WHALETEQ DFS100 Defibrillator Handheld Tester User Manual

June 15, 2024
WHALETEQ

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual

WHALETEQ DFS100 Defibrillator Handheld Tester User
Manual

Copyright © 2013-2023, All Rights Reserved. WhaleTeq Co. LTD

No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language or computer language, in any form, or by any means, electronic, mechanical, magnetic, optical, chemical, manual or otherwise, without the prior written permission of WhaleTeq Co. LTD.

Disclaimer WhaleTeq Co. LTD. provides this document and the programs “as is” without warranty of any kind, either expressed or implied, including, but not limited to, the implied warranties of merchantability or fitness for a particular purpose.

This document could contain technical inaccuracies or typographical errors. Changes are periodically made to the information herein; these changes will be incorporated in future revisions of this document. WhaleTeq Co. LTD. is under no obligation to notify any person of the changes. The following trademarks are used in this document:
is a registered trademark of WhaleTeq Co. LTD All other trademarks or trade names are property of their respective holders.

1 Introduction

1.1 Concept

WhaleTeq DFS100 is designed for AED field tester. AED (Automated External Defibrillator) is a life-saving medical device that delivers defibrillation upon identifying and analyzing the ECG of a resuscitationneeded patient. Because AEDs are installed in public areas where they can be reached before ambulance arrival, the majority of AEDs rely on one-time batteries and requires only minimal training compared to hospital-use defibrillators. Ensuring the functionality of AEDs periodically becomes widely accepted around the world, and DFS100 is the ideal tool to conduct such a regular inspection.

1.2 Applications

WhaleTeq DFS100 is a compact field tester, functioning as an ECG signal simulator and defibrillation energy guarantee equipment for AED regular maintenance.

For ECG patterns simulation, DFS100 simulates NSR (Normal Sinus Rhythm) as well as other arrhythmias that may need AED defibrillation. Arrhythmia convert combination (ECG waveform before vs. after defibrillation) is also available.

For energy guarantee, DFS100 allows User to customize the pass/fail criteria of the delivered defibrillation energy. To ensure the durability of DFS100, protection against multiple defibrillation is implemented. (DFS100 should be used to analyze defibrillation energy at the interval of at least 60 seconds)

2 Specification

2.1 General Specifications

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - General
Specifications

2.2 Compliance Specifications

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Compliance
Specifications

2.3 Energy Measurement Specifications

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Energy
Measurement Specifications

2.4 ECG Signal Specifications

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - ECG Signal
Specifications

3 Instrument Familiarization

3.1 Top

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Figure
1

A. Ventilation Holes: Allow ventilation during operation. Keep this side up when in use.
B. AED Discharge LED:

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - AED Discharge
LED

C. ECG Pattern: There are 3 ECG patterns to choose from: VT (Ventricular Tachycardia), VF (Ventricular Fibrillation), or NSR (Normal Sinus Rhythm).
D. On/Off and Pattern Selection Button:
WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - On Off and
Pattern Selection Button

F. DFS Battery LED:
WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - DFS Battery
LED

3.2 Front and Rear

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Figure
2

  • G. AED Interface Jack (Apex): Connect the Apex jack to the AED pad labeled Apex via DFS accessory “AED Interface Cable”.
  • H. AED Interface Jack (Sternum): Connect the Sternum jack to the AED pad labeled Sternum via DFS accessory “AED Interface Cable”.
  • I. Caution: DFS100 is recommended to be used for analyzing defibrillation shock at the interval of at least 60 seconds apart.
  • J. Caution, the possibility of electric shock: From the time AED starts charging for defibrillation until right after energy delivery, User is recommended to stay clear from DFS100.
  • K. USB Port: Connect to the computer’s USB port for data transferring.
3.3 Back

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Figure
3

  • L. Label: Contain information including model, serial number, manufacturer, and power supply.
  • M. Battery Compartment: Insert a 9V battery for power supply. Both disposable batteries and rechargeable batteries can be used.
3.4 Wiring Diagram

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Figure
4

Note: To plug in properly, please make sure the yellow lines on the black and red banana males that connect to DFS100 totally disappear in the jacks.

4 Getting Started

4.1 PC System Requirements

DFS100 Setup Software uses a Windows PC to connect and modify DFS100 via USB module.
PC requirements:
– Windows PC (Windows 7 or later, suggest to use the genuine version)
– Microsoft .NET 4.0 or higher
– Administrator access (essential for installing software, driver, and Microsoft .Net Framework)
– 1.5 GHz CPU or higher – 1GB RAM or higher1 – USB port

4.2 Software Installation

Please follow the steps below to download and execute DFS software.

  1. Download DFS100 software from WhaleTeq’s website.
  2. Open the file explorer and select the download location.
  3. Double click on WhaleTeqDFS100.exe to execute DFS100 PC software.

Note:
If DFS software cannot be executed properly, or this is the first time using WhaleTeq’s product, please refer to the two sections below (4.3 and 4.4) to confirm that the USB driver and Microsoft .Net Framework 4.0 are both installed.

1 Relative to normal PC processing, there is no special use of PC speed. However, there has been noted a slow increase in system RAM usage over long periods of time up to 30-40MB (related to MS Windows “garbage collection”). PCs with only 512MB or less installed and are running several other programs (in particular, Internet Explorer), may exceed the available RAM, requiring access to the hard drive and dramatically impacting speed. In this case, streaming interruptions and other problems may occur.

4.3 USB Driver Installation

If Windows Device Manager can’t recognize WhaleTeq’s product, please follow the instructions below to install Microchip® USB Driver.

For Microsoft Windows 10

  • As Windows 10 has built-in Microchip® USB Driver, there is no need to install any drivers. It just takes a while for Windows Device Manager to recognize and install the driver.

For Microsoft Windows 8 and Windows 8.1

  • If Windows 8 and Windows 8.1 can’t recognize DFS100 unit, please download “mchpcdc.inf” from WhaleTeq’s website. This driver is provided by Microchip®, and is intended to be used with PIC microprocessors which have built-in USB functions.
  • As mchpcdc.inf provided by Microchip® does not contain a digital signature, please disable the driver signature enforcement in Windows 8 and Windows 8.1. Please click here to watch the tutorial video.
  • When the USB module is connected for the first time, select Manual Installation, and point to the folder containing the above file. Then, continue to follow the instructions to finish the installation. There may be a warning that the driver is not recognized by Windows®, and this can be ignored. Please click here to watch the tutorial video.

For Microsoft Windows 7

  • If Windows 7 can’t recognize DFS100 unit, please download “mchpcdc.inf” from WhaleTeq’s website. This driver is provided by Microchip®, and is intended to be used with PIC microprocessors which have built-in USB functions.
  • When the USB module is connected for the first time, select Manual Installation, and point to the folder containing the above file. Then, continue to follow the instructions to finish the installation. There may be a warning that the driver is not recognized by Windows®, and this can be ignored. Please click here to watch the tutorial video.
4.4 Microsoft .Net Framework Installation

WhaleTeq’s software is developed based on Microsoft .Net Framework 4.0. If DFS software fails to execute properly, please check whether Microsoft .Net Framework 4.0 or a higher version was installed in the operating system. If the PC hasn’t installed Microsoft .Net Framework 4.0 or a higher version, please download it from Microsoft’s website. Please click here to watch the tutorial video (from 2:03).

4.5 PC Software and Firmware Auto-Update

User can download the latest software from WhaleTeq’s website. If User cannot update the DFS software, please contact WhaleTeq’s service team (service@whaleteq.com).

5 PC Software Operation

5.1 Software Function Introduction

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Figure
5

The DFS100 PC Software is solely for customized settings. It does not include testing functions.

  • A. Default Pattern When On: User can choose from VT (Ventricular Tachycardia), VF (Ventricular Fibrillation), or NSR (Normal Sinus Rhythm) as the default playing pattern when turning on DFS100.

  • B. Return To Initial ECG Pattern: When using ONE-TIME TEST mode , User has the option to enable the “Auto Return to Initial ECG Pattern” function, which does so upon 30 seconds after receiving defibrillation.
    Note: If User doesn’t enable the “Auto Return to Initial ECG Pattern” function, User will need to manually switch the pattern from NSR to VT or VF if needed for the next inspection.

  • C. Joule Threshold: Depending on different models of DUT (AED), the delivered defibrillation is different. DFS100 allows User to selfdefine the passing criteria of the defibrillation. If the joule of the defibrillation equals to or is higher than this value, the AED Discharge LED lights green; otherwise, lights red.
    Note: User can set the parameter from 40 to 360 Joules.

  • D. Auto Off: DFS100 will automatically be turned off after being idle for a certain time. The default value is 3 minutes.
    Note: User can set the time from 3 to 10 minutes.

  • E. Check for Updates: Click to see if the latest version of firmware and/or software has been released. Please note that the USB cable used for firmware upgrade must be file-transferrable.

  • F. About: Check the software version and the device function introductions.

6 Standalone Operation

6.1 One-Time Test (Convert) ⇆

In this mode, the test starts immediately upon DFS100 being turned on, and it simulates the ECG pattern of a resuscitated-needed patient.
Ensure the AED and DFS100 are connected properly via the interface cable. Turn on both DFS100 and the AED, and follow the AED instructions. The default setting of DFS100 is sending VT to the AED. After the AED delivers defibrillation shock, DFS100 shows the pass/ fail of the defibrillation energy by lighting different LED colors for 30 seconds.
Note: If the test passes, the LED lights green. If the test fails, the LED lights red.
At the same time, DFS100 simulates the patient having been resuscitated by changing the ECG pattern to NSR.

6.2 Continuous Arrhythmia Test ♻

In this mode, DFS100 continuously sends user-selected arrhythmia to the AED.
Ensure the AED and DFS100 are connected properly via the interface cable. Turn on both DFS100 and the AED, and follow the AED

instructions. The default setting of DFS100 is sending VT to the AED. After the AED delivers defibrillation shock, DFS100 shows the pass/ fail of the defibrillation energy by lighting different LED colors for 30 seconds.
Note: If the test passes, the LED lights green. If the test fails, the LED lights red.
Then, User may connect DFS100 to the other AED for the next test. The test starts without User pressing any button on DFS100.

7 Calibration and Validation

It is recommended to get DFS100 calibrated annually. Calibration and validation both require traceable equipment. Steps for Calibration/ Validation methods are available upon request. Please contact Whaleteq for more details (See section 13).

8 Troubleshooting

  1. If after the AED defibrillation shock, and the DFS100 “Discharge” LED does not light up, check the AED interface cable connection.
  2. If User is certain that the AED defibrillation energy matches its spec, but the “Discharge” LED lights red, check the PC software setting (section 5.1 (C)) to ensure the joule threshold is set correspondingly.
  3. When DFS100 “Battery” LED blinks red, change the battery at User’s earliest convenience.

9 Cautions

  1. Follow AED instructions during testing with DFS100.
  2. Follow the color code and / or labeling when connecting DFS100 and the AED via the AED interface cable.
  3. When DFS100 is in use, the face containing LEDs and buttons must face up to allow ventilation.
  4. To allow ventilation, DFS100 should be used to analyze defibrillation energy at the interval of at least 60 seconds.
  5. USB cable used for firmware upgrade must be file-transferrable.
  6. Battery, whether one-time use or rechargeable, must be included in DFS100 for accurate energy measurement.
  7. When Battery LED blinks red, the Discharge LED may dim and blink at a faster frequency. This is a warning to change the battery as soon as possible in order to ensure accurate test results.
  8. USB Port cannot be used to charge the battery.
  9. For best performance, rechargeable battery capacity is recommended to be 800 mAh or above.
  10. When not in use for a longer period, remove the battery from DFS100 to ensure a longer lifespan of the battery.
  11. Clean the external case of the DFS100 with a clean dry cloth. Do not allow liquid inside DFS100 or near the I/O sockets.
  12. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

10 Ordering Information

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Ordering
Information WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Ordering
Information

  • Customization of AED interface cable adapter is available. Please contact WhaleTeq (service@whaleteq.com)

11 Package Contents

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Package
Contents

DFS100 Optional Accessories

✔ USB transfer cable: Type-A to Type-C (Male to Male), 1.8 meter
✔ DFS carrying bag

12 Revision History

WHALETEQ DFS100 Defibrillator Handheld Tester User Manual - Revision
History

13 Contact WhaleTeq

WHALETEQ Co., LTD
service@whaleteq.com | (O)+886 2 2517 6255
8F., No. 125, Songjiang Rd., Zhongshan Dist., Taipei City 104474, Taiwan

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals