BOSCAROL IMMOBILIZATION DEVICES
WIND AND OB HEAD IMMOBILIZER
OPERATING INSTRUCTIONS
Wind and OB Head Immobilizer
Medical devices compliant with the European Regulation on Medical Devices 745/2017
PRODUCED BY:
OSCAR BOSCAROL SRL
Enzo Ferrari 29
39100 Bolzano
Italy
Tel. +39 0471 932893
Fax: +39 02 57760140
info@boscarol.it
www.boscarol.it
Manufacturer and medical device information:
- Oscar Boscarol applies a quality management system (QMS) according to international ISO 13485 and ISO 9001 standards
- Medical devices of the “Immobilization” category (in all their configurations) comply with the European MDR Regulation 2017/745 and bear the CE marking in accordance with risk class I (see Annex VIII of the Regulation)
- The medical device meets the safety and performance requirements (GSPR) described in Annex I to the European Regulation 2017/745
About these operating instructions:
- This document contains information that is important for safe, effective and compliant use of your medical device
- Use the information below to train users and confirm their training
- It is not allowed to modify (even partially) this manual. Only the device manufacturer can make changes to it
- These instructions must always accompany the device. It is recommended to use the electronic version and make it available on PDAs, tablets and mobile phones of operators
These operating instructions apply to the following devices: IMM121640/ IMM121628/ IMM121631B
MEANING OF SYMBOLS AND PICTOGRAMS
0.1. Symbols used in these operating instructions to draw the reader’s attention
| Danger: Important safety information on the correct use of the head
immobilizer to prevent injury to the operator or patient and/or damage to the
device itself
---|---
| Warnings: Information that requires special attention
| Notes or information for proper use and to prevent damage to your device or
others. Activate the correct prevention measures
1.| List of actions to perform: follow them step by step
| These operating instructions
| Required maintenance service (contact the manufacturer and/or his
authorized service centers)
0.2. Symbols used on the device
| Use the head immobilizer only within the indicated temperature range. The
use of the head immobilizer outside these limits may decrease its functional
performance and damage it too
---|---
| Moisture-related usage limits
| Read these operating instructions carefully and completely
| Indicates the user’s need to consult these operating instructions for the
presence of information, such as warnings and precautions that cannot be
displayed on the medical device
| CE mark in accordance with the European MDR Regulation 2017/745 for medical
devices in class I (of risk)
| Manufacturer of the medical device
| Order number (device code)
| Please read the operating instructions in other languages available on the
website indicated near the symbol
| Indicates that the head immobilizer is a medical device
| Example of a UDI-DI and UDI-PI code of the medical device:
(01) Manufacturer and associated device identifier
(11) Date of production
(20) Product variant
(10) Lot number
INTENDED USE
Device name | Head Immobilizer Boscarol OB e WIND |
---|---|
Primary use | Medical device intended for head immobilization in lifting and |
transport devices
Other uses| There are no known uses other than that for which it was designed
Medical purpose| Immobilization and stabilization of the patient’s head before
transport. It can be used with the aid of a cervical collar
Application part in the human body| Patient’s head
Type of patients| Children and adults of both genders starting from 25 Kg
Application time on the same patient| Short-term use (maximum 30 days of
consecutive use)
Usage information| • Boscarol head immobilizer (OB model and WIND model) can
be used on all types of patients, excluding those under 25 kg in weight due to
head size which may be too small
• It can be used simultaneously with the cervical collar and other
immobilization and transport systems such as the scoop stretcher or spinal
stretcher
• Immobilization of a patient’s head with injuries of various kinds must be
carried out by professional rescuers, trained and aware of the specific
technologies of immobilization and transport of patients
• It is recommended to use the head immobilizer with at least two trained
rescuers
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION
| Read carefully
These operating instructions were prepared using simple and easy-to-understand
language. If you have difficulty interpreting these instructions, please
contact the manufacturer for further clarification.
Phone +39 0471 93 28 93
info@boscarol.it
---|---
**LATEX| Boscarol head immobilizers are made without the use of latex. The
materials used are latex-free, but it is not excluded that along the entire
production chain they may have come into contact with substances or traces of
latex.
Contaminated device| Warning in case of contamination of the device:
immobilization of the patient’s head on the head immobilizer may be a source
of contamination. For this reason, after each use, the device must be cleaned
and disinfected to eliminate any residual risk. Follow the instructions in
this user manual.
If you have any concerns or before sending a device for repair, contact
Boscarol Technical Service by emailing
info@boscarol.it** or calling +39 0471 932893
IMPORTANT INFORMATION TO KNOW BEFORE USE
Boscarol head immobilizer has been designed or tested to comply with the requirements imposed by the European Medical Devices Regulation 745/2017. WIND and OB head immobilizers are medical devices in class of risk I
| If the user or patient becomes aware of a user hazard, a side effect, an
accident caused by the device or a criticality (operational and constructive)
not dealt with in these instructions for use, he must immediately report it to
the manufacturer at the email address:
raq@boscarol.it
PERIODIC SAFETY INSPECTION| Preventive maintenance and periodic safety
inspection:
Boscarol head immobilizer must be checked periodically (full functional
verification is recommended at least once a week) especially on removable
parts and on Velcro attachments that allow the fixing of the side wedges, the
chin guards and fastening straps.
In the event of a failure, you must contact the manufacturer or replace them
asap
****LIFE SPAN| Boscarol head immobilizers have a lifespan of 5 years from
the date of purchase if stored and used in accordance with these operating
instructions
Responsibility operators/users| • Take due precautions in case of contact of
the device with the child/adult body and interpose special sterile sheets
(biocompatibility)
• The Boscarol head immobilizer is designed for the emergency health service
and must therefore always ready to use, at any time and in any situation
• Replace it immediately if there are obvious failures or failures to the
structure, fastening straps and welds (in the case of the WIND head
immobilizer). The head immobilizer must be stored in a place inaccessible to
children
• Do not leave the patient alone after immobilizing him. It must always be
properly assisted
• Dispose of the packaging in accordance with current regulations and make
sure it is out of reach of children. Straps are considered a possible toy that
is dangerous for children and for this reason they must never be left within
their reach
• Tampering, alterations and modifications of the device are not allowed
without the authorization of the manufacturer
• Operators must be trained and aware of the legal rules and provisions on
safety at work (use of PPE)
• Rescuers not properly trained to use the head immobilizer can create wounds
and damage to the patient as well as to themselves. The application of the
device always provides for maximum cooperation within the rescue team
CONTRAINDICATIONS (DO NOT USE FOR)
****Contraindications| • Direct contact of the patient’s skin (injured or
intact) with the device. It is mandatory to always place a sterile cloth to
ensure insulation from the materials of the device
• Do not apply the head immobilizer if the patient’s head is too small and the
immobilization operations are made complicated due to the size of the medical
device
---|---
SIDE EFFECTS (POSSIBLE DURING USE)
****Collateral effects| • Skin irritation and reactivity phenomena due to
the patient’s direct contact with PVC (WIND) or vinyl paint applied to the
surface of the OB device
• Nervous stress on conscious person that could cause sudden movements by the
patient during immobilization and after
---|---
FERMACAPO BOSCAROL
Boscarol head immobilizers are marketed “ready to use” and are complete with
all their functional parts. Upon receipt, problems relating to the use or lack
of certain arts provided must be verified and possibly contested. The head
immobilizers do not include accessories or spare parts.
Boscarol head immobilizers type available:
IMM121640 – WIND Head Immobilizer
IMM121628 – OB Head Immobilizer (yellow color version)
IMM121631B – Black OB Head Immobilizer (black color version)
COMPOSITION OF DEVICES
WIND HEAD IMMOBILIZER
The WIND head immobilizer type has been designed and built to be as compact as
possible when not used and stored in the emergency vehicles. It is made by of
a semi-rigid base covered of a layer of PVC (1 in the figure on the side) to
ensure functional stability and durability. The fastening straps are made of
polyethylene and Velcro.
The two chin guards (2 in the figure on the side) are soft and made in the
same way as the base in PVC. Two other Velcro straps are available for fixing
(see 3 in the picture on the side). The sewing of the belts to the chin guards
guarantees a high tensile strength. The two side wings are welded at the base
but can be raised to hold the head and secure it.
High-frequency welding of PVC guarantees stability and durability over time
even in case of prolonged use. The BOSCAROL WIND head immobilizer is radio-
transparent and translucent and ensuring radiological investigations without
having to be removed from the patient.
Inside the base of the device is inserted a recycled cardboard of appropriate
thickness, coated with a thin layer of foam, that guarantees structural
rigidity. The two chin guards are identical and are used to ensure that the
forehead and chin are good fastened to the device.
OB HEAD IMMOILIZER
Boscarol head immobilizer type “OB” (available in two versions that differ
only in color: yellow or black) while maintaining the functional
characteristics of the “WIND” model, it is realized with elastic expanded
polyurethane better known by the name of “foam rubber”. Thanks to the specific
density and the possibility of obtaining geometric pieces of defined shape it
is possible to make the base of the head immobilizer (number 1 in the photo
beside), the two side wedges with holes for access to the ears (details 2 in
the photo) and the chin guards (see 3 in the photo beside). The base is made
in the same way (see 1 in the photo on the right). The base also includes
eight slots in which straps are applied to fasten the head immobilizer to the
stretcher. The two straps (number 4 in the photo) are used as auxiliary straps
if needed.
How to use Boscarol head immobilizers
Before applying the device, it is good to immobilize the patient’s neck
through a cervical collar of an appropriate size. This allows to safely raise
the patient head enough to place it on the base of the device. Never improvise
rescuers if you have no experience, skill and expertise. Normally the number
of rescuers for these operations is two. Below are the operations necessary
for the immobilization of the head:
- Always warn a conscious patient about what we are doing and calm them down as much as possible
- One rescuer manually stabilizes the patient’s head and neck (to which the cervical collar had previously been applied correctly). These instructions do not deal with the application of the collar that refers to other manufacturers and user manuals
- The second rescuer applies a sterile sheet on the head immobilizer (both models) that will serve to isolate the head and the skin from the device itself and avoid any allergies or irritations. Then apply the device by slightly raising the patient’s head and insert the head immobilizer (applies to both the WIND and OB version)
- Make sure the base is centered relative to the patient’s head
- Now if you are using the WIND head immobilizer it is enough to raise the two wings and apply the two chin guards respectively to fix the chin and forehead. In the OB head immobilizer, it is enough instead to apply the two wedges on the Velcro of the base, taking care not to tighten the head excessively (the wedges must be adherent)
****BIOCOMPATIBILITY| Sterile sheet placed between the head immobilizer and the patient’s head is necessary to prevent skin irritation and reactivity due to the patient’s direct contact with PVC (WIND) or vinyl paint applied to the surface of the device (OB). It also establishes a thermal barrier in case the temperature of the device is very high or very low
---|--- - Always check the patient’s condition during all immobilization operations and during transport to the hospital or nearby rescue center. The photo on the side illustrates the correct use of the immobilizer WIND applied to a patient already equipped with a cervical collar
REUSE OF BOSCAROL HEAD IMMOBILIZER
After each use it is necessary to proceed with the cleaning and disinfection
of the medical device. These operations are of greater importance if the
patient’s pathological condition is not known and direct contamination maybe
present.
The user must always take protective measures and means to protect his or her
safety.
OB and WIND head immobilizers must be washed with water not exceeding 40°C of
temperature. Never use metal or very hard abrasive brushes that could ruin the
fabric and coating of PVC.
Pressure washer| Do not use high pressure washers that could damage the
surface layers of the head immobilizer and damage Velcro and polyethylene
flaps
---|---
Attention| Do not cut or change the structure and shape of Boscarol head
immobilizer so as not to damage it irreparably
Remove the sheet that was placed between the patient’s head and the device and discard it. After separated the wedges (by removing from the Velcro strap), the chin guards and the fasten belts of the OB head immobilizer (the WIND head immobilizer is composed of a single piece) clean and remove with the use of water all the substances present. If necessary, use a non-abrasive sponge to remove any fouling. Before proceeding with disinfection remove any traces of blood and/or organic left from the patient. Disinfect the head immobilizer (the entire surface) with substances suitable for that purpose (do a test on one side of the device to check does not damage). Do not use bleach and iron brushes, steel wool and blades of any kind to eliminate fouling. Colored disinfectants could irreparably stain the surfaces of the device itself. Before storing the device, make sure it is completely dry to avoid mold forming on the PVC.
Disinfection| Always comply with the provisions of local and regional
provisions about the correct disinfection operations
---|---
Sterilization| Boscarol OB and WIND head immobilizers cannot be
sterilized
DEMOLITION OF THE DEVICE
The device may be scrapped in accordance with national and local guidelines for the disposal of substances based on PVC, foam and polyethylene. The inner cardboard can be disposed of in the appropriate containers for recycling purposes. All materials comply with REACH and do not contain dangerous substances.
TECHNICAL SERVICE AND SPARE PARTS
In the event of a functional failure or if auxiliary information is required regarding reuse and/or storage and transport operations, please contact the manufacturer at +39 0471 932893 or by sending an email to info@boscarol.it.
TECHNICAL DATA AND COMPLIANCE FOR BOSCAROL HEAD IMMOBILIZERS
Classification of medical devices (in accordance with MDR 2017/745) | I |
---|---|
Basic UDI number (in conformity a MDR 2017/745) | 805240088IMMGK |
Technical specifications | Device produced on design with technical materials |
compliant with REACH Regulation
Degree of protection against the ingress of liquids and solids (IEC 529):|
IP65 in the absence of abrasion or breakage
Risk assessment (technical documentation)| ISO 14971:2019
Lifespan| 5 years from the date of manufacture
Dimension head immobilizer
Fermacapo Boscarol WIND| 40×30.5×16 (h) cm (with wings raised)| Device weight:
1.28 kg
---|---|---
Fermacapo Boscarol OB| 42.5×26.5×16 (h)| Device weight: 0.92 kg
Conditions of preservation and use
| Temperature range for transport, use and storage| -10 to 50° C (from -33.8
to 122° F)
---|---|---
| Humidity range for transport, use and storage| 5÷95 % U.R. not condensed
| Keep away from direct sunlight
| For further technical or storage information, please contact the
manufacturer ( info@boscarol.it ).
| On all sizes of the head immobilizer the tolerance is ±5 cm (due to the
couplings). Of the remaining measures, it is 5%.
Declaration of conformity
**| The declaration of conformity is kept by the manufacturer together with
all traceability data applicable to materials and production processes. You
can request it in copy from the manufacturer by sending an email to:
info@boscarol.it**
---|---
GUARANTEE
Oscar Boscarol guarantees the head immobilizer (all models) for a period of 1
year or from the date of purchase from the original distributor or
manufacturer. The company guarantees that the device is free of defective
materials and/or other defects due to production processes.
The warranty does not cover: normal wear of the device, discoloration and any
other cosmetic irregularities that do not affect the operation of the unit.
If, throughout the 1-year warranty period, the product is found defective, it
must be sent to Oscar Boscarol S.r.l. (Ltd) with a note describing the defect.
Oscar Boscarol S.r.l. (Ltd) repairs or replaces at its discretion defective
parts and/or the entire unit. All shipping costs are borne by the customer.
Warranty conditions:
To qualify for the warranty, the registration form in the product
documentation must be completed and returned by post, fax or e-mail at the
following address:
OSCAR BOSCAROL SRL V. E. Ferrari, 29 – 39100 BOLZANO, ITALY
Fax: +39 0257760142 – Email:
info@boscarol.it
To validate the warranty process, the customer must provide evidence of the following documentation:
- copy of the invoice and/or purchase receipt containing the lot number of the device and the date of purchase
- confirmation by the manufacturer or a person representing him that it is indeed a failure due to the production process or defective components from the time of their supply
- absence of tampering, modifications and/or something that does not conform to the original product
In terms of safety, reliability and function of the WIND and OB head immobilizer, Oscar Boscarol S.r.l. can only be held liable if:
- all planned technical operations, repairs, modifications and checks were carried out by Oscar Boscarol S.r.l. or an authorized service center
- the device has been and is being used correctly, strictly following the instructions given in these operating instructions
With reference to what is described in these warranty conditions, Oscar Boscarol S.r.l. cannot be held responsible for direct or indirect accidental damage, if modifications, repairs, unauthorized technical interventions or any of its parts have been damaged by accident and improper use. There are no other express or limited guarantees of marketability, suitability or other guarantees on the WIND and OB head immobilizer outside those described in this user manual.
Via Enzo Ferrari 29, 39100 Bolzano IT
Tel. +39 0471932893 – Fax: +39 0257760140
info@boscarol.it
www.boscarol.it
Printed in Italy by Oscar Boscarol Srl (Ltd)
ED01_REV03-2021 IFU HEAD IMMOBILIZER – ENG
Language of editing: English