Little Doctor LD-60 Blood Pressure Monitor Instruction Manual
- June 16, 2024
- Little Doctor
Table of Contents
- Little Doctor LD-60 Blood Pressure Monitor
- Product Information
- GENERAL INFORMATION
- PARTS AND COMPONENTS
- CUFF PREPARATION
- MEASUREMENT PROCEDURE
- CARE, STORAGE, REPAIR AND DISPOSAL
- WARRANTY
- TECHNICAL SPECIFICATIONS
- COMPLETENESS
- CERTIFICATION AND STATE REGISTRATION
- ADDITIONAL ACCESSORIES
- Contact
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Little Doctor LD-60 Blood Pressure Monitor
Product Information
Specifications
- Models: LD-60, LD-61, LD-70, LD-70NR, LD-71, LD-71A, LD-80
- Type: Mechanical Blood Pressure Monitor
General Information
- The LD Blood Pressure Monitor is a mechanical device used for measuring arterial blood pressure.
- It is important to follow the correct measurement procedures to ensure accurate results.
Frequently Asked Questions
- Q: Can I use the LD Blood Pressure Monitor without consulting my doctor?
- A: It is recommended to consult your doctor before using the device, especially if you are undergoing specific treatments or taking certain medications. Your doctor can provide guidance on proper usage and interpretation of the measurements.
- Q: Can I use the LD Blood Pressure Monitor on my right arm?
- A: Yes, you can use your right arm if measuring on your left arm is difficult. However, please note that the readings may differ slightly between the two arms.
- Q: How tight should I wrap the cuff around my arm?
- A: The cuff should fit snugly on your arm without being too tight or too loose. Improper cuff placement may result in inaccurate readings.
- Q: Can I measure my blood pressure while standing?
- A: It is recommended to measure your blood pressure in a sitting or lying position. If measuring in a sitting position, ensure that your hand with the cuff is positioned at the level of your heart.
GENERAL INFORMATION
This Instruction Manual is designed to assist the user with safe and effective operation of the Device for the measurement of blood pressure (hereinafter – the “Device”). Use this Device according to the rules described in this Manual. Operate the Device only as intended. Do not use the Device for any other purposes. Read and understand the whole Instruction Manual, in particular «RECOMMENDATIONS ON CORRECT MEASUREMENT».
INDICATIONS FOR USE
The device is designed to measure arterial pressure of human according to Korotkov method. The device is recommended for use under conditions of clinics and hospitals, as well as in household conditions as supplement to medical observation. Pressure is measured by means of auscultation of Korotkov’s tones by stethoscope and by taking readings on manometer.
RECOMMENDATIONS ON CORRECT MEASUREMENTS
- Do not use the device without preliminary consultation with Your doctor, if You undergo treatment by hemodialysis or by anticoagulants, antithrombocytes or steroids. Use of device in these cases may cause internal hemorrhage.
- For correct measurement you should know that THE BLOOD PRESSURE IS SUBJECT TO SHARP VARIATIONS EVEN WITHIN THE SHORT TIME INTERVALS. The blood pressure depends on many factors. It is usually lower in summer and higher in winter. The blood pressure varies together with the atmospheric pressure, depends on physical loads, emotional excitement, stresses and dietary regime. Drugs, drinking alcohol and smoking produce significant effect. Even the very procedure of blood pressure measurement in a polyclinic sends the blood pressure high in many people, thus, the blood pressure measured at home often differs from the values received in a polyclinic. As the blood pressure tends to rise at low temperatures, make measurements at an indoor temperature (approximately 20°C). If this Device stayed under a low temperature, keep it for at least 1 hour at an indoor temperature before use, otherwise the measurement result may be incorrect. During a day the difference in readings for healthy people may be 30-50 mmHg of systolic pressure and to 10 mmHg of diastolic pressure. The dependence of the blood pressure on various factors is individual for each person. Accordingly, it is recommended to keep a special book with blood pressure records. ONLY A CERTIFIED DOCTOR USING YOUR RECORDS IS CAPABLE TO ANALYZE THE TENDENCY OF YOUR BLOOD PRESSURE VARIATIONS.
- At cardiovascular and some other diseases requiring blood pressure monitoring make measurements in the hours fi xed by your attending doctor. REMEMBER THAT THE DIAGNOSTIC AND ANY TREATMENT OF HYPERTENSION MAY BE CONDUCTED ONLY BY A CERTIFIED DOCTOR ON THE BASIS OF BLOOD PRESSURE VALUES OBTAINED BY THIS DOCTOR. TAKING OF DRUGS AND THEIR DOSES SHOULD BE PRESCRIBED ONLY BY YOUR ATTENDING DOCTOR.
- KEEP QUIET DURING A MEASUREMENT TO OBTAIN THE ACCURATE VALUES OF YOUR BLOOD PRESSURE WITH THE ELECTRONIC DEVICE. Measure your blood pressure in the calm and comfortable conditions at the indoor temperature. No eating an hour before measurement; no smoking, taking tonic agents, alcohol 1.5-2 hours before measurement.
- The accuracy of blood pressure measurement depends on whether the cuff matches the size of your arm. THE CUFF SHOULD NOT BE TOO SMALL OR TOO LARGE.
- Wait 3 minutes between measurements for the blood to restore its circulation. However, the persons with prominent atherosclerosis due to considerable loss of vascular elasticity may need to increase the wait time between measurements (10-15 minutes). This also refers to the patients suff ering for long from diabetes. For more accurate determination of blood pressure it is recommended to make a series of 3 consecutive measurements and to use the average value.
PARTS AND COMPONENTS
- For completeness of set see section COMPLETENESS.
CUFF PREPARATION
- Insert the cuff end for about 5 cm into a metal ring as shown in the fi g. 3 (if cuff is with ring).
- Apply the cuff to your left upper arm so that the air tube is directed to your palm (fi g. 4). If the measurement on your left arm is difficult, you may use your right arm. In this case remember that the readings may differ by 5-10 mmHg and even more.
- Wrap the cuff around your upper arm so that the bottom of the cuff is approximately 2-3 cm above your elbow. The sign “ARTERY” should be over the arm artery as shown in the fig. 5.
- Fix the cuff so that it fi ts tightly to the arm, but see that it is not overtight. Too tight or too free placement of the cuff may give inaccurate readings.
- On the fi xed cuff the sign «INDEX» should point to the area «NORMAL» (fig 7). It means that the cuff is chosen correctly and fits the size of your upper arm. If the sign points to the area marked « » or to the left, the cuff is too small and the readings will be higher. If the sign points to the area marked « » or to the right, the cuff is too large and the readings will be lower.
- If the arm has a conic form, the cuff should be put on with a spiral movement as shown in the fig. 8.
- If the rolled-up sleeve squeezes the arm interfering with free blood flow the Device may give inaccurate figures not corresponding to your actual blood pressure (fig. 9).
- Place the stethoscope head so that it was located in depression, somewhat higher than elbow bend (Fig. 4). If You use the device with a built-in stethoscope (for models LD-60 and LD-71A), to reach better auscultation of pulse on full hand, the cuff may be turned round the hand approximately by quarter of circle (60-90˚) in such way that stethoscope head would be positioned on inner side of hand (more close to the body).
- Pressure shall be measured in sitting or lying position of human. IN SITTING POSITION WATCH THAT PART OF HAND WITH CUFF WOULD BE LOCATED AT THE LEVEL OF HEART, AND HAND WOULD BE FREELY LOCATED ON THE TABLE AND WOULD NOT MOVE.
MEASUREMENT PROCEDURE
- Insert metal tubes of stethoscope into ears. Close air valve on air pump, having turned it clockwise. Squeezing air pump, pump up the cuff , listening to pulse by stethoscope. After You stopped to listen to pulse, pump up the cuff by 30 mm Hg more.
- Slowly opening air valve, turning it counterclockwise, bleed down pressure in the cuff . Watch that pressure in the cuff would fall down at speed 2-4 mm Hg per second. This is necessary to get accurate result.
- As soon as You hear weak pulse beats, memorize manometer reading. This is Your systolic (upper) arterial pressure.
- Pressure in the cuff is continuing falling down at the same speed (2-4 mm Hg per second). You are continuing to listen to pulse. Sounds which You hear will change. Unlike the fi rst beats, they will become softer, resembling rustling. At the moment when You practically stop seize pulse, memorize manometer reading. This is Your diastolic (lower) arterial pressure.
CARE, STORAGE, REPAIR AND DISPOSAL
- Keep this Device from exposure to higher humidity, direct sunlight, shocks.
- Do not keep and use this Device near heating installations and open fire.
- Keep the Device clean and protect it from dust.
- Contact of device with aggressive solutions is not allowed.
- Protect the arm cuff and air tubes from contacting on sharp things.
- This Device does not contain special controls to adjust the measurement accuracy. Independent opening of manometer is prohibited. Repair the Device only in authorized organizations.
- Service life of device is indicated in section TECHNICAL SPECIFICATIONS. Service life is determined from the moment of commodity delivery to the customer. On expiration of the warranted service life apply from time to time to authorized repair organizations to check the technical condition of the Device.
- Dispose of the Device and its components according to the application local regulations. No special requirements to disposal of this Device are defined by the manufacturer.
- The arm cuff may withstand multiple sanitary treatments. The internal tissue surface of the arm cuff (contacting on arm) may be cleaned with cotton ball soaked in 3%-solution of hydrogen peroxide. At long use the partial color fading of the tissue coating of the arm cuff is possible. Washing and ironing of the arm cuff are not allowed.
WARRANTY
- The following LD product is covered by warranty for the period specifi ed in the warranty card.
- The warranty period for the arm cuff is 12 months from the date of sale.
- The warranty liabilities are contained in the warranty card given at the sale of this Device to a purchaser.
- The addresses of organizations for warranty maintenance are given in the warranty card.
TECHNICAL SPECIFICATIONS
Pressure measurement range, mmHg | from 0 to 300 (pressure in an arm cuff ) |
---|---|
Pressure indication range, mmHg | from 0 to 300 |
Range of admissible absolute error at measure- ment of air pressure in an arm
cuff, mmHg| ±3
Operation conditions: Temperature, оС Relative humidity, % Rh| from 10 to 40
85 and lower
Storage and transportation conditions: Temperature, оС
Relative humidity, % Rh
| from minus 34 to 65
85 and lower
Weight (without package, case):
model LD-70 / LD-70NR, g model LD-71, LD-71A, g model LD-60 / LD-61, g model LD-80, g
| 264/237
328
390/309
351
Dimensions (package), mm| 115 x 185 x 75
Service life of device (without taking into account pneumatic chamber and air
pump), years from the moment of its delivery to the customer Service life of
pneumatic chamber and air pump, years from delivery to the customer|
7
3
Year of manufacture| Year the manufacture is given in the bottom of the Device
body in a serial number after symbols “AA”.
Country of manufacture| PRC
CUFF SIZE OF THE DEVICE
modification LD-70 modification LD-71 modification LD-71A| adult’s| for upper arm
circle from 25 to 36 cm
---|---|---
modification LD-70NR| adult’s| for upper arm circle from 25 to 40 cm
modification LD-60| big adult’s| for upper arm circle from 33 to 46 cm
modification LD-61| child’s| for upper arm circle from 18 to 26 cm
modification LD-80| for newborn infants for infants child’s| for upper arm
circle from 7 to 12 cm for upper arm circle from 11 to 19 cm for upper arm
circle from 18 to 26 cm
COMPLETENESS
Name of part| Composition
Manometer LD-S013
| LD-60
1
| LD-61
1
| LD-70
1
| Quantity
LD-70NR
1
| LD-71
1
| LD-71A
1
| LD-80
1
---|---|---|---|---|---|---|---|---
Manometer
Adult’s cuff with ring| Cuff N2AR Pneumatic chamber LD-S02A| –| –| 1| –| 1| 1| –
Adult’s cuff with ring| Cuff N2A Pneumatic chamber LD-S02A| –| –| –| 1| –| –| –
Big adult’s cuff with ring| Cuff N2LR Pneumatic chamber LD-S02L| 1| –| –| –| –|
–| –
Child’s cuff| Cuff C2C Pneumatic chamber LD-S02C| –| –| –| –| –| –| 1
Child’s cuff| Cuff N2C Pneumatic chamber LD-S02C| –| 1| –| –| –| –| –
Cuff for newborn infants| Cuff C2N Pneumatic chamber LD-S02N| –| –| –| –| –| –|
1
Cuff for infants| Cuff C2I Pneumatic chamber LD-S02I| –| –| –| –| –| –| 1
Stethoscope with built-in head into cuff| LD Prof-Plus| 1| –| –| –| –| 1| –
Stethoscope| LD Prof-Plus| –| 1| –| –| 1| –| –
Air pump assembly| Air pump LD-S014 Air valve LD-S015
Non-return valve LD-S016
| 1| 1| 1| 1| 1| 1| 1
Storage Case| Storage Case LD-S059| 1| 1| 1| 1| 1| 1| 1
Instruction Manual| Instruction Manual| 1| 1| 1| 1| 1| 1| 1
Warranty Card| Warranty Card| 1| 1| 1| 1| 1| 1| 1
Package| Package| 1| 1| 1| 1| 1| 1| 1
CERTIFICATION AND STATE REGISTRATION
This Device manufacturing is certifi ed according to international standard ISO 13485:2003.
Complaints and requests should be addressed to: Little Doctor Europe Sp. z
o.o. 57G Zawila Street, 30-390, Krakow, Poland
Service phone: +48 12 2684748, 2684749
Manufactured under control and for Little Doctor International (S) Pte. Ltd.,
35 Selegie Road #09-02 Parklane Shopping Mall, Singapore 188307, Singapore.
Postal address: Yishun Central P.O. Box 9293 Singapore 917699.
Manufacturer: Little Doctor Electronic (Nantong) Co. Ltd., No.8, Tongxing
Road Economic & Technical Development Area, Nantong 226010, Jiangsu, PEOPLE’S
REPUBLIC OF CHINA
Distributor in Europe
- Little Doctor Europe Sp. z o.o. 57G Zawila Street, 30-390, Kraków, Poland
- Sales Office phone: +48 12 2684746, 12 2684747, fax: +48 12 268 47 53
- E-mail: biuro@littledoctor.pl
- www.LittleDoctor.pl
Authorized Representative in the EU:
- Little Doctor Europe Sp. z o.o.
- 57G Zawila Street Krakow 30-390 Poland
ADDITIONAL ACCESSORIES
Cuffs “LD-Cuff ” of the following sizes may be acquired in addition to available cuffs:
- TPU – Thermoplastic polyurethane.
** Metal fixing ring on cuff for convenient independent putting on of cuff.
Contact
LITTLE DOCTOR INTERNATIONAL (S) PTE. LTD.
- Yishun Central P.O. Box 9293 Singapore 917699,
- Fax: 65-62342197
- E-mail: ld@singaporemail.com
EC/REP
- Little Doctor Europe Sp. z o.o. 57G Zawila Street Krakow 30-390 Poland
® Registered trade marks of Little Doctor International (S) Pte. Ltd.
© Little Doctor International (S) Pte. Ltd., 2007-2016
References
- Little Doctor :: Digital Blood Pressure Monitors, Sphygmomanometers, Stethoscopes, Nebulizers, Thermometers
- Manual-Hub.com – Free PDF manuals!
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>