Display | LCD |
Accuracy | ± 0.6 % |
Resolution | 0.1 |
Power | 9V, battery 6F22 |
Battery life | approx. 5000 measurements |
Low bat signal | yes |
Auto power off | yes, after 5 minutes |
Size | |
instrument | 165 80 33 mm |
case | 270 180 55 mm |
Preparing the instrument for measurements
Measurements of wrapping paper G=45 g/m2 or other paper with similar gsm and composition.
To prepare the instrument for measurements:
- set the SELECT knob to CALIBRATION,
- hold the instrument in the air and use CALIBRATION knob to reset the displayed value to 1.0 (acceptable error margin is ± 0.1),
- set the SELECT knob to MEASURE (LCD should display 2.0).
Measurements of other types of paper
To prepare the instruments for measurements of other types of paper perform
once the calibration procedure as described below:
- prepare a sample of the paper,
- find the moisture contents of the sample using the dry oven test, preferably for moisture contents within the range of 5%÷7%,
- set the knob to MEASURE,
- conduct the measurement as described in chapter ‘4. Measurements’ of this users manual,
- between each measurement either increase or decrease the calibration value,
- use the CALIBRATION knob to set the result displayed on the LCD to the result of the dry oven test,
- increasing calibration decreases the result and vice versa,
- read the calibration value for this type of paper (e.g. 0.65),
- write the type of paper and it’s calibration value into the calibration table,
- for subsequent measurements of this type of paper use the calibration value from the table i.e. before measurements hold the instrument in the air, set the knob to CALIBRATION, and set the appropriate calibration value (e.g. 0.65) instead of the standard value of 1.0.
CALIBRATION TABLE
Paper type| Calibration value (instrument in the air, knob set to CALIBRATION)
Measurements
Conduct the measurements of moisture contents of paper in at least several places. These places should be smooth and clean. The thickness of the layer of paper should be larger than 50 mm. The final result is the average of all measurements.
Hold the instrument as show on fig. 1. Set the SELECT knob to MEASURE. Press the metal electrodes against the paper. Read the result. Results that are out of range (below 2% and above 15%) are indicated as follows:
- below 2% – 2.0%
- above 15% – 15%.
Important
To prevent faulty results in measurements the Producer recommends that in due time you check your meter readings with a dry oven test.
Limitations
This warranty becomes null and void if you fail to pack your WIP-15A in a manner consistent with the original product packaging and dam-age occurs during product shipment. This warranty does not cover: circumstance beyond our company’s control; service required as the result of unauthorized modifications or service; misuse, abuse; failure to follow our company operating or maintenance instructions. Repair or replacement without charge is our company’s only obligation under this warranty. Our company will not be responsible for any special, consequential or incidental damages resulting from the purchase, use, or improper functioning of this equipment regardless of the cause. Such damages for which our company’s will not be responsible include, but are not limited to, loss of revenue or profit, downtime costs, loss of use of the equipment, cost of any substitute equipment, facilities of services, or claims of your customers for such damages.
Changing the battery and storage
The device is equipped with a power level control circuit. When the power drops below acceptable level a sign ”BAT” appears on the left side of the LCD. This indicates that the battery has expired and should be replaced with a new one. To replace the battery unscrew the screw on the back of the instrument and carefully remove the back panel. Store the device in a dry place.
Guarantee and Service
The Moisture Meter for Paper WIP-15A has a 12 months guarantee.
- all repairs during the guarantee period will be done within 10 days from the date of delivery,
- all other repairs will be done within 14 days from the date of delivery.
TANEL Electronic warrants the Moisture Meter for Paper WIP-15A to be free from malfunction and defects in both materials and workmanship for one year (12 months) from the date of purchase.
If the Moisture Meter WIP-15A does not function properly during the warranty period due to defects in either materials or workmanship, our company will, at its option, either repair or replace the instrument without charge, subject to the conditions and limitations stated herein. Such repair service will include any necessary adjustments and replacement part.
CUSTOMER SUPPORT
Producer
TANEL
Electronics & IT General Partnership
Poland , 44-100 Gliwice, Kopernika 121
tel. +48 32 238-16-15, 234-96-15
https://www.tanel.com.pl/
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>