FLUKE 787 ProcessMeter User Manual
- June 16, 2024
- FLUKE
Table of Contents
- LIMITED WARRANTY & LIMITATION OF LIABILITY
- Introduction
- Contacting Fluke
- Safety Information
- How to Get Started
- Getting Acquainted with the Meter
- Measuring Electrical Parameters
- Using the Current Output Functions
- Power-Up Options
- Battery Life
- Using the Holster and Flex-Stand
- Maintenance
- General Maintenance
- Replacement Parts and Accessories
- Specifications
- Documents / Resources
Advanced Test Equipment Corp.
www.atecorp.com
800-404-ATEC (2832)
787 ProcessMeter
Users Manual
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for
three years from the date of purchase. This warranty does not cover disposable
batteries or damage from accident, neglect, misuse or abnormal conditions of
operation or handling. Resellers are not authorized to extend any other
warranty on Fluke’s behalf. To obtain service during the warranty period, send
your defective meter to the nearest Fluke Authorized Service Center with a
description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A
PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM
ANY CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or limitation of an
implied warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
Fluke Corporation P.O.
Box 9090
Everett, WA
98206-9090 USA| Fluke Europe B.V.
P.O. Box 1186
5602 B.D. Eindhoven
The Netherlands
---|---
Introduction
Warning
Read “Safety Information” before you use the meter. Your Fluke 787
ProcessMeter™ (referred to as “the meter”) is a handheld, battery-operated
tool for measuring electrical parameters and supplying steady or ramping
current to test process instruments. It has all the features of a digital
multimeter, plus current output capability.
Your meter is shipped with a Flex-Stand™ holster, one set of TL75 test leads,
one set of AC70A Alligator Clips, this manual, and a laminated Quick Reference
Card that fits inside the holster.
If the meter is damaged or something is missing, contact the place of purchase
immediately.
Contact your Fluke distributor for information about DMM accessories. To order
replacement parts or spares, see Table 13 near the end of this manual.
Contacting Fluke
To order accessories, receive operating assistance, or get the location of the
nearest Fluke distributor or Service
Center, call:
USA : 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Address correspondence to:
Fluke Corporation
P.O. Box 9090,
Everett, WA 98206-9090
USA| Fluke Europe B.V.
P.O. Box 1186,
5602 BD Eindhoven
The Netherlands
---|---
Or visit us on the World Wide Web: www.fluke.com
Safety Information
The meter complies with IEC1010-1, ANSI/ISA S82.01-1994 and CAN/CSA C22.2 No.
1010.1-92 Overvoltage Category III. Use the meter only as specified in this
manual, otherwise the protection provided by the meter may be impaired.
A Warning identifies conditions and actions that pose hazard(s) to the user; a
Caution identifies conditions and actions that may damage the meter or the
equipment under test.
International symbols used on the meter and in this manual are explained in
Table 1.
Warning
To avoid possible electric shock or personal injury:
- Do not use the meter if it is damaged. Before you use the meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
- Make sure the battery door is closed and latched before you operate the meter.
- Remove test leads from the meter before you open the battery door.
- Inspect the test leads for damaged insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the meter.
- Do not use the meter if it operates abnormally. Protection may be impaired. When in doubt, have the meter serviced.
- Do not operate the meter around explosive gas, vapor, or dust.
- Use only a single 9V battery, properly installed in the meter case, to power the meter.
- When servicing the meter, use only specified replacement parts.
Caution
To avoid possible damage to meter or to equipment under test:
Disconnect the power and discharge all highvoltage capacitors before testing
resistance or continuity.
Use the proper jacks, function, and range for your measurement or sourcing
application.
To protect yourself, adhere to the following guidelines:
- Use caution when working above 30V ac rms, 42V ac pk, or 60V dc. Such voltages pose a shock hazard.
- When using the probes, keep your fingers behind the finger guards on the probes.
- Connect the common test lead before you connect the live test lead. When you disconnect test leads, disconnect the live test lead first.
Table 1. International Symbols
Symbol | Meaning | Symbol | Meaning |
---|---|---|---|
Alternating current | **** | Earth ground | |
**** | Direct current | Fuse | |
**** | Alternating or direct current | Conforms to European Union directives | |
Refer to the manual for information about this feature. | Conforms to |
relevant Canadian Standards Association directives
| Battery| | Double insulated
| Meets Underwriters’ Laboratories safety requirements| | Inspected and
licensed by TÜV Product Services
CAT III| Overvoltage (Installation) Category III, Pollution Degree 2 per
IEC1010-1 refers to the level of Impulse Withstand Voltage protection
provided. Typical locations include; Mains, wall outlets, main distribution
levels connected closer to the supply system but less than the primary supply
system (CAT IV).
How to Get Started
If you are familiar with the Fluke 80 Series DMM, read “Using the Current
Output Functions,” review the tables and figures in “Getting Acquainted with
the Meter,” and begin using your meter.
If you are unfamiliar with Fluke 80 Series DMMs, or DMMs in general, read
“Measuring Electrical Parameters” in addition to the sections referenced in
the previous paragraph.
The sections following “Using the Current Output Functions” contain
information about the power-up options, and battery and fuse replacement
instructions.
Later, use the Quick Reference Card to refresh your memory about the various
functions and features that you can use.
Getting Acquainted with the Meter
To become familiar with the features and functions of the meter, study the following figures and tables.
- Figure and Table 2 describe the input/output jacks.
- Figure and Table 3 describe the input functions you get with the first five rotary switch positions.
- Figure and Table 4 describe the output functions you get with the last two rotary switch positions.
- Figure and Table 5 describe the functions of the pushbuttons.
- Figure and Table 6 explain what all the elements of the display indicate.
Table 2. Input/Output Jacks
Item| Jack| Measurement Functions| Source Current
Function| Simulate Transmitter Function
---|---|---|---|---
①| A| Input for current to 440 mA continuous. (1A for up to 30 seconds.)
Fused with a 440 mA fuse.| Output for dc current to 24 mA.|
②| mA| Input for current to 30 mA. Fused with a 440 mA fuse.| Common for
dc current output to 24 mA.| Output for transmitter simulation to 24 mA. (Use
in series with an external loop supply.)
3| VΩ| Input for voltage to 1000V, W, continuity, and diode test.| |
④| COM| Common for all measurements.| | Common for transmitter simulation to
24 mA. (Use in series with an external loop supply.)
Table 3. Rotary Switch Positions for Measurements
No. | Position | Function(s) | Pushbutton Actions |
---|---|---|---|
① | OFF | Meter off | |
② | V ~ | Default: measure ac V | |
Frequency counter | Selects a MIN, MAX, or AVG action (see pg. 18) |
Selects a fixed range (hold 1 second for auto range)
Toggles TouchHold
Toggles relative reading (sets a relative zero point)
3| V| Measure dc V| Same as above
④| mV| Measure dc mV| Same as above
⑤| | Default: measure Ω
for continuity
BLUE test| Same as above, except diode test has only one range
⑥| | High test lead in
A: measure A dc BLUE selects ac
High test lead in
mA: measure mA dc| Same as above, except there is only one range for
each input jack position, 30 mA or 1A
No. | Position | Default Function | Pushbutton Actions |
---|---|---|---|
① | OUTPUT | _Test leads in |
_ SOURCE:
Source 0% mA
Test leads in
SIMULATE:
Sink 0% mA| % STEP : Adjusts output up or down to the next 25% step
COARSE : Adjusts output up or down 0.1 mAFINE : Adjusts output up or down
0.001 mA
②| OUTPUT mA| Test leads in
SOURCE:
Source repeating 0% -100%-0% slow ramp (M)
Test leads in
SIMULATE:
Sink repeating 0% -100%-0% slow ramp (M)| BLUE cycles through:
· Fast repeating 0% -100% – 0% ramp (V on display)
· Repeating 0% -100% – 0% ramp in 25% steps (N on display)
· Slow repeating 0% -100% – 0% ramp (M on display)
Table 5. Pushbuttons
No. | Pushbutton | Function(s) |
---|---|---|
① | Toggles the backlight | |
② | (BLUE) | Rotary switch in Lposition and test lead plugged into ****A |
jack: Toggles between ac and dc ampere measure
Rotary switch in position: Selects diode test function ( ) Rotary switch in
OUTPUT mA position: Cycles through
· Slow repeating 0% -100% – 0% ramp ( on display)
· Fast repeating 0% -100% – 0% ramp ( on display)
· Repeating 0% -100% – 0% ramp in 25% steps ( on display)
Table 5. Pushbuttons (cont.)
No. | Pushbutton | Function(s) |
---|---|---|
C | % STEP | Measuring: Selects a MIN, MAX, or AVG action (see pg. 18) |
mA Output: Adjusts mA output up to the next higher 25% step
④| COARSE| Measuring: Selects a fixed range (hold for 1 second for auto
range)
mA Output: Adjusts output up 0.1 mA
⑤| FINE| Measuring: Toggles TouchHold, or in MIN MAX recording, suspends
recording
mA Output: Adjusts output up 0.001 mA
⑥| FINE| Measuring: Toggles between frequency counter and ac voltage
measurement functions
mA Output: Adjusts output down 0.001 mA
⑦| COARSE| Measuring: Toggles relative reading (sets a relative zero point)
mA Output: Adjusts output down 0.1 mA
⑧| % STEP| Measuring: Toggles between W measure and continuity functions
mA Output: Adjusts mA output down to the next lower 25% step
Table 6. Display
No. | Element | Meaning |
---|---|---|
① | Percentage display | Shows the mA measured value or output level in %, in a |
0-20 mA or 4-20 mA scale (change scales with power-up option)
②| OUTPUT| Lights when mA output (source or simulate) is active
3| | Lights in diode test function
④| | Lights in continuity function
⑤| | Lights when relative reading is on
⑥| | Lights when the battery is low
⑦| Numerals| Show the input or output value
⑧| | Lights when TouchHold is on
⑨| MINMAXAVGR| MIN MAX recording status indicators:
MIN means the display is showing the minimum recorded value. MAX means the
display is showing the maximum recorded value.
AVG means the display is showing the average value since starting recording
(up to about 35 hours continuous recording time).
means MIN MAX recording is on.
No.| Element| Meaning
---|---|---
10| mA, DC, mV, AC, M or k W , kHz| Show the input or output
units and multipliers associated with the numerals
11| Auto 400100030| Range status indicators:
Auto means autoranging is on.
The number plus the unit and multiplier indicate the active range.
12| | One of these lights in mA ramping or step output (rotary switch position
mA ):
means slow continuous 0% – 100% – 0% ramping.
means fast continuous 0% – 100% – 0% ramping.
means ramping in 25% steps.
Measuring Electrical Parameters
The proper sequence for taking measurements is as follows:
- Plug the test leads into the appropriate jacks.
- Set the rotary knob.
- Touch the probes to the test points.
Input Impedance
For the voltage measurement functions, input impedance is 10 MΩ. See the
specifications for more information.
Ranges
A measurement range determines the highest value the meter can measure. Most
meter measurement functions have more than one range (see the Specifications).
Being in the right range is important:
- If the range is too low, the display shows OL (overload).
- If the range is too high, the meter will not be displaying its most accurate measurement .
The meter normally automatically selects the lowest range that will measure
the applied input signal (Auto showing on the display). Press if you want to
lock the range. Each time you press , the meter selects the next higher range.
If you have locked the range, the meter resumes auto ranging when you change
to another measurement function or you press and hold it for 1 second.
Measuring a Composite Signal
Because the input is dc-coupled, to meaure an ac voltage or frequency with a
dc bias, you must manually select the range specified in Table 7. For example,
to measure 100 mV ac with 20 V dc superimposed, select the 4 V range.
Table 7. Range Requirements for Measuring a Composite Signal
Range (ac) | Max. Allowable AC + DC |
---|---|
400.0 mV | 3 V |
4.000 V | 30 V |
40.00 V | 300 V |
400.0 V | 400 V |
1000 V | 1000 V |
Testing Diodes
To test a single diode:
- Insert the red test lead into the jack and black test lead into the COM jack.
- Set the rotary switch to .
- Press the BLUE pushbutton so that the symbol is on the display.
- Touch the red probe to the anode and the black probe to the cathode (side with band or bands). The meter should indicate the appropriate diode voltage drop.
- Reverse the probes. The meter should display OL, indicating a high impedance.
- The diode is good if it passes the tests in steps 4 and 5.
Displaying Minimum, Maximum, and Average
MIN MAX recording stores the lowest and highest measurements, and maintains
the average of all measurements.
Press to turn on MIN MAX recording. Readings are stored until you turn the
meter off, switch to another measurement or source function, or turn MIN MAX
off.
The beeper sounds when a new maximum or minimum is recorded. Auto power-off is
disabled and auto ranging is turned off during MIN MAX recording. Press again
to cycle through the MAX, MIN, and AVG displays. Press and hold for 1 second
to erase stored measurements and exit .
If MIN MAX recording is on continuously for over 40 hours, minimum and maximum
readings are still recorded, but the displayed average no longer changes.
In MIN MAX recording, press to suspend recording; press again to resume
recording.
Using TouchHold
Note
You must have MIN MAX recording off to use TouchHold.
Warning
To avoid possible electric shock, do not use TouchHold to determine if
dangerous voltage is present. TouchHold will not capture unstable or noisy
readings.
Activate TouchHold™ if you want the meter to freeze the display on each new
stable reading (except in the frequency counter function). Press to activate
TouchHold. This feature allows you to take measurements in situations in which
it is difficult to look at the display. The meter beeps and updates the
display with each new stable reading.
Compensating for Test Lead Resistance
Use the relative reading feature ( on the display) to set the present
measurement as a relative zero. A common use for this is to compensate for
test lead resistance when measuring Ω.
Select the Ω measure function, touch the test leads together, then press .
Until you press again, or switch to another measurement or source function,
the readings on the display will subtract the lead resistance.
Using the Current Output Functions
The meter provides steady, stepped, and ramped current output for testing 0-20
mA and 4-20 mA current loops.
You can choose source mode, in which the meter supplies the current, or
simulate mode, in which the meter regulates current in an externally-powered
current loop.
Source Mode
Source mode is selected automatically by inserting the test leads into the
SOURCE + and – jacks as shown in Figure 7. Use source mode whenever you need
to supply current into a passive circuit such as a current loop with no loop
supply. Source mode depletes the battery faster than simulate mode, so use
simulate mode whenever possible.
The display looks the same in source and simulate modes. The way to tell which
mode is in use is to see which pair of output jacks is in use.
Simulate Mode
Simulate mode is so named because the meter simulates a current loop
transmitter. Use simulate mode when an external dc voltage of 24 to 30V is in
series with the current loop under test.
Caution
Set the rotary switch to one of the mA output settings BEFORE you connect the
test leads to a current loop. Otherwise, a low impedance from the other rotary
switch positions could be presented to the loop, causing up to 50 mA to flow
in the loop.
Simulate mode is selected automatically by inserting the test leads into the
SIMULATE + and – jacks as shown in Figure 8. Simulate mode conserves battery
life, so use it instead of source mode whenever possible.
The display looks the same in source and simulate modes. The way to tell which
mode is in use is to see which pair of output jacks is in use.
Changing the Current Span
The meter’s current output span has two settings (with overrange to 24 mA):
- 4 mA = 0%, 20 mA = 100% (factory default)
- 0 mA = 0%, 20 mA = 100%
To find out which span is selected, short the OUTPUT SOURCE + and – jacks,
turn the rotary switch to OUTPUT , and observe the 0% output level.
To toggle and save the current output span in nonvolatile memory (retained
when the power is turned off):
- Turn off the meter.
- Hold down the pushbutton while you turn the rotary switch to OUTPUT .
- Wait at least 2 seconds, then release .
Producing a Steady mA Output
When the rotary switch is in the OUTPUT position, and the OUTPUT jacks are
connected to an appropriate load, the meter produces a steady mA dc output.
The meter begins sourcing or simulating 0%. Use the pushbuttons to adjust the
current as shown in Table 8.
Select either sourcing or simulating by choosing the SOURCE or SIMULATE output
jacks.
If the meter cannot deliver the programmed current because the load resistance
is too high or the loop supply voltage is too low, dashes (—–) appear on the
numeric display. When the impedance between the SOURCE jacks is low enough,
the meter will resume sourcing.
Note
The STEP pushbuttons described on the next page are available when the meter
is producing a steady mA output. The STEP pushbuttons go to the next multiple
of 25%.
Table 8. mA Output Adjust Pushbuttons
Pushbutton | Adjustment |
---|---|
Adjusts up 0.1 mA | |
Adjusts up 0.001 mA | |
Adjusts down 0.001 mA | |
Adjusts down 0.1 mA |
Manually Stepping the mA Output
When the rotary switch is in the OUTPUT position, and the OUTPUT jacks are
connected to an appropriate load, the meter produces a steady mA dc output.
The meter begins sourcing or simulating 0%. Use the pushbuttons to step the
current up and down in 25% steps as shown in Table 9. See Table 10 for mA
values at each 25% step.
Select either sourcing or simulating by choosing the SOURCE or SIMULATE output
jacks.
If the meter cannot deliver the programmed current because the load resistance
is too high or the loop supply voltage is too low, dashes (—–) appear on the
numeric display. When the impedance between the SOURCE jacks is low enough,
the meter will resume sourcing.
Note
The COARSE and FINE adjustment pushbuttons described on the previous page are
available when you are manually stepping the mA output.
Table 9. mA Stepping Pushbuttons
Pushbutton | Adjustment |
---|---|
Adjusts up to the next higher 25% step | |
Adjusts down to the next lower 25% step |
Table 10. mA Step Values
Step | Value (for each span setting) |
---|---|
4 to 20 mA | 0 to 20 mA |
0% | 4.000 mA |
25% | 8.000 mA |
50% | 12.000 mA |
75% | 16.000 mA |
100% | 20.000 mA |
125% | 24.000 mA |
120% |
Auto Ramping the mA Output
Auto ramping gives you the ability to continuously apply a varying current
stimulus from the meter to a transmitter, while your hands remain free to
test the response of thetransmitter. Select either sourcing or simulating by
choosing the SOURCE or SIMULATE jacks.
When the rotary switch is in the OUTPUT mA position, the meter produces a
continuously repeating 0% – 100% – 0% ramp in your choice of three ramp
waveforms:
0% – 100% – 0% 40-second smooth ramp, (default)
0% – 100% – 0% 15-second smooth ramp
0% – 100% – 0% Stair-step ramp in 25% steps, pausing 5 seconds at each step.
Steps are listed in Table 10.
The ramp times are not adjustable. Press the BLUE pushbutton to cycle through
the three waveforms.
Note
At any time during auto ramping, you can freeze the ramp simply by moving the
rotary switch to the position. Then you can use the COARSE, FINE, and % STEP
adjust pushbuttons to make adjustments.
Power-Up Options
To select a power-up option, hold down the pushbutton shown in Table 11 while
turning the rotary switch from OFF to any on position. Wait 2 seconds before
you release the pushbutton after powering up the meter. The meter beeps to
acknowledge the power-up option.
Only the setting for current span is retained when the power is turned off.
The others have to be repeated for each operating session.
You may activate more than one power-up option by holding down more than one
pushbutton.
Table 11. Power-Up Options
Option | Pushbutton | Default | Action Taken |
---|---|---|---|
Change current span 0% setting | Remembers last setting | Toggles between 0 |
and 4 mA
Disable beeper| | Enabled| Disables beeper
Disable auto power-off| BLUE| Enabled| Disables the feature that turns off the
meter power after 30 minutes of inactivity. Auto power off is disabled
regardless of this option if MIN MAX recording is on.
Battery Life
Warning
To avoid false readings, which could lead to possible electric shock or
personal injury, replace the battery as soon as the battery indicator (B)
appears.
Table 12 shows typical alkaline battery life. To preserve battery life:
- Use current simulation instead of sourcing when possible.
- Avoid using the backlight.
- Do not disable the automatic power-off feature.
- Turn the meter off when you are not using it.
Table 12. Typical Alkaline Battery Life
Meter Operation | Hours |
---|---|
Measuring any parameter or simulating current | 80 |
Sourcing 12 mA into 500W | 12 |
Using the Holster and Flex-Stand
The meter is supplied with a snap-on holster that absorbs shocks and protects
the meter from rough handling. You can turn the meter over in the holster to
protect the face of the meter from scratches when carrying the meter.
The holster is equipped with a Flex-Stand bail. Some uses of the holster with
Flex-Stand are shown in Figure 9.
Maintenance
This section provides some basic maintenance procedures. Repair, calibration, servicing not covered in this manual must be performed by qualified personnel. For maintenance procedures not described in this manual, contact a Fluke Service Center.
General Maintenance
Periodically wipe the case with a damp cloth and detergent; do not use
abrasives or solvents.
Calibration
Calibrate your meter once a year to ensure that it performs according to its
specifications. Contact a Fluke Service Center for instructions.
Replacing the Battery
Warning
To avoid electrical shock, remove test leads from the meter before you open
the battery door.
Close and latch the battery door before you use the meter.
Remove test leads from the meter before you open the battery door.
Replace the battery as follows. Refer to Figure 10. Use an alkaline 9V
battery, type ANSI/NEDA 1604A or IEC 6LR61.
- Remove the test leads and set the rotary switch to OFF.
- With a standard blade hand screwdriver, turn each battery door screw counterclockwise so that the slot is parallel with the screw picture molded into the case.
- Lift off the battery door.
Replacing a Fuse
Warning
To avoid personal injury or damage to the meter, use only the specified
replacement fuse, 440 mA 1000V fast-blow, Fluke PN 943121.
Both current input jacks are fused with a separate 440 mA fuse. To determine
if a fuse is blown:
- Turn the rotary switch to mA A .
- Plug the black test lead into COM, and the red test lead into ****A.
- Using an ohmmeter, check the resistance between the meter test leads. If the resistance is about 1Ω, the fuse is good. An open means the fuse is blown.
- Move red test lead to ****mA.
- Using an ohmmeter, check the resistance between the meter test leads. If the resistance is about 14Ω, the fuse is good. An open means the fuse is blown.
If a fuse is blown, replace it as follows. Refer to Figure 11 as necessary:
- Remove the test leads from the meter and turn the rotary switch to OFF.
- Remove the battery door.
- Remove the three Phillips-head screws from the case bottom and turn the case over.
- Gently lift the bottom of the front of the case (nearest the input/output jacks) until the top unsnaps from the rear half of the case.
- Replace the blown fuse with the exact type specified: 440 mA 1000V fast-blow fuse, Fluke PN 943121. Both fuses are the same type.
- Make sure the rotary switch is in the OFF position.
- Fit the top of case together, engaging the two snaps (item 1). Make sure that the gasket is properly seated.
- Close the case and reinstall the three screws.
- Replace the battery door.
If the Meter does not Work
- Examine the case for physical damage. If there is damage, make no further attempt to use the meter, and contact a Fluke Service Center.
- Check the battery, fuses, and test leads.
- Review this manual to make sure you are using the correct jacks and rotary switch position.
If the meter still does not work, contact a Fluke Service Center. If the meter is under warranty, it will be repaired or replaced (at Fluke’s option) and returned at no charge. See the Warranty on the back of the title page for terms. If the warranty has lapsed, the meter will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price.
Replacement Parts and Accessories
Warning
To avoid personal injury or damage to the meter, use only the specified
replacement fuse, 440 mA 1000V fast-blow, Fluke PN 943121.
Note
When servicing the meter, use only the replacement parts specified here.
Replacement parts and some accessories are shown in Figure 12 and listed in
Table 13. Many more DMM accessories are available from Fluke. For a catalog,
contact your nearest Fluke distributor.
To find out how to order parts or accessories use the telephone numbers or
addresses shown on page 1 of this manual.
Table 13. Replacement Parts
Item | Description | Fluke PN or Model no. | Quantity |
---|---|---|---|
BT1 | 9V battery, ANSI/NEDA 1604A or IEC 6LR61 | 614487 | 1 |
CG81Y | Holster, Yellow | CG81G | 1 |
F1, 2 | Fuse, 440 mA, 1000V fast-blow | 943121 | 2 |
MP85 | Case top | 619962 | 1 |
MP86 | Case bottom | 619939 | 1 |
H2, 3, 4 | Case screw | 832246 | 3 |
MP89, 90 | Non-skid foot | 824466 | 2 |
MP8 | O-ring for input/output receptacle | 831933 | 1 |
MP92 | Battery door | 619947 | 1 |
H5, 6 | Battery door fasteners | 948609 | 2 |
S1 | Keypad | 646932 | 1 |
TL75 | Standard test lead set | TL75 | 1 |
AC70A | Alligator clips for use with TL75 test lead set | AC70A | 1 |
TL20 | Industrial test lead set | TL20 | Option |
TM1 | Product Overview Manual | 1586717 | 1 |
TM2 | Users Manual (CD-ROM) | 1586721 | 1 |
TM3 | Calibration Manual (not shown) | 641891 | Option |
Specifications
All specifications apply from +18°C to +28°C unless stated otherwise.
All specifications assume a 5 minute warmup period.
The standard specification interval is 1 year.
Note
“Counts” means number of increments or decrements of the least significant
digit.
DC Volts Measurement
Range (V dc)| Resolution| Accuracy, ± (% of Reading +
Counts)
---|---|---
4.000| 0.001V| 0.1% + 1
40.00| 0.01V| 0.1% + 1
400.0| 0.1V| 0.1% + 1
1000| 1V| 0.1% + 1
Input impedance: 10 M W (nominal), < 100 pF
Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz Common mode rejection
ratio: >120 dB at dc, 50 Hz, or 60 Hz Overvoltage protection: 1000V
DC Millivolts Measurement
Range (mV dc) | Resolution | Accuracy (% of Reading + Counts) |
---|---|---|
400.0 | 0. 1 mV | 0.1% + 1 |
AC Volts Measurement
Range (ac) | Resolution | Accuracy, ± (% of Reading + Counts) |
---|---|---|
50 Hz to 60 Hz | 45 Hz to 200 Hz | 200 Hz to 500 Hz |
400.0 mV | 0.1 mV | 0.7% + 4 |
4.000V | 0.001V | 0.7% + 2 |
40.00V | 0.01V | 0.7% + 2 |
400.0V | 0.1V | 0.7% + 2 |
1000V | 1V | 0.7% + 2 |
Specifications are valid from 5% to 100% of amplitude range. AC conversion:
true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add ± (2% reading + 2% f.s.) typical Input
impedance: 10 M W (nominal), < 100 pF, ac-coupled Common mode rejection
ratio: >60 dB at dc, 50 Hz, or 60 Hz
AC Current Measurement
Range
45 Hz to 2 kHz
| Resolution| Accuracy, ± (% of Reading + Counts)| Typical
Burden Voltage
---|---|---|---
1.000A (Note)| 0.001A| 1% + 2| 1.5V/A
Note: 440 mA continuous, 1A 30 seconds maximum
Specifications are valid from 5% to 100% of amplitude range. AC conversion:
true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add ± (2% reading + 2% f.s.) typical
Overload protection 440 mA, 1000V fast-blow fuse
DC Current Measurement
Range| Resolution| Accuracy, ± (% of Reading + Counts)|
Typical Burden Voltage
---|---|---|---
30.000 mA| 0.001 mA| 0.05% + 2| 14 mV/mA
1.000A (Note)| 0.001A| 0.2% + 2| 1.5V/A
Note: 440 mA continuous, 1A 30 seconds maximum
Overload protection: 440 mA, 1000V fast-blow fuse
Ohms Measurement
Range| Resolution| Measurement Current| Accuracy, ± (%
of Reading + Counts)
---|---|---|---
400.0W| 0. 1W| 220 mA| 0.2% + 2
4.000 kW| 0.001 kW| 59 mA| 0.2% + 1
40.00 kW| 0.01 kW| 5.9 mA| 0.2% + 1
400.0 kW| 0.1 kW| 590 nA| 0.2% + 1
4.000 MW| 0.001 MW| 220 nA| 0.35% + 3
40.00 MW| 0.01 MW| 22 nA| 2.5% + 3
Overload protection: 1000V Open circuit voltage: <3.9V
__**Frequency Counter Accuracy**
Range | Resolution | Accuracy, ± (% of Reading + Counts) |
---|---|---|
199.99 Hz | 0.01 Hz | 0.005% + 1 |
1999.9 Hz | 0.1 Hz | 0.005% + 1 |
19.999 kHz | 0.001 kHz | 0.005% + 1 |
Display updates 3 times/second at >10 Hz
Frequency Counter Sensitivity
Input Range | **Minimum Sensitivity (rms Sinewave) 5 Hz to 5 kHz*** |
---|---|
1 V | 0.1 V |
4 V | 1 V |
40 V | 3 V |
400 V | 30 V |
1000 V | 300 V |
- Usable 0.5 Hz to 20 kHz with reduced sensitivity.
Diode Test and Continuity Test
Diode test indication: display voltage drop: 0.2 mA
nominal test current at 0.6V: 2.4V full scale, accuracy ±(2% + 1 count)
Continuity test indication: continuous audible tone for test resistance <100Ω
Open circuit voltage: <3.9V
Short circuit current: 1.2 mA typical
Overload protection: 1000V rms
DC Current Output
Source mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA
Accuracy: 0.05% of span
Compliance voltage: 12V with battery voltage >8.5V
Simulate Mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA
Accuracy: 0.05% of span
Loop voltage: 24V nominal, 30V maximum, 15V minimum
Compliance voltage: 21V for 24V supply
Burden voltage: <3V
General Specifications
Maximum voltage applied between any jack and earth ground: 1000V
Storage temperature: -40°C to 60°C
Operating temperature: -20°C to 55°C
Operating altitude: 2000 meters maximum
Temperature coefficient: 0.05 x specified accuracy per °C for temperatures
<18°C or >28°C
Accuracy adders for use in RF Fields: In an RF field of 3V/m, change the
accuracy specifications as follows:
For DC Millivolts Measrement, add 0.03% of range
For AC Volts Measurement, add 0.37% of range
For DC Current Measurement, 30.000 mA range, add 0.14% or range
For DC Current Output, add 0.02% of span
Accuracy for all meter functions is not specified in RF fields > 3V/m.
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, and 35% up
to 55°C
Vibration: Random 2g, 5 to 500 Hz
Shock: 1 meter drop test
Water and dust protection: Complies with IEC529 IP52 (normal operating vacuum
used for dust test)
Safety: Complies with IEC1010-1, ANSI/ISA S82.011994 and CAN/CSA C22.2 No.
1010.1-92 Overvoltage Category III.
Certifications: CSA, UL, TÜV
Power requirements: Single 9V battery (ANSI/NEDA 1604A or IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L);
With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in
W x 7.93 in L)
Weight: 369 g (13 oz);
With holster and Flex-Stand: 638 g (22.5 oz)
April 1997, Rev.3, 12/01
© 1997, 1898, 2000, 2001 Fluke Corporation, All rights reserved. Printed in
U.S.A.
All product names are trademarks of their respective companies.
Documents / Resources
|
FLUKE 787
ProcessMeter
[pdf] User Manual
787 ProcessMeter, 787, ProcessMeter
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>