MELING YC-55L Pharmacy Medical Refrigerator Instruction Manual
- June 1, 2024
- MELING
Table of Contents
- YC-55L Pharmacy Medical Refrigerator
- Specifications
- Product Usage Instructions
- 1. Installation and Commissioning
- 2. Operating Instructions
- 3. Maintenance and Service
- 4. Troubleshooting and Maintenance Services
- Q: Can I adjust the temperature settings of the
- Q: How often should I clean the interior of the
YC-55L Pharmacy Medical Refrigerator
Specifications
-
Models: YC-55L, YC-55EL, YC-56L, YC-56EL, YC-75L, YC-75EL,
YC-76L, YC-76EL, YC-130L, YC-130EL, YC-315L, YC-315EL, YC-330L,
YC-330EL, YC-395L, YC-395EL, YC-400L, YC-400EL, YC-525L, YC-525EL,
YC-725L, YC-725EL, YC-1015L, YC-1015EL, YC-1320L, YC-650L,
YC-1505L -
Manufacturer: Zhongke Meiling Cryogenics Company Limited
-
Website: www.melingbiomedical.com
-
Email: zkmeiling@zkmeiling.com
Product Usage Instructions
1. Installation and Commissioning
4.1 Installation Environment
Ensure the installation environment is well-ventilated and away
from direct sunlight or rain exposure to prevent short circuits or
overheating.
4.2 Installation Site
Place the refrigerator on a solid and flat surface to prevent
tipping over which may lead to equipment damage or personal
injury.
4.3 Preparation before Use
Check that the power cord is undamaged and securely plugged into
a grounded outlet with a copper conductor cross-sectional area of
at least 4mm².
4.4 First Power-on
Switch on the refrigerator and allow it to stabilize for a few
minutes before placing any items inside.
4.5 Operation after Power Failure
After a power failure, check the contents for spoilage and
verify the temperature settings before continuing to use the
refrigerator.
2. Operating Instructions
6.1 Function Introduction
Familiarize yourself with the temperature controls and ensure
proper storage of pharmaceuticals or medical supplies according to
the manufacturer’s guidelines.
6.2 Optional Function
Explore any optional functions such as temperature alarms or
data logging for enhanced monitoring of stored items.
3. Maintenance and Service
7.1 Equipment Maintenance
Regularly clean the interior of the refrigerator and check for
any signs of wear or malfunction that may require professional
servicing.
7.2 Equipment Discontinuation
If discontinuing use, follow proper disposal procedures or
contact the manufacturer for guidance on equipment disposal.
7.3 Maintenance, Replacement, and Recovery of Rechargeable
Batteries
If applicable, follow recommended practices for maintaining
rechargeable batteries to ensure optimal performance.
4. Troubleshooting and Maintenance Services
If you encounter any issues with the refrigerator, refer to the
troubleshooting section of the manual or contact authorized service
providers for assistance.
Frequently Asked Questions (FAQ)
Q: Can I adjust the temperature settings of the
refrigerator?
A: Yes, you can adjust the temperature settings based on the
requirements of the stored items. Refer to the user manual for
instructions on temperature control.
Q: How often should I clean the interior of the
refrigerator?
A: It is recommended to clean the interior of the refrigerator
regularly, at least once a month, to maintain a hygienic storage
environment for pharmaceuticals or medical supplies.
Pharmacy/Medical Refrigerator
Operation Manual
The following product models are applicable:
YC-55L YC-55EL YC-56L YC-56EL YC-75L YC-75EL YC-76L YC-76EL YC-130L YC-130EL
YC-315L YC-315EL YC-330L YC-330EL YC-395L YC-395EL YC-400L YC-400EL YC-525L
YC-525EL YC-725L YC-725EL YC-1015L YC-1015EL YC-1320L YC-650L YC-1505L
Zhongke Meiling Cryogenics Company Limited
Better science, Better life!
www.melingbiomedical.com zkmeiling@zkmeiling.com
Table of Contents
1. Application Notes
1
2. Safety Instructions
2
3. Precautions in Use
4
4. Installation and Commissioning
4
4.1 Installation Environment
4
4.2 Installation Site
4
4.3 Preparation before Use
5
4.4 First Power-on
5
4.5 Operation after Power Failure
6
5. Product Composition and Overview
7
6. Operating Instructions
10
6.1 Function Introduction
10
6.2 Optional Function
18
7. Maintenance and Service
19
7.1 Equipment Maintenance
19
7.2 Equipment Discontinuation
19
7.3 Maintenance, Replacement, and Recovery of Rechargeable Batteries
19
8. Troubleshooting and Maintenance Services
20
9. Specifications
21
10.Packing List
26
1. Application Notes
Thank you for choosing and using MELING BIOMEDICAL products! For your safe and
convenient use and reasonable maintenance of the equipment, please read the
Operation Instructions carefully before use and keep it properly for
reference. The equipment operator can copy some chapters of this operation
manual, but only for internal use, for example, to instruct the user how to
deal with emergencies. These chapters are clearly marked in the catalog of the
manual. MELING BIOMEDICAL has no obligation and responsibility for any
instrument damage caused by the user’s failure to use the equipment according
to the instructions or the method specified by the manufacturer. Due to the
rapid improvement of MELING BIOMEDICAL products, the functions described in
the instructions may be inconsistent with those of the products you purchased.
Please refer to the physical functions.
® Please read carefully the Attention and Safety Precautions (in 2. Safety
Instructions). ® During transportation or use, no violent vibration or
collision is allowed and the refrigerator shall
be kept away from rain. Store in a clean room with humidity no more than 80 %,
no corrosive gas and good ventilation. ® Pharmacy/Medical refrigerators
(hereinafter referred to as equipment) can only be operated by trained and
authorized personnel. ® Maintenance of the equipment can only be completed by
MELING BIOMEDICAL or an agent authorized by MELING BIOMEDICAL. ® If the
operator encounters any situation not mentioned in this manual, please contact
MELING BIOMEDICAL or the agent authorized by MELING BIOMEDICAL for the correct
handling method. ® If the equipment is not used according to the method
specified in the specification, it may be damaged. ® Try to use the
accessories provided by MELING BIOMEDICAL. If users would like to use other
accessories, MELING BIOMEDICAL will not be responsible for the adverse
consequences caused therefrom. ® Equipment must be inspected and maintained
regularly to ensure good operation of the equipment. ® Do not use mechanical
devices or other means to accelerate the defrosting process, other than those
recommended by the manufacturer.
Tips
¿ Properly use protective equipment (including protective clothing, protective
gloves, goggles, etc.)
¿ Keep good hygiene habits.
¿ Please pay attention to safety when using this product.
¿ Keep clear of obstruction all ventilation openings in the appliance
enclosure or in the structure for building-in.
¿ Do not use mechanical devices or other means to accelerate the defrosting
process, other than those recommended by the manufacturer.
¿ Do not damage the refrigerating circuit.
¿ Do not use electrical appliances inside the food/ice storage compartments
unless they are of the type.
¿ Recommended by the manufacturer.
1
2. Safety Instructions
When using this product for the first time, please pay attention to the
meaning of the following warning signs and carefully read the safety
precautions, so that you can use this equipment safely and correctly.
Beware of fire.Warning;flammable material
Warning: crushing of hands.
Manufacturer
Series NO
Date of production
Warning: Failure to observe the precautions may result in serious personal
injury or death.
Do not touch the equipment with wet hands to avoid electric shock; Equipment
shall not be grounded through gas pipeline, water supply pipeline, telephone
line or lightning rod, which may easily cause electric shock.
This equipment can only be installed by professional technicians or after-
sales maintenance personnel, it may cause electric shock or fire.
Be sure to install the equipment on a solid and flat ground and take due care
to prevent tipping over. If the ground is not solid enough or the installation
location is not appropriate, it may cause the equipment to fall over and cause
equipment damage or personal injury.
Please handle the power cord carefully to avoid short circuits or open
circuits. Please turn off the power before pulling out the power plug. Hold
the power plug carefully and pull it out. Do not pull the wires of the power
plug. Otherwise, it may cause electric shock or fire due to the short circuit.
Don’t bundle the power cord, don’t press it under furniture or heavy objects,
and don’t get close to heat sources such as compressors. Please insert the
power plug into the outlet tightly to ensure firmness and reliability, so as
to avoid electric leakage; After installation, the power plug must be within
reach, so that the power plug can be unplugged in time in case of emergency.
Separate special outlets must be used and grounded reliably. The cross-
sectional area of the copper conductor in the wall connected with the outlet
must be more than 4mm2. Do not lengthen the power cord without authorization
to avoid heating or fire.
Do not use the power supply that is not specified in the equipment design, so
as to avoid overheating, short circuit and other faults. For example,
connecting 110 V rated voltage products to 220 V power supply may cause faults
such as overheating and equipment burning. For detailed input voltage AC
please refer to Specification(Rated voltage ±10%).If the operating voltage is
too low or too high, a suitable automatic voltage stabilizer must be installed
for cooperative use.
Please place the equipment stably and avoid shaking.
Do not place the equipment in a dangerous area, and do not operate the
equipment near flammable items to prevent explosion or fire accidents.
2
Do not place the equipment in areas exposed to the sun or rain, so as to
prevent danger such as short circuit or overheating.
Do not tilt or lay the equipment sideways, and do not impact the equipment
body; Refrigeration systems are installed in the equipment, which is easy to
be damaged by tilt or impact.
Please place the equipment in a dry and dust-free environment to avoid risks
such as overheating, and short circuit.
In case of unexpected sound, smell, smoke, etc. when the power is turned on,
please unplug the power in time and contact the manufacturer or supplier.
Please place the equipment in a dry and ventilated environment, and ensure
that the equipment vents and instrument surfaces are not blocked or shielded
by walls or other objects; Do not use it in a poorly ventilated environment to
prevent damage caused by heat released by equipment. It is forbidden to
disassemble and modify this equipment without authorization, so as to avoid
potential safety hazards. In this case, MELING BIOMEDICAL will not bear any
responsibility for quality accidents.
It is forbidden to put inflammable and explosive dangerous goods, strong
corrosive acids, alkalis and other items unsuitable for the equipment in the
equipment.
When storing toxic, harmful or radioactive materials, please use the equipment
in safe areas. Improper use may cause harm to human health or environment.
Metal objects such as nails or iron wires shall not be inserted into any
aperture and gap or any outlet of the equipment, otherwise electric shock or
injury may be caused due to accidental contact between the above objects and
moving parts. In order to ensure the normal operation and ventilation and heat
dissipation of the equipment, the back, left and right sides of the cabinet
shall be at least 30cm away from the wall, and the air inlet and air outlet
must not be blocked by obstacles.
This equipment must be connected to a ground wire.
Note: Failure to observe the precautions may result in personal injury or
equipment failure and
related property losses. It is forbidden to store living animals, flowers or
other items with strict temperature requirements in the equipment.
When the equipment is running, do not touch the inner surface of the cabinet
without wearing protective gear.
Hold the handle and close the door to avoid pinching your fingers; When the
equipment is not used for a long time, please unplug it and pack it for
storage.
When restarting the equipment after power failure or power off, please check
the equipment settings first, otherwise the stored items may be damaged due to
the change of settings.
The equipment can be used for item preservation, not as production equipment.
Keep the keys properly, so as to avoid accidents when children open the door
accidentally.
When handling the equipment, please be careful not to tip over the equipment,
so as to prevent equipment damage or personal injury. When handling, it shall
be lifted from the bottom, with the inclined plane not be greater than 45°,
and it shall be handled with care. Please use the equipment in safe areas.
Improper use may cause harm to human health or environment.
3
3. Precautions in Use ® Before putting the items into the equipment, please
confirm that the temperature in the freezer storage chamber has
reached the set value first, and then put the articles in batches. Every time
you put in items, they shall not exceed 1/3 of the inner volume of the freezer
storage chamber so as to prevent excessive temperature rise.
® The equipment temperature display value is the temperature at the
temperature sensor in the freezer storage
chamber. There is a certain gap between the displayed temperature and the
actual temperature at the center of the equipment when the equipment just
starts running, but as the equipment enters a stable state, the displayed
temperature will gradually approach the actual temperature.
® Please use a diluted neutral cleaner to clean the equipment, and do not use
brushes, acid, gasoline, soap powder,
polishing agent or hot water to clean the equipment, otherwise the painted
surface and plastic rubber parts may be damaged. Be careful not to wipe
plastic rubber parts with volatile solvents such as gasoline.
® When the equipment is not used for a long time, the power supply shall be
cut off.
4. Product Installation
4.1 Installation Environment
ò Ambient temperature: 16 ~ 32, the most ideal temperature is 18 ~ 25, and the
air conditioning system shall be used when necessary.
ò Relative humidity: 80%RH. ò There is no strong vibration and corrosive gas
around. ò Avoid the existence of a large amount of dust. ò Avoid rocking or
shaking the equipment. ò Elevation of the working position of the equipment:
less than 2000m. ò Indoor use, pollution degree 2, and overvoltage category
II. ò For detailed input voltage AC please refer to Specification(Rated
voltage ±10%). ò There is no direct sunlight, other cold and heat sources and
strong electromagnetic interference, which will
affect the normal operation of the control system and directly damage the
system in severe cases.
4.2 Installation Site
In order to operate the equipment normally and obtain the best performance
level, the installation site of the equipment shall meet the following
requirements: ò It cannot be installed in a narrow and closed space, and the
door of the room shall not be
smaller or lower than this equipment, which shall at least ensure the normal
access of the equipment, so as to avoid the maintenance difficulties in case
of equipment failure, which may result in damage to stored items due to the
failure to repair the equipment in time; ò The installation floor must be
solid, flat, non-combustible and able to bear the weight of the equipment
during operation; ò It shall be with good ventilation, and direct sunlight
shall be avoided; ò Each equipment needs to use a power outlet independently.
Please ensure that the plug and outlet are firmly connected; ò Check the
working voltage before use. In areas with unstable voltage, consider using a
voltage stabilizer suitable for the equipment load to ensure that the input
voltage requirements in the installation environment are met;
Attention: Since the ambient temperature has great influence on the equipment,
if the above
environmental requirements cannot be met, the equipment may not run normally.
Please improve the environment before using the equipment; The equipment is
operated intermittently.
4
4.3 Preparation before use
1.Remove the outer packaging of all products (including the protective foam in the packaging box).
Warning: Don’t put the plastic bags within the reach of children, so as to prevent suffocation
accidents.
2.Inventory of accessories: Please check the accessories and materials according to the packing list.
3. Cleaning: Clean the product once before use.
4.Before use, remove the small wire shelf and put it at the bottom of the cabinet (applicable to YC-55L/YC-
55EL /YC-56L/YC-56ELand YC-75L/YC-75EL/YC-76L/YC-76EL).
5. Installation of door handles (YC-650L,YC-725L, YC-725EL, YC-1015L,YC- 1015EL,YC-1320L,
YC-1505L).
In order to install handles, the handles shall be placed together with accessories.
Handle installation method:
Firstly, open the door handle as shown in the following figure;
Then, remove the screw for installing the door handle on the cabinet, and install the handle seat with this screw;
Finally, install the handle cover, insert both ends of the handle cover into the corresponding grooves of
the handle seat, and then press it to close it.
6.Install the fixing plates(Apply to YC-725LYC-725EL(110V 60Hz)).
a.Firstly take out the two fixing plates and check if consistent.
b.Remove the bolts from the fixing plate,and then fix the plate on the back of the refrigerator.
c.Connect the fixing plate to the immovable wall or bracket on the back of the refrigerator and finally
check whether it is fixed.
Door handle
Screw
Fixing plate
Mounting hole
Screw Mounting hole
4.4 First Power-on
When using the equipment for the first time, please follow these steps: 1.
After the equipment is placed, leveled and cleaned, it shall stand for more
than 24 hours, and then power it on to
ensure the normal operation of the equipment. 2.Under no-load condition,
connect the power cord to a special outlet with appropriate specifications.
3.After powering on, turn on the power switch of the equipment.Toggle the
on/off battery switch to the ON position
located in the back of the unit. 4.Check whether the operating temperature of
the equipment reaches the required value, observe the normal start
and stop of the equipment for more than 24 hours, and put a small amount of
items in the Refrigerator after confirming the normal performance. 5.Please
store items in batches, with the items not exceeding 1/3 of the cabinet volume
each time. Ensure that the equipment is running properly after shutdown for
more than 12 hours before putting in the next batch of items). 6. Try not to
open the door during cooling, otherwise the temperature will rise.
5
4.5 Operation after Power Failure
1. The equipment has a memory function for the set value. When the power is
restored after power failure, the equipment will continue to operate before
power failure.
2. Once the equipment is powered off, it requires 5 minutes before it can be
switched on again, so as to avoid damaging the compressor.
3. We guarantee the normal operation of this equipment under certain
conditions, but we are not responsible for any loss or damage of stored items
after power failure.
Notes:
® A special person shall be responsible for checking and recording the running
status of the equipment every
day (record and check once every 2-4 hours). In case of failure or shutdown,
the temperature in the Refrigerator will rise. If it cannot be repaired in a
short time, please take out the stored items and transfer them to a place that
meets the temperature requirements for storage to avoid damage to the items.
® Before putting items into the equipment, it shall be confirmed in advance
whether the temperature range of
the equipment meets the temperature requirement of the items, so as to avoid
damage to stored items due to the difference between the settable temperature
of the equipment and the required temperature of the items. Please pay
attention not to block the air outlet and air inlet when putting items into
the equipment.
® Due to the refrigeration inertia, there is a certain difference between the
actual display temperature and the
set temperature of the equipment, which is a normal phenomenon.
® The equipment is an item storage equipment, which cannot be used for routine
production operations. It is
strictly forbidden to put too many items which are relatively hot into the
equipment at one time, otherwise the compressor will run for a long time, and
be burned due to high temperature. Items must be put in batches, so as to
ensure that the Refrigerator is cooled step by step until the temperature
required for storing items is reached.
® Electrical appliances without production license shall not be used inside
the equipment. ® Do not change the set temperature frequently in a short time,
otherwise the expected setting effect may not
be achieved due to the large temperature inertia; Ensure that there is a
certain air circulation space around the cabinet when putting in items,
especially do not block the temperature sensor in the cabinet (for collecting
the temperature of the cabinet), otherwise it will affect the stability and
accurate control of the temperature in the cabinet!
® Items shall not be placed directly at the bottom of the Refrigerator, but on
the bottom shelf, otherwise the
refrigeration effect of the equipment will be affected.
® When putting in items, if the moisture content of the items is too much or
too little, it will affect the humidity
change in the cabinet, so it is best to keep the items sealed; The humidity of
the working environment will affect the change of humidity in the cabinet,
especially if the door is opened too frequently and the door is not closed
properly.
Warning:
® Children are not allowed to play with this equipment as a game prop,
otherwise the injury or loss caused
therefrom will be at their own risk.
® To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective
earth.
® No modification of this equipment is allowed. ® Do not modify this equipment
without authorization of the manufacturer. ® Protection impairment if used in
a manner not specified by the manufacturer. ® The instructions concerning
persons (including children) with reduced physical, sensory or mental
capabilities, or lack of experience and knowledge and children playing with
the appliance are not required.
® Keep all ventilation openings in the enclosure or, in the structure for
building in, clear of obstruction. ® Do not use mechanical devices or other
means to accelerate the defrosting process, other than those
recommended by the manufacturer.
® Do not damage the refrigerant circuit. ® Protection impairment if used in a
manner not specified by the manufacturer.
6
5. Product composition and overview
Display panel Door lock
USB interface Door body
Adjust feet
Figure I YC-55L (It can be used as reference for YC-55EL, YC-56L,YC-56EL,YC-
75L,YC-76L, YC-75EL, YC-76EL) The YC-55L/YC-56L/YC-55EL/YC-56EL refrigerator
has two shelves, each with a capacity of 13Kg. The YC-75L/YC-76L/YC-75EL/YC-
76EL refrigerator has three shelves, each with a capacity of 13Kg.
Door lock
USB interface
Display panel
Graph recorder (optional)
Door handle Cabinet
Door body
Adjust feet Caster
Figure II YC-130L (It can be used as reference for YC-130EL) The YC-130L/ YC-
130ELrefrigerator has 2 shelves, the carrying capacity of each shelf and
suspension frame is 21Kg.
7
access port
Model
Shelves
Loading capacity (kg)
YC-315L
4+1
25
Caster
Adjust feet
YC-330L
5
20
YC-395L YC-395EL
6+1
25
YC-400L YC-400EL
6
22
YC-525L YC-525EL
6
32
Figure III YC-395L(It can be used as reference for YC-315L,YC-315EL,YC-330L ,YC-330EL, YC-395EL, YC-400L, YC-400EL and YC-525L,YC-525EL)
Door body Door handle
Model YC-650L
Shelves
Loading capacity (kg)
5
36
Caster
Figure VI YC-650L
8
Door lock
Display panel (power switch on the back)
USB interface
access port
Door handle
Cabinet
Door body
Model
YC-725L YC-725EL
YC-1015L YC-1015EL
Shelves 12 12
Loading capacity (kg) 21
27.5
YC-1320L
10
35
Caster
Figure V YC-725L((It can be used as reference forYC-725EL,YC-1015L,YC-1015EL
,YC-1320L)
access port
Door handle
Cabinet
Door body
Model YC-1505L
Shelves 18
Loading capacity (kg)
32
Caster
Figure IV YC-1505L
Except for the metal shell at the bottom, the contact time of the external accessible parts is t1 min, while the contact time of the metal at the bottom is 1 s t < 10 s.
- Due to the improvement of products and model differences, the actual products may be different from the diagram. Please refer to the actual products! The diagram is only used for functional parts description.
Intended use: * Structure and composition: The product consists of cabinet, door (glass door structure or foam door structure),
refrigeration system and control system. - Scope of application: It is suitable for storing items in hospitals, pharmacies, epidemic prevention stations, research institutions, biopharmaceuticals and other units.
9
6. Operating Instructions
6.1 Function Introduction
This series of products include two control systems, based on which different
adjustment methods apply. Please select the proper adjustment method according
to the type of control system with the storage box you purchased.
Printer (optional)
Type A control panel 1. Function description of Type A control panel (Except YC-130L,YC-130EL)
- Description of display icon indicator (as shown in the right diagram);
Door switch
Power failure
WIFI Print
Defrosting Key lock
Door heating
Refrigeration
a. Door switch indicator
Mute
Electric Serial quantity port
Fan
When the door is opened, the door switch indicator is on; When the door is closed, the door switch indicator is off.
b. WIFI indicator
WIFI Pilot Lamp is on when WIFI is connected; it’s off when WIFI is disconnected.
c.Defrosting indicator
When the refrigerator enters the defrosting state, the defrosting indicator is always on; When the refrigerator exits the defrosting state, the defrosting indicator is always off.
d. Key lock indicator When the keys are in lock state, none of the keys
respond, and the key lock lights up. In this case, after pressing the up
key+down key for 3s, you will be prompted to enter the password, which is
defaulted as “005.” After entering it correctly, press the Set/Mute Multiplex
key, and the key lock will be released. At this time, the key lock indicator
will be off. In the unlock state, if no key is pressed for 60s, the key lock
is started, and the key lock indicator is on. Long press the up key + down key
for 3s, and the keys are locked.
e.Power-off indicator
The refrigerator is normally powered by 220V/110V. When the input power is
turned off, the buzzer will be triggered, the digital tube flashes the power-
off code “PF” alternately at 3s intervals, and the power-off indicator is on.
When the input power is turned on, it returns to normal, and the power-off
indicator is off.
f.Print indicator
When the printer is not working, the indicator is off; When the printer is working, the indicator is on. g.Door heating indicator
When the door heating is turned off, the door heating indicator is off; When
the door heating is turned on, the door heating indicator is on.
h.Refrigeration indicator
If the compressor is in working condition, the refrigeration indicator is on;
If the compressor is in a shutdown state, the cooling indicator is off.
i.Mute indicator
When the alarm tone key is muted, the indicator is on; When the alarm mute
function is canceled, the indicator is off.
j.Low battery level indicator
When the battery voltage is less than 8V, the buzzer will be triggered, the
low battery indicator will be on, and the digital tube will flash the low
battery code “BL” alternately at intervals of 3s; When the battery voltage is
greater than 12V, the buzzer will be turned off, the low battery indicator
will be off, and the digital tube will resume normal display.
k.Serial port indicator
When the reserved RS-485 serial port is not connected to the equipment, the serial port indicator will be off; When the equipment is successfully connected to the reserved RS-485 serial port, the serial port indicator will be on.
l.Fan indicator
When the evaporator fan is turned on, the fan indicator is on; when the evaporator fan is turned off, the fan indicator is off.
10
temperature display window, which displays the average temperature inside the cabinet in °C under normal
operation;
View ambient temperature:
In the key lock state, press key, and digital tube displays the ambient temperature, and returns to normal
display after 5 s without key operation, or pressing and . In the key unlock state, press key, and digital
tube displays ambient temperature, and returns to normal display after 5s without key operation.
Humidity check: key unlocked state, long press and , digital tube display humidity, no press any Key
operation after 5 seconds or press and , return to normal display.
- is the set/mute key;
In case of no alarm state and key unlock state, press , and display the ambient temperature for 5s and then
return to normal display; In the unlock state, press for more than 3s, and enter the user menu.
When buzzer is triggered (including cabinet high temperature alarm, door opening alarm, sensor failure alarm, etc.)
and in the key unlock state, press for the first time, and the buzzer stops ringing, and the ambient temperature
is displayed for 5s, after which the normal display is resumed (pressing the mute button is only to turn off the buzzer
for alarming this abnormal state, for example troubleshooting, and the buzzer will be triggered next time for any
abnormality). Then press again, trigger the buzzer, display the ambient temperature for 5s, and resume the
display of chamber temperature and alarm state. In the key unlock state, can be used as a setting key.
In unlock state and parameter setting mode, press this key to display parameter values and parameter names. If the
pressing time is longer than 3 seconds, save the settings and return to the normal interface.
- is a up key;
In parameter setting mode, move to the next parameter or increase the parameter value. For example, when setting
the set temperature, increase the set temperature value. When setting the parameter value, long press the up
button, and the parameter will increase rapidly. Under normal conditions, long press the up key for 3 seconds to
import the data of the USB flash drive in 12 months.
- is a down key;
In parameter setting mode, move to the previous parameter or decrease the parameter value.
For example, when setting the set temperature, reduce the set temperature.
When setting the parameter value, long press the down button, and the parameter will decrease rapidly.
- is a print key;
The system can keep 7 days of data for printing, and press the print key to print the temperature within the set time
period.
- is a light switch key;
When the equipment is powered on, the lights are turned off by default.
- USB data export;
Automatic export: when the U disk is connected to the USB interface, the recorder buzzer will chirp once and
display “on”. PDF files of data that not currently exported will be generated in the U disk. After data transmission,
the buzzer will chirp once again and display “End”.After 6s, it will return to normal display.
Note: When there is less data, the digital tube will not display “on” and “End”.
Manual export: In the key unlock state, and when the USB flash drive is connected and the file is not being
generated, press the key up for 3 seconds, and the digital tube will display “d01.” Press the up key or down key to
adjust “d00~d12,” and press key to obtain the file generation (d00) or generate the PDF file of the record data
of the previous months (1-12).
11
2. Function setting of Type A control panel
-
After powering on, the equipment can enter the working state;
-
User parameter settings:
Unlock:under normal operating state,simultaneously press and keys for 3s and the digital tube will
display the parameter code”0000″;By pressing to enter the password”0005″and holding to unlock.
Then press key for 3s,the digital tube will display the parameter code”PS1″and enter the setting and
adjustment parameters.Use or key to scroll the parameters;
a. Use or key to scroll the parameters;
b. Press key to display the corresponding parameter value;
c. Use or key to scroll the parameters;
d. Use to temporarily store the modified values and return to the display parameters;
e. If other parameters are modified, repeat steps to
f. Press
for more than 3s, save the modified parameters and return to the display parameter category.
- Press
for more than 3s, or press no key in 60s to exit the parameter setting program.
4)Parameter display
No. Menuitem
Parameter Range
Suggested settings
Remarks
1
MAX
2
MIN
3
CLR
4
Set
5
H
_ 0.0-10.0
0.0-10.0
6
L
0.0-10.0
7
n
Set logger module time – year
8
y
Set logger module time – month
9
r
Set logger module time – day
10
s
Set logger module time – hour
11
F
Set logger module time – minute
12
Pt
13
tH1
0-240min 20.0-50.0
5.0
5.0
5.0 20 40.0
The highest temperature since last clerance
The lowest temperature since last clerance
Clearance of the Max and Min temperature records
Temperature setting
Set value of high temperature alarm set+H; When H =0, High temp alarm is
disabled; When the alarm is over high temp alarm set, H1 will be displayed on
the controller
Set value of low temperature alarm set-L; When L =0, Low temp alarm is
disabled; When the alarm is below low temp alarm set, L1 will be displayed on
the controller
Print interval
Upper limit of ambient temperature alarm
12
No. Menuitem
Parameter Range
Suggested settings
Remarks
14
P1
1.Automatic heating mode 1 2.Automatic heating mode 2 3.Automatic heating mode
3
4.Always on 5. Always off
1.(Set to 4 when
the door is with condensation) (Set to 1 when the door is a
glass door) (Set to 5 when the door is a foaming door)
Mode 1: It shall be judged as once after the door is opened and closed once and heated for 5min (time setting). If the door is opened and closed again during the heating period, the heating time will be updated again; Mode 2: When the compressor runs, heater is on;When the compressor stops, the heater Will automatically off after one min Mode 3: When the humidity in the cabinet is more than 80%, the door heating is on, and when the humidity in the cabinet is moderately less than 60%, the door heating is off; Mode 4: Door heating is always on; Mode 5: Door heating is always off.
15
Ps1
16
b1
17
b2
0000-9999
0005
User menu password settings Repair Information 1 Repair Information 2
Quick setting of time after power on After the power-on self-test on the display board is completed, the quick setting menu is displayed.
Menu item Menu
Menu description
Set range
Default
Unit
n Set logger module time – year
10~50
/
y Set logger module time – month
1~12
—
/
r Set logger module time – day
01~31
—
/
Quick Settings
menu
S Set logger module time – hour F Set logger module time – minute Pt Print interval
00~23 00~59 0~240
—
/
—
/
20
min
0~240
SCY Temperature data recording period
0: shielded
10
min
recorder
If there is no operation for 60 seconds under the quick setting menu, it will automatically exit the quick setting menu and return to normal display.
4. Alarm display
Code
Error Description
H1
High temperature alarm
L1
Low temperature alarm
H2
Alarm for high ambient temperature
H3
Condenser overheat alarm
do
Door opening alarm
PF
Power failure alarm
bL
Battery low alarm
Er
The recorder is not connected
LoF
Recorder did not start
EE
Communication failure
13
a
b
c
°C
Ref.
DOOR ALARM
USB
TEMPERATURE CONTROL PRINT
i (optional)
h g
f
e d (optional)
Type B control panel 1. Function description of Type B control panel (Applicable to YC-130L YC-130EL)
a.
It is temperature display window, which displays the average temperature inside the cabinet in °C under
normal operation; Different prompt characters can be displayed in the setting state (see below for details). b.Door opening indicator: When the refrigerator door is opened, the indicator lights up. After more than 1 minute,
the door opening alarm indicator will be on and the buzzer triggered,
displaying “do”. c.Fault indicator: When the product runs normally, the
indicator is off; The indicator is on for operation abnormality.
d.”LIGHT” : After the machine is powered on, the light is off by default, and
the on-off of the light can be adjusted by
the on-off key. e. “PRINT ” (optional): The system can keep 7 days of data for printing. Press the print key to print the temperature
within the set time period. f. : In parameter setting mode, reduce the parameter value. For example, when setting the set temperature,
reduce the set temperature. When setting the parameter value, long press the down button, and the parameter will
decrease rapidly. g. : In parameter setting mode, increase the parameter value. For example, when setting the set temperature,
increase the set temperature value. When setting the parameter value, long press the up button, and the parameter
will increase rapidly. Under normal conditions, long press the up key for 3 seconds to import the data of the USB
flash drive in 12 months. h. is the set/mute key; In case of no alarm state and key unlock state, press
, and display the ambient
temperature for 5s and then return to normal display; In the unlock state, press for more than 3s, and enter the
user menu. When buzzer is triggered (including cabinet high alarm, door opening alarm, sensor failure alarm, etc.) and in the
button unlock state, press for the first time, and the buzzer stops ringing, and the ambient temperature is
displayed for 5s, after which the normal display is resumed (pressing the mute button is only to turn off the buzzer
for alarming this abnormal state, for example trouble removal, and the buzzer will be triggered next time for any
abnormality). Then press again, trigger the buzzer, display the ambient temperature for 5s, and then resume the
display of cabinet temperature and alarm state. In the key unlock state, can be used as a setting key.
14
In unlock state and parameter setting mode, press this key to display
parameter values and parameter names. If the pressing time is longer than 3
seconds, save the settings and return to the normal interface. i. USB
interface Automatic export of USB data: When the USB interface is connected to
the USB flash drive, the buzzer of the recorder beeps once, displaying “ON”,
and PDF files of the data of the current month and the previous month are
generated in the USB flash drive. After the data transmission is completed,
the buzzer beeps once, displaying “End,” and the normal display resumes after
6s.
Note: When there is less data, “on” and “end” prompts are not displayed.
Manual export of USB data: In the key unlock state, when the USB flash drive
is connected and the file is not being generated, press the up key for 3
seconds, and the digital tube in the lower chamber will display “d01.” Press
the key up or down key to adjust “d00~d12,” and press key to cancel the file
generation (d00) or generate the PDF file of the record data of the previous
months (1-12).
Note: When the alarm of the digital tube flashes and displays “LoF,” the
recorder is not started;
Meanwhile press and key for 3sand “LoF” disappears, the recorder is started.
2. Function setting of Type A control panel
-
After powering on, the equipment can enter the working state;
-
User parameter settings:
Unlock:under normal operating state,simultaneously press and keys for 3s and the digital tube will
display the parameter code”0000″;By pressing to enter the password”0005″and holding to unlock.
Then press key for 3s,the digital tube will display the parameter code”PS1″and enter the setting and
adjustment parameters.Use or key to scroll the parameters;
a. Use or key to scroll the parameters;
b. Press key to display the corresponding parameter value;
c. Use or key to scroll the parameters;
d. Use to temporarily store the modified values and return to the display parameters;
e. If other parameters are modified, repeat steps to
3)Press
for more than 3s, save the modified parameters and return to the display parameter category.
- Press
for more than 3s, or press no key in 60s to exit the parameter setting program.
15
3. Parameter display
No. Menuitem
Parameter Range
Suggested settings
Remarks
1
MAX
2
MIN
3
CLR
4
Set
5
H
_ 0.0-10.0
0.0-10.0
6
L
0.0-10.0
7
n
Set logger module time – year
8
y
Set logger module time – month
9
r
Set logger module time – day
10
s
Set logger module time – hour
11
F
Set logger module time – minute
12
Pt
13
tH1
0-240min 20.0-50.0
5.0
5.0
5.0 20 40.0
The highest temperature since last clerance
The lowest temperature since last clerance
Clearance of the Max and Min temperature records
Temperature setting
Set value of high temperature alarm set+H; When H =0, High temp alarm is
disabled; When the alarm is over high temp alarm set, H1 will be displayed on
the controller
Set value of low temperature alarm set-L; When L =0, Low temp alarm is
disabled; When the alarm is below low temp alarm set, L1 will be displayed on
the controller
Print interval
Upper limit of ambient temperature alarm
14
P1
15
Ps1
16
b1
1.Automatic heating mode 1 2.Automatic heating mode 2 3.Automatic heating mode
3
4.Always on 5. Always off
1.(Set to 4 when
the door is with condensation) (Set to 1 when the door is a
glass door) (Set to 5 when the door is a foaming door)
Mode 1: It shall be judged as once after the door is opened and closed once and heated for 5min (time setting). If the door is opened and closed again during the heating period, the heating time will be updated again; Mode 2: When the compressor runs, heater is on;When the compressor stops, the heater Will automatically off after one min Mode 3: When the humidity in the cabinet is more than 80%, the door heating is on, and when the humidity in the cabinet is moderately less than 60%, the door heating is off; Mode 4: Door heating is always on; Mode 5: Door heating is always off.
0000-9999
0005
User menu password settings
_
_
Repair Information 1
17
b2
_
_
Repair Information 2
16
Quick setting of time after power on After the power-on self-test on the display board is completed, the quick setting menu is displayed.
Menu item Menu
Menu description
Set range
Default
Unit
Quick Settings
menu
n Set logger module time – year y Set logger module time – month r Set logger module time – day S Set logger module time – hour F Set logger module time – minute Pt Print interval
10~50
/
1~12
—
/
01~31
—
/
00~23
—
/
00~59
—
/
0~240
20
min
0~240
SCY Temperature data recording period
0: shielded
10
min
recorder
If there is no operation for 60 seconds under the quick setting menu, it will automatically exit the quick setting menu and return to normal display.
4. Alarm display Alarm Code H1 L1 H2 H3 do PF bL Er LoF EE
Error Description High temperature alarm Low temperature alarm High ambient
temperature alarm Condenser overheat alarm
Door opening alarm Power failure alarm Battery low alarm The recorder is not
connected Recorder did not start Communication failure
17
6.2 Optional function
6.2.1 Printer
The printer has been installed with a paper roll at the factory. If the paper roll is used up after a long period of use, you can buy the same paper roll (size: thermal paper, paper width: 57.5±0.5mm, the outer diameter of the reel: not greater than 40mm, i.e., RM57*40 back-roll paper)
Description of printer panel:
Open button. Press to open the cover;
SEL button, indicator light, for factory setting. Do not press it;
Lf button. The green indicator is the power indicator and it is normally on when the power is turned on;
Paper roll
Install the paper roll:
Press the open button 1, open the cover, install the paper roll, close the cover plate and allow the head of the roll paper to slightly extend out of the cover plate.
6.2.2 Remote alarm terminal, RS485 interface and network interface
Remote alarm terminal, RS485 interface and network interface are installed in
the lower part of the back of the freezer body. Note: These configurations are
standard for some models.
N.O. COM. N.C.
Normally Common Normally open terminal closed
Remote alarm terminal
RS-485
6.2.3 Chart recorder
It is an optional function. If the function is selected, please refer to the
“Operation Instructions for Graph Recorder” in the accessory bag.
Chart recorder 18
7. Maintenance and Service
Warning:
® In order to prevent people from getting electric shock or injury, please cut
off the power supply of the equipment before any repairs and maintenance.
® Make sure you don’t inhale drugs or suspended particles around it when
maintaining the equipment, otherwise it will harm your health.
7.1 Equipment Maintenance
Defrosting: The equipment will automatically defrost during the working
process for the convenience of use. Cleaning and maintenance: The equipment
should be cleaned and maintained regularly (for safety, please unplug the
power plug), and the inner and outer surfaces of the cabinet should be wiped
with a warm, damp soft cloth. condenser cleaning : clean the dust on the
condenser regularly to avoid affecting the cooling effect (cleaning method:
first open the screws of the ventilation hood or front hood and use a brush to
gently brush away the dust on the surface of the condenser).
Notes:
® Do not sprinkle water directly on the cabinet, lest the insulation
performance of electrical components decrease and metal parts rust.
® Do not use hot water, corrosive detergent or organic solvent to clean the
cabinet.
® Do not place heavy objects on top of the equipment, as the equipment may
deform under pressure.
® The integrated overload protecter within our products is programmed to shut
down the power supply automatically with respect to short circuit and/or
system overload for the sake of protecting the entire system and crucial
components such as the compressor.
7.2 Equipment Discontinuation
Deactivation: If the equipment is stored in an unsupervised area for a long
time and not used, the power supply should be cut off and the inner and outer
surfaces of the cabinet should be cleaned with a warm, damp soft cloth, aired
and sealed. The equipment must be locked to ensure that children cannot open
the cabinet door. Scrapping: When the equipment reaches the end of its service
life, it should be scrapped and must be handed over to a qualified
professional recycling agency for disposal per local regulations. Non-
professionals are not allowed to disassemble and break down the equipment
without authorization. The scrapped equipment should be placed in a designated
area inaccessible to children to avoid danger.
7.3 Maintenance, Replacement, and Recovery of Rechargeable Batteries
Battery installation position: top of the cabinet and bottom inside the
electrical box. 1. Battery maintenance In order to prolong the service life of
the battery and avoid the product being left unused for a long time, the
product must be operated for more than 24 hours every month to facilitate
charging; The equipment should be connected to power supply regularly
(generally once a month) and turn on the power switch and battery switch for a
period of time to complete the charging if it’s not used for a long time,and
each charging time should not be less than 24 hours; Power switch and battery
switch should be turned off in time after discontinuation of equipment,
otherwise it will cause battery capacity loss and in serious cases battery
permanent damage could happen; The main power switch must be turned off during
a long-term power outage or during transportation, otherwise longterm
discharge causes power loss in battery and even permanent damage to the
battery, and the display is abnormal after re-energizing; The battery is
consumable, with a service life of about 2-3 years. If the battery is used
improperly, such as power loss or reaching the end of battery life, it will
cause a low battery alarm (refrigeration is not affected, but there is an
alarm failure and influence to the use function of the printer, please contact
the after-sales staff of local distributors for replacement). 2. Battery
replacement and recovery Turn off the power switch and pull out the power plug
from the outlet; Remove the screws on the electrical box with a screwdriver.
(Note: There are high-voltage electrical components in the electrical box,
therefore before opening, turn off the power supply, unplug the power plug and
turn off the power lock switch of the refrigerator, and the electrical box can
only be opened by qualified engineers or maintenance personnel); Pull out the
battery connecting cable; (Before unplugging the battery fixing cable, pay
attention to the cathode/anode of the battery and the sequence of the
connecting cables, so as to prevent burning down the control system due to
installation of the new battery with cathode/anode reversed. Generally, the
red wire is connected to the cathode, and the black wire is connected to the
anode; Use a screwdriver to remove the fastening screws on the battery fixing
plate and take out the battery; For a replaced battery that can be recycled,
please contact the local battery recycling agency. Tips:
In order to effectively ensure that the battery replacement meets the
requirements of the control system and avoid the influence of improper
operation on the system, please contact our after-sales personnel for
replacement or guidance.
When the circuit is short circuit or overloaded, the overload protector will
disconnect the power supply of the device. Please contact after sales at this
time.
19
8. Troubleshooting and Maintenance Services
Any product may fail. Please observe the operation of equipment in time during
use. If there is any abnormality, please check and handle it according to the
following table first. If the abnormality can’t be changed, please inform our
service center in time, and we will serve you wholeheartedly to avoid losses.
Problems
Equipment does not work
Compressor is not running
The temperature does not reach the
set value
High noise
Condensation on cabinet surface
Undesirable odor
The door is not closed properly, and
the cool air leaks
The alarm lamp flashes and the buzzer is triggered Lighting damage
(refrigeration
equipment)
Causes and solutions
Please make sure that the outlet is energized. Please make sure power plug is
plugged in,not loose. Please make sure the power fuse is not disconnected.
Please make sure the supply voltage is appropriate,not too low or too high.
Please make sure that the temperature is set correctly. Please check whether
the temperature inside the cabinet is too low.
Please make sure that the door is closed tightly and don’t open it too many
times during a short time. Please don’t put too many items in at one time.
Please make sure that the ambinet temperature is not too high.
Please make sure that the cabinet is placed on a flat ground. Please make sure
that the cabinet does not contact the wall.
In rainy and humid seasons, door condensation is normal, and it shall be wiped
off with a dry cloth.
Equipment needs to be cleaned. Items with heavy smell are not packed.
Please make sure that the temperature is set correctly. Please check whether
the temperature inside the cabinet is too low.
After the equipment is used for a period of time, the door seal becomes hard
and deformed. Maintenance method: Blow the deformed part of the hot seal with
a blower to soften it, and then close and compress it after the door seal
becomes soft.
Please call the after-sales telephone and contact after-sales service
personnel of MELING BIOMEDICAL for replacement. Please do not replace the
parts by yourself.
The following conditions are not faults When the compressor starts and stops,
the equipment parts will make a slight impact sound; When overheated items are
put in after the door is opened, high temperature alarm and high humidity
alarm will be triggered in the control system (if this function is available,
please refer to the alarm display table); Solution: Put the items into the
equipment after they are naturally cooled. Put the items in small quantities
in batches, and do not put too much at a time. After the system runs stably,
the high temperature alarm and high humidity alarm will be released
automatically. Slight sound of running water caused by refrigerant flowing in
the pipeline.
Warning:
¿ The system contains refrigerant under high pressure. Do not tamper with the
system. It must be serviced by suitably qualified persons only.
¿ Connect to potable water supply only. ¿ Do not store explosive substances
such as aerosol cans with a flammable propellant in this appliance. ¿ If
symbol ISO 7000-1701 (2004-01) is used, its meaning shall be explained. ¿ In
order to reduce flammability hazards the installation of this appliance must
only be carried out by a suitably
qualified person. ¿ The refrigeration system is under high pressure. Do not
tamper with it. Contact qualified service personal before
disposal.
20
Notes:
¿ The equipment can only be repaired, maintained or improved by the engineers
certified by MELING BIOMEDICAL, so as to ensure the normal operation of the
equipment and the compliance with corresponding safety standards.
¿ Please clean and disinfect the equipment before notifying the maintenance
engineer; During the warranty period of the equipment, the Company will not
undertake the warranty obligation if the fault or damage is caused by improper
use of the user.
¿ Ambient temperature for storage: -40°C-+55°C, relative humidity: 10% ~ 90%.
9. Specifications
Model
Ambient Climate
temperature type (°C)
Refrigerant and loading
amount
(Rated) voltage
(V~)
Rated frequency
(Hz)
YC-55L Spraying aluminum liner
16 ~32
N R600a/16g 220-240
60
YC-75L Spraying aluminum liner
16 ~32
N R600a/18g 220-240
60
YC-330L Spraying aluminum liner
16 ~32
N R600a/50g 220-240
60
YC-330L Stainless steel liner
16 ~32
N R600a/50g 220-240
60
Temp Range (
2~8
2~8
2~8
2~8
Volume (L)
Rated current
(A)
Weight (kg)
Exterior dimensions
(DxWxH) (mm)
55
0.56 35 560X540X632
75
0.56 41 560X540X764
330 0.85 76 592X620X1937
330 0.85 81 592X620X1937
21
Model
Ambient temperature
Climate type
Refrigerant Rated
andloading voltage
amount
(V~)
Rated frequency
(Hz)
Temp Range
(°C)
YC-55L Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
YC-55L Stainless steel liner
16 ~32
N
R600a/ 18g
110
60
2~8
YC-55EL Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
YC-55EL Stainless steel liner
16 ~32
N
R600a/ 18g
110
60
2~8
YC-56L Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
Volume (L)
Rated current (A)
Weight (kg)
Exterior dimensions
(DxWxH) (mm)
55
1.16
35 560X540X632
55
1.16
38 560X540X632
55
0.79
32 560X540X632
55
0.79
35 560X540X632
56
1.16
35 565X542X632
YC-56L Stainless steel liner
16 ~32
N
R600a/ 18g
110
60
2~8
56
1.16
38 565X542X632
YC-56EL Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
56
0.79
32 565X542X632
YC-56EL Stainless steelliner
16 ~32
N
R600a/ 18g
110
60
2~8
56
0.79
35 565X542X632
YC-75L Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
75
1.24
41 560X540X764
YC-75L Stainless steelliner
16 ~32
N
R600a/ 18g
110
60
2~8
75
1.24
45 560X540X764
YC-75EL Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
75
0.85
35 560X540X764
YC-75EL Stainless steel liner
16 ~32
N
R600a/ 18g
110
60
2~8
75
0.85
39 560X540X764
YC-76L Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
76
1.24
41 566X542X764
YC-76L Stainless steel liner
16 ~32
N
R600a/ 18g
110
60
2~8
76
1.24
45 566X542X764
YC-76EL Spraying aluminum liner
16 ~32
N
R600a/ 18g
110
60
2~8
76
0.85
35 566X542X764
YC-76EL Stainless steelliner
16 ~32
N
R600a/ 18g
110
60
2~8
76
0.85
39 566X542X764
YC-130L Spraying aluminum liner
16 ~32
N
R290/ 26g
110
60
2~8
130
2.28
51 625X650X810
YC-130L HIPS
16 ~32
N
R290/ 26g
110
60
2~8
130
2.28
50 625X650X810
YC-130EL Spraying aluminum liner
16 ~32
N
R290/ 26g
110
60
2~8
130
1.66
45 625X650X810
YC-130EL HIPS
16 ~32
N R290/26g 110
60
2~8
130
1.66
44 625X650X810
YC-315L YC-315EL
16 ~32
N R290/60g 110
60
2~8
315
3.83
87 673X650X1762
16 ~32
N R290/60g 110
60
2~8
315
3.45
83 673X650X1762
YC-395L Compressor A
16 ~32
N R290/60g 110
60
2~8
395
3.77
95 673X650X1992
YC-395L Compressor B
16 ~32
N R290/60g 110
60
2~8
395
4.76
95 673X650X1992
YC-395EL Compressor A
16 ~32
N R290/60g 110
60
2~8
395
3.77
93 652X650X1992
YC-395EL Compressor B
16 ~32
N R290/60g 110
60
2~8
395
4.76
93 652X650X1992
YC-400L
16 ~32
N R290/55g 110
60
2~8
400
4.74
116 645X700X2010
YC-400EL
16 ~32
N R290/55g 110
60
2~8
400
3.52
107 645X700X2010
22
Model
Ambient temperature
Climat type
e
Refrigerant andloading
amount
Rated voltage (V~)
Rated frequency
(Hz)
Temp Range
(°C)
YC-525L Spraying aluminum liner
16 ~32
N R290/75g 110
60
2~8
YC-525L Stainless steel liner
16 ~32
N R290/75g 110
60
2~8
Volume (L)
Rated current (A)
Weight (kg)
Exterior dimensions
(DxWxH) (mm)
525
5.1
141 810X720X1973
525
5.1
148 810X720X1973
YC-525EL Spraying aluminum liner
16 ~32
N R290/75g 110
60
2~8
525
3.84
127 810X720X1973
YC-525EL Stainless steel liner
16 ~32
N R290/75g 110
60
2~8
525
3.84
134 810X720X1973
YC-650L
Spraying aluminum liner 16 ~32 Compressor A
N
R290/ 110g
110
60
2~8
650
5.77
142 890X715X1985
YC-650L
Stainless steel liner Compressor A
16 ~32
N
R290/ 110g
110
60
2~8
650
5.77
157 890X715X1985
YC-650L
Spraying aluminum liner Compressor B
16 ~32
N
R290/ 80g
110
60
2~8
650
5.77
142 890X715X1985
YC-650L
Stainless steel liner C o m p r e s s o r B
16 ~32
N
R290/ 80g
110
60
2~8
650
5.77
157 890X715X1985
YC-725L Spraying aluminum liner
16 ~32
N
R134a/ 245g
110
60
2~8
725
8
171 718X1093X1992
YC-725L Stainless steel liner
16 ~32
N
R134a/ 245g
110
60
2~8
725
8
189 718X1093X1992
YC-725EL Spraying aluminum liner
16 ~32
N
R134a/ 245g
110
60
2~8
725
6.92
161 718X1093X1992
YC-725EL Stainless steel liner
16 ~32
N
R134a/ 245g
110
60
2~8
725
6.92
179 718X1093X1992
YC-1015L Spraying aluminum liner
16 ~32
N
R134a/ 280g
110
60
2~8
1015
8.2
185 852X1180X1990
YC-1015L Stainless steel liner
16 ~32
N
R134a/ 280g
110
60
2~8
1015
8.2
223 852X1180X1990
YC-1015EL Spraying aluminum liner
16 ~32
N
R134a/ 280g
110
60
2~8
1015
7.09
173 852X1180X1990
YC-1015EL Stainless steel liner
16 ~32
N
R134a/ 280g
110
60
2~8
1015
7.09
202 852X1180X1990
YC-1320L Spraying aluminum liner
16 ~32
N
R290/ 150g
110
60
2~8
1320
10.55
235 826X1453X1998
YC-1320L Stainless steel liner
16 ~32
N
R290/ 150g
110
60
2~8
1320
10.55
258 826X1453X1998
23
Model
Ambient temperatur
(°C)
e
Climat type
e
Refrigerant and loading
amount
Rated) voltage
(V~)
Rated frequency
(Hz)
Temp Range
(°C)
Volume (L)
Rated
current A
Weight (kg)
Exterior dimensions
(DxWxH) (mm)
YC-55L YC-55EL Spraying aluminum liner
YC-55L YC-55EL Stainless steel liner YC-56L Spraying aluminum liner
YC-56EL Spraying aluminum liner
YC-56L Stainless steel liner
YC-56EL Stainless steel liner
YC-75L YC-75EL Spraying aluminum liner
YC-75L YC-75EL Stainless steel liner
YC-76L Spraying aluminum liner
YC-76EL Spraying aluminum liner
YC-76L Stainless steel liner
YC-76EL Stainless steel liner
16 ~32
16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32
16 ~32 16 ~32 16 ~32 16 ~32 16 ~32
N R600a/16g 220-240
50
N R600a/16g 220-240
50
N R600a/16g 220-240
50
N R600a/16g 220-240
50
N R600a/16g 220-240
50
N R600a/16g 220-240
50
N R600a/18g 220-240
50
N R600a/18g 220-240
50
N R600a/18g 220-240
50
N R600a/18g 220-240
50
N R600a/18g 220-240
50
N R600a/18g 220-240
50
2~8
55
2~8
55
2~8
56
2~8
56
2~8
56
2~8
56
2~8
75
2~8
75
2~8
76
2~8
76
2~8
76
2~8
76
0.9
35 560X540X632
0.9 0.9 0.9 0.9 0.9 0.92
38 560X540X632 35 565X542X632 32 565X542X632 38 565X542X632 35 565X542X632 41 560X540X764
0.92
0.92 0.92 0.92 0.92
45 560X540X764
41 566X542X764 35 566X542X764 45 566X542X764 39 566X542X764
YC-130L YC-130EL Spraying aluminum liner
16 ~32
N R600a/55g 220-240
50
2~8
130
0.98
51 625X650X810
YC-130L YC-130EL Stainless steel liner
16 ~32
N R600a/55g 220-240
50
2~8
130
0.98
57 625X650X810
YC-315L HIPS
YC-315EL HIPS YC-330L YC-330EL Spraying aluminum liner
YC-330L YC-330EL Stainless steel liner
YC-395L HIPS
YC-395EL HIPS
YC-525L YC-525EL Spraying aluminum liner
YC-525L YC-525EL Stainless steel liner
16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32
16 ~32
N R600a/22g 220-240
50
N R600a/22g 220-240
50
N R600a/50g 220-240
50
N R600a/50g 220-240
50
N R600a/25g 220-240
50
N R600a/25g 220-240
50
N R290/70g 220-240
50
N R290/70g 220-240
50
2~8
315
1.35
87 652X650X1762
2~8
315
1.35
87 652X650X1762
2~8
330
1.31
76 592X620X1937
2~8
330
1.31
81 592X620X1937
2~8
395
1.8
95 673X650X1992
2~8
395
1.8
95 652X650X1992
2~8
525
2.49
141 810X720X1973
2~8
525
2.49
148 810X720X1973
YC-725L YC-725EL Spraying aluminum liner
16 ~32
N R290/85g 220-240
50
2~8
725
3.90
171 718X1093X1992
YC-725L YC-725EL Stainless steel liner
16 ~32
YC-1015L Stainless steel liner
YC-1015L Spraying aluminum liner
16 ~32 16 ~32
N R290/85g 220-240
50
N R290/90g 220-230
50
N R290/90g 220-230
50
2~8
725
3.90
189 718X1093X1992
2~8
1015
3.21
223 852X1180X1990
2~8
1015
3.21
185 852X1180X1990
24
Model
Ambient
temperatur °C
e
Climate type
Refrigerant andloading
amount
(Rated) voltage
(V~)
Rated frequency
(Hz)
Temp Range
(°C)
YC-130L Spraying aluminumliner
(without disc recorder)
YC-130L (Stainless steel liner)
(without disc recorder)
YC-130L Spraying aluminumliner
(with disc recorder)
YC-130L (Stainless steel liner)
(with disc recorder)
YC-315L HIPS
YC-395L HIPS
YC-400L Spraying aluminumliner
YC-400L (Stainless steel liner)
YC-525L Spraying aluminumliner
YC-525L (Stainless steel liner)
16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32 16 ~32
N R134a/110g 220-240 50/60
2~8
N R134a/110g 220-240 50/60
2~8
N R134a/110g 220-240 50/60
2~8
N R134a/110g 220-240 50/60
2~8
N
R290/25g 220-240 50/60
2~8
N
R290/28g 220-240 50/60
2~8
N
R290/65g 220-240 50/60
2~8
N
R290/65g 220-240 50/60
2~8
N
R290/80g 220-240 50/60
2~8
N
R290/80g 220-240 50/60
2~8
Volume (L)
(Rated) current
(A)
Weight (kg)
Exterior dimensions
(DxWxH) (mm)
130
1.1
51 625X650X810
130
1.1
57 625X650X810
130
1.1
130
1.1
51 625X650X920 57 625X650X920
315 1.77 87 673X650X1762 395 1.39 95 673X650X1992
400 2.23 110 645X700X2010 400 2.23 116 645X700X2010 525 2.17 141 810X720X1973 525 2.17 148 810X720X1973
YC-650L Spraying aluminumliner 16 ~32
CompressorA
N R290/105g 220-240 50/60
2~8
650 3.06 142 890X715X1985
YC-650L Spraying aluminumliner 16 ~32
Compressor B
N
R290/85g 220-240 50/60
2~8
650 3.06 142 890X715X1985
YC-650L (Stainless steel liner)
C o m p r e s s o r A
16 ~32
N R290/105g 220-240 50/60
2~8
650 3.06 157 890X715X1985
YC-650L (Stainless steel liner)
Compressor B
16 ~32
YC-725L Spraying aluminumliner
16 ~32
YC-725L (Stainless steel liner)
16 ~32
YC-1015L Spraying aluminumliner
16 ~32
YC-1015L (Stainless steel liner)
16 ~32
YC-1320L Spraying aluminumliner
16 ~32
YC-1320L (Stainless steel liner)
16 ~32
YC-1505L (Stainless steel liner)
16 ~32
N
R290/85g 220-240 50/60
2~8
650 3.06 157 890X715X1985
N
R290/90g 220-240 50/60
N
R290/90g 220-240 50/60
N
R290/90g 220-240 50/60
N
R290/90g 220-240 50/60
N R290/150g 220-240 50/60 N R290/150g 220-240 50/60
N R290/150g 220-240 50/60
2~8
725
3.9
171 718X1093X1992
2~8
725
3.9
189 718X1093X1992
2~8
1015 3.78 185 852X1180X1990
2~8
1015 3.78 223 852X1180X1990
2~8
1320 6.2
235 826X1453X1998
2~8
1320 6.2
258 826X1453X1998
2~8
1505 5.2
322 832X1798X1997
- The foaming material of this product is cyclopentane.
25
10. Packing List
Item
Operation Manual
Certificate of Conformity
Accessory package
Key
Quantity
1
1
1
2
- The door handle accessory applies to: models YC-650L,YC-725L, YC-1015L, YC-1320L, YC-1505L; * The specific packing list shall be subject to the physical objects received.
26
Zhongke Meiling Cryogenics Company Limited
Address: No 1862 Zishi Road, Hefei City, Anhui, P.R. China Production Address:
No.1862 Zishi Road, Economic and Technological Development Zone, Hefei City
Post Code: 230601 Material Code: 890375561 Email: zkmeiling@zkmeiling.com;
technical.service@zkmeiling.com Website: www.melingbiomedical.com Production
Date: See nameplate on the freezer body Prepared in: August 2023
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>