new art laboratorices NADAL® SARS-CoV-2/Influenza
A+B/RSV Ag Test
(test cassette)
REF 243502N-20
SARS-CoV-2 Influenza A+B RSV Ag Test
nal von minden GmbH
Carl-Zeiss-Strasse 12
47445 Moers
Germany
Moers
Tel: +49 (2841) 99820-0
Fax: +49 (2841) 99820-1
Regensburg
Tel: +49 941 29010-0
Fax: +49 941 29010-50
www.nal-vonminden.com
info@nal-vonminden.com
Directors:
Sandra von Minden
Roland Meißner
Thomas Zander
Commercial reg. Kleve
HRB 5679
Steuer-Nr. 244/133/00130
UST-ID-Nr. DE 189 016 086
Version 1.0, 2022-05-09
Intended Use
The NADAL® SARS-CoV-2/Influenza A+B/RSV Ag Test is a rapid chromatographic immunoassay for the qualitative detection of SARS-CoV-2, influenza A and B as well as respiratory syncytial virus (RSV) antigens in nasopharyngeal swab specimens from individuals with a suspected SARS-CoV-2/influenza/RSV infection in conjunction with clinical presentation and the results of other laboratory tests. The NADAL® SARS-CoV2/Influenza A+B/RSV Ag Test is designed for in-vitro diagnostic and professional use only. It is intended for use in clinical laboratories and by healthcare professionals for point-of-care testing only. Not for home-use.
Introduction and Clinical Significance
COVID-19 (coronavirus disease) is a disease caused by the SARS- CoV-2 virus.
It is highly contagious and spreads quickly. COVID19 most often causes
respiratory symptoms that can feel muchlike a cold, the flu, or pneumonia.
COVID-19 may affect morethan the lungs and respiratory system.
Influenza is an acute viral infection of the upper or lower respiratory tract
marked by fever, chills, and a generalised feeling of weakness and pain in the
muscles, along with varying degrees of soreness in the head and abdomen.
Respiratory syncytial virus, or RSV, is a common respiratory virus that
usually causes mild, cold-like symptoms. Most people recover in a 1-2 weeks,
but RSV can be serious. Infants and older adults are more likely to develop
severe RSV symptoms and more likely to require hospitalisation. Vaccines are
available to protect older adults from a severe course of RSV. Preventative
options are also available to protect infants and young children from severe
RSV symptoms.
Test Principle
The NADAL® SARS-CoV-2 Ag Test has one line of anti-SARS-CoV- 2 antibodies in
the test line region ‘T’ and one line of antimouse IgG antibodies in the
control line region ‘C’. When an extracted specimen is added to the specimen
well, it reacts with the labeled antibodies to form complexes. The mixture
then migrates along the membrane by capillary action and interacts with the
precoated anti-SARS-CoV-2 antibodies in the test line region ‘T’. If the
specimen contains SARS-CoV-2 antigens, a red test line ‘T’ will appear,
indicating a positive SARS-CoV-2 antigen result. Otherwise, the test result is
to be considered negative. The test cassette also contains a control line ‘C’,
which should turn red for all valid tests. If the quality control line ‘C’
does not appear, the test result is to be considered invalid, even if the test
line ‘T’ appears.
The NADAL® Influenza A+B Ag Test has one line of anti-influenza A antibodies
in the test line region ‘A’, one line of anti-influenza B antibodies in the
test line region ‘B’ and one line of antimouse IgG antibodies in the control
line region ‘C’. When an extracted specimen is added to the specimen well, it
reacts with the labeled antibodies to form complexes. The mixture then
migrates along the membrane by capillary action and interacts with the
precoated anti-influenza A antibodies and antiinfluenza B antibodies in the
corresponding test line regions. If the specimen contains influenza A or
influenza B antigens, a red test line will appear, indicating the presence of
influenza A or influenza B antigens. Otherwise, the test result is to be
considered negative. The test cassette also contains a control line ‘C’ which
should turn red for all valid tests. If the control line ‘C’ does not appear,
the test result is to be considered invalid, even if the test line appears.
The NADAL® RSV Ag Test has one line of anti-RSV antibodies in the test line
region ‘T’ and one line of anti-mouse IgG antibodies in the control line
region ‘C’ When an extracted specimen is added to the specimen well, it reacts
with the labeled antibodies to form complexes, the mixture then migrates along
the membrane by capillary action and interacts with the coated anti-RSV
antibodies in the test line region ‘T’. If the specimen contains RSV antigens,
a red test line will appear, indicating the presence of RSV antigens.
Otherwise, the test result is to be considered negative. The test cassette
also contains a control line ‘C’ which should turn red for all valid tests. If
the control line ‘C’ does not appear, the test result is to be considered
invalid, even if the test line appears.
Reagents and Materials Supplied
-
20 NADAL® SARS-CoV-2/Influenza A+B/RSV Ag test cassettes
-
Additional material provided in accordance with 93/42/EEC:
Due to possible supply shortages of accessory medical products, the swab manufacturer might change. Therefore, the supplied swabs are from one of the manufacturers listed below.
a) 20 sterile swabs, CE 0197
CITOTEST LABWARE MANUFACTURING CO., LTD No. 339 Beihai West Road, Haimen, 226100 Jiangsu, P.R. China (authorised EU representative: WellKang Ltd, Enterprise Hub, NW Business Complex, 1 Beraghmore Rd., Derry, BT48 8SE, Northern Ireland)
b) 20 sterile swabs, CE 0197
Hangzhou Yiguoren Biotechnology Co. Ltd., Room
402, Building 2, No. 2628, Yuhangtang Road, Cangqian Street, Yuhang District, Hangzhou, 311121 Zhejiang, China (authorised EU representative: Zoustech S.L., Paseo de la Castellana 141, 28049 Madrid, Spain) -
20 extraction tubes containing extraction solution (approximately 600 μL each)*
-
20 dropper caps
-
1 reagent holder
-
1 package insert
*contains the following preservative: ProClin™ 300: <0.03%.
No hazard labelling for ProClin™ 300 is required according to Regulation (EC) Nº 1272/2008 CLP. Concentrations are below the exemption threshold of <0.03%.
Additional Materials Required
- Timer
Storage & Stability
Store the test kit in a cool, dry place between 2-30°C. Keep away from light. Exposure to temperatures and/or humidity outside the specified conditions may cause inaccurate results. Do not freeze. Use the test kit at temperatures between 1530°C. Use the test kit between 10-90% humidity. Do not use the test kit beyond the expiration date (printed on the foil pouch and box).
Note: All expiration dates are printed in Year-Month-Dayformat. 2022-06-18 indicates June 18, 2022.
Warnings and Precautions
- For in-vitro diagnostic use only.
- This package insert must be read completely before performing the test. Failure to follow the instructions provided in the package insert may lead to inaccurate test results.
- Do not open the sealed pouch unless ready to conduct the assay.
- Do not use expired tests.
- Bring all reagents to room temperature (15-30°C) before use.
- Do not use the components from any other type of test kits as a substitute for the components in this test kit.
- Wear protective clothing and disposable gloves while handling the kit reagents and clinical specimens. Wash hands thoroughly after performing the test.
- Do not smoke, drink, or eat in areas where specimens or kit reagents are being handled.
- Dispose of all specimens and materials used to perform the test as biohazardous waste.
Specimen Collection and Preparation
Consider any materials of human origin as infectious and handle them using standard biosafety procedures.
Specimen Collection:
Use the nasopharyngeal swab supplied in the test kit. It is important to
obtain as much secretion as possible. Therefore, to collect a nasopharyngeal
swab sample, carefully insert the sterile swab into the nostril that presents
the most secretions under visual inspection. Keep the swab near the floor of
the nasal septum while gently pushing it into the posterior nasopharynx.
Rotate the swab several times, and then remove it from the nasopharynx.
Test Procedure
Bring tests and extraction solutions to room temperature (15-30°C) prior to testing.
- Open the extraction solution (in the sealed tube).
- Collect a specimen as outlined in the section ‘Specimen Collection and Preparation’.
- Insert the swab with the collected specimen into the extraction tube filled with extraction solution. Roll the swab 5 times while pressing the tip against the bottom and wall of the extraction tube.
- Remove the swab while squeezing the wall of the tube to extract the liquid from the swab. Try to expunge as much liquid as possible. Dispose of the used swab as biohazardous waste.
- Attach a dropper cap to the extraction tube.
- Take out a test cassette from its sealed foil pouch and put it on a clean and level surface.
- Apply 3 drops of the extracted specimen into each specimen well. Please avoid bubbles forming during application.
- Read the test result after 15 minutes. Do not read results after more than 20 minutes.
Result Interpretation
For SARS-CoV-2:
Positive result: Both the control line ‘C’and the test line ‘T’ appear.
Negative result:
Only the control line ‘C’ appears, no line appears in the test line region
‘T’.
Invalid result:
The control line ‘C’ fails to appear, indicating the test is invalid no matter
whether the test line ‘T’ appears or not. Collect a new specimen and perform
another test with a new test cassette.
For Influenza A+B:
Positive results:
Positive for influenza A antigens:
Both the control line ‘C’ and the influenza A test line ‘A’ appear, while the influenza B test line ‘B’ does not appear.
Positive for influenza B antigens:
Both the control line ‘C’ and the influenza B test line ‘B’ appear, while the
influenza A test line ‘A’ does not appear.
Positive for influenza A and B antigens:
All 3 lines appear, including the control line ‘C’ as well as the influenza A
and influenza B test lines ‘A’ and ‘B’.
Negative result:
Only the control line ‘C’ appears, no lines appear in the influenza A and
influenza B test line regions ‘A’ and ‘B’. This indicates the test result is
negative for both influenza A and influenza B antigens.
Invalid results:
The control line ‘C’ fails to appear, indicating that the test is invalid,
nomatter whether the influenza A orinfluenza B test line appears or not.
Collect a new specimen and perform another test using a new test cassette.
For RSV:
Positive result: Both the control line ‘C’ and the test line ‘T’ appear.
Negative result:
Only the control line ‘C’ appears, no line appears in the test line region
‘T’.
Invalid result:
The control line ‘C’ fails to appear, indicating that the test is invalid, no
matter whether the test line ‘T’ appears or not. Collect a new specimen and
perform another test with a new test cassette.
Quality Control
Internal procedural controls are included in the test. A coloured line appearing in the control region ‘C’ is an internal procedural control. This procedural control line indicates that sufficient flow has occurred, and the functional integrity of the test cassette has been maintained. Control standards are not supplied with this test kit; however, it is recommended that positive and negative controls be tested as good laboratory practice to confirm the test procedure and to verify proper test performance.
Limitations
- The NADAL® SARS-CoV-2/Influenza A+B/RSV Ag Test is for in-vitro diagnostic use only and should not be re-used.
- The used test should be treated as potentially infectious material and be disposed of properly.
- The test kit should be kept away from direct sunlight, moisture and heat.
- Please check whether the test kit has any damage and check the expiry date before use.
- The sample volume may affect the accuracy of the test result. Inaccurate sample volume may cause a false positive or negative result.
- Test results must be evaluated in conjunction with other clinical data available to the physician.
- Positive test results do not rule out co-infections with other pathogens.
- Positive and negative predictive values are highly dependent on prevalence. The local prevalence should be taken into consideration when interpreting diagnostic test results.
- Please be very careful when collecting nasopharyngeal swab specimens from children.
- Components from different batches are not to be used in combination.
Performance Characteristics
Clinical performance
Diagnostic sensitivity and specificity
A total of 362 specimens were tested using the NADAL® SARSCoV-2/Influenza
A+B/RSV Ag Test. These specimens were obtained by collecting nasopharyngeal
swabs from symptomatic patients. The performance of the NADAL®
SARSCoV-2/Influenza A+B / RSV Ag Test was compared to a commercialised
molecular assay.
Table 1: SARS-CoV-2 performance of the NADAL® SARS-CoV2/Influenza A+B/RSV
Ag Test as compared to a PCR Test.
| PCR
---|---
NADAL®| | Positive| Negative| Total
SARS-CoV2/Influenza| Positive| 102| 1| 103
A+B/RSV Ag| Negative| 4| 255| 259
Test (for SARS-CoV-2)| Total| 106| 256| 362
Diagnostic sensitivity: 96.23% (102/106, 95%CI, 90.70% – 98.52%)
Diagnostic specificity: 99.61% (255/256, 95%CI, 97.82% – 99.93%)
Overall agreement: 98.62% (357/362, 95%CI, 96.81% – 99.41%)
Table 2: Influenza A performance of the NADAL® SARS-CoV2/Influenza A+B/RSV Ag Test as compared to a PCR Test.
| PCR
---|---
NADAL®| | Positive| Negative| Total
SARS-CoV2/Influenza| Positive| 55| 4| 59
A+B/RSV Ag| Negative| 3| 300| 303
Test (for SARS-CoV-2)| Total| 58| 304| 362
Diagnostic sensitivity: 94.83% (55/58, 95%CI, 85.86% – 98.23%)
Diagnostic specificity: 98.68% (300/304, 95%CI, 96.67% – 99.49%)
Overall agreement: 98.07% (355/362, 95%CI, 96.06% – 99.06%)
Table 3: Influenza B performance of the NADAL® SARS-CoV2/Influenza A+B/RSV Ag Test as compared to a PCR Test.
| PCR
---|---
NADAL®| | Positive| Negative| Total
SARS-CoV2/Influenza| Positive| 32| 5| 37
A+B/RSV Ag| Negative| 1| 324| 325
Test (for SARS-CoV-2)| Total| 33| 329| 362
Diagnostic sensitivity: 96.97% (32/33, 95%CI, 84.68% – 99.46%)
Diagnostic specificity: 98.48% (324/329, 95%CI, 96.49% – 99.35%)
Overall agreement: 98.34% (356/362, 95%CI, 96.43% – 99.24%)
Table 4: RSV performance of the NADAL® SARS-CoV-2/Influenza A+B/RSV Ag Test as compared to a PCR Test.
| PCR
---|---
NADAL®| | Positive| Negative| Total
SARS-CoV2/Influenza| Positive| 111| 2| 113
A+B/RSV Ag| Negative| 2| 247| 249
Test (for SARS-CoV-2)| Total| 113| 249| 362
Diagnostic sensitivity: 98.23% (111/113, 95%CI, 93.78% – 99.51%)
Diagnostic specificity: 99.20% (247/249, 95%CI, 97.12% – 99.78%)
Overall agreement: 98.90% (358/362, 95%CI, 97.19% – 99.57%)
Analytical performance
Detection limit
Virus type/ subtype | Concentration |
---|---|
SARS-CoV-2 | 75.5 TCID50 /mL |
Influenza A (H1N1) | 1320 TCID50 /mL |
Influenza A (H3N2) | 568 TCID50 /mL |
Influenza B (Victoria lineage) | 1540 TCID50 /mL |
Influenza B (Yamagata lineage) | 2180 TCID50 /mL |
RSV | 93.6 TCID50 /mL |
Measuring range
No prozone effect was observed when testing SARS-CoV-2 (up to 1.51 x 10⁵
TCID50/mL), influenza A (up to 5.68 x 10⁷ TC1D50/mL), influenza B (up to 4.36
x 10⁷ TCID50/mL) and RSV (up to 1.17 x 10⁵ TC1D50/mL).
Cross-reactivity
No cross-reactivity was observed with the following, potentially cross-
reactive pathogens when tested using the NADAL® SARSCoV-2/Influenza A+B/RSV Ag
Test: Influenza A, influenza B, adenovirus, respiratory syncytial virus,
coronavirus, MERSCoronavirus, parainfluenza virus, rhinovirus A16, Legionella
pneumophila, Mycobacterium tuberculosis, Streptococcus pneumoniae,
Streptococcus pyogenes, Mycoplasma pneumoniae, Chlamydophila pneumoniae,
Staphylococcus aureus, human metapneumovirus, enterovirus, Haemophilus
influenza, Candida albicans, Bordetella pertussis, Staphylococcus epidermidis,
Pneumocystis jirovecii, pooled human nasal wash.
The NADAL® SARS-CoV-2 Ag Test may cross-react with SARSCoV-1 which caused the 2002-2004 SARS outbreak.
Interfering substances
The following substances showed no interference with the NADAL® SARS-
CoV-2/Influenza A+B/RSV Ag Test:
Antiviral drugs: Zanamivir, Oseltamivir, Artemether/ Lumefantrine, Doxycycline
hyclate, Quinine, Lamivudine, Ribavirin, Daclatasvir.
Respiratory specimens: Mucin from bovine submaxillary glands, type I-S; EDTA-
anticoagulated human blood; Biotin.
Nasal sprays or drops: Neosynephrin® (phenylephrine), Afrin® Nasal Spray
(oxymetazoline), Saline nasal spray.
Homeopathic allergy relief medicine: homeopathic Zicam® Allergy Relief Nasal
Gel, sodium cromoglycate, olopatadine hydrochloride.
Anti-inflammatory medication: Paracetamol (Acetaminophen), acetylsalicylic
acid, ibuprofen.
Antibiotics: Mupirocin, tobramycin, erythromycin, ciprofloxacin.
Precision
Repeatability
Repeatability was established by testing 20 replicates of specimens using 3
lots of the NADAL® SARS-CoV-2/Influenza A+B/RSV Ag Test. The NADAL® SARS-
CoV-2/Influenza A+B/RSV Ag Test demonstrated acceptable repeatability.
Reproducibility
Reproducibility was established by testing 5 replicates of specimens. Testing
was performed by 5 operators using 3 NADAL® SARS-CoV-2/Influenza A+B/RSV Ag
test lots at 3 different sites on 5 days. The NADAL® SARS-CoV-2/Influenza
A+B/RSV Ag Test demonstrated acceptable reproducibility.
References
- Cui J, Li F, Shi ZL, Origin and evolution of pathogenic coronaviruses, Nat Rev Microbiol 2019; 17:181-192.
- Su S, Wong G, Shi W, et al, Epidemiology, genetic recombination, and pathogenesis of coronaviruses, TrendsMicrobiol 2016;24:490-502.
- Weiss SR, Leibowitz JL, Coronavirus pathogenesis, Adv Virus Res 2011; 81:85-164.
- Murphy, B.R., and R.G. Webster, 1996, Orthomyxoviruses, pp.1397-1445. In: Fields, Virology, 3rd edition, B.N. Fields, D.M. Knipe, P.M. Howley, et al. (eds.), LippincottRaven, Philadelphia.
- Biosafety in Microbiological and Biomedical Laboratories, 4th Edition. U.S. Department of Health and Human Services, CDC, NIH, Washington, DC (1999).
- CDC website: http://www.cdc.gov/flu/
- Anne Moscona. Neuraminidase Inhibitors for Influenza, 2005. The New England Journal of Medicine, 353 (13):1363-1373.
- Chanock, R.M., and L. Findberg. 1957. Recovery from infants with respiratory illness of a virus related to chimpanzee coryza agent (CCA). II. Epidemiologic aspects of infection in infants and young children. Am. J. Hyg.66 : 291-300.
- Chanock, R.M., H.W. Kim, A.J. Vargosko, A. Deleva, K.M. Johnson, C.Cumming, and R.H. Parrott. 1961. Respiratory syncytial virus. I. Virus recovery and other observations during 1960 outbreak of bronchiolitis, pneumonia, and minor respiratory diseases in children. J. Am. Med. Assoc. 176 : 647-653.
- Hall, C.B, R.G. Douglas, and J.M. Geiman. 1976. Respiratory syncytial virus infections in infants: quantitation and duration of shedding. J. Pediatr. 89 : 1443-1447.
- Hall, C,B,J.T. McBride, E.E. Walsh, D.M. Bell, C.L. Gala, S.Hildreth, L.G. Teneyck, and WW.J. Hall. 1983. Aerosolized ribavirin treatment of infants with respiratory syncytial virus infection. N.Engl. J. Med. 308 : 1443-1447.
- Taber, L.H.V, Knight, B.E. Gilbert, H.W. McClung, S.Z. Wilson, H.J. Norton, J.M. Thurson, W.H. Gordon, R.L. Atmar and W.R. Schlaudt. 1983. Ribavirin aerosol treatment of bronchiolitis associated with respiratory syncytial virus infection in infants. Pediatrics 72 : 613-618.
- Ahluwalia, G.J. Embree, P. McNicol, B.Law, and G.W. Hammond. 1987. Comparison of nasopharyngeal aspirate and nasopharyngeal swab specimens for respiratory syncytial virus diagnosis by cell culture, indirect immunofluorescence assay, and enzyme-linked immunoabsorbent assay. J.Clin. Microbiol. 257 : 763-767.
Rev. 0, 2022-05-09 OM
nal von minden GmbH
• Carl-Zeiss-Strasse 12
• 47445 Moers
• Germany
• info@nal-vonminden.com
• www.nal-vonminden.com
References
- nal von minden GmbH - Spezialist für hochwertige in-vitro Diagnostik
- Influenza (Flu) | CDC
- nal von minden GmbH - Spezialist für hochwertige in-vitro Diagnostik
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>