GRAS 246AE SysCheck2 Software Development Kit Instruction Manual
- June 13, 2024
- GRAS
Table of Contents
GRAS 246AE SysCheck2 Software Development Kit
SysCheck2™ Software Development Kit
Any feedback or questions concerning this document are welcome at gras@grasacoustics.com
Revision | Date | Description |
---|---|---|
1 | 11 Mar 2022 | SDK version 2.0.8 Layout changes only |
Instruction Manual
Copyright Notice ©2021 GRAS Sound & Vibration A/S
grasacoustics.com
Any technical documentation that is made available by GRAS is the copyrighted
work of GRAS and is owned by GRAS.
The content in this document is subject to change without notice. GRAS Sound &
Vibration is not liable or responsible for any errors or inaccuracies that may
appear in this document.
Trademarks
Any product names mentioned in this document may be trademarks or registered
trademarks of their respective companies and are hereby acknowledged.
Introduction
SDK for 246AE and 246AO microphones overview
SysCheck2™ is a GRAS-patented technology for verifying measurement chain
integrity. This verification tool performs remote health check on microphones,
channel gain and cable integrity. The verifications are made on each
SysCheck2-enabled microphone connected to a CCP power module with transducer
electronic data sheet (TEDS) support and measurement software with one click.
This software development kit (SDK) enables access to the SysCheck2
functionality.
You will need an analyzer with the ability to connect to a CCP-based
microphone and the ability to read and write to TEDS. Alternatively, you can
use a CCP-capable power analyzer with a non-CCP-capable analyzer or sound card
with TEDS read-write capability. Because the required hardware is built into
the microphone itself (Fig 1), no further specialized hardware is needed.
246AE, 246AO specification
The acoustic specification is like GRAS 46AE.
Syscheck2 generator specification (over temperature range -30°C to 85°C):
Frequency, sine: 250Hz
Frequency stability: ±3%
Amplitude level stability: ±0.03dB
Amplitude level tolerance: ±1dB
Harmonic distortion: -40dB
Output level with no microphone: -1.1dBV
Output level with 246AE (Cmic: 14pF): -27dBV equivalent to 93 dBspl
Output level with 246AO (Cmic: 20pF): -27.5dBV equivalent to 105.5 dBspl
Environment sensor: (operational -40°C to 85°C)
Temperature : ±2°C (0°C to 65°C)
Pressure static: ±1.5hPa (0°C to 65°C, 300hPa to 1100hPa)
Humidity relative: ±4% RH (0°C to 60°C, 0 to 100%)
Temperature sensor in CPU: (operational -40°C to 125°C)
Temperature : typical ±3°C (-40°C to 125°C)
Syscheck2 behavior
Syscheck2 has a very well-defined coupling of the test signal to the
microphone, due to precision coupling capacitor with a guarded signal path
from the generator in close proximity, resulting in reliable test results over
a wide frequency range due to no stray coupling of test signal to
preamplifier, polarization voltage or in transmission cable.
Syscheck2 generator level has very little influence from of temperature, the
Syscheck2 measurement therefore reflect the change of capacity of the
microphone. The membrane capacity has a slight frequency dependency, but not
enough to verify the frequency response. Causes to change in microphones
capacity are change in temperature, change in polarization voltage, change
membrane tension and damages to the microphone housing or the membrane.
Testing the capacity at medium frequencies e.g. 250Hz, can verify the general
sensitivity of the microphone. The Syscheck2 can reveal sensitivity change due
to temperature, but not to change in pressure or humidity.
The microphone sensitivity depends on temperature and static pressure; therefore, temperature and static pressure need to be accounted for. An environment sensor is included for this purpose.
Syscheck2 dependency of polarization voltage
Influence of polarization voltage on sensitivity of the microphone and
Syscheck2 measurement.
Decrease in sensitivity of 0.3dB due to reduction of polarization voltage
results in increased Syscheck2 voltage of 0.1dB.
Syscheck2 dependency of temperature
Influence of temperature on sensitivity of the microphone and Syscheck2
measurement.
40AE r20: is the Syscheck2 measurement with the load of 40AE Piston cr20: is the measurement of the 40AE sourced by the pistonphone.
Syscheck2 usages
In the simple use where environment condition has not changed, a relative
check with Syscheck2 @250Hz before and after will ensure that the sensitivity
has not changed more than 4 times the Syscheck2 change in dB (in other words
if the Syscheck2 has changed 0.05dB the microphone sensitivity is expected to
be within 4*0.05dB ~ ±0.2 dB.
Since the change can originate from different things, such as change in
membrane tension, change in mechanical or change of charge, the sign of the
change cannot be determined.
If temperature, ambient pressure, and humidity has changed, the sensitivity of
the microphone might also have changed. Initial when the reference Syscheck2
measurement is made the temperature, ambient pressure and humidity can read
from 246AE or 246AO and memorized together with the measurement. If only
temperature has changed the Syscheck2 verification can be corrected for this.
Note:
No temperature correction is made automatically in neither 246AE or 246AO.
The validation with Syscheck2 corrected for temperature, verifies that the
microphone is working correct, but the sensitivity might have changed, the
microphone sensitivity will need to be corrected for temperature. The
temperature coefficient of the microphone set 246AE and 246AO is -0.01dB/°C.
To avoid ambient noise disturbing the Syscheck2 verification, the ambient
noise must be reasonable low.
Notes:
- using a 1/3 octave narrow band measurement the noise level of 60dB during test, will typically influence less than 0.2dB.
- If the microphone is calibrated to -25.5dBV/Pa @ 23°C, the sensitivity correction of 246AE microphone @ 35°C is -0.01*(35-23)dB = -0.12dB, the sensitivity @ 35°C will be -25.5+(- 0.12)dBV/Pa = -25.62dBV/Pa.
Example of use case:
-
Together with a calibration of the microphone, make a Syscheck2 reference measurement:
a. Measure the Syscheck2 level, e.g -27.20dBV
b. Read the temperature from the TEDS, e.g. 25°C These data can be used to verify the microphone later. -
In the field when a verification of the microphone is needed:
a. Measure the Syscheck2 level, e.g -27.03dBV
b. Read the temperature from the TEDS, e.g. 35°C
c. Calculate temperature corrected verification 1): SC_level_corrected = SC_level_measured – ((Ta)
2** TC2 + Ta TC – ((RT)
2** TC2 + RT TC) )
→ SC_level_corrected = -27.03 – ((35)2 -96.0E-6 + 35 16.1E-3 – ((25)2 -96.0E-6 + 25 16.1E-3) ) = -27.13 dB
Verification deviation (Difference to Syscheck2 level): DSL = | SC_level_corrected – RL | = | -27.13 – (-27.20) | = 0.07 dB Use DSL result for evaluating Acceptance level according to the following:
Acceptance Level | Syscheck2 Level (DSL) |
---|---|
Green | Red |
0.3 | DSL <= 0.08 |
0.5 | DSL <= 0.13 |
0.8 | DSL <= 0.21 |
I.e. the microphone is expected to measure correct within 0.3 dB according to
the above scheme (when corrected for change in the environmental condition).
Note: If the microphone is calibrated to -25.5dBV/Pa @ 23°C, the sensitivity
correction of 246AE microphone @ 35°C is -0.01*(35-23)dB = -0.12dB, the
sensitivity @ 35°C will be -25.5+(- 0.12)dBV/Pa = -25.62dBV/Pa.
- Read TC2 and TC from the user data, can be recalled by command tc2 and tc, in this example is used tc2 = -96.0E-6 and tc = 16.1E-3
246AE, 246AO firmware 1.8
Firmware change log
Firmware 1.8 bug fix: Added space between t command and response
Uppercase gto to Gto after use
Control via user data in template UDID I27-0-0-0U
In user data the Syscheck2 communication is encapsulated, starting with “{:“,
and the communication terminated with “}”, before and after this the user
data is free to use.
Behavior:
In digital mode:
Normal IEEE 1451.4 TEDS mode, chip used DS2431.
In analog mode:
Reads and act upon content of the TEDS.
If generator has been enabled in TEDS, it will turn on in analog mode the and
the “f” in the user data will be changed to “F”, ensuring that the generator
will not turn on again automatically, except if generator timeout has been
specified, in this case the generator will always start and turn of after the
specified time.
The line drive capacity is reduced with 1.2mA when LED is on, and with 0.7mA
with generator on Normal analog mode is not affected, Syscheck2 idle current
is typical 5uA.
The user data commands defined for are: The command string is preceded by
“{:” and terminate with “}”
Each command in the command string must be followed by the needed spaces for
the response, shown in each command in parentage.
Commands are in general disabled after first use, this is done by uppercasing
the first command letter. For the LED command’s this can be disabled by the
command “a”.
The protocol for control and response in user data used by GRAS is
identified with a Pid: The Pid is a HEX number which defines the
capabilities, each bit in this number indicates if a set of command / response
is supported.
Definition is open to append more bits, once the interpret of a bit is defined
by GRAS it’s definition will not later be changed.
246AE, 246AO will return “Pid 00003F”.
Commands shall be separated by a space.
The figure in parenthesis designates the total needed field width of the
command.
pid (11) : Will return protocol ID in hex, e.g. 246AE returns Pid 00003F.
f : Generates a sinewave at frequency 250Hz, it will be disabled after
used, by changing f to F.
tc2 (14) : Will return the tc2 in the user data, e.g. Tc2 -96.0E-6
tc (13) : Will return the tc in the user data, e.g. Tc – 16.1E-3
gto # : Generator time out in sec., max value 225sec, gto 60 will turn of
the generator after 1 min.
fw (8) : Will return firmware version in user data, e.g. Fw 1.8
hw (8) : Will return hardware version in user data, e.g. Hw 3.0
t (8) : Requests the temperature, the TEDS will be updated when Analog
mode is entered with the CPU temperature in °C (accuracy is reduced, but
operational up to 125°C ) e.g. “t 90.3 ” env (18) : Requests environmental
condition in the preamplifier, the TEDS will be updated when Analog mode is
entered with the temperature in °C, the ambient pressure in hPa and relative
humidity in %, e.g. “env 23.4 1009 43 “.
r # : Turn on LED red 2,3)
g # : Turn on LED green 2,3)
b # : Turn on LED blue 2,3)
x # : Turn on all LED one at a time 2,3)
a : disables disable after first use for LED’s
RL # : Ref level in db : RL # 1)
RF # : Ref frequency in Hz : RF # 1)
RT # : Ref temperature in °C : RT # 1)
RP # : Ref pressure in hPa : RP # 1)
1) The Capital letters are not interpreted by 246AE and 246AO.
2) If xrgb is followed with a # the LED will turn off after # sec. (max
600sec.), if no # is applied it will turn off after 5sec.
3) the noise floor will be a bit higher while the LED is on.
Factory defaults for 246AE and 246AO looks like this:
User data for 246AE and 246AO:
“246AE {: Pid 00003F F Env 23.0 1013 50 RL -27.00 RT 23.0 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 G 010 }”
“246AO {: Pid 00003F F Env 23.0 1013 50 RL -27.00 RT 23.0 RP 1013 Tc2
-85.0E-6 Tc 10.2E-3 G 010 }”
Note: both padded with spaces to allocate space for response, total
length 101 chars.
Examples:
User data: “{: f RL -26.20 RT 24 rgb }”, the last } stops the interpret
of commands. This will flash r, g, b led each in 5 sec. (default), and turn on
test signal, which will stay on while in analog mode.
User data: “{: f RL -26.20 RT 24 g2 }”
This will hold green LED on for 2 sec and turn on test signal, which will stay
on while in analog mode. User data: “{: RL -26.20 RT 24 bg2 }”
This will hold blue LED on for 5 sec followed the green LED on for 2 sec, and
normal measurement can be done, the noise floor will be a bit higher while the
LED is on. User data: “{: f gto 45 RL -26.20 RT 24 g2 }”
This will hold green LED on for 2 sec and turn on test signal in 45sec. and
return to normal measurement in analog mode.
User data: “{: f gto 45 RL -26.20 RT 24 env g2 }” This will measure
temperature, air pressure and relative humidity in the preamplifier and write
these into user data, then hold green LED on for 2 sec and turn on test signal
in 45sec. and return to normal measurement in analog mode.
Description of the interactions between integration software and
Syscheck2 mic set
Abstract:
This chapter is made to help system integrators to retrieve information and
communicate with a Syscheck2 microphone set. It is a step by step instruction
based on three different functions that needs to be included in the
integration software.
Notes: In this description we use a 246AE. The procedures are unchanged*) when
using another type of Syscheck2 microphone set (i.e 246AO).
*) only change is for temperature and pressure coefficients
Requirement Specification for User Interface:
Temperature will be handled in Celcius. Unit conversion to Fahrenheit shall
only be informal to the user. Acceptance level:
User must be able to select an acceptance level based on a choice list (drop
down list?). Current acceptance limit thresholds are; 0.3dB, 0.5dB & 0.8dB
Syscheck2 Result:
User interface needs to show the Syscheck2 result, e.g. with 2 colors; Green &
Red depending on the Syscheck2 Level result compared to the acceptance level.
Criterias for colored verdict read out:
Acceptance Level | Syscheck2 Level (DSL) |
---|---|
Green | Red |
0.3 | DSL <= 0.08 |
0.5 | DSL <= 0.13 |
0.8 | DSL <= 0.21 |
Warning about change in temperature and static pressure
A warning message should be displayed with an indication if difference between
calibration conditions and daily conditions results in measurement error
larger than 0.2dB.
Message text example: “It is advised to compensate measurements, as change in
temperature and static pressure since Syscheck2 reference were made may cause
a measurement correction of more than 0.2dB”.
In the following, small changes in Syscheck2 user data is tracked by marking
changes in red color.
Function «Retrieve env parameter»
This function is used in the case where the user wants to retrieve
temperature, humidity and atmospheric pressure from the sensor present in the
microphone set.
The outcome of this function is readouts of current temperature (in °C),
humidity and atmospheric pressure:
General Timeline
Step by step Execution
// in this example we use a 246AE, user data contains
//“246AE {: Pid 00003F F Env 23.4 1008 47 RL -27.00 RT 23.1 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 }”
// Only data between “{:” and “ }” in user data is modified in the following.
All other TEDS
// data is preserved.
//note: If Pid is not present, function routine shall be aborted and user
notified that “system can not decide whether a Syscheck2 transducer is
present. Encourage to >Make Syscheck2 reference< to clear out”
1) Enter DIGITAL mode
2) Read from TEDS user data into local string
3) Localize Syscheck2 data Pid, Tc2,Tc, RL, RT and RP between “{:” and “
}”, these values are needed later on.
// Now, we want to setup TEDS to retrieve the env parameters
// concatenate Pid, Tc2, Tc, RL, RT and RP together with the command for
env
4) Replace Syscheck2 data in TEDS user data with: “{: pid 00003F F env
23.4 1008 47 RL -27.00 RT 23.1 RP 1013 Tc2 -96.0E-6 Tc 16.1E-3 }”
// I.e. only the upper case Pid and Env is changed with the lower case
pid and env
// note: preserve user data outside Syscheck2 data i.e. the content of
TEDS user data will in
// this example be
// “246AE {: pid 00003F F env 23.4 1008 47 RL -27.00 RT 23.1 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 }”
5) Exit DIGITAL mode → Enter ANALOG mode
6) Wait 2000 ms in order for Syscheck 2 to process
// In this example Syscheck 2 process will update the user data with
environment information
// “246AE {: pid 00003F F env 23.4 1008 47 RL -27.00 RT 23.1 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 }”
// to
// “246AE {: Pid 00003F F Env 27.4 1008 40 RL -27.00 RT 23.1 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 }”
7) Exit ANALOG mode -> Enter DIGITAL mode
8) Read user data eg.: “246AE {: Pid 00003F RT 23.1 RP 1013 Tc2 – 96.0E-6
Tc 16.1E-3 }”, where t=27.4, p=1008 and h=40 are the newly sensor measured
values of temperature, pressure and humidity.
//note: If env is not changed to Env by the sensor during the procedure in 6),
environmental data are not updated.
9) Log Env data locally (t, p, h) + standard TEDS information (Model &
serial no & sensitivity) + other data of interest (e.g. time stamp or other
system integrator parameters
10) Exit DIGITAL mode → Enter ANALOG mode
11) * allow minimum 5 seconds for ANALOG microphone state to stabilize
before starting measurement
Function «Make Syscheck2 reference»
This function is used in the case where the user wants to create a reference
Syscheck2 level (after the user has ensured that the system is measuring
correct i.e. calibration has been verified).
The integrator software will analyze the level of the sinusoidal tone produced
by the microphone set internal generator.
The outcome of this function is to generate a new Syscheck2 Reference Level
and Reference temperature to be stored in TEDS.
Only the data present in the user data allocated place in the TEDS chip will
be actively used.
General Timeline
Step by step Execution
// in this example user data contains
// “246AE {: Pid 00003F F Env 23.4 1008 47 RL -27.00 RT 23.1 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 G 10 }”
// Only data between “{:” and “ }” in user data is modified in the
following. All other TEDS
// data is preserved.
1) Enter DIGITAL mode
2) Read from TEDS user data and store into local string
3) It should be tested whether a Syscheck2 transducer is present
3.1 Inform user that user data will be overwritten in the initialization
process
3.2 Write following to user data: “{: pid 00003F }”
// note: pid will be updated to Pid by Syscheck2 enabled transducer
3.3 Enter analog mode for 2000 ms and switch back to digital mode to let
transducer update
3.4 Read user data and check if pid is updated with “Pid 00003F” (which
means Syscheck2 is alive) and jump to 5)
3.5 If pid is not updated, restore TEDS user data from local string in 2)
into user data of the transducer for bringing user data back to untouched
state. Inform user that no Syscheck2 transducer is present and exit >>Make
Syscheck2 reference<< function.
4) Now, we want to setup TEDS to retrieve env parameters and generate a
tone from generator at the default frequency 250Hz
5) Replace Syscheck2 data in TEDS user data with: “{: Pid 00003F f env
27.4 1008 40 RL -27.00 RT 23.1 RP 1013 tc2 -96.0E-6 tc 16.1 E-3 }”, note:
env, tc2 and tc will be updated by 246AE/O
6) Exit DIGITAL mode → Enter ANALOG mode
7) Wait 5000 ms (The system is asynchronous and we need to wait long
enough for fetching and writing the env data, for the generator to start and
for the microphone analog mode to stabilize)
8) Acquire 3000 ms of measurement data;
// Find dBV level of measurement data around 250Hz +/- 3% by
// system integrators “standard” routine for calibration with narrow band
filter to reduce noise.
// store dBV level as “measured_SC_level” *see note
9) Exit ANALOG mode → Enter DIGITAL mode
10) Read user data eg.: “{: Pid 00003F F Env Tc2 -96.0E-6 Tc 16.1E-3 }”
11) Localize Syscheck2 data Tc2, Tc, and derive p and t from Env between
“{:” and “ }”, these values are needed for step 13
12) Set RL=measured_SC_level; eg. -26.95 Set RT=t; eg. 21.5Set RP=p; eg.
1013
13) Write TEDS Syscheck2data: “{: Pid 00003F F Env 21.5 1013 RL -26.95 RT
21.5 RP 1013 Tc2 -96.0E-6 Tc 16.1E-3 b3}”
// b3 meaning blue LED ON for 3 seconds, indicating Syscheck2 ref is
done.
14) Exit DIGITAL mode → Enter ANALOG mode
15) Log (optionally) data locally (RL, RT, RP, h) + standard TEDS
information (Model & serial no & sensitivity) + e.g. time
16) Present “Syscheck2 reference captured” screen to user.
Function «Syscheck2»
This function is used in the case where the user has already created a
reference Syscheck2 level and wants to compare it to the actual microphone
set.
The system integrator software will analyze the level of a tone produced by
the microphone set, correct it with the current temperature, and compare it to
the Syscheck2 reference level.
Also, a warning should indicate if temperature is too far from reference level
temperature.
The outcome of this function is to perform a system check.
General Timeline
Step by step Execution
// in this example user data contains
// “246AE {: Pid 00003F F Env 27.4 1008 40 RL -27.00 RT 23.1 RP 1013 Tc2
-96.0E-6 Tc 16.1E-3 G 010 }”
// note: If “Pid 00003F” is not present, function routine shall be
aborted and user notified that “system can not decide whether a Syscheck2
transducer is present. Encourage to perform >>Make Syscheck2 reference<< to
clear out”
1) Enter DIGITAL mode
Only data between “{:” and “ }” in user data is modified in the following. All
other TEDS data is preserved.
2) Read from TEDS user data into local string
3) Localize Syscheck2 data Pid, Tc2, Tc, RL, RT and RP between “{:” and “
}”, store each variable locally. If Tc2, Tc, RL, RT and RP is not present in
user data, inform user that Syscheck2 data is missing and encourage to perform
Make Syscheck2 reference<< to clear out” and exit >>Syscheck2<< function.
// If Tc2, Tc, RL, RT and RP are present, we want to proceed and setup TEDS to retrieve env parameters and generate a tone from generator
4) Write string to TEDS user data: “246AE {: Pid 00003F f env 27.4 1008 40 RL -27.00 RT 23.1 RP 1013 Tc2 -96.0E-6 Tc 16.1E-3 G 010 }”?
// F and Env is changed to lower case f and env
5) Exit DIGITAL mode → Enter ANALOG mode
6) Wait 5000 ms (wait long enough for fetching and writing the env data,
for the generator to start and for the microphone analog mode to stabilize)
7) Acquire 3000 ms of measurement data;
// Find dBV level of measurement data around 250Hz +/- 3% by
// system integrator “standard” routine for calibration with narrow band
filter to reduce noise.
// store dBV level as “SC_level_measured” ?
*see note
8) Exit ANALOG mode → Enter DIGITAL mode
// TEDS user data should now contain “246AE {: Pid 00003F F Env
RL -27.00 RT 23.1 RP 1013
Tc2 -96.0E-6 Tc 16.1E-3 G 010 }
9) Read userdata: “Env” and split into t, p and h variables and store
locally for use in the following.
10) Calculate SC_level_corrected SC_level_corrected = SC_level_measured –
((t) 2 Tc2 + t Tc – ((RT)2 Tc2 + RT Tc) )
Note: Tc and Tc2 has been read from Syscheck2 transducer.
Use the following criteria for evaluation whether microphone is within
chosen acceptance level: Calculate DSL, which is the distance from recent
SC_level_corrected to the microphone Syscheck2 reference level RL. DSL = |
SC_level_corrected – RL |
Acceptance Level | Syscheck2 Level (DSL) |
---|---|
Green | Red |
0.3 | DSL <= 0.08 |
0.5 | DSL <= 0.13 |
0.8 | DSL <= 0.21 |
11) ENTER DIGITAL mode and write evaluation to TEDS
// if GREEN: write userdata:
//“246AE {: “246AE {: Pid 00003F F Env 21.5 1008 54 RL -27.00 RT 23.1 RP
1013 Tc2 – 96.0E-6 Tc 16.1E-3 g 010 }”
// If RED: write userdata:
//“246AE {: “246AE {: Pid 00003F F Env 21.5 1008 54 RL -27.00 RT 23.1 RP
1013 Tc2 – 96.0E-6 Tc 16.1E-3 r 010 }”
12) Exit DIGITAL mode
13) Log data (optionally) locally (t, p, h, SC_level_corrected) +
standard TEDS information (Model & serial no & sensitivity) + time On system
integrator side
14) Present Syscheck2 evaluation to user.
// Green = Syscheck2 OK, Red = Syscheck2 NOK If microphone sensitivity
correction (see notes) is more than 0.2dB then display the message: “It is
advised to compensate microphone sensitivity, as change temperature and static
pressure since Syscheck2 reference were made may cause a measurement
correction of more than 0.2dB”.
NOTES:
- If either one of the following variables Pid, Tc2, Tc, RL, RT and RP are not present in the TEDS user data the Syscheck2 can not be evaluated and should terminate with a note to the user.
- All variables (Pid, Tc2, Tc, RL, RT and RP) can be created by calling >>Make Syscheck2 reference<<
- Background noise should be lower than 65dBspl for 246AE and 77dBspl for 246AO
- The temperature sensor has reduced specifications outside [0…65 °C]
- If the temperature is higher than 85 °C, the env value is set to 85. User should be alerted.
- If the current temperature and atmospheric pressure results in change in microphone sensitivity of more than 0.2dB, since Syscheck2 reference were made, the user should be notified;
Sensitivity correction = |(t-RT)sensTC + (p-RP)sensPC |
If Sensitivity correction is more than 0.2dB then display the message :
“It is advised to compensate microphone sensitivity, as change temperature and
static pressure since Syscheck2 reference were made may cause a measurement
correction of more than 0.2dB”.
RL = Syscheck2 reference level, designate with 2 decimals
RT= Syscheck2 reference temperature, designate with 1 decimals
RP= Syscheck2 reference atmospheric pressure, designate with 0 decimals
Vmeas= measured output from analysator [Vrms], uncorrected for
environment.
t = Measured output from the temperature sensor (env)
TC = Syscheck2 temperature coefficient, read from user data
TC2 = Syscheck2 temperature coefficient 2, read from user data
| sensTC| sensPC
---|---|---
246AE| -0.01| 0.0014
246AO| -0.01| 0.0007
Environment correction coefficients, valid for 250Hz only.
SC_level_corrected = SC_level_measured – ((t)2 TC2 + t TC – ((RT)2 TC2 + RT TC) )
GRAS TEDS editor SW0051
SW0051 is available as download on request.
After startup of editor, select interface and Read TEDS:
Example of TEDS with commands in user data: