FACEIL JPD-HA200 Electronic Blood Pressure Monitor User Manual

June 14, 2024
FACEIL

Electronic Blood Pressure MonitorFACEIL JPD-HA200 Electronic Blood
Pressure Monitor JPD-HA200 User Manual
Version: V2.2
Revision Date: 2022.07

Product Composition

The product is comprised of the host machine and the cuff of Electronic Blood Pressure Monitor.
Application Scope
The Electronic Blood Pressure Monitor is used to measure the diastolic pressure, systolic pressure and pulse rate of adults by oscillometric method for diagnostic reference.
Intended use
The Electronic Blood Pressure Monitor is intended to measure the systolicand dlastolic blood pressure as well as the pulse rate of adult person vla non- invaslveoscillometric technique at medical facilities or at home.

Safety Precautions

The warnings and illustrations shown in the User Manual enable you to use the product safely and correctly, thus preventing you and others from being injured, specifically as follows:
Legend, mark and meaning

| Waming mese
---|---
| Anti-clectric shock degree is Type BF of the application part
| When the product life expires and the end users discard the products, send them o the designated collecting and separating place for disposal according to the requirements from the local environmental protection authority.
| Consult the instructions for use.
| This product complies with the MDD93/42/EEC requirements.
| Degree of protection against the ingress of water.
| Information of manufacturer
| Date of manufacture
| Authorized European Representative

Quick Start Guide

Avoid smoking, cating, drinking caffeinated drinks or exercising for 30 minutes before taking measurement.

  1.  Sit upright in a chair with both feet on the floor.
  2. Remove tight fitting clothing from your upper arm along with any thick clothing.
  3. Pull on the end of the cuff until it wraps securely around your upper arm. Place your arm on a table so – that the cuff will be at the same level as your heart.
  4. The cuff will automatically inflate and the measurement will start, upon pressing the “ ” button.
  5. When the measurement is complete, the cuff will automatically deflate and your systolic and diastolic pressure values and pulse rate will be displayed.

FACEIL JPD-HA200 Electronic Blood Pressure Monitor -
Button

Overview

Monitor

FACEIL JPD-HA200 Electronic Blood Pressure Monitor -
USB

Features

  • Universal Cuff
  • Irregular Heartbeat Detector
  • Average of last 3 readings
  • 99 Readings* 2 Users
  • Auto-off Power Saving Feature
  • Operated by 4 AAA batteries (USB 5V optional)

Display

FACEIL JPD-HA200 Electronic Blood Pressure Monitor -
Display

Packing List

No. Name Quantity
I Electronic Blood Pressure Monitorifensiometre electronique I
2 Cuff I
3 Dry Battery ( AAA) 4
4 User Manual I
5 Easy-to-carry Storage Bag 1

Preparation

2.1 Battery Installation
a) Open the battery cover as shown in the — picture.
b) Place 4 AAA dry batterics. Pay attention to the battery electrode indication 2.2 Battery power indication and replacement
After the product is turned on, if low power symbol lappears on the screen , the measurement cannot be performed, and the battery must be replaced.
DO not use any expired battery;
If the product is not used for over 3 months, please take out the batteries.
2.3 USB Power Supply
USB line can be connected for power supply of the product without battery.
2.4 Cuff
The applicable arm circumference range of the cuff is 22-36cm or 22-42cm .FACEIL JPD-HA200 Electronic Blood Pressure Monitor -
Air Connection: Insert the air plug of cuff air tube into air jack of the Electronic Blood Pressure Monitor. Figure shown as above:

Function Setting Mode

3.1 User mode
In the shutdown state, press the “User button for 3 seconds, the screen displays “User 1 or “User 2 ”, and flickers continuously. Press the “Memory button to switch to the other user. (default user: User 1 ) FACEIL JPD-HA200
Electronic Blood Pressure Monitor - Button

(Note: In the shutdown state, press “User button to view the current time and the current user)
3.2 Year/Month/Date/Time setting
In the shutdown state, press the “User  ” ” button for 3 seconds to enter the setting mode. Press the “User & button to enter the year setting mode ( or automatically enter the year setting mode after the batteries are reloaded and the device is powered on ), the screen displays 20XX and flickers continuously. Press the “Memory button to increase the value which can be switched among 2019 and 2039. Press the “User button for confirmation and enter the month setting mode.
After entering the month setting mode, the screen will display xx-xx and – flash on the month. Press the “Memory button to increase the value which can be switched among 1 and 12, press the “User *” button for confirmation and enter the date setting mode. After entering the date setting mode, XX-XX is displayed on the screen and flickers on the date display. Press the “Memory ” ” button to increase the value which can be switched among 01 and 31, press the “User ” ” button for confirmation and enter the time seiting mode. After entering the time setting mode, the screen will display XX: XX and it will flicker continuously, press the “Memory ” ” button to increase the value which can be switched among 0 and 23, press the “User ” ” button for confirmation and enter the minute setting mode, press the “Memory ” ” button to increase the value which can be switched among 00 and 59, press the “User ” ”button for confirmation and enter the talking function setting. FACEIL JPD-HA200
Electronic Blood Pressure Monitor - Comfirm

Note:
When setting time values, continue pressing the “Memory”  button till the correct values are well set. Value ranges are year (2019~2039), month (01~ 12), date (01~31), hour (1~12 AM/PM) and minute (00~59).
When the ceiling value of a certain time sector is reached, the relevant value of the sector will start from the minimum individually.
3.3 Talking function setting (voice version only)
After entering talking function setting mode, & will show up at the “« ” SYS area, with the current voice status “ ” and voice icon (not displayed, when talking function mode is off) at the DIA arca. Press the “Memory button to switch the voice status, and press the “User button for confirmation and enter the unit display setting.![FACEIL JPD-HA200 Electronic Blood Pressure Monitor

  • Comfirm 1](https://manuals.plus/wp-content/uploads/2023/11/FACEIL-JPD-HA200 -Electronic-Blood-Pressure-Monitor-Comfirm-1.jpg) 3.4 Unit setting
    After entering the unit setting mode, the SYS area will display and unit area will display mmHg, it means the unit mmHg is selected (system default unit: mmHg), At this time, press of the “Memory button, the unit will be switched, it means the unit a is selected. After selecting the unit, press the “User button again to save the unit Setting .FACEIL JPD-HA200 Electronic Blood
Pressure Monitor - Comfirm 2

Correct Method of Use

4.1 How to use the cuff

  1. Place the cuff flat on the table, leave the hook & loop downwards, and pass the end of cuff through the metal ring to form a loop. The hook & loop shall point outwards.
  2. Pull the cuff through the upper arm to be measured, and wear the cuff correctly based on the downward icon ” ”, the air tube runs down the inside of your arm.
  3. Hook up the cuff on the upper arm according to the illustration, ensure that the lower edge of cuff is 2~3cm away from the elbow joint.
  4. Tighten the free edge of cuff to stick the hook & loop.
  5. The cuff should be wrapped on the upper arm comfortably, with tight space for two fingers. Before measurement, remove tight fitting clothing from your upper arm along with any thick clothing.
  6. Place the lower arm flat on the desktop, leaving the center of palmnaturally upwards, sitting upright, and ensuring the center of cuff and the heart are at the same level. Note that the tube of cuff cannot be folded or bent.

Note: If you cannot use the left arm for measurement, please use the right arm for measurement. All the measurements must be performed on the same arm for comparison. FACEIL JPD-HA200 Electronic Blood Pressure Monitor -
Cuff 1

4.2 Starting measurement

After you wear the cuff correctly, you can start the measurement:

  1. Press the “Power ‘ ” button, and the device will return to zero automatically, the air pump will start to inflate the cuff, and the screen will display the change of the pressure in the cuff.
  2. When reaching the stable pressure upon inflation, the air pump will stop the inflation, and the pressure in the cuff will be reduced gradually and displayed on the screen. If the inflated pressure is insufficient, the device will reinflate the cuff automatically for a higher pressure;
  3. When the pulse is measured, the screen will display the “heart” symbol and start flashing. The flashing “heart” symbol will be displayed on the screen.
  4. Upon measurement completion, the measured values of systolic pressure, diastolic pressure and pulse will be displayed on the screen.
  5. The screen will continue to display the measurement results, unless your long-press on the “Power’ ” button to turn off the device. If there is no operation, the device will be powered off automatically in 30s.

4.3 Using memory function
This blood pressure monitor is vailable for 2 users with 99 memories per user. Track the memory data by pressing “Memory button from the latest measurement, recorded as 1″, Then track “2″, “3” or more via the same operation. When there are over 3  measurement data, in power off state, press “Memory button for the average values of the latest 3 measurement data, recorded as “AVG 03”.
4.4 Clearing stored measurement data
In the shutdown state, Tong-press “Memory button and “User AL button for 3 seconds until “ALL” shows in the SYS area and “ ~ shows in the low DIA area, thus all the measurement data of the current user are deleted.
4.5 Classification standard for blood pressure condition
There is no definition of hypotension yet. Generally, if the systolic blood pressure is less than 100mmHg (13.3kPa), it is hypotension.FACEIL JPD-HA200
Electronic Blood Pressure Monitor - Diastolic

Contraindications, Precautions, Warnings and Prompt

(1) Warning:
Keep the device out of the reach of the children under 12 and people who can’t express their intention. When children of 12~18 use the device, they should be accompanied by the adults. Pregnant women shall use under the guidance of doctors.
Do not wrap the cufFover a wound, as this can cause further injury.
This model of Electronic Blood Pressure Monitor s suitable for the arm circumference range of 22-36cm or 22-42cm, and if the arm circumference exceeds this range, you might fail to obtain the correct measured value of blood pressure;
This model of Electronic Blood Pressure Monitor is not suitable for newborns or young children;
The blood pressure is constantly changing. You shall not judge the blood pressure condition with just one measurement result. The repeated measurement data over a period of time will be more reliable;
Do not make self-diagnosis according to the measurement results. Pleconsult your professional doctor with the measurement result record(s). The treatment based on the sclf-diagnosis of measurement results is very dangerous;
(2) Precautions:
Do not repair, disassemble or modify the Electronic Blood .Pressure Monitor without permission;
Do not collide or drop the host to avoid collision or strong impact on the device;
Do not mix old and new batteries of different brands for use.
Indoor temperature, environment, noise, user’s body position, speech or exercise might affect the blood pressure measurements. The body movement, magnetic field and improper wearing of cuff and sensor will affect the measurement as well.

Troubleshootin

  • If you have trouble using the device, please check the following instruction.
Problem Possible cause How to correct

No display when
press the Power
button| Low battery| Replace new batteries or use the USB for power supply.
The polarities of batteries are installed wrongly.| Install the batteries in correct polarities.
No pressurizing| The air plug is loosely installed.| Make sure the air plug is securely inserted in the main unit.
The air tube is broken or leaked.| Purchase a new cuff
“Err I ” displayed| The leakage is too fast or the pulse signal is too weak.| Please check the cuff, tie it up and try again.
” EIT 2″ displayed| The blood pressure signal cannot be
detected due to too much noises.| Please remove the noise sources and measure again.
“Err 3” displayed| The test result is abnormal.| Please measure again.
“Err p ” displayed| The inflation fails.| Please check the cuff, tie it up and try again.
“HI “displayed| The inflation pressure is greater than 3078 mmHg (41 kPa] kPa).| Please measure again.

Note: If your problem cannot be solved by the above, please contact customer service. Do not disassemble the device!

Storage & Maintenan

  • Keep the device away from direct sunlight, extreme temperatures, humidity or moisture.
  • Use adry, soft cloth to clean the device, or if desired, use a cloth lightly dampened with water.
  • Do not use corrosive cleaner, benzene, thinner or other volatile liquids to clean the device.
  • Do not wash or expose the arm cuff to liquid.
  • Remove batteries from the device when it will not be used for more than 3 months.

Disinfection:

  • Recommended disinfecting agent:
  • Is opropanol solution with 70% concentration
  • Medical alcohol with 75% concentration
Product Name Electronic Blood Pressure
Model JPD-HA200
Display Mode Digital display mode
Measuring Mode Oscillographic method
Measuring Body Part Upper arm Haut du bras
Measuring Range Pressure value Valeur de pression
Pulse value 40-199 pulse beats/min
Static measurement Accuracy Pressure value
Pulse value ± 5% of read value
LCD display Pressure
Pulse Pulse rate per minute, displaying three digits
--- ---
Storage capacity It can store 99 groups of measurement data for each of two

users
Power supply| 4 AAA dry batteries /DC 5V USB external power supply
Power Off Mode| Manual power-ofd Auto power-off
Device weight (without batteries)| About 195g
Monitor Size| 120mm (length) 90mm (width) 43mm (height)
Screen Size| 59mm (length) *47mm (width)
Cuff| Upper-arm-type cuff (measured arm circumference 22-36cm or 22-42cm)
---|---
Annexure| Cuff, User Manual, Dry Battery, Easy-to-carry Storage Bag
Battery Life| High-performance dry battery can be used for about 300 times at normal temperature
Service Life| 5 years|
Date of Production| See label|
Operating Environment| Temperature Condition| 5 ‘C -40 °C| If the device is stored and used in the environment
out of the designated
Humidity| 15%45 %RH

After-sale Service
The service life of this product is 1 years and we offer free maintenance service of I year.
Notes: For any failure caused by the user personal reasons or any damage resulted from the unauthorized disassembly, the free maintenance service is not provided.

http://www.faceil.net

E-mail: support@faceil.net
Website: www.faceil.net
Authorized European Representative:
MedPath GmbH
** Mies-van-der-Rohe-Strasse 8, 80807 Munich,
Germany
Private Label Manufacturer:
I LIKELINKS TRADING LIMITED
Room 32, 11 /F, Lijia industrial building, 8 Wufang street,
HONG KONG China999077
E-mail: support@faceil.net
Website: www.faceil.net