ALLmeter ALL011 6000 Counts Pen Type Digital Multimeter Intelligent Voltage Tester Instruction Manual

June 16, 2024
ALLmeter

ALLmeter ALL011 6000 Counts Pen Type Digital Multimeter Intelligent Voltage Tester

ALLmeter ALL011 6000 Counts Pen Type Digital Multimeter Intelligent Voltage
Tester

Overview

6000 counts is a pocket 3 5/6 true effective value, pen type intelligent multimeter. It does not need to turn the dial to select the function. According to the difference of input voltage/resistance/, the instrument will automatically identify and measure.
The machine has stable performance, high accuracy, high reliability, clear reading and overload protection function. Driven by a AAA 1.5V battery, the instrument uses a large screen LCD display with boost power supply. Even at the edge of a 0.8V low battery, the backlight and the ultra-high brightness of the flashlight can be guaranteed. The meter is portable and is extremely popular with users. This series of instruments can automatically identify DC voltage and AC voltage, resistance, without any switching, and can also manually switch to measure diodes, Non-contact voltage measurement and zero- fire line measurement are excellent tools and meters, and are ideal tools for laboratories, factories, radio lovers and families.

Safety matters

The design of this series of instruments conforms to IEC1010 (safety standards issued by the International Electrotechnical Commission).

Please read the safety precautions before use:

  1. When measuring the voltage, do not input the limit voltage that exceeds the effective value of DC 600V or AC 600V;
  2. The voltage below 36V is the safe voltage;
  3. When changing the function and range, the probe should leave the test point;
  4. Select the correct function and range, and beware of wrong operation. Although this series of instruments has full range protection function, please pay more attention for safety
  5. The safety symbol indicates that
    ” ” dangerous voltage exists,
    ” ” grounding,
    ” ” double insulation,
    ” ” The operator must refer to the instructions,
    ” ” low voltage symbol.

Characteristic

General characteristics

Display mode: LCD;
Maximum display: 5999 (3 5/6) bit automatic polarity display;
Measurement method: double integral AID conversion;
Sampling rate: about 3 times per second;
Over-range display: “OL” is displayed at the highest position;
Low voltage display: ” ” symbol appears;
Working environment: (0-40) t:, relative humidity < 80%;
Power supply: AAA 1 • 1.5V battery;
Volume (size): 170 x twenty-four x 21 mm (length x wide x High);
Weight : about 50g (including 1.5V battery);
Accessories: one operation manual, one certificate of conformity, one outer packing box, one pair of pens, and one AAA1 .5V battery,

Technical characteristics

Accuracy: ± (a% of reading + the lowest significant digit), guaranteed accuracy ambient temperature: (23 ± 5) C, relative humidity < 75%, calibration warranty period is one year from the dale of delivery.

Performance (note “ ” indicates that this table has this function)

function

|
---|---
DCV|
ACV|
resistance/diode/On-off test|
Non-contact phase sequence measurement type B|
Color display type B|
Black and white display type A|
NCV|
Zero line/live line test|
Full unit symbol|
Backlight manual/automatic off|
True RMS measurement|
Temperature (ºC/ºF ) normal temperature display|
Flashlight lighting|

Description of operation panel

  1. Test pen tip: voltage, resistance, positive end test point of diode;
  2. Input end test protective adhesive.
  3. Flashlight;
  4. Signal indicator;
  5. LCD display;
  6. Power and function selection key SELECT {long press for power on and off, short press for manual DC voltage/AC voltage respectively/Diode);
  7. Lock the HOLD key; {If the display LCD is black and white, press and hold to turn the backlight on and off)
  8. Electric field induction measurement/ zero-fire line measurement:
  9. Pen hanging;
  10. Measure the negative input COM;
    Description of operation panel
Technical indicators

DC voltage/AC voltage automatic scanning test (DCV/ACV)

Accuracy

|

6000 counts

|

Resolution

---|---|---

Range

DC/AC6V| ± (0.5%+4)| 0.001V
DC/AC60V| 0.01V
DC600V/ AC600V| ± (1.2%+10)| 0.1V

Input impedance: 10M O; Overload protection: true RMS measurement, frequency response between 50Hz and 800Hz,600V AC peak.

The specific operation is as follows:

  1. Press and hold POWER for more than 2s, and the power-on display will be “AUTO”.
  2. Insert the black pen into the “COM” tail socket, and the positive pole is the front pen lip; The test pen lip is in reliable contact with the measured point.
  3. When the measured voltage between the input port “COM” and “pen lip” is greater than 0.8V, regardless of the AC voltage or the DC voltage, the instrument will compare the DC component and the AC component, lake its larger component signal, and then automatically switch between 6V/60V/600V DC according to the size of the measured value, and automatically switch  between 6V/60V/600V AC, and then display the measured value on the LCD.

Note:

  1. The input voltage shall not exceed DC600V or AC600V. If it exceeds, ii may damage the instrument circuit; In case of high voltage circuit, pay special attention to avoid electric shock;
  2. After completing all measurement operations, disconnect the probe from the circuit under teslo

resistance (Ω)

Accuracy 6000 counts Res o lu t ion

Range
6kΩ| ± ( 1.8% + 10 )| 1Ω
60kΩ| ± ( 1. 2 %+10 )| 10Ω
600kΩ| 100Ω
6MΩ| 1kΩ
60MΩ| ± ( 5%+ 2 0 )| 10kΩ

Input impedance: 10M Ω; Overload protection: DC600V or 600V AC peako

The specific operation is as follows:

  1. The boot display is “AUTO”
  2. Insert the black pen into the “COM” tail socket, and the positive pole is the front pen lip; The test pen tip is in reliable contact with the measured point.
  3. If the measured resistance at both ends of the probe is less than 50 Ω, the buzzer will emit a continuous sound.
  4. If the closed circuit resistance is measured, the resistance at both ends of the resistance to be measured must be discharged. Otherwise, if the voltage in the circuit is greater than 0.8V, the instrument will mistakenly consider ii as a voltage measurement and enter the voltage measurement mode.
  5. Input the resistance measurement value between the input port “COM” and “pen tip”. The instrument will automatically switch between 6k Ω/60k ΩI 600k Ω/6M Ω/60M Ω according to the size of the resistance measurement value, and then display the measured value on the LCD. be careful:
  6. When measuring low resistance, the probe will bring internal resistance. In order to obtain accurate readings, you can first record the short circuit value of the probe, and subtract the value of the short circuit from the measured reading;
  7. When measuring the online resistance, all power supplies of the circuit under test must be turned off and all capacitors must be fully discharged to ensure the correctness of the measured value;
Range Display value Test conditions

AUTO

|

Diode forward voltage drop

| The forward DC current is about 0.5mA,and the open-circuit voltage is about3V,

NCV measure

The operation is as follows:
The power-on status is automatic scanning; Trigger the “NCV/LIVE” key; Enter electric field measurement EF/zero fire/measurement respectively, and cycle between automatic/EF/LIVE/ NCV measurement: trigger the “NCV/LIVE” key;
When entering the EF measurement, the LCD displays “EF”. When the pen lip is close lo the power lest point, the measured frequency is 50Hz/60Hz. According lo the strength of the signal, the LCD will display different segments —- and the buzzer will also emit different sounds. At the same time, the indicator light will also emit red light according to the strength of the signal. LIVE measurement: trigger the “NCV/LIVE/” key twice;
When entering LIVE measurement, the LCD will display “LIVE”.
When the pen lip reliably contacts the test point of the live wire, the LCD will display—-, the buzzer will emit a continuous sound, and the indicator light will emit a red light.

Automatic switch on/off

When the instrument stops using for about 15 minutes, the instrument will automatically power off and enter the sleep state;
To restart the power supply, press and hold the “power” key for more than 2 seconds.

Troubleshooting

If your instrument does not work properly, the following methods can help you quickly solve general problems. If the fault still cannot be eliminated, Please contact the service center or dealer.

Symptoms Inspection position and method
Not displayed The power supply is not connected

Replace the battery
Large resistance display error| The pen is not in good contact

This manual is subject to change without notice;
The contents of this manual are considered to be correct. If the user finds any errors or omissions, please contact the manufacturer;
The company will not bear the accidents and hazards caused by user’s mis operation;
The functions described in this manual are not used as reasons for using the product for special purposes.

Caution

Please turn off the item when it is not in use in order to prolong its working life.

Warranty

Thanks for using all meter product, all all meter product comes with warranty of 18 months from the day of purchase, if you need any support, please feel free to contact us.

Contact

Email: LivehomeShop@hotmail.com

Symbols

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals