midmark 153637 Ceiling Mounted Dental Light Instructions
- June 13, 2024
- Midmark
Table of Contents
- midmark 153637 Ceiling Mounted Dental Light
- Product Information
- Product Usage Instructions
- Equipment Alert
- Wood Frame Ceiling
- Suspended Ceiling
- Halogen Shown
- Step 1: Marking hole location.
- Step 2: Routing power line.
- Step 3: Secure ceiling plate.
- Step 4: Wiring Installation.
- Step 5: Securing power line to ceiling plate.
- Step 6: Installation of suspension tubing.
- Step 7: Installation of suspension tubing.
- Step 8: Leveling suspension tubing.
- Step 9: Installation of Ceiling Cover.
- Step 10: Installation Light Arm.
- Step 11: Installation Fixed Arm.
- Step 12: Installation of Stop Ring & Pin.
- Step 13: Verify Operation of Light.
- Tension Adjustments
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
midmark 153637 Ceiling Mounted Dental Light
Product Information
The Ceiling Mounted Light is a lighting system designed for medical facilities. It is available in both Halogen and LED styles. The product is suitable for use with Models 153637 (-001) and 153905 (-001). The maximum weight capacity of the lighting system is 70 lbs (31.7 kg).
Installation Note
It is recommended to mount the light on the opposite side of the doctor’s normal work position.
Equipment Alert
The orientation of the ceiling plate is crucial to ensure proper range of motion. Rotate the ceiling plate until the strain relief bracket is positioned as shown.
Ceiling Support Structure
Note that Midmark does not supply the ceiling support structure.
Consult a qualified professional to install the appropriate support structure
for your facility. The illustrations provided in the manual are for suggestion
only.
Ceiling Plate Deflection
The ceiling structure must support the light system without deflection. To prevent the suspension arm from drifting, the ceiling plate must deflect no more than 1/16 in (1.6 mm) over a 12 in (30.5 cm) distance.
Product Usage Instructions
Step 1: Mounting the Light
- Choose the appropriate side to mount the light, opposite to the doctor’s normal work position.
- Rotate the ceiling plate until the strain relief bracket is positioned as shown in the illustrations.
- If using a wood frame ceiling, follow the suggested installation method using joists and nails or consult a qualified professional for guidance.
- If using a suspended ceiling, refer to the provided illustration for the suggested ceiling support structure using metal masonry anchors and bolts.
Step 2: Power Supply
- Ensure a suitable power supply is available near the installation location.
- Connect the light system to the power supply following the manufacturer’s instructions.
Step 3: Adjusting the Suspension Tube
The suspension tube length may need to be adjusted based on your specific installation requirements. Refer to the table below for suggested lengths:
Light Style | Suggested Suspension Tube Length |
---|---|
Halogen | 96 in. (243.8 cm) to 132 in. (335.3 cm) |
LED | >73 in. (>185.4 cm) with suggested suspension tube of 50 |
cm
Adjust the suspension tube length accordingly to achieve the desired positioning of the light.
For any further assistance or clarification, please refer to the complete user manual provided with the product.
Note
Midmark recommends to mount the light on the opposite side of the doctor’s
normal work position.
Equipment Alert
The orientation of the ceiling plate is crucial to ensure proper range of motion. Rotate the ceiling plate until the strain relief bracket is positioned as shown.
NOTE: This install guide applies to Halogen and LED-style lights.
Note
Midmark does not supply the ceiling support structure. The illustrations below
are for suggestion only. Consult a qualified professional to install the
appropriate support structure for the facility.
Wood Frame Ceiling
Suspended Ceiling
Lighting Systems (ONLY)
-
Weight (max.)
70 lbs. (31.7 kg.) -
Torque (max.)
70 ft-lbs. (94.9 N•m)
Note
The ceiling structure must support the light system without deflection. To
prevent the suspension arm from drifting, the ceiling plate must deflect no
more than 1/16 in. (1.6 mm) over a 12 in. (30.5 cm) distance.
Halogen Shown
Suspension Tube Recommendations
Mounting Height| Suspension Tube Part Number|
Suspension Tube Length
96 to 98 in. (243.8 to 248.9 cm)| 030-1845-00-216| 11.71 in. (29.7cm)
99 to 102 in. (251.5 to 259.1 cm)| 030-1845-01-216| 15.71 in. (39.9 cm)
103 to 106 in. (261.6 to 269.2 cm)| 030-1845-02-216| 19.71 in. (50.1 cm)
107 to 110 in. (271.8 to 279.4 cm)| 030-1845-03-216| 23.71 in. (60.2 cm)
111 to 114 in. (281.9 to 289.5 cm)| 030-1845-04-216| 27.71 in. (70.4 cm)
115 to 118 in. (292.1 to 299.7 cm)| 030-1845-05-216| 31.71 in. (80.5 cm)
119 to 122 in. (302.3 to 309.9 cm)| 030-1845-06-216| 35.71 in. (90.7cm)
123 to 126 in. (312.4 to 320.0 cm)| 030-1845-07-216| 39.71 in. (100.9 cm)
127 to 130 in. (322.6 to 330.2 cm)| 030-1845-08-216| 43.71 in. (111 cm)
130 in. ( >330.2 cm)| 030-1845-09-216| 47.71 in. (121.2 cm)
WARNING
Disconnect facility power before performing this installation. Failure to
comply could result in serious personal injury.
Step 1: Marking hole location.
- A) Position ceiling plate or template (supplied on carton.)
- B) Mark three mounting holes and power line hole.
- C) Drill holes.
Note
The customer must provide the wiring to power the lighting system. The
lighting system requires 115/230 VAC input voltage. Wiring should be two
conductor w/ground, 14 gauge copper, rated for 300 VAC (mini-mum). Connect and
route wiring according to state and local codes.
Step 2: Routing power line.
- A) Route power line thru ceiling access hole.
- B) Pull power line thru strain relief bracket on ceiling plate.
Step 3: Secure ceiling plate.
- A) Align ceiling plate to ceiling support structure.
- B) Secure ceiling plate via 1/2” holes with appropriate hardware [not provided].
Step 4: Wiring Installation.
- A) Use label for insert wires from power line into terminal block.
- B) Secure wires with terminal block screws.
Equipment Alert (Halogen ONLY)
If ceiling light unit is being powered by 230 VAC (nominal), disconnect wire
from 115 VAC transformer terminal
and reconnect to 230 VAC transformer terminal.
Step 5: Securing power line to ceiling plate.
- A) Secure facility power line to ceiling plate with cable clamp.
- B) Tighten strain relief screw(s) to secure power line.
Note: Assure adequate facility line slack prior to mounting.
Step 6: Installation of suspension tubing.
- A) Screw suspension tube into mounting flange until tight, and the top of mount and top of collar are at least flush.
- B) Secure suspension tube in mounting flange with set screws.
- C) Check that tube will not unscrew after set screws are tightened.
Note: If suspension tube is 26 in. (66 cm) or longer, connect wire harness extension(s) [included w/suspension tube] before performing Step 7.
WARNING
These are easily cross-threaded.Ensure suspension tubes are fully
threaded in to castings and all set screws are installed in place and securely
tightened. Failure to do so could allow monitor or light tubes to separate
from trolley castings, resulting in serious personal injury to patient or
staff.
Step 7: Installation of suspension tubing.
- A) Route wire harness(es) down thru suspension tube while sliding mounting flange onto three bolts.
- B) Install three lockwashers and three hex nuts.
Step 8: Leveling suspension tubing.
- A) Place level on suspension tube.
- B) Adjust hex nuts to level suspension tube.
- C) Tighten jam nuts to secure mounting flange.
Step 9: Installation of Ceiling Cover.
- A) Slide ceiling cover onto suspension tube.
- B) Check to make sure ceiling cover is flush with mounting plate then tightening set screws.
Caution – Equipment Alert
To ensure quiet operation, apply a thin coat of lubricant
(included w/light) to inside of suspension tube and bearing shaft of fixed arm
before performing the following steps.
Step 10: Installation Light Arm.
- A) With rounded edge of collar upward, slide collar onto suspension tube.
- B) Tighten set screw (1) to temporarily hold collar in place.
- C) Fold excess length of wire harness and push up into suspension tube as shown.
Step 11: Installation Fixed Arm.
- A) Connect wire harness to connector in the end of fixed arm.
- B) Insert fixed arm into suspension tube, then install key thru hole so that it fits into groove of fixed arm.
- C) Loosen set screw (1) and slide collar all the way down, and tighten set screw (1).
Step 12: Installation of Stop Ring & Pin.
- A) Position tab of rotation stop ring in line with flex arm so that it is facing lighthead.
- B) Insert flex arm shaft into curved arm assembly and secure with set screw (2).
- C) Lift cover ring and install stop pin into hole in curved arm assembly.
Step 13: Verify Operation of Light.
Note: If necessary, refer to User’s Manual for operation.