MARSDEN M-100 Integrated Height Measure for Precision Weighing User Manual

May 15, 2024
MARSDEN

MARSDEN M-100 Integrated Height Measure for Precision Weighing

Specifications

  • Model: M-100
  • Accuracy Class: Class III
  • Capacity/Division: 300kg x 50g / 100g
  • Weight of Scale: Approximately 16kg
  • Units of Measure: Kilograms (kg)
  • Function Keys: Not specified
  • Stabilization Time: Not specified
  • Operating Temperature: Not specified
  • Power Supply: Rechargeable battery pack,non-rechargeable battery pack, or AC adaptor (12V 2A)
  • Indicator Display: Not specified
  • Warranty Duration: Not specified

Product Usage Instructions

Safety Instructions

  • Cleaning: Follow the recommended cleaning procedures in the user manual.
  • Maintenance: Regular maintenance is advised to ensure accurate weighing.
  • Disposing of the Scale: Follow local regulations for proper disposal.
  • Intended Use: Use the scale only for its intended purpose as described in the manual.

Power Supply and Battery
The scale can be powered by a rechargeable or non-rechargeable battery pack or via an AC adaptor. Follow these steps for installing or replacing the battery pack:

  1. Remove the battery housing.
  2. Slide the rechargeable battery pack into the housing, ensuring proper connection.
  3. Place the housing back in the indicator and close the cover securely.

Please take time to read these instructions before starting to use the scale.

Introduction

Thank you for purchasing a Marsden professional medical scale. This is a precision Class III weighing instrument and considerate use will result in many years of accurate weighing. The scale has a maximum load capacity of 300kg which must not be exceeded.

Product Specification

Model M-100
Accuracy Class Class III
Capacity/Division 300kg x 50g<150kg>100g
Weight of Scale Approximately 16kg
Units of Measure Kg
Function Keys ON/OFF, HOLD, TARE, BMI, UNIT, 0-9
Stabilization Time 1-2 Seconds
Operating Temperature 0 to 40˚C
Power Supply Rechargeable battery pack
Indicator Display 2.5cm LCD display with 5 active digits
Warranty Duration 8 years

Safety Instructions

  • Before putting the device into use, please read with care the information given in this user manual, which contains important instructions for proper installation, use and maintenance of the device.

Marsden and/or the manufacturer shall not be liable for damages arising from failure to heed the following instructions:

  • When using electrical components under increased safety requirements, always comply with appropriate regulations.
  • Inappropriate installation/use will render the warranty null and void.
  • Ensure the voltage marked on the power supply unit matches your main supply.
  • This device is designed for use indoors only.
  • Observe the permissible ambient temperatures for use.
  • The device meets the requirements for electromagnetic compatibility. Do not exceed the maximum values specified in the applicable standards.
  • Batteries should be kept away from small children. If swallowed, promptly seek urgent medical assistance.

If you have any problems with this scale, please contact Marsden/your local dealer/your service partner.
If a serious incident occurs concerning this device, it should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established.

Cleaning

  • We recommend using alcohol-based wipes or similar when cleaning the scale.
  • Please do not use corrosive liquids, large amounts of water or high-pressure washers.
  • Always disconnect the scale from the mains power supply before cleaning.

Maintenance

  • The scale does not require any routine maintenance. However, we recommend checking the scale’s accuracy at regular intervals. If any inaccuracies occur, please contact your local dealer or service partner.
  • If you are in the UK, service contracts are available from Marsden to keep your scale accurate and reliable for longer. Call 01709 364296 for more information.

Disposing of the Scale

  • This product should not be treated as regular household waste but should be handed into an electrical/electronic equipment recycling centre.
  • You can obtain further details from your local council, your municipal waste disposal company or from where you purchased the product.
  • Alternatively, you can return this product to Marsden – we will recycle this free of charge,

Intended Use

  • This scale is intended for use to determine the weight of patients, supported by professional personnel and in rooms intended for carrying out healthcare. The weighing value can be read after a stable weighing value has been obtained. Before use, the scale must be checked by an authorised person to ensure it’s in a suitable condition.
  • Device is intended to measure one subject at a time.

Explanation of Graphic Symbols

Power Supply and Low Battery

  • The indicator uses a rechargeable battery pack, a non-rechargeable battery pack, or can be powered from the mains via the AC adaptor.
  • Make sure the battery pack is installed in the battery box of the indicator. Alternatively, plug the AC adaptor (12V 2A) into the port on the side of the indicator.

Installing and Replacing the Battery Pack

  1. Take out the battery housing.
  2. The rechargeable battery pack will slide into, or out of, the housing.
  3. Check that the housing pin is connecting to the right point inside the indicator.
  4. Place the housing back in the back of the indicator and close the battery housing cover.

Switching on the Scale

Switching off the Scale

Setting the Scale to Zero

Hold Function

Setting Auto Hold Function (Optional)

Body Mass Index (BMI) Function

Tare and Pre-Set Tare Functions

Setting the Date

Press HOLD for three seconds to access the time setting mode. The period digit that is flashing can be changed by using the numeric keys. The period to be edited is selected by pressing HOLD.
E.g., To input 25 December 2008, 8:00 a.m.:

Using the Scale with a Printer

  • An optional Marsden external printer (model TP-2100) is available for all models. With the printer fitted, the patient’s weight, height, and BMI result can be printed.
  • Once the person has been weighed and their BMI calculated, simply press PRINT to produce the following ticket:

Connecting the TP-2100 Thermal Printer

Plug the cable to the printer, and then connect its 9D connector to the indicator.

Using the Scale with Wi-Fi/Bluetooth
If your scale has device connectivity, the universal wireless symbol will be on the main indicator display. If your scale does have Wi-Fi or Bluetooth connectivity, we do not supply the software to capture the data from the weighing scale. We do however provide the protocols for you to implement the devices into your own software.

Bluetooth Connection

Wi-Fi Connection

EMC Guidance and Manufacturer’s Declaration

  • Guidance and manufacturer’s declaration – electromagnet emissions.
  • The M-100/M-110 is intended for use in the electromagnetic environment specified below. The customer or user of this scale should ensure that it is used in such environment.

Emission Test| Compliance| Electromagnetic environment- guidance
---|---|---
RF emissions CISPR 11| Group 1| This scale uses RF energy only for its internal function. Therefore, its RF emissions are very low and not likely to cause any interference in nearby

electronic equipment.

RF emissions CISPR 11| Class B| This scale is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC 61000-3-2| Class A
Voltage fluctuations/flicker emissions IEC 61000-3-3| Compliance

Guidance and manufacturer’s declaration – electromagnetic immunity.
The M-100/M-110 is intended for use in the electromagnetic environment specified below. The customer or the user of this scale should ensure that it is used in such an environment.

Guidance and manufacturer’s declaration – electromagnetic immunity.
This scale is intended for use in the electromagnetic environment specified below. The customer or the user of the scale should ensure that it is used in such an environment.

Immunity Test| IEC 60601 test level| Compliance level| Electromagnetic environment- guidance
---|---|---|---

Conducted RF IEC 61000-4-6

|

3 Vrms

150 KHx to 80 MHz

|

3 Vrms

| Portable and mobile RF communications equipment should be used no closer to any part of the scale including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Radiated RF IEC 61000-4-3

|

3 V/m 80 MHz to 2,5 GHz

|

3 V/m

| Recommended separation distance:

d = 1,2 √ P

d = 1,2 √ P 80MHz to 800 MHz d = 2,3 √ P 800MHz to 2,5 GHz

Where P is the maximum output power rating of the transmitter in watts (w) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.

Interference may occur in the vicinity of equipment marked with the following symbol:

---|---|---|---

  • NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
  • NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
    • A) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the scale is used exceeds the application RF compliance level above, the scale should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the scale.
    • B) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distance between portable and mobile RF communications equipment and the M-100/M-110.
This scale is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the scale can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the scale as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter.

W

| Separation distance according to frequency of transmitter m
---|---
150 kHz to 80 MHz

d = 1,2√ P

| 80 MHz to 800 MHz

d = 1,2√ P

| 800 MHz to 2,5

GHz

d = 2,3√ P

0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the maximum output rating of the transmitter in watts (w) according to the transmitter manufacturer.

  • NOTE1) At 80 MHz and 800 MHz, the separation distance for the high frequency range applies.
  • NOTE2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

Error Messages

FAQ

Frequently Asked Questions

  • Q: What should I do if the scale does not turn on?
    • A: Check the battery pack or power supply connections and ensure they are properly installed. If using batteries, check for proper charge or replace them if necessary.
  • Q: How often should I calibrate the scale?
    • A: Calibration frequency may vary based on usage and environmental factors. It is recommended to calibrate the scale regularly according to the user manual guidelines.