FLUKE 713 Pressure Calibrator Instruction Manual
- June 13, 2024
- FLUKE
Table of Contents
FLUKE 713 Pressure Calibrator
Product Information
The 713 Pressure Calibrator is a versatile tool used for measuring and calibrating pressure in various units. It features a wide pressure range and high resolution, allowing for accurate readings in different applications. The calibrator is designed with safety in mind, providing protection to the user and the equipment under test. It is important to follow the specified instructions to ensure proper usage and avoid any hazards or damage.
Pressure Range and Resolution
Model | Pressure Range | Resolution |
---|---|---|
Model 713 30G | 30.000 psi | 830.4 inH2O, 2109.0 cmH2O, 2.0685 bar, 2068.5 mbar, |
206.85 kPa,
61.080 inHg, 1551.3 mmHg, 2.1090 kg/cm2
Model 713 100G| 100.00 psi| 2768.0 inH2O, 7030 cmH2O, 6.895 bar, 6895 mbar,
689.5 kPa,
203.6 inHg, 5171 mmHg, 7.030 kg/cm2
Safety Information
The calibrator should only be used as specified in the instruction sheet to maintain its protective features. Failure to do so may impair the calibrator’s protection. The instruction sheet provides important warnings and cautions to ensure user safety and prevent damage to the calibrator or the equipment being tested.
Explanation of International Symbols
The instruction sheet also includes a list of international symbols used on the calibrator. These symbols provide additional information about specific features, such as earth ground, fuse, battery, double insulation, and compliance with relevant standards and directives.
Product Usage Instructions: 713 Pressure Calibrator
- Press the green O pushbutton to turn the calibrator on and off.
- The calibrator will display both pressure and current
- measurements simultaneously.
- Use the U pushbutton to select a different pressure unit.
- The upper part of the display shows the gage pressure applied to the pressure fitting.
- The lower part of the display shows the current (up to 24 mA) applied to the current inputs.
- Press D to turn on/off pressure reading damping. Damping slows down the display update.
- Press Z to zero the pressure display. Vent the pressure fitting before pressing Z.
- Press and hold M to read the maximum pressure and current readings since power on or Z was pressed.
- Press and hold m to read the minimum pressure and current readings since power on or Z was pressed.
- Press h to freeze the display. The g symbol will appear on the display.
- Press c to clear the MIN and MAX memories.
Introduction
Fluke Model 713 30G and 713 100G Pressure Calibrators (hereafter called “calibrator”) calibrate P/I (pressure to current) transmitters, or measure pressure or current. The calibrator simultaneously displays pressure and current measurements. Full scale pressure input is as follows:
- Model 713 30G: 30 psi (206.85 kPa, 2.0685 bar)
- Model 713 100G: 100 psi (689.5 kPa, 6.895 bar)
The calibrator measures and displays current up to 24 mA, and the pressure
applied to the 1/8-inch NPT pressure fitting. The calibrator measures pressure
in the units shown in the next table.
Your calibrator is supplied with a Flex-Stand holster, an installed 9 V
alkaline battery, one set of TL75 test leads, one set of AC70A alligator
clips, and this instruction sheet. If the calibrator is damaged or something
is missing, contact the place of purchase immediately. Contact your Fluke
distributor for information about accessories. See “How to Contact Fluke.” To
order replacement parts or spares, see “Replacement Parts.” .
Pressure Range and Resolution
Displayed Pressure Units| Model 713 30G Range and Resolution|
Model 713 100G Range and Resolution
---|---|---
psi| 30.000 psi| 100.00 psi
inH2O at 4°C| 830.4 inH2O| 2768.0 inH2O
inH2O at 20°C| 831.9 inH2O| 2772.9 inH2O
cmH2O at 4°C,| 2109.0 cmH2O| 7030 cmH2O
cmH2O at 20°C| 2113.0 cmH2O| 7043 cmH2O
bar| 2.0685 bar| 6.895 bar
mbar| 2068.5 mbar| 6895 mbar
kPa| 206.85 kPa| 689.5 kPa
inHg| 61.080 inHg| 203.6 inHg
mmHg| 1551.3 mmHg| 5171 mmHg
kg/cm2| 2.1090 kg/cm2| 7.030 kg/cm2
Safety Information
Use the calibrator only as specified in this instruction sheet, otherwise the protection provided by the calibrator may be impaired. A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the calibrator or the equipment under test
Warning
To avoid possible electric shock or personal injury:
- Never apply more than 30 V between the mA terminals, or between either of the mA terminals and earth ground
- Remove the test leads from the calibrator before you open the battery door.
- Make sure the battery door is closed and latched before you operate the calibrator.
- Do not operate the calibrator if it is damaged.
- Do not operate the calibrator around explosive gas, vapor, or dust. To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before you attach or detach the calibrator pressure fitting to the pressure line.
- When servicing the calibrator, use only specified replacement parts
Caution
To avoid overpressure damage, do not apply
pressure that exceeds the following:
- Model 713 30G: 3X top of range (90 psi, 620 kPa, 6.2 bar)
- Model 713 100G: 2X top of range (200 psi, 1380 kPa, 13.8 bar) To avoid corrosion in the pressure sensor, use the calibrator only with media compatible with glass, ceramic, silicon, RTV, type 316 stainless steel, and nickel. To avoid mechanically damaging the calibrator, do not apply torque between the pressure fitting and the calibrator case. See the figure for the proper use of tools.
Explanation of International Symbols
The symbols in the table below are used on the calibrator or in this instruction sheet
International Symbols
Symbol | Meaning |
---|---|
. | Earth ground |
Fuse | |
Battery | |
Refer to this instruction sheet for information about this feature. | |
Double insulated | |
**** | Conforms to relevant Canadian Standards Association directives. |
Conforms to European Union directives |
Getting Acquainted with the Calibrator
Press the green O pushbutton to turn the calibrator on and off. The calibrator
displays pressure and current measurements simultaneously. See the figure
below. The upper part of the display shows the gage pressure applied to
the pressure fitting. Press U to select a different pressure unit. When you
cycle the power off and on, the calibrator retains the pressure unit you last
used. The lower part of the display shows the current (up to 24 mA) applied to
the current (mA) inputs. The current inputs are fused with a 0.125 A, 250 V
fast fuse (Littelfuse type 2AG).
Pushbutton Functions
Pushbutton | Function |
---|---|
Selects a different pressure unit. | |
Turns pressure reading damping on and off. With damping on, the display does |
not update as quickly.
l| Press this to zero the pressure display. Vent the pressure fitting on the
calibrator to atmosphere before you press this pushbutton.
| Press and hold to read the minimum pressure and current readings since the
power was turned on or I was pressed.
| Press to clear the MIN and MAX memories.
| Press and hold to read the maximum pressure and current readings the power
was turned on or I was pressed.
1| Press 1 to freeze the display. The lll’ symbol appears on the display.
Press 1 again to resume normal operation.
Calibrating a P/I Transmitter
To calibrate a P/I (pressure to current) transmitter, you apply a pressure to the transmitter and measure the transmitter’s current loop output. Connect the calibrator to the transmitter as shown in the figure.
Maintenance
For maintenance procedures not described in this instruction sheet, or if the calibrator needs repair, contact a Fluke Service Center.
In Case of Difficulty
- Check the battery, test leads, and pressure tubing. Replace as necessary.
- Review this instruction sheet to make sure you are using the calibrator correctly. If the calibrator needs repair, and the calibrator is under warranty, see the warranty statement for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use
abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs according to
its specifications. A calibration manual is available (PN 686540). Call
1-800-526-4731 from the USA and Canada.
In other countries, contact a Fluke Service Center.
Replacing the Battery
When the Bsymbol appears on the display, replace the battery with a 9 V alkaline battery.
Replacing the Fuse
Warning
To avoid personal injury or damage to the calibrator, use only a 0.125A 250V
fast fuse, Littelfuse 2AG. Fuse F1 is probably blown if the mA measurement
display reading does not respond to current applied to the current (mA)
inputs.
Replace the fuse as follows:
- Remove the test leads and turn the calibrator off.
- Remove the battery door.
- Remove the three Phillips-head screws from the case bottom and turn the case over.
- Gently lift the top cover from the end nearest the current (mA) inputs until it unsnaps from the bottom cover.
- Replace the fuse with a 0.125 A 250 V fast fuse, Littelfusetype 2AG.
- Carefully fit the case top and circuit board assembly together, making sure that the O-ring is properly seated between the pressure sensor and the pressure fitting on the case top. Fit the case bottom onto the case top, engaging the two snaps near the display end of the case. Reinstall the three screws.
- Replace the battery door.
Replacement Parts and Accessories
Item
| ****
Description
| Part or model no.| ****
Qty.
---|---|---|---
BT1| 9V battery, ANSI/NEDA 1604A or IEC 6LR61| 614487| 1
CG81Y| Holster, Yellow| CG81Y| 1
& F1| Fuse, 125 mA, 250V fast| 686527| 1
MP85| 713 30G case top| 620218| 1
MP85| 713 100G case top| 620226| 1
MP86| Case bottom| 620168| 1
H2, 3, 4| Case screw| 832246| 3
MP89, 90| Non-skid foot| 824466| 2
OR1| O-ring for pressure input| 146688| 1
SP1| Spacer for pressure input| 687449|
MP92| Battery door| 619947| 1
H5, 6| Battery door fasteners| 948609| 2
S1| Keypad| 687068| 1
TL75| Test lead set| TL75| 1
AC70A| Alligator clips| AC70A| 1
–| 713 Instruction Sheet| 650298| 1
TL20| Industrial test lead set| TL20| Option
–| 71X Series Calibration Manual| 686540| Option
Specifications
Specifications are based on a one year calibration cycle and apply for ambient temperature from +18 °C to +28 °C unless stated otherwise. “Counts” are the number of increments or decrements of the least significant digit.
Pressure Input, 713 30G
Range | Accuracy |
---|---|
30 psi (206.85 kPa) | ±0.05% of range |
Maximum nondestructive pressure: 3X top of range (90 psi, 620 kPa,
6.2 bar)
Temperature coefficient: 0.01% of range per °C for temperature ranges
-10 °C to 18 °C and 28 °C to 55 °C
Pressure Input, 713 100G
Range | Accuracy |
---|---|
100 psi (689.5 kPa) | ±0.05% of range |
Maximum nondestructive pressure: 2X top of range (200 psi, 1380 kPa,
13.8 bar)
Temperature coefficient: 0.01% of range per °C for temperature ranges
-10 °C to 18 °C and 28 °C to 55 °C
DC mA Input, 713 30G and 713 100G
Range| Resolution| Accuracy,
± (% of Reading + Counts)
---|---|---
24 mA| 0.001 mA| 0.025 + 1
Overload protection: 125 mA, 250V fast acting fuse
Temperature coefficient: 0.005% of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C
General Specifications
- Maximum voltage applied between either mA terminal and earth ground or between the mA terminals: 30 V
- Storage temperature: -40°C to 60°C
- Operating temperature : -10°C to 55°C
- Operating altitude: 3000 meters maximum
- Relative humidity : 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, and 35% up to 55°C
- Vibration: Random 2 g, 5 Hz to 500 Hz
- Shock: 1 meter drop test
- Safety: Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992. Complies with ANSI/ISA S82.01-1994.
- Power requirements : Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
- Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x 7.35 in L); With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
- Weight : 369 g (13 oz); With holster and Flex-Stand: 624 g (22 oz
How to Contact Fluke
To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call: 1-800-44FLUKE (1-800-443-5853) in U.S.A. and Canada
- + 31-402-678-200 in Europe
- + 1-425-356-5500 from other countries
Address correspondence to:
- Fluke Corporation P.O. Box 9090, Everett, WA 98206-9090 USA
- Fluke Europe B.V. P.O. Box 1186, 5602 BD Eindhoven The Netherlands
Visit us on the World Wide Web at: www.fluke.com
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for
three years from the date of purchase. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect, misuse or abnormal
conditions of operation or handling. Resellers are not authorized to extend
any other warranty on Fluke’s behalf. To obtain service during the warranty
period, send your defective calibrator to the nearest Fluke Authorized Service
Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A
PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM
ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion
or limitation of an implied warranty or of incidental or consequential
damages, this limitation of liability may not apply to you.
https://manual-hub.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>