Hunan Accurate Bio-Medical HS20 Handheld Pulse Oximeter Instruction Manual
- June 12, 2024
- Hunan Accurate Bio-Medical
Table of Contents
- Safety
- The Basics
- Getting Started
- Basic Operations
- Notices
- Measuring SpO2
- Battery
- Maintenance and Cleaning
- Accessories
- Clinical study summary
- A Product Specifications
- B EMC
- C Factory Defaults
- D Notice Messages
- E SpO2 Sensor Model
- F Symbols and Abbreviations
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Hunan Accurate Bio-Medical HS20 Handheld Pulse Oximeter Instruction Manual
Product Information
Product Name: Pulse Oximeter
Product Model: HS20A, HS20A-LI
Manufacturer: Hunan Accurate Bio-Medical Technology Co., Ltd.
Address: Accurate Industrial Park, No.108, Zhixian Road, Xuelian
Community, Xueshi Street of Yuelu District, 410208 Changsha, Hunan Province,
PEOPLE’S REPUBLIC OF China
Tel: +86-731-85598539
Software Version :V1.0
Shelf life:2 years
EUROPEAN REPRESENTATIVE:
Name:Shanghai International Holding Corp. GmbH (Europe)
ADD:Eiffestrasse 80, 20537 Hamburg,Germany
Tel: +49-40-2513175
Fax: +49-40-255726
Revision History
This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical specification change. Contents of this manual are subject to change without prior notice.
- Revision number: 1.1
- Release time: 2023-8
© Copyright
Hunan Accurate Bio-Medical Technology Co., Ltd.
All rights reserved.
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely
and in accordance with its function and intended use.
Observance of this manual is a prerequisite for proper product performance and
correct operation and ensures patient and operator safety.
This manual is based on the maximum configuration and therefore some contents
may not apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close
to the equipment so that it can be obtained conveniently when needed.
This product is a reusable medical device.
Intended Audience
This manual is geared for clinical professionals who are expected to have a
working knowledge of medical procedures, practices and terminology as required
for monitoring of critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not
necessarily reflect the setup or data displayed on your pulse oximeter.
Conventions
Italic text is used in this manual to quote the referenced chapters or
sections.
→ is used to indicate operational procedures.
[ ] is used to enclose screen texts.
Safety
1.1 Safety Information
DANGER
- Indicates an imminent hazard that, if not avoided, will result in death or serious injury.
- Indicates a statement that there may be hazards in using different alarm presets for the same or similar devices in any signal area.
WARNING
- Indicates a potential hazard or unsafe practice that, if not avoided, will result in death or serious injury.
- Indicates a warning to the effect that probes and cables are designed for use with specific monitors.
- Indicates a warning statement to the effect that the responsible organization or operators needs to verify the compatibility of the monitor, probe or cables before use, or patient injury can result.
- Indicates a warning statement to the effect that misapplication of a probe with excessive pressure for prolonged periods can induce pressure injury
CAUTION
- Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.
- Indicates a statement to the effect that a FUNCTIONAL TESTER cannot be used to access the ACCURACY of a pulse oximeter probe or a pulse oximeter monitor
NOTE
- Provides application tips or other useful information to ensure that you get the most from your product.
- The pulse oximeter is not part of the body or tissue.
1.1.1 Dangers
There are no dangers that refer to the product in general. Specific “Danger” statements may be given in the respective sections of this manual.
1.1.2 Warnings
WARNINGS
-
Before putting the system into operation, verify that the equipment, connecting cables and accessories are in correct working order and operating condition.
-
To avoid explosion hazard, do not use the equipment in the presence of flammable anesthetics, vapors or liquids.
-
Do not open the equipment housings; electric shock hazard may exist. All servicing and future upgrades must be carried out by the personnel trained and authorized by our company only.
-
Do not rely exclusively on the audible notice system for personnel monitoring. Adjustment of notice volume to a low level or off may result in a hazard to the personnel. Remember that notice settings should be customized according to different or personnel situations and always keeping the personnel under close surveillance is the most reliable way for safe personnel monitoring.
-
The physiological data and notice messages displayed on the equipment are for reference only and cannot be directly used for diagnostic interpretation.
-
To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and secure excess cabling to avoid risk of entanglement or strangulation by personnel.
-
No modification of this equipment is allowed.
-
The maximum temperature that the applied part can achieve is 42.2℃ at an environmental temperature of 40℃.
-
The continuous test time of one part shall not exceed 2 hours, and the test part shall be replaced.
-
Applied part having contact with the patient for a time “t”, 10 min ≤ t≤2 hours.
-
Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
-
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation.
If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally. -
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
-
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment could result. -
Use only accessories specified in this manual. Using other accessories may cause damage to the pulse oximeter.
-
Disposable accessories are designed for single-person use only. Reuse of them may cause a risk of contamination and affect the measurement accuracy
-
Check the accessories and their packages for any sign of damage. Do not use them if any damage is detected.
-
The SpO2 sensor mentioned in this chapter meet the bio-compatibility requirements and complies with ISO 10993-1, ISO 10993-5 and ISO10993-10 standards.
-
It should change the measurement part per 2 hours when user continuing using for detecting.
-
The measured value is best when the pulse rate waveform is smooth and stable after normalization. Data update period: ≤3 pulse rate cycles, < 30s.
-
FUNCTIONAL TESTER cannot be used to assess the ACCURACY of a PULSE OXIMETER .
-
The nature of Reminder signals is continue to turning on for 1s. And the intervals between Reminder signals is turn on 1 seconds and turn off 4 second. One cycle has 5 seconds.
1.1.3 Cautions
CAUTIONS
-
To ensure safety, use only parts and accessories specified in this manual.
-
At the end of its shelf life, the equipment, as well as its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have any questions concerning disposal of the equipment, please contact us.
-
Magnetic and electrical fields are capable of interfering with the proper performance of the equipment. For this reason make sure that all external devices operated in the vicinity of the equipment comply with the relevant EMC requirements.
Mobile phone, X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation. -
Always install or carry the equipment properly to avoid damage caused by drop, impact, strong vibration or other mechanical force.
1.1.4 Notes
NOTES
-
Put the equipment in a location where you can easily see the screen and access the operating controls.
-
Keep this manual in the vicinity of the equipment so that it can be obtained conveniently when needed.
-
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or re-orienting the equipment. -
This manual and the pictures on the color box are schematic diagrams, and the product is subject to the actual product.
1.2 Equipment Symbols
The Basics
2.1 Introduction
2.1.1 Intended Use
Monitoring of functional arterial oxygen saturation (SpO2) and pulse rate.
WARNING
- This pulse oximeter is intended for use only by clinical professionals or under their guidance. It must only be used by persons who have received adequate training in its use. Anyone unauthorized or untrained must not perform any operation on it.
2.1.2 Intended operator
The intended operator of the pulse oximeter are clinical professionals or under their guidance. It must only be uesed by persons who have received adequate training in its use. Anyone unauthorized or untrained must not perform any operation on it .
2.1.3 Intended Patient Population
The Pulse Oximeter is intended for adult, adolescent, child, infant and
neonatus patients.
2.1.4 Medical Condition
The Pulse Oximeter is intended to be used in hospitals and clinics.
2.1.5 Essential Performance
- Pulse rate.
- Blood oxygen saturation concentration.
WARNING
- Blood oxygen and pulse rate inaccurate measurement probably may lead to the doctors do false diagnosis.
2.1.6 Contraindications
Do not use oximeter in a magnetic resonance ( MR or CT )environment.
2.1.7 Components
This pulse oximeter consists of a main unit and an SpO2 sensor.
2.1.8 Operator Position
This device is a handheld portable pulse oximeter. Generally, the operator
should use it within half a meter from the device.
2.1.9 Validation
Generally, after operator turn on the device , and the beep will continue to
turn on for 1 second prove that it is working normally .
2.2 Main Unit
2.2.1 Front View
-
Display screen
-
Power button
♦ Press this button to turn the pulse oximeter on after the batteries are installed.
♦ Press and hold it for 2 seconds to turn off the pulse oximeter.
♦ Press this button for 1 second, the menu screen can quickly return to measurement interface. -
Up button
♦ Press this button to change the value of select menu item or raise the beat volume. -
Down button
♦ Press this button to change the value of select menu item or lower the beat volume. -
Power Indicating Lamp(Only for lithium)
It is a LED that lights green and yellow. The status of the LED is specified as follows:
♦ Green: The lithium battery of the device has been fully charged when the power adapter is connected to the device;
♦ Yellow: The lithium battery of the device is being charged when the power adapter is connected to the device;
♦ Off: The device configured with the lithium battery is not connected to the power adapter. -
Power Supply Connector(Only for lithium)
It is used to connect the AC adapter. -
Speaker
-
Setting button
Press this button to enter the main menu. -
Notice Silence Button
Press this button to pause or reactivate the notice sound. -
Notice indicating lamp
When an notice occurs, this lamp will light up as defined below:
- High level notices: the lamp quickly flashes red.
- Medium level notices: the lamp slowly flashes yellow.
2.2.2 Rear View and Right View
-
Sensor connector
It is used to connect an SpO2sensor to measure the oxygen saturation. -
Cord hole
-
Battery door
2.3 Display Views
The following figures show the layout of the normal screen.:
- Alarm information Area
- SpO2 label
- SpO2(%) Value
- Wave Area
- Tech Information Area
- User Type
- Function ICON Area
- PR(bpm) label
- User ID
- PR Value
- Pulse Bar
- Time
- Battery level
2.3.1 SpO2Area
- SpO2label and unit
- Oxygen saturation reading
2.3.2 PR Area
- PR label and unit
- Pulse rate reading
2.3.3 Information Area
- Physiological Alarm Area
- Technology Alarm Area
- Battery level
Getting Started
3.1 Unpacking and Checking
Before unpacking, examine the packing case carefully for signs of damage. If
any damage is detected, contact the carrier.
If the packing case is intact, open the package and remove the equipment and
accessories carefully. Check all materials as per the packing list and check
for any mechanical damage. Contact us in case of any problem.
NOTE
- Save the packing case and packaging material as they can be used if the equipment must be reshipped.
WARNING
- Keep the packing material out of children’s reach. Disposal of the packaging material should observe the applicable waste control regulations.
- The equipment might be contaminated during storage and transport. Before use, please verify whether the packages, especially the packages of the single use accessories, are intact. In case of any damage, do not apply it to patient.
3.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual. When the equipment is moved from one place to another, condensation may occur as a result of temperature or humidity difference. In this case, never start the system before the condensation disappears.
WARNING
- Make sure that the operating environment of the equipment meets the specific requirements. Otherwise the equipment may not meet the specifications claimed in this manual and unexpected consequences, e.g. damage to the equipment, could result.
3.3 Power On the Pulse Oximeter
- Before use, check whether there is any mechanical damage on the pulse oximeter.
- Install the alkaline batteries or the lithium ion battery and ensure that the batteries have sufficient power.
- Plug the SpO2extension cable into the multifunctional connector.
- Press the Power button. The pulse oximeter enters the main screen.
WARNING
- Do not use the pulse oximeter for monitoring if it is mechanically damaged or appears abnormal. Contact your service personnel or Accurate Bio-Medical.
3.4 Power Off the Pulse Oximeter
Shut down the pulse oximeter:
- Confirm that the person monitoring is finished.
- Disconnect the SpO2extension cable from the pulse oximeter.
- Press and hold the Power button for 2 seconds to shut down the device.
Basic Operations
4.1 Adjusting the Pulse Volume
To adjust the pulse volume:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [System].
- Press the [SET] button, enter the System Menu.
- Press the [Up]/[Down] button, select the [Sound volume].
- press the [SET] button, enter the Sound Volume setting.
- Press the [Up]/[Down] button, set a value between 0 and 10.
- Press the [SET] button, quit the Sound Volume setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
NOTE
- 0 means the pulse volume is turned off, and 10 is the maximum volume.
- Increase/decrease the pulse volume by pressing the Up/Down button in the case that no menu is opened.
- During SpO2 monitoring, the pitch of the pulse tone changes as the patient’s oxygen saturation level changes.
- The pitch of the tone rises as the oxygen saturation level increases and falls as the oxygen saturation level decreases.
4.2 Selecting Patient Type
WARNING
- Be sure to select correct patient type setting for your patient before measurement. Wrong patient category may result in patient hazard due to improper notice limits.
To admit a patient:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Patient].
- Press the [SET] button, enter the Patient Menu.
- Press the [Up]/[Down] button, select the [Patient Type].
- Press the [SET] button, enter the Patient Type setting.
- Pressing the [Up]/[Down] button, set to a value: ADU(Adult), PED(pediatric),NEO(neonate).
- Press the [SET] button, exit the Patient Type setting.
- Press the Power button for 1 second, the menu screen can quickly return to the measurement interface.
4.3 Setting Alarm Limit
4.3.1 Setting PR High Alarm Limit
To adjust the PR high alarm limit:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Alarm].
- Press the [SET] button, enter the Alarm Menu.
- Long press the [Up] button for 3s, and repeat three times.
- Press the [Up]/[Down] button, select the [PR Alarm High].
- press the [SET] button, enter the PR Alarm High setting.
- Press the [Up]/[Down] button, set a value between (low limit+1) and 250.
- Press the [SET] button, quit the PR Alarm High setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
4.3.2 Setting PR Low Alarm Limit
To adjust the PR low alarm limit:
- .Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Alarm].
- Press the [SET] button, enter the Alarm Menu.
- Long press the [Up] button for 3s, and repeat three times.
- Press the [Up]/[Down] button, select the [PR Alarm High].
- press the [SET] button, enter the PR Alarm Low setting.
- Press the [Up]/[Down] button, set a value between 25 and (high limit – 1).
- Press the [SET] button, quit the PR Alarm Low setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
4.3.3 Setting SPO2 High Alarm Limit
To adjust the SpO2 high alarm limit:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Alarm].
- Press the [SET] button, enter the Alarm Menu.
- Long press the [Up] button for 3s, and repeat three times.
- Press the [Up]/[Down] button, select the [SPO2 Alarm High].
- press the [SET] button, enter the SPO2 Alarm High setting.
- Press the [Up]/[Down] button, set a value between (low limit+1) and 100.
- Press the [SET] button, quit the SPO2 Alarm High setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
4.3.4 Setting SPO2 Low Alarm Limit
To adjust the SpO2 low alarm limit:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Alarm].
- Press the [SET] button, enter the Alarm Menu.
- Long press the [Up] button for 3s, and repeat three times.
- Press the [Up]/[Down] button, select the [SPO2 Alarm Low].
- press the [SET] button, enter the SPO2 Alarm Low setting.
- Press the [Up]/[Down] button, set a value between 88 and (high limit – 1).
- Press the [SET] button, quit the SPO2 Alarm Low setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
4.4 Data Storage
- It can save 10 patents’ data. There are 600 pcs storage of measured result for each patient.
- If the measured result have been saved successfully, user can see the graph or table in the menu “Graph” or “Table”
4.4.1 Date Storage ON/OFF
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Data].
- Press the [SET] button, enter the Data Menu.
- Press the [Up]/[Down] button, select the [Storage].
- Press the [SET] button, enter Storage setting.
- Press the [Up]/[Down] button, set to a status: ON or OFF
- Press the [SET] button, quit the Storage setting.
- Press the [Power] button for 1 second, the menu screen can quickly return to the measurement interface.
4.4.2 Date Storage Time Step
Save a measured result every Data Storage Time Step
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Data].
- Press the [SET] button, enter the Data Menu.
- Press the [Up]/[Down] button, select the [Storage Time].
- Press the [SET] button, enter Storage Time setting.
- Press the [Up]/[Down] button, set to a value between 6 to 60,step 2.
- Press the [SET],quit the Storage Time setting.
- Press the [Power] button for 1 second, the menu screen can quickly return to normal operation display.
Notices
Notices triggered by a vital sign that appears abnormal or by technical problems of the Pulse Oximeter, are presented to the user by visual and audible notice indications.
5.1 Notice Categories
By nature, the pulse oximeter’s notices can be classified into three categories: physiological notices, technical notices and prompt messages.
-
Physiological notices
Physiological notices, also called patient status notices, are triggered by a monitored parameter value that violates set notice limits or an abnormal patient condition. -
Technical notices
Technical notices, also called system status notices, are triggered by a device malfunction or a patient data distortion due to improper operation or system problems. -
Prompt messages
In fact, prompt messages are not notice messages, which are displayed in the technical notice area. Apart from the physiological and technical messages, the pulse oximeter will show some messages which indicate the system status.
5.2 Notice Levels
By severity, the pulse oximeter’s physiological notices can be classified into three categories: high level notices, medium level notices.
-
High level notices
Indicate that the patient is in a life threatening situation and an emergency treatment is demanded. -
Medium level notices
Indicate that the patient’s vital signs appear abnormal and an immediate treatment is required.
5.3 Notice Indicators
When an notice occurs, the pulse oximeter will indicate it through the following indications:
- Notice lamp
- Notice tone
- Notice message
The notice lamp, tone and messages will be indicated differently according to different notice levels.
5.3.1 Notice Lamp
If a technical or a physiological notice occurs, the notice lamp will flash. The flashing color and frequency match the notice level as follows:
- High level notices: the lamp quickly flashes red.
- Medium level notices the lamp slowly flashes yellow.
5.3.2 Notice Tones
When a technical or a physiological notice occurs, the pulse oximeter presents different notice tone patterns to match the notice level:
- High level notices: triple + double + triple + double beep
- Medium level notices: triple beep.
5.3.3 Notice Messages
When an notice occurs, an notice message will appear in the technical or physiological notice area.
NOTE
When multiple notices of different levels occur simultaneously, the pulse
oximeter will select the notice of the highest level and give visual and
audible notice indications accordingly.
5.4 Notice Tone Configuration
5.4.1 Switch ON/OFF Notice Volume
To switch ON/OFF notice volume:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Alarm].
- Press the [SET] button, enter the Alarm Menu.
- Long press the [Up] button for 3s, and repeat three times.
- Press the [Up]/[Down] button, select the [Alarm].
- press the [SET] button, enter the Alarm ON/OFF setting.
- Press the [Up]/[Down] button,set to a status :ON or OFF.
- Press the [SET] button, quit the Alarm setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
WARNING
-
When the notice sound is switched off, the pulse oximeter will give no audible notice tones even if a new notice occurs.
Therefore the user should be very careful about whether to switch off the notice sound or not. -
Do not rely exclusively on the audible notice system for patient monitoring. Adjusting notice volume to a low level may result in a hazard to the patient. Always keep the patient under close surveillance.
5.4.2 Pausing the Notice Tones
To pause the notice tones, press the key for 1 second.
In this case:
- The audible notice is paused, but the key remains lit and the notice messages remain displayed.
- The remaining notice pause time is 60 seconds.
Pause time can be set as follow:
- Press [SET] button, enter the Main Menu.
- Press the [Up]/[Down] button, select the [Alarm].
- Press the [SET] button, enter the Alarm Menu.
- Long press the [Up] button for 3s, and repeat three times.
- Press the [Up]/[Down] button, select the [Alarm Pause Time].
- press the [SET] button, enter the Alarm Pause Time setting.
- Press the [Up]/[Down] button, set to value:30s,60s,90s,120s.
- Press the [SET] button, quit the Alarm Pause Time setting.
- Press the Power button for 1 second, the menu screen will quickly return to the measurement interface.
5.5 When an Notice Occurs
When an notice occurs, observe the following steps to take proper actions:
- Check the patient’s condition.
- Confirm the notice parameter or notice category.
- Identify the source of the notice.
- Take proper action to eliminate the notice condition.
- Make sure the notice condition is corrected.
For troubleshooting specific notices, see appendix D Notice Messages.
Measuring SpO2
6.1 Introduction
SpO2measuring is a non-invasive technique used to measure the amount of oxygenated haemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light emitted by a red and infrared light-emitting diodes passes through the tissue and is converted into electrical signals by a photodiode. This device is calibrated to display functional oxygen saturation.
The Pulse Oximeter provides:
- Oxygen saturation of arterial blood (SpO2): It is the percentage of oxygenated hemoglobin in relation to the sum of oxyhemoglobin and deoxyhemoglobin.
- Pulse rate (PR): It is the pulsations per minute derived from the Pleth wave.
- Pulse Bar : The number of segments indicates the pulse strength and signal quality.
6.2 Safety
WARNINGS
-
Use only SpO2sensors specified in this manual. Follow the SpO2 sensor’s instructions for use and adhere to all warnings and cautions.
-
Check the SpO2 sensor and its package for any sign of damage before use. Do not use the sensor if any damage is detected.
-
When a trend toward deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely understand the person’s condition.
-
Do not use the pulse oximeter and the SpO2sensor during magnetic resonance imaging (MRI). Induced current could cause burns.
-
Reusable sensor must be moved to new site at least every four hours. Because individual skin condition affects the ability of skin to tolerate sensor placement, it may be necessary to change the sensor site more frequently with some patients. If skin integrity changes, move the sensor to another site.
Do not apply tape to secure the sensor in place or to tape it shut; venous pulsations may lead to inaccurate saturation measurements. -
As with electrosurgical unit, carefully route patient cabling to avoid entanglement.
-
Do not use the SpO2sensor on a limb with an intravenous infusion or arterial catheter in place.
-
Do not use the SpO2sensor on a limb where the NIBP cuff is applied. This may result in inaccurate SpO2reading due to blocked blood flow during cuff inflation.
6.3 Applying the Sensor
- Select an appropriate SpO2sensor according to the person category and weight.
- Apply the SpO2 sensor to the patient.
- Connect the SpO2extension cable to the pulse oximeter.
- This equipment is recommended to use ACCARE Disposable SpO2 Sensor or Reusable SpO2 Sensor(see Annex C).Operator needs to verify the compatibility of the monitor, sensor and cable before use, or patient injury can result; misapplication of a sensor with excessive pressure for prolonged periods can induce pressure injury.
6.4 Settings
6.4.1 Setting Patient Type
Please see to the “4.2 Selecting Patient Type”
6.4.2 Adjusting the Notice Limits
Please see to the “4.3 Setting Alarm Limit”
6.4.3 Setting VOL
Press button [SET]→[VOL], then pressing the Up/Down button to set the expected
Pulse Beat volume.
6.5 Measurement Limitations
If you doubt the SpO2measurements, check the personnel or person’s vital signs
first. Then check the pulse oximeter and SpO2 sensor.
The following factors may influence the accuracy of measurements:
- Ambient light
- Physical movement (person or imposed motion)
- Diagnostic testing
- Low perfusion
- Electromagnetic interference, such as MRI environment
- Electrosurgical units
- Dysfunctional haemoglobin, such as carboxyhemoglobin (COHb) and methemoglobin (MetHb)
- Presence of certain dyes, such as methylene and indigo carmine
- Inappropriate positioning of the SpO2 sensor, or use of incorrect SpO2
- Drop of arterial blood flow to unmeasurable level due to shock, anemia, low temperature or vasoconstrictor
Battery
7.1 Overview
The pulse oximeter is designed to operate on three 1.5V alkaline AA batteries or a rechargeable lithium battery.
NOTE
-
Remove the batteries prior to shipping or if the pulse oximeter is not likely to be used for an extended period of time.
WARNING -
Keep the batteries out of children’s reach.
-
Use only batteries specified in this manual.
7.2 Installing the Batteries
7.2.1 Opening the Battery Door
1. Press the battery door, push it downwards and remove the battery door.
7.2.2 Installing the Alkaline Batteries
- Insert the AA alkaline batteries in the battery compartment, aligning the + on each battery with the + shown inside the battery compartment.
- Close the battery door and push it upwards.
Caution
- Do not run the pulse oximeter using alkaline batteries of different types or capacities at the same time.
7.2.3 Charging the Lithium Batteries
To charge the lithium battery.
-
Connect the plug of AC adaptor to the power supply connector of the pulse oximeter.
-
Plug the AC adaptor cord into the AC mains.
WARNINGS
- Do not use the charger stand when the alkaline batteries is depleted or no battery is installed.
- Monitoring a patient while the battery is being charged is not recommended.
7.3 Checking The Lithium Battery
The performance of rechargeable lithium battery may deteriorate over time. To check the performance of a battery, follow this procedure:
- Disconnect the pulse oximeter from the patient and stop all monitoring and measuring procedures.
- Place the pulse oximeter in the Charger stand and connect the AC mains. Allow the battery to be charged uninterruptedly for above 2 hours.
- Disconnect AC mains and allow the pulse oximeter to run on the battery until shuts off.
The operating time of a battery reflects its performance directly. If the operating time of a lithium battery is noticeably shorter than that stated in the specifications, replace it or contact your service personnel.
NOTES
- Lithium batteries can be recharged 200 times and have a lifespan of approximately 2 years. Improper use of batteries may lead to shortened battery life. It is recommended to replace the lithium battery every 2 years or when the charging frequency exceeds 200 times.
- The built-in lithium battery of the device must be replaced by authorized maintenance personnel from our company. Do not replace it on your own.
- The operating time of a lithium battery depends on the configuration and operation of the pulse oximeter.
7.4 Disposing Of The Batteries
Batteries that are damaged or depleted should be replaced and discarded properly. Dispose of used batteries according to local regulations.
WARNINGS
- Do not disassemble the battery, throw it into fire, or short-circuit it, as it may cause accidental injury.
- Replacing lithium batteries by personnel without sufficient training may lead to battery burning, explosion, or leakage, which may cause accidental injury.
Maintenance and Cleaning
Use only the substances approved by us and methods listed in this chapter to clean or disinfect your equipment. Warranty does not cover damage caused by unapproved substances or methods. We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the method to control infection, consult your hospital’s Infection Control Officer or Epidemiologist.
Keep you equipment and accessories free of dust and dirt. To avoid damage to the equipment, follow these rules:
- Always dilute according the manufacturer’s instructions or use lowest possible concentration.
- Do not immerse part of the equipment in the liquid.
- Do not pour liquid onto the equipment or accessories.
- Do not allow liquid to enter the case.
- Never use abrasive materials (such as steel wool or silver polish), or erosive cleaners (such as acetone or acetone based cleaners)
WARNING
- Be sure to shut down the system and disconnect all power cables from the outlets before cleaning the equipment.
CAUTION
- If you spill liquid onto the equipment or accessories, contact us or your service personnel.
NOTE
- To clean or disinfect reusable accessories, refer to the instructions delivered with the accessories.
8.1 Safety Checks
Before first use, or at least every year, or whenever your pulse oximeter is repaired or upgraded, a thorough inspection should be performed by qualified service personnel to ensure the reliability.
Follow these guidelines when inspecting the equipment:
- Make sure that the environment and power supply meet the requirements.
- Inspect the equipment and its accessories for mechanical damage.
- Inspect all power cords for damage, and make sure that their insulation is in good condition.
- Make sure that only specified accessories are applied.
- Inspect if the notice system functions correctly.
- Make sure that the batteries meet the performance requirements.
- Make sure that the pulse oximeter is in good working condition.
In case of any damage or abnormity, do not use the pulse oximeter. Contact your hospital’s biomedical engineers or your service personnel immediately.
8.2 Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations for cleaning the equipment.
Recommended cleaning agents are:
- Mild soap (diluted)
- Ammonia (diluted)
- Sodium hypochlorite bleach (diluted)
- Hydrogen peroxide (3%)
- Ethanol (70%)
- Isopropanol (70%)
To clean your equipment, follow these rules:
- Shut down the pulse oximeter and disconnect it from the power.
- Clean the display screen using a soft, clean cloth dampened with a glass cleaner.
- Clean the exterior surface of the equipment using a soft cloth dampened with the cleaner.
- Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
- Dry your equipment in a ventilated, cool place.
8.3 Disinfecting
Disinfection may cause damage to the equipment and is therefore not recommended for this pulse oximeter unless otherwise indicated in your hospital’s servicing schedule. Clean the pulse oximeter before disinfecting it. The recommended disinfectants include: ethanol 70%, isopropanol 70%, glutaraldehyde-type 2% liquid disinfectants to wiping equipment.
CAUTION
- Never use EtO or formaldehyde for disinfection.
8.4 Disposal
Dispose of the pulse oximeter in accordance with local environment and waste
disposal laws and regulations. For the disposal of SpO2 sensor, follow local
regulations regarding disposal of hospital waste.
Follow local disposal and recycling laws for the oximeter and its components,
including the battery, packaging waste.
Accessories
Clinical study summary
This Pulse oximeter has completed the clinical study in Xiangya Third Hospital of Central South University. The study included 12 subjects – 8 women and 4 men. Subjects are healthy, 22-28 years old, and their skin color includes Medium, Medium dark, dark and light.
The below table shows statistic conclusion of an invasive controlled desaturation study which guided by “ISO 80601-2-61,Annex EE, Guideline for evaluating and documenting SpO2 accuracy in human subjects”. The statistic result displayed the accuracy distribution between the range of 70%~100%, which maybe helpful to user.
A Product Specifications
B EMC
The device meets the requirements of IEC 60601-1-2
Transmission power:0dBm
Wireless frequency range:2402MHz~2480MHz
Hereby, [Hunan Accurate Bio-Medical Technology Co., Ltd.], declares that this
[HS20A,HS20A-LI] is in compliance with the essential requirements and other
relevant provisions of RE Directive 2014/53/EU. A copy of the full DoC is
attached.
CAUTION
- Using accessories, transducers and cables other than those specified may result in increased electromagnetic emission or decreased electromagnetic immunity of the person monitoring equipment.
- The device or its components should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device or its components should be observed to verify normal operation in the configuration in which it will be used.
- The device needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided below.
- Other devices may affect this pulse oximeter even though they meet the requirements of CISPR.
- When the inputted signal is below the minimum amplitude provided in technical specifications, erroneous measurements could result.
- Portable and mobile communication equipment will have impact on the performance of the pulse oximeter.
WARNINGS
- This device should not be used in the vicinity or on the top of other electronic equipment such as cell phone, transceiver or radio control products. If you have to do so, the device should be observed to verify normal operation.
- The use of accessories and power cord other than those specified, with the exception of cables sold by the manufacturer of the equipment or system as replacement parts for internal components, may result in increased emissions or decreased immunity of the equipment or system.
C Factory Defaults
C.1 System Setup
C.2 SpO2 Setup
D Notice Messages
This section lists only the most important physiological and technical notice messages. In the tables below, “H” means high, “M” means medium. The “Cause and actions” column gives recommendations to instruct you to troubleshoot the problems. If the problem persists, contact your service personnel.
D.1 Physiological Notice Messages
D.2 Technical Notice Messages
E SpO2 Sensor Model
E.1 Disposable SpO2 Sensor
E.2 Reusable SpO2 Sensor
F Symbols and Abbreviations
F.1 Units
F.2 Symbols
– minus
– negative
% percent
/ per; divide; or
+ plus
= equal to
< less than
greater than
≤ less than or equal to
≥ greater than or equal to
± plus or minus
× multiply
© copyright
F.3 Abbreviations
CISPR International Special Committee on Radio Interference
EEC European Economic Community
EMC Electromagnetic Compatibility
ID Identification
IEC International Electrotechnical Commission
LCD Liquid Crystal Display
LED Light Emitting Diode
MDD Medical Device Directive
PC Personal Computer
PR Pulse Rate
RF Radio Frequency
SpO2 Arterial Oxygen Saturation from Pulse Oximeter
Copy Right. Hunan Accurate Bio-Medical Technology Co., Ltd
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>