HANNA Instruments HI96707C Nitrite Low Range ISM Instruction Manual
- June 13, 2024
- HANNA instruments
Table of Contents
- HI96707C Nitrite Low Range ISM
- PRELIMINARY EXAMINATION
- ABBREVIATIONS
- SPECIFICATIONS
- PRECISION AND ACCURACY
- PRINCIPLE OF OPERATION
- FUNCTIONAL DESCRIPTION
- ERRORS AND WARNINGS
- GENERAL TIPS FOR AN ACCURATE MEASUREMENT
- STARTUP
- MEASUREMENT PROCEDURE
- VALIDATION PROCEDURE
- CALIBRATION PROCEDURE
- GLP
- BATTERY REPLACEMENT
- ACCESSORIES
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
HI96707C
Nitrite Low Range ISM
INSTRUCTION MANUAL
HI96707C Nitrite Low Range ISM
Dear Customer,
Thank you for choosing a Hanna Instruments product.
Please read this instruction manual carefully before using the instrument.
This manual will provide you with the necessary information for correct use of
the instrument, as well as a precise idea of its versatility.
If you need additional technical information, do not hesitate to e‑mail us at
tech@hannainst.com or view our worldwide contact
list at www.hannainst.com.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instruments Inc.,
Woonsocket, Rhode Island, 02895, USA.
PRELIMINARY EXAMINATION
Please examine this product carefully. Make sure that the instrument is not
damaged. If any damage occurred during shipment, please contact your local
Hanna Instruments Office.
Each HI96707 Ion Selective Meter is supplied complete with:
- Sample Cuvettes and Caps (2 pcs.)
- CAL Check™ standard cuvettes
- 9V Battery
- 9V Battery
- Scissors
- Cloth for wiping cuvettes
- Instrument quality certificate
- Instruction Manual and Quick Referance Guide
- Rigid comying cose
Note: Save all packing material until you are sure that the instrument
works correctly. Any defective item must be returned in its original
packing.The HI96707 is an auto diagnostic portable
microprocessor meter that benefits from Hanna Instruments’ years of experience
as a manufacturer of analytical instruments. It has the advanced optical
system based on a special Tungsten Lamp and a narrow band interference filter
that allows most accurate and repeatable readings. All instruments are factory
calibrated and the electronic and optical design minimizes the need of
frequent calibration. With the powerful CAL CheckTM validation function, you
are able to validate good performance of your instrument at any time. The
validation procedure is extremely user friendly. Just use the exclusive Hanna
Instruments ready‑made, NIST traceable standards to verify the performance of
the instrument and recalibrate if necessary.
All instruments are splash proof and the lamp and filter units are protected
from dust or dirt by a transparent cup. This makes the instruments fulfill
field applications. Display messages aid the user in routine operation. The
meter has an auto‑shut off feature that will turn off the instrument after 10
minutes of non use in measurement mode or after 1 hour if left in calibration
mode. The meter uses an exclusive positive‑locking system to ensure that the
cuvette is in the same position every time it is placed into the measurement
cell. It is designed to fit a cuvette with a larger neck making it easier to
add both sample and reagents. The cuvette is made from special optical glass
to obtain best results.
The HI96707 meter measures the Nitrogen‑Nitrite (N02‑ ‑N) content in the
samples in the 0.000 to 0.600mg/L (ppm) range, in drinking, surface and saline
water samples and in domestic and industrial wastes. The method is an
adaptation of an EPA approved method.
The reagent is in powder form and it is supplied in packets. The amount of
reagent is precisely dosed to ensure the maximum repeatability.
ABBREVIATIONS
°C: degree Celsius
°F: degree Fahrenheit
mg/L: milligrams per liter. mg/L is equivalent to ppm (parts per million)
mL: milliliter
mV: millivolt
ASTM: American Society for Testing and Materials
SPECIFICATIONS
Range | 0.000 to 0.600 mg/L (NO2-N) |
---|---|
Resolution | 0.001 mg/L |
Accuracy
©25 °C (77 1)| ±0.020 mg/L ±4% of reading
Light Source| Tungsten lamp
Light Detector| Silicon Photocell with narrow band interference filter ©525nm
Method| Adaptation of an EPA approved method. The reaction between Nitrite and
reagent causes a pink tint in the sample.
Environment| 0 to 50 °C (32 to 122 °F); max 95% RH non-condensing
Battery Type| 9V (1 pc.)
Auto-Shut off| After 10′ of non-use in measurement mode; after 1 hour of non-
use in calibration mode; with last reading reminder.
Dimensions| 192 x 104 x 69 mm (1.6 x 4.1 x 2.7″)
Weight| 320 g (11.3 oz.).
REQUIRED REAGENTS
Code| Description| Quantity/test
HI93707-0| Nitrite Low Range Reagent| 1 packet
PRECISION AND ACCURACY
Precision is how closely repeated measurements agree with each other.
Precision is usually expressed as standard deviation (SD).
Accuracy is defined as the nearness of a test result to the true value.
Although good precision suggests good accuracy, precise results can be
inaccurate. The figure explains these definitions.
In a laboratory using a standard solution of 0.100 mg/L Nitrogen‑Nitrite and a
representative lot of reagent, an operator obtained with a single instrument a
standard deviation of 0.001 mg/L.
PRINCIPLE OF OPERATION
Absorption of Light is a typical phenomenon of interaction between
electromagnetic radiation and matter. When a light beam crosses a substance,
some of the radiation may be absorbed by atoms, molecules or crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the
optical path length through the matter and on the physical‑chemical
characteristics of the substance according to the Lambert‑Beer Law: Where:
‑log I/Io = Absorbance (A)
Io = intensity of incident light beam
I = intensity of light beam after absorption
ελ = molar extinction coefficient at wavelength λ
c = molar concentration of the substance
d = optical path through the substance
Therefore, the concentration “c” can be calculated from the absorbance of the
substance as the other factors are known.
Photometric chemical analysis is based on the possibility to develop an
absorbing compound from a specific chemical reaction between sample and
reagents. Given that the absorption of a compound strictly depends on the
wavelength of the incident light beam, a narrow spectral bandwidth should be
selected as well as a proper central wavelength to optimize measurements.
The optical system of Hanna Instruments’ HI96 series colorimeters is based on
special subminiature tungsten lamps and narrow‑band interference filters to
guarantee both high performance and reliable results.HI96 series block diagram (optical layout)
A microprocessor controlled special tungsten lamp emits radiation which is
first optically conditioned and beamed to the sample contained in the cuvette.
The optical path is fixed by the diameter of the cuvette. Then the light is
spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of
intensity Io or I.
The photoelectric cell collects the radiation I that is not absorbed by the
sample and converts it into an electric current, producing a potential in the
mV range.
The microprocessor uses this potential to convert the incoming value into the
desired measuring unit and to display it on the LCD.
The measurement process is carried out in two phases: first the meter is
zeroed and then the actual measurement is performed.
The cuvette has a very important role because it is an optical element and
thus requires particular attention. It is important that both, the measurement
and the calibration (zeroing) cuvettes, are optically identical to provide the
same measurement conditions. Whenever possible use the same cuvette for both.
It is necessary that the surface of the cuvette is clean and not scratched.
This to avoid measurement interference due to unwanted reflection and
absorption of light. It is recommended not to touch the cuvette walls with
hands.
Furthermore, in order to maintain the same conditions during the zeroing and
the measuring phases, it is necessary to close the cuvette to prevent any
contamination.
FUNCTIONAL DESCRIPTION
INSTRUMENT DESCRIPTION![HANNA Instruments HI96707C Nitrite Low Range ISM
-
parts5](https://manuals.plus/wp-content/uploads/2023/10/HANNA-Instruments- HI96707C-Nitrite-Low-Range-ISM-parts5.jpg)
- GLP/ key
- CAL CHECK key
- ZERO/CFM key
- READ /TIMER key
- ON/OFF key
- Liquid Crystal Display (LCD)
- Cuvette alignment indicator
- Cuvette holder
KEYPAD DESCRIPTION
- ON/OFF: to turn the meter on and off.
- ZERO/CFM: this is a bi‑functional key. Just press to zero the meter prior to measurement, or confirm edited values. In calibration mode press to confirm factory calibration restore.
- READ /TIMER: this is a multi‑functional key. In measurement mode, press to make a measurement, or press and hold for three seconds to start a pre‑programmed countdown prior to measurement. In GLP mode press to view the next screen.
- CAL CHECK: this is a bi‑functional key. Just press to perform the validation of the meter, or press and hold for three seconds to enter calibration mode.
- GLP/ : this is a bi‑functional key. Just press to enter GLP mode. In calibration mode press to edit the date and time.
OPERATING MODES
- Measurement mode: default operation mode, enables both validation and measurement.
- Calibration mode: may be entered by keeping CAL CHECK pressed for three seconds (the “CAL” tag appears), it enables calibration of the instrument.
- GLP mode may be entered by pressing GLP/ (“GLP” appears), it enables consulting of user calibration date or restore factory calibration.
DISPLAY ELEMENTS DESCRIPTION
- The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or reading measurement
- Error messages and warnings
- The battery icon indicates the charge state of the battery
- The hourglass appears when an internal check is in progress
- Status messages
- The chronometer appears when the reaction timer is running
- The month, day and date icons appear when a date is displayed
- Four digit main display
- Measuring units
- Four digit secondary display
ERRORS AND WARNINGS
The instrument shows clear messages when erroneous condition appears. Messages
are also displayed when the obtained values are outside expected range. The
beeper is playing a beep on errors.
a) on zero reading
Light High: There is too much light to perform a measurement. Please check
the preparation of the zero cuvette. Light Low: There is not enough light
to perform a measurement. Please check the preparation of the zero cuvette.
No Light: The instrument cannot adjust the light level. Please check that
the sample does not contain any debris. b) on sample reading
Inverted Cuvettes: The sample and the zero cuvette are inverted. Zero:
A zero reading was not taken. Follow the instructions of the measurement
procedure for zeroing the meter. Under range: A blinking “0.000” indicates
that the sample absorbs less light than the zero reference. Check the
procedure and make sure you use the same cuvette for reference (zero) and
measurement. Over Range: A flashing value
of the maximum concentration indicates an over range condition. The
concentration of the sample is beyond the programmed range: dilute the sample
and re‑run the test. c) during calibration procedure
Standard Low: The standard reading is less than expected. Standard High:
The standard reading is higher than expected. d) other errors and warnings
Cap error: Appears when external light enters in the analysis cell. Assure
that the cuvette cap is present. Cooling lamp: The instrument waits for
the lamp to cool down. Battery low: The battery must be replaced soon.
Dead battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the meter.
GENERAL TIPS FOR AN ACCURATE MEASUREMENT
The instructions listed below should be carefully followed during testing to ensure best accuracy.
-
Color or suspended matter in large amounts may cause interference, therefore these should be removed by treatment with active carbon and by prior filtration.
-
For a correct filling of the cuvette: the liquid in the cuvette forms a concavity on the top; the bottom of this concavity must be at the same level of the 10 mL mark.
-
Proper use of the powder reagent packet:
(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet. -
It is important that the sample does not contain any debris. This would corrupt the reading.
-
Each time the cuvette is used, the cap must be tightened to the same degree.
-
In order to avoid reagent leaking and to obtain more accurate measurements, it is recommended to close the cuvette first with the supplied HDPE plastic stopper and then with the black cap.
-
Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI731318 or a lint‑free cloth prior to insertion.
-
Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
-
Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
-
It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement.
-
After the reading it is important to discard immediately the sample, otherwise the glass might become permanently stained.
-
All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time should be increased for temperatures lower than 20 ºC (68 ºF), and decreased for temperatures higher than 25 ºC (77 ºF).
-
In order to maximize accuracy, prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.
STARTUP
Prepare the instrument for measurement as follows:
- Unpack the instrument by removing the dust protection sleeve from the instrument cuvette holder.
- Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
- Place the instrument on a flat table.
- Do not place the instrument under direct sun light.
MEASUREMENT PROCEDURE
To compensate the meter for the sample turbidity or color, the measurement takes place in two phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the reacted sample is measured.
-
Turn the meter on by pressing ON/OFF. The display briefly shows all tags on.
-
When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking “ZERO” indicates that the instrument needs to be zeroed first.
-
Fill the cuvette with 10 mL of unreacted sample, up to the mark, and replace the cap.
-
Place the cuvette into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
-
Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.
-
After a few seconds, the display will show “-0.0-”. The meter is now zeroed and ready for measurement.
-
Remove the cuvette.
-
Add the content of one pack of HI93707‑0 Nitrite Low Range reagent. Replace the cap and shake gently for 15 seconds.
-
Replace the cuvette into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
-
Press and hold READ TIMER for three seconds. The display will show the countdown prior to measurement. The beeper is playing a beep at the end of countdown period.Alternatively, wait for 15 minutes and just press READ /TIMER.
In both cases, the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. -
At the end of measurement, the instrument directly displays concentration in mg/L of nitrogen ‑ nitrite (NO2 ¯‑N) on the LCD.
-
To convert the NO 2 ¯‑N concentration to the nitrite ion concentration (NO2 ¯), multiply the reading by factor of 3.29.
-
To convert the NO2 ¯‑N concentration to sodium nitrite concentration (NaNO2 ), multiply the reading by factor of 4.93.
INTERFERENCES
Highly buffered samples (alkalinity above 500 ppm) or with extreme pH may
exceed the buffering capacity of the reagent: neutralize acidic samples with
diluted NaOH or alkaline samples with diluted HCl, prior to addition of the
reagent.
Turbidity must be removed by prior filtration.
The following ions may cause negative errors:
cupric ions (Cu 2+): above 5 ppm
ferrous ions (Fe 2+): above 1 ppm
ferric ions (Fe 3+): above 2 ppm
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly
calibrated.
Warning: Do not validate the instrument with any standard solutions other
than the Hanna Intruments CAL CheckTM Standards, otherwise erroneous results
will be obtained. For accurate validation, please perform test at room
temperature: 18 to 25 °C (64.5 to 77.0 °F)
- Turn the meter on by pressing ON/OFF.
- When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
- Place the CAL CheckTM Standard HI96707‑11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
- Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.
- After a few seconds, the display will show “-0.0-”. The meter is now zeroed and ready for validation.
- Remove the cuvette.
- Place the CAL CheckTM Standard HI96707‑11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
- Press CAL CHECK and the lamp, cuvette and detector icons together with “CAL CHECK” will appear on the display, depending on the measurement phase.
- At the end of the measurement the display will show the validation standard value.
The reading should be within specifications as reported in the CAL CheckTM Standard Certificate. If the value is found out of the specifications, please check that the cuvettes are free of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications, then recalibrate the instrument.
CALIBRATION PROCEDURE
Note: It is possible to interrupt calibration procedure at any time by
pressing CAL CHECK or ON/OFF keys.
Warning: Do not calibrate the instrument with standard solutions other
than the Hanna Instruments CAL CheckTM Standards, otherwise erroneous results
will be obtained. For accurate calibration, please perform test at room
temperature: 18 to 25 ºC (64.5 to 77.0 ºF).
- Turn the meter on by pressing ON/OFF.
- When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
- Press and hold CAL CHECK for three seconds to enter calibration mode. The display will show “CAL” during calibration procedure. The blinking “ZERO” asks for instrument zeroing.
- Place the CAL CheckTM Standard HI96707‑11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
- Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.
- After a few seconds the display will show “-0.0-”. The meter is now zeroed and ready for calibration. The blinking “READ” asks for reading calibration standard.
- Remove the cuvette.
- Place the CAL CheckTM Standard HI96707‑11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove.
- Press READ TIMER and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase.
- After measurement the instrument will show for three seconds the CAL CheckTM Standard value. Note: If the display shows “STD HIGH”, the standard value was too high. If the display shows “STD LOW”, the standard value was too low. Verify that both CAL CheckTM Standards HI96714-11 Cuvettes, A and B are free fingerprints or dirt and that they are inserted correctly.
- Then the date of the last calibration (e.g.: ”01.08.2009”) appears on the display, or “01.01.2009” if the factory calibration was selected before. In both cases the year number is blinking, ready for date input.
DATE INPUT
- Press GLP/ to edit the desired year (2009‑2099). If the key is kept pressed, the year number is automatically increased.
- When the correct year has been set, press ZERO/CFM or READ TIMER to confirm. Now the display will show the month blinking.
- Press GLP/ to edit the desired month (01‑12). If the key is kept pressed, the month number is automatically increased.
- When the correct month has been set, press ZERO/CFM or READ /TIMER to confirm. Now the display will show the day blinking.
- Press GLP/ to edit the desired day (01‑31). If the key is kept pressed, the day number is automatically increased. Note: It is possible to change the editing from day to year and to month by pressing READ /TIMER.
- Press ZERO/CFM to save the calibration date.
- The instrument displays “Stor” for one second and the calibration is saved.
- The instrument will return automatically to the measurement mode by displaying dashes on the LCD.
GLP
In the GLP mode, the last user calibration date can be verified and the
factory calibration can be restored.
LAST CALIBRATION DATE
To display the calibration date:
- Press GLP/ ****to enter GLP mode. The calibration month and day will appear on the main display and the year on the secondary display.
- If no calibration was performed, the factory calibration message, “F.CAL” will appear on the main display and the instrument returns to measurement mode after three seconds.
FACTORY CALIBRATION RESTORE
It is possible to delete the calibration and restore factory calibration.
- Press GLP/ to enter GLP mode.
- Press READ /TIMER to enter in the factory calibration restore screen. The instrument asks for confirmation of user calibration delete.
- Press ZERO/CFM to restore the factory calibration or press GLP/ again to abort factorycalibration restore.
- The instrument briefly indicates “donE” upon restoration of factory calibration prior to returning to measurement mode.
To save battery, the instrument shuts down after 10 minutes of non‑use in
measurement mode and after 1 hour of non‑use in calibration mode.
If a valid measurement was displayed before auto shut off, the value is
displayed when the instrument is switched on. The blinking “ZERO” means that a
new zero has to be performed. One fresh battery lasts for around 750
measurements, depending on the light level.
The remaining battery capacity is evaluated at the instrument startup and
after each measurement.
The instrument displays a battery indicator with three levels as follows:
- 3 lines for 100 % capacity
- 2 lines for 66 % capacity
- 1 line for 33 % capacity
- Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken anymore, the
instrument shows“dEAd bAtt” and turns off.
To restart the instrument, the battery must be replaced with a fresh one.
BATTERY REPLACEMENT
To replace the instrument’s battery, follow the steps:
- Turn the instrument off by pressing ON/OFF.
- Turn the instrument upside down and remove the battery cover by turning it counterclockwise.
- Extract the battery from its location and replace it with a fresh one.
- Insert back the battery cover and turn it clockwise to close.
ACCESSORIES
Reagent Set
HI93707-01| Reagents for 100 tests
HI93707-03| Reagents for 300 tests
Other Accessories
HI96707-11| CAL Check™ Standard Cuvettes (1 set)
HI740029P| 9V battery (10 pcs.)
HI731318| Cloth for wiping cuvettes (4 pcs.)
HI731331| Glass cuvettes (4 pcs.)
HI731335| Caps for cuvettes
HI740318| Carrying case
HI93703-50| Cuvette cleaning solution (230 mL)
MAN96707 01/18
Recommendations for Users
Before using these products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other
electronic equipments, this requiring the operator to take all necessary steps
to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the
instruments’ EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. For
yours and the instrument safety do not use or store the instrument in
hazardous environments.
Warranty
HI96707 is warranted for two years against defects in workmanship and
materials when used for its intended purpose and maintained according to the
instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed maintenance
are not covered.
If service is required, contact your local Hanna Instruments Office. If under
warranty, report the model number, date of purchase, serial number and the
nature of the failure. If the repair is not covered by the warranty, you will
be notified of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department and
then send it with shipment costs prepaid. When shipping any instrument, make
sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
Hanna Instruments reserves the right to modify the design, construction or
appearance of its products without advance notice.
World Headquarters
Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com
Local Office
Hanna Instruments USA
270 George Washington Highway
Smithfield, RI 02917
Phone: 800.426.6287
Fax: 401.765.7575
e‑mail: tech@hannainst.com
MAN96707
Printed in ROMANIA
https://manual-hub.com/
References
- pH Meters, Photometers, Titrators, Controllers - Hanna Instruments
- Manual-Hub.com – Free PDF manuals!
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>