AILE X5 Electronic Sphygmomano Meters Instruction Manual

June 15, 2024
AILE

AILE X5 Electronic Sphygmomano Meters

Preface

  • Dear users, We appreciate for your purchase of our Digital arm blood pressure monitor.
  • Before using this monitor, read the Manual carefully and use it correctly. Please keep the
  • Manual properly after reading for easy check and reference at any time.
  • This monitor is for adults use, but not for newborns.

Introduction

  • It is an Electronic Sphygmomanometers consisting of the Main Unit and Cuff. It is suitable for measurement of systolic blood pressure, diastolic blood pressure and pulse of the human body. The blood pressure measured by it is equivalent to that measured by auscultation method, and the error is controlled within the range specified in EN IEC 80601-2-30
  • Non-invasive automated monitor. This Electronic Sphygmomanometers employs the intelligent pressurization to automatically pressurize to a suitable pressure value based on your blood pressure. It is designed to effectively reduce discomfort caused by incorrect pressurization, shorten measurement time, and extend the life of the Cuff.

Range of application

Untended Use

  • Electronic Sphygmomanometers measures the diastolic pressure and systolic pressure and pulse rate of people over 12 years old in home environments by using a non-invasive
  • Oscillometric technique with a single upper arm cuff.
  • The measurement position is at human being’s upper arm.
  • All values can be read out in one LCD panel.
  • The device is designed for home use and recommended for use by adults over aged 18 with upper arm circum­ference ranging from 22cm~42cm.

Scope of application of product

  • This product is used for measurement of blood pressure of people over 12 years old and not for newborns, not for pregnant, including pre-eclamptic patients;

Intended Users

  • Adult, except contraindications
  • Patient populations
  • People over 12 years, except contraindications

Parts of the body/tissues, duration

  • This device is a non-invasive Electronic Sphygmoma­nometers, the cuff will be ware in upper arm by user during measuring.

Single Use/Reusable

  • This product is an active and reusable device.

Contraindications

  1. The product shall not be used for measuring for under 12 years old;
  2. Those who suffer from severe arrhythmia shall not use the this product;
  3. Do not use this monitor on the injured arm or the arm under medical treatment;
  4. Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.

Important Safety Information

Warnings
To measure the blood pressure accurately and reliably, followings shall be noted when measuring!

  1. Don’ t disassemble the monitor or arm cuff. This may cause an inaccurate reading.
  2. Do not apply the arm cuff on the arm while being on an intravenous drip or blood transfusion.
  3. Consult your physician before using this monitor on the arm with an arterio-venous (A-V) shunt.
  4. Don’ t use the Cuff on the arm on the side of a mastectomy or lymph node clearance.
  5. Do not use this monitor with other medical electrical (ME) equipment simultaneously. This may result in incorrect operation of the monitor and/or cause an inaccurate reading.
  6. Measurements may be impaired if the device is used close to televisions, microwave ovens, cellular telephones, X-ray or other device with strong electrical fields.
  7. Don’ t use this monitor on the injured arm or the arm under medical treatment, as this can cause further injury.
  8. Do not use in a location with moisture, or a location where water may splash on this monitor. This may damage this monitor.
  9. Use only the approved arm cuff for th is monitor. Use of other arm cuffs may result in incorrect measurement results.
  10. Do not fold or kink the air tube while taking a measurement. This may cause harmful injury by interrupting blood flow.
  11. Do not measure while standing, walking, or your body is under pressure.
  12. Do not measure after smoking, drinking alcohol, or coffee (black tea).
  13. Do not measure after exercise or bathing.
  14. Do not speak or move your body during measurement.
  15. Do not move, shake your arms or bend your fingers during measurement.
  16. Do not measure under the excessively cold or hot condition, or the environment suffering drastic changes.
  17. Do not measure in a moving vehicle (car, airplane).
  18. Measure too frequently can cause injury to the patient due to blood flow interference, (it is recom­mended that the interval between two adjacent measurements be no less than 3 minutes)
  19. Do not move the monitor during the measurement.
  20. Do not measure within 1 hour after meal.
  21. Note: It would be dangerous by making self-judg­ment and self-treatment according to the measure­ment resuIts. Please follow the doctor’s advice.
  22. The Mains supply adaptor is not provided by this product. DC SV Input port only can be powered by approved (IEC/EN 60601-1) DC adaptor (DC SV/500 mA). Use of other DC adaptor may damage this monitor.
  23. Avoid compression or restriction of the connection tubing.

Notes

  1. Avoid eating, smoking and strenuous exercise for 30 minutes prior to blood pressure measurement.
  2. Sitting is the correct measurement posture for measurement of blood pressure, and the Cuff and the heart should be on the same level. Therefore, the blood pressure value measured at the front of the counter may be slightly different from value that is measured correctly.
  3. Any blood pressure measurement is affected by the posture and physical condition of the subject. Please be quiet while measuring, and do not move your arms or body.
  4. Maintain the same state and same posture for each time of measurement.
  5. The monitor of this model should be tied on the levee
  6. When continuous measurement is made on the same person, loosen the Cuff to have the arm rested for at least 3 minutes before next measurement.
  7. The Cuff provided by this product is suitable for arms of people over 12 years old.
  8. Avoid electromagnetic interference above lOV/m during measurement. The possible sources of interfer­ence may be from TV, mobile phone, microwave oven, X-ray, etc.
  9. Do not measure the blood pressure of newborn babies using this product.
  10. Do not have this product used for purposes other than blood pressure measurement.
  11. Remove the batteries when the product is not in use for a long time to prevent the battery from leakage.
  12.  When the bladder of cuff is over-pressurized constantly, the pain, numbness and even congestion may occur to the arm. Do not pressurize when it exceeds 300 mHg, and do not maintain the cuff pressure higher than 15mmHg for more than 3 minutes.
  13. Arrhythmia such as PAC, PVC, and atrial fibrillation during the measurement process may lead to the detection of irregular pulse rate. The electronic monitor can still work normally, but the measured blood pressure value may be inaccurate. When irregularities occur frequently, please consult a doctor.
  14. 1f the monitor is used and stored beyond the temperature and humidity range specified in this Manual, its safety or performance may be affected or become invalid, or even be totally damaged.
  15. When replacing the Cuff, use the Cuff supplied by the manufacturer. If the original part is replaced with a Cuff not supplied by the manufacturer, measurement errors may occur.
  16. Contact the manufacturer for static pressure test.
  17. Do not share the Cuff with others to avoid cross-contamination.
  18. Don’ t use the device in the environment with higher than 40°C or lower than 5°C, and over RH 15%~ 90%.
  19. The protective glass outside the LCD frame is very important and also the fragile part of the instrument, so it must be used carefully.
  20. Please do not charge the non-rechargeable battery and do not throw the battery into the fire.
  21. Please do not expose the product to the sun or touch water.
  22. To help avoid strangulation, keep the air tube away from infants, toddlers or children.
  23. Keep components out of the reach of infants, toddlers or children. This product contains small parts that may cause a choking hazard if swallowed by infants, toddlers or children.
  24. Keep batteries out of the reach of infants, toddlers or children.
  25. Do not store your monitor and other components:
    • If the monitor and other components are wet.
    • In locations exposed to extreme temperatures, humidity, direct sunlight, dust or corrosive vapors such as bleach.
    • In locations exposed to vibrations or shocks.

Potential adverse reaction

  1. Patients with severe blood circulation disorders and blood diseases, Inflate the arm cuff than necessary may cause blood stasis or paralysis of the arm;
  2. Patients with severe blood circulation disorders and blood diseases may cause acute internal bleeding due to the compression of the arm.

Waste disposal

  1. Direct disposal of electronic products and batteries in garbage cans will cause harm to the environment. Please dispose them according to the local laws.
  2. Don’t throw the Electronic Sphygmomanometers in the garbage can at the end of use; please dispose it according to the local laws or contact with the manufac­turer for recycling.

Features and Specification

List of product

No. Part Name Model Quantity
1) Main unit XS 1 set
2) Cuff URN-2242 l set
3) Instructions Manual I 1 copy
D evice Name Electronic Sphygmomanometer
--- ---
Model X5
Display Digital LCD Digital LCD
Method of measurement Weight Dimensions Oscillometric About 245 g

Appr. 135 (L) X 100 (W) X 68 mm (H)

Cuff| Appr. 150 mm (W) x 450 mm (L)
Cuff Measurement perimeter of the upper ann| 220mm,.,420mm

  • Material of cuff Measuring pressure range
  • Measuring pressure accuracy

|

  • Nylon polyester

  • Blood pressure: 0,.,295 mmHg, 0~40 kPa

  • Pressure: within ±3 mmHg (±0.4 kPa), or 2% of the reading

  • Pulse range Pulse accuracy

  • Pressure measurement

|

  • (40-160) times per minute
  • Within ±5% of the reading Resistance-type pressure sensor

Working mode| Continuous operation

  • Number of memory
  • Air-blow device

|

  • 2 people x 99 groups
  • Blow-off valve
Inflation Auto pressurization by air pump
Venting Uniform-speed decompression by solenoid valve
Auto power-off Two minutes
Power supply MA batteries (4pc) or DC 5V 500 mA
Operating condition
  • Temperature: (+5 ~ +40}°C
  • Relative Humidity: (15% ~ 90%)RH
  • Atmospheric’/pressure: 70 kora – 106 kpa
  • Power: DC 5 (4 AAA batteries}

Storage and transportation environment|

  • Environment temperature: -25°C~ +55°C;
  • Relative humidity: 10% – 95% (non-con- denting)
  • No corrosive gas, well ventilated room.
  • Transportation requirements must conform to regulations of the order contract, but severe impacts vibration and rain and snow splash during transportation must be prevented.

IP class| IP 21
Applied parts| Type BF
Electromagnetic compatibility| Group 1 and Class B device
Accessories| Cuff (for arm circumference of 22 cm ~ 42cm) !FU
Software release version| Vl.O

Structure and descriptions of product
Product Picture

Product structure
The product consists of a main unit and a cuff. See below Figures.

AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-2

Description of LCD Screen
The product consists of a main unit and a cuff. See below Figures.

Settings

  1. Memory clear
    Press and hold the ” MEM
    key for 3 seconds in the reading memory mode to clear the memory value stored in the product.

  2.  Set the date/time

    • “Year” When the monitor is turned on but not measured, press and hold the
    • “SET” key for about 3 seconds, the Year on screen starts to flash, and press the
    • “MEM ” key to set the year.
    • “Month” After the year is adjusted, press the “SET “key again, the Month starts to flash at this time, and press the “MEM” key to make adjustment. 11Day”
      After the month is adjusted, press the “SET “key again, the Date starts to flash at this time, and press the “MEM” key to make adjustment. 1Hours11/”Minutes” Upon date adjustment, press the “SET” key again and the Hour starts to flash, press the “MEM key to adjust. Set the “Minutes” by following the above-men­tioned steps.
  3. Unit conversion
    The unit is defaulted as “mm Hg” . After setting “Date/­Time” , press “SET “key and “O” is displayed. If you want to switch unit to “kPa” , press 11,MEM” key and 110.0″ will appear, which means unit conversion is successful. Pressing “MEM” key again to convert from “kPa” to 1mmHg’1

  4. “User”

    • The monitor stores the memory of 2 people x 99 groups, with each group of users enjoying 99 groups of memory. Upon measurement each time, the monitor automatically stores the measurement data. Once the memory capacity is full, the old measurement data will be overwritten by the new one.
    • By pressing the “SET ‘key when the monitor is powered on but no measurement is performed, the screen will display the current user group, and then press the “SET” key again to switch the user (“SET” or “MEM” optional). If no set is made, the User 1 SET is defaulted.
  5. Reading of main memory value

    • Press the “MEM” key after the measurement is finished or in the power-off mode, and the LCD will display and broadcast the averaged result of the last three measurements.
    • Press the “M EM” key again to display the last measurement result. View the measured values
      of 2, 3 … continuously by repeating this step.
  6. Error code description

    • Before using the arm-type electronic monitor, check to see if the monitor has an error code prompt or be unable to display the number (or symbol) after power on. If there’ re
    • error codes prompts or display abnor­mality, please contact the after-sales service center to eliminate the abnormality as per the prompts.

AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-5

Preparation

  • Open the Electronic Sphygmomanometers. The device case should be uniformly colored with no obvious defects such as scars and scratches.

Please keep it properly when carrying and storing.

  1. To load the batteries
  2. Open the Battery Case cover on the side of the unit and put the 4 batteries (AAA) correctly.
  3. Put the Battery Case cover back properly.

Notes:

  • When the depleted battery symbol () appears on the display, turn the monitor off and remove all the batteries. Replace with 4 new batteries at the same time. Long life alkaline batteries are recommended.
  • The measurement values continue to be stored in memory even after the batteries are replaced.
  • The supplied batteries may have a shorter life.
  • &Disposal of used batteries should be carried out in accordance with the national/local regulations for the disposal of batteries.

To tie the Cuff

  1. The standard accessory is the cuff (M), which is suitable for the subject with arm circumference of 22cm ~42cm.

  2. Insert the hose top plug into the opening located on the side of the Main Unit. (Only the original standard Cuff can be used for this Main Unit.)

  3. Wrap the arm Cuff firmly in place around your left upper arm.
    It is advisable to tie the Cuff properly that the arm will not feel physical tension.

  4. The bottom edge of the arm cuff should be 2 to 3 cm above the elbow.
    Air tube is on the inside of your arm and aligned with your middle finger.

  5. Secure closed with the fabric fastener.

Notes:

  • When you take a measurement on the right arm, the air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube. 
  • The blood pressure can differ between the right arm and the left arm, also the measured blood pressure values can be different. Al CARE recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your measure­ments.
  • Do not tie the Cuff too tight, otherwise the measure­ment will be inaccurate.

Refer to the range of arms indicated on the Cuff to determine if a properly sized Cuff is used.

  • Do not place the arm Cuff over thick clothes.

How to Sit Correctly

  • The correct posture is also extremely important to measurement.
  • To take a measurement, you need to be relaxed and comfortably seated at comfortable room temperature.
  1. Sit in a comfortable chair with your back and arms supported, and maintain the right posture; do not force with your fingers, and keep them stretched.
  2. Keep your feet flat and your legs uncrossed.
  3. Both arms should be resting comfortably on a table.
  4. Place your monitor close enough to touch with your bent elbows.
  5. The arm cuff should be placed on your arm at the same level as your heart for a blood pressure measure­ment.
  6. Please measure while the body is relaxed, and keep it in peace for 4 ~ 5 minutes before measurement. Otherwise, the measurement results will be affected. At the same time, make blood pressure measurement with the same arm and the same posture every day.
  7. Judge the measurement results under the guidance of a doctor to help you manage your health.
  8. When the person being measured suffers from arrhythmia (i.e. PAC, PVC, and atrial fibrillation), the measurement results may be inaccurate.

For the accuracy of the measurement, we recom­mend

  1. Do not have intense activities 30 minutes before the measurement, do not smoke or drink coffee or alcohol;
  2. Have rest and sit in a comfortable and quiet atmosphere for at least 5 minutes prior to measure­ment;
  3. Generally, the blood pressure is measured in a sitting position. If a prone position is required, remember to tell the doctor;
  4. Measure the blood pressure at least twice each time, with the interval of not less than S minutes. A longer interval may be required depending on your physical condition.
  5. Patients suffering from arrhythmia and arteriosclero­sis should be measured by medical workers so as to make professional diagnosis.
  6. 6)Avoid any electromagnetic interference or noise interfer­ence during measurement.

Method to measure the blood pressure

  1. Start measuring after you have worn the cuff by pressing the “Start/Stop” key. After that, start the measurement.
  2. Since the monitor also checks the pulse during the inflation process, do not move any part of your body during measurement.
  3. When the monitor is slowly vented, the value on the display keeps decreasing and the heartbeat icon (heart shaped) keeps flashing until the measurement results appear.
    • During During During inflation inflation venting
    • During Measuring venting result appears AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-14

Reading of measured values and make repeated measurements After the automatic measurement is finished, the display screen will automatically display the systolic pressure and diastolic pressure of the person under measurement in mm Hg, and simultane­ously display the measured pulse rate (in times/min­ute).

  • Press the “Start” key again to make another measure­ment.
  • Notes 1: Due to the fluctuation of blood pressure, each measurement result may vary. If the user’s blood pressure fluctuates greatly, it is best to measure 3 times and take the average value as the measured value. Compared with the casual blood pressure as measured, the blood pressure measured using the above method will be the true value of the human blood pressure. Note 2: The effectiveness of the device in measuring blood pressure is verified by auscultation.
  • Note 3 : If your left arm is not available for measure­ment, use the right arm. However, in any case, each measurement must be performed on the arm of the same side for an effective comparison of measure­ments.
  • Guidance for assessment of adult hypertension
  • Blood pressure is the pressure that blood is transmitted to the arteries through the heart-throb (contraction and relaxation).
  • The pressure at which the blood is delivered to the artery during systole is called systolic pressure, also as the higher pressure; and the pressure the time of diastole after the blood that circulates throughout the body is returned to the heart is called diastolic pressure, also as the lower pressure.
  • Figure S offers the standard blood pressure values of WHO. Hypotension is not defined. A blood pressure reading lower than 100 mmHg for the top number (systolic) is generally considered as hypotension. Guidelines for the classification of hypertension (regardless of age and gender) is provided. Please note that factors such as diabetes, obesity, and smoking would affect blood pressure measurements. If you have any questions,
  • please consult your doctor, and do not adjust your treatment or medication just based on one of the two measurements. AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-15

Variation in blood pressure:

  • Individual blood pressures vary greatly both on a daily and a seasonal basis. These variations are even more pronounced in hyper tense patients. Normally the blood pressure rises while at work and is at its lowest during sleeping period. (hypertense: means a person who has high blood pressure symptom.)
  • The graph below illustrated the variations in blood pressure over a whole day with measurement taken every five minutes.
  • The thick line represents sleep. The rise in blood pressure at 4PM (A in the graph) and 12 PM (B in the graph) correspond to an attach of pain.

Requirements of Storage and mainte­nance

  1. Avoid falling or strong collisions.
  2. Avoid high temperatures or exposure to direct sunlight.
  3. on’ tin any twist the cuff.
  4. Do not press “START/STOP” button if the cuff is not wrapped around your upper arm.
  5. Avoid immersing the unit and the Cuff into water.
  6. Please leave the product indoors for one hour before using if it is brought into a warm room from a cold environment.
  7. Remove the battery when the product will not be used for 3 months or more.
  8. Use only four “AAA” batteries with this Main Unit. Don’t use new and used batteries together.
  9. Do not implement the maintenance procedure for equipment during measurement.
  10. Put in a dry place and avoid sunshine, avoid immersing it in the water.
  11. Make sure the cuff is completely dry before using.
  12. Do not attempt to disassemble or change any parts of the monitor, including arm cuff, due to substitution of a component different from that supplied might result in measurement error.
  13. After each use, wipe the cuff with a thoroughly wrung-out cloth that is dampened with water or a mild detergent, then dry with a dry cloth. If necessary, gently wipe it with absorbent cotton soaked with 75% alcohol for disinfection.
  14. 1f the product is not used for a long period of time, it is recommended to start the product once every two months. After the monitor is running, clean it with a soft and dry cloth.
    If the main unit gets dirty, clean it with a soft and dry cloth. If the main unit is extremely dirty, wipe its body thoroughly with a wrung-out cloth that is dampened with water or a mild detergent, and dry with a dry cloth. If necessary, gently wipe it with absorbent cotton soaked with alcohol for disinfection.

Calibration and Service

  1. The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life.
  2. It is generally recommended to have the device inspected every 2years to ensure correct functioning and accuracy. Please consult your local Al CARE representative.
  3. Service life information:
    • Main unit: 5 years (6 times/day);
    • Cuff: The Cuff is a consumable. The life of the Cuff is about 1 year when measured 6 times a day (3 times in the morning and 3 times in the evening), (under the test conditions of manufacturer);
    • In order to measure blood pressure correctly, it is recommended to replace it in time.
    • In case of air leakage, replace the cuff with a new one.

Instructions for Waste Disposal

  • Disposal of used batteries should be carried out in accordance with the national/local regulations for the disposaI of batteries.
  • Dispose of this Main Unit, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmen­tal pollution. in the garbage can at the end of use; please dispose it according to the local laws or contact with the manufac­turer for recycling.

Correct Disposal of This Product (Waste Electrical & Electronic Equipment)

  • This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life.
  • To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.
  • Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.
  • Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial waste for disposal.

Symbol Description

Service information

  • Name of after-sales service unit: Jiangxi AICARE Medical Technology Co., Ltd.
  • Contact information: +86-794-6577516
  • Address: (Building No.1, New Era Home Group, Le’ an County) South Side of South Ring Road, Le’ an County, Fuzhou City 344300, Jiangxi Province, China.

EMC Declaration
DECLARATION

  • All types of electronic equipment may cause electromag­netic interference to other equipment through the air or the cable connected with it. The term EMC (electromag­netic compatibility) refers to the ability of a device to be unaffected by electromagnetic interference from other devices and not to affect other devices through similar electromagnetic radiation.
  • To fully achieve the specified EMC performance, users should follow the detailed described in the service manual to properly install the product.

WARNING

  • Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.

WARNING

  • Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.

WARNING

  • Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the RF Generators, including cables specified by the manufacturer.
  • Otherwise, degradation of the performance of this equipment could result.

CAUTION

  • Please do not modify the equipment by yourself or any other unauthoritative organization. Unauthorized changes or modifications could void the user’s authority to operate the device.

CAUTION

  • Unless the device provided by our company, Devices which intrinsically transmit radio waves such as cellular phones, radio transceivers, mobile radio transmitters, radio-controlled toys, and so on, should preferably not be operated near the unit.

CAUTION

  • This device that is used may emit some radio frequency energy. This device may cause radio frequency interference to other medical, non-medical and radio communications’ device. To effectively prevent such interference, this product conforms to the radio frequency emission limit specified in IEC 60601-1-2 standard for class A of group 1. However, the company does not guarantee that there will be no interference in individual installation environments.

ESSENTIAL PERFORMANCE

AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-23 AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-24 AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-22

Recommended minimum separation distances

  • Nowadays, many RF wireless equipment’s have being used in various healthcare locations where medical equipment and/or systems are used. When they are used in close proximity to medical equipment and/or systems, the medical equipment and/or systems’ basic safety and essential performance may be affected. This RF Generators has been tested with the immunity test level in the below table and meet
    the related requirements of IEC 60601-1-2:2014. The customer and/or user should help keep a minimum distance between RF wireless communications equipment and this RF

Generators as recommended below.

AILE-X5-Electronic-Sphygmomano-Meters-IMAGE-27

Manufacture
[MANUFACTURER]

  • Jiangxi AICARE Medical Technology Co., Ltd.
  • (Building No.1, New Era Home Group, Le’ an County) South Side of South Ring Road, Le’ an County, Fuzhou City 344300, Jiangxi Province, China.

European Representative

  • EUROPEAN REPRESENTATIVE B=J Caretechion GmbH
    • Niederrheinstr. 71, 40474 Duesseldorf, Germany.