01DV3000 Demand Valve Resuscitator
User Manual
DEMAND VALVE 01DV3000
DEMAND VALVE RESUSCITATOR
01DV2000
Introduction
1.1. Introduction
O-Two oxygen powered Demand Valve provides trained individuals with a safe and
effective means of delivering 100% oxygen to spontaneously breathing patients
in respiratory distress, at the fl ow rate demand by the patient.
O-Two oxygen powered Demand Valve Resuscitator provides a fast, simple and
effective means to ventilate the non-breathing patient and to deliver 100%
oxygen to spontaneously breathing patients in respiratory distress, at the flow
rate demand by the patient.
O-Two Demand Valves and Demand Valve Resuscitators are lightweight, portable,
and extremely durable. Designed for the demands of the pre-hospital
environment, they can be operated anywhere a regulated 50 PSI oxygen supply is
available.
1.2. Warranty
WARRANTY
O-Two equipment is manufactured from the fi nest quality materials. Each
individual part is subject to strict quality control tests to ensure
exceptionally high standards.
The manufacturer warrants to the purchaser of DEMAND VALVE (RESUSCITATOR) that
its component parts are free from defects in material and workmanship for a
period of two years from the date of purchase. The manufacturer will replace
and /or repair all parts of the resuscitator at its option for two years from
the date of purchase at no cost to the purchaser, upon the notifi cation of
the defects, in writing by the purchaser. All shipping costs shall be borne by
the purchaser. The manufacturer shall be liable under this warranty only if
the device and its parts have been used and serviced in the normal manner
described in the instruction manual. There are no other expressed or implied
warranties. This warranty gives no specific legal rights. You may also have
other rights which may vary according to local regulations.
1.3. Safety Precautions
The O-Two Demand Valve and Demand Valve Resuscitator are designed to provide
emergency ventilatory support to victims of respiratory and/or cardiac arrest
and demand breathing capability or those in respiratory distress. They are
intended for use by suitably trained and qualified personnel. The following
precautions should always be observed:
CAUTION
- FEDERAL LAW IN THE U.S.A. RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A LICENSED PRACTITIONER.
- WHEN THE UNIT IS IN USE, DO NOT SMOKE OR USE NEAR OPEN FLAME EITHER DURING USE OR WHEN CHANGING THE CYLINDER.
- WHEN NOT IN USE, ALWAYS TURN OFF THE CYLINDER.
- NEVER ALLOW OIL OR GREASE TO COME INTO CONTACT WITH ANY PART OF THE CYLINDER, REGULATOR OR DEMAND VALVE.
- DO NOT DISASSEMBLE ANY PART OF THE DEMAND VALVE (RESUSCITATOR) EXCEPT WHERE DESCRIBED IN THIS MANUAL AS ANY UNAUTHORIZED DISASSEMBLY WILL INVALIDATE THE WARRANTY.
- AFTER USE, ALWAYS ENSURE THAT ALL COMPONENTS ARE CLEANED IN ACCORDANCE WITH THE INSTRUCTIONS PROVIDED IN THIS MANUAL.
- ALWAYS USE THE CHECK LIST TO ENSURE THAT ALL COMPONENTS ARE REASSEMBLED CORRECTLY AND READY FOR USE.
- AFTER USE, ALWAYS ENSURE THAT A FULL GAS CYLINDER IS ATTACHED BEFORE RETURNING THE UNIT TO ITS NORMAL STORAGE POSITION.
1.4. Performance Specification
Manual Flowrate (01DV2000 only) | 40 L/min |
---|---|
Peak Inspiratory Flow rate @-8 cmH2O | >120 L/min |
Demand Breathing Flow Rate | 0 -140 L/min |
Exhalation Resistance | 0 to 6 cmH2O @ 60 L/min |
Inhalation Resistance | 0 to -6 cmH2O @ 60 L/min |
Input Pressure | 50 PSI |
Max Pressure Relief Valve (01DV2000 only) | 60 cmH2O |
Audible Over Pressure Alarm (01DV2000 only) | YES |
Triggering Pressure | 0 to -3.5 cmH2O |
Operating Temperature | -180C to 500C (00F to 1220F) |
Storage Temperature | -400C to 600C (-400F to 1400F) |
Operating and Storage Humidity | RH 15% to 95% |
Input Connection | 9/16” DISS |
Patient Connector | 15/22 mm |
Patient Valve Dead Space | 8 ml |
Weight (01DV2000) | 230 grams (8.2 oz) |
Weight (01DV3000) | 210 grams (7.4 oz) |
Dimensions (01DV2000) | Dia. 51mm x 90mm (Dia. 2.0” x 3.5”) |
Dimensions (01DV3000) | Dia. 51mm x 81mm (Dia. 2.0” x 3.2”) |
Gas Supply
2.1. Gas Supply Connections
The O-Two Demand Valve and Demand Valve Resuscitator are designed to operate
on an input pressure of 50 PSI of compressed medical oxygen from either a
cylinder or piped-in system.
The inlet fitting on the Demand Valve and Demand Valve Resuscitator is a
standard 9/16” DISS fitting.
Operating Procedure
3.1. Demand Breathing
The O-Two Demand Valve and Demand Valve Resuscitator are equipped with a
Demand System enabling spontaneously breathing patients to demand oxygen.
An inspiratory eff ort by the patient will activate the demand valve and
oxygen will fl ow to the patient at a rate and volume in line with their
inspiratory effort.
3.2. Manual Resuscitation (01DV2000 only)
The O-Two Demand Valve Resuscitator (01DV2000) has been fitted with a Manual
Button to allow manual ventilation of the nonbreaching or respiratory
distressed patient.
By using the manual button, the operation of the resuscitator can be easily
timed with chest compressions during CPR. This will avoid the potential
problem of the aspiration of stomach contents due to gastric distension which
may occur if overlap of chest compression and ventilation occurs. (In
patients that are intubated this overlap of compression and ventilation may
increase cardiac output without the danger of gastric distension.)
WARNING
At all times during resuscitation the rise and fall of the patients’ chest
should be monitored to ensure adequate ventilation.
3.3. Operating Instructions
- Ensure that the Demand Valve/ Regulator connection is secured and no leaks present.
- Turn the cylinder hand wheel counter clockwise to the ON position (at least one full turn).
- Ensure that the regulator contents gauge reads at least 500 psi. If not, replace with a full cylinder.
- Attach the Universal Facemask to the Demand Valve outlet.
- Place the mask on the patient’s face, covering both the nose and mouth and hold the mask on his/her face with a light pressure. The sound of gas fl owing should be heard with patient’s inspiratory eff orts.
- Should manual ventilation is required for the non-breathing or respiratory distressed patient, simply press the Manual Button.
- Ensure that the cylinder is turned OFF after each use.
Note: Demand Valves and Demand Valve Resuscitators are considered critical devices, and their components considered critical components. Only those individuals trained in CardioPulmonary Resuscitation and operation of oxygen- powered resuscitators should use this equipment. Thoroughly review the instruction manual before use.
Servicing
4.1. Routine Maintenance
WARNING
The O-Two Demand Valve and Demand Valve Resuscitator are designed to provide
respiratory support in all emergency situations. Failure to follow the
maintenance and inspection routine could result in incorrect operation of the
resuscitator.
To ensure proper operation of the Demand Valve and Demand Valve Resuscitator,
regular inspection and checking of the demand valve and accessories for
correct function should be undertaken by a responsible member of staff on a
regular basis, and it is recommended that a service record be maintained for
each unit.
This check is to ensure that all of the accessories and demand valve
components are present, the oxygen cylinder is full and that the demand valve
is in working order.
MAINTENANCE
Resuscitation parameters should be checked at least every six months, and more
frequently in high use applications. Units with test parameters outside of the
ranges listed in the product specifications should not be used. O-Two Demand
Valve are not designed for field disassembly or service outside that indicated
in this manual. Any malfunctioning units should be returned to the
manufacturer or an Authorized O-Two Medical Technologies Dealer. Unauthorized
repairs will nullify the product warranty.
4.2. Cleaning and Disinfection
Routine cleaning and/or disinfection of the O-Two Demand Valve and Demand
Valve Resuscitator should be undertaken after each patient use to maintain the
equipment in a clean condition.
WARNING
Do not immerse the demand valve into cleaning solution or disinfectant.
Discontinue to use and discard the face mask, the patient valve and the
diaphragm should discoloration or deterioration be observed.
Cleaning / Disinfection Procedure
- Operate the manual button to blow out any contaminant from the patient valve (01DV2000 only).
- Ensure that the cylinder is turned off (completely clockwise) and bleed any remaining gas out of the system by depressing the manual ventilation button (01DV2000 only).
- Remove the facemask from the Demand Valve (Resuscitator), and then disconnect the unit from the regulator. This is accomplished by counter-clockwise turning of the connection nut at the regulator end of the hose. A wrench may be required.
- Unscrew patient valve (A) from the unit body being careful to ensure that the diaphragm (B) is retained (Fig1).
- Remove the diaphragm and anti-entrainment ring (C) from the patient valve swivel housing and wash these components thoroughly in a mild soap solution;
- Disinfect the face mask, patient valve and diaphragm using commonly available cold chemical disinfecting solutions. Rinse thoroughly with water after disinfection.
- Sterilization, if required, may be achieved by other methods compatible with the materials of manufacture of these components including Ethylene Oxide Gas and Cidex.
- The demand valve body (D) and gas supply hose can be wiped over with a soft cloth and mild soap solution.
- Dry all components thoroughly.
- Re-assemble and check performance.
Replacement Parts
ORDER N0 | PART |
---|---|
17MP1528 | Silicone Diaphragm |
02FM4999-Cs | Universal Face Mask (case/12)* |
01RO01O23-B | Oxygen Pressure Regulator |
17MP9039 | Non-rebreathing Valve Assembly |
01FG6520 | Head Harness System |
01FV4303-AFNR | O-Two 6 Foot (1.85 Meter) O2 Supply Hose with AFNOR probe and |
9/16″ DISS Nut Ventilator Connection
01FV4303-AGA| O-Two 6 Foot (1.85 Meter) O2 Supply Hose with AGA probe and
9/16″ DISS Nut Ventilator Connection
01FV4303-DIN| O-Two 6 Foot (1.85 Meter) O2 Supply Hose with DIN probe and
9/16″ DISS Nut Ventilator Connection
01FV4304| O-Two 6 Foot (1.85 Meter) O2 Supply Hose with 9/16 DISS Nut and
9/16″ DISS Nut Ventilator Connection
01FV4303-UNFR| O-Two 6 Foot (1.85 Meter) O2 Supply Hose with UNIFOR probe and
9/16″ DISS Nut Ventilator Connection
01FV4303-BM| O-Two 6 Foot (1.85 Meter) O2 Supply Hose with BRITISH probe and
9/16″ DISS Nut Ventilator Connection
- Other pressure regulators and face masks are available.
Select the required oxygen supply hose for the country of use.
Contact your nearest O-Two Medical authorized distributor for a complete catalogue on all O-Two Medical manufactured products.
Your Representative is:
O-TWO MEDICAL TECHNOLOGIES INC.**
For your nearest Authorized O-Two Distributor
In North America call Toll Free 1-800-387-3405
MedNet EC-REP IIb GmbH
Borkstrasse, 10
48163 Münster, Germanywww.otwo.com