- SECTION 1 GENERAL INFORMATION
- Introduction
- About This Manual
- About the Readout
- Readout Specifications
- SECTION 2 INSTALLATION
- Mounting the Readout
- Electrical Connections
- Reading Direction
- Linear Scaling
- Maintenance
- SECTION 3 OPERATION
- Keypad Functions
- + (PLUS)
- 0 (ZERO)
- – (MINUS)
- ABS / INC
- RS-485 Bus Commands
- System Programming
- Programming Parameters
- What does “no Enc” mean?
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
LED Readout
LED Readout
Installation & Operation
Installation & Operation
WARRANTY
Accurate Technology, Inc. warrants the ProScale Systems against defective parts and workmanship for 1 year commencing from the date of original purchase. Upon notification of a defect, Accurate Technology, Inc., shall have the option to repair or replace any defective part. Such services shall be the customer’s sole and exclusive remedy. Expenses incidental to repair, maintenance, or replacement under warranty, including those for labor and material, shall be borne by Accurate Technology, Inc. (Including freight or transportation charges during the first 30 days). Except as expressly provided in this warranty, Accurate Technology, Inc. does not make any warranties with respect to the product, either expressed or implied, including implied warranties of merchantability or fitness for a particular purpose, except as expressly provided in this agreement. Accurate Technology, Inc. shall not be liable for any special, incidental, or consequential damages or for loss, damage or expense directly or indirectly arising from the customer’s use of or inability to use the equipment either separately or in combination with other equipment, or for personal injury or loss or destruction of other property, or from any other cause. To request repair work, (either warranty qualified parts or not) contact Accurate Technology, Inc. directly by phone, fax, or e-mail. A Returned Merchandise Authorization (RMA) number is required before returning a product for repair.
SAFETY WARNING
Before installing Readout on any machinery:
Turn off machine and LOCK-OUT POWER.
SECTION 1 GENERAL INFORMATION
Introduction
This LED Digital Readout is one of many Digital readouts available for use with ProScale® Measuring Systems.
ProScale digital measuring systems are affordable precision electronic devices for making linear measurements with speed and accuracy. ProScale consists of a SCALE, an ENCODER (also called READHEAD) and a DIGITAL READOUT (DRO).
This readout supports ProScale Models 150G, 250G, 190, 290, 390 and 590.
About This Manual
This manual includes Operating and Programming information for:
1/8 DIN Red LED In-Panel Mounted Digital Readout, Part Number 700-1600-091
with Firmware version L3.000 or higher
(Firmware version is displayed on power-up, ie L3.000)
(This manual does NOT include installation information for any other ProScale product.)
1/8 DIN LED Readout, 700-1600-091
About the Readout
-
Two independent absolute measurement modes (M1 and M2) with programmable incremental mode (M3). Incremental mode can be programmed as:
· Incremental mode with offset adjustment.
· Third independent absolute mode. -
Programmable offsets for both M1 and M2 absolute modes and M3 incremental mode.
-
Programmable to display millimeters or decimal inches.
-
Offset lockout in both incremental and absolute modes.
-
Offset programming disable via system programming.
-
System configuration using keyboard programming.
-
Programming lockout capability using jumper selection.
-
Program functions stored in non-volatile memory (EEPROM).
-
RS-485 communication bus allows up to 32 individual readouts to be connected on single twisted pair cable.
-
Programmable scaling factor (.00001 to 9.99999).
-
Programmable scale direction for sign correction.
-
Programmable upper and lower limits with relay output.
-
Multi-stage transient voltage protection.
Readout Specifications
Display Range: + 999.999 in; or + 9999.99 mm
Resolution: .1inch .1mm
.01inch .01mm
.001inch .01mm
.0005inch .01mm
Repeatability: .01mm; .001in
Size: Fits 1/8 DIN cutout (50 x 100mm)
Operating Power: 12-24VDC @ 50mA max
Operating Temperature: 32 to 120 oF, 0 to 50 oC
Output: RS485
Manual Part # 800-1095-010
SECTION 2 INSTALLATION
Mounting the Readout
A cutout should be made in the panel that is at least 45mm x 90mm (1.75 x 3.55 inches).
Cut-out opening size: 45mm x 90mm
The back half of the plastic housing sandwiches the sheet/panel using the supplied 4-40 machine screws.
NOTES:
1. Only use the longer screws if your sheet/panel is great than 12mm (.25
inches) thickness.
2. Do not overtighten rear panel screws or damage to the keypad can occur.
3. Use of thread lock on these screws is recommended.
Electrical Connections
This illustration shows the electrical connections to the ProScale LED readout. Power supply voltage is connected at pins 1 (plus) and 2 (ground). RS-485 bus is connected to pins 3 (Data A) and 4 (Data B). Relay output (for limits) is connected to pins 5 and 6.
Reading Direction
Once the system has been put into operation, if the counting direction of readings is opposite the desired direction, the counting direction can be reversed. See Programming Parameter (Pr5).
Linear Scaling
The readout has the ability to “scale” the actual measurement. This function is useful when the actual measurement must be multiplied or divided before being displayed. NOTE: Care should be taken when using this function since invoking it will cause the unit to display a reading different than the actual measured value. This function has a range of: 0.0001 to 99.9999. See Programming Parameter (Pr8).
Maintenance
ProScale is designed to operate in a dry environment with non-conductive debris such as plastic or sawdust. The system should be cleaned of excess debris when necessary. Do not use any liquid lubricants on the scale assembly, as this may impede the encoder’s ability to operate properly and could attract other contaminants to the scale.
Manual Part # 800-1095-010
Page 7 of 16
SECTION 3 OPERATION
Keypad Functions
MODE
Activates the following functions depending on operating state:
- Selects absolute M1 or M2 modes or incremental M3 mode of operation during normal use.
- Press and hold for offset lock.
- Advances to next programming option while in programming mode.
- Acts as a “SHIFT” key to execute special operations:
o Mode and 0 = Programming Mode entry and exit.
o Mode and + = Add segment in multi-segment mode.
o Mode and = Subtract segment in multi-segment mode.
+ (PLUS)
- Adds to position offset value in normal use.
- Increments value of programming option while in program mode.
0 (ZERO)
· Sets the current displayed position to the DATUM value programmed. M1 = Pr 1, M2 = Pr 2, M3 = Pr 3 (when used as an ABS coordinate system). The factory default value = 0.0000.
– (MINUS)
· Subtracts from position-offset value in normal use. · Decrements value of programming option while in program mode.
ABS / INC
The independent absolute measurement modes M1 and M2 are used to display a position from a fixed reference point on a machine or coordinate system. The PLUS, ZERO and MINUS keys can be used to apply a position offset while the offset LOCK feature is not enabled.
The INC coordinate system can be operated in one of two modes: – Automatic
offset adjustment – Third coordinate system similar to ABS M1 and M2
In automatic offset adjustment, an offset is programmed using the Plus, Zero
and Minus keys (assuming LOCK is not enabled). Each single depression of the
MODE key causes the programmed offset to be re-displayed as the current
position on the scale. Pressing the MODE key twice within 2 seconds will
return the display back to the ABS M1 mode of operation.
When programmed as a third independent absolute coordinate system, the incremental mode M3 operates exactly like ABS M1 and M2 modes. This can be useful in applications where independent coordinate systems are needed.
Offset Locking
After approximately 2 seconds, the LED will flash if offsets are unlocked (or
glow steady if the offsets are locked). If offset change has been disabled in
the programming mode (Pr 6), ABS mode and INC mode will always be in the
locked state (pressing the MODE key will have no effect).
Segment Adjustment
If multi-segment mode has been enabled (ProScale Models 150 & 250 with
measuring ranges longer than 17 inches), pressing and holding the MODE key
along with either the PLUS or MINUS key will adjust the segment offset by
+/one segment length for each key depression. (This feature is disabled if
LOCK mode is on.)
Displayed Measurement Units
This function is selected via programming option Pr 4 and can be set to
millimeters or decimal inches. See the System Programming section for
additional details.
Programmable Output
The ProScale LED readout provides a low current output relay that will be
activated whenever the value of absolute coordinate system M1 exceeds the user
defined limits. An independent upper and lower limit exists for this
functionality and can be accessed via the programming mode. This output is
limited to 0.5A @ 125 VAC/ 1A @ 12VDC. See Programming Parameters Pr 11, Pr
12, Pr 13, Pr 14, Pr 15
Manual Part # 800-1095-010
Page 9 of 16
RS-485 Bus Commands
The ProScale LED readout can be queried for its current position and status data via the RS-485 two-wire bus. Commands are issued using ASCII data transmitted in a packet form. The position query command (from host) is shown below:
See Programming Parameters Pr 16, Pr 17, Pr 18
Byte # Definition
1 Readout Address (ASCII 1 through ASCII b)
2 Position Query Command (ASCII A)
3 Data Bytes To Follow (ASCII 0)
The LED readout will respond with the following string:
Byte # Definition
1 Echo of display address (ASCII 1 through ASCII b) 2 Number of data bytes to follow (ASCII : 10 bytes) 3 Current coordinate system (ASCII 0 = M1, 1 = M2, 2 = M3) 4 Units mode (ASCII 0 = mm, 1 = inches)
5 – 12 Position data in ASCII. Position data based on the currently selected resolution
13 Carriage return character, value = 13 decimal or 0x0D in hex.
The readout expects individual characters to be received no more than 100 milliseconds apart. A minimum inter-character time should be no less than 3 milliseconds. If the inter-character time exceeds 100 milliseconds, the data packet will be ignored.
The readout will begin sending the requested position data within 30 milliseconds of a valid query command.
The word format is 8 data bits, no parity and 1 stop bit.
1/8 DIN LED Readout, 700-1600-091 Page 10 of 16
System Programming
Several functions of this digital readout are user programmable. The following instructions describe what features are available and how to change the system’s factory defaults to customize the readout for your application.
The keys have multiple functions. How long a key is depressed (and the combination of the keys pressed) is important.
This manual uses the term “momentarily” to describe a key press of less than 1 second. Whereas “press and hold” is used imply a key press of longer than 1.5 seconds. As an example; when using a computer keyboard to type a CAPITAL letter you “press and hold” the SHIFT key and “momentarily” press the appropriate letter key.
The “function” associated with the readout key pressed is sometimes executed on the key RELEASE, not the key DEPRESS. This is important since some keys execute different functions based on how long they are held down. These key operations, once tried, quickly become intuitive.
Programming can be Disabled by moving the jumper located on the circuit board so pins are NOT connected. Programming is Enabled when these pins are connected. All program parameters are maintained in non-volatile memory.
To enter programming mode, press and hold the MODE key and then momentarily press and release the 0 (zero) key.
Once in programming mode, The LED will briefly display:
PG on (Programming On), then Pr 1, (indicating Programming Parameter #1) for about 1 second, then it will display the current value of the parameter.
To change the current value of the parameter, press the PLUS or MINUS key to increase or decrease the displayed value. (Each parameter has a limited range that cannot be exceeded by additional key presses of the PLUS or MINUS keys.)
Press the ZERO key to revert to the factory default setting for that parameter.
To move to the next programming parameter, momentarily press the MODE key. Parameters are numbered 0 through 21. When the last parameter has been reached, pressing the MODE key again will restart the parameter list at 0.
To exit programming mode, press and hold the MODE key and then momentarily press and release the 0 (zero) key.
NOTE: The readout will automatically exit programming mode after 60 seconds of no key activity.
Manual Part # 800-1095-010
Page 11 of 16
Programming Parameters
The Digital Readout Programming Parameters are listed below. Values in [ ] are the range of values available for that Parameter. Factory defaults values are shown in Bold Red.
Pr1 – Datum preset for M1 [0.00 ] Sets the value measurement mode M1 will be forced to when the ZERO (0) key is pressed.
Pr2 – Datum preset for M2 [0.00 ] Sets the value measurement mode M2 will be forced to when the ZERO (0) key is pressed.
Pr3 – Datum preset for M3 [0.00 ] Sets the value measurement mode M3 will be forced to when the ZERO (0) key is pressed.
Pr4 – Measurement Units [0 ,1] 0 = Millimeters
1 = Inches
Pr5 – Encoder Direction [0 ,1] Changes the sign operation on the readout to alter which direction measurements are increasing or decreasing.
Pr 6 – Key Lockout [0 or 1] This parameter controls the operation of the +, – and datum keys. If enabled, (set to 1), these keys will not function (M1, M2, M3 LEDs remain ON). This prevents accidental changes when depressing these keys during normal operation.
Pr 7 – Display Resolution [1, 2, 3 or 4] This parameter sets the number of places to the right of the decimal point on the display. The readout will auto-range to the next resolution if the value is too large to be displayed in the current resolution but is displayable in an alternate resolution.
A value of 1 will display x.x.
A value of 2 will display x.xx
A value of 3 will display x.xxx (inches only)
A value of 4 will display x.xxxx (inches only)
1/8 DIN LED Readout, 700-1600-091 Page 12 of 16
Pr 8 – Scaling [0.00001 to 9.99999, 1.00000 ] This parameter sets a linear multiplier correction factor in the readout that is applied to the actual measurement prior to any offsets and then the adjusted measurement is displayed on the LED. This is useful, for instance, if you would like to display half, double, or any other value that can be achieved by multiplying the amount of actual movement by the value of the parameter setting.
Pr 9 – ProScale Compatibility [0 or 1 ] This enables or disables compatibility with long ProScale ABS systems. Disable [set to 0] for all ProScale 150-10, 180, 280, 380 and 580 systems.
Enabled [set to 1] for ProScale Model ABS systems longer than 16 inches (Model 150-18 and all Model 250 systems).
Pr 10 – M3 Operation. [0 or 1 ] Configures the operation of the third
coordinate system (M3). 0 = standard ABS (Absolute)
1 = Incremental mode
Pr 11 – Use Limits. [0 or 1] Enables or disables the use of soft limits. If enabled, a message on the LED is displayed when the measuring system is operating outside the programmed (Pr 12 & Pr 13) position limits.
0 = off
1 = on
Pr 12 –Lower Limit . [+ 999.999in , + 9999.99mm] References the M1 absolute coordinate system and activates the output if the value of M1 decreases below this value.
Default = 0.000
Pr 13 –Upper Limit [+ 999.999in , + 9999.99mm] References the M1 absolute coordinate system and activates the output if the value of M1 increases above this value.
Default = 5.000 inches, (127mm)
Pr14 – State of the Output signal for limits [0 or 1] When M1 position is within the limits set in Pr12 and Pr13:
0 = The output terminals are closed
1 = The output terminals are open
Manual Part # 800-1095-010
Page 13 of 16
Pr 15 –Output Signal Operation for limits [0 or 1] Selects the signal output operation.
0 – No output.
1 = Monitor limits
Pr 16 – RS-485 Bus address [0 to 32] This value identifies the address of the readout when used on a multi point bus.
Pr 17 – Bus Mode [0 or 1] 0 = poll for data.
1 = data is streamed at the encoder sampling rate.
Pr 18 – RS-485 Bus baud rate [0, 1, 2 , 3, 4] 0 = 1200 baud
1 = 2400 baud
2 = 9600 baud
3 = 19200 baud
4 = 115200 baud
Pr 19 – LED brightness [0 to 15, 10 ] Larger than 10 results in brighter LEDs
Smaller than 10 results in dimmer LEDs
Pr 20 – 4th decimal place rounding in INCHES [0 or 1 ] 0 = 4th decimal place is shown as value 0-9
1 = 4th decimal place is shown as value 0 or 5 only
Pr 21 – Reset to factory settings [0 or 1 ] If this value is changed to 1, then programming menu is exited, the readout will reboot and revert all programming values to factory settings.
1/8 DIN LED Readout, 700-1600-091 Page 14 of 16
SECTION 4 MISCELLANEOUS Frequently Asked Questions
What does “no Enc” mean?
If the encoder is off the Scale, or the encoder cable is unplugged from the readout, an “no Enc” will appear on the Display. To clear error: 1. Be sure the encoder is on the scale and properly oriented. 2. Unplug the connector from the readout for 5 seconds. 3. Reconnect the encoder cable to the readout.
Can I mount the scale/encoder without the connector link/guide clip? The connector link and guide clip serve to provide an accurate method of transferring the movement of the moving part to the encoder or scale, while also absorbing any stresses that may occur. If they are not used, the warranty may be voided.
The readout displays numbers, but they seem to be random. Be sure the encoder is oriented correctly on the scale. If your scale is Model 150G or 250G, one end of the scale is black. Be sure that the arrow on the encoder is pointed towards the Black End of the Scale.
Manual Part # 800-1095-010
Page 15 of 16
Thank you for choosing a ProScale Product,
IT WAS PROUDLY MADE IN THE USA
Accurate Technology, Inc.
270 Rutledge Rd. Unit E
Fletcher, NC 28732 USA
www.proscale.com
P/N 800-1095-010, Rev 22-09-14, Copyright © 2022,
Accurate Technology, Inc. All rights reserved.
1/8 DIN LED Readout, 700-1600-091 Page 16 of 16
References
- Digital Measuring Devices | Accurate Technology Inc.
- ProScale 150G Absolute Linear Encoder (Grounded)
- ProScale 190 Incremental Linear Encoder
- ProScale 250G Absolute Linear Encoder (Grounded)
- ProScale 290 Incremental Linear Encoder
- ProScale 390 Incremental Linear Encoder
- ProScale 590 Incremental Linear Encoder with Bearing System