milwaukee pH54 Waterproof pH Tester User Manual
- June 14, 2024
- Milwaukee
Table of Contents
milwaukee pH54 Waterproof pH Tester
CONDITIONING
Remove the cap and immerse the pH54 in MA9015 storage solution for 2 hours to activate the electrode.
CALIBRATION
Unscrew the battery compartment cap on the top of the meter and perform one of
the following two-point calibration procedure, according to the required
application:
Acid calibration for measurements below pH7:
-
Immerse the electrode in pH7.01 (M10007) buffer solution.
-
Allow the reading to stabilize and with a small screwdriver turn the “7” calibration trimmer inside the battery compartment until the display shows “7.01”.
-
Rinse the electrode with tap water and immerse it in pH4.01 (M10004) buffer solution.
-
Allow the reading to stabilize and with a small screwdriver turn the “4” calibration trimmer until the display shows “4.01”.
Alkaline calibration for measurements above pH7:
Immerse the electrode in pH7.01 ( M10007 ) buffer solution.
- Allow the reading to stabilize and with a small screwdriver turn the “7” calibration trimmer until the display shows “7.01”.
- Rinse the electrode with tap water and immerse it in pH10.01 ( M10010 ) buffer solution.
- Allow the reading to stabilize and with a small screwdriver turn the “4” calibration trimmer until the display shows “10.01”.
Replace the battery compartment cap.
OPERATION
-
Remove the protective cap.
-
Turn the pH54 on by pressing the ON key.
-
Immerse the electrode in the solution to be tested.
-
Stir gently and wait for the reading to stabilize.
-
After use, rinse the electrode with water to minimize contamination.
-
Store the electrode with a few drops of storage (MA9015) or pH7 (M10007) solution in the protective cap.
-
Always replace the protective cap after use. Never use distilled or deionized water for storage purposes.
ELECTRODE REPLACEMENT
The electrode can be easily replaced in the following way:
-
Remove the protective cap.
-
Unscrew the plastic ring on the top of the electrode.
-
Pull out the MA73600 pH electrode and replace it with a new one (aligning the colored pin towards the LCD side).
-
y Make sure the gaskets are in place before screwing back the ring.
BATTERY REPLACEMENT
The pH54 is going to turn off when the batteries are to weak. To replace
the batteries unscrew the battery compartment cap and replace all three 1.5V
batteries while paying attention to their polarity. Make sure the gasket is in
place before screwing back the cap.
Batteries should only be replaced in a nonhazardous area using the battery
type specified in this instruction manual.
SPECIFICATIONS
Range | 0.0 to 14.0 pH |
---|---|
Range | 0.01 pH |
Resolution | ±0.1 pH |
Accuracy(@25°C) | Manual, at 2 points |
through trimmers
Calibration| MA73600
pH electrode| 0 to 50 °C; 100% RH
Environment| 3 x 1.5V, alkaline
Battery life| more than 1500 hours
of continuous use
Dimensions| more than 1500 hours
of continuous use
Weight| 80 g
ACCESSORIES
MA73600 | Replaceable pH electrode |
---|---|
M10004B | pH 4.01 buffer, 20 ml sachet, 25 pcs. |
M10007B | pH 7.01 buffer, 20 ml sachet, 25 pcs. |
M10010B | pH 10.01 buffer, 20 ml sachet, 25 pcs. |
MA9015 | Electrode storage solution, 230 ml |
MA9016 | Electrode cleaning solution, 230 ml |
M10000B | Electrode rinse solution, 20 ml sachet, 25 pcs. |
CERTIFICATION
Milwaukee Instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment. Do not treat this product as household waste. Hand it over to the appropriate collection point for the recycling of electrical and electronic equipment.
Disposal of waste batteries. This product contains batteries. Do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.
Please note: proper product and battery disposal prevents potential negative consequences for human health and the environment. For detailed information, contact your local household waste disposal service or go to www.milwaukeeinstruments.com (USA & CAN) or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any modification introduced by the user to the supplied equipment may compromise the meter’s performance. For your and the meter’s safety do not use or store the meter in hazardous environment. To avoid damage or burn, do not perform any measurement in microwave ovens.
WARRANTY
This instrument is warranted against defects in materials and manufacturing for a period of 2 years from the date of purchase. Electrode is warranted for 6 months. This warranty is limited to repair or free of charge replacement if the instrument cannot be repaired. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered by warranty. If service is required, contact your local Milwaukee Instruments Technical Service. If the repair is not covered by the warranty, you will be notified of the charges incurred. When shipping any meter, make sure it is properly packaged for complete protection.
Milwaukee Instruments reserves the right to make improvements in design, construction and appearance of its products without advance notice.
References
- Shop Milwaukee Instruments - the Official EU Site
- Shop Milwaukee Instruments - Official US Site
- Shop Milwaukee Instruments - the Official EU Site
- Shop Milwaukee Instruments - Official US Site
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>