TRIPLETT CM1070 TRMS AC-DC Clamp User Manual
- June 13, 2024
- TRIPLETT
Table of Contents
TRIPLETT CM1070 TRMS AC-DC Clamp User Manual
Introduction
Congratulations on your purchase of the CM1070 True RMS Clamp Meter. This meter measures AC Current, DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency, Diode Test, Duty Cycle and Continuity. Special features include Dual Input Thermocouple Temperature and Non-Contact Voltage detector. The double molded case is designed for heavy duty use. This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Safety
International Safety Symbols
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in
which measures are taken to limit the transient overvoltages to an appropriate
low level.
Note – Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be
supplied from the fixed installation.
Note – Examples include household, office, and laboratory appliances.
OVERVOLTAGE CATEGORY III
Equipment of OVERVOLTAGE CATEGORY III is equipment in fixed installations.
Note – Examples include switches in the fixed installation and some equipment
for industrial use with permanent connection to the fixed installation.
OVERVOLTAGE CATEGORY IV
Equipment of OVERVOLTAGE CATEGORY IV is for use at the origin of the
installation.
Note – Examples include electricity meters and primary over-current protection
equipment.
SAFETY NOTES
- Do not exceed the maximum allowable input range of any function.
- Do not apply voltage to meter when resistance function is selected.
- Set the function switch OFF when the meter is not in use.
- Remove the battery if meter is to be stored for longer than 60 days.
WARNINGS
- Set function switch to the appropriate position before measuring.
- When measuring volts do not switch to current/resistance modes.
- Do not measure current on a circuit whose voltage exceeds 600V.
- When changing ranges always disconnect the test leads from the circuit under test.
CAUTIONS
- Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter.
- Always remove the test leads before replacing the battery or fuses.
- Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use.
- Use great care when making measurements if the voltages are greater than 25VAC rms or 35VDC. These voltages are considered a shock hazard.
- Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or Continuity tests.
- Voltage checks on electrical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts. Other means should be used to ensure that the terminals are not “live”.
- If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
- This device is not a toy and must not reach children’s hands. It contains hazardous objects as well as small parts that the children could swallow. In case a child swallows any of them, please contact a physician immediately.
- Do not leave batteries and packing material lying around unattended; they can be dangerous for children if they use them as toys.
- In case the device is going to be unused for an extended period of time, remove the batteries to prevent them from draining.
- Expired or damaged batteries can cause cauterization on contact with the skin. Always, therefore, use suitable hand gloves in such cases
- See that the batteries are not short-circuited. Do not throw batteries into the fire.
- Do not directly view or direct the laser pointer at an eye. Low power visible lasers do not normally present a hazard but may present some potential for hazard if viewed directly for extended periods of time.
Description
Meter Description
- Current clamp
- Clamp opening trigger
- HOLD/BACKLIGHT button
- MODE / oC/ oF
- Backlit LCD Display
- Type K input jacks
- Multimeter input jacks
- Non-Contact Voltage Detector
- LAMP
- NCV LED indicator
- LAMP/ZERO button
- Function switch
- MAX/MIN button
- RANGE/PEAK / Thermocouple display button
- Input shutter
Display icons Description
Operation
NOTES: Read and understand all Warning and Caution statements in this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use.
Input Shutter
The Input Shutter inhibits simultaneous connection to the thermocouple jacks
and the DMM input jacks. This is a safety feature which prevents a potentially
hazardous condition from existing during high voltage measurements. Slide the
shutter up for test lead measurements or slide it down for thermocouple
temperature measurements.
Non-Contact Voltage Detector
WARNING: Risk of Electrocution. Before use, always test the Voltage Detector
on a known live circuit to verify proper operation.
- Rotate the Function switch to any measurement position.
- Place the detector probe tip on the conductor to be tested.
- If AC voltage is present, the NCV detector light will turn on with a steady red light.
NOTE: The conductors in electrical cord sets are often twisted. For best results, move the probe tip along a length of the cord to assure placing the tip near the live conductor.
NOTE: The detector is designed with high sensitivity. Static electricity or other sources of energy may randomly trip the sensor. This is normal operation.
AC/DC Current Measurements
WARNING: Disconnect the test leads before making clamp measurements.
- Rotate the Function switch to the 1000AAC/DC position
- Press the MODE button to select AC or DC.
- Press the trigger to open jaw. Fully enclose only one conductor.
- Read the current value in the display.
- If the value is less than 50A, rotate the function switch to the 50AAC/DC position to improve resolution.
DCA Zero
The Zero feature removes offset values and improves accuracy for DC current
measurements. To perform a zero, select ADC and, with no conductor in the jaw,
press and hold the MODE ZERO button for two beeps. The display will zero. The
offset value is now stored and removed from all measurements.
Frequency
When ACV is selected, the measured frequency can be viewed in the lower
display.
AC/DC Voltage Measurements
CAUTION: Do not measure voltages if a motor on the circuit is being switched
ON or OFF. Large voltage surges may occur that can damage the meter.
- Slide the input shutter to the up position.
- Rotate the function switch to the V position.
- Press the MODE button to select AC or DC Voltage.
- Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack.
- Touch the black test probe tip to the negative side of the circuit. Touch the red test probe tip to the positive side of the circuit.
- Read the voltage value in the display.
Frequency
When ACA is selected, the measured frequency can be viewed in the lower
display.
Resistance Measurements
Note: Remove power from the device under test before making resistance
measurements
- Slide the input shutter to the up position.
- Set the function switch to the position.
- Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive V jack.
- Touch the black test probe tip to one side of the device. Touch the red test probe tip to the other side of the device.
- Read the resistance value in the display.
Continuity Test
- Connect as described for resistance measurements.
- Press the MODE button to select continuity•))).
- Touch the test probe tips across the circuit or component under test.
- If the resistance is < 50, a tone will sound.
Diode Test
Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor before measuring.
Frequency and Duty Ratio Measurements
-
Slide the input shutter to the up position.
-
Rotate the function switch to the Hz % Position.
-
Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive Hz jack. -
Touch the black test probe tip to one side of the device.
Touch the red test probe tip to the other side of the device. -
Read the Frequency value on the upper large display.
Read the Duty Ratio on the lower small display. -
Press the MODE button to display the Duty Ratio on the large display.
µA DC/AC Current Measurements
- Slide the input shutter to the up position.
- Rotate the function switch to the µ A position.
- Press the MODE button to select AC or DC.
- Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive A jack.
- Turn power to the circuit under test off and make a break in the circuit.
- Insert the meter in series with the circuit; Touch the black test probe tip to the negative side of the break. Touch the red test probe tip to the positive side of the break.
- Turn circuit power on.
- Read the current value in the display.
Type K Temperature Measurements
- Slide the input shutter to the down position.
- Rotate the function switch to the TYPE K temperature position.
- Press the MODE button to select °F or °C.
- Insert the Temperature Probe(s) into the T1 and/or T2 type k sockets.
- Place the temperature probe tip(s) where needed.
- Read the temperature on the display.
- Press the T1-T2 button to step through the display combinations:
Data Hold
To freeze the LCD reading, press the HOLD button. While data hold is active,
the HOLD icon appears on the LCD. Press the HOLD button again to return to
normal operation.
MAX/MIN
- Press the MAX/MIN button to activate the MAX/MIN recording mode. The display icon “MAX” will appear. The meter will begins recording and displaying the maximum value measured.
- Press the MAX/MIN button and “MIN” will appear. The meter will display the minimum value measured during the recording session.
- Press the MAX/MIN button and “MAX MIN” will appear. The meter will display the present reading, but will continue to update and store the max and min readings.
- To exit MAX/MIN mode press and hold the MAX/MIN button for 2 seconds.
Peak Hold
When ACA or ACV is selected, press and hold PEAK button 2 seconds enables the
peak capture circuit. The meter will now capture and display the maximum and
minimum peaks
of the waveform.
To exit Peak Hold mode press and hold the Peak Hold button for 2 seconds.
RANGE
In the Voltage, Resistance, Capacitance, Frequency or uA function the meter
automatically
selects the best range for the measurements being made. For measurement
situations requiring that a range be manually selected, perform the following:
- Press the RANGE button. The “AUTO” display icon will turn off.
- Press the RANGE key to step through the available ranges. Observe the decimal point and units displayed until the preferred range is located.
- To exit the Manual Ranging mode and return to Autoranging, press and hold the RANGE key for 2 seconds.
Maintenance
WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals, and turn OFF the meter before opening the case. Do not operate the meter with an open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for 60 days or more, remove the battery and store it separately.
Battery Replacement
- Remove the Phillips head screw that secures the rear battery door
- Open the battery compartment
- Replace the 9V battery
- Secure the battery compartment door
You, as the end user, are legally bound (Battery ordinance) to return all used
batteries and accumulators; disposal in the household garbage is prohibited!
You can hand over your used batteries / accumulators, gratuitously, at the
collection points for our branches in your community or wherever batteries /
accumulators are sold!
Disposal
Follow the valid legal stipulations in respect of the disposal of the device at the end of its lifecycle
Fuse Replacement
- Remove the battery
- Remove the Phillips head screws (2) that secures the rear cover.
- Replace the fuse with one of equal rating. (500mA, 660V fast blow [SIBA 70-180-40])
- Replace the rear cover and battery
Specifications
Warranty
Triplett / Jewell Instruments extends the following warranty to the original purchaser of these goods for use. Triplett warrants to the original purchaser for use that the products sold by it will be free from defects in workmanship and material for a period of (1) one year from the date of purchase. This warranty does not apply to any of our products which have been repaired or altered by unauthorized persons in any way or purchased from unauthorized distributors so as, in our sole judgment, to injure their stability or reliability, or which have been subject to misuse, abuse, misapplication, negligence, accident or which have had the serial numbers altered, defaced, or removed. Accessories, including batteries are not covered by this warranty.
Copyright © 2023 Triplett www.triplett.com
Read More About This Manual & Download PDF:
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>