IET LABS HARS-X High Accuracy Resistance Substituter User Manual
- June 13, 2024
- IET LABS
Table of Contents
♦ PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT ♦
HARS-X and X2
High Accuracy Resistance
Substituter
User and Service Manual
HARS-X High Accuracy Resistance Substituter
Copyright © 2018 IET Labs, Inc.
HARS-X im/October 2018
HARS-X
WARRANTY
We warrant that this product is free from defects in material and workmanship
and, when properly used, will perform in accordance with applicable IET specifi
cations. If within one year after original shipment, it is found not to meet
this standard, it will be repaired or, at the option of IET, replaced at no
charge when returned to IET. Changes in this product not approved by IET or
application of voltages or currents greater than those allowed by the specifi
cations shall void this warranty. IET shall not be liable for any indirect,
special, or consequential damages, even if notice has been given to the
possibility of such damages. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
HARS-X and X2
WARNING
OBSERVE ALL SAFETY RULES
WHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES.
Dangerous voltages may be present inside this instrument. Do not open the case
Refer servicing to qualifi ed personnel
HIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENT WHENEVER
HAZARDOUS VOLTAGES (> 45 V) ARE USED, TAKE ALL MEASURES TO AVOID ACCIDENTAL
CONTACT WITH ANY LIVE COMPONENTS.
USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS
INSTRUMENT.
Use extreme caution when working with bare conductors or bus bars.
WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS AND KEEP UNREQUIRED
PERSONNEL SAFELY AWAY.
CAUTION
DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THIS INSTRUMENT
IN EXCESS OF THE MAXIMUM LIMITS INDICATED ON THE FRONT PANEL OR THE OPERATING
GUIDE LABEL.
Figures and Tables
Figure 1-1. HARS-X Series High Accuracy Resistance Substituter ……….1
Figure 1-2. Single-Decade HARS Unit ……………………………………………..1
Figure 2-1. Typical Operating Guide Affi xed to Unit ………………………….4
Figure 3-1 Optimal 4-Wire Kelvin Lead Connection …………………………..6
Figure 4-1. HARS Series Schematic Diagram ……………………………………8
Table 4-2: Replacement List …………………………………………………………….9
Figure 4-2. HARS-X Replaceable Parts …………………………………………….9
Chapter 1 INTRODUCTION
1.1 Introduction
The High-Accuracy Resistance Substituter (HARS) series of resistance decade
substituters is a family of instruments providing a very broad choice of high-
performance resistance sources. Any number ofdecades from one to eleven is
available in a choice
of accuracies. See Figures 1-1 and 1-2.
The HARS substituter is a precision resistance sourcewith excellent
characteristics of stability, temperature coeffi cient, power coeffi cient, and
frequency response.
The HARS Series employs very-low-resistanceswitches with silver-alloy
contacts. A special design keeps zero resistance to less than 1 mΩ per decade.
Self cleaning keeps the silver contacts from becomingtarnished when unused, or
when only low currents are passed through them. This is most often the case
when only minute test currents are drawn by digital multimeters or other test
instruments. Contact resistance is stable and remains low and repeatable.
High-quality gold-plated tellurium-copper binding posts serve to minimize the
thermal emf effects whichwould artifi cially refl ect a change in dc resistance
measurements. All other conductors within the instrument, as well as the
solder employed, contain no
metals or junctions that could contribute to thermal emf problems.
The standard models offer a choice of one through eleven decades. The panels
are clearly labeled showing the step size and maximum voltage and current
limitations for each decade.
With a resolution as low as 1 mΩ and a maximum available resistance of over
111 MΩ, the HARS series may be used for exacting precision measurement
applications requiring high accuracy, good stability, and low zero-resistance.
They can be used as compo- nents of dc and ac bridges, for calibration, as
transfer standards, and as RTD simulators.
Single-decade units may be panel-mounted and combined with additional units to
form po ten ti om e ter cir cuits or other con fig u ra tions.
The larger units may be rack-mounted to serve as components in measurement and
control systems.
Chapter 2 SPECIFICATIONS
For convenience to the user, the pertinent specifi cations are given in an
OPERATING GUIDE affi xed to the case of the instrument. Figure 2.1 shows a
typical example.
SPECIFICATIONS
Resistance per step
| Total decade resistance| ****
Stability (±ppm/yr)
| Long-term stability (±ppm/3 yrs)| Temperature coefficient (±ppm/°C)| ****
Resistor type
| HARS-X| HARS-X2
---|---|---|---|---|---|---|---
Max current| Max
voltage (per step)
| Max power (per step)| Max current| Max
voltage (per step)
| Max power (per step)
1 m W| 10 mW| 50| 75| 50| ****
Resistance wire
| 8.0 A| 5 mV| 0.04 W| 9.0 A| 9 mV| 0.08 W
10 m W| 100 mW| 50| 75| 20| 4.0 A| 40 mV| 0.16 W| 6.3 A| 63 mV| 0.4 W
100 m W| 1 W| 50| 75| 20| 1.6 A| 0.16 V| 0.25 W| 2.2 A| 0.3 V| 0.5 W
1 W| 10 W| 20| 25| 20| ****
Wirewound, non-inductive
| 0.8 A| 0.8 V| 0.6 W| 1.1 A| 1.1 V| 1.2 W
10 W| 100 W| 20| 25| 15| 0.25 A| 2.5 V| 0.6 W| 0.35 A| 3.5 V| 1.2 W
100 W| 1 kW| 20| 25| 5| 80 mA| 8 V| 0.6 W| 110 mA| 11 V| 1.2 W
1 k W| 10 kW| 20| 25| 5| 23 mA| 23 V| 0.5 W| 35 mA| 35 V| 1.2 W
10 k W| 100 kW| 20| 25| 5| 7 mA| 70 V| 0.5 W| 11 mA| 110 V| 1.2 W
100 k W| 1 MW| 20| 25| 5| 2.3 mA| 230 V| 0.5 W| 3 mA| 500 V| 1 W
1 M W| 10 MW| 20| 25| 5| 0.7 mA| 700 V| 0.5 W| 1 mA| 1000 V| 1 W
10 M W| 100 MW| 50| 100| 10| Metal oxide film| 0.1 mA| 1000 V| 0.1 W|
0.1 mA| 1000 V| 0.1 W
*Subject to maximum of 2000 V to case
Accuracy:
≤1 MΩ steps: ±(0.01% + 2 mΩ)
10 MΩ steps: ±0.03%
After subtraction of zero- resistance, at 23°C; traceable to SI
Zero resistance:
≤1 MΩ steps: ≤1 mΩ per decade
10 MΩ steps: ≤3 mΩ
Switch type:
Multiple solid silver alloy contacts, continuous rotation
Switch capacitance:
<1 pF between contacts
Operation:
If switches have not been operated for an extended period of time, they should
be rotated a few times in both directions to restore contact resistance to
specifi cations.
Environmental conditions:
Operating conditions: +10 to +40°C, <80% RH
Storage conditions: -20 to +65°C
Terminals:
3 Low-thermal-emf tellurium-copper binding posts with standard 3/4 inch
spacing, plus shield terminal; connections on the rear of the instrument are
available (RO option). -K Option: Kelvin connection is available as an option
with 5 binding posts
ORDERING INFORMATION
**Model*| Total resistance**
(Ω)
| No of
Dials
| Resolution
(Ω)
---|---|---|---
HARS-X-1-0.001| 0.01| 1| 0.001
HARS-X-1-0.01| 0.1| 1| 0.01
HARS-X-1-0.1| 1| 1| 0.1
HARS-X-1-1| 10| 1| 1
HARS-X-1-10| 100| 1| 10
HARS-X-1-100| 1 k| 1| 100
HARS-X-1-1K| 10 k| 1| 1 k
HARS-X-1-10K| 100 k| 1| 10 k
HARS-X-1-100K| 1 M| 1| 100 k
HARS-X-1-1M| 10 M| 1| 1 M
HARS-X-1-10M| 100 M| 1| 10 M
HARS-X-2-0.001| 0.11| 2| 0.001
HARS-X-2-0.01| 1.1| 2| 0.01
HARS-X-2-0.1| 11| 2| 0.1
HARS-X-2-1| 110| 2| 1
HARS-X-2-10| 1.1 k| 2| 10
HARS-X-2-100| 11 k| 2| 100
HARS-X-2-1K| 110 k| 2| 1 k
HARS-X-2-10K| 1.1 M| 2| 10 k
HARS-X-2-100K| 11 M| 2| 100 k
HARS-X-2-1M| 110 M| 2| 1 M
HARS-X-3-0.001| 1.11| 3| 0.001
HARS-X-3-0.01| 11.1| 3| 0.01
HARS-X-3-0.1| 111| 3| 0.1
HARS-X-3-1| 1.11 k| 3| 1
HARS-X-3-10| 11.1 k| 3| 10
HARS-X-3-100| 111 k| 3| 100
HARS-X-3-1K| 1.11 M| 3| 1 k
HARS-X-3-10K| 11.1 M| 3| 10 k
HARS-X-3-100K| 111 M| 3| 100 k
Mechanical:
Model | Dimensions | Weight |
---|---|---|
1 decade | 7.7 cm W x 7.7 cm H x 8.4 cm D (3″ x 3″ x 3.3″) |
0.45 kg (1 lb)
2-3 decade| 31 cm W x 8.9 cm H x 10.2 cm D (12.2″ x 3.5″ x 4″)|
1.7 kg (3.8 lb)
4-5 decade| 37.5 cm W x 8.9 cm H x 10.2 cm D (14.75″ x 3.5″ x 4″)|
2.0 kg (4.3 lb)
6 decades| 43.9 cm W x 8.9 cm H x 10.2 cm D (17.3″ x 3.5″ x 4″)| 2.2 kg
(4.8 lb)
7 decades| 2.4 kg (5.3 lb)
8 decades|
48.3 cm W x 17.8 cm H x 17.8 cm D (19″ x 7″ x 7″)
| 3.4 kg (7.5 lb)
9 decades| 3.5 kg (7.7 lb)
10 decades| 3.6 kg (7.9 lb)
11 decades| 3.7 kg (8.1 lb)
**Model*| Total resistance (Ω)| No of Dials|
Resolution (Ω)
---|---|---|---
HARS-X-4-0.001| 11.11| 4| 0.001
HARS-X-4-0.01| 111.1| 4| 0.01
HARS-X-4-0.1| 1.111 k| 4| 0.1
HARS-X-4-1| 11.11 k| 4| 1
HARS-X-4-10| 111.1 k| 4| 10
HARS-X-4-100| 1.111 M| 4| 100
HARS-X-4-1K| 11.11 M| 4| 1 k
HARS-X-4-10K| 111.1 M| 4| 10 k
HARS-X-5-0.001| 111.11| 5| 0.001
HARS-X-5-0.01| 1.1111 k| 5| 0.01
HARS-X-5-0.1| 11.111 k| 5| 0.1
HARS-X-5-1| 111.11 k| 5| 1
HARS-X-5-10| 1.1111 M| 5| 10
HARS-X-5-100| 11.111 M| 5| 100
HARS-X-5-1K| 111.11 M| 5| 1 k
HARS-X-6-0.001| 1.111 11 k| 6| 0.001
HARS-X-6-0.01| 11.1111 k| 6| 0.01
HARS-X-6-0.1| 111.111 k| 6| 0.1
HARS-X-6-1| 1.111 11 M| 6| 1
HARS-X-6-10| 11.1111 M| 6| 10
HARS-X-6-100| 111.111 M| 6| 100
HARS-X-7-0.001| 11.111 11 k| 7| 0.001
HARS-X-7-0.01| 111.1111 k| 7| 0.01
HARS-X-7-0.1| 1.111 111 M| 7| 0.1
HARS-X-7-1| 11.111 11 M| 7| 1
HARS-X-7-10| 111.1111 M| 7| 10
HARS-X-8-0.001| 111.111 11 k| 8| 0.001
HARS-X-8-0.01| 1.111 111 1 M| 8| 0.01
HARS-X-8-0.1| 11.111 111 M| 8| 0.1
HARS-X-8-1| 111.111 11 M| 8| 1
HARS-X-9-0.001| 1.111 111 11 M| 9| 0.001
HARS-X-9-0.01| 11.111 111 1 M| 9| 0.01
HARS-X-9-0.1| 111.111 111 M| 9| 0.1
HARS-X-10-0.001| 11.111 111 11 M| 10| 0.001
HARS-X-10-0.01| 111.111 111 1 M| 10| 0.01
HARS-X-11-0.001**| 111.111 111 11 M| 11| 0.001
- Use “X2” for higher power model
OPTIONS
– RM Rack mountable case for standard 19” rack
– K Kelvin type 4-terminal binding posts
– RO Rear output binding posts
HARS-X and X2
HARS-X SERIES HIGH ACCURACY DECADE RESISTANCE SUBSTITUTER
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
Resistance per step| Total deacde resistance| Max current|
Max voltage| Power per/step (W)| Stability (±ppm/yr)| Long-
term stability (±ppm/3 yrs)| Temp. coeffieient (±ppm/°C)| Resistor
type
---|---|---|---|---|---|---|---|---
1 m Ω| 10 mΩ| 8 A| 5 mV| 0.4| 50| 75| 50| __
Resistance Wire
10 m Ω| 100 mΩ| 4 A| 40 mV| 0.16| 50| 75| 20
100 m Ω| 1 Ω| 1.6 A| 0.16 V| 0.25| 50| 75| 20
1 Ω| 10 Ω| .8 A| 0.8 V| 0.6| 20| 25| 20| __
__
__
Wirewound, non-inductive
10 Ω| 100 Ω| .25 A| 2.5 V| 0.6| 20| 25| 15
100 Ω| 1 kΩ| 80 mA| 8 V| 0.6| 20| 25| 5
1 k Ω| 10 kΩ| 23 mA| 23 V| 0.5| 20| 25| 5
10 k Ω| 100 kΩ| 7 mA| 70 V| 0.5| 20| 25| 5
100 k Ω| 1 MΩ| 2.3 mA| 230 V| 0.5| 20| 25| 5
1 M Ω| 10 MΩ| 0.7 mA| 700 V| 0.5| 20| 25| 10
10 M Ω| 100 MΩ| 0.1 mA| 1000 V| 0.1*| 50| 100| 10| Metal oxide film
Subject to maximum of 2000 V to case
Accuracy: £ 1 M L steps: ±(0.01% + 2 mL)
10 M L steps: ±0.03%
After subtraction of zero resistance, at 23°C; traceable to SI
Zero Resistance:
£ 1 M L steps: £1 mL per decade
10 M L steps: approx. 3 mL
Switch Type:
Multiple solid silver alloy contacts, continuous rotation
Switch Capacitance:
<1 pF between contacts
Operating Conditions:
+10 to + 40°C, <80% RH
Operation: If switches have not been operated for an extended period,
they should be rotated a few times to restore contact resistance to
*Subject to maximum of 2000 V to case __
specifications.
MODEL: HARS-X-6-0.01 SN: C3-09404334
__ WARNING
Observe all safety rules when working with high voltages or line voltages.
Connect the (G) terminal to earth ground in order to maintain the case at a
safe voltage. Whenever hazardous voltages (>45 V) are used, take all measures
to avoid accidental contact with any live components: a) Use maximum
insulation and minimize the use of bare conductors. b) Remove power when
adjusting switches. c) Post warning signs and keep personnel safely away.
Figure 2-1. Typical Operating Guide Affi xed to Unit
Chapter 3 OPERATION
3.1 Initial Inspection and Setup
This instrument was carefully inspected before shipment. It should be in
proper electrical and mechanical order upon receipt.
An OPERATING GUIDE is attached to the case of the instrument to provide ready
reference to speci- fi cations.
3.2 Connection
3.2.1 General Considerations
The HARS Series Decade unit provides three terminals labeled H (high), L
(low), and G (ground). The H and L terminals are connected to the ends of the
resistance being set; the G terminal is connected to the case. The G terminal
may be used as a guard or shield terminal. It may also be connected (using a
shorting link) to the L terminal to allow two-terminal as opposed to three-
terminal measurements. In order to make the most stable measurements,
determine which is the more sensitive of the two user leads, i.e. the one
going into a higher impedance. This lead should be connected to the more
protected one of the two HARS terminals. That would either be the HARS
terminal that is shorted to the case, or the LOW HARS terminal whenever
neither is connected to the case. If switches have not been operated for an
extended period, they should be rotated a few times to restore contact
resistance to specifi cations.
3.2.2 Electrical Considerations
In order to make proper use of the full performance capabilities of the HARS
unit, especially if low resistance or low-resistance increments are important,
take care when connecting to the terminals of the decade box. In particular,
in order to keep contact resistance to a minimum, take the most substantial
and secure connection to the binding posts. They accept banana plugs,
telephone tips, spade lugs, alligator clips, and bare wire. The largest or
heaviest mating connection should be made, and, where applicable, the binding
posts should be securely tightened. These considerations may be relaxed
whenever single milliohms are considered insignifi cant for the task being
performed.
3.2.3 Four-Wire Kelvin Lead Connections
Whenever possible, 4-wire Kelvin leads, the ideal connection, should be
employed. Such a connection minimizes the effects of contact resistance and
approaches ideal performance.
If the four terminals are available as clamps similar to alligator clips, they
may be connected to the necks of the binding posts. If the four terminals are
available separately, the optimal connection is shown in Figure
3.1, where the current leads are introduced into the top of the binding posts,
and the voltage leads at the necks.
HARS-X and X2
3.2.4 Thermal emf Considerations
The highest-quality low-emf components are used in the HARS Series. There
nevertheless may be some minute thermal emf generated at the test leads where
they contact the gold-plated binding posts.
This emf will not refl ect itself if an ac measurement instrument is employed.
It will also be eliminated if a meter with a “True Ohm” capability is used.
Otherwise it may represent itself as a false component of the dc resistance
measurement. It is also possible to take a second measurement with the leads
reversed and average the reading.
3.3 Dial Setting
Whenever the dials are used in positions 0-9, the resulting resistance is read
directly. Both the decimal point and the steps are clearly marked on the
panel. For additional fl exibility and range, each decade provides a “10”
position setting. This “10” position
on any one decade equals the “1” position on the next higher decade. It adds
about 11% to the nominal total decade resistance.
To determine the resistance obtained when one or more “10” settings are used,
simply add “1” to the next higher decade. For example, a setting of 3-610-0-10
Ω becomes:
3.4 Environmental Conditions
For optimal accuracy, the decade box should be used in an environment of 23 o
C. It should be allowed to stabilize at that temperature after any signifi cant
temperature variation.
Humidity should be maintained at laboratory conditions. This is especially
important if high resistances are involved.
Chapter 4 MAINTENANCE
4.1 Verifi cation of Performance
4.1.1 Calibration Interval
The HARS Series instruments should be verifi ed for performance at a
calibration interval of twelve (12) months. This procedure may be carried out
by the user if a calibration capability is available, by IET Labs, or by a
certifi ed calibration laboratory.
If the user should choose to perform this procedure, then the considerations
below should be observed.
4.1.2 General Considerations
It is important, whenever testing the HARS Series Decade Units, to be very
aware of the capabilities and limitations of the test instruments used. A
resistance bridge may be employed, and there are direct-reading resistance
meters or digital multimeters available that can verify the accuracy of these
units, especially when used in conjunction with standards that can serve to
confi rm or improve the accuracy of the testinginstrument Such test
instruments must be signifi cantly more accurate than ±(l00ppm+2 mΩ) for all
applicable ranges, allowing for a band of uncertainty of the instrument
itself.
A number of commercial bridges and meters exist that can perform this task;
consult IET Labs.
It is important to allow both the testing instrument and the HARS Substituter
to stabilize for a number of hours at the nominal operating temperature of 23
O C, and at nominal laboratory conditions of humidity.
There should be no temperature gradients across the unit under test.
Substantial Kelvin-type 4-wire test terminals should be used to obtain
accurate low-resistance readings. It is convenient, once the zero resistance
has been determined, to subtract it from the remaining measurements.
This can be automatically done in many instruments which have an offset
subtraction capability.
4.1.3 Calibration Procedure
-
Confirm the zero resistance of the unit.
-
Determine the allowable upper and lower limits for each resistance setting of each decade based on the specifi ed accuracy.
For the HARS series, these limits for any resistance “R” are [R±(0.0001 R + 2 mΩ)]. -
Confi rm that the resistances fall within these limits after subtraction of the zero resistance.
-
If any resistances fall outside these limits, the associated switch assembly may require service or replacement.
4.2 Schematic
Refer to Figure 4-1 for a schematic of the HARS decade unit.
4.3 Replaceable Parts List
Table 4.2: Replacement List
Model Ref | IET Pt No | Description |
---|---|---|
1 | BP-1000-RD | Binding Post, Red |
2 | BP-1000-BK | Binding Post, Black |
3 | BP-1000-GN | Binding Post, Green |
4 | HARS-X-4300-KNB | Knob Assembly |
Not Shown | HARS-X-3100 | Foot |
Not Shown | HARS-4100-X-.001 | 1 mΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-0.01 | 10 mΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-0.1 | 100 mΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-1 | 1 Ω/step Decade Switch Assembly |
Not Shown | HARS-4100-X-10 | 10 Ω/step Decade Switch Assembly |
Not Shown | HARS-4100-X-100 | 100 Ω/step Decade Switch Assembly |
Not Shown | HARS-4100-X-1k | 1 kΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-10k | 10 kΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-100k | 100 kΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-1M | 1 MΩ/step Decade Switch Assembly |
Not Shown | HARS-4100-X-10M | 10 MΩ/step Decade Switch Assembly |
- Use “X2” for higher power model, for example HARS-4100-X2-100 for 100 Ω/step Decade X2 Switch Assembly
Maintenance
1.888.610.7664
www.calcert.com
sales@calcert.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>