KLEIN TOOLS CL100 AC Clamp Meter Instruction Manual

June 13, 2024
KLEIN TOOLS

KLEIN-TOOLS-LOGO

KLEIN TOOLS CL100 AC Clamp Meter

KLEIN-TOOLS-CL100-AC-Clamp-Meter-PRODUCT

Product Information

The Klein Tools CL100 is a multi-featured clamp meter designed for measuring AC/DC voltage, AC, resistance, diodes, and continuity. It has a 1999 count LCD and can measure up to 600V and 600A. The clamp meter is equipped with CAT III- rated leads or higher for safety purposes.

Electrical Specifications

DC Voltage Measurement:

  • Range : 200mV, 2V, 20V, 200V, 600V
  • Resolution : 0.1mV, 0.001V, 0.01V, 0.1V, 1V
  • Input Impedance : 10M

AC Voltage Measurement:

  • Range : 200mV, 2V, 20V, 200V, 600V
  • Resolution : 0.1mV, 0.001V, 0.01V, 0.1V, 1V
  • Input Impedance : 10M
  • Response : Averaging

AC Current Measurement:

  • Range : 2A, 20A, 200A, 600A
  • Resolution : 0.001A, 0.01A, 0.1A, 1A
  • Frequency Range : 50 to 60Hz
  • Response : Averaging

Resistance Measurement:

  • Range : 200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 20MΩ
  • Resolution : 0.1Ω, 0.001kΩ, 0.01kΩ, 0.1kΩ, 0.001MΩ, 0.01MΩ
  • Overload Protection: 250V
  • Open Circuit Voltage: 0.25V

Diode Test:

  • Forward DC Current : Approximately 1mA

Continuity Test:

  • Overload Protection: 250V
  • Resolution : 0.001V
  • Resolution: 0.1mΩ
  • Reverse DC Voltage : Approximately 1.5V
  • Open Circuit Voltage: Approximately 0.45V

Usage Instructions

  1. DC Voltage Measurement:
    • Connect the test leads to the appropriate terminals on the clamp meter.
    • Select the desired range using the range switch.
    • Touch the test leads to the circuit or component you want to measure.
    • Read the measurement value from the LCD.
  2. AC Voltage Measurement:
    • Follow the same steps as for DC voltage measurement, but ensure the range switch is set to the AC voltage range.
  3. AC Current Measurement:
    • Turn off the power to the circuit being measured
    • Open the clamp and position it around the conductor carrying the current.
    • Close the clamp securely.
    • Turn on the power to the circuit.
    • Read the measurement value from the LCD.
  4. Resistance Measurement:
    • Disconnect power from the circuit being measured.
    • Connect the test leads to the appropriate terminals on the clamp meter.
    • Select the desired range using the range switch.
    • Touch the test leads to the circuit or component you want to measure.
    • Read the measurement value from the LCD.
  5. Diode Test:
  6. Disconnect power from the circuit being tested.
  7. Connect the test leads to the appropriate terminals on the clamp meter.
  8. Set the clamp meter to the diode test mode. d. Touch the test leads to the diode being tested.
  9. The clamp meter will display the forward DC value.
  10. Continuity Test:
    • Disconnect power from the circuit being tested.
    • Connect the test leads to the appropriate terminals on the clamp meter.
    • Set the clamp meter to the continuity test mode.
    • Touch the test leads to the circuit being tested.
    • The clamp meter will indicate continuity if the circuit is closed.

Please refer to the user manual for additional information and safety instructions.

ELECTRICAL SPECIFICATIONS

DC Voltage Measurement

Range Resolution Accuracy
200mV 0.1mV ****

± (0.8% + 2 digits)

2V| 0.001V
20V| 0.01V
200V| 0.1V
600V| 1V| ± (1.0% + 2 digits)

Overload Protection:

  • 200mV range: 250V DC / AC RMS
  • 2V-600V ranges: 600V DC / AC RMS
  • Input Impedance: 10MΩ

AC Voltage Measurement

Range Resolution Accuracy
200mV 0.1mV ± (1.0% + 10 digits)
2V 0.001V ****

± (1.0% + 5 digits)

20V| 0.01V
200V| 0.1V
600V| 1V| ± (1.5% + 5 digits)

Overload Protection:

  • 200mV range: 250V DC / AC RMS
  • 2V-600V ranges: 600V DC / AC RMS
  • Input Impedance: 10MΩ
  • Response: Averaging

AC Current Measurement

Range Resolution Accuracy
2A 0.001A ± (3.5% + 20 digits) ≤ 0.5A
2A 0.001A ± (3.0% + 10 digits) > 0.5A
20A 0.01A ± (3.0% + 10 digits)
200A 0.1A ± (2.5% + 10 digits)
600A 1A ± (1.5% + 5 digits)
  • Frequency Range: 50 to 60Hz
  • Response : Averaging

Resistance Measurement

Range Resolution Accuracy
200Ω 0.1Ω ****

± (1.2% + 2 digits)

2kΩ| 0.001kΩ
20kΩ| 0.01kΩ
200kΩ| 0.1kΩ
2MΩ| 0.001MΩ
20MΩ| 0.01MΩ| ± (2.0% + 5 digits)

  • Overload Protection: 250V
  • Open Circuit Voltage: 0.25V

Diode Test

Forward DC Current Resolution Reverse DC Voltage
Appx. 1mA 0.001V Appx. 1.5V

Continuity Test

Overload Protection Resolution Open Circuit Voltage
250V 0.1mΩ Appx. 0.45V

GENERAL SPECIFICATIONS
The Klein Tools CL100 is a multi-featured clamp meter. It measures AC / DC voltage, AC, and resistance. It can also test diodes and continuity.

  • Operating Altitude: 2000 meters
  • Relative Humidity : 75% max
  • Operating Temperature: 32°F ~ 104°F (0°C~ 40°C)
  • Storage Temperature: 10°F ~ 122°F (-10°C ~ 50°C
  • Accuracy Temperature: 64°F ~ 82°F (18°C ~ 28°C)
  • Temperature Coefficient : 0.1*(specified accuracy) / °C
  • Sampling Frequency: 2.5 samples per second
  • Dimensions: 8″ x 3″ x 1″ (203 mm x 76 mm x 25 mm)
  • Weight : 10 oz. (283 g)
  • Calibration : Accurate for one year
  • Safety Rating : CAT III 600V
  • Accuracy : ± (% of reading + # of least significant digits)

WARNINGS
To ensure the safe operation and service of the tester, follow these instructions. Failure to observe these warnings can result in severe injury or death.

  • Before each use, verify meter operation by measuring a known voltage or current.
  • Never use the meter on a circuit with voltages that exceed the category-based rating of this meter.
  • Do not use the meter during electrical storms, or in wet weather.
  • Do not use the meter or test leads if they appear to be damaged.
  • Ensure meter leads are fully seated, and keep fingers away from the metal probe contacts when making measurements.
  • Do not open the meter to replace batteries while the probes are connected.
  • Use caution when working with voltages above 60V DC, or 25V AC RMS. Such voltages pose a shock hazard.
  • To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
  • Unless measuring voltage or current, shut off and lock out power before measuring resistance or capacitance.
  • Always adhere to local and national safety codes. Use individual protective equipment to prevent shock and arc blast injury where hazardous live conductors are exposed

SYMBOLS

KLEIN-TOOLS-CL100-AC-Clamp-Meter-FIG- \(1\)

FEATURE DETAILS

KLEIN-TOOLS-CL100-AC-Clamp-Meter-FIG- \(2\)

  • A. Use CAT III rated leads or higher
    • Do not attempt to measure more than 600V or any current on lead input.
  • B. Keep hands below the line when measuring high current levels.
  • C. Select Functionality Button
    • Switch between AC and DC.
    • Switch between and .
  • D. Max Hold
    • Press to hold the maximum value on the display.
    • Press again to return to live reading.
  • E. Auto/Manual Range
    • Press repeatedly to cycle through manual ranges.
    • Hold for 2 seconds to return to auto-ranging mode.
    • AUTO is displayed on LCD only during auto-ranging mode.
  • F. Data Hold
    • Press to hold the current input on the display.
    • Press again to return to live reading.
  • G. H. Battery Replacement
    • When the indicator is displayed on the LCD, the batteries must be replaced.
    • Remove the back screw and replace 3 x AAA batteries.
  • I. Backlight / Worklight
    • Press the button to activate the worklight and backlight.
    • Lights will turn off automatically after 15 seconds.
    • The worklight only operates during AC measurement.
    • Using lights drains the battery significantly.

Auto Power-Off

  • The device will power off after 15 minutes of non-use.
  • Turn the dial or press a button to wake up.
  • Press the HOLD button while turning on to disable the Auto Power-Off feature.

FUNCTION INSTRUCTIONS

  1. AC Current: < 600A
    Features : HOLD MAX RANGE WORKLIGHTKLEIN-TOOLS-CL100-AC-Clamp-Meter-
FIG- \(7\)

    • Centre wire in guides for best accuracy.
    • Opposing currents cancel. Use line-splitter (USA only) when necessary
  2. AC / DC Voltage: < 600V
    Features : HOLD MAX RANGEKLEIN-TOOLS-CL100-AC-Clamp-Meter-FIG-
\(8\)

    • Select AC or DC voltage source
  3. Diode / Continuity

  4. Diode Features: HOLD
    Display shows:

  * Forward voltage drop if forward biased.
  * “O.L.” if reverse biased
2.  **Continuity Features** : HOLD
  * The display shows resistance.
  * Buzzer sounds if less than 60Ω.
  1. Resistance : < 20MΩ
    Features : HOLD RANGE

    • Do not measure resistance on a live circuit.

SYMBOLS USED ON LCD

WARRANTY
www.kleintools.com/warranty

CLEANING
Turn instrument off and disconnect test leads. Clean the instrument by using a damp cloth. Do not use abrasive cleaners or solvents.

STORAGE
Remove the batteries when instrument is not in use for a prolonged period of time. Do not expose to high temperatures or humidity. After a period of storage in extreme conditions exceeding the limits mentioned in the Specifications section, allow the instrument to return to normal operating conditions before using it.

DISPOSAL / RECYCLE

Caution : This symbol indicates that equipment and its accessories shall be subject to a separate collection and correct disposal.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

KLEIN TOOLS User Manuals