Labco Scientific 550.001.005 Pipette Filler Electronic User Manual
INTRODUCTION
The Pipette Filler has been carefully designed and manufactured as a lightweight, cordless Pipette Filler. Pipette Filler offers simple, efficient pipetting performance and maximum pipetting comfort. It can be used with all plastic or glass pipettes between 0.1-100ml.
PRODUCT CODE | DESCRIPTION |
---|---|
550.001.005 | Pipette Filler Electronic |
550.001.022 | Filter for Pipette Controller Electronic 0.45um |
WARRANTY
The Pipette Filler is warranted for one year against defects in materials and
workmanship (Battery is warranted for three months only). Should your Pipette
Filler fail to function according to specifications at any time, please
contact the local representative immediately. The warranty will not cover
defects caused by normal wear or by a breach of the manufacturer’s
recommendations.
Each Pipette Filler is tested before shipping by the manufacturer. The Quality
Assurance Procedure guarantees that the Pipette Filler you have purchased is
ready for use.
DELIVERY
The Pipette Filler package contains 1 x Pipette Filler, AC-adapter, 0.45µm hydrophobic filter, User Manual, Quality control certificate, Wall-mounted Holder & a Two-position stand.
CHARGING THE PIPETTE FILLER
- Before the first use the battery must be charged for 10 hours. Allow the unit to charge anywhere from 2 to 3 hours for all subsequent recharging. More than 8 hours of intermittent use is typical when fully charged.
- Before connecting the AC-adapter to the electrical outlet, check that its voltage setting corresponding to the voltage of your electrical outlet.
- Connect the AC-adapter to the electrical outlet and to the electrical fitting at the base of the handle of the Pipette
Filler (Fig. 1).
Note: The Pipette Filler should only be charged with the AC-adapter
provided.
An icon on the LCD display indicates the level of battery charge – the more
bars, the greater the charge.
Special circuitry prevents overcharging of the battery and allows the Pipette
Filler to be charged or partially charged at any convenient time.
ELECTRICAL SPECIFICATIONS
Battery: 3.7 V, 1100mA, rechargeable lithium-ion battery. Charging time
10 hours for completely drained batteries.
ACadapter: Input AC 100-240V Output voltage 4.2V, 500mA.
PIPETTE FILLER OVERVIEW (FIG. 2)
PIPETTE FILLER DESCRITION
Operating buttons
The Pipette Filler has two operating buttons located at the front of the
handle. The upper button enables aspiration for as long as the button is held
down. The lower button allows for dispensing. The aspiration and dispensing
speeds can be finetuned during operation by applying varying finger pressure
on the operating buttons.
Speed adjustment button
The speed adjustment button is located at the upper part of the handle. The
desired speed is set by pushing the speed adjustment button.
Nose cone and silicone adaptor
The pipette adapter provided with the Pipette Filler allows the use of all
standard plastic and glass pipettes (graduated and non-graduated) including
Pasteur pipettes. The pipette adapter consists of the nose cone and silicone
adapter. The silicon adapter enables a pipette to fit securely in place, thus
creating a proper vacuum level and prevents leakage.
Filters
Removing the nose cone and silicone adaptor exposes a hydrophobic filter (0.45
micron), located at the top of the nose piece. The filter protects the Pipette
Filler from accidental over-aspiration. It is strongly recommended to use only
the original filters supplied by the manufacturer with the Pipette Filler.
Note: the Pipette Filler cannot be operated without a filter. Wetting of
filter due to overfilling will slow or prevent fluid movement. If overfilling
occurs, change the filter immediately.
OPERATING THE PIPETTE FILLER
Setting the operating speed
The button located under the digital display at the upper part of the handle,
allows you to set your aspiration and dispensing speeds
(Fig. 3).
Eight speeds are available for each function. Simply push + for faster and –
for slower speeds. Your selection will be displayed on the LCD screen.
The Pipette Filler can self-regulate the aspiration and dispensing speeds by
circuit feedback.
Note: The aspiration and dispensing speeds can be fine-tuned during
operation by applying a varying amount of finger pressure to the operating
buttons.
Inserting a serological pipette
Gently push a serological pipette into the silicone adapter and ensure that it fits securely. To insert the serological pipette safely, hold the Pipette Filler firmly by the nose cone and the serological pipette from the top end of the tube.
Aspiration
- Select a speed and fill the pipette by pressing the aspirating button. Fill the pipette just above the desired level, check level by the meniscus.
- Remove the serological pipette from the sample.
- Adjust the volume of the sample by pressing the dispensing button until the meniscus is at the required level.
Dispensing
The Pipette Filler offers two different dispensing modes:
- If the dispensing button is pressed only partially to a point where a check can be felt, the dispensing function will be based on the force of gravity and the liquid will simply flow out.
- Pressing the dispensing button further in will engage the motor and activates the blow-out function.
Note: Finger pressure can be varied to fine-tune the speed within each mode.
Other pipetting recommendations
- Use the instrument within the limits of its chemical resistance.
- Never use with highly flammable liquids.
- Observe the reagent manufacturer’s safety information.
- Never use the instrument in an atmosphere with a danger of explosion.
- Never use force with the Pipette Filler.
- Should the instrument fail to function properly, stop pipetting and check the trouble shooting section.
MAINTENANCE
The Pipette Filler has been designed for easy in-house maintenance.
Decontamination
To decontaminate, lightly spray the outside of the Pipette Filler with the
decontamination spray and wipe dry with a clean cloth.
Changing the nose cone, silicone adapter and filter
(Fig. 4)
To change the nose cone, silicone adapter and filter firmly hold the nose
cone by the ribbed collar, rotate the nose cone counter-clockwise until
completely loose and pull out the nose cone and the silicone adapter. The
filter can be removed by simply pulling it out once the nose cone and the
silicone adapter have been disengaged. Position the new filter with the
coloured side upwards. To reassemble, insert the adapter and nose cone and
twist the nose cone gently clockwise.
Note: It is recommended to check and clean the silicone adapter every six months, or when necessary, to prevent a loss of vacuum. If there is a loss of vacuum, please contact the manufacturer or your local distributor.
The Pipette Filler cannot be operated without a filter. Wetting of filter due to overfilling will slow or prevent fluid movement. If overfilling occurs, change the filter immediately. Only use filters supplied by the manufacturer.
Changing the battery (Fig. 5 and Fig. 6)
If the battery needs be replaced follow these steps to replace the battery:
- Take the lid off the battery compartment to expose the battery compartment
(Fig. 5).
-
Disconnect the plug of the Lithium-ion battery from the Printed Circuit Board (PCB) connector inside the battery compartment exposing a 2pin connector on the PCB
-
(Fig. 6).
-
Connect the 2 pin plug of the new Lithium-ion battery to the exposed 2 pin connector on the PCB. The side of the plug with the exposed metal tabs should face away from the
PCB,
Fig 7. If the battery is connected with the plug in the wrong orientation, you will not see a display on the LCD screen. -
Replace the battery compartment lid.
-
Dispose of the battery appropriately.
Note: Only use the original battery supplied by the manufacturer
Fig.8
Side with expose metal tabs
STORAGE
The Pipette Filler should not be stored for a long period of time without removing the pipette. It is recommended to disconnect the Pipette Filler from the AC-adapter if the unit is not used for a lengthy period of time.
TROUBLE SHOOTING
PROBLEM | CAUSE | SOLUTION |
---|---|---|
REDUCED ASPIRATION CAPACITY OR THE PIPETTE FILLER IS BLOCKED | -Filter blocked, | |
dirty or damaged-Battery low | -Replace the filter-Recharge | |
SEROLOGICAL PIPETTE DRIPS | -Serological pipette not securely attached-Silicone | |
adapter or nose cone loose-Silicone adapter or nose cone damaged | -Reattach |
serological pipette-Reassemble silicone adapter or nose cone-Replace silicone
adapter or nose cone
REDUCED OPERATING TIME| -Battery worn| -Replace battery
NO DISPLAY| -Dead battery-Battery plugged in the wrong orientation-Failure of
the Printed Circuit Board| -Replace battery-Refer to section 6.3 of manual to
plug battery in correctly.-Call supplier for Tech Support