Palintest Kemio Disinfection Benchtop Kit Instructions
- June 13, 2024
- Palintest
Table of Contents
Palintest Kemio Disinfection Benchtop Kit
Product Information
The Kemio Test Instructions provide detailed information on performing various tests using the Kemio multiparameter instrument. The instrument has three variants: Kemio Disinfection, Kemio Heavy Metals, and Kemio Multi.
- Kemio Disinfection can test for chlorine, chlorine dioxide, chlorite, and peracetic acid.
- Kemio Heavy Metals can test for lead, cadmium, and copper.
- Kemio Multi can test for all of these parameters.
The instructions cover the methodology and care points for each test type.
Product Usage Instructions
General Guidelines
Performing a Test:
- On the home page, press “PERFORM TEST” or open the lid.
- Scan the sensor barcode using the barcode scanner. Hold the barcode about 12cm (5 inches) from the instrument. Align the red dot with the center of the barcode. When the scan is complete, the instrument proceeds to the next screen.
- Select the test type. This screen does not appear if only one test type is available.
- Select the correct batch code number from the list. This screen does not appear if the sensor was scanned.
- Select the label. This screen does not appear if only one label is defined for this test type or if tagging labels is disabled.
- Select or scan the user. This screen does not appear if only one user is defined or tagging users is disabled.
- If dilution is required for the test and programmed into the label, perform the dilution with the volumes on the screen.
- Follow the on-screen test instructions specific to this test type.
Free Chlorine, Total Chlorine, and ‘Free and Total Chlorine’ Tests
Palintest’s chlorine sensors can be used to measure free and total chlorine simultaneously. On Kemio, select either free chlorine or total chlorine to display a single result, or free & total to measure and display both parameters.
When using labels with the free and total test, labels set for both free chlorine and total chlorine are listed. The selected label limits apply only to the parameter they were set for in the configuration menu.
Kemio Test Instructions
Kemio is a multiparameter instrument with three variants.
Kemio Disinfection can test for chlorine, chlorine dioxide, chlorite and
peracetic acid. Kemio Heavy Metals can test for lead, cadmium and copper.
Kemio Multi can test for all of these.
These instructions explain the methodology and care points in detail for each
of these test types.
General Guidelines
· Do not touch the sensor surface.
· Keep the instrument clean and dry during use.
· Keep the instrument on a level surface free from vibrations.
· Sensors perform within a specified temperature range. If the sample
temperature is outside of the specified temperature range for the sensor, the
measurement will proceed as normal but a note will be attached to the test
result.
· For optimum performance we recomend sensors are stored at < 4° C (39° F) for
over one year and < 20° C (68° F) for less than one year.
Performing a Test
5
Every test on Kemio begins with the same steps. These are used to enrich test results with application information, enable decision-making with pass/fail results, and ensure traceability.
1
On the home page, press
PERFORM TEST or open
the lid.
2
Scan the sensor barcode
using the barcode scanner.
Hold the barcode about 12cm
(5 inches) from the instrument.
Align the red dot with the
center of the barcode. When
the scan is complete, the
instrument proceeds to the
next screen.
6
Scanning Errors
`
If the barcode is out of date, a warning will flag. Choose to scan a different sensor or continue anyway. The warning flag will be attached to the test results.
If the barcode refers to a new batch that has not yet been entered into the system, the instrument will prompt the user to add that batch using the QR code, see p.20 of the Kemio Manual. After the batch is added, restart the test from the home screen.
If the barcode is from the wrong sensor type or is not a sensor barcode, or if there is a problem reading the barcode, the instrument prompts the user to use a suitable sensor.
The scanner times out after five seconds. If this happens, press TRY AGAIN to restart the scan or press SELECT to select the batch code from a list.
3
Select the test type.
This screen does not appear if only one test type is available. See p. 25 of the Kemio manual for details about selecting tests.
7
4
Select the correct batch code number from the list.
This screen does not appear if the sensor was scanned. See p. 30 of the Kemio
manual for details about sensor batch codes.
5
Select the label.
This screen does not appear if only one label is defined for this test type or if tagging labels is disabled. See p.15 of the Kemio Manual for details about labels and limits.
6
Select or scan the user.
This screen does not appear if only one user is defined or tagging users is disabled. See p. 19 of the Kemio Manual for details about users.
8
7
If dilution is required for the
test and programmed into the
label, perform the dilution with
the volumes on the screen.
8
Follow the on-screen test
instructions below specific to
this test type.
Free Chlorine, Total Chlorine, and ‘Free and Total Cl’ Tests
9
Palintest’s chlorine sensors can be used to measure free and total chlorine
simultaneously. On Kemio, select either free chlorine or total chlorine to
display a single result, or “free & total” to measure and display both
parameters.
When using labels with the free and total test, labels set for both free
chlorine and total chlorine are listed. The selected label limits apply only
to the parameter they were set for in the configuration menu.
Four different sensor types can be used to measure chlorine:
CLO sensors are used for lower concentrations of chlorine. · 0.02 mg/L – 10
mg/L of free chlorine · 0.02 mg/L – 75 mg/L of total chlorine
CHR sensors are used for higher concentrations of chlorine at higher
temperatures · 0.1 mg/L – 25 mg/L of free chlorine (5 °C – 25 °C) · 1 mg/L –
500 mg/L of total chlorine (5 °C – 30 °C)
CLT sensors are used for higher concentrations of chlorine at lower
temperatures (2 °C – 15 °C) · 0.1 mg/L – 25 mg/L of free chlorine · 1 mg/L –
500 mg/L of total chlorine
FHR sensors are used for high range free only chlorine · 0.1 mg/L – 100 mg/L
of free chlorine (2 °C – 20 °C)
Care Points
If chlorine dioxide is present in solution, it may interfere with the chlorine
reading. Use the degassing step from the chlorite test to produce a chlorine
dioxide-free solution. Ensure the vessel is washed with water between each
test.
10
Results
Free chlorine and total chlorine tests display one parameter each. The free
and total test result screen displays both results. The data log stores one
entry for each test, requiring twice as much space in memory.
The chlorine test methods display the concentration of free and total chlorine
in units of mg/L or ppm Cl-.
Test Method 1
Lift the arm. If the arm is already open, the screen is skipped.
2
Tear each side of the packet,
then twist and tear the short
side from the packet, revealing
the sensor tracks. Do not touch
the dosed half of the sensor.
3
Hold the sensor through
11
the foil packet only, with
the white side facing up.
Slide it all the way in, until
the end stops against the
back of the slot.
If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.
4
Fill the right side of the vessel
to the line. If there is excess
sample, it will pour into the
left side of the vessel. Do not
overfill both sides or water
can splash onto the sensor
contacts and interfere with the
test.
5
Pull the foil off the sensor.
Close the arm. The test
will start automatically.
Wait one minute while the instrument performs the electrochemical measurement.
12
Total Bromine Test
Total Bromine is measured using Palintest’s BRS sensors for concentrations of
bromine ranging from 0.2 mg/L – 10 mg/L across the temperature range of 5 – 30
°C.
Care Points
If other oxidants are present in solution, they will interfere with the
bromine reading.
Results
The total bromine test method displays the concentration of bromine in units
of mg/L or ppm bromine.
Test Method 1
13
Lift the arm. If the arm is already open, the screen is skipped.
2
Tear each side of the packet,
then twist and tear the short
side from the packet, revealing
the sensor tracks. Do not touch
the dosed half of the sensor.
3
Hold the sensor through
the foil packet only, with
the white side facing up.
Slide it all the way in, until
the end stops against the
back of the slot.
14
If the sensor is not fully inserted, push it in all the way using the edges of
the sensor only. Do not touch the dosed chemicals on the electrode surface.
4
Fill the right side of the vessel
to the line. If there is excess
sample, it will pour into the
left side of the vessel. Do not
overfill both sides or water
can splash onto the sensor
contacts and interfere with the
test.
5
Pull the foil off the sensor.
Close the arm. The test
will start automatically.
Wait one minute while the instrument performs the electrochemical measurement.
15
Peracetic Acid (Peroxyacetic acid or PAA) Test
16
Peracetic acid (Peroxyacetic acid or PAA) is measured using Palintest’s PAA
sensors for concentrations of peracetic acid ranging from 5 mg/L to 2000 mg/L.
Two sensor types can be used to measure peracetic acid:
PAA sensors are used for higher concentrations. · 10 mg/L – 2000 mg/L of
peracetic acid
PAL sensors are used for lower concentrations. · 0.02 mg/L – 50 mg/L of
peracetic acid
Care Points
Hydrogen peroxide is found in solution with peracetic acid. Palintest’s PAA
and PAL sensors are not strongly affected by the presence of hydrogen
peroxide.
Certain fungicides are known to interfere with the PAA sensor. Contact your
Palintest representative when measuring water samples with fungicides.
Results
The peracetic acid test method displays the concentration of peracetic acid in
units of mg/L or ppm PAA.
17
Test Method
1
Lift the arm. If the arm is
already open, the screen is
skipped.
2
Tear each side of the
packet, then twist and tear the
short side from the packet,
revealing the sensor tracks. Do
not touch the dosed half of the
sensor.
3
Hold the sensor through the
foil packet only, with the white
side facing up. Slide it all the
way in, until the end stops
against the back of the slot.
18
If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.
4
Fill the right side of the vessel
to the line. If there is excess
sample, it will pour into the
left side of the vessel. Do not
overfill both sides or water can
splash onto the sensor contacts
and interfere with the test.
5
Pull the foil off the sensor.
Close the arm. The test will
start automatically.
Wait one minute while the instrument performs the electrochemical measurement.
19
Chlorine Dioxide Test
20
Chlorine dioxide is measured using Palintest’s CDX sensors, for concentrations of chlorine dioxide ranging from 0.02 mg/L to 50 mg/L.
Care Points
· If there is chlorine present in the sample, use Glycine reagent. If there is
no chlorine present, skip the glycine step on p. 16 · Chlorine dioxide is a
volatile chemical that leaves a sample very easily. Perform a test quickly
after taking a sample. Do not agitate the sample.
Results
The chlorine dioxide test method displays the concentration of chlorine
dioxide in units of mg/L or ppm ClO2.
21
Test Method
1
Lift the arm. If the arm is
already open, the screen is
skipped.
2
Tear each side of the packet,
then twist and tear the short
side from the packet, revealing
the sensor tracks. Do not touch
the dosed half of the sensor.
3
Hold the sensor through the foil
packet only, with the white side
facing up. Slide it all the way in,
until the end stops against the
back of the slot.
If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.
22
4
Fill the right side of the vessel
to the line. If there is excess
sample, it will pour into the
left side of the vessel. Do not
overfill both sides or water
can splash onto the sensor
contacts and interfere with
the test.
5
If chlorine or copper is present
in the solution, add three drops
of glycine reagent to prevent
them from interfering with the
measurement. Use the stir rod
to mix the sample. If no chlorine
or copper are present, skip
this step.
6
Pull the foil off the sensor. Close
the arm. The test will start
automatically.
Wait one minute while the instrument performs the electrochemical measurement.
214337
Chlorite Test
24
The chlorite test method involves three separate measurements of various
chemical combinations in the sample, followed by some calculations to
determine the total chlorite concentration. The test procedure must be
followed carefully to ensure the chemical reactions and measurements occur
correctly.
A chlorine dioxide measurement forms a part of the chlorite test. However, in
the chlorite test type, only labels and limits for chlorite can be selected.
Care Points
· The chlorite test uses three CDX sensors. These must all be from the same
batch of sensors. If sensors from different batches are used, it may produce
inaccurate results. · If there is chlorine present in the sample, use glycine
reagent during the first CDX measurement. If there is no chlorine present,
skip the glycine step on p. 20. · Chlorine dioxide is a volatile chemical that
leaves a sample very easily. Perform the first test quickly after taking a
sample for an accurate measurement. Do not agitate the sample. · If the
chlorite test method is interrupted, or if an error occurs partway through the
test procedure, the entire test must restart from the beginning to ensure
accurate results. · Use separate vessels for the first CDX test before the
degassing step, and the second and third tests with CR-1 and CR-2. Do not mix
up the two vessels. · Do not let the kit stay wet. Trapped humidity could
damage the kit components.
25
Results
The chlorite results screen displays two readings: the calculated chlorite
result and the chlorine dioxide result measured by the first sensor.
The chlorite test method displays the concentration of chlorine dioxide in
units of mg/L or ppm ClO2, and the concentration of chlorite in units of mg/L
or ppm ClO2-.
Test Method 1
Lift the arm. If the arm is already open, the screen is skipped.
2
Tear each side of the packet,
then twist and tear the short
side from the packet, revealing
the sensor tracks. Do not touch
the dosed half of the sensor.
26
3
Hold the sensor
through the foil packet
only,with the white side
facing up. Slide it all the
way in, until the end stops
against the back of the slot.
If the sensor is not fully inserted, push it in all the way using the edges of the sensor only. Do not touch the dosed chemicals on the electrode surface.
4
Fill the right side of the vessel to
the line. If there is excess
sample, it will pour into the left
side of the vessel. Do not overfill
both sides or water can splash
onto the sensor contacts and
interfere with the test.
5
If chlorine or copper is present
in the solution, add three drops
of Glycine reagent to prevent
them from interfering with the
measurement. Use the stir rod to
mix the sample. If no chlorine or
copper are present, skip this step.
27
6
7
8
Pull the foil off the sensor. Close the arm. The test will start
automatically.
Wait one minute while the instrument performs the electrochemical measurement.
During the one minute electrochemical measurement period, set up the degasser
as shown. Ensure the two plastic halves of the degasser are properly
connected. Ensure the filter is installed correctly.
Fill the degasser vessel up to the mark with fresh sample.
28
9
Press the start button
on the pump to begin
degassing the chlorine dioxide
from the sample.
It is not a problem to run the degasser for longer than the timer displayed on the screen.
10
Press the START TIMER
button to begin the timer. The
displayed time is calculated
based on the concentration of
chlorine dioxide measured in the
previous step.
11
While the degasser is running,
prepare for the next
sensor test. Open the lid,
remove the sensor, and discard
the old sample.
At the end of the degassing period, perform a test on the degassed sample.
29
12
13 14
Tear each side of the second CDX sensor packet, then twist and tear the short
side from the packet, revealing the sensor tracks. Do not touch the dosed half
of the sensor.
Hold the sensor through the foil packet only, with the white side facing up.
Slide it all the way in, until the end stops against the back of the slot.
Fill the right side of the vessel to the line with degassed water.
Add three drops of CR-1 reagent.
30
15
Pull the foil off the
sensor. Close the arm.
The test will start automatically.
Wait one minute while the instrument performs the electrochemical measurement.
16
Open the arm, discard the
sensor. KEEP the sample from
the previous test.
17
Add three drops of CR-2
reagent.
31
18
19
20
Close the arm with no sensor inserted.
Wait two minutes while chlorite reacts with the CR-2 reagents. At the end of
the two minute period perform the final test.
Lift the arm. If the arm is already open, the screen is skipped.
32
21
Open the foil packet.
22
Hold the final CDX sensor
through the foil packet only, with
the white side facing up. Slide it
all the way in, until the end stops
against the back of the slot.
23
Pull the foil off the sensor. Close
the arm. The test will start
automatically.
Wait one minute while the instrument performs the electrochemical measurement.
33
Metals Tests
34
(Lead, Cadmium, Copper, Arsenic)
The following metal sensors are available for use with Kemio Heavy Metals and Kemio Multi.
Metal Lead Cadmium Copper Arsenic
Sensor Type MPB MCD MCU MAS
Range 1 – 100 µg/L (or ppb) 0.5 – 10 µg/L (or ppb) 0.1 – 2.0 mg/L (or ppm) 2.0 – 250 µg/L (or ppb)
Care Points
· Lead, Cadmium and Copper sensors contain a very small amount of mercury.
Check your local laws for disposal. In most regions there are no special
requirements. In some regions the capped disposable vessel must be disposed of
per regional laws.
· Avoid touching the chemicals on the sensor surface. · Avoid cross-
contamination between samples ensuring all
equipment used is clean. · Ensure the Soluprep tablets are completely
dissolved before
starting the test.
Results
The lead, cadmium and copper results screens display a single reading, in
units of µg/L or ppb Pb, Cd or Cu. Separate Arsenic tests are available for
either Total Arsenic or Arsenic (III). Selecting, Speciated Arsenic will
provide results for Total Arsenic, Arsenic (III) and Arsenic (V) in units of
µg/L or ppb As.
35
1
Test Method for Lead, Cadmium and Copper
Lift the arm. If the arm is already open, the screen is skipped.
2
Tear each side of the packet,
then twist and tear the short
side from the packet, revealing
the sensor tracks. Do not touch
the dosed half of the sensor.
3
Hold the sensor through the foil
packet only, with the white side
facing up. Slide it all the way in,
until the end stops against the
back of the slot.
4
Fill the L-shaped vessel to
the line. This is so that the
temperature probe is in
contact with the sample for
temperature compensation.
36
5
Fill the disposable vessel
to the shoulder. Do not
overfill, or water can splash
onto the sensor contacts and
interfere with the test.
6
Add one Soluprep SP-A tablet
into the disposable vessel.
Crush and stir using the
crushing rod until the tablet is
fully dissolved.
7
Insert both vessels into the
instrument.
8
Remove the foil and close the
arm to start the test.
Depending upon the test, Kemio will take around 3-5 minutes to perform the measurement and then display the result.
37
Test Methods for Arsenic
Total Arsenic or Arsenic(III) can be selected as standalone tests. Arsenic(V)
can then be calculated as follows. Arsenic (V) = Total Arsenic Arsenic (III)
Speciated Arsenic. Kemio will guide through the complete procedure for a Total
Arsenic test followed by an Arsenic (III) test. The instrument will then use
these results to also calculate and display the Arsenic (V).
Test Method for Total Arsenic
1
Lift the arm. If the arm is already open, the screen is skipped.
38
2
Fill syringe with sample to the 10 mL line
10 ml
3
Attach the ion exchange cartridge
10 ml
to the syringe.
(Two cartridges are provided in a pack of one hundred sensors. Therefore, it is recommended that the cartridge is used up to 50 times before replacing with a new one.)
4
6 ml
4 ml
1x DROP/ SECOND
6 ml
Flush out the cartridge by expressing 4 mL of sample into a waste container. Then, express the remaining 6 mL into the disposable sample holder. IMPORTANT: Use a gentle pressure, so that droplets of sample emerge slowly (around 1 per second).
39
5
6
7
Add TWO Soluprep SP-T tablets. Crush and stir thoroughly.
Fill the L-Shaped vessel with further unfiltered sample. This will enable
Kemio to measure the temperature and compensate the result.
Place both the L-shaped vessel and the disposable sample vessel in the
instrument.
40
8
Take a Kemio MAS sensor and tear
the foil on each side. Then, twist,
tear and remove the short end of
the foil from the sensor to reveal
the sensor contacts. Do not touch
the active surfaces of the sensor.
Ensure the sensor is pushed into
the remaining foil packaging to keep
the active surfaces protected.
9
Holding the sensor by the foil, with the white side up, slide it fully into
the instrument.
10
Remove the foil from the sensor and gently close the arm. The instrument will
take about two minutes to perform the measurement and then display the
concentration of Total Arsenic.
41
Test method for Arsenic (III) 1
Lift the arm. If the arm is already open, the screen is skipped.
2
Fill syringe with sample to the 10 mL line
10 ml
3
Attach the ion exchange cartridge
to the syringe.
10 ml
(Two cartridges are provided in
each pack of one hundred sensors.
Therefore, it is recommended that
the cartridge is used up to 50 times
before replacing with a new one.)
4
6 ml
4 ml
5
42
1x DROP/ SECOND
6 ml
Flush out the cartridge by expressing 4 mL of sample into a waste container. Then, express the remaining 6 mL into the disposable sample holder. IMPORTANT: Use a gentle pressure, so that droplets of sample emerge slowly (around 1 per second).
Add a Soluprep SP-A tablet. Crush and stir thoroughly.
6
Fill the L-Shaped vessel with further unfiltered sample. This will enable
Kemio to measure temperature and compensate the result.
7
Place both the L-shaped vessel and the disposable sample vessel in the
instrument.
43
8
9 10
Take a Kemio MAS sensor and tear the foil on each side. Then, twist, tear and
remove short end of the foil from the sensor to reveal the sensor contacts. Do
not touch the active surfaces of the sensor. Ensure the sensor is pushed into
the remaining foil packaging to keep the active surfaces protected.
Holding the sensor by the foil, with the white side up, slide it fully in to
the instrument.
Remove the foil from the sensor and close the arm. The instrument will take
about 5 minutes to perform the measurement and then display the concentration
of Total Arsenic.
Arsenic Testing Critical Care Points
44
4
3
2
1
If the sample is turbid (cloudy) then it will need to be filtered to avoid
blocking of the ion exchange cartridge. This can be done simply by inserting a
particulate filter to the syringe in front of the ion exchange cartridge.
Samples that have been
acidified must be neutralized
before analysis. Acceptable
pH
range is pH 4 – 7.
Ensure that sample is within
10°C
35°C the temperature range
10°C – 35°C.
Always store the ion exchange cartridge in the tube it is supplied in.
45
Darkening of the inlet end of the ion exchange cartridge is normal with some
types of water. If this color extends the entire length of the filter then it
should be discarded.
Speciated Arsenic requires two sensors to be used during the test procedure.
For valid results, these two sensors must have the same batch number.
Check and adhere to all local regulations regarding the disposal of the sensors, ion exchange cartridge and sample water.
46
1
Results
At the end of the countdown, the test results will display on screen. The
number of results displayed depends on the test type.
Disinfection test results are displayed in mg/L or ppm. Metals tests results
are displayed in µg/L or ppb.
Dilution
If a label with a dilution factor was used during the test, the factor will
have applied to the number on screen. For example, if a 50mg/L solution was
diluted 5X and measured by Kemio using a label with a 5X dilution factor, the
result displayed on the screen will be 50mg/L.
It is critical that the correct dilution is performed when a label with a
dilution factor is used.
47
No Limits
If no label or a label with no limits was used during testing, the page title
will be DETAILS on a blue background.
Limits
If a label with limits was used, either a green PASS or a red FAIL will be
displayed at the top of the screen. This is simple traffic light system
enables a quick and clear decision by the operator.
2
Press DETAILS to view the
complete information for the
test.
If a warning or flag was raised before or during testing, a note will be added to the test record. The result will still be displayed. A reading that falls within the test limits will still be a pass. However, the note will indicate in the record that the test was not normal.
48
1
Details
The first line displays the test type and the measured concentration (after
dilution if applicable).
Up to three test types and concentrations can be displayed in this way.
Label indicates which label was selected at the start of the test. If no label
was selected, DEFAULT is displayed.
Limits indicates the minimum and maximum acceptable concentrations set by the
label that was selected. If a limit is not set on the label or if no label was
selected, “-” is displayed.
Temperature indicates the temperature of the sample in the vessel. Note: this
may differ from the sample temperature.
User indicates which user was selected at the start of the test. If no user
was selected, DEFAULT is displayed.
49
Time indicates the time when the test was performed. Date indicates the date
on which the test was performed. Batch Code indicates the batch used for the
test. Dilution factor indicates the multiplier on the measured result to
produce the displayed result. Default is 1.
Press ADD REFERENCE to add information from a barcode (or the information can
be typed in manually) to enrich the data record. See p. 9 in the instruction
manual for details on producing and using barcodes.
Press ADD NOTES to type information about the test using the on-screen
keyboard. Maximum of 48 characters. After a failed test, it is recommended to
write an explanatory note or explain the corrective actions taken, as the
information will be stored alongside the test in memory.
Error messages are stored in the note field automatically.
50
Final Steps
After the test results are displayed, the user must complete the test by
removing the sensor, disposing of the sample and cleaning the instrument. This
protects the instrument, prevents microbial growth, prevents spills, and
prepares the instrument for the next test. Kemio will prompt the user to
complete these actions.
After viewing the results, the user has the choice to:
A Finish the test and clean for the instrument B Return to the home screen C
Perform another test D Shut down the instrument
A. Finish The Test (Disinfection Test)
1
Open the lid.
51
2
Remove the sensor.
3
Discard the sample.
4
If Kemio is wet, wipe it down
with a cloth according to
the care and maintenance
suggestions. Take special
care of the electrical
contacts. These can be
accessed by opening the
front lid of the arm.
52
5
Close the lid. Kemio will return to the results screen.
Kemio does not detect whether the sensor and sample have been removed – only
that the lid has been opened to do so. Take care of the instrument and ensure
they are properly removed.
A. Finish The Test (Metals Test)
1
Open the front lid to drop the sensor into the disposable vessel.
53
2
Open the lid.
3
Remove both vessels from
Kemio
4
Place a cap on the
disposable vessel. If required
by local regulations, dispose
of the sample, sensor and
vessel in a secure manner.
54
5
Discard the sample from the L-Shaped vessel
Kemio does not detect whether the sensor and sample have been removed – only
that the lid has been opened to do so. Take care of the instrument and ensure
they are properly removed.
B. Return To The Home Screen
If the “Finish the test” instructions have not yet been completed, Kemio will
prompt the user to to do these. After the lid is closed, Kemio will return to
the home screen.
If the lid has already been opened, Kemio will proceed directly to the home
screen.
55
C. Perform Another Test
Press REPEAT TEST.
When retesting, Kemio reuses the same label, user and batch information from
the previous test.
If testing a sample that should use a different label, if a different user is
testing or if the next sensor comes from a new batch, DO NOT press REPEAT
TEST. This could lead to an inaccurate result or incorrect information.
Instead, press the HOME button and PERFORM TEST.
D. Shut Down The Instrument
Press the power button.
If the user presses and holds the power button before opening the lid, Kemio
will display a warning message reminding the user to discard the sensor and
sample, and then turn off.
If the lid has already been opened, the regular “Goodbye’ screen will display
and Kemio will turn off.
Sensor Performance
Sensor Test type Range type
CLO
Free Chlorine
Total Chlorine
0.02-10 (mg/L)
0.05-75 (mg/L)
CHR
Free
0.1 – 25
Chlorine (mg/L)
Temp Resolution Range (mg/L) (°C)
Precision (mg/L)
2 – 35 0.01 up to 5.0 0.05 @ 1.0 mg/L, 20°C 0.1 up to 10
2 – 35 0.01 up to 5.0 0.5 @ 10 mg/L, 20°C 0.1 up to 20 1 up to 75
5 – 25 0.01 up to 5.0 2.0 @ 20 mg/L, 10°C 0.1 up to 25
Total
1 – 500
Chlorine (mg/L)
5 – 30 0.1 up to 20 1 up to 500
10.0 @ 200 mg/L, 10°C
CLT
Free
0.1 – 25 2 – 15 0.01 up to 5.0 2.0 @ 20 mg/L, 5°C
Chlorine (mg/L)
0.1 up to 25
Total
1 – 500
Chlorine (mg/L)
2 – 15 0.1 up to 20 1 up to 500
10.0 @ 200 mg/L, 5°C
FHR CDX
Free Chlorine
Chlorine Dioxide
Chlorite
0.1 – 100 2 – 20 0.02 up to 10 2.25 @ 30 mg/L, 2°C
(mg/L)
0.1 up to 100 5.63 @ 75 mg/L, 14°C
0.02 – 50 5 – 40 0.01 up to 10.0 0.05 @ 1.0 mg/L, 20°C
(mg/L)
0.1 up to 50
0.02 – 50 5 – 40 0.01 up to 2.5 0.05 @ 1.0 mg/L, 20°C
(mg/L)
0.1 up to 20
1.0 up to 50
Sensor Performance Cont.
57
Sensor Test type Range Temp Resolution
type
Range (mg/L)
(°C)
Precision (mg/L)
PAA
Peracetic 5 –
Acid
2000
(mg/L)
5 – 30 0.1 up to 20.0 1 up to 500 5 up to 2000
1.25 @ 25 mg/L, 10°C 10 @ 200 mg/L, 20°C 75 @ 1500 mg/L, 20°C
PAL Peracetic 0.02 – 10 2 – 30 0.01 up to 5
Acid
(mg/L)
0.1 up to 20
0.04 @ 0.2 mg/L, 20°C 0.05 @ 1.0 mg/L, 5°C
PAH
Peracetic 400 – 10 – 60 1 up to 1000
Acid
4000
10 up to 4000
(mg/L)
125 @ 2500 mg/L, 50°C 175 @ 3500 mg/L, 50°C
MPB Lead
1 – 100 5 – 35 1 up to 100 µg/L 1.0 µg/L @ 5 ug/L, 20°C (µg/L)
MCD Cadmium 0.5 – 10 5 – 35 0.1 up to 10.0
(µg/L)
µg/L
1.0 µg/L @ 5 ug/L, 20°C
MCU
Copper
0.1-2.0 (mg/L)
5 – 35 0.001mg/L up to 0.1mg/L @ 0.5mg/L, 20°C 0.6mg/L
0.005mg/L up to 2.0 mg/L
BRS
Bromine 0.2 – 10 5 – 30 0.01 up to 1 mg/L 20% up to 1 mg/L
(mg/L)
0.1 up to 10 mg/L 10% 1 – 10 mg/L
MAS Arsenic 2 -250 10 – 35 0.1 µg/L (µg/L)
20% up to 5 µg/L 10% 5 – 250 µg/L
58
Reorder Codes
Description Chlorine Free High Range Sensors (FHR) (100 pack) Chlorine Free High Range Sensors (FHR) (500 pack) Chlorine Standard Range Sensors (CLO) (100 pack) Chlorine Standard Range Sensors (CLO) (500 pack) Bromine Sensors (BRS) 100 Pack Chlorine High Range Sensors (CHR) (100 pack) Chlorine High Range Sensors (CHR) (500 pack) Chlorine High Range Low Temp Sensors (CLT) (100 pack) Chlorine High Range Low Temp Sensors (CLT) (500 pack Chlorine Dioxide and Chlorite Sensors (CDX) (100 pack) Chlorine Dioxide Sensors and Chlorite (CDX) (500 pack) Glycine Reagent CR-1 Reagent CR-2 Reagent Peracetic Acid Sensors (PAA) (100 pack) Peracetic Acid Sensors (PAA) (500 pack) Peracetic Acid Sensors Low Range (PAL) (100 pack) Peracetic Acid Sensors Low Range (PAL) (500 pack) Peracetic Acid Sensors High Range (PAL) (100 pack) Peracetic Acid Sensors High Range (PAL) (500 pack) Lead Sensors (MPB) (10 pack) Cadmium Sensors (MCD) (10 pack) Total Arsenic Sensors (100 pack) Speciated Arsenic (III & V) Sensors (50 Tests) Copper Sensors (MCU) (10 pack)
Product Code KEM21FHR KEM25FHR KEM21CLO KEM25CLO KEM21BRS KEM21CHR KEM25CHR KEM21CLT KEM25CLT KEM21CDX KEM25CDX PT549 PT546 PT547 KEM21PAA KEM25PAA KEM21PAL KEM25PAL KEM25PAH KEM25PAH KEM22MPB KEM22MCD KEM100-TAS KEM100M-SAS KEM22MCU
5599
www.palintest.com
KEM TI V2 05/23
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>