ABEYE ABLT02 LEXILENS App User Guide

June 16, 2024
ABEYE

ABEYE ABLT02 LEXILENS App - logo 2User Guide

ABLT02 LEXILENS App

Getting started
Detailed instructions available at https://www.lexilens.com

  1. Install the Lexilens® app
    to configure your glassesABEYE ABLT02 LEXILENS App - Lexilens®
app Please read and keep all safety, security and use instructions.

  2. Press button
    to activate Lexilens®.ABEYE ABLT02 LEXILENS App -
buttonLED green: battery level above 10%
    LED flashing red: battery low

  3. Hold down button
    until LED blue light appears to activate Bluetooth®.
    Follow the app instructions to connect Lexilens®.ABEYE ABLT02 LEXILENS App -
Hold down buttonFlashing blue LED: Bluetooth® activated
    Fixed blue LED: Lexilens® connected to app

  4. Adjust Lexilens® speed to make it easier for you to read.
    Try the setting 0, then -5 and finally 5.If it is still not easy to read, try -10 and +10. Lexilens® saves your settings: you can now read more easily with Lexilens® without having to activate Bluetooth® or connect to the app.

  5. You’re all set!
    You can now deactivate Bluetooth. Your settings have been saved in your Lexilens® glasses. Next time, you will not need your telephone.

Optical setting

OPTION

  1. Remove nose pad*
    (Only if your model has a removeable nose pad, otherwise go straight to step
  • Free the pad with a paper clip, then remove the nose pad. Do the same when you clean the lenses once the optical pad is in place.ABEYE ABLT02 LEXILENS App - Remove nose pad 2. Install the optical clip
    to adapt Lexilens® to your eyesight, if necessary.![ABEYE ABLT02 LEXILENS App

Important Safety Instructions

Indications for Use
The ­Lexilens® eyewear filter light using electronic lenses to improve the reading quality of people with reading disabilities by detaching and disentangling letters, words, lines.
When the ­Lexilens® eyewear is correctly adjust- ed for the user, faster reading, better under-standing of texts and less concentration effort are obtained in most cases. ­Lexilens® active lenses do NOT provide sunlight protection. Lexilens® optical clip accessory can optionally be mounted on ­Lexilens® eyeframe to provide eye sight correction with prescription lenses.

Contraindications and adverse effects
Do not use if you are prone to epilepsy:
Lexilens ® glasses can cause epileptic seizures for photosensitive people. ­Lexilens® can cause headaches and nausea. Do not use the ­Lexilens® eyewear without prior consultation if the user has a known ocular pathology (strabismus, amblyopia, etc.). It is recommended to consult an ophthalmologist if in doubt.

  1. Read and keep these instructions.
  2. Heed all warnings and follow all instructions
  3. Do not use this apparatus near water.
  4. Clean only with a dry cloth.
  5. Only use attachments/accessories specified by the manufacturer
  6. Unplug this apparatus during lightning storms or when unused for long periods of time.
  7. Do not use this apparatus to protect from the sunlight.
  8. Do not use as eye protection against mechanical impact hazards.
  9. Refer all servicing to qualified personnel. Servicing is required when the apparatus has been damaged in any way, such as power supply cord or plug is damaged, liquid has been spilled or objects have fallen into the apparatus, the apparatus has been exposed to rain or moisture, does not operate normally, or has been dropped.
  10. Do not attempt to service the apparatus yourself, do not open the device.
  11. Do not wear the ­Lexilens® eyewear while charging.

WARNINGS/CAUTIONS

Do not use the ­Lexilens® eyewear for an extended period of time. It is recommended to have a break every hour to avoid headaches.
Do not use the ­Lexilens® eyewear under neon light to avoid flickering of the lenses. Visible flickering of the ­Lexilens® lenses could cause headaches, dizziness or epileptic seizures.
In the event of visible flickering of the lenses of the ­Lexilens® frame under a flicker-free light source (incandescent lamp, LED), contact your retailer or ABEYE support.
Contains small parts which may be a choking hazard. Not suitable for children under age 5.
Use only the USB charging cord supplied with Lexilens®.
This product contains magnetic material. Consult your physician on whether this might affect your implantable medical device. To reduce the risk of fire or electrical shock, do NOT expose this product to rain, liquids, or moisture.
Do NOT expose this product to dripping or splashing, and do not place objects filled with liquids, such as vases, on or near the product.
Keep the product away from fire and heat sources. Do NOT expose the product to naked flame sources, such as lighted candles, on or near the product.

Do NOT make unauthorized alterations to this product.
Use this product only with an agency-approved power supply that meets local regulatory requirements (e.g. UL, CSA, VDE, CCC).
Do not expose products containing batteries to excessive heat (e.g. from storage in direct sunlight, near fire or simliar heat sources). Do NOT wear the ­Lexilens® eyewear while charging. Do NOT wear the ­Lexilens® eyewear while driving, walking or practicing sports. The ­Lexilens® smartphone app is available for iOS 11 and later, and for Android 5.1 and later.

Regulatory Information

NOTE : This equipment has been tested and complies with class I medical device standards.
The ­Lexilens® eyewear complies with the requirements of the IEC 60601 standard relative to the security of electrical medical devices. The Lexilens® eyewear complies with FCC and Industry Canada radiation exposure limits set forth for general population.


ABEYE SAS hereby declares that this product is in compliance with the essential requirements and other relevant provisions of Directive (EU) 2017/745 and all other applicable EU directive requirements. The complete declaration of conformity can be found at:
https://www.lexilens.com/compliance


This symbol means the product must not be discarded as household waste, and should be delivered to an appropriate collection facility for recycling.
DO NOT attempt to remove the rechargeable lithium-ion battery from this product. Contact your local retailer or other qualified professional for removal.

Manufacturer: ABEYE, 27 rue Buffon, 21200 Beaune, France
Date of Manufacture:
See label: YYYY-MM-DD

Legal Information
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc.
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth® SIG, Inc. and any use of such marks by ABEYE is under license. Google Play and the Google Play logo are trademarks of Google LLC.

©2021 ABEYE.
Lexilens® market launch date : 10/2020
Last document revision : 06/2021

Guide to symbols

**** CE marking Manufacturer
**** FCC marking **** Manufacture date
IC marking Keep in a dry place
Commercial reference Temperature limit
Series number (label).
Please note for future use: Humidity limit
Lot number Pressure limit
Please read instructions carefully Bluetooth®
**** Electronic waste Type BF applied part

WARRANTY

Your ­Lexilens® eyewear is covered by a one-year limited warranty from the date of original purchase of your ­Lexilens® eyewear (the “Warranty Period”). If you purchased your Spectacles in the EU, the Warranty Period is two years.
This Warranty covers faulty materials or workmanship in the hardware that comprises your ­Lexilens® eyewear. We warrant that the hardware will be free from manufacturing defects and, under normal and intended use, function substantially in accordance with our technical specifications or product documentation (“Warranted Functionality”) during the Warranty Period. Our Warranty does not cover damage resulting from accident, or from normal wear and tear. Our Warranty does not cover damage resulting from improper storage, misuse or abuse, neglect including scratches, cracks, or surface damage, or exposure to extreme conditions including extreme heat, dirt, or sand. “Misuse” includes without limitation use in a manner or environment not in compliance with the recommendations in this document, our online user manual, or related documentation.
Our Warranty does not cover normal depletion of consumable parts such as batteries, unless occurring due to a defect in original materials or workmanship. As an example, only batteries that exhibit fully charged capacity below 80% of the actual rated capacity, or that leak, during the warranty period will be considered defective for purposes of the warranty. Our Warranty does not cover products that have been physically or electronically altered, modified, undergone unauthorized repairs, or exhibit third-party software changes. Our Warranty does not cover software or firmware embedded in or distributed with the ­Lexilens® eyewear, and we do not warrant that such software or firmware will operate without error, that it will be compatible with any third-party hardware or software, or that content captured or distributed through your Lexilens ­ ® eyewear will not be lost or corrupted.

LIMITATION OF LIABILITY.
WE DO NOT, UNDER THIS LIMITED WARRANTY, ASSUME ANY LIABILITY FOR ANY SPECIAL, INDIRECT, INCIDENTAL, PUNITIVE, OR CONSEQUENTIAL DAMAGES OF ANY KIND WHATSOEVER, INCLUDING BUT NOT LIMITED TO LOSS OF PROFITS OR REVENUES, LOSS OF DATA, LOSS OF USE OF THE PRODUCT OR ANY ASSOCIATED EQUIPMENT, COST OF ANY REPLACEMENT GOODS OR SUBSTITUTE EQUIPMENT, OR LOSS OF USE DURING THE PERIOD THE PRODUCT IS BEING REPLACED OR REPAIRED. WITHOUT LIMITING THE GENERALITY OF THE FOREGOING, IN THE EVENT OF A PRODUCT REPLACEMENT, WE ARE NOT LIABLE FOR ANY ASSOCIATED SETTINGS LOSS OR CONSEQUENTIAL DAMAGES OF ANY KIND. FURTHER, IN NO EVENT WILL WE BE LIABLE FOR ANY ACCIDENT, INJURY, DEATH, LOSS, OR OTHER CLAIM RELATED TO OR RESULTING FROM THE USE OF THIS PRODUCT. WE SPECIFICALLY ADVISE YOU AGAINST USING THE ­Lexilens® EYEWEAR IN A MANNER INCONSISTENT WITH THIS DOCUMENT, OUR ONLINE USER MANUAL, AND OTHER ACCOMPANYING DOCUMENTATION. IN NO EVENT WILL WE BE LIABLE FOR DAMAGES IN EXCESS OF THE PRODUCT PURCHASE PRICE.

If any term herein is held to be illegal or unenforceable under applicable law, such term shall be severed from this warranty and to the extent legally permitted, the remaining terms shall not be affected.

WARRANTY REPAIR & REPLACEMENT SERVICE INSTRUCTIONS
Please follow the instructions at https://www.lexilens.com/support

Electromagnetic compatibility (60601-1-2 Ed4)

All the information provided below is derived from regulatory requirements to which manufacturers of electro-medical devices are subject, as defined in standard IEC60601-1-2 Ed4. The medical device complies with the current electromagnetic compatibility standards; however, the user must ensure that potential electromagnetic interferences do not create any additional risks, such as radiofrequency emitters or other electronic devices.
In this chapter, you will find necessary information to ensure the installation and commissioning of your medical device under the best possible conditions in terms of electromagnetic compatibility.
The different cords of the medical device should be kept apart from each other. Certain types of mobile telecommunication devices, such as cell phones, may interfere with the medical device. Therefore, the recommended separation distances in this chapter must be strictly adhered to.
The medical device should not be used near or placed on another device. If this cannot be avoided, it is necessary to verify its proper functioning under the intended conditions of use before any usage. The use of accessories other than those specified or sold by ABEYE as replacement parts may result in increased emissions or decreased immunity of the medical device.

Cable length

Cables and accessories| Maximum length| Type of test| In compliance with:
---|---|---|---
Cables / …| < 3m| RF emission| CISPR 11, Class B
Current harmonics emission| IEC 61000-3-2
Voltage fluctuation and flickers (voltage flicker)| IEC 61000-3-3
Electrostatic discharge immunity| IEC 61000-4-2
Radiated immunity – Electromagnetic fields| IEC 61000-4-3
Immunity to Fast Transient Bursts| IEC 61000-4-4
Shock wave immunity| IEC 61000-4-5
Conducted Immunity – Radiofrequency disturbance| IEC 61000-4-6
Radiated immunity – Magnetic fields| IEC 61000-4-8
Immunity to Voltage Dips, Short Interruptions, and Voltage Variations| IEC 61000-4-11

Recommended Separation Distances
The medical device is intended for use in an electromagnetic environment where RF disturbances are controlled. The medical device’s user can help avoid any electromagnetic interference by maintaining a minimum distance, depending on the maximum power of the RF transmitting equipment. Portable RF communication devices (including peripherals such as antenna cables and external antennas) should not be used within 30 cm (12 inches) of any part of the Lexilens® frames, including cables specified by the manufacturer. Otherwise, the performance of these devices may be affected.

Electromagnetic emissions
The medical device is intended for use in the electromagnetic environment described in the table below. Therefore, the user must ensure that the medical device is used in the environment described below.

Emission test Compliance Electromagnetic environment / notes

Electromagnetic Radiation
Disturbance (Radiated Emissions) (CISPR 11)| Group 1| The medical device uses RF energy for its internal operation.
Disturbance Voltage at Power
Supply Terminals (Conducted Emissions) (CISPR 11)| Class B|
Harmonic current Emission
(IEC61000-3-2)| N/A|
Voltage variations, Voltage fluctuations and flickering (IEC61000-3-3)| N/A|

Magnetic and electromagnetic immunity
The medical device is intended to be used in the magnetic and electromagnetic environment described in the table below. The user must ensure compliance with the electromagnetic environment.

Immunity test| Test level according to IEC60601| Compliance level| Electromagnetic environment / notes
---|---|---|---
Electrostatic discharge (ESD)
(IEC61000-4-2)| ± 8 kV contact
± 15 kV air discharge| ± 8 kV contact
± 15 kV air discharge| Home healthcare environment
Fast Transient Bursts (IEC61000-4-4)| ± 2 kV for power supply lines| N/A| N/A
Shock waves (IEC61000-4-5)| ± 1 kV in differential mode
± 2 kV in common mode| N/A| N/A
Industrial Frequency Magnetic
Field(IEC61000-4-8)| 30 A/m| 30 A/m| Home healthcare environment
Voltage Dips, Short Interruptions, and Voltage Variations
(IEC61000-4-11)| 0% UT
(Uninterrupted Time)
for 0.5 cycles at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0% UT for 1 cycle
70% UT for 25 cycles at 50 Hz for 30 cycles at 60 Hz Single phase: at 0°| N/A| N/A
Voltage interruptions
(IEC61000-4-11)| 0% UT; for 250 cycles at 50 Hz for 300 cycles at 60 Hz| N/A| N/A

Electromagnetic and Radiofrequency (RF) Immunity
The medical device is intended to be used in the magnetic and electromagnetic environment described in the table below. The user must ensure compliance with the electromagnetic environment.

Immunity test| Test level| Compliance level| Electromagnetic environment / notes
---|---|---|---
WARNING Portable RF communications devices (including peripherals such as antenna cables and external antennas) should not be used closer than 30 cm (12 inches) of any part of the Lexilensit) frame, including cables specified by the manufacturer. Otherwise, performance of these devices may be affected.
Radiated Radiofrequency Electromagnetic Fields (IEC61000-4-3)| 10 V/m
80 MHz at 27 Ghlz
80 % modulation depth at 1 kHz| 10 V/m
80 MHz at 2.7 GHz
80 % modulation depth at 1 kHz| Home healthcare environment
Proximity Fields Emitted by RF Wireless Communication Devices (IEC 61000-4-3 provisional method)| 9 V/m
710 MHz 745 MHz, 780 MHZ,
5240 MHz, 5550 MHz, 5785 MHz 27 V/m 385 MHz 28 V/m
450 MHz 810 MHz, 870 MHz 930 MHz 1720 MHz 1845 MHz,
1970 MHz, 2450 MHz 5785 MHz| 9 V/m
710 MHz 745 MHz
780 MHZ,
5240 MHz,
5550 MHz
27 V/m
385 MHz
28 V/m
450 MHz 810 MHz,
870 MHz 930 MHz
1720 MHz,
1845 MHz,
1970 MHz,
2450 MHz| Home healthcare environment
Conducted disturbances, induced by RF fields (IEC610004-6)| 3V
150KHz to 80MHz 6 V in ISM band and frequency range
between 0.15 MHz and 80 MHz
including amateur radio bands.
80% modulation depth at 1 kHz| WA| WA

WARNING TO USERS IN THE UNITED STATES

Federal Communication Commission Interference Statement (47 CFR Section 15.105(b))
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver.
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.

This device Lexilens® complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

NO UNAUTHORIZED MODIFICATIONS (47 CFR Section 15.21)

CAUTION : This equipment may not be modified, altered, or changed in any way without signed written permission from ABEYE. Unauthorized modification may void the equipment authorization from the FCC and will void the ABEYE warranty.
FCCID : 2A9XX-ABLT02

WARNING TO USERS IN THE CANADA

This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada.
To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication.
IC: 30009-ABLT02
HVIN: ABLT02

www.lexilens.com/support
© ABEYE SAS 2021 – 27 rue Buffon – 21200 Beaune – France