O-Two 01CV7000 Controlled Ventilation User Manual
- June 15, 2024
- O-Two
Table of Contents
01CV7000 Controlled Ventilation
Product Information
Specifications
- Model: 01CV6000, 01CV7000
- Intended Use: Pre-hospital, intra-hospital, inter-hospital, and
air ambulance settings
Introduction
The ventilator is a medical device designed for use by suitably
trained medical personnel in various healthcare settings. It is
important to read and understand the user manual before attempting
to use the ventilator.
Warnings
-
Read the manual prior to using the ventilator.
-
Avoid smoking or open flames near the unit during use or when
changing the cylinder. -
Always turn off the gas supply when the ventilator is not in
use. -
Avoid contact between oil/grease and any part of the cylinder,
regulator, or ventilator. -
Do not use heated humidifiers with this device.
-
Not recommended for use on neonates and pregnant or nursing
women. -
Unauthorized disassembly will void the warranty.
-
Clean all components after use according to the instructions
provided. -
Ensure correct reassembly of all components before using the
ventilator again. -
Have an alternative means of ventilation available in case of
gas supply failure. -
Do not leave the patient unattended during use.
-
Avoid using the device in areas with hazardous or explosive
ambient air. -
Use patient circuits supplied by O-two Medical for optimal
performance. -
Use gas pressure regulators that maintain the required output
pressure and flowrate. -
When using the air mix mode, ensure the ventilator is not
obstructed and in its normal operating orientation to avoid
increased oxygen concentration.
Preparation for Use
Before using the ventilator, follow these steps:
- Refer to the component list provided in the user manual.
- Connect the hoses according to the instructions.
- Perform a pre-use functional check as described in the
manual.
Operating Instructions
To set the ventilation parameters and operate the
ventilator:
-
Refer to the user manual for detailed instructions on setting
the ventilation parameters. -
Use the frequency/minute volume chart provided as a
reference. -
Follow the instructions for operating in extreme conditions, if
applicable.
After Use Cleaning and Testing
After each use, follow these steps for cleaning and testing:
-
Dismantle the ventilator and clean all components according to
the instructions. -
Change the air intake filter if necessary.
-
Perform post-use and monthly checking as specified in the
manual.
Preventive Servicing
Regular preventive servicing is recommended to maintain the
optimal performance of the ventilator. Refer to the manual for
specific instructions.
Troubleshooting Chart
If you encounter any issues with the ventilator, consult the
troubleshooting chart provided in the manual for possible
solutions.
FAQ
Q: Can the ventilator be used on neonates?
A: No, the ventilator is not recommended for use on
neonates.
Q: Can heated humidifiers be used with this
device?
A: No, this device is not designed for use with heated
humidifiers.
Q: Can I disassemble any part of the
ventilator?
A: Unauthorized disassembly of any part of the ventilator will
invalidate the warranty. Only disassemble parts as described in the
manual.
USER MANUAL
CAREvent® ATV+ 01CV6000
CAREvent® MRI 01CV7000
CAREvent® ATV+ & MRI
01CV6000 01CV7000
01CV6000
01CV7000
CAREvent® ATV/ATV+ & MRI User Manual – REV 26 -July 2023
TABLE OF CONTENTS
1. Introduction
4
1.1. Warnings
4
1.2. Intended Use
6
1.3. Terms and Definitions
6
1.4. General Information
7
1.5. CAREvent® MRI
7
1.6. Device Description
8
1.7. Principles of Operation
9
1.8. Modes of Operation
9
1.9. Alarms and Gauges
11
1.10. Accessories
13
1.11. Control Adjustment
14
1.12. Technical Data
15
1.13. Warranty
2. Preparation for Use
17
2.1. Component List
17
2.2. Connection of Hoses
18
2.3. Pre-use Functional Check
18
3. Operating Instructions
22
3.1. Setting of the Ventilation Parameters
22
3.2. Frequency/Minute Volume Chart
22
3.3. Operation in Extreme Conditions
23
4. After Use Cleaning and Testing
23
4.1. Dismantling and Cleaning
23
4.1.1. Changing the Air Intake Filter
24
4.2. After Use and Monthly Checking
24
5. Preventive Servicing
24
6. Troubleshooting Chart
26
7. CAREvent® Accessories
27
2
CAREvent® ATV/ATV+ & MRI – User Manual
1. Introduction 1.1. Warnings
The CAREvent® ATV+ and MRI are intended for use by suitably trained and
qualified personnel.
The following precautions should always be observed: 1. Read this manual prior
to attempting to use the ventilator.
2. When the unit is in use, do not smoke or use near open flame either during
use or when changing the cylinder.
3. When not in use, always turn off the gas supply.
4. Never allow oil or grease to come into contact with any part of the
cylinder, regulator or ventilator.
5. This device is not designed for use with heated humidifiers.
6. This resuscitator is not recommended for use on neonates and pregnant or
nuring women.
7. Do not disassemble any part of the ventilator except where described in
this manual as any unauthorized disassembly will invalidate the warranty.
8. After use, always ensure that all components are cleaned in accordance
with the instructions provided in this manual. (See section 4)
9. Always use the check list to ensure that all components are reassembled
correctly and ready for use.
10. It is recommended that an alternative means of ventilating the patient be
available in case of gas supply failure.
11. During use, the patient should not be left unattended.
12. The use of this device in areas where the ambient air may be hazardous or
explosive should be avoided as entrainment of ambient air during the use of
the air mix mode will permit the patient to inhale atmospheric gas.
13. Only the CAREvent® MRI unit may be used in a magnetic resonance imaging
department.
14. Only patient circuits supplied by O-two Medical should be used, as the
use of other circuits may adversely affect the output performance of the ATV.
CAREvent® ATV/ATV+ & MRI – User Manual
3
15. The use of gas pressure regulators that do not maintain a minimum output
pressure and flowrate in line with the requirements of the specification may
cause the device to fail resulting in the patient not being ventilated.
16. The use of this device in carring case of any type may, when used in the
60% air mix mode, result in an increase in oxygen concentration. If the air
mix mode is to be used it is recommended that the ventilator be placed in its
normal operating orientation and that the air intake on the rear of the
ventilator is not obstructed.
1.2. Intended Use
The CAREvent® ATV+ and MRI are pneumatically powered, time cycled,
intermittent, positive pressure automatic transport ventilator used to provide
resuscitation and/or ventilatory support to a widerange of patients (from
above 5 kg (11 lbs.) and higher) during transport.
They are designed to be used by suitably trained medical personnel in the pre-
hospital, intra-hospital, inter-hospital, and air ambulance settings.
In addition, CAREvent® MRI is suitable for use in 1.5 – 3.0 Tesla MRI medical
imaging suite environments, where ventilatory support may also be required
during an MRI imaging procedure.
The ventilator is suitable for use in:
· Pulmonary resuscitation during respiratory and/or cardiac arrest.
· Short term ventilatory support in the pre-hospital, intrahospital, inter-
hospital, and air ambulance transport of nonbreathing patients.
· In addition, CAREvent® MRI only is suitable for use in 1.5 3.0 Tesla MRI
medical imaging suite environments, where ventilatory support may also be
required during an MRI imaging procedure.
4
CAREvent® ATV/ATV+ & MRI – User Manual
1.3. Terms and Definitions
Airway Resistance: Pressure drop across the airway per unit flow.
ATV:
Automatic Transport Ventilator
CMV:
Controlled Mandatory Ventilation.
CPAP: Demand Valve: Expiratory Phase:
A positive pressure applied to the lungs during all ventilation phases
(Continuous Positive Airway Pressure).
A valve that delivers gas to the patient at a flowrate equivalent to that
demanded by the patient’s inspiratory effort.
The time period from the end of the inspiratory flow to the end of the
expiratory flow.
Expiratory Time (Te): Duration of the expiratory phase.
Frequency (f): Inspiratory Flow: Inspiratory Phase:
The number of breaths in one minute (also expressed as BPM).
The flow delivered to the patient by the ventilator during the inspiratory
phase.
The interval from the start of the inspiratory flow to the start of the
expiratory phase.
Inspiratory Time (Ti): Duration of the inspiratory phase.
Lung Compliance:
Volume added per unit pressure increase when gas is added to a human or artificial lung.
Maximum Patient Inflation Pressure:
The maximum airway pressure delivered by the ventilator to the patient.
Minute Volume (Vm):
The total volume of gas delivered to the patient in one minute.
The material used in this device contains “DEHP”.
Oxygen Concentration: Proximal Airway Pressure:
Patient Valve:
The oxygen content of the inspired gases expressed as a percentage.
The airway pressure measured at the patient valve.
The valve which directs the flow of gas into the lungs and out of the
expiratory port to atmosphere during expiration.
CAREvent® ATV/ATV+ & MRI – User Manual
5
PEEP Valve:
Pressure Relief Valve: Tidal Volume (Vt):
G05® :
A device which, when attached to the expiratory port of the patient valve,
holds a positive pressure in the patients airway at the end of the expiratory
phase. (Positive End Expiratory Pressure)
Valve which limits the maximum lung inflation pressure by venting excess gas
to atmosphere.
Volume of gas delivered to the patient during each inspiratory phase.
Guidelines 2005 compliance. Complies with the requirements of the ERC and AHA
for 30:2 compression : ventilation ratio.
1.4. General Information
The CAREvent® ATV+ and MRI are designed for use in the prehospital,
intrahospital, interhospital and air ambulance settings.
Designed to be used by suitably trained personnel, the various modes of operation of the device support the resuscitation and transportation of a wide range of patients (from infants to adults) who require ventilatory support. The features and controls offered by the device comply with the various International Standards for Transport Ventilators.
The G05® models have been manufactured to comply with the latest Guidelines 2005 for CPR and ECC from the European Resuscitation Council and the American Heart Association.
1.5. CAREvent® MRI
The CAREvent® MRI has been designed to meet the requirements of ventilation
within the MRI environment. The CAREvent® MRI has been tested in a 1.5 Tesla
MRI environment (unshielded magnet spatial gradient of <23Mt/m/sec and a slew
rate of 120 T/m/sec at an RF transmitter power of 2000 watts) and a 3.0 Tesla
MRI environment (unshielded magnet – spatial gradient of < 40Mt/m/ sec and a
slew rate of 150 T/m/sec at an RF transmitter power of 8000 watts) during a
head phanton spectroscopy test scan with the device located no less than 12
inches from the magnet aperture which would be its normal use position during
clinical use.
In addition, the CAREvent® MRI model can be attached to the CAREvent® MRI
mounting plate or attachment to a cart or wall/ medirail mounting bracket.
6
CAREvent® ATV/ATV+ & MRI – User Manual
1.6. Device Description
The CAREvent® ATV+ and MRI consist of a ruggedly constructed, portable control
module, input hose and optional single use (or reusable where available)
patient circuit. The ergonomically designed control groupings facilitate the
selection and setting of the breathing parameters. The colour groupings
provided on the controls adds to this ease of use concept.
The unique pneumatic alarm for low input pressure improves the level of safety
for the patient by warning the operator of any problems of insuffIcient or
failed gas supply.
The air mix capability of the ventilator improves the operating time on
bottled oxygen for long duration transports. The extremely low drive gas
consumption ensures full utilization of the cylinder contents.
Controls The controls provided are as follows:
· Ventilation Frequency · Pressure Relief (Proximal Airway Pressure) · Minute
Volume · Delivered Oxygen Concentration (%) · Manual Ventilation · Manual/AUTO
Selector · CPAP/PEEP Control · BSI Alarm Silence Button
Visual Indicators · Proximal Airway Pressure Gauge (cmH2O) · Gas Supply Status
Indicator · BSI Alarm Indicator
Connections Connections are provided on the side of the ventilator for the
Oxygen Input and the Gas Output.
1.7. Principles of Operation
The CAREvent® ATV+ and MRI are time-cycled, Intermittent Positive Pressure
devices providing a range of frequencies of ventilation and Minute Volume
settings to provide a comprehensive range of delivered tidal volumes and
ventilation rates. The design of the micro-pneumatic circuitry in the
CAREvent® ATV/ATV+ & MRI – User Manual
7
Automatic Transport Ventilator maintains a consistent I:E ratio of 1:2 across
the Minute Volume/frequency combinations to optimise the exchange of gases in
the alveoli. This helps avoid the risk of inconsistent ventilations and
protracted expiratory times that may cause additional physiological acidosis
or short inspiratory times that may not allow for complete alveolar filling or
may generate high airway pressures.
Note: The design of these ventilators does not incorporate a negative pressure
phase during either automatic or manual ventilation.
1.8. Modes of Operation
Automatic Ventilation Mode In the automatic mode (selected by turning the
Manual/AUTO Selector on the ventilator control panel to AUTO) the ventilator
will supply the patient with positive pressure ventilations of a frequency and
minute volume as selected using the Ventilation Frequency and Minute Volume
Selectors on the front panel. The I:E ratio is maintained at a constant 1:2 to
ensure that good gas exchange takes place in the lung.
Manual Ventilation Mode In the AUTO mode, when the Manual Override Button is
depressed, the ventilator will cease automatic cycling and will deliver a
constant flowrate equivalent to that selected on the flowrate selector for as
long as the button is depressed. Releasing the button will allow the
ventilator to recycle into an automatic ventilation mode following a delay
period of approximately 20 seconds. This delay allows the operator sufficient
time to provide chest compressions or initiate a further manual breath should
they so wish. If the Manual Override Button is depressed for too long the
patient is protected from high airway pressure by the Pressure Relief system.
Note: During the provision of chest compressions during mask Ventilated CPR,
the BSI alarm may activate if the mask seal is not maintained or if there is
insufficient positive intra-pulmonary pressure created during the compression
phrase to disable the alarm. Under these circumstances the BSI alarm acts as a
warning of the impending recommencement of automatic cycling.
8
CAREvent® ATV/ATV+ & MRI – User Manual
In the Manual mode, depressing the Manual override button will allow the
operator to manually ventilate the patient at a flowrate equivalent to the
Minute Volume selected.
Demand Breathing During automatic ventilation, the Demand Breathing mode will
allow the patient to commence spontaneous breathing through the ventilator
while causing the automatic cycling to cease. As with the Manual Override,
there is an increase in the expiratory time before the next automatic breath
is delivered to allow the patient to continue demand breathing. This system
ensures that the patient receives a sufficient minute volume should their
inspiratory efforts be erratic. Demand Breathing is available to the patient
in both the inspiratory and expiratory phases of the ventilation cycle.
In the Manual mode the patient can demand breathe at their own rate and
volume.
Note: It is important to constantly observe the patient’s respirations to
ensure that adequate perfusion is occurring.
Note: Small children may not be able to demand at a sufficient flowrate to
disable the automatic cycling.
Note: If used in conjunction with external chest compressions the Auto Circuit
Shut Off function may be actuated should the depth of compression create
sufficient tidal volume on chest recoil. Under these circumstances a manual
breath may be delivered (if deemed necessary) between sets of compressions.
Adjustable Pressure Relief This control allows the operator to vary the
maximum delivered airway pressure. This is achieved by diverting the excess
delivered volume to atmosphere should the set airway pressure be reached.
Air Mix Mode To conserve oxygen during long transports or when a patient’s
respiratory condition demands an oxygen concentration of less than 100%, the
air mix mode can be selected. This reduces the oxygen concentration to 60%. In
this mode the ventilator entrains
CAREvent® ATV/ATV+ & MRI – User Manual
9
ambient air into the system and automatically reduces the oxygen flow to
maintain a consistent tidal volume. By using this mode the operating time on
the cylinder supply is significantly increased.
Note: Due to the design of the micro-pneumatic circuit, increasing pulmonary
resistance has little effect on the delivered tidal volume or respiratory
rate. Research has shown that decreasing pulmonary compliance in the infant
Minute Volume settings may give a minimal increase in the delivered oxygen
concentration when used in the Air Mix mode.
Emergency Air Intake System for Gas Supply Failure In accordance with the
International Standards for this type of device, the ventilator is equipped
with a failsafe Emergency Air Intake circuit which allows the spontaneously
breathing patient to draw ambient air through the circuit should the gas
supply fail.
1.9. Alarms and Gauges
The alarm systems in the Automatic Transport Ventilator provide an audible
indication of any inconsistencies in the patients breathing and warn of a loss
of driving pressure from the gas supply. The alarms function as follows:
Low Input Pressure Alarm – Gas Supply A low frequency oscillating alarm occurs
when the gas supply reaches the minimum safe operating pressure. (See also Gas
Supply Status indicator).
p = Audible Pulsed Tone t = Time
Fig. 1 – Graphic illustration of the audible alarm signal
WARNING
When this alarm sounds, immediately change the gas supply.
10
CAREvent® ATV/ATV+ & MRI – User Manual
Airway Over Pressure Alarm Pressure Relief Continuous tone of a low pitch
during the inspiratory phase of the ventilation that indicates that the
maximum airway pressure selected has been reached.
Inspiration
Expiration
p = Audible Pulsed Tone t = Time
Fig. 2 – Graphic illustration of the Pressure Relief audible alarm signal
Breathing System Integrity (BSI) Alarm Pulsed tone of a high pitch which
activates if the patient circuit becomes disconnected or the airway pressure
achieved does not reach 10 cmH2O. Alarm can be silenced for 15 seconds by
depression of the BSI Alarm Mute Button (See also BSI Visual Indicator).
p = Audible Pulsed Tone t = Time
Fig. 3 – Graphic illustration of the BSI audible alarm signal
Gas Supply Status Indicator Located on the front panel, this visual indicator
will show green when gas is supplied to the ventilator. Used in conjunction
with the Low input pressure alarm, this indicator provides an additional
reference for the operator as to the ventilators gas supply status.
BSI Alarm Visual Indicator Located on the front panel above the airway
pressure gauge, this visual indicator will flash red in time with the BSI
Audible Alarm when the patient circuit is disconnected or insufficient airway
pressure is being produced.
CAREvent® ATV/ATV+ & MRI – User Manual
11
Airway Pressure Gauge Located on the front panel of the ventilator, this gauge
provides the operator with a visual indication of the airway pressure being
reached during the ventilation cycle.
1.10. Accessories
Supply Hose The supply hose is a standard armoured oxygen hose with a 9/16
DISS threaded connection for the ventilator.
Single Use and Reusable Patient Circuits The Single Use Deluxe Transport
Ventilation Circuits and Deluxe Reusable Patient Circuits are comprised of a
patient valve housing and 22 mm tubing. The expiratory port is configured to
accept either 30 or 19 mm PEEP valves. The 22 mm corrugated tubing is easily
attached to the patient circuit output connector on the side of the
ventilator. (fig 5.)
Note: 1. The choice of using either a Single Use or Reusable circuit is
at the discretion of the user. Only patient circuits supplied by O-Two Medical
Technologies Inc. should be used. 2. The use of a Bacterial/Viral filter is
recommended to reduce the potential for cross infection and contamination of
the reusable patient circuit. If used, the Bacterial/Viral filter must be used
on a single patient basis. 3. The use of a Bacterial/Viral Filter in
conjunction with a reusable patient circuit does not guarantee that the
patient circuit will remain free of contamination. 4. The use of a
bacterial/viral Filter on any of the CAREvent® circuits will increase
inspiratory resistance and if used on the distal side of the patient valve
will increase deadspace.
Note: Careful consideration should be given to the use of a reusable patient
circuit on more than one patient. Single use circuits are not designed for use
on more than one patient and must be discarded after each use. The
manufacturer recommends that circuits are used on a per patient basis and
reusable circuits should be cleaned and single use circuits discarded, after
each use.
Note: The patient valve assembly is flow direction sensitive. Ensure that the
patientcircuit is attached to the ventilator correctly.
12
CAREvent® ATV/ATV+ & MRI – User Manual
1.11. Control Adjustment
The ventilator is equipped with a number of selectors depending on the model.
Each selector is actuated by the following method (Fig 4).
Note: All selector positions are as viewed from the front of the ventilator.
1. Ventilation Frequency (BPM): Located on the top right hand side of the
control panel. Rotary control with an anti-clockwise rotation from low to
high.
2. Minute Volume (Litres): Located on the bottom right hand side of the
control panel. Rotary control with a clockwise rotation from high to low.
3. Pressure Relief (Maximum Delivered Airway Pressure (cmH2O): Located on the
top of the control panel slightly left of center. Rotary control with a
clockwise rotation from low to high.
4. Delivered O2 Concentration(%): Located on the bottom left hand side of the
control panel. Rotary control with a clockwise rotation from low to high.
5. Manual Ventilation: Located on the lower center section of the control
panel activated by depression of the button.
6. Manual/Auto Selector: Located on the front panel of the ventilator. Rotary
control with a clockwise rotation from Manual to Auto.
7. CPAP/PEEP Control: Located on the front panel of the ventilator. Rotary
control with a clockwise rotation from 0 – 20 cmH2O.
8. BSI Alarm Mute Button: Located on the front panel of the ventilator above
the pressure gauge. Activated by depression of the button.
CPAP/PEEP Control
Adjustable Pressure
Relief
Bsi Alarm Silence Control
Bsi Alarm Visual
Indicator
Airway Pressure
Frequency (BPM) Selector
Minute Volume Selector
Manual/Automatic Selector
Manual Air Mix Ventilation Selector Button
Gas Supply Status Indicator
Fig. 4 – Front Panel Layout for ATV+ and MRI models
CAREvent® ATV/ATV+ & MRI – User Manual
13
1.12. Technical Data
(All specifications are subject to a tolerance of +/- 10% except the I:E Ratio
which is subject to a tolerance of +/- 20% and Maximum Airway Pressure
+0/-15%)
Minute Volume Range:
2 – 14 L/min
Frequency Range:
8 – 40 BPM
I:e Ratio:
1:2
Input Pressure:
45 – 87 PSI (3 – 6 Bar)
Automatic Flowrate:
6 – 42 L/min
Manual Flowrate:
6 – 42 L/min
Manual Overide Delay Time:
17 – 23 sec
Demand Breathing Flowrate:
100 L/min @ -6 cmH20 (hPa)
Demand Breathing Triggering Pressure: -2 cmH20 max
Auto Shut Off Delay Time:
5 – 8 sec
CPAP/PEEP: Oxygen Concentration:
0 – 20 cmH20 60 or 100 %
Inspiratory/Expiratory Resistance: Operating Temperature: Storage Temperature: Gas Supply Status Indicator: Pressure Relief Range:
< 6 cmH20 (hPa) @ 60 L/min
-18oC to + 50oC / 0oF to +122oF – 40oC to + 60oC / – 40oF to + 140oF
Provides a visual Indication of gas Supply status. 20 – 60 cmH20 19.6 TO 58.8
mBar
14
CAREvent® ATV/ATV+ & MRI – User Manual
Low Input Pressure Alarm:
Audible pulsed alarm to indicate input pressure is dropping below the minimum requirement. Cannot
be switched off.
Airway Pressure Alarm:
Audible alarm to indicate that the selected peak airway pressure has been reached.
BSI Alarm:
Pressure Gauge Accuracy:
Audible high frequency pulsed alarm indicating that the patient circuit is
disconnected or insufficient ventilation pressure (<10 cmH2O) has been
attained
± 2% Full Scale
Input Connection:
9/16″ DISS
Patient Connection:
15 / 22 mm
Dimensions:
9.27 X 7.6 X 4.4 INCHES 236 X 194 X 112 MM
Weight:
2.84 Kg Approx for ATV+ 3.0 Kg Approx for MRI
Patient Valve Dead Space: 8 ml
Cylinder Duration (Aluminum “D” size cylinder containing 415 Litres of oxygen.):
On 100% (a) 2 Litre Mv Setting: (b) 14 Litre Mv
On 60% (a) 2 Litre Mv Setting: (b) 14 Litre Mv
207 minutes 30 minutes 345 minutes 50 minutes
MRI Test Condition Parameters: The CAREvent® MRI has been tested in a 1.5 Tesla MRI environment (unshielded magnet – spatial gradient of < 23Mt/m/sec and a slew rate of 120 T/m/sec at an RF transmitter power of 2000 watts) and a 3.0 Tesla MRI environment (unshielded magnet – spatial gradient of < 40Mt/m/sec and a slew rate of 150 T/m/sec at an RF transmitter power of 8000 watts) during a head phanton spectroscopy test scan with the device located no less than 12 inches from the magnet aperture which would be its normal use position during clinical use.
Note: When tested as indicated above, the spectroscopy traces were not affected in any way by the CAREvent® MRI ventilator, and the CAREvent® MRI ventilator functioned to specification.
CAREvent® ATV/ATV+ & MRI – User Manual
15
1.13. Warranty
WARRANTY
The CAREvent® ATV+ and MRI are manufactured from the finest quality materials.
Each individual part is subject to strict quality control tests to ensure
exceptionally high standards. The manufacturer warrants to the purchaser of
the Automatic Transport Ventilator that its component parts are free from
defects in material and workmanship for a period of two years from the date of
purchase. The manufacturer will replace and/or repair all parts of the
Automatic Transport Ventilator at its option for two years from the date of
purchase at no cost to the purchaser, upon the notification of the defects, in
writing by the purchaser and confirmation of said defect by the manufacturer.
All shipping costs shall be borne by the purchaser. The manufacturer shall be
liable under this warranty only if the Automatic Transport Ventilator and its
parts have been used and serviced in the normal manner described in the
instruction manual. There are no other expressed or implied warranties. This
warranty gives no specific legal rights.
Note: THIS DEVICE MUST BE SERVICED BY AN O-TWO MEDICAL AUTHORIZED SERVICE
CENTER.
2. Preparation for use 2.1. Component List
Having unpacked the ventilator from its shipping carton, use the following
list to ensure that all components have been received:
1. Operating Manual 2. Ventilator 3. Supply Hose. 4. Single Use or Reusable
Patient Circuit with Patient Valve 5. Calibrated Test Lung
Note: If any components are missing from the shipping carton, immediately call
the supplier quoting the packing slip number, your original purchase order
number and the description of the item which is missing.
16
CAREvent® ATV/ATV+ & MRI – User Manual
2.2. Connection of Hoses
The supply hose provided is attached to the input connection on the side of
the ventilator “finger tight” (fig 5).
WARNING
Using a wrench or excessive force in tightening the supply hose may damage the
seal or the thread of the connection.
The patient circuit is attached to the gas outlet on the right hand side of the control module by simply pushing the 22 mm taper over the outlet.
Supply Hose
Patient Circuit
Fig 5. – Connecting the Supply hose and Patient Circuit.
2.3. Pre-use Functional Checks
Along with the contents of the shipping cartons you will require the following
items to enable you to undertake the pre-use functional check:
1. Full oxygen cylinder 2. Oxygen regulator with a 60 PSI 9/16 DISS outlet.
The regulator
must be able to output a minimum of 100 L/min at no less than 40.6 PSI (2.8
Bar). Calibrated Test Lung (supplied with ventilator)
Having connected the supply hose to the regulator, ensure that the ventilator
Manual/AUTO Selector is in the Manual position and turn on the oxygen supply.
Using a mild soap solution, spray the input connection to the ventilator to
check for leaks. If any leak is present, tighten the connection and re-test.
Once no leaks are found, connect the Test Lung to the 15/22 mm patient
connector on the Patient Circuit. Using the selectors on the front panel,
select a frequency of ventilation and a flowrate. Turn the Manual/AUTO Control
to the Auto position and the ventilator will commence cycling.
CAREvent® ATV/ATV+ & MRI – User Manual
17
Testing of the Individual Features of the Ventilator. The following features
can be individually tested during the preuse Functional Check:
1. Airway Over Pressure Alarm, Pressure Gauge Function and Pressure Relief
Adjustment Function
2. Low Input Pressure Alarm 3. Frequency Adjustment 4. Flow Adjustment 5. Air
Mix 6. Manual Ventilation 7. Demand Valve Function and Automatic Circuit Shut
Off 8. CPAP/PEEP Control 9. BSI Alarm
1. Airway Over Pressure Alarm, Pressure Gauge Function and Pressure Relief
Adjustment Function With the ventilator connected to the 60 PSI source, set
the ventilator cycling at 12 BPM with a Minute Volume control setting of 14 L.
Set the adjustable Airway Pressure Control to 20 cmH2O. Remove the Test Lung
from the Patient Connector and occlude the Patient Connector outlet. This will
cause the Airway Over Pressure Alarm to activate and a continuous tone will be
heard during the inspiratory cycle of the ventilator.
To confirm the pressure relief setting, observe the needle on the Airway
Pressure Gauge during the inspiratory cycle and ensure that the needle reaches
the 20 cmH2O mark. Now turn the Adjustable Airway Pressure Selector to the 60
cmH2O position. Repeat the test and observe that the needle reaches the 60
cmH2O mark.
2. Low Input Pressure Alarm
Note: To fully test this function it is necessary to have a supply regulator
with an adjustable output pressure. (Not supplied) Checking of the alarm can
be undertaken by simply slowly closing the cylinder valve.
With the adjustable outlet pressure regulator set with an outlet pressure of
60 PSI, set the ventilator cycling at 12 BPM with a Minute Volume control
setting of 14 L.
18
CAREvent® ATV/ATV+ & MRI – User Manual
As the ventilator cycles, gradually reduce the outlet pressure of the
regulator until you hear the Low Input Pressure Alarm activate. This will be a
slow, mid pitched, pulsed tone. Continue to decrease the regulator outlet
pressure and the tone will gradually slow in frequency until it becomes a
continuous tone. This indicates that the drive gas is now exhausted to the
point where the ventilator will no longer function correctly.
3. Frequency Adjustment With the ventilator connected to the 60 PSI source,
set the ventilator cycling at 8 BPM with a Minute Volume control setting of 6
L. Connect the Test Lung to the Patient Connector and turn the ventilator
Manual/Auto Selector to the Auto position.
After every 5 breaths rotate the frequency control counter clockwise to the
next setting and observe the increasing frequency of ventilation in the Test
Lung. Repeat for every setting. If required, a stopwatch may be used to check
the frequency in each setting.
4. Minute Volume Adjustment With the ventilator connected to the 60 PSI
source, set the ventilator cycling at 18 BPM with a Minute Volume control
setting of 2 L. Connect the Test Lung to the Patient Connector and turn the
ventilator Manual/Auto Selector to the Auto position.
After every 5 breaths rotate the Minute Volume control counter clockwise to
the next setting and observe the increasing expansion of the Test Lung. Repeat
for every setting. The increasing expansion of the lung indicates an increase
in the delivered tidal volume, as the frequency remains constant so the
delivered Minute Volume (tidal volume x frequency) will increase with each
increase in flowrate.
Note: When testing the lower Minute Volume settings using the test lung
provided, there may be insufficient volume delivered with each breath to
attain an airway pressure that will cause the BSI Alarm to be shut off. Under
these circumstances simply depress the BSI Mute Button to silence the alarm.
CAREvent® ATV/ATV+ & MRI – User Manual
19
5. Airmix
Note: To fully test this function a calibrated 0 – 100% oxygen monitor is
required (not supplied).
With the ventilator connected to the 60 PSI medical oxygen source, set the
frequency control to 8 BPM and the Minute Volume Selector set at 14 L, attach
the oxygen monitor to the outlet of the patient connector. Turn the Air Mix
control to the 100% setting and verify that the oxygen concentration, using
the oxygen monitor, is 100%.
Now turn the Air Mix control to the 60% setting and again verify (using the
oxygen monitor) that the oxygen concentration is in the range of 54 to 66%.
6. Manual Ventilation With the ventilator connected to the 60 PSI source turn
the Manual/AUTO Selector to the AUTO position. Set the frequency control to 14
BPM and the Minute Volume Selector to the 6 L position. Allow the ventilator
to cycle automatically for 5 breaths.
Depress the Manual Ventilation Button located on the front of the ventilator.
Hold the button in and gas will flow to the test lung for as long as the
button is depressed.
Release the button and the automatic cycling will restart with a delay of
approximately 20 seconds.
7. Demand Valve Function and Automatic Circuit Shut Off With the ventilator
connected to the 60 PSI source, set the ventilator cycling at 12 BPM with a
Minute Volume control setting of 14 L. Turn the Manual / AUTO Selector to AUTO
and with the ventilator cycling, apply a vacuum to the patient connector
equivalent to a flowrate of 30 L/min for a minimum of 1 second. The demand
valve will provide a flowrate equivalent to that demanded and the automatic
cycling will cease.
Remove the vacuum from the patient connector and check for the automatic
cycling restart after a delay of 5-8 seconds.
Note: The vacuum source can be as simple as a 500 ml calibration syringe.
20
CAREvent® ATV/ATV+ & MRI – User Manual
8. BSI Alarm With the Patient Circuit attached to the ventilator attach a
test lung to the 15/22 mm adapter. Set the frequency control to 10 BPM and the
Minute Volume Selector to the 6 L position. Turn the ventilator to AUTO and
allow it to cycle for 5 breaths. Remove the test lung and time the delay until
the BSI alarm sounds. Depress the silence button and time the delay until the
alarm re-starts.
9. CPAP/PEEP Control With the Patient Circuit attached to the ventilator
attach a test lung to the 15/22 mm adapter. Set the frequency control to 20
BPM and the Minute Volume Select or to the 2 L position. Turn the ventilator
to AUTO and allow it to cycle for 5 breaths. Slowly open the CPAP/PEEP control
and monitor the gauge. As the control is rotated the baseline pressure should
increase.
3. Operating Instructions 3.1. Setting of the Ventilation Parameters
The setting of the ventilation parameters is dependent upon the patient’s
size, condition and the clinical parameters required to be provided to the
patient by the operator. The controls should be set according to the
established protocols to which the operators perform their tasks.
3.2. Frequency/Minute Volume Chart (Fig.6)
Infant (3.33-20 kgs) Child/Teen (11.3-67 kgs) Adult (40-180 kgs) FREQUENCY OF
VENTILATION (BPM)
40 36 32 28 24 20 18 16 14 12 10 8 2.0 .050 .056 .063 .071 .08 .10 .11 .13 .14
.17 .20 .25 2.5 .063 .069 .078 .089 .10 .13 .14 .16 .18 .21 .25 .31 3.0 .075
.083 .094 .107 .13 .15 .17 .19 .21 .25 .30 .38 3.5 .088 .097 .109 .125 .15 .18
.19 .22 .25 .29 .35 .44 4.0 .100 .110 .130 .14 .17 .20 .22 .25 .29 .33 .40 .50
5.0 .13 .14 .16 .18 .21 .25 .28 .31 .36 .42 .50 .63 6.0 .15 .17 .19 .21 .25
.30 .33 .38 .43 .50 .60 .75 7.0 .18 .19 .22 .25 .29 .35 .39 .44 .50 .58 .70
.88 8.0 .20 .22 .25 .29 .33 .40 .44 .50 .57 .67 .80 1.0 10 .25 .28 .31 .36 .42
.50 .56 .63 .71 .83 1.0 1.3 12 .30 .33 .38 .43 .50 .60 .67 .75 .86 1.0 1.2 1.5
14 .35 .39 .44 .50 .58 .70 .78 .88 1.0 1.2 1.4 1.8
TIDAL VOLUMES (LITRES)
Recommended bodyweight range: 5 Kg to 180 Kg (@ 10ml/Kg)
MINUTE VOLUME (LITRES)
CAREvent® ATV/ATV+ & MRI – User Manual
21
This chart gives an easy reference to the delivered tidal volumes provided by
the different combinations of frequency of respiration and minute volume.
3.3. Operation in Extreme Conditions
Operation of the ventilator in environmental conditions outside of those
detailed in this manual may result in a reduction in the ventilator’s
performance. In extreme cold weather a slowing down of the frequency of
ventilation may be seen with a corresponding increase in the delivered tidal
volume. In high temperature environments the effect is not noticeable in terms
of delivered ventilations but may cause excessive wear in the ventilator
components over time.
Operation of the ventilator under supply pressures outside those detailed in
this manual may result in a reduction in the ventilator’s performance. Input
pressures below the minimum stated will cause the low input pressure alarm to
function and will, as the pressure falls to the point where the alarm is
constant, cause the ventilator to cease cycling.
Input pressures higher than that recommended in this manual may result in a
risk of internal component failure if the pressure exceeds three times the
maximum working pressure stated.
4. After Use Cleaning and Testing 4.1. Dismantling and Cleaning
After use turn off the ventilator and disconnect the supply hose from the gas
supply. Disconnect the patient circuit from the gas outlet on the side of the
ventialtor and either (a) discard the circuit if it is designed for single use
or (b) process the reusable circuit following your normal protocols. Disinfect
using a legally marketed, commercially available, hard surface disinfectant
solution which is compatible with the materials in accordance to local
protocols.The ventilator can then be cleaned using a mild soap solution and
warm water and a soft nail brush to remove ingrained contamination.
WARNING
Do not immerse the ventilator in any solutions. Do not use chlorine based
cleaning agents. If the ventilator is accidentally submerged in any liquid it
should be shaken to remove as much of the liquid
as possible and then the ventilator should be returned to the manufacturer for
factory service.
22
CAREvent® ATV/ATV+ & MRI – User Manual
4.1.1. Changing the Air Intake Filter To change the air intake filter, undo
the locking screw on the filter cover, which is located on the rear panel of
the unit. Remove the old filter and discard following local protocols. Replace
the filter with a new one ensuring that the filter is properly located and
then reattach the cover, screwing the cover on finger tight.
Note: It is recommended that the air intake filter is changed after each use
of the ventilator using the 60% air/oxygen mix mode to avoid using a partially
occluded or contaminated filter for the next patient.
Fig. 7 – Rear Panel Layout
4.2. After Use and Monthly Checking
· After Use After each use the ventilator should be cycled to confirm
function. The selectors should be rotated through all positions and all alarm
functions should be checked to ensure function (see section 2.3).
5. Preventive Servicing
· Monthly Checking Every month (or more frequently if local protocols dictate)
the ventilator should be put through a checking procedure to ensure full
function of all features. The checking protocol detailed in section 2.3 “Pre-
use Functional Check” should be followed for this purpose.
· Annual It is recommended that the ventilator is returned to a service centre
(authorized by the manufacturer to undertake service and repair of this
device) for an Annual Preventative Maintenance inspection (more frequently in
high use areas). The inspection will incorporate
CAREvent® ATV/ATV+ & MRI – User Manual
23
a full diagnostic test of all parameters. When complete, a certificate of
compliance will be issued to cover the device for the next service period.
This maintenance inspection, providing it is carried out by an authorized
service centre, will not affect the product warranty.
· Full Service
Level II Service The device shall be returned to the manufacturer or a service
center authorized by the manufacturer for annual Level II Service every year.
Manufacturer Full Service The device shall be returned to O-Two Medical
Technologies for Manufacturer Full Service every 5 years.
Monthly
DESCRIPTION Function Check
PROCEDURE User Manual Chapter 2.3
Annual
Level II Service
Service Manual
Every 5 years
Full Service
Service Manual
CRITERIA Within
specification
SCHEDULE Monthly
BY User
Meet product specifications
Every year
Manufacturer/ Service Center
Meet product specifications
Every 5 years
Manufacturer
24
CAREvent® ATV/ATV+ & MRI – User Manual
6. Troubleshooting Chart
Note: If any of the remedies do not resolve the problem you are experiencing
please contact your nearest Approved Service Centre.
Symptom Ventilator does not function in any mode
Ventilator does not function in automatic mode Insufficient ventilations Low
Input Pressure Alarm sounds intermittently or continuously Automatic circuit
shut-off not activating on demand breathing Breathing System Integrity Alarm
Sounds
Probable Cause Gas supply exhausted
Gas supply not turned on fully
Supply hose not connected Gas pressure regulator faulty Gas supply hose kinked
or cut Manual/AUTO switch in the Manual position. Internally clogged orifice
in the Frequency or Minute Volume controls Other control knobs may be
incorrectly positioned Incorrect control setting selection Other control knobs
may be incorrectly positioned Internal blockage in Minute Volume Control
Pressure relief setting too low
Gas supply low
Cylinder valve not fully opened Gas supply hose
kinked or cut Gas supply regulator faulty Poor patient circuit integrity/
loose connections or faulty patient valve
Poor facemask seal
Patient circuit has become disconnected
Tidal volume being delivered is too low
Remedy Replace cylinder Open cylinder valve Connect supply hose to regulator
outlet Replace pressure
regulator Replace supply hose
Switch to AUTO
Switch selector knobs to a different position Check all selector knob
positions Check selector positions Check selector positions
Switch to another Minute Volume setting
Check selector position Change Cylinder Open Valve fully
Replace supply hose Replace pressure regulator
Check all circuit connections and hose for tears/holes
Re-position facemask Re-connect circuit
Adjust ventilator settings
CAREvent® ATV/ATV+ & MRI – User Manual
25
Symptom Demand flowrate low or not functioning Airway Pressure Alarm
Probable Cause Supply pressure too low
(Low Input Pressure Alarm will sound) Gas supply cylinder low or empty.
Regulator output insufficient
Cylinder valve not fully open Brand of circuit used may be entraining room air
through expiratory port Incorrectly selected airway pressure Airway obstructed
Selected Minute Volume too high/Frequency of ventilation too slow
Remedy Faulty gas supply regulator Replace cylinder Replace regulator Open valve fully Use only O-Two approved circuits Select a higher pressure Correct the obstruction Change the settings selected
7. CAREvent® Accessories
ORDER N0 01CV8015
PART
CAREvent® Deluxe Single Use Transport Ventilation Circuit with PEEP Port
01CV8016-CS
O-Two CAREvent® MRI Ventilation Circuit with PEEP Port and 10 Foot Hose
17MP7010
Single Use PEEP Valve
17MP7327-Cs
Air Intake Filter for CAREvent® ATV+ & MRI
01CV7025
CAREvent® MRI Mounting Bracket
01CV7026
CAREvent® MRI Stand and Mounting Bracket
01CV7030
CAREvent® ATV+ & MRI Ambulance Mounting Bracket
26
CAREvent® ATV/ATV+ & MRI – User Manual
CAREvent® ATV/ATV+ & MRI – User Manual
27
Your Representative is:
O-TWO MEDICAL TECHNOLOGIES INC.
For your nearest Authorized O-Two Distributor In North America call Toll Free
1-800-387-3405
SERIAL N0:
EC REP
MedNet EC-REP IIb GmbH Borkstrasse, 10 48163 Münster, Germany
www.otwo.com
45A Armthorpe Road, Brampton, ON, Canada, L6T 5M4 Telephone: +1 905 792-OTWO
(6896) N.A. Toll Free: +1 800 387 3405
Facsimile: +1 905 799 1339 Email: resuscitation@otwo.com