KLEIN TOOLS IR5 Dual Laser Infrared Thermometer Instruction Manual
- June 4, 2024
- KLEIN TOOLS
Table of Contents
INSTRUCTION MANUAL
Dual-laser Infrared
Thermometer
IR5 Dual Laser Infrared Thermometer
- DUAL-LASER TARGETING
- 12:1 DISTANCE TO-SPOT RATIO
- AUTO-SCAN
- MAX/MIN/AVG/DIFF
- HIGH/LOW ALARM
- BACKLIT DISPLAY
- AUTO POWER-OFF
- ADJUSTABLE EMISSIVITY
| |
---|---|---
GENERAL SPECIFICATIONS
Klein Tools IR5 is a professional dual-laser targeting infrared thermometer. It offers a wide measurement range, a tight distance-to-spot ratio, dual targeting lasers, and several calculation modes to facilitate different temperature measurement applications.
- Operating Altitude: 6562 ft. (2000 m)
- Relative Humidity: <85% non-condensing
- Operating Temp: 32° to 122°F (0° to 50°C)
- Storage Temp: -4° to 140°F (-20° to 60°C)
- Measurement Range: -22° to 752°F (-30° to 400°C)
- Units: Settable to °F or °C
- Emissivity: 0.10-1.00 adjustable
- Optical Resolution (Distance-to-spot) : 12:1
- Dimensions: 6.83″ x 4.53″ x 1.85″ (173 x 115 x 47 mm)
- Weight: 9.84 oz. (279 g) including battery
- Battery Type: 1 x 9V battery
- Battery Life: (Estimates assume 9V Zinc-Carbon Battery) 10 hours continuous use w/laser and backlight on.
- Display: Backlit LCD with white backlight
- Display Resolution: 0.1°F (0.1° C)
- Calibration: Accurate for one year
- Lasers: FDA and IEC Class II
- Standards: Certified to IEC EN 61326-1:2013, EN 61326-2-3:2013, IEC EN 60825-1:2014 Complies with 21 CFR 1040.10 and 1040.11 except for conformance with IEC 60825-1 ED. 3, as described in Laser Notice No. 56 dated May 8, 2019
- Pollution degree: 2
- Drop Protection: 6.6 ft. (2m)
- Electromagnetic Environment: IEC EN 61326-1. This equipment meets requirements for use in basic and controlled electromagnetic environments like residential properties, business premises, and light-industrial locations. Specifications are subject to change.
MEASUREMENT SPECIFICATIONS
- Response Time: <250ms
- Data hold: Yes
- MAX/MIN/Average/Differential: Yes
- Spectral Response: 8000-14000nm
≥32°F (≥0°C) | ±4°F (±2°C) or ±2% (whichever is greater) |
---|---|
<32°F (<0°C) | ±(4°F + 0.2°F per degree below 32°F) |
±(2.2°C + 0.2°C per degree below 0°C)
- Repeatability: ±0.8% of reading or ±2°F(±1°C) (whichever is greater).
- Temperature Correction Coefficient: ±0.2°F per °F (±0.1°C per °C), or ±0.2% of reading (whichever is greater) when the ambient temperature is above or below 70 – 77°F (21 – 25°C). Specifications are subject to change.
WARNINGS
To ensure safe operation and service of the meter, follow these
instructions.
Failure to observe these warnings can result in severe injury or death.
- Do not direct the laser beam into eyes, as this can cause permanent eye damage.
- Do not use the meter if the case is damaged in any way.
- Replace the battery as soon as the low battery indicator appears.
- The thermometer is NOT intended for use on people or animals.
- Service the meter immediately if it is acting abnormally.
- Avoid using the meter around strong electromagnetic fields.
CAUTION
- BURN HAZARD. Reflective materials may have a higher actual temperature than the measured temperature. Set emissivity to match the object being measured (see Emissivity section).
SAFE PRACTICES
This meter is designed for professionals who understand the hazards associated with their trade. While this meter causes no foreseeable dangers beyond its targeting lasers; the objects being measured, as well as the environment in which they reside, can be hazardous. Common safety practices to follow when operating near temperature-critical environments are:
- Follow the manufacturer’s maintenance procedures when servicing equipment.
- Before using this meter to determine if an area is safe, verify correct operation by measuring a known temperature value of a comparable object.
- Properly maintain this meter and calibrate it regularly.
Warning label on side of the meter
FEATURE DETAILS
NOTE: There are no user-serviceable parts inside the meter.
- Control button – Selects mode and moves down through menus
- SET button – Sets selection
- Control button – Turns to target lasers on/off, turns the backlight on/off, moves up through menus
- LCD Display
- Trigger
- Battery compartment door
- IR temperature sensor
- Targeting lasers
SYMBOLS ON LCD
| Indicates active measurement
---|---
| Targeting lasers active
| Backlight active
HOLD
| Data Hold
| Low battery indicator
°F °C
| Indicates active temperature scale
| Indicates measurement mode
| Emissivity
| High / Low-temperature limit indicators
| Enable/disable audible temperature limit alarms
OPERATING INSTRUCTIONS
TEMPERATURE MEASUREMENT
To measure temperature with the IR5, aim the meter at the object to be
measured, pull the trigger 5, and hold it depressed for at least 2
seconds. Releasing the trigger initiates Data Hold, “HOLD” will show on
the display, and the measurement will be held on the display. If the display
shows “OL” or “–OL” following a measurement, this indicates that the
surface temperature of the object being measured is either above or below the
measurement range of the meter.
TARGETING
The IR5 features dual lasers 8 to assist in targeting the measurement
area. The distance between the two laser spots on the surface of the object
being measured approximates the diameter of the circular measurement area from
which the infrared sensor is collecting data. Measurement areas located far
away from the meter will be larger than those close to the meter.
The meter is configured with a 12:1 optical resolution (distance-to-spot ratio). The distance-to-spot ratio defines the size of the measurement area relative to the distance between the measurement location and the IR sensor 7. Typical diameters of the measurement area as a function of the distance between the meter and the target area are depicted below for 12:1 optical systems (Fig. 1).
TARGETING LASERS
The targeting lasers 8 may be turned on/off by long presses of control
button 3. When on, the laser icon “ ” will be visible on the display.
BACKLIGHT
The backlight may be toggled on/off by short presses of the control button
3. When on, the backlight icon will be visible on the display.
MODE SELECTIONS
When measuring temperature the meter continuously samples the object being
measured. Following a measurement, repeatedly press the mode control button
1 to cycle through:
- the maximum temperature value measured (“MAX”).
- the average value measured (“AVG”).
- the difference between maximum and minimum values measured (“DIFF”).
- the minimum value measured (“MIN”).
- Press once more to exit the MODE menu.
SETTINGS
User adjustable settings may be defined using the SET control button 2.
Press SET 2 to enter the settings menu, subsequent presses of the SET cycle
through the following list of options:
- emissivity: set the numerical value of the emissivity to match the surface being measured (see EMISSIVITY section below). When in this setting the emissivity icon will flash on the display.
- units: Switch between °F (Fahrenheit) and °C (Celsius). When in this setting the “°F” or “°C” icon will flash on the display, indicating the temperature scale that is currently active.
- mute temperature limit alarms: (turn audible limit alarms on/off). When in this setting the audible icon “ “will flash on the display. When on, the high /low icons will be visible, and when off they will not be visible.
- high-temperature limit setting: assign a numerical value to the high-temperature limit. When in this setting, the high icon will flash on the display.
- low-temperature limit setting: assign a numerical value to the low-temperature limit. When in this setting, the low icon “ Lo ” will flash on the display.
When in any setting, control buttons 1 and 3 function as up/down to adjust settings and they may be used to select units, turn alarms on or off, or adjust the numerical values of the respective parameters up or down.
EMISSIVITY
Emissivity is a measure of the ability of a surface to emit thermal energy by
radiation. Different types of surfaces (metals, masonry, wood, etc.) emit
thermal energy through radiation at different efficiencies. Accordingly, these
materials have different emissivity coefficients which must be considered in
order to make accurate measurements with an infrared thermometer.
Emissivity on the IR5 may be adjusted from 0.10 to 1.00 to enable accurate
measurement of the temperature of most types of materials. Generally speaking,
shiny bright surfaces such as chrome, whiteboards, etc. exhibit lower
emissivity than flat black materials.
For guidance only, the chart below may be used to estimate emissivity for many
different types of materials. However, the emissivity of surfaces is dependent
upon many parameters such as surface finish, temperature, the shape of the
object, etc.
This chart should be used for guidance only.
Material | Emissivity |
---|---|
Asphalt | 0.93 |
Red brick | 0.93 |
Gray brick | 0.75 |
Porcelain ceramic | 0.92 |
Fired clay | 0.91 |
Rough concrete | 0.94 |
Cotton cloth | 0.77 |
Smooth glass | 0.92 – 0.94 |
Granite | 0.45 |
Gravel | 0.28 |
Smooth ice | 0.97 |
Smooth white marble | 0.56 |
Black paint | 0.96 |
Hard rubber | 0.94 |
Wood | 0.80 – 0.90 |
Matte Copper | 0.22 |
Commercial sheet aluminum | 0.09 |
Cold rolled steel | 0.75 – 0.85 |
Find a comprehensive list of emissivity values at https://www.kleintools.com/emissivity.
MAINTENANCE
BATTERY REPLACEMENT
When the indicator is displayed on
the LCD, the battery must be replaced.
- Open battery compartment 6 by pulling the battery compartment cover at the indentations away from the trigger.
- Remove the exhausted 9V battery and dispose of it appropriately.
- Replace the 9V battery and return the battery compartment cover, ensuring that it locks into place.
CLEANING
Be sure the meter is turned off and wipe with a clean, dry lint-free cloth.
Do not use abrasive cleaners or solvents. Take care to keep the sensor
lens clean at all times. If required, loose debris may be removed from the
lens using clean compressed air. The lens may also be cleaned using a soft
cloth or cotton swab with water or rubbing alcohol only. The lens must be
allowed to completely dry prior to use.
STORAGE
Remove the battery when the meter is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the General Specifications section, allow the meter to return to normal operating conditions before use.
FCC AND IC COMPLIANCE
See this product’s page at www.kleintools.com for
FCC compliance information.
Canada ICES-003 (B) / NMB-003 (B)
WARRANTY
www.kleintools.com/warranty
DISPOSAL / RECYCLE
Do not place equipment and its accessories in the trash.
Items must be properly disposed of in accordance with local regulations.
Please see www.epa.gov/recycle for additional
information.
CUSTOMER SERVICE
KLEIN TOOLS, INC.
450 Bond Street Lincolnshire, IL 60069
1-800-553-4676
customerservice@kleintools.com
www.kleintools.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>