ACCU SCOPE EXI-410 Series Inverted Microscope Instruction Manual
- June 13, 2024
- ACCU SCOPE
Table of Contents
MANUAL
EXI-410
INVERTED
MICROSCOPE SERIES
SAFETY NOTES
-
Open the shipping carton carefully to prevent any accessory, i.e. objectives or eyepieces, from dropping and being damaged.
-
Do not discard the molded shipping carton; the container should be retained should the microscope ever require reshipment.
-
Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments.
Ensure the microscope is located on a smooth, level and firm surface. -
If any specimen solutions or other liquids splash onto the stage, objective or any other component, immediately disconnect the power cord and wipe up the spillage. Otherwise, the instrument may be damaged.
-
All electrical connectors (power cord) should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations.
-
Avoid blocking the natural air circulation for cooling. Ensure objects and obstructions are at least 10 centimeters from all sides of the microscope (the only exception is the table on which the microscope sits).
-
For safety when replacing the LED lamp or fuse, be sure the main switch is off (“O”), remove the power cord, and replace the LED bulb after the bulb and the lamp house has completely cooled.
-
Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The use of a different input voltage other than indicated will cause severe damage to the microscope.
-
When carrying this product, firmly grasp the microscope with one hand in the recess in the lower front of the main body and the other hand in the recess in the rear of the main body. Refer to the figure below.
Do not grasp or hold using any other parts (such as the illumination pillar, focus knobs, eyetubes or stage) when carrying the microscope. Doing so may result in dropping the unit, damage to the microscope or failure of proper operation.
CARE AND MAINTENANCE
- Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
- Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces should be cleaned with a damp cloth. More persistent dirt should be removed using a mild soap solution. Do not use organic solvents for cleansing.
- The outer surface of the optics should be inspected and cleaned periodically using an air stream from an air bulb. If dirt remains on the optical surface, use a soft cloth or cotton swab dampened with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a circular motion. A small amount of absorbent cotton wound on the end of a tapered stick such as cotton swabs or Q-tips, makes a useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick up grease making cleaning more difficult. Oil immersion objectives should be cleaned immediately after use by removing the oil with lens tissue or a clean, soft cloth.
- Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not in use.
- CCU-SCOPE® microscopes are precision instruments which require periodic preventative maintenance to maintain proper performance and to compensate for normal wear. An annual schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized ACCU-SCOPE® distributor can arrange for this service.
INTRODUCTION
Congratulations on the purchase of your new ACCU-SCOPE® microscope. ACCU- SCOPE® microscopes are engineered and manufactured to the highest quality standards. Your microscope will last a lifetime if used and maintained properly. ACCU-SCOPE® microscopes are carefully assembled, inspected and tested by our staff of trained technicians in our New York facility. Careful quality control procedures ensure each microscope is of the highest quality prior to shipment.
UNPACKING AND COMPONENTS
Your microscope arrived packed in a molded shipping carton. Do not discard the carton: the carton should be retained for reshipment of your microscope if needed. Avoid placing the microscope in dusty surroundings or in high temperature or humid areas as mold and mildew will form. Carefully remove the microscope from the EPE foam container by its arm and base and place the microscope on a flat, vibration-free surface. Check the components against the following standard configuration list:
- Stand, which includes the supporting arm, focusing mechanism, nosepiece, mechanical stage (optional), condenser with iris diaphragm, illumination system, and phase contrast accessories (optional).
- Binocular viewing head
- Eyepieces as ordered
- Objectives as ordered
- Stage plate inserts, green and yellow filters (optional)
- Dust cover
- 3-prong electric power cord
- Camera adapters (optional)
- Fluorescence filter cubes (optional)
Optional accessories such as optional objectives and/or eyepieces, slides sets, etc., are not shipped as part of the standard equipment. These items, if ordered, are shipped separately.
COMPONENTS DIAGRAMS
EXI-410 (with Phase Contrast)
1. Phase Contrast Slider
2. Eyepiece
3. Eyetube
4. Viewing Head
5. Emboss Contrast Slider
6. Power Indicator
7. Illumination Selector
8. Main Frame
9. LED Lamp (transmitted)
10. Illumination Pillar| 11. Condenser Set Screw
12. Field Iris Diaphragm
13. Condenser
14. Objective
15. Stage
16. Mechanical Stage with Universal Holder (optional)
17. Mechanical Stage Control Knobs (XY movement)
18. Focus Tension Adjustment Collar
19. Coarse Focus
20. Fine Focus
---|---
EXI-410 (with Phase Contrast)
1. Illumination Pillar
2. Field Iris Diaphragm
3. Phase Contrast Slider
4. Condenser
5. Mechanical Stage with Universal Holder (optional)
6. Objective
7. Nosepiece
8. Power Switch| 9. Eyepiece
10. Eyetube
11. Viewing Head
12. Light Path Selector
13. Camera Port
14. Power Indicator
15. Illumination Selector
16. Illumination Intensity Adjustment Knob
---|---
EXI-410 (with Phase Contrast)
1. Viewing Head
2. Stage
3. Emboss Contrast Slider
4. Main Frame
5. Focus Tension Adjustment Collar
6. Coarse Focus
7. Fine Focus
8. Condenser Set Screw| 9. Phase Contrast Slider
10. Condenser
11. Illumination Pillar
12. Rear Hand Grasp
13. Mechanical Stage (optional)
14. Nosepiece
15. Fuse
16. Power Outlet
---|---
EXI-410-FL
1. Phase Contrast Slider
2. Eyepiece
3. Eyetube
4. Viewing Head
5. Fluorescence Light Shield
6. Emboss Contrast Slider
7. Power Indicator
8. Illumination Selector
9. Main Frame
10. LED Lamp (transmitted)
11. Illumination Pillar
12. Condenser Set Screw
13. Field Iris Diaphragm| 14. Condenser Centering Screw
15. Condenser
16. Light Shield
17. Objective
18. Stage
19. Mechanical Stage with Universal Holder (optional)
20. Fluorescence Illumination
21. Fluorescence Turret
22. Mechanical Stage Control Knobs (XY movement)
23. Tension Adjustment Collar
24. Coarse Focus
25. Fine Focus
---|---
EXI-410-FL
1. Illumination Pillar
2. Field Iris Diaphragm
3. Phase Contrast Slider
4. Condenser
5. Mechanical Stage with Universal Holder (optional)
6. Objective
7. Nosepiece
8. Fluorescence Turret
9. Fluorescence Turret Access Door| 10. Power Switch
11. Eyepiece
12. Eyetube
13. Viewing Head
14. Light Path Selector (Eyepieces/Camera)
15. Camera Port
16. Power Indicator
17. Illumination Selector
18. Illumination Intensity Adjustment Knob
---|---
EXI-410-FL
1. Viewing Head
2. Fluorescence Light Shield
3. Emboss Contrast Slider
4. Main Frame
5. Focus Tension Adjustment Collar
6. Coarse Focus
7. Fine Focus
8. Condenser Set Screw
9. Phase Contrast Slider| 10. Condenser
11. Illumination Pillar
12. Light Shield
13. Rear Hand Grasp
14. Mechanical Stage (optional)
15. Nosepiece
16. LED Fluorescence Light Source
17. Fuse
18. Power Outlet
---|---
MICROSCOPE DIMENSIONS
EXI-410 Phase Contrast and Brightfield
EXI-410-FL with Mechanical Stage
ASSEMBLY DIAGRAM
The diagram below shows how to assemble the various components. The numbers
indicate the order of assembly. Use the hex wrenches that are supplied with
your microscope when required. Be sure to keep these wrenches for changing out
components or making adjustments.
When assembling the microscope, make sure that all parts are free of dust and
dirt, and avoid scratching any parts or touching glass surfaces.
ASSEMBLY
Condenser
To install the condenser:
- Unscrew the condenser set screw sufficiently to allow the condenser tube to slide over the dovetail groove of the condenser hanger.
- Lightly press the condenser into position and tighten the set screw.
Phase Contrast Slider
To install the phase contrast slider:
- With the printed notations on the slider facing up and readable from the front of the microscope, insert the phase contrast slider horizontally into the condenser slot. The orientation of the slider is correct if the edge of the slider facing the operator has adjusting screws visible.
- Continue to insert the slider until an audible “click” indicates that one position of the 3-postion phase contrast slider is aligned with the optical axis. Insert the slider further into the slot or backward to the desired slider position.
Mechanical Stage (optional)
To install the optional mechanical stage:
- Install the mechanical according to path ① (as shown in the figure). First, align the edge A of the mechanical stage with the edge of the flat/plain stage surface. Align the mechanical stage with the plain stage until the two set screws in the bottom of the mechanical stage align with screw holes in the bottom of the plain stage. Tighten the two set screws.
- Install the universal holder according to path ② (as shown in the figure). Start by placing the flat universal holder plate on the plain stage surface. Align the two screw holes on the universal holder plate with the set screws on the lateral movement ruler of the mechanical stage. Tighten the two set screws.
Objectives
To install the objectives:
- Turn the coarse adjustment knob ① until the revolving nosepiece is at its lowest position.
- Remove the nosepiece cap ② closest to you and thread the lowest magnification objective onto the nosepiece opening, then rotate the nosepiece clockwise and thread the other objectives from low to high magnification.
NOTE:
- Always rotate the nosepiece by using the knurled nosepiece ring.
- Keep the covers on any unused nosepiece openings to prevent dust and dirt from getting inside.
Stage Plate
Insert the clear glass stage plate ① into the opening on the stage. The clear
glass allows you to view the objective in position.
Eyepieces
Remove the eyetube plugs and fully insert the eyepieces ① into the eyepiece
tubes ②.
Camera (optional)
To install the optional camera:
-
Remove dust cover from 1X relay lens.
-
Thread the camera into the relay lens as shown.
NOTE:
● Always keep one hand on the camera to prevent it from falling. -
Several camera relay lens magnifications are available depending on application and/or camera sensor size.
a. A 1X lens is standard and included with the microscope. This magnification is suitable for cameras with sensor diagonal sizes of 2/3” and larger.
b. A 0.7X lens (optional) will accommodate camera sensors of ½” to 2/3”. Larger sensors may result in images with significant vignetting.
c. A 0.5X lens (optional) accommodates ½” camera sensors and smaller. Larger sensors may result in images with significant vignetting.
Fluorescence Filter Cubes
(EXI-410-FL models only)
SEE PAGES 17-18 FOR EXACT POSITIONING
To install a fluorescence filter cube:
- Remove the cover from the filter cube mounting port on the left side of the microscope.
- Rotate the filter turret to a position that accepts a filter cube.
- If replacing an existing filter cube, remove that filter cube first from the position into which the new filter cube will be placed. Align the filter cube with the guide and groove before inserting. Insert completely until an audible “click” is heard.
- Replace the filter turret cover.
NOTE:
- Fluorescence filter sets must match the fluorescence LED excitation light source and the fluorescence probes used in the application. Please contact with ACCU-SCOPE with any questions about compatibility.
Installing Fluorescence Filter Cubes
Installing Fluorescence Filter Cubes\
- To install a filter cube, align the cube notch with the securing pin on the inside right of the turret receptacle and carefully slide the cube in until it clicks into place.
- Shown here, the filter cube is properly seated and installed.
NOTE
- Never touch any area of the filter cube other than the black casing.
- Be sure to carefully reinstall the turret cover to avoid breakage.
Power Cord
VOLTAGE CHECK
Confirm that the input voltage indicated on the rear label of the microscope
corresponds to your line voltage. The use of a different input voltage than
indicated will cause severe damage to your microscope.
Connecting the Power Cord
Make sure the On/Off Switch is “O” (the off position) before connecting the
power cord. Insert the power plug into the power outlet of the microscope;
make sure the connection is snug. Plug the power cord into a power supply
receptacle.
NOTE: Always use the power cord that came with your microscope. If your
power cord becomes damaged or lost, please call your authorized ACCU-SCOPE
dealer for a replacement.
OPERATION
Powering On
Plug the 3-prong line cord into the microscope power outlet and then into a
grounded 120V or 220V A.C. electrical outlet. Usage of a surge suppressor
outlet is highly recommended. Turn the illuminator switch ① to “―”, then press
the illumination selector ② to toggle the light to on (power indicator ③ will
light). For longer lamp life, always turn the illuminator variable intensity
knob ④ to the lowest illumination intensity setting possible before turning
the power on or off.
Adjusting the Illumination
The light level may need adjustment depending upon the specimen density and
objective magnification. Adjust the light intensity for comfortable viewing by
turning the light intensity control knob ④ clockwise (towards the operator) to
increase brightness. Turn counterclockwise (away from the operator) to
decrease brightness.
Adjusting Interpupillary Distance
To adjust the interpupillary distance, hold the left and right eyetubes while
observing a specimen. Rotate the eyetubes around the central axis until the
fields of view of both eyetubes coincide completely. A complete circle should
be seen in the viewing field when viewing the specimen slide. An improper
adjustment will cause operator fatigue and will disrupt the objective
parfocality.
Where the “●” ① on the eyepiece tube lines up, that is the number for your
interpupillary distance. The range is 5475mm. Make note of your interpupillary
number for future operation.
Adjusting the Focus
To ensure that you obtain sharp images with both eyes, (since eyes vary,
especially for those wearing glasses) any eyesight variation can be corrected
in the following manner. Set both diopter collars ② to “0”. Using your left
eye only and the 10X objective, focus your specimen by adjusting the coarse
adjustment knob. When the image is in view, refine the image to its sharpest
focus by turning the fine adjustment knob. Rotate the diopter collar to obtain
the sharpest focus. To obtain the same sharp image using your right eye, do
not touch the coarse or fine adjustments. Instead, rotate the right diopter
collar until the sharpest image appears. Repeat several times to check.
IMPORTANT: do not counter rotate the focusing knobs as this will cause
severe problems and damage to the focusing system.
Focusing on a Specimen
To adjust focus, rotate the focus knobs on the right or left sides of the
microscope to move the objective up and down. Coarse focus ① and fine focus ②
knobs are identified in the figure to the right.
The figure to the right illustrates the relationship between the rotational
direction of the focus knobs and the vertical motion of the objective.
Focus travel: The default focus travel from the surface of the plain stage is
up 7mm and down 1.5mm. The limit can be increased up to 18.5mm by adjusting
the limit screw.
Adjusting the Focusing Tension
If the feel is very heavy when focusing with the focusing knobs ②③, or the
specimen leaves the focus plane after focusing, or the stage lowers by itself,
adjust the tension with the tension adjustment ring ①. The tension ring is the
inner most ring with the focus knobs.
Turn the tension adjustment ring clockwise to loosen or counterclockwise to
tighten according to user preference.
Using the Stage Plates (Optional)
NOTE: for optimal viewing, ensure the thickness of the container, dish or
slide matches the thickness marked on each objective (0.17mm or 1.2mm). For
modern objectives, coverglass is optimally 0.17mm thick (No. 1½), whereas most
tissue culture vessels are 1-1.2mm thick. A mismatch between slide/vessel
thickness and that for which the objective was designed will likely present an
outof-focus image.
With the mechanical stage ①, a user can use any of the optional stage plates
for flasks, well plates, culture dishes or slides. The figure to the right
illustrates the combination 60mm Petri dish/microscope slide holder ② mounted
in the universal holder of the mechanical stage. The specimen holder can then
be moved by turning the X③ and Y④ stage movement controls.
Selecting the Light Path
The EXI-410 is outfitted with a binocular viewing head with one camera port
for digital imaging. You must select the appropriate light path for observing
and imaging specimens.
When the light path selection slider ① is set to the “IN” position (pushed all
the way in to the microscope), the light path sends 100% of the light to the
binocular eyepieces.
When the light path selection slider is in the “OUT” position (pulled all the
way to the left, away from the microscope), 20% of the light is sent to the
binocular eyepieces and 80% of the light is directed to the camera port for
observation and imaging with a digital camera.
For fluorescence units, the light path is configured for either 100% to the
binocular viewing head (“IN” position), or 100% to the camera port (“OUT”
position).
Using the Aperture Diaphragm
The iris diaphragm determines the numerical aperture (N.A.) of the
illumination system in bright field observation.
When the N.A. of the objective and the illumination system match, you obtain
the optimal balance of image resolution and contrast, as well as an increased
depth of focus.
To check the iris diaphragm: remove the eyepiece and insert the centering
telescope (if you purchased one).
When observing through the eyepiece, you will see the field of view as shown
in the figure on the right. Adjust the iris diaphragm lever to the desired
contrast.
When observing a dyed specimen, set the iris diaphragm ② to 70-80% of the N.A.
of the objective ① in use. However, when observing a live culture specimen
that is not dyed (has virtually no color), set the iris diaphragm to 75% of
the N.A. of the objective in use.
NOTE: An iris diaphragm that is closed too far will give optical
artifacts in the image. An iris diaphragm that is too open may make the image
appear too “washed out”.
Phase Contrast Observation
Depending on the configuration ordered, the EXI-410 can be used for phase
contrast observation with LWD phase contrast objectives: 4x, 10x, 20x and 40x.
For phase contrast observation, replace the normal objectives with phase
contrast objectives on the nosepiece – refer to page 8 for objective
installation instructions. Brightfield observation can still be performed with
phase contrast objectives, but phase contrast observation requires phase
contrast objectives.
Phase Contrast Slider
The adjustable phase slider is pre-aligned at our facility, so further
adjustment is not typically necessary. If the phase ring is not centered, you
can adjust it by centering the bolt with the 2mm hex wrench provided with the
microscope – see instructions below.
The EXI-410-PH includes a 3-position phase slider.
Position 1 is for the 4x objective; Position 2 is for the 10x/20x/40x
objectives. Position 3 is “open” for use with optional filters.
Match the 4x and 10x/20x/40x light annuli with phase contrast objectives of
matching magnifications.
Installing the Phase Slider (Optional) (Refer to Page 14)
Centering the Light Annulus
The phase slider is pre-aligned at our facilities. If realignment is
necessary, follow these steps:
- Place a specimen on the stage and bring it into focus.
- Replace the eyepiece in the eyepiece tube with the centering telescope (optional).
- Make sure the magnification of the objective in the light path matches that of the light annulus on the phase slider.
- While observing through the centering telescope, adjust its focus on the phase annulus ② of the objective and corresponding light annulus ①. Refer to figure on previous page.
- Insert the 2mm hex wrench into the two centering screw holes on the phase slider ③. Tighten and loosen the centering screws until the light annulus is superimposed on the phase annulus of the objective.
- Repeat the steps above to adjust centering with other objectives and corresponding light annuli.
NOTES:
- Halo-like ghost images of the light annulus may sometimes appear. It this occurs, superimpose the brightest light annulus image over the phase annulus.
- When a thick specimen is moved or replaced, the light annulus and the phase annulus may deviate. This is usually due to the amount of media or certain wellplate inconsistences. This can reduce image contrast. If this occurs, repeat steps 1-5 for readjustment.
- The centering procedure may have to be repeated in order to get the best possible contrast if a specimen slide or the bottom surface of a culture vessel is not flat. Center the light annulus using objectives in the order of lower to higher magnifications.
Emboss Contrast Observation
Emboss contrast microscopy requires a condenser-side emboss contrast slider
and eyepiecetube-side emboss contrast slider. These were shipped with the
microscope and the installation and operation instructions are below.
Condenser-side Emboss Contrast Slider
The condenser-side emboss contrast slider is equipped with a sector diaphragm.
Attaching a centering telescope to the eyepiece tube enables you to view a
sector diaphragm image.
You can change the direction of image contrast by rotating the condenser-side
emboss contrast slider adjuster to turn the sector diaphragm.
To use the condenser-side emboss contrast slider, first remove the phase
contrast slider from the condenser.
Then insert the condenser-side emboss contrast slider into the condenser
slider slot ①.
Eyetube-side Emboss Contrast Slider
The eyepiece-tube-side emboss contrast slider has several position markings
corresponding to the objective magnification, and several stop positions to
ensure alignment of the apertures with the light path. For emboss contrast
microscopy, insert the slider into the microscope until it reaches the
position of the same number as the magnification of the objective. To switch
back to brightfield microscopy, pull out the slider up to the hollow position.
Slider position ❶ corresponds with aperture ①, ❷ with ②, and so forth.
For observation without emboss contrast, ensure the condenser-side emboss
contrast slider is in the open position, and the eyetube-side slider is in
position ❶.
Using a Microscopy Camera (Optional)
Installing Couplers (Refer to Page 16)
Selecting the Light Path for Observation/Imaging with a Camera (Refer to Page
21)
Using Fluorescence (EXI-410-FL only)
If you purchased your EXI-410 with fluorescence, your complete fluorescence
system is pre-installed, aligned and tested by our trained technicians to your
specifications prior to shipment.
The complete fluorescence illumination light path includes:
- Integrated LED fluorescence illumination modules
- Dovetail filter slider
- 3 position fluorescence filter turret.
Each position of the filter turret features positive click stop ball-bearing
positioning and printed markings above the knurled wheel identifying the
turret position in the light path.
Refer to pages 8-10 for component diagrams of the EXI-410-FL.
The EXI-410-FL is not available with alternate light sources for fluorescence.
Various filter sets are also available for installation. Selection of filter
sets depends upon the available LED fluorescence modules in your microscope.
Contact your authorized ACCU-SCOPE dealer, or call us at 631864-1000 for a
list of available and recommended filter sets.
Operating Fluorescence (EXI-410-FL only)
Epi-fluorescence illumination
As right figure shown, press the illumination selector button to switch
between epi-fluorescence illumination and transmitted illumination modes.
The intensity of the fluorescence LED illumination will increase when rotating
direction of illumination intensity adjustment knob as in the figure on the
right, the same as when using transmitted LED illumination.
NOTE: To minimize photobleaching of the specimen and avoid
“autofluorescence” from the transmitted LED light module, ensure that the
light shield is rotated into its down position (as shown in the figure to the
right.
Fluorescence Cube Turret
The fluorescence cube turret directs excitation illumination light from the
fluorescence LED unit into the objective. The turret accepts up to three
filter cubes.
Change the filter in the light path by rotating the filter cube turret. When
the filter cube is switched, the fluorescence LED unit is also automatically
switched.
Brightfield positions on the turret are indicated by a symbol and alternate with the three fluorescence filter cube positions. Detents on the turret indicate when a filter cube or brightfield position is engaged. The position of the filter turret is visible on the edge of the turret wheel from both the left and right sides of the microscope. When switching the filter cube, check that the turret clicks at the desired filter cube or brightfield position.
NOTE: a UV light shield is included with the EXI-410-FL version to reduce extraneous light from the fluorescence sample.
TROUBLESHOOTING
Under certain conditions, performance of this unit may be adversely affected
by factors other than defects. If a problem occurs, please review the
following list and take remedial action as needed. If you cannot solve the
problem after checking the entire list, please contact your local dealer for
assistance.
OPTICAL
PROBLEM | CAUSE | SOLUTION |
---|---|---|
The illumination is on, but the field of view is dark. | The LED bulb is burnt |
out. The brightness is set too low.
Too many filters are stacked.| Replace it with a new one.
Set it to the appropriate position.
Reduce them to the minimum required number.
The edge of the field of view is obscured or not evenly illuminated.| The
nosepiece is not in the located position.
The color filter is not inserted fully.
The phase contrast slider is not located in the proper position.| Turn the
nosepiece into the position where you can hear it engaged.
Push it in all the way.
Move the slider until it clicks into place.
Dirt or dust is visible in the field of view.
– Or –
The image has glare.| Dirt/dust on the specimen.
Dirt/dust on the eyepiece.
The iris diaphragm is closed too much.| Clean or replace the specimen.
Clean the eyepieces.
Open the iris diaphragm more.
| The objective is not correctly engaged in the light path.| Turn the
nosepiece into the engaged position.
Visibility is poor
• Image is not sharp
• Contrast is poor
• Details are indistinct| The aperture diaphragm is opened or stopped down too
far in brightfield observation.
The lens (condenser, objective, ocular, or culture dish) become dirty.
In phase contrast observation, the bottom thickness of the culture dish is
more than 1.2mm.
Using a brightfield objective.
The light annulus of the condenser does not match the phase annulus of the
objective.
The light annulus and the phase annulus are not centered.
The objective used is not compatible
with phase contrast observation.
When looking at the edge of the culture dish, the phase contrast ring and the
light ring are deviated from each other.| Adjust the aperture diaphragm
properly.
Clean it thoroughly.
Use a culture dish whose bottom thickness is less than 1.2mm, or use a long
working distance objective.
Change to a phase contrast objective.
Adjust the light annulus so that it matches the phase annulus of the
objectives
Adjust the centering screws to center it.
Please use a compatible objective.
Move the culture dish until you obtain the phase contrast effect. You may
also remove the phase contrast slider, and set the field diaphragm lever to “
”
Phase contrast effect cannot be obtained.| The objective is not in the center
of the light path.
The specimen is not correctly mounted on the stage.
The optical performance of the culture vessel bottom plate is poor (profile
irregularity, etc.).| Confirm that the nosepiece is in the “clicked” position.
Place the specimen on the stage correctly.
Use a vessel with a good profile irregularity characteristic.
MECHANICAL PART
PROBLEM | CAUSE | SOLUTION |
---|---|---|
The coarse adjustment knob is too difficult to rotate. | The tension adjustment | |
ring is tightened too much. | Loosen it appropriately. | |
The image goes out of focus during observation. | The tension adjustment collar | |
is too loose. | Tighten it appropriately. |
ELECTRICAL SYSTEM
PROBLEM | CAUSE | SOLUTION |
---|---|---|
The lamp doesn’t light | No power to the lamp | Check the power cord is |
connected correctly
NOTE: Lamp Replacement
The LED illuminator will provide approximately 20,000 hours of illumination
under normal use. If you should need to replace the LED bulb, please contact
an authorized ACCU-SCOPE service
center or call ACCU-SCOPE at
1-888-289-2228 for an authorized
service center near you.
The light intensity is not bright enough| Not using a designated lamp.
The brightness adjustment knob is not adjusted properly.| Use n designated
lamp.
Adjust the brightness adjustment knob in a correct way.
MISCELLANEOUS
The field of view of one eye does not match that of the other| The
interpupillary distance is not correct.
The diopter is not right.
Your view is not accustomed to the microscope observation and widefield
eyepieces.| Adjust the interpupillary distance.
Adjust the diopter.
Upon looking into eyepieces, try looking at the overall field before
concentrating on the specimen range. You may also find it helpful
to look up and into distance for a moment before looking into the microscope
again.
---|---|---
The indoor window or the fluorescence lamp is imaged.| The stray light enters
through the eyepieces and is reflected to the camera.| Cap/cover both the
eyepieces before imaging.
MAINTENANCE
Please remember to never leave the microscope with any of the objectives or eyepieces removed and always protect the microscope with the dust cover when not in use.
SERVICE
ACCU-SCOPE® microscopes are precision instruments which require periodic servicing to keep them performing properly and to compensate for normal wear. A regular schedule of preventative maintenance by qualified personnel is highly recommended. Your authorized ACCU-SCOPE® distributor can arrange for this service. Should unexpected problems be experienced with your instrument, proceed as follows:
- Contact the ACCU-SCOPE® distributor from whom you purchased the microscope. Some problems can be resolved simply over the telephone.
- If it is determined that the microscope should be returned to your ACCU-SCOPE® distributor or to ACCU-SCOPE® for warranty repair, pack the instrument in its original Styrofoam shipping carton. If you no longer have this carton, pack the microscope in a crush-resistant carton with a minimum of three inches of a shock absorbing material surrounding it to prevent in-transit damage. The microscope should be wrapped in a plastic bag to prevent Styrofoam dust from damaging the microscope. Always ship the microscope in an upright position; NEVER SHIP A MICROSCOPE ON ITS SIDE. The microscope or component should be shipped prepaid and insured.
LIMITED MICROSCOPE WARRANTY
This microscope and its electronic components are warranted to be free from
defects in material and workmanship for a period of five years from the date
of invoice to the original (end user) purchaser. LED lamps are warranted for a
period of one year from the date of original invoice to the original (end
user) purchaser. The mercury power supply is warranted for a period of one
year from the date of invoice to the original (end user) purchaser. This
warranty does not cover damage caused in-transit, misuse, neglect, abuse or
damage resulting from improper servicing or modification by other than ACCU-
SCOPE approved service personnel. This warranty does not cover any routine
maintenance work or any other work, which is reasonably expected to be
performed by the purchaser. Normal wear is excluded from this warranty. No
responsibility is assumed for unsatisfactory operating performance due to
environmental conditions such as humidity, dust, corrosive chemicals,
deposition of oil or other foreign matter, spillage or other conditions beyond
the control of ACCU-SCOPE INC. This warranty expressly excludes any liability
by ACCU-SCOPE INC. for consequential loss or damage on any grounds, such as
(but not limited to) the non-availability to the End User of the product(s)
under warranty or the need to repair work processes. Should any defect in
material, workmanship or electronic component occur under this warranty
contact your ACCU-SCOPE distributor or ACCU-SCOPE at
631-864-1000. This warranty is limited
to the continental United States of America. All items returned for warranty
repair must be sent freight prepaid and insured to ACCU-SCOPE INC., 73 Mall
Drive, Commack, NY 11725 – USA. All warranty repairs will be returned freight
prepaid to any destination within the continental United States of America,
for all foreign warranty repairs return freight charges are the responsibility
of the individual/company who returned the merchandise for repair.
ACCU-SCOPE is a registered trademark of ACCU-SCOPE INC., Commack, NY 11725
ACCU-SCOPE®
73 Mall Drive, Commack, NY 11725
631-864-1000 (P)
631-543-8900 (F)
www.accu-scope.com
info@accu-scope.com
v071423
Documents / Resources
| ACCU
SCOPE EXI-410 Series Inverted
Microscope
[pdf] Instruction Manual
EXI-410 Series Inverted Microscope, EXI-410, Series Inverted Microscope,
Inverted Microscope, Microscope
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>