DW-YL90 Ultra Low Temperature Biomedical Freezer
“`html
Product Information:
Specifications:
-
Models: DW-YL90, DW-FL90, DW-YL270, DW-FL270, DW-YL450,
DW-FL450 -
Manufacturer: Meling Biomedical
-
Website: www.melingbiomedical.com
-
Email: zkmeiling@zkmeiling.com
Product Usage Instructions:
1. Cabinet Fixing:
Twirl the support leg or lock the universal caster to ensure the
cabinet does not move during use.
2. Installation of Fixing Plates:
For models DW-FL270 and DW-YL450 (110V 60Hz):
-
Take out the two fixing plates and check if they are
consistent. -
Remove the bolts from the fixing plate and fix it on the back
of the refrigerator. -
Connect the fixing plate to an immovable wall or bracket on the
back of the refrigerator and ensure it is securely fixed.
3. Power On:
Turn on the power switch after connecting the power source. If
applicable, turn on the battery switch as well.
4. Door Alarm and Temperature Control:
Refer to the control panel functions for setting up door alarms,
USB data export, and temperature control.
Frequently Asked Questions (FAQ):
Q: How do I unlock the keyboard for parameter adjustment?
A: Under normal operating state, simultaneously press and hold
specific keys for 3 seconds to unlock the keyboard. Follow the
on-screen instructions to enter the password and adjust
parameters.
Q: What is the quick setting menu for after power on?
A: After the power-on self-test, you can access the quick
setting menu to adjust various parameters such as time settings,
print intervals, and temperature data recording periods.
Q: How do I use the optional printer function?
A: The system can store data for up to 7 days for printing.
Press the print key to print temperature data within a specific
time period.
“`
DW-YL90 DW-FL90 DW-YL270 DW-FL270 DW-YL450 DW-FL450
Better science, Better life!
www.melingbiomedical.com zkmeiling@zkmeiling.com
8
8
6.2 Optional Accessories
18
18 18 19 19
20
21 24
Thank you for choosing and using MELING BIOMEDICAL products! For your safe and
convenient use and reasonable maintenance of the equipment, please read the
Operation Manual carefully before use and keep it properly for reference. The
equipment operator can copy some chapters of this operation manual, but only
for internal use, for example, to instruct the user how to deal with
emergencies. These chapters are clearly marked in the catalog of the manual.
MELING BIOMEDICAL has no obligation and responsibility for any instrument
damage caused by the user’s failure to use the equipment according to the
instructions or the method specified by the manufacturer. Due to the rapid
improvement of MELING BIOMEDICAL products, the functions described in the
instructions may be inconsistent with those of the products you purchased.
Please refer to the physical functions.
® Please read carefully the Attention and Safety Precautions (in 2. Safety
Instructions). ® During transportation or use, no violent vibration or
collision is allowed and the freezer shall be kept away
from rain. Store in a clean room with humidity no more than 80 %, no corrosive
gas and good ventilation. ® Ultra low temperature freezer and biomedical
freezer (hereinafter referred to as equipment) can only be
operated by trained and authorized personnel. ® Maintenance of the equipment
can only be completed by MELING BIOMEDICAL or an agent authorized
by MELING BIOMEDICAL. ® If the operator encounters any situation not mentioned
in this manual, please contact MELING
BIOMEDICAL or the agent authorized by MELING BIOMEDICAL for the correct
handling method. ® If the equipment is not used according to the method
specified in the specification, it may be damaged. ® Try to use the
accessories provided by MELING BIOMEDICAL. If users would like to use other
accessories, MELING BIOMEDICAL will not be responsible for the adverse
consequences caused therefrom. ® Equipment must be inspected and maintained
regularly to ensure good operation of the equipment. ® Do not use mechanical
devices or other means to accelerate the defrosting process, other than those
recommended by the manufacturer.
1
Beware of fire.Warning;flammable material Beware of low temperature.
Warning: crushing of hands.
Do not use the power supply that is not specified in the equipment design, so
as to avoid overheating, short circuit and other faults. For example,
connecting 110 V rated valtage products to 220 V power supply may cause faults
such as over heating and equipment buming. For detailed input voltage AC
please refer to Specification(Rated voltage ±10%). If the operating voltage is
too low or too high, a suitable automatic voltage stabilizer must be installed
for cooperative use.
2
3
For detailed input voltage AC please refer to Specification(Rated voltage
±10%).
In order to operate the equipment normally and obtain the best performance
level, the installation site of the equipment shall meet the following
requirements: ò It cannot be installed in a narrow and closed space, and the
door of the room shall not be smaller or lower than this
equipment, which shall at least ensure the normal access of the equipment, so
as to avoid the maintenance difficulties in case of equipment failure, which
may result in damage to stored items due to the failure to repair the
equipment in time; ò The installation floor must be solid, flat, non-
combustible and able to bear the weight of the equipment during operation; ò
It shall be with good ventilation, and direct sunlight shall be avoided; ò
Each equipment needs to use a power outlet independently. Please ensure that
the plug and outlet are firmly connected; ò Check the working voltage before
use. In areas with unstable voltage, consider using a voltage stabilizer
suitable for the equipment load to ensure that the input voltage requirements
in the installation environment are met;
4
4. Cabinet fixingtwirl the support leg or lock the universal caster to ensure the cabinet not moving
when in use.
5.Install the fixing plates(Apply to DW-FL270,DW-YL450(110V 60Hz)).
a.Firstly take out the two fixing plates and check if consistent.
b.Remove the bolts from the fixing plate,and then fix the plate on the back of the refrigerator.
c.Connect the fixing plate to the immovable wall or bracket on the back of the refrigerator and finally
check whether it is fixed.
fixing plate
Screw Mounting hole
Turn on the power switch after the power is on(If there is a battery, turn on
the battery switch).
5
6
7
°C
Ref.
DOOR ALARM
USB TEMPERATURE CONTROL PRINT
Type A control panel Function description of Type A control panel(Applicable to DW-YL90,DW-FL90,DW-FL270,DW-YL270)
”
”
to export the data thru U disk (up to12 months’ data once time)
8
When the keyboard was unlocked already Unlock:under normal operating
state,simultaneously press and keys for 3s and the digital tube will display
the parameter code”0000″;By pressing to enter the password”0005″and holding to
unlock. Then press key for 3s,the digital tube will display the parameter
code”PS1″and enter the setting and adjustment parameters.Use or key to scroll
the parameters;
9
YL
-10~ -25
-23
FL
-20~ -40
-40
YL
10
FL
10
YL
5
FL
5
10
Quick setting of time after power on After the power-on self-test on the display board is completed, the quick setting menu is displayed.
Menu item Menu
Menu description
Set range
Default
Unit
n
Set logger module time – year
10~50
/
y
Set logger module time – month
1~12
—
/
Quick Settings
menu
r
Set logger module time – day
S Set logger module time – hour
F Set logger module time – minute
Pt Print interval
01~31
—
/
00~23
—
/
00~59
—
/
0~240
20
min
0~240
SCY Temperature data recording period
0: shielded
10
min
recorder
If there is no operation for 60 seconds under the quick setting menu, it will automatically exit the quick setting menu and return to normal display.
11
a
bc
°C
Ref.
°C DOOR
Fr.
ALARM
USB
TEMPERATURE CONTROL PRINT
h
(USB)
g f ed (optional printer)
Type B control panel
Function description of Type B control panel(ApplicabletoDW-FL450,DW-YL450)
a. 4.0 -20.0 is a temperature display window, which displays the average
temperature of the upper chamber and the lower chamber of the freezer body in
°C under normal operation. (The upper and lower chambers of DW-YL450 are set
at -10 ~ -25°C; The upper and lower chambers of DW-FL450 are set at -20 ~
-40°C; Different prompt characters can be displayed in the setting state (see
below for details);
b. Door opening indicator: When the door is opened, the indicator lights up.
After more than 1 minute, the door opening alarm indicator light will be on
and the buzzer triggered, displaying “door.”
c. Fault indicator light: When the product runs normally, the indicator light
is off; The indicator light is on for operation abnormality.
d. “PRINT ” (optional printer): The system can keep 7 days of data for
printing, and press the print key to print the temperature within the set time
period. e. : In parameter setting mode, reduce the parameter value. For
example, when setting the set temperature, reduce the set temperature. When
setting the parameter value, press and hold the down button, and the parameter
will decrease rapidly.
f. : In parameter setting mode, increase the parameter value. For example,
when adjust the set temperature, press this button to increase the set
temperature value. When setting the parameter value, press and hold the up
button, and the parameter will increase rapidly. Under normal conditions,
press and hold the up key for 3 seconds to export the data thru U disk (up to
12 months’ data once time).
g. : is the set/mute key; In case of no alarm state and key unlock state,
press display the ambient temperature for 5s and then return to normal
display; In the unlock state, press for more than 3s, can enter the user menu.
When buzzer is triggered (including cabinet including high temperature alarm,
door opening alarm, sensor fault alarm, etc.) and in the button unlock state,
press for the first time, and the buzzer stops ringing, and the ambient
temperature is displayed for 5s, after which the normal display is resumed
(pressing the mute button is only to turn off the buzzer for alarming this
abnormal state, if the fault is not eliminated after 5 minutes of mute, the
buzzer will alarm again.) Then press , and the buzzer alarms, and after
displaying the ambient temperature for 5s, the cabinet temperature and alarm
state are resumed. In the key unlock state, can be used as a setting key.
12
In unlock state and parameter setting mode, press this key to display
parameter values and parameter names. If the pressing time is longer than 3
seconds, save the settings and return to the normal interface. H. USB
interface Automatic export of USB data: When USB is connected to USB
interface, buzzer beeps once after data transmission is completed, “USB” is
displayed in the upper chamber, “on” is displayed in the lower chamber, PDF
files of data of current month and last month are generated in USB, buzzer
beeps once after data transmission is completed, “End” is displayed, and
normal display resumes after 6 seconds.
Note: When there is less data, “USB,” “on” and “End” prompts are not
displayed. Manual export of USB data: In the key unlock state, when the USB
flash drive is connected and the file is not being generated, press the up key
for 3 seconds, and the digital tube in the lower chamber will display “d01.”
Press the key up or down key to adjust “d00~d12,” and press key to cancel the
file generation (d00) or generate the PDF file of the record data of the
previous months (JAN-DEC).
Note: When the alarm of the digital tube flashes and displays “LoF,” the
recorder is not started; Meanwhile press and key for 3sand “LoF” disappears,
the recorder is started.
2.Function Settings 1.After powering on, the machine will enter into the
working state; 2.User parameter settings; Unlock: Under normal operation,
press and key for 3 seconds at the same time, and the digital tube in the
upper chamber will display “PS1” and the digital tube in the lower chamber
will display the parameter code “0000.” Press , enter the password “0005”
(when entering the user menu password, enter “0099” to restore the key lock
password to the default “0005.” ) , and holding to unlock. Then press key for
3s, and the digital tube will display the parameter code “PS1,” and enter the
setting and adjustment parameters. Use to scroll the parameters; Use or to
increase or decrease the value; Use to temporarily store the modified values
and return to the display parameters; If other parameters are modified, repeat
steps ~; Press for more than 3s, save the modified parameters and return to
the display parameter category. 3. Press for more than 3s, or press no key in
60s to exit the parameter setting program;
13
4. Parameter display
No. Menu item
1
b1
2
b2
Parameter Range —
Suggested settings —
3
Set1
DW-YL series: -10~-25 DW-YL series: -23 DW-FL series: -20~-40 DW-FL series: -40
Remarks Hardware version Software version
Upper chamber temperature setting
4
H01
DW-YL:0.0~10.0 DW-FL:0.0~10.0
DW-YL:10.0 DW-FL:10.0
Set value of high temperature alarm set+H; When H =0, High temp alarm is disabled; When the alarm is over high temp alarm set, H1 will be displayed on the controller
5
L01
0.0~10.0
Set value of low temperature alarm set-L; When L =0, Low
5.0
temp alarm is disabled; When the alarm is below low temp
alarm set, L1 will be displayed on the controller
6
Set2
DW-YL series: -10~-25 DW-YL series: -23 DW-FL series: -20~-40 DW-FL series: -40
Lower chamber temperature setting
7
H02
DW-YL:0.0~10.0 DW-FL:0.0~10.0
DW-YL:10.0 DW-FL:10.0
Set value of high temperature alarm set+H; When H=0, cancel the alarm; When the temperature is too high, the high temperature alarm displays H02
8
L02
9
Pt
10
tH1
0.0~10.0
0-240 min 20.0~50.0
Set value of low temperature alarm set-L; When L=0,
5.0
cancel the alarm; When the temperature is too high, the
high temperature alarm displays L02
20
Print interval
40
Upper limit of ambient temperature alarm
Mode 1: It shall be judged as once after the door is
opened and closed once and heated for 5min (time
Automatic heating mode
1
Automatic heating mode
11
P1
2 Automatic heating mode
3
Always on4
Always off5
2
setting). If the door is opened and closed again during the heating period, the heating time will be updated again; Mode 2: After the compressor is running. Heater on; When the compressor stops, the heater is delayed for 1 minute for turn-off; Mode 3: When the humidity in the cabinet is more than 80%, the door heating is on, and when the humidity in the cabinet is moderately less than 60%, the door heating is off;
Mode 4: Door heating is always on;
Mode 5: Door heating is always off.
12
Ps1
13
Max
14
NIN
15
CLR
09999 –
5
User menu password settings
–
The highest temperature since last clerance
–
The lowest temperature since last clearance
–
Clearance of the Max and Min temperature records
14
Quick setting of time after power on After the power-on self-test on the display board is completed, the quick setting menu is displayed.
Menu item Menu
Menu description
Set range
Default
Unit
n Set logger module time – year
10~50
/
y Set logger module time – month
1~12
—
/
r
Set logger module time – day
01~31
—
/
Quick Settings
menu
S Set logger module time – hour F Set logger module time – minute
00~23
—
/
00~59
—
/
Pt Print interval
0~240
20
min
0~240
SCY Temperature data recording period
0: shielded
10
min
recorder
If there is no operation for 60 seconds under the quick setting menu, it will automatically exit the quick setting menu and return to normal display.
5.Alarm display
Alarm Code H01 L01 H02 L02 H03 door
Default Description Upper chamber high temperature alarm Upper chamber low
temperature alarm Lower chamber high temperature alarm Lower chamber low
temperature alarm
High ambient temperature alarm Door opening alarm
PF
Power failure alarm
bL
Low battery alarm
EE
Communication failure
E2
Upper chamber temperature sensor alarm
E4
Lower chamber temperature sensor alarm
Er
Data Logger not connected
Pr
USB Data logging failure
LoF
Data Logger not started
15
Note:This thermostat cannot be configured with a printer,remote alarm terminal,RS485 interface 16
17
6.2 Optional function
1.Printer
The printer has been installed with a paper roll at the factory. If the paper roll is used up after a long period
of use, you can buy the same paper roll (size: thermal paper, paper width: 57.5±0.5mm, the outer diameter
of the reel: not greater than 40mm, i.e., RM57*40 back-roll paper)
Description of printer panel:
Open button. Press to open the cover;
SEL button, indicator light, for factory setting. Do not press it;
Lf button. The green indicator is the power indicator and it is normally on when the power is turned on;
Paper roll
Install the paper roll:
Press the open button 1, open the cover, install the paper roll,
close the cover plate and allow the head of the roll paper to
slightly extend out of the cover plate.
2. Remote alarm terminal, RS485 interface
Remote alarm terminal, RS485 interface and network interface
are installed in the lower part of the back of the freezer body.
N.O. COM. N.C.
Note: These configurations are standard for some models.
Normally Common Normally
RS-485
open terminal closed Remote alarm terminal
18
The integrated overload protecter within our products is programmed to shut
down the power supply automatically with respect to short circuit and/or
system overload for the sake of protecting the entire system and crucial
components such as the compressor.
7.2 Equipment Discontinuation
Deactivation: If the equipment is stored in an unsupervised area for a long
time and not used, the power supply should be cut off and the inner and outer
surfaces of the cabinet should be cleaned with a warm, damp soft cloth, aired
and sealed. The equipment must be locked to ensure that children cannot open
the cabinet door. Scrapping: When the equipment reaches the end of its service
life, it should be scrapped and must be handed over to a qualified
professional recycling agency for disposal per local regulations. Non-
professionals are not allowed to disassemble and break down the equipment
without authorization. The scrapped equipment should be placed in a designated
area inaccessible to children to avoid danger.
7.3 Maintenance, Replacement, and Recovery of Rechargeable Batteries
Battery installation position: top of the cabinet and bottom inside the
electrical box. 1. Battery maintenance In order to prolong the service life of
the battery and avoid the product being left unused for a long time, the
product must be operated for more than 24 hours every month to facilitate
charging; The equipment should be connected to power supply regularly
(generally once a month) and turn on the power switch and battery switch for a
period of time to complete the charging if it’s not used for a long time,and
each charging time should not be less than 24 hours; Power switch and battery
switch should be turned off in time after discontinuation of equipment,
otherwise it will cause battery capacity loss and in serious cases battery
permanent damage could happen; The main power switch must be turned off during
a long-term power outage or during transportation, otherwise long-term
discharge causes power loss in battery and even permanent damage to the
battery, and the display is abnormal after re-energizing; The battery is
consumable, with a service life of about 2-3 years. If the battery is used
improperly, such as power loss or reaching the end of battery life, it will
cause a low battery alarm (refrigeration is not affected, but there is an
alarm failure and influence to the use function of the printer, please contact
the after-sales staff of local distributors for replacement). 2. Battery
replacement and recovery Turn off the power switch and pull out the power plug
from the outlet; Remove the screws on the electrical box with a screwdriver.
(Note: There are high-voltage electrical components in the electrical box,
therefore before opening, turn off the power supply, unplug the power plug and
turn off the power lock switch of the refrigerator, and the electrical box can
only be opened by qualified engineers or maintenance personnel); Pull out the
battery connecting cable; (Before unplugging the battery fixing cable, pay
attention to the cathode/anode of the battery and the sequence of the
connecting cables, so as to prevent burning down the control system due to
installation of the new battery with cathode/anode reversed. Generally, the
red wire is connected to the cathode, and the black wire is connected to the
anode; Use a screwdriver to remove the fastening screws on the battery fixing
plate and take out the battery; For a replaced battery that can be recycled,
please contact the local battery recycling agency. Tips:
In order to effectively ensure that the battery replacement meets the
requirements of the control system and avoid the influence of improper
operation on the system, please contact our after-sales personnel for
replacement or guidance.
When the circuit is short circuit or overloaded, the overload protector will
disconnect the power supply of the device. Please contact after sales at this
time.
19
8. Troubleshooting and Maintenance Services
Any product may fail. Please observe the operation of equipment in time during
use. If there is any abnormality, please check and handle it according to the
following table first. If the abnormality can’t be changed, please inform our
service center in time, and we will serve you wholeheartedly to avoid losses.
Problems
Equipment does not work
Compressor is not running
The temperature does not reach the
set value
High noise
Condensation on cabinet surface Undesirable odor
The door is not closed properly,
The alarm lamp flashes and the buzzer is triggered Lighting damage
(refrigeration
equipment)
Causes and solutions
Please make sure that the outlet is energized. Please make sure power plug is
plugged in,not loose. Please make sure the power fuse is not disconnected.
Please make sure the supply voltage is appropriate,not too low or too high.
Please make sure that the temperature is set correctly.
Please check whether the temperature inside the cabinet is too low.
Please make sure that the door is closed tightly and don’t open it too many
times during a short time. Please don’t put too many items in at one time.
Please make sure that the cabinet is placed on a flat ground. Please make sure
that the cabinet does not contact the wall. In rainy and humid seasons, door
condensation is normal, and it shall be wiped off with a dry cloth.
Equipment needs to be cleaned. Items with heavy smell are not packed.
Please make sure that the temperature is set correctly. Please check whether
the temperature inside the cabinet is too low.
After the equipment is used for a period of time, the door seal becomes hard
and deformed. Maintenance method: Blow the deformed part of the hot seal with
a blower to
Please call the after-sales telephone and contact after-sales service
personnel of MELING BIOMEDICAL for replacement. Please do not replace the
parts by yourself.
Warning:
The following conditions are not faults When the compressor starts and stops,
the equipment parts will make a slight impact sound; When overheated items are
put in after the door is opened, high temperature alarm and high humidity
alarm will be triggered in the control system (if this function is available,
please refer to the alarm display table); Solution: Put the items into the
equipment after they are naturally cooled. Put the items in small quantities
in batches, and do not put too much at a time. After the system runs stably,
the high temperature alarm and high humidity alarm will be released
automatically. Slight sound of running water caused by refrigerant flowing in
the pipeline. ¿ The system contains refrigerant under high pressure. Do not
tamper with the system. It must be serviced by suitably qualified persons
only. ¿ Connect to potable water supply only. ¿ Do not store explosive
substances such as aerosol cans with a flammable propellant in this appliance.
¿ If symbol ISO 7000-1701 (2004-01) is used, its meaning shall be explained. ¿
In order to reduce flammability hazards the installation of this appliance
must only be carried out by a suitably qualified person. ¿ The refrigeration
system is under high pressure. Do not tamper with it. Contact qualified
service personal before disposal.
20
¿ The equipment can only be repaired, maintained or improved by the engineers
certified by MELING BIOMEDICAL, so as to ensure the normal operation of the
equipment and the compliance with corresponding safety standards.
¿ Please clean and disinfect the equipment before notifying the maintenance
engineer; During the warranty period of the equipment, the Company will not
undertake the warranty obligation if the fault or damage is caused by improper
use of the user.
¿ Ambient temperature for storage: -40°C-+55°C, relative humidity: 10% ~ 90%.
Model
Ambient temperature
(°C)
Climate type
Refrigerant and
loading amount
(Rated) voltage
(V~)
Rated frequency
(Hz)
Temp Range ( )
Volume (L)
(Rated) current
(A)
Weight (kg)
Exterior dimensions (DxWxH)
(mm)
DW -FL90 (Stainless steel liner)
16~32
DW -YL90 (Stainless steel liner)
DW -FL270 (Stainless steel liner)
16~32 16~32
DW -YL90 (Spraying aluminum liner
16~32
DW -YL270
16~32
DW -YL450 for condenser structure A 16~32
DW -YL450 for condenser structure B 16~32
DW -FL90 (Spraying aluminum liner
16~32
DW -FL270 (Spraying aluminum liner
16~32
DW -FL450
16~32
N
R290/45g 110
60
-20~-40 90 4.19
62 669×700×854
N R600a/55g 110
60
-10~-25 90 1.66
59 669×700×854
N R290/110g 110
60
-20~-40 270 4.22
98 640×700×1792
N R600a/55g 110
60
-10~-25 90 1.66
54 669×700×854
N R600a/97g 110
N
R600a/ 65gx2
110
N
R600a/
110
50gx2
N
R290/45g 110
60
-10~-25 270 2.51
90 640×700×1792
60
-10~-25 450 2.8
136 735×810×1960
60
-10~-25 450 2.8
136 735×810×1960
60
-20~-40 90 4.19
54 669×700×854
N R290/110g 110 N R290/42gx2 110
21
60
-20~-40 270 4.22
90 640×700×1792
60
-20~-40 450 5.26 144 735×810×1960
Model
Ambient temperature
()
Climate type
Refrigerant and
loading
amount
(Rated)
voltage (V~)
Rated
frequency (Hz)
Temp
Range (°C)
Volume (L)
(Rated)
current Weight
(A)
(kg)
Exterior dimensions (DxWxH) (mm)
DW-YL90 (Spraying aluminum liner
16-32
N
R600a/ 55g
220-230
50
-10~-25 90
0.98 54 669×700×854
DW-YL90 Stainless steel
liner
16-32
N
R600a/ 55g
220-230
50
-10~-25 90
0.98 59 669×700×854
DW-YL270 Spraying aluminum
liner
16-32
N
R600a/ 97g
220-230
50 -10~-25 270
1.53
90 640×700×1792
DW-YL270 Stainless steel
liner
16-32
N
R600a/ 97g
220-230
50 -10~-25 270
1.53
98 640×700×1792
DW-YL450 Spraying aluminum liner
16-32
N
R600a/ 65g×2
220-240
50
-10~-25 450
1.9 136 741×810×1960
DW-YL450 Stainless steel
liner
16-32
N
R600a/ 65g×2
220-240
50
-10~-25 450
1.9 162 741×810×1960
DW-YL450 H I P S
16-32
N
R600a/ 65g×2
220-240
50
-10~-25 450
1.9 128 741×810×1960
DW-FL90 Stainless steel
liner
16-32
N
R290/50 g
220-240
50
-20~-40 90
1.33 62 669×700×864
DW-FL90 H I P S
16-32
N
R290/50 g
220-240
50
-20~-40 90
1.33 52 669×700×864
DW-FL90 Spraying aluminum liner
16-32
DW-FL270 Stainless steel
liner
16-32
N
R290/50 g
220-240
50
-20~-40 90
1.33 54 669×700×864
N
R290/ 115g
220-240
50
-20~-40 270
2.8
98 640×700×1792
DW-FL270 H I P S
16-32
N
R290/ 115g
220-240
50
-20~-40 270
2.8
86 640×700×1808
DW-FL270 Spraying aluminum liner
16-32
DW-FL450 Stainless steel
liner
16-32
N
R290/ 115g
220-240
50
-20~-40 270
2.8
90 640×700×1792
N
R290/ 50g×2
220-240
50
-20~-40 450
4.50 162 735×810×1960
DW-FL450 H I P S
16-32
N
R290/ 50g×2
220-240
50
-20~-40 450
4.50 133 735×810×1960
DW-FL450 Spraying aluminum liner
16-32
N
R290/ 50g×2
220-240
50
-20~-40 450
4.50 144 735×810×1960
22
MMooddeell
tetemAmAm(pmpebebriraei)aetnutnturtree
CCtayltilyptmiempeaete
RReefrfirgigeerarannt t aannddloloaaddiningg
aammoouunnt t
(RRaatteedd)
vvoollttaaggee ((VV~~))
RRaatteedd
ffrreeqquueennccyy ((HHzz))
Temp
Range (°C)
Volume (L)
(Rated)
current Weight
(A)
(kg)
Exterior dimensions (DxWxH) (mm)
DDWW–YYLL9900 ((SSpprraayyiinngg aalluummiinnuummlliinneerr
1166–3322
NN R60R065a05/0g5a5/g 222200–224400
6600
-10~-25 90
DDWW–YYLL9900 SSttaaiinnlleesssssstteeeell
lliinneerr
1166–3322
NN R60R065a05/0g5a5/g 222200–224400
6600
-10~-25 90
0.59 54 678×700×864 0.59 62 678×700×864
DDWW–YYLL9900 ((HHIIPPSS))
DDWW–YYLL227700 SSpprraayyiinngg aalluummiinnuumm
lliinneerr
1166–3322 NN R60R065a05/0g5a5/g 222200–224400
6600
-10~-25 90
1166–3322 NN R60R069a07/0g9a7/g 222200–224400
6600
-10~-25 270
0.59 1.05
52 678×700×864 90 640×700×1792
DDWW–YYLL227700 SSttaaiinnlleesssssstteeeell
lliinneerr
1166–3322
NN R60R069a07/0g9a7/g 222200–224400
6600
-10~-25 270
1.05
98 640×700×1792
DDWW–YYLL227700 ((HHIIPPSS))
1166–3322 NN R60R069a07/0g9a7/g 222200–224400
6600
-10~-25 270
DDWW–YYLL445500 SSpprraayyiinngg aalluummiinnuummlliinneerr
1166–3322
NN
R60R06a0/06a5/g 6×52g×2
222200–224400
6600
-10~-25 450
1.05 86 640×700×1808 1.4 136 741×810×1960
DDWW–YYLL445500 SSttaaiinnlleesssssstteeeell
lliinneerr
1166–3322
NN
R60R06a0/06a5/g 6×52g×2
222200–224400
6600
-10~-25 450
1.4 162 741×810×1960
DDWW–YYLL445500 HHIIPPSS
1166–3322
NN
R60R06a0/06a5/g 6×52g×2
222200–224400
6600
-10~-25 450
aalluu(Dm(DmSSWiWpinpnr-ur-auFaFmymLyLi9inl9nli0ing0gneerr))
16~32
N R290/50g 220-240 60 -20~-40 90
1.4 128 741×810×1960 1.76 54 669×700×864
DW-FL90 (HIPS)
16~32
N R290/50g 220-240 60 -20~-40 90
1.76 52 669×700×864
DW-FL90 (Stainless steel
liner)
DW-FL270 (Spraying aluminum liner)
16~32 16~32
DW-FL270 (HIPS)
16~32
DW-FL270 (Stainless steel
liner)
DW-FL450 (Spraying aluminum liner)
16~32 16~32
DW-FL450 (Plastic liner)
16~32
DW-FL450 (Stainless steel
liner)
16~32
N R290/50g 220-240 60 -20~-40 90
1.76 62 669×700×864
N
R290/ 115g
220-240
60
-20~-40 270
2.22
90 640×700×1792
N
R290/ 115g
220-240
60
-20~-40 270
2.22
86 640×700×1808
N
R290/ 115g
220-240
60
-20~-40 270
2.22
98 640×700×1792
N
R290/ 45g ×2
220-240
60
-20~-40 450
3.71 144 735×810×1960
N
R290/ 45g ×2
220-240
60
-20~-40 450
3.71 133 735×810×1960
N
R290/ 45g ×2
220-240
60
-20~-40 450
3.71 162 735×810×1960
23
24
Zhongke Meiling Cryogenics Company Limited
Address: No 1862 Zishi Road, Hefei City, Anhui, P.R. China Production Address:
No.1862 Zishi Road, Economic and Technological Development Zone, Hefei City
Post Code: 230601 Email: zkmeiling@zkmeiling.com;
technical.service@zkmeiling.com Website: www.melingbiomedical.com Material
Code: 890375568 Production Date: See nameplate on the freezer body Prepared
in: August 2023
28