BIOCARE MEDICAL ONCORE Pro X Fully Open Slide Stainer User Manual
- June 16, 2024
- BIOCARE MEDICAL
Table of Contents
ONCORE Pro X Fully Open Slide Stainer
ONCORETM Pro X
System Specifications
Oncore Pro X Automated Slide Staining System:
-
Product Code: ONCPRX00001
-
Software Version: 4.0
-
Computer Configuration: Laptop or Desktop Computer with
Mouse -
Instrument Communications Interface: USB 2.0
-
USB Cable must be Tools>Move Arm to Middle to safely move
the robotic arm aside to access the buffer.
Product Usage Instructions
Buffer Replacement
-
Disconnect the buffer container and carefully remove it from
the instrument. Avoid making contact with the robotic arm or
deforming the buffer adapter tubing. -
Refill the container with fresh buffer, then reconnect the
buffer container and place it back onto the instrument. Clean up
any leaks or spills. -
Click System Utilities>Tools>Prime Z1 and Z2 to prime the
inline tubing, then visually inspect the lines to ensure that air
bubbles are not present.
Waste Container Emptying
-
Click System Utilities>Tools>Move Arm to Middle to safely
move the robotic arm aside from the waste stations. -
Carefully pour some water through the waste stations to flush
out the waste tubing. -
Disconnect the tubing from the Waste Container caps and empty
the waste containers. Dispose of the waste according to local
regulations. -
Reconnect the waste tubing: the black tubing is designated for
hazardous waste and the blue tubing is designated for non-hazardous
waste. -
Verify that the waste tubing is draining properly, making sure
that the inserted end of the tubing clears the maximum waste level
line. Adjust the angle and free length of the tubing as
needed.
Reagent Check: Scan Reagents
- Remove all reagent vial caps before proceeding.
- From the Main screen, click Scan Reagents to open the Reagent
Check screen. The reagent map is displayed on the left and a table
of reagents required for the run is displayed on the right.
FAQ
Q: What is the maximum capacity of the Waste Containers?
A: The standard Waste Containers have a maximum capacity of
4000mL, with a secondary containment provided for each waste
container.
Q: How should I dispose of the waste?
A: Dispose of the waste according to local regulations.
ONCORETM Pro X
Operating Manual
PathCom Systems Corporation 6400 Sierra Court, Suite E Dublin, CA 94568
Document Number Form 751-02-03 Revision L4 2023-10-27 For Research Use Only
Tel. 800-799-9499 | www.biocare.net |
Fax: 925-603-8080
Distributed by: Biocare Medical
60 Berry Drive Pacheco, CA 94553
Table of Contents
Section 1. Introduction ………………………………………………………………………………………………………………………… 7
Section 2. System Specifications ……………………………………………………………………………………………………………. 8
2.1. Instrument Overview ………………………………………………………………………………………………………………………..8 2.2.
Module Overview……………………………………………………………………………………………………………………………11 2.3.
Technical Specifications …………………………………………………………………………………………………………………..12 2.4.
Accessories Included with the Instrument………………………………………………………………………………………….13
2.5. Slide Specifications………………………………………………………………………………………………………………………….14
Slide Dimensions ………………………………………………………………………………………………………………………………14 Tissue
Thickness ……………………………………………………………………………………………………………………………….14 Tissue Placement
……………………………………………………………………………………………………………………………..14 Section 3. Installation
Requirements and Instructions …………………………………………………………………………….. 15 3.1.
Installation Requirements………………………………………………………………………………………………………………..15 3.2.
Installation Instructions …………………………………………………………………………………………………………………..15
Contacts for Assistance ……………………………………………………………………………………………………………………..16
Section 4. Instructions for Use …………………………………………………………………………………………………………….. 17
4.1. Starting Up the Slide Stainer
…………………………………………………………………………………………………………….17 4.2. Preparing and Printing
Labels …………………………………………………………………………………………………………..19 Add New Labels in the
Barcode Editor…………………………………………………………………………………………………19 Print Labels in the
Slide Label Editor……………………………………………………………………………………………………21 Move Slides to
Printed Slides (for use with LIS only) ……………………………………………………………………………..22 4.3.
Loading Slides and Installing Chambers ……………………………………………………………………………………………..22
Load/Unload Slides …………………………………………………………………………………………………………………………..22
Install/Remove Chambers ………………………………………………………………………………………………………………….23 4.4.
Assigning Protocols: Scan Slides and Manually Assign Protocols
…………………………………………………………..24 4.5. Loading Reagents
……………………………………………………………………………………………………………………………26 4.6. Vialing
Reagents……………………………………………………………………………………………………………………………..29 7mL Reagent Vial
Specifications………………………………………………………………………………………………………….29 15mL Reagent Vial
Specifications ……………………………………………………………………………………………………….29 Instructions for
Use…………………………………………………………………………………………………………………………..29 4.7. Managing Buffer and
Waste …………………………………………………………………………………………………………….31 Refill the Wash Buffer
Container ………………………………………………………………………………………………………..31 Empty the Waste
Containers ……………………………………………………………………………………………………………..32 4.8. Reagent Check:
Scan Reagents …………………………………………………………………………………………………………33 4.9. Starting the Run
……………………………………………………………………………………………………………………………..37 Gantt
Chart………………………………………………………………………………………………………………………………………39 4.10. Auto
Start…………………………………………………………………………………………………………………………………….40 4.11. Completing a
Run………………………………………………………………………………………………………………………….40 Runtime
Monitoring………………………………………………………………………………………………………………………….41 End of
Run……………………………………………………………………………………………………………………………………….43 Aborting a Run
…………………………………………………………………………………………………………………………………44 4.12. Continuous Loading
………………………………………………………………………………………………………………………45 Stop the Current Run
………………………………………………………………………………………………………………………..45 Restart the Run
………………………………………………………………………………………………………………………………..46 4.13. Shutting Down the
Slide Stainer ……………………………………………………………………………………………………..47 Section 5. System
Utilities…………………………………………………………………………………………………………………… 49 5.1.
Tools……………………………………………………………………………………………………………………………………………..49 Prepare Slides
with Baking Slides Function…………………………………………………………………………………………..49 Load a Saved
Slide Map File ……………………………………………………………………………………………………………….50 Verify the
Calibration of SP1-Z1 or SP1-Z2……………………………………………………………………………………………50
ONCORE Pro X, Operating Manual, Rev. L
2
Verify the Calibration of the Barcode Reader ……………………………………………………………………………………….50 Clean Z1/Z2 Tubing …………………………………………………………………………………………………………………………..51 (Optional) Access the Tubing Clean Log ……………………………………………………………………………………………….52 (Optional) Reset the Module/Chamber Usage Counter………………………………………………………………………….52 Generate Reports……………………………………………………………………………………………………………………………..53 Move Robot XYZZ Arm to Middle………………………………………………………………………………………………………..53 Perform System Initialization ……………………………………………………………………………………………………………..53 Check for Module Malfunction …………………………………………………………………………………………………………..53 Flush Tubing with Prime Pump Z1 and Z2 Function ……………………………………………………………………………….55 Prime Pump Z1 or Pump Z2 ……………………………………………………………………………………………………………….55 Lock or Unlock Door ………………………………………………………………………………………………………………………….55 Backup Report Data ………………………………………………………………………………………………………………………….55 Replace Pump 1 Syringe or Pump 2 Syringe (Distributor Access only) ……………………………………………………..55 5.2. Editors …………………………………………………………………………………………………………………………………………..56 Editing Protocols in the Protocol Editor: Overview………………………………………………………………………………..56 Adjusting Protocol Customizations: IHC……………………………………………………………………………………………….58 Adjusting Protocol Customizations: ISH/CISH/FISH ……………………………………………………………………………….60 Refreshing Protocol Customizations ……………………………………………………………………………………………………61 Changing the Assigned Protocol Template …………………………………………………………………………………………..62 Adding New Protocols in the Protocol Editor ……………………………………………………………………………………….63 Creating and Modifying Special Protocols (Distributor Access only) ………………………………………………………..64 Creating and Modifying Protocol Templates (Distributor Access only) …………………………………………………….66 Editing Protocol Parameters in the Advanced Protocol Editor ………………………………………………………………..68 Configuring the Protocol Editor ………………………………………………………………………………………………………….71 Importing Protocols in the PR Manager……………………………………………………………………………………………….72 Exporting Protocols in the PR Manager (Distributor Access only) ……………………………………………………………73 Adding New Reagents in the Reagent Editor ………………………………………………………………………………………..75 Preparing RFID Reagent Tags with the RFID Tag Editor ………………………………………………………………………….78 Assigning Negative Controls……………………………………………………………………………………………………………….80 Creating Custom Groups ……………………………………………………………………………………………………………………81 Creating Custom Panels …………………………………………………………………………………………………………………….82 5.3. Security …………………………………………………………………………………………………………………………………………83 Change Password ……………………………………………………………………………………………………………………………..83 User Administration ………………………………………………………………………………………………………………………….83 5.4. Settings …………………………………………………………………………………………………………………………………………84 Baking Slide Settings …………………………………………………………………………………………………………………………84 2D Barcode Format …………………………………………………………………………………………………………………………..85 Label Format ……………………………………………………………………………………………………………………………………86 Edit Doctor Name List ……………………………………………………………………………………………………………………….87 Check Heater Settings ……………………………………………………………………………………………………………………….87 Save All Customizations …………………………………………………………………………………………………………………….87 Device Version………………………………………………………………………………………………………………………………….87 Baking Slides Option………………………………………………………………………………………………………………………….88 Detect Waste Overflow Option (Distributor Access only)……………………………………………………………………….88 Run Without RFID Reader Option (Distributor Access only) ……………………………………………………………………88 Reagent Consumption Priority ……………………………………………………………………………………………………………88 Software Version………………………………………………………………………………………………………………………………88 Allowed Reagent Assignment (Distributor Access only) …………………………………………………………………………88 Interlock Options (Distributor Access only) ………………………………………………………………………………………….88 Language …………………………………………………………………………………………………………………………………………88 Tubing Clean Tracking Option (Distributor Access only) …………………………………………………………………………89 Module Tracking Option (Distributor Access only) ………………………………………………………………………………..89 5.5. Network ………………………………………………………………………………………………………………………………………..90
ONCORE Pro X, Operating Manual, Rev. L
3
5.6. LIS/LIMS ………………………………………………………………………………………………………………………………………..92 Section
6. Reports……………………………………………………………………………………………………………………………… 93
6.1. Generate a Report ………………………………………………………………………………………………………………………….93 Slide
Barcode……………………………………………………………………………………………………………………………………93 Patient
Information…………………………………………………………………………………………………………………………..95 Staining
Runs……………………………………………………………………………………………………………………………………96
6.2. Report Types ………………………………………………………………………………………………………………………………….97 Slide
Report……………………………………………………………………………………………………………………………………..97 Reagent Usage in
Run Report …………………………………………………………………………………………………………….99 Reagent Usage Report
…………………………………………………………………………………………………………………….100 Slide Summary
Report……………………………………………………………………………………………………………………..101
6.3. Print and Save a Report …………………………………………………………………………………………………………………102
6.4. Backup Report Data ………………………………………………………………………………………………………………………102
Section 7. System Maintenance …………………………………………………………………………………………………………. 103
7.1. Daily Maintenance Recommendation ……………………………………………………………………………………………..103
Modules ………………………………………………………………………………………………………………………………………..103 Chamber
……………………………………………………………………………………………………………………………………….. 103 Tubing and
Pumps…………………………………………………………………………………………………………………………..103 7.2. Weekly
Maintenance Recommendation ………………………………………………………………………………………….104 Modules and
Deck Plate…………………………………………………………………………………………………………………..104 Wash Station and
Z1/Z2 Probes………………………………………………………………………………………………………..104 Tubing and
Pumps…………………………………………………………………………………………………………………………..104 7.3. Monthly
Maintenance Recommendation…………………………………………………………………………………………105 Chambers
………………………………………………………………………………………………………………………………………105 Tubing
…………………………………………………………………………………………………………………………………………..105 Robotic Arm
…………………………………………………………………………………………………………………………………..105 Computer
………………………………………………………………………………………………………………………………………105 7.4. Annual/Semi-Annual
Maintenance………………………………………………………………………………………………….105 Section 8. General
Precautions ………………………………………………………………………………………………………….. 106
ONCORE Pro X, Operating Manual, Rev. L
4
Instrument Information
Oncore Pro X Automated Slide Staining System: ONCPRX00001 Software Version:
4.0
Computer Configuration: Laptop or Desktop Computer with Mouse Instrument
Communications Interface: USB 2.0 USB Cable must be <3 meters (10 feet)
The use of this instrument is fully licensed under U.S. Patent no. 5,839,091;
U.S. Patent no. 7,476,543; U.S. Patent no. 7,635,453; and U.S. Patent no.
7,977,086 B2.
Software Product License
IMPORTANT: Please read the license carefully before using this Software. The
right to use this Software is granted only if the Customer agrees to the terms
of this license
HOWEVER, USE OF THIS SOFTWARE INDICATES YOUR ACCEPTANCE OF THESE TERMS AND
CONDITIONS. In return for the payment of the applicable one time license fee
included with the purchase price of the instrument, BioCare grants the user a
license to use one copy of the Software subject to the following terms and
conditions. No title or ownership of the Software is conferred with this
license.
1. The Software may be used on one Oncore Pro X Automated Slide Staining
System. 2. The Software may not be copied. 3. The Software may not be
disassembled, decompiled, decrypted or reverse engineered. 4. BioCare reserves
the right to terminate this license upon breach. In the event of termination
at BioCare’s
election, all copies of the Software must be returned to BioCare. 5. No title
or ownership of the Software is conferred with this license.
ONCORE Pro X, Operating Manual, Rev. L
5
Safety
Report any serious incidents related to this device to your local Oncore Pro X
representative and the state’s competent authority. Contact your local
representative if you have any questions or concerns regarding safety.
Protection from High Temperatures The instrument is equipped with a safety
interlock system so that the door remains locked during instrument operation.
However, the heaters inside may reach high temperatures. Care must be
exercised to make sure that the heaters have cooled down sufficiently before
removing the slides.
Protection from Moving Parts The instrument is equipped with a safety
interlock system so that the door remains locked during instrument operation.
The interlock will automatically engage to restrict access to the inside of
the instrument while it is the instrument is in operation. Never tamper with
the safety interlock system.
Use of Hazardous Reagents Some reagents used on the instrument may be
considered hazardous. Hazardous reagent wastes must be disposed of according
to local, state and federal regulations. Wear appropriate Personal Protective
Equipment to prevent exposure. It is the responsibility of the user to comply
with the local, state and federal regulations.
Maintenance Do not attempt to service the instrument unless instructed to do
so by Technical Support. Doing so will void the warranty. Removing the back
panel may also electrical hazard. The instrument must be unplugged from the
power source before the outer housing can be removed. The power supply is
connected to 120/240V power source and the power cord must be disconnected
from the AC power port before any maintenance work is started.
Decontamination Use bleach at 1:10 dilution to decontaminate the instrument.
Do not use bleach with any other chemicals because it may react and create
toxic fumes. Decontaminate in a well ventilated room.
ONCORE Pro X, Operating Manual, Rev. L
6
Section 1. Introduction
Intended Use of the Equipment
The Oncore Pro X Automated Slide Staining System is a fully-automatedfully
automated slide processing system for staining paraffin-embedded and frozen
tissue sections, cytospins, cell smears and fine needle aspirates. This
universal system is designed to automate the manual staining methods routinely
used in immunohistochemistry, in situ hybridization and related applications.
The intended use of the instrument is RUO (Research Use Only).
Quick Start-up
The Oncore Pro X Service Team installs and tests each system to ensure optimal
performance. The system is factory-installed with a set of standard protocols
optimized for use with the Oncore Pro X detection kits and its comprehensive
line of ready-to-use primary antibodies and probes.
Flexible Programming
The Oncore Pro X Automated Slide Staining System software implements user-
customized protocols and slidespecific programming. The system provides
intuitive programming routines, reagent and slide loading maps and a
convenient run time display. Each staining run can process 1 to 36 microscope
slides. Each slide can be programmed to run a different protocol.
Economic Use
The Oncore Pro X Automated Slide Staining System maximizes efficiency by
operating with small volumes of reagent and wash buffer, effectively reducing
supply and waste disposal costs. The system provides the flexibility to
optimize reagent volumes and wash steps.
Reagent RFID Tags
Reagent RFID tags store programmable information such as reagent name, lot
number and expiration date. The on-board RFID reader scans and identifies
reagent vials to automatically determine if there are sufficient resources to
complete a selected staining protocol.
Slide 2D Barcode Labels
The Oncore Pro X Automated Slide Staining System software provides a simple
user interface to generate and print 2D Barcode labels for slides. The on-
board 2D barcode reader scans and identifies slides to automatically program a
run and track slides. All printed labels are heat and chemical-resistant,
allowing users to apply slide labels prior to starting the staining procedure.
ONCORE Pro X, Operating Manual, Rev. L
7
Section 2. System Specifications
The Oncore Pro X Automated Slide Staining System employs a unique integration
of hardware, software, and reagent chemistry to provide full automation of
immunohistochemical staining methods. This section provides an overview of the
instrument and details the technical specifications.
2.1. Instrument Overview
Figure 1
1. Slide Deck holds 36 modules, a reagent rack, and 2 probe wash stations. 2.
X axis mechanism moves the robotic arm from left to right. 3. Y axis mechanism
carries the Z head assembly and moves the arm from front to back 4. Z head
Assembly consists of 2 independent Z axes (Z1 and Z2) situated on the Y arm.
5. Power Button turns the instrument on/off. 6. Door controls access to the
instrument. 7. Linear Solenoid locks the door. 8. Safety Interlock Switch
blocks unsafe access to the instrument while the robotic arm is in motion. 9.
Liquid Level Sensor (Z1 only) detects the liquid level in the reagent vial and
waste station during runtime.
The system will issue a warning when a reagent vial has insufficient volume or
when a waste station is overflowing.
ONCORE Pro X, Operating Manual, Rev. L
8
Figure 2
1. 2D Barcode Reader scans the labelled slides on the slide deck to assign
protocols for the staining run. 2. RFID Reader scans the RFID-tagged reagent
vials to determine the reagents’ positions on the rack and
updates the reagent usage information to the RFID tags during the run. 3.
Reagent Rack holds the reagent vials. There are 40 available reagent vials
positions. 4. Reagent Vials holds the reagents. 15mL and 7mL size options. 5.
Pumps (Z1 and Z2) draw and dispense buffer and reagents through the in-line
tubing (max 5mL capacity) 6. Wash/Waste Station is an assembly located in the
rear left corner of the instrument and is used as a
cleaning station for the reagent probes. There is a purging position for the
reagent waste and a percolating position for cleaning the outside of the probe
each time a new reagent is used. 7. Hazardous Wash/Waste Station is a second
probe wash station dedicated to hazardous reagents. Hazardous reagents will be
separately dispensed into the hazardous waste station and will be collected in
the Hazardous Waste Container. 8. Z1 Probe is a TeflonTM-coated stainless
steel probe that aspirates the reagents from the vials/in-line wash buffer and
dispenses the reagents into the chambers. 9. Z2 probe is a TeflonTM-coated
stainless steel probe that extracts the reagent waste from the chambers and
aspirates/dispenses reagent during on-board mixing. 10. Module is an assembly
that holds 1 glass slide. 36 modules per instrument.
Not pictured: 11. Tubing (Z1 and Z2) connects the pumps with the probes to
transfer buffer, reagents, and waste.
ONCORE Pro X, Operating Manual, Rev. L
9
Figure 3
1. Feet adjusted to level the instrument on the work surface. 2. Main Power
Supply connects to the wall outlet and provides power to the instrument. 3.
Power Supply Switch turns the power supply on/off. 4. AC Power Port is the
position to plug in the power cord. 5. Fan provides cooling to the instrument
and its electrical components. 6. USB Port is the position to plug in the USB
cable. 7. U-Ring is the position to attach an anti-theft chain to the
computer. 8. Waste Tubing (Blue) connects the waste station to the waste
container to transfer waste (gravity-fed). 9. Hazardous Waste Tubing (Black)
connects the waste station to the hazardous waste container to transfer
hazardous waste (gravity-fed).
ONCORE Pro X, Operating Manual, Rev. L
10
2.2. Module Overview
Figure 4
1. Side Retainer Clips (2) secure the module onto the instrument deck plate.
2. LED Light displays heater status green (inactive) or orange (active).
3. Hinge secures the slide at the back. 4. Chamber Chimney reagent
dispensation/waste
extraction point. 5. Slip Holder holds the chamber. Moves to controlled
angles to facilitate incubation and reagent dispensation/waste extraction. 6.
Chamber (consumable) holds reagent/buffer on the slide during the incubation
and washes. 7. Side Rails (2) secure the slide on the left/right. 8. Heater
maintains temperature during incubation. 9. Slide Clip secures the slide at
the front.
Not pictured: 10. Glass Slide is a standard microscope slide to which
specimen(s) are mounted for processing on the instrument. Standard slide
dimensions are 25mm x 75mm x 1mm.
ONCORE Pro X, Operating Manual, Rev. L
11
2.3. Technical Specifications
Dimensions W x H x D 36.5in x 22in x 24in/93cm x 56cm x 61cm (with door closed)
Weight 125lb (57kg)
36.5in x 36in x 24in/93cm x 91cm x 61cm (with door open)
Electrical requirements 120V 110/120V (±10%) 60Hz (±2Hz) 850 watts 220V
220/240V (±10%) 50Hz (±2Hz) 850 watts Mains Connection: IEC320
Normal operating temperature 18°C-26°C (64°F-79°F)
Slide capacity 1 – 36 glass slides
Reagent capacity 40 different reagents (15 mL/7mL reagent vial)
Heating capacity Room temperature to 103°C. Maximum 110°C.
Waste Separation 1 hazardous waste, 1 non-hazardous waste
Reagent dispense volumes 130uL standard test volume (65 minimum-400 µl maximum)
Reagent probe volume capacity 20 µL minimum-4500µL maximum
Recommended Computer Controller Specifications The manufacturer reserves the right to change the computer controller specifications at any time. Intel Core i5-2410M CPU 2.3GHz, or equivalent 8GB RAM 500GB Hard drive or 256GB SSD drive 4 USB 2.0 communication ports, or equivalent
Display Specifications 16:9 aspect ratio, minimum resolution 1366 x 768
Operating System Windows 10 Pro (64 bit), English (US Region)
Label Printer (ZPL) Zebra GX430t, Zebra ZD620t, or Zebra ZD621t
Surge Protector 120/240 VAC, 15 Amps, 50/60 Hz
LIS Connectivity Compatible with SQL Database and HL7 messaging standards.
Operating Logic Designed to calculate the most time-efficient sequence of
steps to complete a programmed staining run.
Protocol Logic Protocols are generated from a common protocol template
optimized for each detection system. End-users have the limited ability to
customize key protocol steps. Advanced users have full flexibility to
add/remove and customize each step in the protocol template and to generate
special protocols.
ONCORE Pro X, Operating Manual, Rev. L
12
2.4. Accessories Included with the Instrument
Reagent Rack Two removable reagent racks. Each reagent rack holds a maximum of
40 reagent vials. Removable Vial Holder Installed on the reagent rack for
programming RFID-tagged reagent vials. Buffer Container 2L plastic bottle with
quick disconnect fittings. Stores the in-line wash buffer. Waste Container Set
of two 4L plastic bottles with cap. Stores the reagent waste generated during
the run. Label Printer 2D barcode printer, thermal transfer.
Labels Size 19.05mm (0.75in) x 25.4mm (1in), 1000 labels/roll. Ribbon Media
ribbon for the printer. USB Cable Connects the instrument and computer. Power
Cord, Instrument Computer Pre-configured for use with instrument. Computer
Mouse Lint-free Alcohol Wipes Chamber Set of 36 pre-installed on the modules.
ONCORE Pro X, Operating Manual, Rev. L
13
2.5. Slide Specifications
To ensure staining quality and reliability, slides should be prepared
following the instrument’s guidelines.
Slide Dimensions
Standard slide dimensions are 25mm x 75mm x 1mm
Width Length Thickness
Material Glass, ISO 8037/1
Dimensions (mm)
2526.0 mm (0.981.02 in) 74.7575.5 mm (2.942.97 in) 0.951.1 mm
(0.0380.043 in)
Note: Use of non-standard slides may affect staining area and staining quality.
Tissue Thickness
The recommended tissue thickness is 2-5 microns.
Tissue Placement
The circled area is the recommended tissue placement area. The yellow
highlighted region is the staining area range. This range may be expanded to
the blue highlighted regions; however, staining results may not be ideal in
these areas.
Figure 5
The IHC staining area extends to the 40 mm line on the slide. The ISH staining
area extends to the 30 mm line on the slide.
ONCORE Pro X, Operating Manual, Rev. L
14
Section 3. Installation Requirements and Instructions
This section provides an overview of the instrument installation to be
performed by the field service engineer. The following requirements should be
met to ensure proper instrument function.
3.1. Installation Requirements
Location Place the system in an area where it is easy to operate and easy to
connect to the instrument.
Surface Prior to unpacking, ensure that the work area for the instrument is a
solid, level surface that can safely
support the weight of the system.
Size Minimum dimensions for the work area are 36.5 inch (93 cm) width x 36
inch (91 cm) height x 24 inch (61 cm)
depth clearance.
Environment The work area must be in an environment with an ambient
temperature between 18°C-26°C (64°F-
79°F). The area must be properly ventilated with a maximum relative humidity
of 80% for temperatures up to 31oC; Pollution Degree II or less. The equipment
will produce less than 85 decibels at peak operation.
Power source The instrument and computer must be connected to a 120V/240 15A
supply with grounding (3
prong). It is recommended to use a dedicated power source to prevent
interference from other instruments or equipment. It is also recommended to
use a UPS (uninterruptible power supply) with minimum 1500VA. This will help
to prevent unexpected system failures due to an unstable power grid system or
in the event of a temporary power outage.
3.2. Installation Instructions
Unpacking The system is packaged in a wooden crate mounted on a pallet.
Unlatch all fasteners and remove the
lid on top of the crate first. Remove the computer, accessories box and label
printer from the crate. Remove the sides of the crate.
Installation Requires 2 people. Each person should pick up one end of the
instrument and carefully lift it out of
the crate. Carefully move the instrument to the operating area and place it on
the bench. Level the instrument surface using a bubble level. Remove all
packing material and the packaging securing the robotic arm. Verify that all
axes can move freely before operating the instrument. Unpack and install the
computer and accessories.
Hooking up Connect the USB cable from the designated port on the computer to
the back of the instrument.
Do not plug and unplug the USB cable while the PC or Instrument is on.
***Always turn off the instrument and PC before plugging and unplugging the
USB cable.
Connect the Computer to the power source or UPS. Follow the instructions on
the UPS, as some may require powering up for several hours to charge the
internal battery. Using a standard IEC320 compatible power cord, connect the
AC power port at the rear of the instrument to the power source or UPS.
Cut the waste tubing to size and route to the waste bottles on the floor. The
ideal configuration is straight down to the waste bottle to minimize the
chance of overflow and waste buildup.
Avoid immersing the waste tubing in the waste bottle or flattening and kinking
the tubing as this may disrupt the gravity fed drainage and result in waste
overflow.
ONCORE Pro X, Operating Manual, Rev. L
15
Starting up Turn on the instrument and computer.
Power on the power supply switch on the back of the instrument. Power on the
Slide Stainer by pushing on the power button at the front of the instrument.
Log in to the computer. Default User Name: PathCom Default Password: 6747l
Note: The user may change the computer user name and password after initial
log in.
Disable the Wi-Fi radio and local internet connection (if applicable). This
feature may cause unexpected crashes during instrument operation.
The internet access should only be enabled for a remote support session, email
communication or web conference call. Disable the internet access while the
instrument is in operation to prevent unexpected crashes.
Disable automatic time synchronization. This feature may cause unexpected
crashes during instrument operation.
For Windows 10 systems: Configure Time & Language settings within Windows
Settings. Toggle Off the Date and time options as shown.
Figure 6
Do not change the Windows Settings without consulting Technical Support, as
this may cause unexpected crashes and performance issues. This includes
settings for input language, OS language, region, regional formatting, power
options, Windows Update etc.
Setup the barcode label printer. Load the ribbon and media according to the
manufacturer’s instructions. Connect the printer to the designated port on the
computer and switch the power on. The printer driver will automatically
install within a few minutes. Calibrate the media, if needed, using the Zebra
Setup Utilities.
The on-site field representative will perform the necessary system checks on
robot, modules, and pumps.
Contacts for Assistance
Customer Service: 925-603-8000
Technical Support: 1-800-799-9499 Opt.
3
Replacement of Consumable Materials Replacements for all consumable materials can be obtained from: Website: www. biocare.net
ONCORE Pro X, Operating Manual, Rev. L
16
Section 4. Instructions for Use
This section provides the instructions for daily routine operation of the
instrument. Please check all Requirements p.7 before beginning instrument
operation.
4.1. Starting Up the Slide Stainer
Turn on the Slide Stainer using the power button and then turn on the computer. Note: It is important to follow this exact power on sequence.
Double-click the shortcut for the Oncore Pro X Automated Slide Staining System UI
to access the login
screen.
Select “Yes” when prompted by Windows User Account Control to grant access permissions.
“Do you want to allow this app from an unknown publisher to make changes to your device?”
PCStainer.exe
Login to the main user interface. Default User ID: Supervisor Default Password: Supervisor
Figure 7
Refer to Security, p. 83, for information on managing User IDs and Passwords.
Close the door of the Slide Stainer and click “Enter” to access the Main
screen and begin system initialization. The system will take a few minutes to
verify all connections between the instrument and the computer, and then home
the robotic arm, modules, and pumps.
Figure 8
ONCORE Pro X, Operating Manual, Rev. L
17
Tip: If the system fails to start or is stopped/interrupted while performing a task, re-initialize the system and try again. Refer to Perform System Initialization, p.53, for more information. An interactive slide map corresponding to the 36 slide positions is displayed on the Main screen.
Figure 9
All routine functions are readily accessed from the Main screen. To perform a
typical run, click:
Prepare Labels to print labels for slides. Each label is printed with a 2D
barcode required to identify the slide. Load/Unload Slides to lift and lower
lids for inserting and removing slides. Auto Start to automatically start the
run after loading the labeled slides and tagged reagents. Scan Slides to scan
the barcode labels on the slides. Assign Protocols to manually assign a
protocol to a slide. Scan Reagents to scan the RFID tags on the reagent vials.
System Utilities to access additional features and utility functions. Clean
Screen to clear the slide map. Exit to close the application.
ONCORE Pro X, Operating Manual, Rev. L
18
4.2. Preparing and Printing Labels
From the Main screen, click “Prepare Labels” to open the Slide Label Editor:
Figure 10
Labels available for printing are listed on the screen under New Slides. A.
Select New Slides to create or view new labels. B. Select Printed Slides to
review or reprint a label. C. Select Scanned Slides to review labels that have
already been scanned. D. Select Processed Slides to review labels that have
already been processed.
Add New Labels in the Barcode Editor
Labels must be prepared in the Barcode Editor and added to the Slide Label
Editor before they can be printed.
To add new labels to the Slide Label Editor, click “Add” to open the Barcode
Editor. Enter the Patient ID, Case Number, and other required label
information into the designated fields.
Note: caret ^ and tilde ~ characters are not permitted in the entry fields.
(Optional) Select the Date Time of the label.
(Optional) Select the name of the Doctor from the dropdown list. To add new
names to the list, refer to 2D Barcode Format p. 84.
(Optional) To print the “Block ID” field to the slide label, the user must
change the default label format. Select “Container Identifier” from the drop-
down list in the desired field of the Label Formatter.
Note: The information entered may not be printable due to the size constraints
of the label. Use abbreviations as needed or select a different label format.
Protocol Name, Patient ID, Date, Case Number, Hospital Name and Description
are printed on the label by default. To change this setting, refer to 2D
Barcode Format p.84.
Figure 11
Next, add protocols to the list of Selected Protocols:
ONCORE Pro X, Operating Manual, Rev. L
19
To add protocols to the Selected Protocols, select protocol(s) from the list
of Available Protocols and click “Add”. Tip: The user may search the list of
protocols by typing the first few letters of the protocol name. Click “Save”
to generate new labels for each protocol in the list of Selected Protocols.
Alternatively, the user may use a customized sublist of available protocols by
selecting a custom group from the Custom Group drop down list.
Figure 13
Refer to Creating Custom Groups, p. 81, for more information on creating
custom groups
Figure 12
To remove protocols from the list of Selected Protocols, select protocol(s)
from the list and click “Remove”. (Optional) To add a positive and/or negative
control to the Selected Protocols, select the protocol and check the boxes Add
With: Positive and/or Negative, and click “Add”. Click “Save” to generate new
labels.
Refer to Assigning Negative Controls, p. 78, for more information on assigning
negative controls to protocols.
Figure 15
Figure 14
ONCORE Pro X, Operating Manual, Rev. L
20
(Optional) To add an entire panel of protocols to the Selected Protocols,
select the custom panel from the list of Available Panels and click “Add”. All
protocols assigned to the panel (including positive and negative controls)
will be added. Click “Save” to generate new labels.
The user must first create a custom panel before the Available Panels field is
made visible in Barcode Editor.
Figure 17
Refer to Creating Custom Panels, p. 81, for more information on creating
Panels.
Figure 16
After saving the current labels, the user may continue preparing labels for
the next case. By default, the input patient information is retained in the
entry fields for the next set of labels. To change this setting, refer to 2D
Barcode Format p. 84. Click “Return” to close the Barcode Editor.
Print Labels in the Slide Label Editor
All unprinted label entries are listed under New Slides in the Slide Label
Editors. This list is automatically sorted by Patient ID; however, the user
may sort the list by the column headers i.e.: Label No., Case Number, Block ID
etc.
Figure 18
To print labels from the Slide Label Editor, select the label(s) or “Select
All” using the checkbox, and click “Print”. Note: Label entries are
automatically removed from the system after 30 days from the date assigned on
the label.
To delete a label from the Slide Label Editor, select the checkbox and click
“Delete”. Note: Once the slide label(s) have been printed, slide processing is
initiated and the slide information will be distributed to all instruments on
the Network. The label entry can no longer be deleted.
Refer to 2D Barcode Format, p. 84, for information on adjusting printer
settings and label format.
ONCORE Pro X, Operating Manual, Rev. L
21
Move Slides to Printed Slides (for use with LIS only)
The instrument can be configured to recognize slide labels generated from an
external source when integrated with an LIS system. Note: The 2D Barcode
reader must be configured to recognize labels of different formats. Please
contact Technical Support for more information.
Figure 19
The LIS system sends the processing order directly to the system computer. The
ordered slides will appear under New Slides in the Slide Label Editor. The
user has the option to 1) print the slide label(s) to initiate processing, or
2) directly move slide label(s) to Printed Slides to initiate slide processing
without printing, or 3) delete the slide labels to cancel the order.
To manually move pre-printed labels to Printed Slides (without printing),
select the label(s) using the checkbox and click “Move To Printed Slides”.
4.3. Loading Slides and Installing Chambers
From the Main screen, click “Load Unload Slides” to raise all module lids.
Peel the labels from the label roll and affix them to the frosted side of the
slide. Slides are loaded individually onto each module.
For slides that must be baked prior to staining, refer to Prepare Slides with
Baking Slides Function p. 49. Refer to Slide Specifications, p. 29, for more
information on tissue location recommendations for slides.
Load/Unload Slides
Verify that the heating plate and module surfaces are clean and dry before
loading a new slide.
To load a slide onto a module, insert the slide into the hinge slot and then
lower it onto the heating plate until it locks in place under the front clip.
If needed, adjust the slide so that it is centered over the heating plate.
Click “Slide Lids Down” to lower all module lids.
Tip1: If the slide is too difficult to insert, clean the hinge area to remove
any salt or debris.
Tip2: The hinge contains springs to accommodate minor differences in slide
length. However, if the slide is too long, the front clip can be adjusted by
loosening the two screws with a small screwdriver.
Figure 20
ONCORE Pro X, Operating Manual, Rev. L
22
To unload the slide from a module, push the slide against the hinge slot until
it is released from under the front clip. Carefully lift the slide out,
keeping it horizontal to prevent excess buffer from spilling onto the module
surfaces.
Transfer the slide to a slide rack and rinse it with distilled water before
post treatment. Wipe the heater, hinge slot, and other affected module
surfaces with a moist paper towel to remove any staining residues, then blot
dry.
Install/Remove Chambers
Chambers are installed individually onto each module. They should be removed
from the module for routine cleaning and inspection to maintain optimal
staining quality. Refer to Daily Maintenance Recommendations, p. 103, for more
information.
Figure 21
To install the chamber, slide the chamber into the recess under the module lid
and push up on the chamber until it locks into place. Verify that all 4
corners of the chamber sit flat in place. Tip: Push up on the underside of the
chamber surface to verify the integrity of the chamber gasket. If the chamber
gasket is loose or damaged, it is time to replace the chamber.
To remove the chamber, gently push down on the chamber to unlock it from the
module lid. Transfer the chamber to a container for cleaning. Tip: After
completing a staining run, gently wipe the chamber surfaces with a lint-free
alcohol wipe to remove any staining residues/debris that may adhere to the
chamber.
ONCORE Pro X, Operating Manual, Rev. L
23
4.4. Assigning Protocols: Scan Slides and Manually Assign Protocols
To automatically assign protocols by scanning labeled slides, close the door
and click “Scan Slides” on the Main screen. The system will automatically scan
all 36 slide positions and assign the protocol(s) of labeled slides to the
appropriate position on the slide map. Refer to Editing Protocols, p. 71, and
Change Protocol Template, p. 71, for more information on generating protocols.
Figure 22
Tip: Click + drag on the slide map to select specific slide positions to scan.
Figure 23
ONCORE Pro X, Operating Manual, Rev. L
24
To manually assign protocols to unlabeled slides, click + drag on the slide map to select the slide position(s). The slide position(s) will be highlighted and display “????????”:
Figure 24
Click “Assign Protocol” to open the Select Protocol screen. A. Click to select
Normal to access the list of normal protocols. B. Click to select Negative to
access the list of negative control protocols. C. Click to select All to
access the list of all protocols.
Scroll through the protocol list, highlight the desired protocol, and click
“Select” to assign the protocol to the highlighted slides. Tip: The user may
quickly scroll through the protocol list by typing the first few letters of
the protocol name.
Click “Clear” to cancel the selection and close the window.
Figure 25
To clear the protocol assignment(s), click + drag to highlight the slide
position(s), and then click + drag once more to remove the assignment(s).
Alternatively, click “Clean Screen” to clear the entire slide map.
ONCORE Pro X, Operating Manual, Rev. L
25
4.5. Loading Reagents
Detection kits, ancillary reagents and antibody/probe products are shipped
ready-to-use. Refer to the manufacturer’s Datasheet and SDS for specific
instructions. Prepare all other reagent solutions following the manufacturer’s
instructions. Note: Only Oncore Pro reagent vials are compatible with the
instrument. Refer to the vialing specifications in Vialing Reagents, p. 29,
for additional information on vialing and recommended reagent volumes. Load
the reagent vials onto the reagent rack and remove all caps. Note: The user
may roughly assess the reagent volume remaining in each vial using the volume
indicator lines located along the side of the vial. The required reagent
volumes for the run are displayed on the right side of the Reagent Check
screen. Refer to Reagent Check: Scan Reagents, p. Error! Bookmark not
defined.. CAUTION! Evaporation will cause gradual loss of volume within the
reagent vial. Avoid leaving the reagent vials open for a prolonged time. Close
all caps and refrigerate reagent vials immediately after use.
Figure 26
Please note that the reagent volume may be significantly reduced by
evaporation after performing several overnight runs and/or after performing
multiple runs using the same reagent vial. This particularly impacts
antibodies that are not routinely run on the system (consuming only 1-2 tests
per run over 20+ runs). Under such circumstances, the reagent vials will empty
prematurely or trigger a low volume warning before the RFID test number
reaches 0. To minimize evaporation in such cases, users should attempt to
aggregate slides sharing the same antibody/protocol to run those slides in
larger batches over a reduced number of staining runs. Meanwhile, users should
monitor the test volume and refill/reprogram those affected vials if the
volume becomes too low.
ONCORE Pro X, Operating Manual, Rev. L
26
Verify that the vials are fully inserted into the rack and oriented correctly,
as shown in the figure below.
A. For ready-to-use reagents, place the reagent in any position on the rack.
B. For on-board mixing reagents, place the two mixing components on the rack,
A and B. Avoid placing the
second mixing component, B, in the last two rows of the rack as the Z2 probe
cannot reach it.
Note: On-board mixing reagents, such as Fast Red 1:1, are specially formulated
to be mixed directly on the slide:
A. The first mixing component, i.e.: Fast Red Substrate, is applied with the
Z1 probe.
B. The second mixing component, i.e.: Fast Red Chromogen, is applied
immediately afterwards with the Z2 probe.
The Z2 probe is specially designated for dispensing the second mixing
component; therefore, those reagents, such as Fast Red Chromogen, can only be
loaded in reagent rack positions #1-30.
Figure 27
C. For on-rack mixing reagents, place all mixing components A and B (and C if
applicable) on the rack. Additionally, place an empty mixing vial (7mL or
15mL) in designated position, ie: #40, as prompted by the system in the
Reagent Check screen.
Note: The system will consume extra tests of each reagent component to fill
the dead volume of the empty mixing vial or to fulfill the minimum mixing
volume requirement.
Figure 28
ONCORE Pro X, Operating Manual, Rev. L
27
Multiple vials of the same reagent may be placed on the rack. The system will
prioritize the use of vials with the earliest expiration date or the least
number of tests remaining. Refer to Reagent Consumption Priority, p. 88.
Load the reagent rack onto the designated reagent plate on the instrument.
Verify that the rack is fully seated on the plate, and is oriented correctly,
as shown in the figure below.
Figure 29
ONCORE Pro X, Operating Manual, Rev. L
28
4.6. Vialing Reagents
The system provides user-fillable reagent vials for special reagents, uncommon
antibodies, and custom ab titers. Each vial is packaged with an RFID tag, cap,
and a blank label.
Use of unauthorized reagents on the system may cause damage to instrument
components and will revoke the manufacturer’s warranty. The clear vial option
is designated for general use. The dark vial option is designated for light-
sensitive reagents such as probes and chromogens. Contact Customer Service for
ordering information.
7mL Reagent Vial Specifications
The 7mL vial is specially designed for efficient use of antibodies and probes
by minimizing dead volume. Max capacity 9mL. The dead volume for a 7mL vial is
~250uL.
Note: Although the 7mL vial can hold up to 9mL, it is recommended to fill
below 7.5mL to avoid triggering liquid volume errors.
Figure 30.
15mL Reagent Vial Specifications
The 15mL vial is designed to hold larger volumes of reagent. Max capacity
16mL. The dead volume for a 15mL vial is ~1000uL.
Figure 31
Instructions for Use
Fill the reagent vial with the appropriate amount of reagent solution. Tip:
Use the Reagent Check screen to quickly obtain a list of required reagent
volumes, then add 1mL dead volume for a 15mL vial or add 250uL for a 7mL vial.
To calculate the required volume: Volume = number of tests x protocol volume
x draw factor 1.08 + dead volume of the vial Note1: It is advised to
fill a lower number of tests per vial for reagents that are volatile or less
chemically stable, and for reagents that are not routinely used. Note2: An
additional 8% extra volume (or more) is required to compensate for reagent
evaporation and the amount adhering to the internal surface of the tubing and
external surface of the aspiration probe. Note2***: The reagent volume may be
significantly reduced by evaporation over several overnight runs or over
multiple staining runs. In such cases, users may need to increase the dead
volume or refill vials more frequently.
Label the reagent vial and cap.
Program the RFID tag. Refer to Preparing RFID Reagent Tags with the RFID Tag
Editor, p. 78, for more details.
ONCORE Pro X, Operating Manual, Rev. L
29
ONCORE Pro X, Operating Manual, Rev. L
30
4.7. Managing Buffer and Waste
The system will prompt the user to verify the required Wash Buffer and Waste
Container volumes before proceeding with the staining run.
Refill the Wash Buffer Container
The standard Wash Buffer Container has a maximum capacity of 2000mL.
The buffer container is stored on the inner left housing structure. Click
System Utilities>Tools>”Move Arm to Middle” to safely move the robotic arm
aside to access the buffer.
Figure 32
Figure 33
Disconnect the buffer container and carefully remove it from the instrument. Avoid making contact with the robotic arm or deforming the buffer adapter tubing.
Refill the container with fresh buffer, then reconnect the buffer container and place it back onto the instrument. Clean up any leaks or spills.
Click System Utilities>Tools>”Prime Z1 and Z2″ to prime the inline tubing, then visually inspect the lines to ensure that air bubbles are not present.
ONCORE Pro X, Operating Manual, Rev. L
31
Empty the Waste Containers
The standard Waste Containers have a maximum capacity of 4000mL, with a
secondary containment provided for each waste container. The waste containers
are stored underneath the instrument, fed by gravity through the waste tubing.
Click System Utilities>Tools>”Move Arm to Middle” to safely move the robotic
arm aside from the waste stations. Carefully pour some water through the waste
stations to flush out the waste tubing.
Figure 34
Disconnect the tubing from the Waste Container caps and empty the waste
containers. Dispose of the waste according to local regulations. Reconnect the
waste tubing: the black tubing is designated for hazardous waste and the blue
tubing is designated for non-hazardous waste. Verify that the waste tubing is
draining properly, making sure that the inserted end of the tubing clears the
maximum waste level line. Adjust the angle and free length of the tubing as
needed.
ONCORE Pro X, Operating Manual, Rev. L
32
4.8. Reagent Check: Scan Reagents
Remove all reagent vial caps before proceeding. From the Main screen, click
“Scan Reagents” to open the Reagent Check screen. The reagent map is displayed
on the left and a table of reagents required for the run is displayed on the
right.
Figure 35
Tip: Click “Print Required Reagents” to print the list of required reagents. A
network or USB printer is required.
Figure 36
ONCORE Pro X, Operating Manual, Rev. L
33
To scan all reagents with the RFID reader, close the door and click “Scan
Reagents”. The robotic arm will move the RFID reader antenna over each reagent
vial position and read out the stored RFID tag information. The system will
continue scanning until all required reagents have been detected or all 40
positions have been scanned. Tip1: Click + drag to select specific reagent
vial positions to scan. Tip2: Click “Stop” to stop the scan while it is in
progress.
Figure 37
The scanned reagent assignments are displayed on the reagent map on the left.
Each scanned reagent vial will be displayed with the following:
A. Reagent name. B. Number of tests remaining. Updates during the staining
process each time reagent is aspirated. C. Expiration date. D. Unique RFID tag
ID. Provides traceability for the individual reagent vials across different
runs.
ONCORE Pro X, Operating Manual, Rev. L
34
After completing the scan, the system will display the current status of the reagents: A. All expired reagents are highlighted in brown and must be replaced. B. All missing reagents are highlighted in pink in the reagent list. C. All reagents with insufficient number of tests are highlighted in yellow. D. All reagents ready for the staining run are highlighted in green. E. All reagents not used will remain with a white background. F. All RFID tags that produced an error during scanning are highlighted in red and must be replaced.
Figure 38
The user must add additional vials when the reagent is insufficient or
missing. Up to 5 vials of the same reagent may be placed on the rack. Manual
assignment or a re-scan of the reagent rack is required to proceed.
Note: A warning will appear when insufficient reagent is detected. Additional
tests may be added either by 1) refilling the reagent vial and reprogramming
the test number, or 2) adding additional vials of reagent. Upon re-scan, the
reagent will highlight in green once the test number is found to be
sufficient.
The user may need to reprogram or replace the RFID tag when the reagent is not
scanned properly. In such circumstances, users must manually verify the volume
remaining in the vial to determine the actual number of tests left.
A. Not Exist. The scanned reagent name does not exist in the system. B. Data
Read Error. The RFID tag failed to scan correctly. C. Tag Not Updated. The
RFID tag information was not updated correctly in the previous run. D. RFID
tag is not found. The RFID tag failed to scan completely.
Note: Contact Technical Support for help troubleshooting recurrent cases of
RFID misreading or failure, as this may indicate an issue with the system’s
RFID Reader system. For more information on reading and writing RFID tags,
refer to Preparing RFID Reagent Tags with the RFID Tag Editor p. 78.
ONCORE Pro X, Operating Manual, Rev. L
35
To manually assign a missing reagent to the Reagent Map, select the missing
reagent in the table on the right, then click a position on the reagent map to
assign the exact number of tests required to this reagent rack position.
Note: The user has the capability to manually assign missing or insufficient
reagents; however, it is recommended to reserve its use only in the case of an
RFID reader or RFID tag failure. Manually assigned reagents cannot be tracked
by the system, so it will be the user’s responsibility to keep track of the
vial contents and gauge the number of tests remaining.
Figure 39
The manual assignments will be saved until the reagent rack is scanned again.
G. All manually added reagents are highlighted in dark green. Always verify
that the manually assigned reagent is placed in its corresponding position on
the Reagent Rack and that it has sufficient volume before proceeding.
To review a previously run Reagent Map, click “Load Reagent Map File”, and
select the map file from the list of reagent map logs. Each log is labeled
with the date and time of the run. Open the log and it will be displayed on
the reagent map on the left.
Figure 40
ONCORE Pro X, Operating Manual, Rev. L
36
4.9. Starting the Run
To immediately start the run, close the door and click “Start Staining
Process” from the Reagent Check screen.
Verify that all caps have been removed from the reagent vials before starting
the staining process. Close the door before starting the run. The door will
lock and block unsafe access to the instrument while it is in operation. The
staining run will not proceed while the door is open.
The system will initialize and lock the door, then proceed to calculate the
most efficient schedule to complete the staining run. Longer protocols are
prioritized over shorter protocols to enable the slides to finish at a similar
time.
Figure 41
Once all calculations are complete, the system will unlock the door and prompt
the user to verify the volume of inline wash buffer and the remaining waste
capacity.
A. The robotic arm will automatically move to the center of the instrument to
allow the user to add more buffer if needed.
B. The waste containers are located under the instrument work surface and
should be emptied as needed. Note: The calculated waste volume reflects the
total volume of waste, combining both hazardous and non-hazardous waste. In
the event of an overflow of the non-hazardous waste station, all waste will be
dispensed into the hazardous waste container. Therefore, it is important to
ensure that both waste containers have sufficient capacity.
Refer to Managing Buffer and Waste, p.31, for more information.
Click “OK” to proceed with the staining process. The staining process will not
proceed until the user completes this verification step.
ONCORE Pro X, Operating Manual, Rev. L
37
As the instrument prepares to start the run, the system will prime each pump
three times over the wash stations to fill the in-line tubing. A run timer
will appear at the top of the screen, counting down the expected time
remaining until the end of the run.
Figure 42
Tip: Runtime will vary depending on the protocol template(s) used and the run
complexity (number of slides and types of protocols selected). Selecting more
slides and several protocols running multiple detection systems will result in
a longer runtime. To increase overall lab throughput, it is recommended to run
shorter runs during the day and save longer, more complex runs for overnight.
To delay the start of the run to a later date or time, click “Delay Start
Staining Process” from the Reagent Check screen:
Figure 43
Enter the Date and Time that the run is expected to finish. The maximum delay
start is 72 hours. The system will calculate the delay time and initiate a
countdown. When Time Left to Start reaches 0:00:00, the system will
automatically begin the staining process.
Figure 44
Tip: Use Delay Start on overnight runs to prevent processed slides from drying
out.
ONCORE Pro X, Operating Manual, Rev. L
38
Gantt Chart
The current run schedule is displayed on the Gantt chart. This may aid in
tracking the progress of the run and determining the wait time for
loading/unloading slides.
To view the Gantt chart, expand the AutoStainer Server window from the taskbar
and click the Gantt tab.
Figure 45
The purple line tracks the progress of the staining run. The time is measured
in thousands of seconds. Tip: Hover the mouse over the bar to display the name
of the task within the protocol.
Each bar represents a batched task for a group of slides. The maximum group
size is 18. 1) The red bar represents a task corresponding to a reagent
addition. 2) The blue bar represents a task corresponding to a quick wash
using wash buffer. 3) The green bar represents a task corresponding to a quick
wash using a reagent 4) The purple bar represents a task corresponding to an
incubated wash. 5) (Special) The brown bar represents two linked tasks for
mixing of 2 RTU reagents.
Sequential tasks are arranged horizontally on the Gantt chart following the
sequence of steps in the staining protocol. One or more tasks may be running
simultaneously within the course of the staining run, based on the grouping of
the slides and protocols. Simultaneously running tasks (from different groups
of slides) are arranged vertically on the Gantt chart.
ONCORE Pro X, Operating Manual, Rev. L
39
4.10. Auto Start
The Auto Start feature may be used to start the staining process directly from
the Main screen. The system will automatically proceed from scan slides to
scan reagents and skip the buffer and waste verification step to immediately
start the run. This may reduce instrument startup time.
This feature requires certain setup conditions for use: A. The user has loaded
all slides onto the instrument and labelled with a 2D barcode. B. The user has
loaded all required reagents onto the instrument and programmed the vials with
RFID tags. C. The user has verified that the buffer volume and waste capacity
are sufficient.
To use Auto Start, close the door and click “Auto Start” on the Main screen
and enter the number of slides loaded onto the instrument.
Figure 46
4.11. Completing a Run
The system will remain locked while the staining run is in progress. Limited
access is provided to prepare labels. In the case of an emergency, click
“Exit” to abort the staining run. All run progress will be lost.
Please heed the following precautions while the run is in progress: Disable
Wi-Fi and local internet connection. Do not attempt to log in/out of Windows
or switch users while the instrument is still in operation. This will
cause a fatal system error and cause the system to crash. Do not attempt to
run other applications in the background when the instrument is in operation
i.e.: Team
Viewer/remote control session, Windows Update, Internet Explorer, Antivirus
etc., as it may result in an unexpected system crash.
Do not attempt to change file/folder settings or screen display/appearance
settings when the instrument is in operation ie: locking the Taskbar, as it
may result in an unexpected system crash.
The Autostainer Server, Robot, Slide Processor, and RFID Processor Terminals
will be running while the instrument is in operation. Do not attempt to close
any of these applications as it may result in an unexpected system crash.
Figure 47
ONCORE Pro X, Operating Manual, Rev. L
40
Runtime Monitoring
While the instrument is running, the user may track the progress of the run
from the Main screen. The slide map displays all 36 slides and displays the
protocol, time, and the current step at each slide position. A system log is
displayed at the bottom of the screen.
Figure 48
The system monitors the instrument for errors during the staining process. If
it encounters any notable errors during its operation, it will attempt to
recover from the error and proceed to the next step. Any slide(s) that may
have been affected by the error will be marked in yellow and labeled with the
step in which the first error occurred. The user is advised to verify the
staining results of potentially affected slides and review the Slide Report.
Please contact Technical Support if encountering persistent system errors.
A. The system monitors the heater temperature during the staining process. If
a heater fails to heat or cool down to the temperature set point, its position
on the slide map will be marked in yellow and display a “Low Temperature” or
“Overheating” warning. The user is advised to perform a heater check after the
run to verify the heater function. Refer to Check for Module Malfunction, p.
53, for more information.
B. The system monitors the module function during the staining process. If a
module fails, its position on the slide map will be marked in red and display
a “Module Malfunction” warning. The module will be disabled and cannot be used
again until the error has been cleared. The user is advised to perform a
module check after the run to verify the module function. Refer to Check for
Module Malfunction, p. 53, for more information.
C. The system verifies the liquid level in the reagent vial before drawing
reagents. The Z1 probe is equipped with a liquid level sensor. If a low volume
or no liquid is detected, the system will mark all affected slides yellow with
a “Low Volume” warning. The user is advised to verify the volume in the
reagent vial(s) after the run to verify the error and identify potentially
affected slides. Reagents that have low conductivity or high viscosity cannot
be reliably detected by the liquid level sensor. The system may show a low
volume warning for such reagents. In this case, the user may visually inspect
the reagent vials to confirm that sufficient volume was available during the
staining run.
D. (If the Overflow Detection option is enabled), the system verifies the
liquid level in the waste overflow bin before dispensing waste. If liquid is
detected, the system will alert the user of a potential overflow. When an
overflow is detected in the non-Hazardous waste, the system will automatically
switch over to use the
ONCORE Pro X, Operating Manual, Rev. L
41
Hazardous Waste to prevent further overflow. The user is advised to verify that the waste tubing is draining properly and/or empty the waste container(s).
ONCORE Pro X, Operating Manual, Rev. L
42
End of Run
The system will highlight the slides that have successfully completed staining
in light green and mark as “Finished”.
Figure 49
After completing the final step of the run, the system will prime both pumps
to clean the syringes, unlock the door and prompt the user: “Run Successfully
Ended”. The slide chambers are filled with wash buffer during the last step of
the staining protocol to prevent the slides from drying out before they are
unloaded from the system. However, if an error or warning was encountered
during the run, the system will instead prompt the user: “Run Ended with
Warnings”.
Figure 50
The user can review the slide map and report data to identify the slide(s)
which may have been affected by the error. Refer to Section 6: Reports, p. 93,
for more information.
Do not close any UI applications or shut down the instrument before the seeing
the Finished Notification message. The instrument may still be in operation.
ONCORE Pro X, Operating Manual, Rev. L
43
Aborting a Run
If the system encounters an irrecoverable error, it will immediately abort the
staining process. The aborted slides will be highlighted in orange or yellow
and the system will prompt the user: “Run Aborted Due to Script Error”. The
staining run progress cannot be recovered upon system abort.
Contact Technical Support immediately to verify that the system abort is not
caused by a serious issue on the instrument.
Figure 51
Note: In an emergency case, the user may manually abort the staining process.
Click “Exit” to close the main interface or power off the system to
immediately stop the instrument.
ONCORE Pro X, Operating Manual, Rev. L
44
4.12. Continuous Loading
The Continuous Loading feature allows users to load more slides after the
staining process has already started. Note: Using this feature may
significantly increase the total run time and delay the finish time for all
remaining slides in the run. This feature should not be used for time
sensitive reagents such as Fluorescence In-situ Hybridization (FISH). Please
call Technical Support for further assistance.
Stop the Current Run
To stop the current run, click “Load New Slides”. The system will stop the
progress of each slide as it reaches a safe stopping point in its staining
protocol. The wait times will vary for each circumstance.
Figure 52
After the system has stopped the progress of all slides, the door will unlock
and a popup message will alert the user “Run Successfully Stopped”. The slides
that have been stopped will be highlighted in light coral and marked
“Stopped”.
Figure 53
ONCORE Pro X, Operating Manual, Rev. L
45
Never exit the Main screen after the run has been stopped. The progress on all
stopped slides will be deleted and cannot be recovered.
Restart the Run
Prepare and load new slides. The system will automatically lift the lids of
all available slide positions, including empty positions and positions marked
as “Finished”. The positions marked as “Stopped” will be set at the lid Down
position and those slides cannot be unloaded from the instrument.
Unload all Finished slides before loading any additional slides.
Scan the new slides to automatically assign the protocols, or manually assign
new protocols to the available slide positions. The positions marked as
“Stopped” are locked and cannot be assigned a new protocol.
Figure 54
Load any additional reagents that may be required by the new slides and re-
scan the reagent rack.
Figure 55
ONCORE Pro X, Operating Manual, Rev. L
46
Re-start the staining process. The system will reschedule the run to incorporate the new slides. The new slides will take priority and will run separately until they catch up with the Stopped slides. The Stopped slides will resume progress when the new slides have reached the same point as the Stopped slides.
Figure 56
Continuous loading can be repeated on successive steps in the new run
schedule; however, the overall runtime for the Stopped slides will increase
with each load of slides.
4.13. Shutting Down the Slide Stainer
From the Main screen, double click “Exit” to close the program. Click “Yes” to
confirm.
Figure 57
This process may take a few minutes as the system waits for all module lids to
go the extract position and for all sub-applications to close. Wait for the
application to close completely and return to the login screen before powering
off the instrument
Push and hold the power button on the lower right side of the instrument to
power down. Wait a few minutes after powering down to ensure the instrument
powers down completely.
(Optional) Turn off the power supply switch on the back of the instrument.
Shut down the computer from the Start Menu.
Do not set the module lids to the down position for any extended period of
time. Modules should rest in the extract position when idle.
ONCORE Pro X, Operating Manual, Rev. L
47
Before shutting down the Slide Stainer for an extended period of time (i.e.: long weekends, holidays, transport), prime the pumps and tubing lines with distilled water to prevent excess buildup of salt. Clean the modules with a moist paper towel to remove any residual buffer that may dry along the hinge slot and slip holder.
ONCORE Pro X, Operating Manual, Rev. L
48
Section 5. System Utilities
From the Main screen, click “System Utilities” to access additional utility
functions and advanced features. Some features may be restricted depending on
user’s security access level. Refer to Security, p. 83, for more information.
5.1. Tools
Perform manual operations with the system using the functions found under the “Tools” tab. *Note: Depending on the configuration of the system, some of these functions may not be available to all users.
Figure 58
1. Baking Slides 2. Load Slide Map File 3. SP1-Z1 Calibration 4. SP1-Z2 Calibration 5. Barcode Calibration 6. Tubing Clean 7. (Option) Tubing Clean Log 8. Move Arm to Middle 9. (Option) Module Chamber Tracking 10. Generate Reports 11. Initialize System 12. Check Modules 13. Prime Pump Z1 and Z2 14. Prime Pump Z1 15. Prime Pump Z2 16. Lock Door 17. Unlock Door 18. Backup Report Data 19. Replace Pump 1 Syringe 20. Replace Pump 2 Syringe
Prepare Slides with Baking Slides Function
Click “Baking Slides” to select slides for baking. Check the boxes to select
an entire row or select individual slides by module position. Refer to Baking
Slide Settings, p. 84, for more information.
Figure 59
Click “Start” to begin baking. The timer on the left will begin counting down
and automatically shut heaters off once the timer reaches 0:00:00. Click
“Stop” to manually stop baking.
Refer to Baking Slides Settings, p. 84, to set up slides for automatic baking
before every run.
ONCORE Pro X, Operating Manual, Rev. L
49
Load a Saved Slide Map File
Click “Load Slide Map File” to load and view a previously run slide map on the
Main screen. The system automatically saves the slide map file for each set of
slides run on the instrument.
Figure 60
Verify the Calibration of SP1-Z1 or SP1-Z2
Click “SP1-Z1 Calibration” to move the Z1 probe to module #1 and verify the
XYZ position of the Z1 probe. Click “SP1-Z2 Calibration” to move the Z2 probe
to module #1 and verify the XY-Z position of the Z2 probe.
Figure 61
Verify the Calibration of the Barcode Reader
Click “Barcode Calibration” to move the Barcode Reader to module #1 and verify
the XY position of the barcode reader.
ONCORE Pro X, Operating Manual, Rev. L
50
Clean Z1/Z2 Tubing
Click “Tubing Clean” to open the tubing cleaning feature. Use the Tubing
Cleaning Kit or fill two 15mL vials with an appropriate cleaning solution.
Place Z2 cleaning solution in reagent rack position #5. Recommendations: Use
Slide Brite to remove wax residues in the Z2 tubing.
Place Z1 cleaning solution in reagent rack position #1. Recommendations: Use
1% Bleach to remove Hematoxylin/DAB precipitate in the Z1 tubing.
Figure 62
Close the door and click “OK” to initiate the tubing clean process. The Z1 and
Z2 probes will both aspirate 5mL of cleaning solution into the tubing from
their respective reagent rack positions. The system will then initiate a 20
minute countdown.
Figure 63
After 20 minutes has elapsed, the system will automatically perform a System
Initialization and purge the cleaning solution waste. Click “Cancel” at any
moment to immediately purge the waste from the tubing.
ONCORE Pro X, Operating Manual, Rev. L
51
(Optional) Access the Tubing Clean Log
If Tubing Clean Tracking is enabled, the system tracks the number of slides
processed since the last tubing clean. If the total number of slides run
exceeds the set threshold, the system will block the user from initiating a
new staining process. Refer to Tubing Clean Tracking Option, p. 88, for more
information.
Click “Tubing Clean Log” to review the tubing cleaning status. Perform a
tubing clean to reset the counter or click “Reset” to manually set the counter
to 0.
Figure 64
Click “Show Tubing Clean Records” to review the entire log of tubing cleans
performed to date. Click “Remove” to clear all cleaning records before the set
Date Before. (Alternative option) Click “Remove All” to clear all tubing
records-to-date.
Figure 65
(Optional) Reset the Module/Chamber Usage Counter
If Module Chamber Tracking is enabled, the system tracks the number of slides
processed on each module. If the total number of slides run exceeds the set
threshold, the system will display a red warning message to notify the user
that it is time to replace the chamber. Refer to Module Tracking Option, p.
89, for more information.
Figure 66
Click “Reset” to reset the counter for selected modules back to 0 after
replacing the chambers.
ONCORE Pro X, Operating Manual, Rev. L
52
Generate Reports
Click “Generate Reports” to open the Report Generator utility. Refer to
Section 6: Reports, p. 93, for more information.
Move Robot XYZZ Arm to Middle
Close the door and click “Move Arm to Middle” to move the robotic arm to the
center of the instrument. This will enable the user to gain access to the left
side of the instrument to fill the buffer bottle and inspect wash stations,
pumps (if applicable) etc.
Perform System Initialization
Close the door and click “Initialize System” to initiate system
initialization. The robot will initialize the XYZZ axes while all 36 module
lids move up to Home. Then each pump will initialize over its respective waste
station.
Check for Module Malfunction
Click “Check Modules” to check the function of the modules. The Check Modules
window will open to display 36 module checkboxes. Select the modules using the
check boxes then select an option to perform the check:
Figure 67
Click “Start” to initialize the system and start the countdown timer. The
modules will heat up to a high temperature >100C.
Take precaution not to touch the modules until the heater check has completed
and the modules have cooled down.
Click “Cancel” to cancel the module check at any time.
Figure 68
ONCORE Pro X, Operating Manual, Rev. L
53
- Select the Heater option to check for heater malfunction only. The system will preheat the heaters to 40C, then heat to 103C and verify the temperatures. The system will highlight any malfunctioning heater in red on the slide map and mark “Low Temperature” or “Overheating”; otherwise, the system will return the message: “All heaters are functioning properly.” The heaters will take several minutes to heat to higher temperatures during the heater check. To avoid burns/injury, do not resume operation until the heaters have cooled down sufficiently.
- Select the Motion option to check for module motion malfunction only. The system will move the lids to verify all module agitation heights: Home, Extract, Inject and A1-A7. The system will highlight any malfunctioning module in red on the slide map and mark “Module Malfunction”; otherwise, the system will return the message: “All modules are functioning properly.”
- Select the Motion and Heater option to check for both module motion and
heater malfunction. The system will perform both the heater check and the
motion check on the selected modules.
Figure 69
Protocols may not be assigned to a slide position with a malfunctioning heater or module until the module is replaced or repaired, and a new module check is performed to clear the error. Contact Customer Service to repair/replace any malfunctioning heaters or modules.
ONCORE Pro X, Operating Manual, Rev. L
54
Flush Tubing with Prime Pump Z1 and Z2 Function
To flush out the tubing, connect a bottle of distilled water. To clear the
tubing lines, connect an empty bottle. Click “Prime Z1 and Z2” to prime each
pump 5x over the waste station.
Prime Pump Z1 or Pump Z2
Pump 1 is connected to the Z1 tubing line. Click “Prime Pump Z1” to check the
function of Pump 1. The Z1 probe will move to the wash station and dispense
one full draw of buffer.
Pump 2 is connected to the Z2 tubing line. Click “Prime Pump Z2” to check the
function of Pump 2. The Z2 probe will move to the wash station and dispense
one full draw of buffer.
Lock or Unlock Door
Click “Lock Door” to manually lock the door. Click “Unlock Door” to manually
unlock the door.
Backup Report Data
Click “Backup Report Data” to make a backup copy of the report data. Refer to
Section 6: Reports, p.93, for more information.
Replace Pump 1 Syringe or Pump 2 Syringe (Distributor Access only)
To prevent leaks/water damage, prime the pump several times using “Prime Z1
and Z2” to clear the fluid in the tubing line before removing the syringe. Dry
the syringe and port valve opening before installing the new syringe.
Click “Replace Pump 1 Syringe” to move the plunger into position before
removing the pump 1 syringe. Click “Replace Pump 2 Syringe” to move the
plunger into position before removing the pump 2 syringe.
ONCORE Pro X, Operating Manual, Rev. L
55
5.2. Editors
Access the protocol, reagent, and RFID editors under the “Editors” tab.
Figure 70
Editing Protocols in the Protocol Editor: Overview
Click “Protocol Editor” to open the Protocol Editor window. The system is pre-
installed with a set of protocols optimized to run on the instrument; however,
the Protocol Editor provides several modes of customization to fine-tune the
protocols to accommodate the individual tissue and staining requirements of
each lab:
a. Adjust key protocol customizations b. Change protocol templates c. Add new
protocols d. Create and modify special protocols (Distributor Access only) e.
Create and modify protocol templates (Distributor Access only)
Figure 71
Protocols are grouped under separate tabs for each detection system type,
i.e.: HRP, AP, ISH, FISH. Note: The available detection system types may vary
based on the system configuration. Every protocol (except those listed under
Special and Template) is assigned a protocol template. Protocol Templates are
grouped under the Templates tab. The protocol template is used as the base to
generate all protocol parameters, excluding the primary reagent and the saved
protocol customizations. Special protocols are grouped under the Special tab.
Special protocols are stand-alone protocols that are not generated from a
protocol template.
ONCORE Pro X, Operating Manual, Rev. L
56
Click “Version” to display the manufacturer’s protocol version control number
(if applicable). The version control number may be updated each time the
manufacturer releases a new set of default protocols and reagents. Note: User-
added protocols do not affect the version control number.
Figure 72
Click “Protocol List” in the main window to display the complete list of all
protocols, including negative control protocols, special protocols, and
templates. Note: Click “Save” to export the list to an external text file.
Figure 73
ONCORE Pro X, Operating Manual, Rev. L
57
Adjusting Protocol Customizations: IHC
Select the appropriate tab in the Protocol Editor to access all protocols
grouped under that detection system type.
Figure 74
Select a protocol by clicking the empty cell in the left column of the editor.
The selection will highlight in blue.
To display the full protocol, click “Edit” to open the Advanced Protocol
Editor. Note: The user cannot edit the protocol directly from this window
because the protocol is generated from a template.
Figure 75
Most parameters displayed in the Advanced Protocol Editor are determined by
the protocol template; however, the user has the option to make further
customization by changing the protocol template or creating a special
protocol. Refer to “Creating and Modifying Protocol Templates”, p. 65, and
“Creating and Modifying Special Protocols”, p. 64, for more information.
ONCORE Pro X, Operating Manual, Rev. L
58
The user may adjust the key protocol parameters directly in the main window: pretreatment, antigen retrieval, peroxidase blocking and antibody incubation.
Figure 76
To customize the pretreatment reagent, select from the dropdown list under DS
Buffer Option. A. DS2-50. This is the default option. Dewax solution 2 is
applied to enhance antigen retrieval. B. DS Buffer. Select this option to
apply wash buffer instead of Dewax solution 2. This option may improve
morphology or improve over-staining by reducing the strength of antigen
retrieval. C. DSE-50. Select this option to apply a protease enzyme, incubated
at 58C, instead of Dewax Solution 2.
To customize the antigen retrieval reagent, select from the dropdown list
under AR Option. A. AR1. This is the default option. Antigen Retrieval
solution 1, high pH 9 is applied. B. AR2. Select this option to apply Antigen
Retrieval solution 2, low pH 6. This may improve morphology or
improve over-staining by reducing the strength of antigen retrieval. C. AR
Buffer. Select this option to apply wash buffer instead of Antigen Retrieval
solution. This option may be
used when antigen retrieval is not required. Note: Set incubation temperature
to 37C to reduce evaporation. D. AR Enzyme. Select this option to apply a
custom enzyme instead of Antigen Retrieval solution.
Note: Set incubation temperature to 37C. Requires 370uL/test fill volume.
Note: It is recommended to create an alternate protocol template or special
protocol because the high volume may cause some overflow.
To customize the antigen retrieval incubation temperature, adjust the Temp (
C). A. 101C is the default temperature for antigen retrieval. B. Reduce the
temperature to 98C /95C (or lower) to reduce the strength of antigen
retrieval. C. Increase the temperature to 103C or higher to increase the
strength of antigen retrieval. D. Reduce the temperature to 37C if an enzyme
or buffer option was selected.
To customize the peroxidase blocking reagent, select from the dropdown list
under Block Option. A. Buffer. This is the default option (no blocking step).
AR1 is formulated to provide sufficient blocking under
most conditions. B. Block. Select this option to apply H2O2 block instead of
wash buffer.
To customize the primary antibody incubation time and temperature, adjust the
Time (hh,mm,ss) and Temp ( C). A. 25C is the default temperature for antibody
incubation. B. Increase the incubation temperature to 37C to increase staining
intensity. C. Decrease the incubation time (minimum 10min) to decrease the
staining intensity. D. Increase the incubation time to 45min/1hr to increase
staining intensity.
Note: Increasing the incubation time (or temperature) may affect the staining
area by increasing evaporation. It is recommended to adjust the antibody
dilution to maintain an incubation time 30min.
ONCORE Pro X, Operating Manual, Rev. L
59
Click “Save” to apply and save the changes to the selected protocol(s). All
protocols that have been modified from the manufacturer’s default protocol
will be “Marked as Changed”.
The user may backup these protocol customizations under System
Utilities>Settings> “Save All Customizations”. Refer to p. 87, Save All
Customizations, for more details.
Figure 77
Click “Cancel” to cancel all unsaved changes. Click “Return” to exit the
Protocol Editor.
Adjusting Protocol Customizations: ISH/CISH/FISH
Select the appropriate protocol tab in the Protocol Editor to access ISH,
CISH, or FISH protocols.
Figure 78
To customize the enzyme incubation time and temperature, adjust the Time
(hh,mm,ss) and Temp ( C). Click “Save” to apply and save the changes to the
selected protocol(s).
The user may backup these protocol customizations under System
Utilities>Settings> “Save All Customizations”. Refer to p. 87, Save All
Customizations, for more details.
ONCORE Pro X, Operating Manual, Rev. L
60
Refreshing Protocol Customizations
The user may save and restore the protocol customizations after making a
change in the Protocol Editor. Backup the protocol customizations before
refreshing the protocols, if necessary. Any recent
customizations that have not been backed up using “Save all Customizations”
will be overwritten during the refresh process. Refer to Save All
Customizations, p. 87, for more details.
Click “Protocol Editor” to open the Protocol Editor window. Select the
protocol (s) by clicking the empty cell in the left column of the editor. The
selection will highlight in blue.
Click “Refresh” and select the desired refresh option, then click “OK”.
A. “Refresh with saved customization” will restore the protocol(s) using the
user’s saved customizations.
Figure 79
B. “Refresh with manufacturer default” will reset the protocol(s) to the
manufacturer’s customizations.
Figure 80
Tip: Select multiple protocols by holding down the Ctrl key. Select all
protocols in the tab by clicking the cell in the upper-left corner.
ONCORE Pro X, Operating Manual, Rev. L
61
Changing the Assigned Protocol Template
Some staining protocols may require further customization beyond the available
parameters accessible in the main Protocol Editor window. In such
circumstances, the user may assign a different protocol template.
Contact Technical Support for assistance. Only users with Distributor access
may create or modify protocol templates. Refer to “Creating and Modifying
Protocol Templates”, p. 64, for more information.
Click “Protocol Editor” to open the Protocol Editor window. Select the
protocol by clicking the empty cell in the left column of the editor. The
selection will highlight in blue.
To select a different protocol template, click “Change Template”. The Select
Template window will open and display the Protocol Templates available for
this Protocol Type. The current assigned protocol template is highlighted in
blue.
Note: Make sure to save all current protocol customizations in Settings before
proceeding. The protocols must be regenerated after changing the protocol
template. Any unsaved customizations will be overwritten.
Figure 81
Select the desired protocol template and click “OK” to change the template
assignment. The system will automatically regenerate the protocol using the
selected template and apply the saved protocol customizations.
Figure 82
ONCORE Pro X, Operating Manual, Rev. L
62
Adding New Protocols in the Protocol Editor
The user may create new protocols to test third-party antibody products,
different antibody titers, and various customization options to aid in
staining optimization.
To add a new protocol, select the appropriate protocol tab of the Protocol
Editor, and click “Add”. Input a unique name in the field for Protocol Name.
Each protocol must have a unique name. Append a short suffix to the end of the
protocol name to distinguish the detection system used or modifications made.
To avoid conflicts with positive or negative controls in the programming
software, please refrain from using the symbols “+” and “-“at the end of the
protocol name.
Figure 83
Select the Reagent Type and Reagent Name of the primary reagent. Refer to
“Adding New Reagents in the Reagent Editor”, p. 75, for more information on
adding new reagents.
Select the Template Name, if more than one template is available. Refer to
“Creating and Modifying Protocol Templates”, p. 64, for more information on
adding new templates.
Click “Add” to generate a new protocol from the selected protocol template.
Note: New protocol(s) will be generated with default customizations based on
the protocol template. The user must customize the new protocol(s) based on
the desired test parameters, vendor’s datasheet, or manufacturer’s
recommendations.
Click “Return” to close the window after all protocols are added.
To delete a protocol, select the protocol from the left-most column and click
“Delete”. Note: The system may restrict the user from deleting certain
protocols created by the manufacturer.
ONCORE Pro X, Operating Manual, Rev. L
63
Creating and Modifying Special Protocols (Distributor Access only)
Special protocols are grouped under the Special tab of the Protocol Editor.
These stand-alone protocols offer users the flexibility to customize each step
of the staining protocol to fulfill the special requirements of new staining
applications, ie: Special Stains, IF, Multiplex etc.
Only users with Distributor access may create special protocols. Contact
Technical Support for assistance. New protocols may be imported to the system
using the Protocols and Reagents Manager. Refer to p. 72, Importing Protocols
in the PR Manager.
Select the Special tab in the Protocol Editor to view the list of special
protocols.
Figure 84
To add a new special protocol, click “Add”. Enter the Protocol Name and click
“Add” to generate a basic one-step protocol.
The protocol name must be unique. Refer to the Protocol List to view the full
list of protocols.
Figure 85
The Advanced Protocol Editor will automatically open to display the details of
the special protocol. Insert additional steps and modify the special protocol
as needed. Refer to “Editing Protocol Parameters in the Advanced Protocol
Editor”, p. 68, for more information.
Protocol Name stores the name of the protocol. Protocol Type identifies the
protocol as Special. Primary Reagent is not identified because special
protocols are not generated from a template. Control Reagent is not identified
because special protocols are not customizable from the main window.
Description stores the description that will be displayed in the main window.
Open To stores the security access level of the protocol.
ONCORE Pro X, Operating Manual, Rev. L
64
Alternatively, right-click and Copy a protocol, template or special protocol
from any tab in the Protocol Editor. Note: It is recommended to select a
similar existing protocol to use as a base for generating the special
protocol.
Figure 86
Right-click and Paste the protocol into the Special tab and enter the Paste
Name.
Figure 87
To modify a special protocol, select the protocol in the Special tab and click
“Edit” to display the details of the protocol. Note: Users with Distributor
access may edit the special protocol directly from this window. Users with
Supervisor access may have limited permissions to edit certain steps only. To
delete a special protocol, select the protocol and click “Delete”. It is
recommended to delete special protocols that are no longer used for system
efficiency. Note: Manufacturer’s protocols may be protected by the system and
cannot be removed.
ONCORE Pro X, Operating Manual, Rev. L
65
Creating and Modifying Protocol Templates (Distributor Access only)
Protocol templates are grouped under the Templates tab of the Protocol Editor.
Users may create additional templates to apply additional protocol
customization(s) to a subset of protocols grouped under the same detection
type. ie: add protein block step, add enzyme digestion step, customize
counterstaining or customize antigen retrieval time.
Only users with Distributor access may create or modify protocol templates
protocols. Contact Technical Support for assistance. New templates may be
imported to the system using the Protocols and Reagents Manager. Refer to p.
72, Importing Protocols in the PR Manager.
Select the Template tab in the Protocol Editor to view the list of protocol
templates.
Figure 88
To add a new protocol template, click “Add”. Enter the Template Name and click
“Add” to generate a basic one-step protocol template with the Primary Reagent
automatically added.
The protocol template name must be unique and must contain the word, ” Temp”
Note: “Temp” is preceded by a space.
The template’s primary reagent must be set be the same as the protocol
template name. Note: This name is used as a space holder to indicate the
position in the template to be replaced by each protocol’s primary reagent
when the template is used to generate protocols.
Figure 89
ONCORE Pro X, Operating Manual, Rev. L
66
The Advanced Protocol Editor will automatically open to display the protocol
template. Select the appropriate Protocol Type or enter a new type. Insert
additional steps and modify the protocol template as needed. Refer to “Editing
Protocol Parameters in the Advanced Protocol Editor”, p. 68, for more
information.
Protocol Name stores the name of the template. Protocol Type identifies the
detection system type assigned to the template. Primary Reagent identifies the
primary reagent (antibody or probe) in the template. Control Reagent
identifies the reagent to be customized for incubation time and temperature in
the main window. Description stores the description that will be displayed in
the main window. Open To stores the security access level of the template.
Alternatively, right-click and Copy a protocol, template or special protocol
from any tab in the Protocol Editor. Note: It is recommended to select a
similar existing protocol to use as a base for generating the template.
Right-click and Paste the protocol into the Template tab and enter the Paste
Name.
To modify a template, select the template in the Template tab and click “Edit”
to display the details of the protocol template. Note: Users with Distributor
access may edit the protocol template directly from this window.
After saving changes to the protocol template, the system will automatically
regenerate all assigned protocols to apply the changes. Note: The user may
manually refresh all protocols with their assigned protocol templates by
clicking “Refresh All Protocols” in the Templates tab.
Figure 90
To delete a template, select the protocol and click “Delete”. All protocols
assigned to the template must be deleted or assigned to a new template before
the template may be deleted. Note: Manufacturer’s templates may be protected
by the system and cannot be removed.
ONCORE Pro X, Operating Manual, Rev. L
67
Editing Protocol Parameters in the Advanced Protocol Editor
Special Protocols and Templates may be modified using the Advanced Protocol
Editor.
Figure 91
Right-click the protocol step(s) or use the tool buttons to add and remove
steps: Select and “Copy” protocol step(s), then “Paste…” at another line in
the protocol. Select and “Cut” protocol step(s), then “Paste…” at another line
in the protocol. Select and “Insert…” a new step at any line in the protocol.
Select and “Delete” a step at any line in the protocol.
Click “Save” to save all changes. Click “Undo” to revert to the previous saved
state. Click “Return” to close the Advanced Protocol Editor.
ONCORE Pro X, Operating Manual, Rev. L
68
Edit the following parameters, as needed, for each step of the protocol or template. Refer to the Specialist manual for more details.
Figure 92
Set. Configure the RunSet pattern to optimize runtime scheduling. Note: It is
recommended to set the same standard RunSet pattern i.e.: AABBC(D variable)
across all protocols and protocol templates to ensure that they can be run
together efficiently in the same staining run.
Extract. Select the option to extract the waste from the previous step before
dispensing reagent.
Link. Select the option to link two consecutive steps. This will ensure that
the current step is immediately executed upon completion of the previous step,
without interruption from other processes running in the same staining run.
Note: Selecting this option may negatively affect overall runtime as it
reduces scheduling flexibility.
Max Group Size. Set the maximum group size smaller (default 18) to improve the
incubation accuracy of timesensitive reagents. The run scheduler will pool
common steps between protocols and execute them together. Groups of 18 slides
will have ~4 min longer “effective” incubation time compared to 1 slide, due
to the longer dispensation time for a larger group. Note: Setting a smaller
group size will negatively affect overall runtime by increasing scheduling
complexity.
Reagent Type and Reagent Name. Select the reagent.
Volume. Set the reagent volume. The volume drawn from the vial is augmented by
a factor of 1.08. The volume dispensed may vary slightly based on the
viscosity level setting.
ONCORE Pro X, Operating Manual, Rev. L
69
Incubation Type: A. Select Normal incubation. The “effective” incubation time
is slightly longer than the set time due to the reagent dispensation time,
which varies based on group size. B. Select Fine incubation to precisely
control the incubation time <1min. The protocol step will be executed
individually for each slide to ensure incubation accuracy (within seconds).
System fluid is dispensed at the end of this process. Note: This option may
greatly affect overall runtime when the incubation time is set >2min. C.
Select AirDry incubation to perform a Fine incubation with drying. No fluid is
dispensed at the end of this process. D. Select Accurate incubation to
precisely control incubation time >1min. This feature is currently in
development and is not available at this time.
Incubation Time. Set the total incubation time. Note: Setting long incubations
times may reduce the staining area, especially for heated processes.
Incubation Temperature Table. Configure the temperature and corresponding
heating time period. Note: Set the heat incubation longer than the Incubation
Time to perform heated washes and to maintain the temperature until the next
Step is executed.
Figure 93
Agitation. Configure the agitation parameters. The module lids will alternate
between two angles, Lid Up Position and Lid Down Position, at a certain
frequency, Lid Up Time and Lid Down Time. This mechanism helps to accelerate
the rate of reaction, improve uniformity, and disperse air bubbles.
Additionally, set the Threshold temperature to stop agitation above a certain
temperature set point. Note: Agitation angles must be set in consideration of
the reagent volume. Agitation frequency must be set in consideration of the
incubation time and temperature.
Figure 94
Wash Buffer Type and Wash Buffer. Select the quick wash reagent. A quick wash
is executed immediately at the end of incubation and will agitate twice
automatically. The default reagent is set to System Fluid (wash buffer).
and Volume. Select the number and volume of quick washes. Note: The quick
wash ensures that the reagent is promptly removed at the end of the Step without interruption from other processes running in the same staining run. Include a quick wash or linked wash step to control timesensitive incubation steps (within minutes).
ONCORE Pro X, Operating Manual, Rev. L
70
Configuring the Protocol Editor
Click “Protocol Editor Config” to configure the main Protocol Editor window.
This feature may be used when configuring the system for new protocol types.
Figure 95
Adjust the Protocol Types Display Order using the up/down arrows to change the
order of the tabs displayed in the Protocol Editor. The display order can be
changed to prioritize frequently used protocol types. Note: Remove unnecessary
Protocol Types by deleting all reagents of that type in the Reagent Editor.
Figure 96
Configure Select Editing Options to change the customization options for each
tab in the Protocol Editor. Select the Protocol Type from the list and select
the desired options: DS Buffer Option AR Option Block Option
Figure 97
Note: The Control Reagent option is always displayed and is set by the
protocol template.
ONCORE Pro X, Operating Manual, Rev. L
71
Importing Protocols in the PR Manager
Export packages are periodically released to the user in order to 1) update
protocol templates and protocols, 2) add additional applications and products,
or 3) add special protocols. These packages may be imported into the system
using the PR Manager utility. Contact Technical Support for more information.
Click “Protocol Reagent Manager” to open the Protocol and Reagent Manager
utility. Login with the User ID and Password.
From the Import tab, click “Browse” and select an export package (compressed
zip folder). Note: Do not unzip/extract the export package.
Tip: It is advised to make a backup copy of the data before making upgrades to
the SW or protocols. Please contact Technical Support for assistance.
The new protocol(s)/template(s)/special protocol(s) to be imported are listed
under Import Protocols. The associated reagent(s) to be imported are listed
under Import Reagents. The associated negative control link(s) to be imported
are listed under Import Negative Control.
Figure 98
ONCORE Pro X, Operating Manual, Rev. L
72
Note: The manufacturer may use the export function to remove obsolete
protocols and discontinued products. The protocols to be deleted are listed
under Protocols to Delete. The reagents to be deleted are listed under
Reagents to Delete.
Click “Import” to import the contents of the export package. A status bar will
be displayed to show the progress of the import process.
After the system has imported a new protocol, the system will automatically
refresh all protocols. Save all customizations before performing an import,
otherwise, some recent protocol modifications may be lost.
All new protocols, templates and reagents will be added to the system. All
existing protocols or templates that share the same name as an imported
protocol will be overwritten. All existing reagents that share the same name
as an imported reagent will be overwritten.
Note: Protocols cannot be transferred between different protocol versions. The
protocol import process will abort if the current protocol version number does
not match the version number of the export package. Please update to the
latest protocol version before importing a new protocol package.
Figure 99
Click “Display Current Protocol Version” to view the current protocol version
number. The version number will automatically update after importing a
manufacturer’s protocol update.
The manufacturer’s protocol updates must be imported in sequential order. The
current protocols must be up-to-date before installing the next manufacturer’s
release.
Click “Exit” to close the utility.
Exporting Protocols in the PR Manager (Distributor Access only)
Click “Protocol Reagent Manager” to open the Protocol and Reagent Manager
utility. Login with the User ID and Password.
Click the Export tab to view the export options:
Click “Browse” to designate the From Folder containing the protocol(s) to be
exported. Click “Browse” to designate the destination Export Folder for the
export package.
Enter the Export File Name.
Select the protocol(s) to be exported under Available Protocols to Export.
Note: Protocols of different detection types, Templates or Special protocols
may be selected using the dropdown list for Type.
ONCORE Pro X, Operating Manual, Rev. L
73
Figure 100
Click “Add” and “Remove” to modify the list of Protocols to Export. The
Reagents to Export and Negative Controls to Export are automatically added.
Ensure that the Importer’s Version and Exporter’s Current Version Control
numbers match. Alternately, select the option Base Version Update to change
the Importer’s Version Control number upon import of the package.
Click “Export” to create an export package in the designated destination
folder.
Figure 101
ONCORE Pro X, Operating Manual, Rev. L
74
Adding New Reagents in the Reagent Editor
The user has the option to add third-party antibody products or other approved
reagents to the system. This may be necessary when certain reagents are not
offered in the Oncore Pro product line.
Click “Reagent Editor” to view and manage the system’s reagents. Note: The
reagent list may vary based on the system configuration and user’s security
access level.
Figure 102
To add a new antibody (or antibody titration), select the Add New tab. Select
the Antibody group listed under Reagent Types. Antibodies are grouped
alphabetically.
Input a unique antibody name in the Name field. Each unique antibody product
must have a unique name i.e.: different clone #, manufacturers,
dilutions etc.
Figure 103
The option “Open To” is set by default to the user’s security access level.
The option “This reagent is ready to use” is set by default. This field stores
the mixing parameters of the reagent.
Check the box Hazardous to designate the reagent as hazardous, if needed. All
hazardous waste will be dispensed in the designated hazardous waste station.
The Viscosity Level is automatically set to 1 by default. This field
determines the liquid handling parameters of the reagent.
Click “Save” to add the new antibody to the system.
ONCORE Pro X, Operating Manual, Rev. L
75
To add a new reagent, select the Add New tab. Select Negative, Others or
Special under Reagent Types. Note: The list of available reagent types may
vary based on the system configuration.
Input a unique reagent name in the Name field. Contact Technical Support
before using third-party reagents. Certain reagents may be chemically
incompatible and may cause damage to the instrument. The option “Open To” is
set by default to the user’s security access level. Set the mixing parameters
of the reagent:
A. Select the option “This reagent is ready to use” for reagents applied
directly to the slide. B. Select the option “This reagent is mixed on-board
with the following parameters” for 2 reagents to be
mixed directly on the slide. Reagent mixtures that have a short activity time
must be mixed on-board. Note: Reagent A and B should have a low mixing ratio,
i.e.: 1: 1.
C. Select the option “This reagent is mixed on-rack with the following
parameters” for 2-3 reagents to be mixed on the reagent rack in a separate
mixing vial before application to the slide. Note: Reagent A, B (and C
optional) may have a higher mixing ratio, i.e.: 1:20 or 1:30; however, the
user must set an appropriate minimum mixing volume to ensure the compositional
accuracy of the mixture. The maximum mixture volume must not exceed 16mL.
Check the box Hazardous to designate the reagent as hazardous, if needed. Set the Viscosity Level. Refer to the Specialist User Manual for more details. Click “Save” to add the new reagent to the system.
ONCORE Pro X, Operating Manual, Rev. L
76
To edit or remove an existing antibody or reagent, select the Edit tab. Select
the reagent from the list under the appropriate Reagent Type and edit the
available fields, as needed. Note: Users granted Supervisor access may only
edit reagents Open to Customer. Reagents designated Distributor or Closed are
protected by the system and cannot be modified or removed. Contact Technical
Support for assistance.
Figure 104
Click “Delete” to delete the reagent from the system, as needed. Note: The
reagent cannot be deleted if it is used in an existing protocol on the system.
Click “Return” to exit.
ONCORE Pro X, Operating Manual, Rev. L
77
Preparing RFID Reagent Tags with the RFID Tag Editor
The system offers user-programmable vials for use with the user’s custom
reagents and third-party antibodies. Each programmable vial contains an RFID
tag which will store all the information associated with that reagent. The
RFID tag may be reprogrammed for multiple uses, if needed. Load the removable
vial holder into bottom right corner of the reagent rack and then load the
rack onto the instrument.
Close the door and click “RFID Editor” to launch the RFID Tag Editor. The
robotic arm will position the RFID antenna over the vial holder. Place a user-
programmable vial in the vial holder and remove the cap.
Figure 105
ONCORE Pro X, Operating Manual, Rev. L
78
To read the information in the RFID tag, click “Read Tag”. The programmed
information will be displayed in the empty fields to the left. Note: Brand-new
RFID tags will not be read correctly, as they do not contain any data, only
the tag manufacturer’s randomly generated Tag ID.
To write information into the RFID tag, complete the entry fields to the right
and click “Write Tag”.
Select the Reagent Type and Reagent Name from the drop down lists. Note: Users
with Supervisor access may only select user-added reagents and reagents Open
To Customer. Some reagents are restricted by the manufacturer and may not
appear on the user-programmable reagent list.
Select the appropriate Bottle Type (15 or 7 mL). The bottle type is required
to ensure accurate liquid detection. 15 ml vials should be programmed for 60
tests, while 7 ml vials should be programmed for 30 tests. All vials must be
filled properly to the line indicated.
Enter the Storage Temperature.
Enter the Number of Tests Left.
Note1: The system may permit the user to reuse/reprogram user-programmed
vials. Please ensure vials are thoroughly cleaned and dried before re-use.
Note2: The system may block access for reprogramming reagent kit vials.
Enter the Lot Number (up to 12 characters) and Expiration Code (yyyymm).
(Optional) Select the Manufacturer Name. Note: Select “—Add or Remove—“, from
the dropdown menu to add a new reagent manufacturer name.
(Optional) Enter the Catalog Number.
The system will immediately read the tag after writing to verify and display
the data in the left field. Do not remove the vial until the tag ID on the
left is highlighted in green to indicate that the tag has been read
successfully.
Note: The system will give a warning if the user chooses to re-use a
programmed vial.
ONCORE Pro X, Operating Manual, Rev. L
79
Assigning Negative Controls
All protocols (except Special) are automatically generated with a
corresponding negative control protocol. All steps of the protocols are
identical except the primary reagent which is substituted with the negative
control reagent. The negative control is set to Universal Negative by default.
Click “Negative Control Editor” to manage and assign negative controls for
each protocol. Note: Add new negative control reagents in the Reagent Editor,
under reagent type Negative. Refer to Reagent Editor, p. 78 for more
information on adding new reagents.
To view all protocols assigned to a negative control, select the negative
control from the list and click “Show Links to Protocols”. All protocols
linked to the negative control will be highlighted under Protocols. Click
“Clear Selected Protocols” to clear the protocol selection(s) displayed on the
screen.
To view the negative control assigned to a protocol, select the protocol from
the list and click “Show Links to Negative”. The negative control linked to
the protocol will be highlighted under Negative Control. Click “Clear Selected
Negative” to clear the negative control selection displayed on the screen.
Figure 106
To assign a negative control to a protocol, select the protocol(s) listed
under Protocols and select a negative control reagent listed under Negative
Control. Tip: Hold down the Ctrl key to select multiple protocols.
To change the negative control assignment of the protocol, select the
protocol(s) listed under Protocols and select a different negative control
reagent. Note: The user cannot remove a negative control assignment, only
change it. A negative control must always be selected for every protocol.
Click “Save” to link the selected protocol(s) to the negative control. Click
“Reset” to reset all assignments back to the default negative control,
Universal Negative.
ONCORE Pro X, Operating Manual, Rev. L
80
Creating Custom Groups
Click “Custom Group Editor” to edit/create custom lists of protocols for quick
access to protocols in the Barcode Label Editor. Refer to Add New Labels in
the Barcode Label Editor, p. 19, for more information.
Select a custom group from the dropdown list Custom Group Name. Protocols in
the custom group will be listed under Selected Protocols.
To add a custom group, click “Add New Custom Group”, enter a New Custom Group
Name, and click “Add”.
To add protocol(s) to a custom group, select protocol(s) from the list under
Available Protocols, click “Add Protocol”, and click “Save”.
To remove protocol(s) from a custom group, select the protocol(s) from the
list under Selected Protocols, click “Remove Protocol”, and click “Save”.
Figure 107
Figure 108
To delete a custom group, select the group from the dropdown list Custom Group Name and click “Delete This Custom Group”.
Click “Return” to exit the Custom Group Editor.
ONCORE Pro X, Operating Manual, Rev. L
81
Creating Custom Panels
Click “Panel Editor” to edit/create custom panels of protocols accessible in
the Barcode Label Editor. Refer to Add New Labels in the Barcode Label Editor,
p. 19, for more information.
Select a panel from the dropdown list Panel Name. Protocols in the panel will
be listed under Selected Protocols.
To add a new panel, click “Add New Panel”, enter a New Panel Name, and click
“Add”.
To add protocol(s) to a panel, select the protocol(s) from the list under
Available Protocols, click “Add Protocol”, and click “Save”. Note: The user
may choose to add a positive and/or negative control with each protocol if
desired.
To remove protocol(s) from a panel, select the protocol(s) from the list under
Selected Protocols, click “Remove Protocol”, and click “Save”.
Figure 109
Figure 110
To delete a panel, select the panel from the dropdown list Panel Name and click “Delete This Panel”.
Click “Return” to exit the Panel Editor.
ONCORE Pro X, Operating Manual, Rev. L
82
5.3. Security
Manage system security features under the “Security” tab.
Figure 111
Change Password
Click “Change Password” to change the current user login password.
Figure 112
User Administration
Click “User Administration” and login to view the list of users. To add a new
user, click “Add User” to open the “Add User” window. Enter the Name, User ID
and Password and set the Access Level. Note: Only users with Distributor or
Supervisor access level may create additional user accounts.
Technician (customer) level will grant access limited to the essential
features needed for routine operation of the instrument.
Supervisor (customer) level will grant access to more advanced features needed
to configure and customize the instrument.
Distributor level (manufacturer) will grant full access to Tools, Editors,
Security and Settings.
Figure 113
To remove a user, select the user in the grid and click “Delete User.”
ONCORE Pro X, Operating Manual, Rev. L
83
5.4. Settings
View system information and adjust system settings under the “Settings” tab.
Figure 114
Baking Slide Settings
Click “Baking Slides Settings” to set the Temperature (°C) and Baking Time
(Minute). Click “Save” and “Return”.
Figure 115
ONCORE Pro X, Operating Manual, Rev. L
84
2D Barcode Format
Click “2D Barcode Format” to open the “Barcode Label Options” and edit
settings for the Barcode Label Editor.
Select the option for the Date Format.
Select the option Copy Patient Information to New Record to retain the patient
information from the previous set of labels for the next set of labels in the
Barcode Label Editor.
(For LIS systems only) Select the option Show Slide UID to display the unique
slide ID received from the LIS system in the Slide Label Editor. This allows
the user to identify slides ordered by the LIS system.
Select the option Print Barcode Using Slide UID to generate the 2D barcode
label using the unique slide ID received from the LIS system instead of the
system-generated ID.
Select the option Record from LIS use LIS label format to use LIS label
formatting.
Figure 116
Configure the layout of the label. Refer to Label Format, p. 86. Manage the list of doctor names that appears in the Barcode Editor. Refer to Edit Doctor Name List, p. 87.
Enter default Label field names for Text Line 1 and Text Line 2 of the label. Enter a default Entry for Text Line 1 and Text Line 2, as needed.
Select the Printer name from the dropdown list. Ensure that a compatible label printer is connected and selected in the Printer dropdown list.
Adjust the Top and Left Margins, as needed. Adjust the print Width, Dark Level and print Speed, as needed.
Click “OK” to close the window.
The default label print-out is displayed below for Label Style #1.
Figure 117
ProtocolName PatientID CaseNum DateTimeEntry TextLine1 TextLine 2
2D Barcode m/n
ONCORE Pro X, Operating Manual, Rev. L
85
Label Format
Click “Label Format” in the Barcode Label Options to open the Label Formatter
and edit the printable data on the label. Select the type of label to be
formatted from the Label Format dropdown list:
Record Entered Locally configures the labels input directly on the system.
Record Received from LIS configures the labels generated by the LIS system
Figure 118
Figure 119
Select the Label Style to change the layout of the label.
Select the data field identifier for each line of the label. The user may change the order and type of data to be printed on each line of the label. Note: Some data types are only available for use with an LIS system.
Click “Save” to save changes to the label format. Note: Please verify that the printed data will not overlap with the 2D barcode label.
ONCORE Pro X, Operating Manual, Rev. L
86
Edit Doctor Name List
Click “Edit Doctor Name List” in the Barcode Label options to display the list
and add/remove entries. To add a new doctor, enter a new name and click “Add”.
Figure 120
To remove an existing doctor, select a name from the dropdown list and click
“Remove”.
Check Heater Settings
The manufacturer configures the temperature, tolerance, and time for the
heater check function.
Save All Customizations
Click “Save All Customizations” to back up all protocol customizations in the
Protocol Editor. The previously saved customizations will be overwritten.
Therefore, the user should verify all
current protocol customizations in the Protocol Editor before proceeding.
Figure 121
Tip: The user may restore an individual protocol or a selected group of
protocols using the saved protocol customizations by clicking “Refresh” from
the appropriate tab in the Protocol Editor. Alternately, the user may restore
all protocol customizations by clicking “Refresh All Protocols” from the
Template tab in the Protocol Editor
Device Version
Click “Get Device Version” to view the system device information.
ONCORE Pro X, Operating Manual, Rev. L
87
Baking Slides Option
Check the box Baking Slides Before Staining to set the system to automatically
begin baking slides before the start of every staining run. Set the time
allotted for cool down after baking. Click “Save”.
Figure 122
Detect Waste Overflow Option (Distributor Access only)
Check the box “Detect Waste Overflow” to enable waste overflow detection of
the wash stations. Click “Save”. When the system detects overflow of a waste
station, it will warn the user with a pop-up message and an alarm. If the non-
hazardous waste station has overflowed, the system will divert waste to the
hazardous waste station.
Run Without RFID Reader Option (Distributor Access only)
Check the box “Run Without RFID Reader” to disable the RFID Reader. Click
“Save”. This option should only be used in the event of an RFID reader
failure.
Reagent Consumption Priority
Set the usage priority when multiple vials of the same reagent are scanned
during Reagent Check. Click “Save”. The option “Expiration Date” will
prioritize the consumption of reagents with an earlier expiration date. The
option “Less Test Numbers” will prioritize the consumption of reagents
containing fewer tests.
Figure 123
Software Version
Contact Technical Support for release notes and information about the latest
software updates.
Allowed Reagent Assignment (Distributor Access only)
Set the allowed number of manually assigned reagent tests. The number must be
reset if the user exceeds the limit. Please contact Technical Support for
assistance.
Interlock Options (Distributor Access only)
Set the safety interlock configuration. The interlock may be enabled/disabled
for the listed processes. Please contact Technical Support for assistance.
Language
Select the system operating language from the dropdown list. Click “Save”. The
system will prompt the user to restart the application to apply the changes.
ONCORE Pro X, Operating Manual, Rev. L
88
Tubing Clean Tracking Option (Distributor Access only)
Check the box Enable Tubing Clean Tracking to track the number of slides run
since the last tubing clean. Enter the Maximum Slides Run. The count is
incremented after each staining run. When the system detects that the counter
has exceeded the set threshold, the system will block the user from starting a
new staining process until the tubing cleaning has been performed.
Figure 124
Module Tracking Option (Distributor Access only)
Check the box Enable Module Tracking to track the number of slides run on each
module since the last reset. Enter the Maximum Slides Run. The counter for
each chamber is displayed on the upper right corner of its grid on the slide
map. It is incremented after each staining run. When the counter has exceeded
the threshold, it will be highlighted in red to alert the user to replace the
chamber(s) and reset the counter.
Figure 125
ONCORE Pro X, Operating Manual, Rev. L
89
5.5. Network
Connect several Slide Stainer systems over the local netw
References
- IHC Primary Antibody Supplier & Manufacturer | Biocare Medical
- IHC Primary Antibody Supplier & Manufacturer | Biocare Medical
Read User Manual Online (PDF format)