Palintest Kemio Disinfection Benchtop Kit Instructions

June 13, 2024
Palintest

Palintest Kemio Disinfection Benchtop Kit

Product Information

The Kemio Test Instructions provide detailed information on performing various tests using the Kemio multiparameter instrument. The instrument has three variants: Kemio Disinfection, Kemio Heavy Metals, and Kemio Multi.

  • Kemio Disinfection can test for chlorine, chlorine dioxide, chlorite, and peracetic acid.
  • Kemio Heavy Metals can test for lead, cadmium, and copper.
  • Kemio Multi can test for all of these parameters.

The instructions cover the methodology and care points for each test type.

Product Usage Instructions

General Guidelines

Performing a Test:

  1. On the home page, press “PERFORM TEST” or open the lid.
  2. Scan the sensor barcode using the barcode scanner. Hold the barcode about 12cm (5 inches) from the instrument. Align the red dot with the center of the barcode. When the scan is complete, the instrument proceeds to the next screen.
  3. Select the test type. This screen does not appear if only one test type is available.
  4. Select the correct batch code number from the list. This screen does not appear if the sensor was scanned.
  5. Select the label. This screen does not appear if only one label is defined for this test type or if tagging labels is disabled.
  6. Select or scan the user. This screen does not appear if only one user is defined or tagging users is disabled.
  7. If dilution is required for the test and programmed into the label, perform the dilution with the volumes on the screen.
  8. Follow the on-screen test instructions specific to this test type.

Free Chlorine, Total Chlorine, and ‘Free and Total Chlorine’ Tests

Palintest’s chlorine sensors can be used to measure free and total chlorine simultaneously. On Kemio, select either free chlorine or total chlorine to display a single result, or free & total to measure and display both parameters.

When using labels with the free and total test, labels set for both free chlorine and total chlorine are listed. The selected label limits apply only to the parameter they were set for in the configuration menu.

Kemio Test Instructions

Kemio is a multiparameter instrument with three variants.
Kemio Disinfection can test for chlorine, chlorine dioxide, chlorite and peracetic acid. Kemio Heavy Metals can test for lead, cadmium and copper. Kemio Multi can test for all of these.
These instructions explain the methodology and care points in detail for each of these test types.

General Guidelines
· Do not touch the sensor surface.
· Keep the instrument clean and dry during use.
· Keep the instrument on a level surface free from vibrations.
· Sensors perform within a specified temperature range. If the sample temperature is outside of the specified temperature range for the sensor, the measurement will proceed as normal but a note will be attached to the test result.
· For optimum performance we recomend sensors are stored at < 4° C (39° F) for over one year and < 20° C (68° F) for less than one year.

Performing a Test

5

Every test on Kemio begins with the same steps. These are used to enrich test results with application information, enable decision-making with pass/fail results, and ensure traceability.

1

On the home page, press

PERFORM TEST or open

the lid.

2

Scan the sensor barcode

using the barcode scanner.

Hold the barcode about 12cm

(5 inches) from the instrument.

Align the red dot with the

center of the barcode. When

the scan is complete, the

instrument proceeds to the

next screen.

6

Scanning Errors

`

If the barcode is out of date, a warning will flag. Choose to scan a different sensor or continue anyway. The warning flag will be attached to the test results.

If the barcode refers to a new batch that has not yet been entered into the system, the instrument will prompt the user to add that batch using the QR code, see p.20 of the Kemio Manual. After the batch is added, restart the test from the home screen.

If the barcode is from the wrong sensor type or is not a sensor barcode, or if there is a problem reading the barcode, the instrument prompts the user to use a suitable sensor.

The scanner times out after five seconds. If this happens, press TRY AGAIN to restart the scan or press SELECT to select the batch code from a list.

3

Select the test type.

This screen does not appear if only one test type is available. See p. 25 of the Kemio manual for details about selecting tests.

7
4
Select the correct batch code number from the list.
This screen does not appear if the sensor was scanned. See p. 30 of the Kemio manual for details about sensor batch codes.

5

Select the label.

This screen does not appear if only one label is defined for this test type or if tagging labels is disabled. See p.15 of the Kemio Manual for details about labels and limits.

6

Select or scan the user.

This screen does not appear if only one user is defined or tagging users is disabled. See p. 19 of the Kemio Manual for details about users.

8

7

If dilution is required for the

test and programmed into the

label, perform the dilution with

the volumes on the screen.

8

Follow the on-screen test

instructions below specific to

this test type.

Free Chlorine, Total Chlorine, and ‘Free and Total Cl’ Tests

9

Palintest’s chlorine sensors can be used to measure free and total chlorine simultaneously. On Kemio, select either free chlorine or total chlorine to display a single result, or “free & total” to measure and display both parameters.
When using labels with the free and total test, labels set for both free chlorine and total chlorine are listed. The selected label limits apply only to the parameter they were set for in the configuration menu.
Four different sensor types can be used to measure chlorine:
CLO sensors are used for lower concentrations of chlorine. · 0.02 mg/L – 10 mg/L of free chlorine · 0.02 mg/L – 75 mg/L of total chlorine
CHR sensors are used for higher concentrations of chlorine at higher temperatures · 0.1 mg/L – 25 mg/L of free chlorine (5 °C – 25 °C) · 1 mg/L – 500 mg/L of total chlorine (5 °C – 30 °C)
CLT sensors are used for higher concentrations of chlorine at lower temperatures (2 °C – 15 °C) · 0.1 mg/L – 25 mg/L of free chlorine · 1 mg/L – 500 mg/L of total chlorine
FHR sensors are used for high range free only chlorine · 0.1 mg/L – 100 mg/L of free chlorine (2 °C – 20 °C)

Care Points
If chlorine dioxide is present in solution, it may interfere with the chlorine reading. Use the degassing step from the chlorite test to produce a chlorine dioxide-free solution. Ensure the vessel is washed with water between each test.

10

Results

Free chlorine and total chlorine tests display one parameter each. The free and total test result screen displays both results. The data log stores one entry for each test, requiring twice as much space in memory.
The chlorine test methods display the concentration of free and total chlorine in units of mg/L or ppm Cl-.

Test Method 1

Lift the arm. If the arm is already open, the screen is skipped.

2

Tear each side of the packet,

then twist and tear the short

side from the packet, revealing

the sensor tracks. Do not touch

the dosed half of the sensor.

3

Hold the sensor through

11

the foil packet only, with

the white side facing up.

Slide it all the way in, until

the end stops against the

back of the slot.

If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.

4

Fill the right side of the vessel

to the line. If there is excess

sample, it will pour into the

left side of the vessel. Do not

overfill both sides or water

can splash onto the sensor

contacts and interfere with the

test.

5

Pull the foil off the sensor.

Close the arm. The test

will start automatically.

Wait one minute while the instrument performs the electrochemical measurement.

12

Total Bromine Test

Total Bromine is measured using Palintest’s BRS sensors for concentrations of bromine ranging from 0.2 mg/L – 10 mg/L across the temperature range of 5 – 30 °C.
Care Points
If other oxidants are present in solution, they will interfere with the bromine reading.

Results
The total bromine test method displays the concentration of bromine in units of mg/L or ppm bromine.

Test Method 1

13
Lift the arm. If the arm is already open, the screen is skipped.

2

Tear each side of the packet,

then twist and tear the short

side from the packet, revealing

the sensor tracks. Do not touch

the dosed half of the sensor.

3

Hold the sensor through

the foil packet only, with

the white side facing up.

Slide it all the way in, until

the end stops against the

back of the slot.

14
If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.

4

Fill the right side of the vessel

to the line. If there is excess

sample, it will pour into the

left side of the vessel. Do not

overfill both sides or water

can splash onto the sensor

contacts and interfere with the

test.

5

Pull the foil off the sensor.

Close the arm. The test

will start automatically.

Wait one minute while the instrument performs the electrochemical measurement.

15

Peracetic Acid (Peroxyacetic acid or PAA) Test

16

Peracetic acid (Peroxyacetic acid or PAA) is measured using Palintest’s PAA sensors for concentrations of peracetic acid ranging from 5 mg/L to 2000 mg/L.
Two sensor types can be used to measure peracetic acid:
PAA sensors are used for higher concentrations. · 10 mg/L – 2000 mg/L of peracetic acid
PAL sensors are used for lower concentrations. · 0.02 mg/L – 50 mg/L of peracetic acid

Care Points
Hydrogen peroxide is found in solution with peracetic acid. Palintest’s PAA and PAL sensors are not strongly affected by the presence of hydrogen peroxide.
Certain fungicides are known to interfere with the PAA sensor. Contact your Palintest representative when measuring water samples with fungicides.

Results
The peracetic acid test method displays the concentration of peracetic acid in units of mg/L or ppm PAA.

17

Test Method

1

Lift the arm. If the arm is

already open, the screen is

skipped.

2

Tear each side of the

packet, then twist and tear the

short side from the packet,

revealing the sensor tracks. Do

not touch the dosed half of the

sensor.

3

Hold the sensor through the

foil packet only, with the white

side facing up. Slide it all the

way in, until the end stops

against the back of the slot.

18

If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.

4

Fill the right side of the vessel

to the line. If there is excess

sample, it will pour into the

left side of the vessel. Do not

overfill both sides or water can

splash onto the sensor contacts

and interfere with the test.

5

Pull the foil off the sensor.

Close the arm. The test will

start automatically.

Wait one minute while the instrument performs the electrochemical measurement.

19

Chlorine Dioxide Test

20

Chlorine dioxide is measured using Palintest’s CDX sensors, for concentrations of chlorine dioxide ranging from 0.02 mg/L to 50 mg/L.

Care Points
· If there is chlorine present in the sample, use Glycine reagent. If there is no chlorine present, skip the glycine step on p. 16 · Chlorine dioxide is a volatile chemical that leaves a sample very easily. Perform a test quickly after taking a sample. Do not agitate the sample.
Results
The chlorine dioxide test method displays the concentration of chlorine dioxide in units of mg/L or ppm ClO2.

21

Test Method

1

Lift the arm. If the arm is

already open, the screen is

skipped.

2

Tear each side of the packet,

then twist and tear the short

side from the packet, revealing

the sensor tracks. Do not touch

the dosed half of the sensor.

3

Hold the sensor through the foil

packet only, with the white side

facing up. Slide it all the way in,

until the end stops against the

back of the slot.

If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.

22

4

Fill the right side of the vessel

to the line. If there is excess

sample, it will pour into the

left side of the vessel. Do not

overfill both sides or water

can splash onto the sensor

contacts and interfere with

the test.

5

If chlorine or copper is present

in the solution, add three drops

of glycine reagent to prevent

them from interfering with the

measurement. Use the stir rod

to mix the sample. If no chlorine

or copper are present, skip

this step.

6

Pull the foil off the sensor. Close

the arm. The test will start

automatically.

Wait one minute while the instrument performs the electrochemical measurement.

214337

Chlorite Test

24

The chlorite test method involves three separate measurements of various chemical combinations in the sample, followed by some calculations to determine the total chlorite concentration. The test procedure must be followed carefully to ensure the chemical reactions and measurements occur correctly.
A chlorine dioxide measurement forms a part of the chlorite test. However, in the chlorite test type, only labels and limits for chlorite can be selected.
Care Points
· The chlorite test uses three CDX sensors. These must all be from the same batch of sensors. If sensors from different batches are used, it may produce inaccurate results. · If there is chlorine present in the sample, use glycine reagent during the first CDX measurement. If there is no chlorine present, skip the glycine step on p. 20. · Chlorine dioxide is a volatile chemical that leaves a sample very easily. Perform the first test quickly after taking a sample for an accurate measurement. Do not agitate the sample. · If the chlorite test method is interrupted, or if an error occurs partway through the test procedure, the entire test must restart from the beginning to ensure accurate results. · Use separate vessels for the first CDX test before the degassing step, and the second and third tests with CR-1 and CR-2. Do not mix up the two vessels. · Do not let the kit stay wet. Trapped humidity could damage the kit components.

25

Results

The chlorite results screen displays two readings: the calculated chlorite result and the chlorine dioxide result measured by the first sensor.
The chlorite test method displays the concentration of chlorine dioxide in units of mg/L or ppm ClO2, and the concentration of chlorite in units of mg/L or ppm ClO2-.

Test Method 1

Lift the arm. If the arm is already open, the screen is skipped.

2

Tear each side of the packet,

then twist and tear the short

side from the packet, revealing

the sensor tracks. Do not touch

the dosed half of the sensor.

26

3

Hold the sensor

through the foil packet

only,with the white side

facing up. Slide it all the

way in, until the end stops

against the back of the slot.

If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.

4

Fill the right side of the vessel to

the line. If there is excess

sample, it will pour into the left

side of the vessel. Do not overfill

both sides or water can splash

onto the sensor contacts and

interfere with the test.

5

If chlorine or copper is present

in the solution, add three drops

of Glycine reagent to prevent

them from interfering with the

measurement. Use the stir rod to

mix the sample. If no chlorine or

copper are present, skip this step.

27
6
7
8

Pull the foil off the sensor. Close the arm. The test will start automatically.
Wait one minute while the instrument performs the electrochemical measurement.
During the one minute electrochemical measurement period, set up the degasser as shown. Ensure the two plastic halves of the degasser are properly connected. Ensure the filter is installed correctly.
Fill the degasser vessel up to the mark with fresh sample.

28

9

Press the start button

on the pump to begin

degassing the chlorine dioxide

from the sample.

It is not a problem to run the degasser for longer than the timer displayed on the screen.

10

Press the START TIMER

button to begin the timer. The

displayed time is calculated

based on the concentration of

chlorine dioxide measured in the

previous step.

11

While the degasser is running,

prepare for the next

sensor test. Open the lid,

remove the sensor, and discard

the old sample.

At the end of the degassing period, perform a test on the degassed sample.

29
12
13 14

Tear each side of the second CDX sensor packet, then twist and tear the short side from the packet, revealing the sensor tracks. Do not touch the dosed half of the sensor.
Hold the sensor through the foil packet only, with the white side facing up. Slide it all the way in, until the end stops against the back of the slot.
Fill the right side of the vessel to the line with degassed water.
Add three drops of CR-1 reagent.

30

15

Pull the foil off the

sensor. Close the arm.

The test will start automatically.

Wait one minute while the instrument performs the electrochemical measurement.

16

Open the arm, discard the

sensor. KEEP the sample from

the previous test.

17

Add three drops of CR-2

reagent.

31
18
19
20

Close the arm with no sensor inserted.
Wait two minutes while chlorite reacts with the CR-2 reagents. At the end of the two minute period perform the final test.
Lift the arm. If the arm is already open, the screen is skipped.

32

21

Open the foil packet.

22

Hold the final CDX sensor

through the foil packet only, with

the white side facing up. Slide it

all the way in, until the end stops

against the back of the slot.

23

Pull the foil off the sensor. Close

the arm. The test will start

automatically.

Wait one minute while the instrument performs the electrochemical measurement.

33

Metals Tests

34

(Lead, Cadmium, Copper, Arsenic)

The following metal sensors are available for use with Kemio Heavy Metals and Kemio Multi.

Metal Lead Cadmium Copper Arsenic

Sensor Type MPB MCD MCU MAS

Range 1 – 100 µg/L (or ppb) 0.5 – 10 µg/L (or ppb) 0.1 – 2.0 mg/L (or ppm) 2.0 – 250 µg/L (or ppb)

Care Points
· Lead, Cadmium and Copper sensors contain a very small amount of mercury. Check your local laws for disposal. In most regions there are no special requirements. In some regions the capped disposable vessel must be disposed of per regional laws.
· Avoid touching the chemicals on the sensor surface. · Avoid cross- contamination between samples ensuring all
equipment used is clean. · Ensure the Soluprep tablets are completely dissolved before
starting the test.
Results
The lead, cadmium and copper results screens display a single reading, in units of µg/L or ppb Pb, Cd or Cu. Separate Arsenic tests are available for either Total Arsenic or Arsenic (III). Selecting, Speciated Arsenic will provide results for Total Arsenic, Arsenic (III) and Arsenic (V) in units of µg/L or ppb As.

35
1

Test Method for Lead, Cadmium and Copper
Lift the arm. If the arm is already open, the screen is skipped.

2

Tear each side of the packet,

then twist and tear the short

side from the packet, revealing

the sensor tracks. Do not touch

the dosed half of the sensor.

3

Hold the sensor through the foil

packet only, with the white side

facing up. Slide it all the way in,

until the end stops against the

back of the slot.

4

Fill the L-shaped vessel to

the line. This is so that the

temperature probe is in

contact with the sample for

temperature compensation.

36

5

Fill the disposable vessel

to the shoulder. Do not

overfill, or water can splash

onto the sensor contacts and

interfere with the test.

6

Add one Soluprep SP-A tablet

into the disposable vessel.

Crush and stir using the

crushing rod until the tablet is

fully dissolved.

7

Insert both vessels into the

instrument.

8

Remove the foil and close the

arm to start the test.

Depending upon the test, Kemio will take around 3-5 minutes to perform the measurement and then display the result.

37
Test Methods for Arsenic
Total Arsenic or Arsenic(III) can be selected as standalone tests. Arsenic(V) can then be calculated as follows. Arsenic (V) = Total Arsenic ­ Arsenic (III)
Speciated Arsenic. Kemio will guide through the complete procedure for a Total Arsenic test followed by an Arsenic (III) test. The instrument will then use these results to also calculate and display the Arsenic (V).
Test Method for Total Arsenic
1
Lift the arm. If the arm is already open, the screen is skipped.

38
2
Fill syringe with sample to the 10 mL line

10 ml

3

Attach the ion exchange cartridge

10 ml

to the syringe.

(Two cartridges are provided in a pack of one hundred sensors. Therefore, it is recommended that the cartridge is used up to 50 times before replacing with a new one.)

4
6 ml
4 ml

1x DROP/ SECOND
6 ml

Flush out the cartridge by expressing 4 mL of sample into a waste container. Then, express the remaining 6 mL into the disposable sample holder. IMPORTANT: Use a gentle pressure, so that droplets of sample emerge slowly (around 1 per second).

39
5
6
7

Add TWO Soluprep SP-T tablets. Crush and stir thoroughly.
Fill the L-Shaped vessel with further unfiltered sample. This will enable Kemio to measure the temperature and compensate the result.
Place both the L-shaped vessel and the disposable sample vessel in the instrument.

40

8

Take a Kemio MAS sensor and tear

the foil on each side. Then, twist,

tear and remove the short end of

the foil from the sensor to reveal

the sensor contacts. Do not touch

the active surfaces of the sensor.

Ensure the sensor is pushed into

the remaining foil packaging to keep

the active surfaces protected.

9
Holding the sensor by the foil, with the white side up, slide it fully into the instrument.

10
Remove the foil from the sensor and gently close the arm. The instrument will take about two minutes to perform the measurement and then display the concentration of Total Arsenic.

41
Test method for Arsenic (III) 1
Lift the arm. If the arm is already open, the screen is skipped.
2
Fill syringe with sample to the 10 mL line

10 ml

3

Attach the ion exchange cartridge

to the syringe.

10 ml

(Two cartridges are provided in

each pack of one hundred sensors.

Therefore, it is recommended that

the cartridge is used up to 50 times

before replacing with a new one.)

4
6 ml
4 ml
5

42

1x DROP/ SECOND
6 ml

Flush out the cartridge by expressing 4 mL of sample into a waste container. Then, express the remaining 6 mL into the disposable sample holder. IMPORTANT: Use a gentle pressure, so that droplets of sample emerge slowly (around 1 per second).

Add a Soluprep SP-A tablet. Crush and stir thoroughly.

6
Fill the L-Shaped vessel with further unfiltered sample. This will enable Kemio to measure temperature and compensate the result.
7
Place both the L-shaped vessel and the disposable sample vessel in the instrument.

43
8
9 10

Take a Kemio MAS sensor and tear the foil on each side. Then, twist, tear and remove short end of the foil from the sensor to reveal the sensor contacts. Do not touch the active surfaces of the sensor. Ensure the sensor is pushed into the remaining foil packaging to keep the active surfaces protected.
Holding the sensor by the foil, with the white side up, slide it fully in to the instrument.
Remove the foil from the sensor and close the arm. The instrument will take about 5 minutes to perform the measurement and then display the concentration of Total Arsenic.

Arsenic Testing ­ Critical Care Points

44

4
3
2
1
If the sample is turbid (cloudy) then it will need to be filtered to avoid blocking of the ion exchange cartridge. This can be done simply by inserting a particulate filter to the syringe in front of the ion exchange cartridge.

Samples that have been

acidified must be neutralized

before analysis. Acceptable

pH

range is pH 4 – 7.

Ensure that sample is within

10°C

35°C the temperature range

10°C – 35°C.

Always store the ion exchange cartridge in the tube it is supplied in.

45

Darkening of the inlet end of the ion exchange cartridge is normal with some types of water. If this color extends the entire length of the filter then it should be discarded.
Speciated Arsenic requires two sensors to be used during the test procedure. For valid results, these two sensors must have the same batch number.

Check and adhere to all local regulations regarding the disposal of the sensors, ion exchange cartridge and sample water.

46
1

Results
At the end of the countdown, the test results will display on screen. The number of results displayed depends on the test type.
Disinfection test results are displayed in mg/L or ppm. Metals tests results are displayed in µg/L or ppb.

Dilution
If a label with a dilution factor was used during the test, the factor will have applied to the number on screen. For example, if a 50mg/L solution was diluted 5X and measured by Kemio using a label with a 5X dilution factor, the result displayed on the screen will be 50mg/L.
It is critical that the correct dilution is performed when a label with a dilution factor is used.

47
No Limits
If no label or a label with no limits was used during testing, the page title will be DETAILS on a blue background.

Limits
If a label with limits was used, either a green PASS or a red FAIL will be displayed at the top of the screen. This is simple traffic light system enables a quick and clear decision by the operator.

2

Press DETAILS to view the

complete information for the

test.

If a warning or flag was raised before or during testing, a note will be added to the test record. The result will still be displayed. A reading that falls within the test limits will still be a pass. However, the note will indicate in the record that the test was not normal.

48
1

Details
The first line displays the test type and the measured concentration (after dilution if applicable).
Up to three test types and concentrations can be displayed in this way.

Label indicates which label was selected at the start of the test. If no label was selected, DEFAULT is displayed.
Limits indicates the minimum and maximum acceptable concentrations set by the label that was selected. If a limit is not set on the label or if no label was selected, “-” is displayed.
Temperature indicates the temperature of the sample in the vessel. Note: this may differ from the sample temperature.
User indicates which user was selected at the start of the test. If no user was selected, DEFAULT is displayed.

49
Time indicates the time when the test was performed. Date indicates the date on which the test was performed. Batch Code indicates the batch used for the test. Dilution factor indicates the multiplier on the measured result to produce the displayed result. Default is 1.
Press ADD REFERENCE to add information from a barcode (or the information can be typed in manually) to enrich the data record. See p. 9 in the instruction manual for details on producing and using barcodes.
Press ADD NOTES to type information about the test using the on-screen keyboard. Maximum of 48 characters. After a failed test, it is recommended to write an explanatory note or explain the corrective actions taken, as the information will be stored alongside the test in memory.
Error messages are stored in the note field automatically.

50

Final Steps

After the test results are displayed, the user must complete the test by removing the sensor, disposing of the sample and cleaning the instrument. This protects the instrument, prevents microbial growth, prevents spills, and prepares the instrument for the next test. Kemio will prompt the user to complete these actions.
After viewing the results, the user has the choice to:
A Finish the test and clean for the instrument B Return to the home screen C Perform another test D Shut down the instrument

A. Finish The Test (Disinfection Test)

1

Open the lid.

51

2

Remove the sensor.

3
Discard the sample.

4

If Kemio is wet, wipe it down

with a cloth according to

the care and maintenance

suggestions. Take special

care of the electrical

contacts. These can be

accessed by opening the

front lid of the arm.

52
5

Close the lid. Kemio will return to the results screen.

Kemio does not detect whether the sensor and sample have been removed – only that the lid has been opened to do so. Take care of the instrument and ensure they are properly removed.
A. Finish The Test (Metals Test)
1
Open the front lid to drop the sensor into the disposable vessel.

53

2

Open the lid.

3

Remove both vessels from

Kemio

4

Place a cap on the

disposable vessel. If required

by local regulations, dispose

of the sample, sensor and

vessel in a secure manner.

54
5

Discard the sample from the L-Shaped vessel

Kemio does not detect whether the sensor and sample have been removed – only that the lid has been opened to do so. Take care of the instrument and ensure they are properly removed.
B. Return To The Home Screen
If the “Finish the test” instructions have not yet been completed, Kemio will prompt the user to to do these. After the lid is closed, Kemio will return to the home screen.
If the lid has already been opened, Kemio will proceed directly to the home screen.

55
C. Perform Another Test
Press REPEAT TEST.
When retesting, Kemio reuses the same label, user and batch information from the previous test.
If testing a sample that should use a different label, if a different user is testing or if the next sensor comes from a new batch, DO NOT press REPEAT TEST. This could lead to an inaccurate result or incorrect information. Instead, press the HOME button and PERFORM TEST.
D. Shut Down The Instrument
Press the power button.
If the user presses and holds the power button before opening the lid, Kemio will display a warning message reminding the user to discard the sensor and sample, and then turn off.
If the lid has already been opened, the regular “Goodbye’ screen will display and Kemio will turn off.

Sensor Performance

Sensor Test type Range type

CLO

Free Chlorine
Total Chlorine

0.02-10 (mg/L)
0.05-75 (mg/L)

CHR

Free

0.1 – 25

Chlorine (mg/L)

Temp Resolution Range (mg/L) (°C)

Precision (mg/L)

2 – 35 0.01 up to 5.0 0.05 @ 1.0 mg/L, 20°C 0.1 up to 10

2 – 35 0.01 up to 5.0 0.5 @ 10 mg/L, 20°C 0.1 up to 20 1 up to 75

5 – 25 0.01 up to 5.0 2.0 @ 20 mg/L, 10°C 0.1 up to 25

Total

1 – 500

Chlorine (mg/L)

5 – 30 0.1 up to 20 1 up to 500

10.0 @ 200 mg/L, 10°C

CLT

Free

0.1 – 25 2 – 15 0.01 up to 5.0 2.0 @ 20 mg/L, 5°C

Chlorine (mg/L)

0.1 up to 25

Total

1 – 500

Chlorine (mg/L)

2 – 15 0.1 up to 20 1 up to 500

10.0 @ 200 mg/L, 5°C

FHR CDX

Free Chlorine
Chlorine Dioxide
Chlorite

0.1 – 100 2 – 20 0.02 up to 10 2.25 @ 30 mg/L, 2°C

(mg/L)

0.1 up to 100 5.63 @ 75 mg/L, 14°C

0.02 – 50 5 – 40 0.01 up to 10.0 0.05 @ 1.0 mg/L, 20°C

(mg/L)

0.1 up to 50

0.02 – 50 5 – 40 0.01 up to 2.5 0.05 @ 1.0 mg/L, 20°C

(mg/L)

0.1 up to 20

1.0 up to 50

Sensor Performance Cont.

57

Sensor Test type Range Temp Resolution

type

Range (mg/L)

(°C)

Precision (mg/L)

PAA

Peracetic 5 –

Acid

2000

(mg/L)

5 – 30 0.1 up to 20.0 1 up to 500 5 up to 2000

1.25 @ 25 mg/L, 10°C 10 @ 200 mg/L, 20°C 75 @ 1500 mg/L, 20°C

PAL Peracetic 0.02 – 10 2 – 30 0.01 up to 5

Acid

(mg/L)

0.1 up to 20

0.04 @ 0.2 mg/L, 20°C 0.05 @ 1.0 mg/L, 5°C

PAH

Peracetic 400 – 10 – 60 1 up to 1000

Acid

4000

10 up to 4000

(mg/L)

125 @ 2500 mg/L, 50°C 175 @ 3500 mg/L, 50°C

MPB Lead

1 – 100 5 – 35 1 up to 100 µg/L 1.0 µg/L @ 5 ug/L, 20°C (µg/L)

MCD Cadmium 0.5 – 10 5 – 35 0.1 up to 10.0

(µg/L)

µg/L

1.0 µg/L @ 5 ug/L, 20°C

MCU

Copper

0.1-2.0 (mg/L)

5 – 35 0.001mg/L up to 0.1mg/L @ 0.5mg/L, 20°C 0.6mg/L
0.005mg/L up to 2.0 mg/L

BRS

Bromine 0.2 – 10 5 – 30 0.01 up to 1 mg/L 20% up to 1 mg/L

(mg/L)

0.1 up to 10 mg/L 10% 1 – 10 mg/L

MAS Arsenic 2 -250 10 – 35 0.1 µg/L (µg/L)

20% up to 5 µg/L 10% 5 – 250 µg/L

58

Reorder Codes

Description Chlorine Free High Range Sensors (FHR) (100 pack) Chlorine Free High Range Sensors (FHR) (500 pack) Chlorine Standard Range Sensors (CLO) (100 pack) Chlorine Standard Range Sensors (CLO) (500 pack) Bromine Sensors (BRS) 100 Pack Chlorine High Range Sensors (CHR) (100 pack) Chlorine High Range Sensors (CHR) (500 pack) Chlorine High Range Low Temp Sensors (CLT) (100 pack) Chlorine High Range Low Temp Sensors (CLT) (500 pack Chlorine Dioxide and Chlorite Sensors (CDX) (100 pack) Chlorine Dioxide Sensors and Chlorite (CDX) (500 pack) Glycine Reagent CR-1 Reagent CR-2 Reagent Peracetic Acid Sensors (PAA) (100 pack) Peracetic Acid Sensors (PAA) (500 pack) Peracetic Acid Sensors Low Range (PAL) (100 pack) Peracetic Acid Sensors Low Range (PAL) (500 pack) Peracetic Acid Sensors High Range (PAL) (100 pack) Peracetic Acid Sensors High Range (PAL) (500 pack) Lead Sensors (MPB) (10 pack) Cadmium Sensors (MCD) (10 pack) Total Arsenic Sensors (100 pack) Speciated Arsenic (III & V) Sensors (50 Tests) Copper Sensors (MCU) (10 pack)

Product Code KEM21FHR KEM25FHR KEM21CLO KEM25CLO KEM21BRS KEM21CHR KEM25CHR KEM21CLT KEM25CLT KEM21CDX KEM25CDX PT549 PT546 PT547 KEM21PAA KEM25PAA KEM21PAL KEM25PAL KEM25PAH KEM25PAH KEM22MPB KEM22MCD KEM100-TAS KEM100M-SAS KEM22MCU

5599
www.palintest.com
KEM TI V2 05/23

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals