HACH DR3900 Laboratory Spectrophotometer User Manual
- June 14, 2024
- HACH
Table of Contents
HACH DR3900 Laboratory Spectrophotometer
Specifications
Performance specifications
Operating mode| Transmittance (%), Absorbance and Concentration
Source lamp| Halogen lamp
Wavelength range| 320–1100 nm
Wavelength Accuracy| ± 1.5 nm (wavelength range 340–900 nm)
Wavelength reproducibility| ± 0.1 nm
Wavelength resolution| 1 nm
Wavelength calibration| Automatic
Wavelength selection| Automatic, based on method selection
Scanning speed| ³ 8 nm/s (in steps of 1 nm)
Spectral bandwidth| 5 nm
Photometric measuring range| ± 3.0 Abs (wavelength range 340–900
nm)
Photometric accuracy| 5 m Abs at 0.0–0.5 Abs
1% at 0.50–2.0 Abs
Photometric linearity| < 0.5% to 2 Abs
£1% at > 2 Abs with neutral glass at 546 nm
Stray light| < 0.1% T at 340 nm with NaNO2
Data log| 2000 readings (Result, Date, Time, Sample ID, User ID)
User programs| 100
Physical and environmental specifications
Width| 350 mm (13.78 in)
Height| 151 mm (5.94 in)
Depth| 255 mm (10.04 in)
Mass| 4200 g (9.26 lb)
Ambient operating requirements| 10–40 °C (50–104 °F), maximum
80% relative humidity (without condensate formation)
Ambient storage requirements| –40–60 °C (–40–140 °F), maximum
80% relative humidity (without condensate formation)
Additional technical data
Power connector via external power supply| Input:100–240 V ± 10
VAC/50–60 Hz
Output: 15 V/40 VA
Interfaces| Use only shielded cable with maximum length of 3 m: 2x USB
type A 1x USB type B
Use only shielded cable (e.g. STP, FTP, S/FTP) with maximum length of 20 m: 1x
Ethernet
Enclosure rating| IP40
Protection class| Class I
Altitude| 2000 m
Pollution degree| 2
Overvoltage category| II
Environmental conditions| Indoor use only
Power supply| External power supply
General Information
Safety information:
Please read the entire manual before unpacking, setting up, or operating this
equipment. Pay attention to all danger and caution statements. Failure to do
so could result in serious injury to the operator or damage to the equipment.
Precautionary labels: Read all labels and tags attached to the instrument.
Personal injury or damage to the instrument could occur if this information is
not observed.
RFID module (not available on all models): The DR 3900 contains an RFID module
to receive and transmit information and data. The RFID module operates with a
frequency of 13.56 MHz. Please contact the distributor for national conditions
of authorization regarding RFID technology.
Warning Labels
- DANGER : Indicates a potentially or imminently hazardous situation that, if not avoided, results in death or serious injury.
- WARNING : Indicates a potentially or imminently hazardous situation that, if not avoided, may result in death or serious injury.
- CAUTION : Indicates a potentially hazardous situation that may result in minor or moderate injury.
- NOTICE : Indicates a situation that, if not avoided, could result in damage to the instrument. Information that requires special emphasis.
Note: Information that supplements points in the main text.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage
to the instrument could occur if this information is not observed.
A symbol, if noted on the instrument, will be included with a danger or
caution statement in the operating instructions.
This symbol, if noted on the instrument, references the user manual for
operation and/or safety information.
Electrical equipment marked with this symbol may not be disposed of in
European domestic or public disposal systems after 12 August 2005. In
conformity with European local and national regulations
(EU Directive 2002/96/EC), European electrical equipment users must now return
old or end-of life equipment to the manufacturer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment manufacturer
or supplier for instructions on how to return end-of-life equipment,
manufacturer-supplied electrical accessories, and all auxiliary items for
correct disposal.
RFID module (not available on all models)
RFID technology is a radio application. Radio applications are subject to
national conditions of authorisation. In case of doubt, please contact the
distributor.
The DR 3900 contains an RFID module to receive and transmit information and
data. The RFID module operates with a frequency of 13.56 MHz.
WARNING
The spectrophotometer may not be used in dangerous environments.
The manufacturer and its suppliers reject any express or indirect guarantee
for the use with High Risk Activities.
Follow the following safety information, in addition to any local guidelines
in force.
Safety information for the correct use of the instrument:
- Do not operate the instrument in hospitals or comparable establishments in proximity to medical equipment, like pace makers or hearing aids.
- Do not operate the instrument near highly flammable substances, such as fuels, highly combustible chemicals and explosives.
- Do not operate the instrument near combustible gases, vapors or dust.
- Do not vibrate or jolt the instrument.
- The instrument can cause interference in immediate proximity to televisions, radios and computers.
- Do not open the instrument.
- Improper use of the instrument voids the warranty.
Certification (integrated RFID module)
FCC ID: YUH-QR15HL / YUH-Q152 IC : 9278A-QR15HL / 9278A-Q152
This device complies with Part 15 of the FCC rules and Industry Canada license
exempt RSS standard(s).
Operation is subject to the following two conditions:
- this device may not cause interference, and
- this device must accept any interference, including interference that may cause undesired operation of the device.
Changes or modifications to this equipment not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a Class
A digital device, pursuant to Part 15 of the FCC rules. These limits are
designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment
generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the
user will be required to correct the interference at their expense.
The following techniques can be used to reduce interference problems:
- Disconnect the equipment from its power source to verify that it is or is not the source of the interference.
- If the equipment is connected to the same outlet as the device experiencing interference, connect the equipment to a different outlet.
- Move the equipment away from the device receiving the interference.
- Reposition the receiving antenna for the device receiving the interference.
- Try combinations of the above.
Chemical and Biological Safety
DANGER
Potential danger in the event of contact with chemical/biological materials.
Handling chemical samples, standards and reagents can be dangerous.
Familiarise yourself with the necessary safety procedures and the correct
handling of the chemicals before the work and read and follow all relevant
safety data sheets.
Normal operation of this instrument may involve the use of hazardous chemicals
or biologically harmful samples.
- Observe all cautionary information printed on the original solution containers and safety data sheet prior to their use.
- Dispose of all consumed solutions in accordance with national regulations and laws.
- Select the type of protective equipment suitable to the concentration and quantity of the dangerous material at the respective work place.
Overview of product
The DR 3900 is a VIS spectrophotometer with a wavelength range of 320 to 1100
nm. The instrument is delivered with a complete series of application programs
and supports several languages.
The spectrophotometer DR 3900 contains the following programs and operating
types:
- Stored programs (pre-installed tests)
- Barcode Programs
- User Programs
- Favorites
- Single Wavelength
- Multi Wavelength
- Wavelength Scan
- Time course
The DR 3900 Spectrophotometer provides digital readouts in direct
concentration units, absorbance or percent transmittance.
When selecting a user-generated or programmed method, the menus and prompts
serve to direct the user through the test.
This menu system can also generate reports, statistical evaluations of
generated calibration curves and to report instrument diagnostic checks.
Installation
WARNING
Electrical and fire hazards.
Only use the supplied benchtop power supply LZV844.
Only qualified experts may perform the tasks described in this section of the
manual, while adhering to all locally valid safety regulations.
Unpack the instrument
The DR 3900 Spectrophotometer comes packaged with the following items:
- DR 3900 spectrophotometer
- Dust Cover
- USB dust cover, fitted as standard
- Benchtop power supply with power cord for US, GB
- Cell adapter A
- Light shield, fitted as standard in the DR 3900
- Operator RFID tag (not available on all models)
- Basic operating instructions DR 3900, operating instructions LINK2SC
Further information, detailed user manuals and documentation are available on
the website of the manufacturer.
Note: If any of these items are missing or damaged, please contact the
manufacturer or a sales representative immediately.
Operating environment
Note the following points so that the instrument will operate perfectly and
have a long service life.
- Place the instrument firmly on an even surface. Do not push any objects under the instrument.
- The ambient temperature must be 10–40 °C (50–104 °F).
- The relative humidity should be less than 80%; moisture should not condense on the instrument.
- Leave at least a 15 cm clearance at the top and on all sides for air circulation, to avoid overheating of electrical parts.
- Do not operate or store the instrument in extremely dusty, humid or wet locations.
- Keep the surface of the instrument, the cell compartment and all accessories clean and dry at all times. Immediately remove splashes or spilt materials on or in the instrument.
NOTICE
Protect the instrument from extreme temperatures from heaters, direct sunlight
and other heat sources.
Front and back view
Front view
1 | USB port type A | 4 | On/off switch |
---|---|---|---|
2 | Cell compartment cover | 5 | RFID module (not available on all |
models)
3| Touch screen
Back view
1 | Ethernet port | 3 | USB port type A |
---|---|---|---|
2 | USB port type B | 4 | Connection for benchtop power supply |
Cell compartments
1 ** Cell compartment for round cells (1)| 2 ** Cell compartment
for rectangular cells (2)
---|---
Power connections
WARNING
Electrical and fire hazards.
Only use the supplied benchtop power supply LZV844.
- Connect the power cable to the benchtop power supply.
- Plug the benchtop power supply cable into the back of the instrument.
- Insert the power cable plug into a mains socket (100–240 V~/ 50–60 Hz).
Turn on the power button next to the display to turn on the power supply.
Startup
Switch on the instrument, startup process
- Connect the power cable to the mains outlet.
- Switch on the instrument by using the power button next to the display.
- The instrument starts automatically with a startup process lasting approximately 45 seconds. The display shows the logo of the manufacturer. At the end of the startup process, a startup melody is heard.
Note: Wait approximately 20 seconds before switching on again so as not to damage the electronics and mechanics of the instrument.
Language selection
The DR 3900 software includes several language options. The first time the instrument is switched on, the language selection screen will be shown automatically after the startup process.
- Select the required language.
- Press OK to confirm the language selection. The self-check will then start automatically.
Change the language setting
The instrument functions in the selected language until the option is changed.
- Turn the instrument on.
- During the startup process, touch any point on the screen and maintain contact until the option for selecting a language is shown (approximately 45 seconds).
- Select the required language.
- Press OK to confirm the language selection. The self-check will then start automatically.
Self-check
Each time the instrument is powered up, a test program begins.
This procedure, which takes approximately two minutes, checks the system,
lamp, filter adjustment, wavelength calibration and voltage. Each test that
functions correctly is marked accordingly.
The Main Menu is displayed when diagnostics are completed.
Sleep mode
The instrument can be put into sleep mode.
-
Briefly press the power button next to the display.
The “Sleep mode” message is shown. The display will then switch off automatically. -
To switch on, press the power button next to the display.
The self-check will start automatically.
After that, the instrument is ready to use.
Power off the instrument
- Press the power button next to the display for approximately 5 seconds.
Standard programs
Tips for the use of the touch screen
The whole screen responds to touch. To choose an option, tap with a
fingernail, fingertip, an eraser or a specialised stylus. Do not press the
screen with a sharp object, such as the tip of a ballpoint pen.
- Do not place anything on top of the screen, to prevent damage or scratches on the screen.
- Press buttons, words or icons to select them.
- Use scroll bars to move up and down long lists very quickly. Press and hold the scroll bar, then move up or down to move through the list.
- Highlight an item from a list by pressing it once. When the item has been successfully selected, it will be displayed as reversed text (light text on a dark background).
Use of the alphanumeric keypad This display is used to enter letters,
numbers and symbols as needed when programming the instrument. Unavailable
options are disabled (grayed out). The icons on the right and left of the
screen are described in Table 1.
The central keypad changes to reflect the chosen entry mode. Press a key
repeatedly until the desired character appears on the screen. A space can be
entered by using the underscore on the YZ_ key.
Press Cancel to cancel an entry, or press OK to confirm an entry.
Note: It is also possible to use a USB keyboard (with US keyboard layout)
or a hand-held USB barcode scanner.
Alphanumeric keypad
Icon / key | Description | Function |
---|---|---|
ABC/abc | Alphabetic | Toggles the character input mode between upper and |
lower case.
# %| Symbols| Punctuation, symbols and numerical sub- and superscripts
may be entered.
123| Numeric| For entering regular numbers.
CE| Clear Entry| Clear the entry.
Left Arrow| Back| Deletes the current character and goes back one
position.
Right Arrow| Next| Navigates to the next space in an entry.
Main menu
A variety of modes may be selected from the Main Menu. The following table
briefly describes each menu option.
There is a toolbar on the right-hand side of the screen. Press to activate the
various functions.
Main Menu options
Option | Function |
---|---|
Stored Programs / Barcode Programs | Stored programs are pre- |
programmed methods that make use of HACH chemicals and vial tests.
Further information, as well as illustrated, step-by-step pro- cess
instructions for analyses using HACH programs, are available on the website of
the manufacturer.
User Programs| User programs make “made to measure analysis” possible:
- Users can program methods they have developed themselves
- Existing HACH methods can be stored as user programs.
Favorites| List of methods/tests created by the user to suit his own
requirements.
Option| Function
---|---
Single Wavelength| Single wavelength readings are:
Absorbance readings: The light absorbed by the sample is measured in
absorbance units.
Transmittance reading (%) : Measures the percent of the original light
that passes through the sample and reaches the detector.
Concentration readings: A concentration factor can be entered to enable
the measured absorbance values to be converted into concentration values.
Multi Wavelength| In the Multi Wavelength mode, absorbance (Abs) or
percentage transmittance (%T) is measured at up to four wavelengths and
absorbance differences and absorbance relationships are calculated. Simple
conversions into concentrations can also be carried out.
Wavelength Scan| A wavelength scan shows how the light from a sample
is absorbed over a defined wavelength spectrum. This function can be used to
determine the wavelength at which the maxi- mum absorbance value can be
measured. The absorbance behavior is displayed graphically during the scan.
Time course| The time scan records the absorbance or % transmittance at
a wavelength over a defined time.
System checks
| The “System check” menu offers a number of options, including optical
checks, output checks, lamp history, instrument update, service time, settings
for analytical quality assurance and instrument backup.
Recall measurement data| Stored data can be called up, filtered,
sent and deleted.
Instrument Setup| In this mode, user-specific or method-specific
settings can be entered: operator ID, sample ID, date & time, sound, PC &
printer, password, energy-saving mode and stored data.
Save, recall, send and delete data
The data log
The data log can store up to 2000 readings saved by the following programs:
- Stored Programs,
- Barcode Programs,
- User programs,
- Favorites,
- Single Wavelength and
- Multi Wavelength.
A complete record of the analysis is stored, including the Date, Time,
Results, Sample ID and Operator ID.
Recall stored data from the data log
-
Press Recall Data in the Main Menu.
-
Press Data Log.
A listing of the stored data is displayed. -
Press Filter: On/Off.
The function Filter Settings is used to search for specific items. -
Activate On. The data can now be filtered using the following selection criteria.
-
Sample ID
-
Operator ID
-
Start Date
-
Parameter or any combination of the four.
-
Press OK to confirm the selection.
The chosen items are listed. -
Press View Details to get more information.
Send data from the data log
NOTICE
Network and access point security is the responsibility of the customer that
uses the wireless instrument. The manufacturer will not be liable for any
damages, inclusive however not limited to indirect, special, consequential or
incidental dam-ages, that have been caused by a gap in, or breach of network
security.
Data is sent from the internal data log as an XML (Extensible Markup Language)
file or as a CSV (Comma Separated Value) file, to a directory with the name
DataLog on a USB mass storage device or a network drive. The file can then be
processed using a spreadsheet program. The file name has the format: DLYear-
Month-
Day_Hour_Minute_Second.csv or DLYear-Month-
Day_Hour_Minute_Second.xml.
-
Plug the USB storage device into the USB type A port on the DR 3900, or connect the DR 3900 to a network drive.
-
Press Recall Data from the Main Menu.
-
Select the data category to be transferred, e.g. Data Log. A list of the selected measurement data is displayed.
-
Press Options and then the PC & Printer icon.
-
Select the data to be sent.
The following options are available:- Single point: only this selected reading will be sent
- Filtered data: only readings that correspond to the set filters will be sent
- All data: all data in the selected data category will be sent.
-
Press OK to confirm. Note : The number in parentheses is the total number of data sets assigned to this selection.
Stored Programs
-
Press Stored Programs in the Main Menu to view an alphabetical list of stored programs with program numbers.
The “Stored Programs” list will appear. -
Highlight the required test.
Hinweis: Select the program by name or scroll through the list using the arrow keys. Highlight the program or press Select by No. to search for a specific program number. Press OK to confirm. -
Press Start to run the program. The respective measurement window is displayed.
Hinweis: All corresponding data (wavelength, factors and constants) is already preset. -
Follow the instructions for chemical procedures in the relevant procedures manual. Further information are available on the website of the manufacturer.
Hinweis: To display the procedural instructions in the display, press the info icon. Not all programs have this option.
Complete a barcode test
- Insert the light shield into cell compartment (2).
- Prepare the barcode test according to the process specification, and insert the cell in cell compartment (1).
- When a coded cell is placed in cell compartment (1) (Figure 3, page 10), the corresponding reading program is automatically activated in the Main Menu.
- Otherwise, press Barcode Programs in the Main Menu and insert a zero cell (depending on process specification) into cell compartment (1).
The reading is started automatically and the results are displayed.
To evaluate other cell tests and other parameters, insert the prepared cell
into the cell compartment and read off the result.
Note: The control bar displayed on the right of the screen shows the
relationship of the reading result to the measuring range. The black bar shows
the measured result independently of any dilution factor that was entered
. Select the chemical evaluation form
The chemical form of the test result of a number of parameters can be selected
individually.
-
In the result display, press the unit (e. g. mg/L) or the chemical representation of the evaluation form (e. g. PO43––P).
A list of available evaluation forms appears. -
Select the required form from the list. Press OK to confirm.
Another way of changing the standard setting is:
-
In the result display, press Options>More>Chemical Form.
A list of available evaluation forms appear. -
Select the required chemical form and press OK to confirm. Hinweis: The selected chemical form is displayed, but does not become the default. To change the default, refer to section Change of the default setting of the chemical form.
Change of the default setting of the chemical form
- Insert the zero cell or sample cell (corresponding to the process specification) into the cell compartment.
- In the result display, press Options>More>Chemical Form.
- A list of available evaluation forms appear. Select the new default setting.
- Press Save as Default.
The current result and all further readings will be displayed in the new chemical form.
Troubleshooting
Error displayed | Definition | Resolution |
---|
Attention!
Please insert the light shield.| Readings with barcode cuvettes
generally require the light shield.| Insert the light shield. Press OK.
Barcode label not read| Barcode faulty| Reinsert the cell.
If the barcode is not recognised, contact the manufacturer or his
representative.
An error occurred when uploading the instrument data.| | Start the
procedure again or contact the manufac- turer or his representa- tive.
An error occurred when reading from the USB memory
stick.| | Start the procedure again or contact the manufacturer or his
representative.
An error occurred when writing to the USB memory stick.|
| Start the procedure again or contact the manufacturer or his
representative.
Please check on the current update file.| Error during update.| Check
the USB memory stick.
Please contact Customer Services.| Error during update.|
Please check network configuration.| |
Please check the connection.| |
Please close the cover.| | Close the cover.
Error displayed| Definition| Resolution
---|---|---
Please insert the USB memory stick.| | Insert a USB memory stick into a
USB A port on the instrument.
Please check the connection and contact the administrator.| Network
setup or FTP error|
Blank value correction not possible!| Blank value correction not
possible with LCW919.|
File for instrument update missing.| Error during update.| Check the USB
memory stick.
File for instrument update is faulty.| Error during update.| Save the
update file again and repeat the procedure.
It’s recommended to execute a Full System Check| Check of the air values
failed| Switch the instrument of and then back on again. If the system check
is not successful, contact the manufacturer or his representative.
Entry invalid!| Password incorrect| Forgotten your password?
Contact the manufacturer or his representative.
Absorbance > 3.5!| The measured absorbance exceeds 3.5| Dilute sample and
measure again
Fault
Barcode check number?
Please update program data!| Deviation to the stored data| Data
updating
Error when calling up the local IP address.| Network setup: DHCP client
has no connection to the DHCP server| Enter the IP address again.
Error displayed| Definition| Resolution
---|---|---
Error during default gateway setup.| Network setup: default gateway
cannot be set for fixed IP address| Try to create the connection again.
Error during network drive setup!| Error during network setup| Check the
settings.
Error during subnet mask setup.| Network setup: Subnet mask cannot be
set for fixed IP address| Enter the subnet mask again.
Error copying from USB memory stick.| Error during
update| Start the procedure again or contact the manufacturer or his
representative.
Error in FTP connection.| FTP error| Make sure that the instrument is
connected to the network.
Fault Program not accessible.
Please update program data!| Barcode test not present| Data updating
Fault
Clean cuvette!| The cuvette is soiled or there are undissolved
particles in the cuvette| Clean the cuvette; allow the particles to settle
Fault
Test program stopped! Please check lamp Close the lid.
Error [xx]| Test program stops when the instrument is started| Check
the lamp and replace it if necessary.
Close lid.
Press Start Again.
Fault
Test program stopped! Please remove the cuvette
Close the lid.| Test program stops when the instrument is started|
Remove the cuvette/ sample cell from the cell compartment.
Press OK.
Error displayed| Definition| Resolution
---|---|---
Error
Selfcheck stopped. Hardware error.
Error [x]| Electronic defect| Contact the manufacturer or a sales
representative and indicate the error number
Error
Too much ambient light!
Move instrument into shade or close the lid| The instrument
sensors detects too much ambi- ent light.| Decrease ambient light. (Avoid
direct sunlight.)
Close lid.
No help function is available for this program.| |
Shelf life exceeded! Use chemicals?| | The analysis is possibly
erroneous. Use new chemicals
No evaluation!| Error in the test data- base / user database| Check
programming
Contact the manufacturer or a sales representative
No barcode!| No barcode found| Reinsert the cell.
If the barcode is not recognised, contact the manufacturer or his
representative.
No instrument backup present!| | Check the USB memory stick.
No valid data for these parameters!| Data analysis not possible, no
measurement data| Change the selection.
No valid data found!| View Data not possible in data log| Change the
selection.
No help function present.| |
Error displayed| Definition| Resolution
---|---|---
No measurement data present!| Data analysis settings cannot be
configured without measurement data.| Change the selection.
Control range not reached!| Data analysis limits not reached| This is a
warning notice. The control limit set was not reached.
Control range exceeded!| Data analysis limits exceeded.| This is
a warning notice. The control limit was exceeded.
Concentration too high!| Calculated concentration is higher than 999999|
Dilute sample and measure again
Over measuring range| The measured absorbance is above the calibration
range of the test| Dilute sample and measure again
Under measuring range| The measured absor- bance is below the cali-
bration range of the test| If possible, select a test with a lower reading
range or use a cuvette with a longer path length
Possible interference by:| Interference Check| The analysis is
possibly erroneous due to interferences.
Possible interference from:| Interference Check| The analysis is
possibly erroneous due to interferences.
Next service is due!| | Contact the manufacturer or his representative
for an inspection of the instrument.
Negative result!| The calculated result is negative| Check concentration
of sample
Network switched off.| Network setup off, when access to instruments
homepage via sidebar| Activate the online connection.
Error displayed| Definition| Resolution
---|---|---
Remote server cannot be reached.| Error during network setup| Make sure
that the instrument is connected to the network.
Unstable lighting conditions!| | Avoid direct sunlight at the
measuring location.
Insufficient memory for update .| Error during update.| Select a memory
with more space.
System check incorrect!| Measurement of air values failed| Switch the
instrument of and then back on again. If the system check is not successful,
contact the manufacturer or his representative.
Temperature too high. Measurement not possible!| | Switch of
the instrument and allow it to cool for a few minutes. If necessary, move it
to a cooler place.
Update file is faulty.| Error during update.| Save the update file again
and repeat the procedure.
USB memory stick is not connected.| Update not possible.| Check the USB
memory stick.
Web server cannot be reached.| Instruments homepage cannot be reached.|
Try the connection again later.
Replacement Parts
Description | Cat. No. |
---|---|
Halogen lamp | LPZ440.99.00007 |
Cell adapter A for 1-cm rectangular and 1-inch round cells | LZV846 |
Cell adapter B for 3-cm rectangular cells, China only | LZV847 |
Light shield | LZV849 |
Bench top power supply | LZV844 |
Power cord EU | YAA080 |
Power cord CH | XLH051 |
Power cord UK | XLH057 |
Power cord US | XLH055 |
Power cable China/Australia | XLH069 |
Dust Cover | LZV845 |
Cell compartment | LZV848 |
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>