ACME YF14 Vacuum Tester User Manual

June 13, 2024
acme

ACME-logo

ACME YF14 Vacuum Tester

ACME-YF14-Vacuum-Tester-product

Product Information

Product Name Vacuum Tester
Product Type Air Tightness Tester
Usage For testing waterproof fixtures using compressed air or vacuum

as the medium for product inspection.
Power Voltage| 100-240V~ 50/60Hz
Power Consumption| 40W
Lux| Not specified
Maximum Positive Pressure| 17L/minute
Maximum Negative Pressure| 80kpa -60kpa
Firmware Upgrade| Firmware Upgrade via USB disk
Construction Features| Dimensions: 355x167x354mm (14×6.6×13.9in)
Display: LCD display
Protection Rating: IP20
Weight: 7kgs (15.4lbs)

Product Usage Instructions

  1. Read the instruction manual carefully before installation, usage, and maintenance.
  2. Keep the User Guide for future reference. If selling the unit to another user, provide them with the instruction manual as well.
  3. Unpack and check the unit for any transportation damage before using it.
  4. The product is for indoor use only and should be used in a dry location.
  5. Ensure proper voltage connection according to the specifications mentioned in the manual or on the product’s label before operation.
  6. The minimum ambient temperature for operation is 0°C and the maximum ambient temperature is 40°C. Do not operate the product at lower or higher temperatures.
  7. Keep flammable materials away from the fixture while operating to avoid fire hazards.
  8. Check the power cord for any damage or crimps. Replace it immediately if damaged.
  9. Avoid allowing flammable liquids, water, or metal to enter the unit. If it happens, immediately cut off the mains power.
  10. Do not operate the unit in a dirty or dusty environment. Clean the fixture regularly.
  11. Avoid entanglement of the power cord with other wires.
  12. If there is a serious operating problem, stop using the unit immediately.
  13. Do not attempt to operate the unit if it becomes damaged. Do not attempt any repairs yourself. Contact the nearest authorized technical assistance center for repairs.
  14. Disconnect the product from its power source before servicing.
  15. Use the original packaging if transporting the device.
  16. Authorized maintenance personnel should refer to the MAINTENANCE TECHNICAL MANUAL for conducting maintenance.
  17. Do not operate the fixture at altitudes higher than 3000 meters.
  18. Ensure humidity levels are between a minimum of 0%RH and a maximum of 70%RH.
  19. Pollution degree: 1

Control Panel

Button Function
Display To show various menus and selected function
Buttons To enter into, move backward, or leave the menu
User Manual Not specified
Pressure/Vacuum To Fixture Not specified
USB Not specified
Stop Not specified
Power In Not specified
Power Switch Not specified
Start Not specified

Safety Instructions

Please read the instruction carefully which includes important information about the installation, usage and maintenance.

Please keep this User Guide for future consultation. If you sell the unit to another user, be sure that they also receive this instruction manual.

Important: Damages caused by the disregard of this user manual are not subject to warranty. The dealer will not accept liability for any resulting defects or problems.

  • Unpack and check carefully to ensure that there is no transportation damage before using the unit.
  • This product is for indoor use only. Use only in a dry location.
  • CAUTION, possibility of electric shock.
  • DO install and operate by qualified operator.
  • DO NOT allow children to operate the fixture.
  • Before operation, ensure that you are connecting this product to the proper voltage in accordance with the specifications in this manual or on the product’s specification label.
  • Minimum ambient temperature TA: 0℃. Maximum ambient temperature TA: 40℃. Do not operate this product at a lower or higher temperature.
  • Keep flammable materials away from the fixture while operating to avoid fire hazard.
  • Make sure the power cord is not crimped or damaged; replace it immediately if damaged.
  • Avoid any flammable liquids, water or metal from entering the unit. Once it happens, cut off the mains power immediately.
  • DO NOT operate in a dirty or dusty environment. DO clean the fixture regularly.
  • Avoid entanglement of the power cord with other wires.
  • In the event of serious operating problem, stop using the unit immediately.
  • DO NOT attempt to operate this unit if it becomes damaged. DO NOT attempt any repairs yourself. Repairs carried out by unskilled people can lead to damage or malfunction. Please contact the nearest authorized technical assistance center if needed.
  • Disconnect this product from its power source before servicing.
  • DO use the original packaging if the device is to be transported. Authorized maintenance personnel must refer to MAINTENANCE TECHNICAL MANUAL to conduct to maintenance.
  • Altitude: Below 3000m. DO NOT operate the fixture at altitudes higher than 3000 meters.
  • Humidity: Min. 0%RH. Max. 70%RH.
  • Pollution degree: 1.

Technical Specifications

Power Voltage 100-240V~ 50/60Hz
Power Consumption 40W
Lux 17L/minute
Maximum Positive Pressure 80kpa
Maximum Negative Pressure -60kpa
Control Firmware Upgrade

Construction

| Display| LCD display|
Protection Rating| IP20|


Features

| Vacuum Tester is a portable air tightness tester designed for testing waterproof fixtures. It utilizes the pressure drop method to conduct non- destructive testing by using compressed air (or vacuum) as the medium for product inspection.
Dimensions| 355x167x354mm| 14″x6.6″x13.9″in
Weight| 7kgs| 15.4lbs

Control Panel

**

ACME-YF14-Vacuum-Tester-fig- \(2\)**

1. Display To show the various menus and the selected function
MENU To enter into move backward or leave the menu

2. Buttons

|  UP| To go backward to move up in the menu
 DOWN| To go forward to move down in the menu
| ENTER| To perform the desired functions
3. User Manual
4. Pressure/Vacuum To Fixture| Used to connect the air hose to pump/transfer gas for the luminaire
5. USB| Non-power output port, only use to upgrade the fixture’s firmware or export the test record
6. Stop| Press to stop testing
7. Power In| To connect to supply power
8. Power Switch| Turns on/off the power
9. Start| Press to start testing

Connections

  1. Connect the power cord to the Vacuum Tester (Figure 1) and plug it in.
  2. Take out the air hose (Figure 2), which has a quick-connect fitting and a bigger nozzle at each end (Figure 3). Connect the air hose quick-connect fitting to the Vacuum Tester (Figure 4). Carefully insert the threaded end of the nozzle into the air relief valve hole of the luminaire and twist it until it is finger tight (Figure 5), then switch the Vacuum Tester on (Figure 6).ACME-YF14-Vacuum-Tester-fig- \(4\)
  3. To switch the Vacuum Tester off, press the power switch again or directly unplug the power cord.

Test Instructions

Attention : The Vacuum Tester should be hermetically connected with the luminaire, otherwise the test will be affected.

ACME  luminaire pressure test

  • To access the control menus, press the [MENU]
  • Navigate the menu structure, using the [ENTER], [ UP] and [ DOWN]
  • To select a menu option or to confirm a selection, press the [ENTER]
  • To return to a higher level in the menu structure without making a change, press the [MENU] button, or wait 30
  1. Enter the Main Menu and use the UP/ DOWN button to select Air Pressure Test (Figure 1), confirm your selection with ENTER
  2. Enter the Test Menu and use the UP/ DOWN button to select Fixture Type (Figure 2), confirm your selection with ENTER
  3. Enter the Fixture Type page and use the UP/ DOWN button to select Moving Lights, Lights or Static Lights (Figure 3), confirm your selection with ENTER.ACME-YF14-Vacuum-Tester-fig- \(7\)
  4. Enter into the test page (Figure 4, Figure 5) and press the green button to start testing.ACME-YF14-Vacuum-Tester-fig- \(8\)
  5. Press the red button to stop testing if there is any problem during the test.
  6. “Test Pass” will be displayed when the test passes (Figure 6) and “Test Failed” will be displayed when the test fails (Figure 7).ACME-YF14-Vacuum-Tester-fig- \(10\)

User-defined pressure  test

  • To access the control menus, press the [MENU]
  • Navigate the menu structure, using the [ENTER], [ UP] and [ DOWN]
  • To select a menu option or to confirm a selection, press the [ENTER]
  • To return to a higher level in the menu structure without making a change, press the [MENU] button, or wait 30
  1. Enter the Main Menu and use the UP/ DOWN button to select Air Pressure Test (Figure 1), confirm your selection with ENTER
  2. Enter the Test Menu and use the UP/ DOWN button to select Manual Setting (Figure 2), confirm your selection with ENTER. Adjust the parameters of the time and pressure on the Manual Setting page (Figure 3), press the ENTER button to restore and it will automatically return to the Test Menu.ACME-YF14-Vacuum-Tester-fig- \(12\)
  3. Enter the Test Menu and use the UP/ DOWN button to select Vacuum Test (Figure 4) or Pressure Test (Figure 5) as needed, confirm your selection with ENTERACME-YF14-Vacuum-Tester-fig- \(13\)
  4. Enter into the test page (Figure 6, Figure 7) and press the green button to start testing.ACME-YF14-Vacuum-Tester-fig- \(14\)
  5. Press the red button to stop testing if there is any problem during the test.
  6. “Test Pass” will be displayed when the test passes (Figure 8) and “Test Failed” will be displayed when the test fails (Figure 9).ACME-YF14-Vacuum-Tester-fig- \(16\)

Air leakage test

  • To access the control menus, press the [MENU]
  • Navigate the menu structure, using the [ENTER], [ UP] and [ DOWN]
  • To select a menu option or to confirm a selection, press the [ENTER]
  • To  return to a higher level in the menu structure without making a change, press  the [MENU] button, or wait 30 seconds
  1. Enter into the Main Menu and use the UP/ DOWN button to select Air Leakage Test (Figure 1), confirm your selection with ENTER
  2. Enter into the test page (Figure 2), press the green button to start, then the Vacuum Tester will continuously fill the luminaire with air.
  3. Put the luminaire into the water to confirm the location of air leakage by checking where the luminaire bubbles.
  4. Take the luminaire out of the water and press the red button to stop testing.

System Setting

  • To access the control menus, press the [MENU]
  • Navigate the menu structure, using the [ENTER], [ UP] and [ DOWN]
  • To select a menu option or to confirm a selection, press the [ENTER]
  • To return to a higher level in the menu structure without making a change, press the [MENU] button, or wait 30
  1. Enter into the Main Menu and use the UP/ DOWN button to select System Setting (Figure 1), confirm your selection with ENTER
  2. Enter into the System Setting page and use the UP/ DOWN button to select Language (Figure 2), confirm your selection with ENTER
  3. Enter into the Language page and use the UP/ DOWN button to select English or (Figure 3), confirm your selection with ENTER.
  4. Enter into the System Setting page and use the UP/ DOWN button to select Test Record (Figure 4), confirm your selection with ENTER. Enter into the Test Record page to view or export the test record, press the MENU button to exit.ACME-YF14-Vacuum-Tester-fig- \(22\)
  5. Enter into the System Setting page and use the UP/ DOWN button to select Firmware (Figure 6), confirm your selection with ENTER. Enter into the ACME V-BOX page to view the firmware version of the Vacuum Tester (Figure 7), press the MENU button to exit.ACME-YF14-Vacuum-Tester-fig- \(23\)

System Upgrade

  • To access the control menus, press the [MENU]
  • Navigate the menu structure, using the [ENTER], [ UP] and [ DOWN]
  • To select a menu option or to confirm a selection, press the [ENTER]
  • To  return to a higher level in the menu structure without making a change, press  the [MENU] button, or wait 30 seconds
  1. Insert the USB drive with the upgrade file into the USB port.
  2. Enter the Main Menu and use the UP/ DOWN button to select System Upgrade (Figure 1), confirm your selection with ENTER
  3. Enter the System Upgrade page and use the UP/ DOWN button to select  Upgrade  Firmware (Figure 2), confirm your selection with ENTER to upgrade fixture’s

Main Functions

  • To access the control menus, press the [MENU]
  • Navigate the menu structure, using the [ENTER], [ UP] and [ DOWN]
  • To select a menu option or to confirm a selection, press the [ENTER]
  • To return to a higher level in the menu structure without making a change, press the [MENU] button, or wait 30

The main functions are shown below:

MENU SUBMENU OPTION

Air Pressure Test

|

Fixture Type

| Moving Lights
Arc. Lights
Static Lights
Manual Setting
Vacuum Test
Pressure Test
Air Leakage Test|

System Setting

| Language| English
中文
Test Record
Firmware Ver.
System Upgrade| Upgrade Firmware