3nh YG60L Gloss Meter Instruction Manual
- June 13, 2024
- 3NH
Table of Contents
Non-contact
Gloss Meter
OPERATION MANUAL
Please read this manual carefully before using the instrument
Overview
The non-contact glossmeter can measure without contacting the sample and will not pollute the sample. It 1s also easy to install and fast to measure. It is very suitable for automatic production line glossiness detection. It can find defective products in time and reduce production costs.
External Structure Description
Button name | Icon | Function |
---|---|---|
Measurement button | Switch ON/OFF, measurement, quick return to measurement |
interface, operation confirmation
OK button| | Display menu, prompt or operation confirmation
Return button| | Return to previous menu or cancel operation
Up button| | Move up menu items
Down button| | Move down menu items
Remark: Use the above icons to represent the corresponding buttons in the instrument operation prompt
Measurement Principle and Introduction
This product conforms to the international standard ISO 2813 and the Chinese
national standard GB/T9754, and acts the requirements of JJG696 working gloss
meter. Under normal use conditions, Non-contact gloss meter is the same as
our company’s gloss meter. With automatic calibration, it has four measurement
modes: basic mode, quality control mode, statistical mode and continuous mode
to meet different measurement needs. At the same time, it has quality
management software.
The special design of Non-contact gloss meter enables the instrument to carry
out contact measurement and non-contact measurement with an interval of 10mm.
As shown in the figure below, the distance between the calibration plate and
the sample to be tested should be 10mm from the Non-contact gloss meter gloss
basic plate. When the instrument 1s connected to a 10mm calibration bracket
(refer to Figure 1), the sample to be tested can be attached to the
calibration bracket to achieve contact measurement. When the instrument is not
connected to the 10mm calibration bracket (refer to Figure 1), the measured
sample is 10mm away from the bottom of the instrument to achieve non-contact
measurement.
There are two sampling procedures for Non-contact gloss meter. One is, 1n the
normal measurement mode, that the lighting source is turned on, the sample to
be tested is illuminated, the detector signal is collected, the lighting
source is turned off, and the glossiness of the sample to be tested is output.
The single measurement time is 1 second; the other is that when the online
measurement mode is turned on, the lighting source will be turned on all the
time, and the detector will collect and output signals uninterruptedly, so
that the measurement period can be as short as possible, and the single
measurement period is 180 milliseconds.
Switch on/off
Startup
The instrument is powered by an external power adapter through the USB
interface (dry batteries can also be used in special cases). When using
battery power, press the measurement button to turn on; if using USB power,
it will automatically turn on when the USB is plugged in and connected to the
power supply.
If the calibration plate is detected during startup, the instrument will be
calibrated automatically.
If the calibration 1s passed, the interface shown in Figure 2 will be
displayed, and at this time, press the measurement button to enter the
measurement interface.
If the calibration is failure, it is as shown in Figure 3.
After the calibration is failure, please choose to press the measurement button to retry or press the return button to directly enter the measurement interfacc. If you try again, please check these first:
- Whether the calibration plate 1s in good connection or not;
- Is the calibration tile clean?
- Calibration tile protection paper is removed.
Shutdown
When using battery power, the instrument will automatically shut down within a
certain period of time (30 seconds by default, which can be changed in the
settings).
It can also be turned off by pressing and holding the measurement button for
more than 3 seconds.
It is not able to turn off when using USB power.
Basic Operation
This chapter introduces some basic operations, which are the basis of the following chapters, and it is best to browse through them.
Menu Operation
Click the OK button on the measurement interface to enter the menu interface.
In the menu interface, the item displayed in reverse color indicates the
currently selected menu item.
Press the up and down buttons to change the currently selected item.
If there is a down arrow ( ) in the lower right corner, it means that there
are more menu items that need to be turned up to see; if there 1s an up arrow
(f) in the upper right corner, it means that there are more menu items that
need to be turned up to see.
If there is a triangle symbol (►) on the right side of the menu item, it means
that you can enter the submenu item by clicking the OK button.
Tip 1: Normally press the measurement button on the menu interface to return
to the measurement interface.
Enter the Main Menu
Press the OK button in the measurement interface to enter the menu of the corresponding mode, and select “Main Menu”, then press the OK button to enter the main menu.
The main menu contains the following items:
- Mode: Select the measurement mode;
- Angle: Select the measurement angle (only for multi-angles gloss meter);
- Data: Turn on/off storage function; check and delete records;
- Standard management: Turn on/off the difference value; check, modify, add and delete the standard;
- Calibration: Calibrate the instrument and modify the calibration parameters;
- Setting: System setting.
Confirm Prompt Interface
The prompt interface is used to prompt to confirm an operation or not like deleting the record as shown in Figure 6.
If you want to confirm the corresponding operation, press the OK button or the
Measure button.
Press the return button if you want to cancel the operation.
Tip 2: If confirming “No” in the prompt interface means pressing the return
button to perform a “No” operation, and “Yes”’ means pressing the OK button to
perform a “Yes” operation.
Input Prompt Interface
The input interface is mainly used for prompting input the sample name or number as shown in Figure 7.
As shown in Figure 7, between the square brackets (between “[” and “]”) is the
input area, and the character on “” is the current input focus.
Press the up or down button to switch the character at the focus position.
Press the OK button to move the input focus to the next position.
Press the measurement button to confirm the entry.
Press the return button to cancel the input operation.
Editing Opcration
The number or character displayed on the right side of some menu items 1s editable. For example, in the item of the average number of statistical modes, when press the OK button, it will display as shown in Figure 9, which 1s an edit area enclosed by square brackets. When display the editing area as shown in Figure 7, press the up and down buttons to switch the character at the focus, and press the OK button to move the edit focus to the next character, or press the return button to cancel the changes and press the measurement button to save the changes.
Figure 8 The number of average measurement is editable
Figure 9 Press OK button to edit the number of average times
Display Item
Figure 10 Display items in edit stats mode will show option box
Some setting values can only be one of several options. For example, in the
display item of the statistics mode, press the OK button to select other
options.
When you edit an item, it is anti-color display and the current item are
enclosed in square brackets.
Press the up or down button to switch the current option.
Press the OK or measurement button to save the changes.
Press the return button to restore the original value.
Instrument Calibration
The calibration plate is automatically detected and calibrated when the
instrument is turned on.
In order to the accuracy of measurement, it has to re-calibrate it after
changing the power supply mode (for battery to USB or from USB to battery
power), and it also needs to re-calibrate the instrument after modifying the
calibration parameter. If the environment changes too much, it 1s also better
to re-calibrate the instrument.
The related functions are under the Instrument Calibration menu.
Enter the main menu and select “Instrument Calibration” then press the OK
button to enter the Instrument Calibration menu.
If you are in the basic measurement mode, you can also find the “calibration”
menu item in the basic mode menu.
Calibrate the Instrument
Before calibration, please confirm that the calibration tile 1s clean and the
calibration plate is in good connection.
During use, enter the “Calibration” menu to calibrate the instrument. Then
select “Calibrate” inthe “Calibration” menu as shown in Figure 11 and press
the OK button.
Figure 11 Press OK button after selecting “Calibrate”’
Figure 12 Press measurement button to start calibrate
This will enter the “Calibration” interface, and then press the measurement or
OK button to start calibrate the instrument as shown in Figure 12.
When the instrument is calibrating, it will display “Calibrating” and flashes
green light.
After the calibration is completed, if the calibration is successful, it is as
shown in Figure 13. At this time, press the measurement button to return to
the measurement interface or press the return button to return to the
“Calibration” menu. If the calibration is failure, it is as shown in Figure
14. Press the measurement button or the OK button to retry, or press the
return button to cancel it.
Figure 13 Calibration pass
Figure 14 Calibration fail
Modify Calibration Parameters
Please modify the calibration plate parameters in the instrument simultaneously if replace or re-calibrate the calibration tile.
Figure 15 Modify the calibration board parameters
Please enter the “calibration” menu to modify the calibration tile parameters,
then select the “modify parameters” menu item and press the OK button to enter
the “modify parameters” interface.
Select the angle to be modified and press the OK button to start editing the
corresponding value. Press the measurement button to save the changes after
finishing the editing
Note: The random change of the calibration tile parameters will lead to the
inaccuracy of measurement. Please consult the factory to re-calibrate the
calibration tile or replace the calibration plate.
Measurement Modes
Switching Measurement Modes
The instrument is in the basic mode when it is turned on for the first time. Please enter the “main menu” and select the “measurement mode” menu item, then press the OK button to the “Mode” interface to change the measurement mode as shown in Figure 16. Select the appropriate measurement mode and press the OK button.
Figure 16 Measurement mode menu
Basic Mode
6.2.1 Measurement
Press the measurement button under the basic mode interface to measure.
During the measurement, the green indicator lights up, while the green
indicator goes out after the measurement is completed and display the
measurement results as shown in Figure 17.
Figure 17 Basic mode measurement results
Figure18 Enter the sample name
If the storage function is turned on, it prompts to enter the sample name
after each measurement is completed as shown in Figure 18. The system will
provide a default name and press the measurement button to confirm the name
and save it or press the return button to cancel it.
Press the return button to delete the measurement result as shown in Figure
19.
Figure 19 Press the return button to delete the measurement result
6.2.2 Basic Mode Menu
Press the OK button at the measurement interface to enter the basic mode menu.
Figure 20 Select “Main Menu” and press the OK button to enter the main menu
The basic mode menu contains the following menu items:
- The main menu: Enter the main menu;
- Delete: Delete the current measurement result;
- Storage: Enable or disable the storage function. If the storage function is turned on, “ON” will be displayed on the right, otherwise “OFF” will be displayed.
- Angle: Switch the instrument measurement angle (only for multi-angles instrument);
- Calibration: Functions related to instrument calibration, see 5 Instrument Calibration;
- Setting: Enter the system setting menu, see 10 System Setting.
Quality Control Mode
6.3.1 Interface Introduction
Figure 21 Figure 20 Select “Main Menu” and press the OK button to enter the main menu
The standard interface of the quality control mode is shown in the figure
above, the upper left corner is the name of the current standard sample, and
the upper right corner is the name of the current sample.
As shown in Figure 21, the “value” column shows the measured value; the “diff”
column shows the difference between the sample value and the standard value;
the “pass” column shows judgement results, and “OK” means within the tolerance
of the standard, “NG” indicates out of tolerance.
If the difference display is not turned on, the measurement interface of the
quality control mode is the same as the basic mode.
6.3.2 Measurement
Press the measurement button in the measurement interface to measure. The
green light will light up during measurement, and will go out after the
measurement is completed and the measurement result will be displayed.
6.3.3 Measure Standard
Press the OK button to enter the quality control mode menu, then select the
measurement standard. After the measurement is ready, press the OK button to
measure.
After the measurement is completed, the measurement result is displayed and
prompted for confirmation as shown in Figure 22.
Figure 22 Confirm the measurement result of standard
Figure 23 Enter the sample name
Press the OK button to confirm or press the return button to discard the
result.
If the storage function is turned on, it prompts to enter the standard name
after the confirmation is finished as shown in Figure 23.
After confirming the name of the standard, press the measurement button to
save the standard or press the return button to cancel the storage. At this
time the standard will be save as a temporary standard.
After the standard 1s measured, it is automatically set as the current
standard; if the differential display is not turned on, it will be
automatically turned on.
Statistical Mode
6.4.1 Interface Introduction
As shown in Figure 24, in the top left corner of the statistics mode, it
displays the current standard name. The first row in the upper right corner
shows the current sample name, and the second row shows the number of
measurements, which is displayed in the form of “m/n”. “m” indicates the
number of measured times and “n” indicates the total numbers of requirement
measurement.
In the figure 24, “avg” , “range” , “stdev” each shows the average value, the
variation range of m measurements, and the standard deviation. They can be
changed in the Display Options in the Statistics Mode menu.
Figure 24 Statistics Mode Measurement Interface
6.4.2 Measurement
After entering the statistical mode measurement interface, press the
measurement button to measure. The measurement result will be updated at each
measurement, and measurement will be automatically performed after the
specified number of measurements. If the storage is turned on, it will prompt
to enter the sample name when measurement 1s finished.
Please press the down button to finish the current measurement if the setting
numbers of measurements are not enough.
Please press the up button (this operation can only be performed when the
measurement is not completed) to delete the previous measurement data.
Please press the return button to delcte the mcasurement result.
6.4.3 Measure Standard
Press the OK button to enter the statistical mode menu, then select the
“Measure Standard” menu item, and press the OK button to enter the standard
measurement interface.
The “m/n” in the top left corner of the measurement interface shows the
current number of measurements and the total number of required measurements.
The “avg” column in the measurement interface shows the average of m
measurements; the “min” column shows the minimum valuc in m measurements, and
“max” shows the maximum value in m measurements.
Figure 25 Statistics mode measurement standard interface
After entering the standard measurement interface, press the measurement
button to measure. After measuring for a specified number of times, if the
storage 1s turned on, it will be prompt to input the standard name, otherwise
it is set as a temporary standard.
During the measurement, press the return button to cancel the measurement or
press the OK button to complete the measurement.
6.4.4 Set the Number of Measurements
Enter the statistics mode menu and select the “Average” menu item, then press
the OK button to modify the number of measurements.
The number of measurements in the statistical mode can be set from 2 to 99.
6.4.5 Setting Display Options
After entering the statistics mode menu, select the “Display Options” menu
item, and then press the OK button to enter the display options setting page.
The statistical mode measurement interface can display up to three columns,
corresponding to the first, second and third items respectively.
The available display items are:
- “ave”: Avcrage;
- “min” : The minimum value of the measured value;
- “max”: The maximum value of the measured value;
- “range” : The value that the maximum value minus the minimum value;
- “stdev” : Standard deviation;
- “diff” : Difference that minus the standard deviation;
- “pass”: The result of comparison with the standard sample, “OK” or “NG”;
- Blank: no display.
The standard deviation calculation formula is:
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>