testo 175 T1 Set Temperature Data Logger Instruction Manual
- June 12, 2024
- testo
Table of Contents
testo 175 · Data loggers
Instruction manual
Safety and the environment
2.1. About this document
Use
Please read this documentation through carefully and familiarize yourself with the product before putting it to use. Pay particular attention to the safety instructions and warning advice in order to prevent injuries and damage to the products.
Keep this document to hand so that you can refer to it when necessary.
Hand this documentation on to any subsequent users of the product.
Symbols and writing standards
Representation | Explanation |
---|---|
Warning advice, risk level according to the signal word: |
Warning! Serious physical injury may occur.
Caution! Slight physical injury or damage to the equipment may occur.
Implement the specified precautionary measures.
| Note: Basic or further information.
1. …
2. …| Action: more steps, the sequence must be followed.
…| Action: a step or an optional step.
– …| Result of an action.
Menu| Elements of the instrument, the instrument display or the program interface.
[OK]| Control keys of the instrument or buttons of the program interface.
… | …| Functions/paths within a menu.
“…”| Example entries
2.2. Ensure safety
> Only operate the product properly, for its intended purpose and within
the parameters specified in the technical data. Do not use any force.
Never use the instrument to measure on or near live parts.
Before each measurement check that the connections are correctly closed with blanking plugs or that appropriate sensors have been correctly plugged in.
The protection class in the technical data specified for the corresponding instrument may otherwise not be reached.
testo 175 T3 : The maximum permissible difference in potential between the sensor inputs is 50 V. Take this into account when using surface sensors with nonisolated thermocouple.
After the final measurement, allow probes and probe shafts to cool down sufficiently in order to avoid burns from the hot sensor tip or the probe shaft.
Temperatures given on probes/sensors relate only to the measuring range of the sensors. Do not expose handles and feed lines to any temperatures in excess of 70 °C unless they are expressly permitted for higher temperatures.
Carry out only the maintenance and repair work on this instrument that is described in the documentation.
Follow the prescribed steps exactly. Use only original spare parts from Testo.
Do not use the device in a polluted environment (heavily dusty, oil, foreign matter, volatile chemicals).
2.3. Protecting the environment
> Dispose of faulty rechargeable batteries/spent batteries in accordance
with the valid legal specifications.
At the end of its useful life, send the product to the separate collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
Specifications
3.1. Use
Data loggers testo 175 are used to store and to read out individual
readings and measurement series. With testo 175 measurement values are
measured, saved and transmitted to the PC via USB cable or SD card, where they
can be read out and analysed with the software testo Comfort Software. With
the software the data loggers can also be individually programmed.
Typical applications
testo 175 T1 and testo 175 T2 are optimally suitable for temperature
measurements in refrigerators, freezers, cold storage rooms and cooling
shelves. testo 175 T3 records two temperatures at the same time and is most
suitable e.g. for monitoring the temperature spreading between feed and return
flow in a heating system.
testo 175 H1 controls the climatic conditions e.g. in warehouses, offices and
in the production area.
3.2. Technical data
testo 175 T1 (0572 1751)
Feature | Values |
---|---|
Measurement parameter | Temperature (°C/°F) |
Sensor type | NTC temperature sensor internal |
Measurement range | -35 to +55 °C |
System accuracy | ±0.4 °C (-35 to +55 °C) ± 1 digit |
Resolution | 0.1 °C |
Operating temperature | -35 to +55 °C |
Storage temperature | -35 to +55 °C |
Measurement parameter | Temperature (°C/°F) |
Battery type | 3x battery type AAA or Energizer L92 AAA-size cells |
Life | 3 years (15 min. measuring cycle, +25 °C) |
Degree of protection | IP 65 |
Dimensions in mm (LxWxH) | 89 x 53 x 27 mm |
Weight | 130g |
Housing | ABS/PC |
Measuring cycle | 10s – 24h (freely selectable) |
Interface | Mini-USB, SD card slot |
Memory capacity | 1 million readings |
EU Directive | 2014/30/EU, complies with the EN standard128306 |
testo 175 T2 (0572 1752)
Feature | Values |
---|---|
Measurement parameter | Temperature (°C/°F) |
Sensor type | NTC temperature sensor internal and extemal |
Measurement range | -35 to +55 °C internal -40 to +120 °C external |
System accuracy
Instrument accuracy| ±0.5 °C (-35 to +55 °C) ± 1 digit ±0.3 °C (-40 to +120
°C) ± 1 digit
Resolution| 0.1 °C
6 Please note that, according to EN 12830, this instrument must be regularly checked and calibrated as specified in EN 13486 (recommendation: every year) Contact us for more information.
Feature | Values |
---|---|
Measurement parameter | Temperature (°C/°F) |
Operating temperature | -35 to +55 °C |
Storage temperature | -35 to +55 °C |
Battery type | 3x battery type AAA or Energizer L92 AAA-size cells |
Life | 3 years (15 min. measuring cycle, +25 °C) |
Degree of protection | IP 65 |
Dimensions in mm (LxWxH) | 89 x 53 x 27 mm |
Weight | 130g |
Housing | ABS/PC |
Measuring cycle | 10s – 24h (freely selectable) |
Interface | Mini-USB, SD card slot |
Memory capacity | 1 million readings |
EU Directive | 2014/30/EU, complies with the EN standard12830 6F7 |
7 Please note that, according to EN 12830, this instrument must be regularly checked and calibrated as specified in EN 13486 (recommendation: every year) Contact us for more information.
testo 175 T3 (0572 1753)
Feature | Values |
---|---|
Measurement parameter | Temperature (°C/°F) |
Sensor type | 2 thermocouples (type K or T) external |
Measurement range | -50 to +400 °C (type T) -50 to +1000 °C (type K) |
Instrument accuracy | ±0.5 °C (-50 to +70 °C) ± 1 digit |
± 0.7% of the measurement value (+70.1 to +1000 °C)
± 1 digit
Resolution| 0.1 °C
Operating temperature| -20 to +55 °C
Storage temperature| -20 to +55 °C
Battery type| 3x battery type AAA or Energizer L92 MA-size cells
Life| 3 years (15 min. measuring cycle, +25 °C)
Degree of protection| IP 65
Dimensions in mm (LxWxH)| 89 x 53 x 27 mm
Weight| 130g
Housing| ABS/PC
Measuring cycle| 10s – 24h (freely selectable)
Interface| Mini-USB, SD card slot
Memory capacity| 1 million readings
EU Directive| 2014/30/EU
testo 175 H1 (0572 1754)
Feature | Values |
---|---|
Measurement parameter | Temperature (°C/°F), moisture (%rF /%RH/ °Ctd/ g/m3) |
Sensor type | NTC temperature sensor, capacitive humidity sensor |
Number of measuring channels | 2x internal (stubs) |
Measuring ranges | -20 to +55 °C -40 to +50 °Ctd |
0 to 100 %rF (not for condensing atmosphere)8
System accuracy 9| ±2%rF (2 to 98%rF) at 25 °C ±0.03 %rF/K ± 1 digit
±0.4 °C (-20 to +55 °C) ± 1 digit
Long-term drift of the sensor under normal conditions| <1 %R1-1/year (ambient
temperature +25 °C)
Usage conditions| All specifications call for an atmosphere with a percentage
of harmful gases that does not exceed the maximum allowable concentration
(MAC). A higher percentage of harmful gases (e.g. ammonia or hydrogen
peroxide) may result in damage to the sensor.
Resolution| 0.1 %rF, 0.1 °C
Operating temperature| -20 to +55 °C
Storage temperature| -20 to +55 °C
8 Long-term condensation in the system may result in damage to the measuring
instrument.
9 The use of sintered caps may affect the response time of the sensor.
Feature | Values |
---|---|
Measurement parameter | Temperature (°C/°F), moisture (%rF /%RH/ °Ctd/ g/m3) |
Battery type | 3x battery type AAA or Energizer L92 AAA-size cells |
Life | 3 years (15 min. measuring cycle, +25 °C) |
Degree of protection | IP 54 |
Dimensions in mm (LxWxH) | 149 x 53 x 27 mm |
Weight | 130g |
Housing | ABS/PC |
Measuring cycle | 10s – 24h (freely selectable) |
Interface | Mini-USB, SD card slot |
Memory capacity | 1 million readings |
EU Directive | 2014/30/EU |
Battery life
The programming windows of the software provide you with typical guide values
for the expected lifetime of the battery. This lifetime is calculated on the
basis of the following factors:
- Measuring cycle
- Number of connected sensors
Since the battery life depends on quite a few other factors, the calculated
data can only serve as guide values.
The following factors have a negative effect on the battery life:
- longer flashing of the LEDs
- frequent reading out (several times per day) via the SD-card
- extreme fluctuations in operating temperature
The following factors have a positive effect on the battery life:
- display switched off
The battery capacity reading in the display of the data logger is based on the calculated values. However, the data logger is switched off when a critical voltage level has been reached. It may therefore happen that:
- readings are still recorded, even though the battery capacity reading says “empty”.
- the measurement program is stopped, even though the battery capacity reading just before indicated a still remaining battery capacity.
In case of an empty battery or a battery change saved readings will not be lost.
First steps
4.1. Unlock the data logger
- Open the lock with the key (1).
- Remove the lock (2) from the locking pin.
- Pull the locking pin (3) out of the holes in the wall bracket.
- Slide the data logger out of the wall bracket (4).
4.2. Inserting batteries
In order to reach the battery life under application temperatures below -10
°C you should use Energizer L92 AAA-size cells.
-
Place the data logger on its front.
-
Loosen the screws on the back of the data logger.
-
Remove the battery compartment cover.
-
Insert the batteries (type AAA). Observe the polarity!
-
Place the battery compartment cover on the battery compartment.
-
Tighten the screws.
– The display shows rST.
4.3. Connecting the data logger to PC
For testo Comfort Software Basic 5:
The software is available in the Internet as a free download requiring
registration: www.testo.com/download-center.
The instructions for the installation and operation of the software can be
found in the testo Comfort Software Basic 5 instruction manual, which can be
downloaded together with the software.
For testo Comfort Software Professional und testo Comfort Software CFR:
-
Install the software testo Comfort Software.
-
Connect the USB cable to a free USB port on the PC.
-
Loosen the screw on the right side of the data logger.
-
Open the cover.
-
Display and control elements
-
Plug the USB cable into the Mini USB port (1).
-
Configure the data logger, see separate operating instructions testo Comfort Software.
Display and control elements
5.1. Display
The display function can be switched on/off via the software testo Comfort
Software.
Depending on the operating status, various information may be shown in the
display. A detailed representation of the information that can be called up
can be found under Menu overview.
For technical reasons the display speed of liquid crystal displays becomes
slower at temperatures below 0 °C (approx. 2 seconds at -10 °C, approx. 6
seconds at -20 °C). This has no influence on the measuring accuracy.
testo 175 T1
-
Highest saved reading
-
Lowest saved reading
-
Reading
-
Units
-
Measurement program over
-
Measurement program running
-
Wait for start of measurement program
-
Start criterion Date/ Time programmed
-
Battery capacity
Icon| Capacity
---|---
| >151 days
| <150 days
| <90 days
| <60 days
| <30 daysRead out data and change battery (see Reading out measurement data).
-
Lower alarm value
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were fallen short of -
Upper alarm value
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were exceeded
testo 175 T2, testo 175 T3, testo 175 H1
-
Reading channel 1
-
Units channel 1
-
Reading channel 2
-
Units channel 2
-
Measurement program over
-
Measurement program running
-
Wait for start of measurement program
-
Start criterion Date/ Time programmed
-
Battery capacity
Icon| Capacity
---|---
| >151 days
| <150 days
| <90 days
| <60 days
| <30 days
| <30 daysRead out data and change battery (see Reading out measurement data).
-
Lower limit value channel 2:
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were fallen short of -
Upper limit value channel 2:
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were exceeded -
Lowest saved reading
-
Highest saved reading
-
Lower limit value channel 1:
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were fallen short of -
Upper limit value channel 1:
• Flashes: programmed alarm value is shown
• Lights: programmed alarm values were exceeded
5.2. LED
Representation | Explanation |
---|---|
Red LED flashes once every 10 seconds | Remaining battery capacity has dropped |
below 30 days
Red LED flashes twice every 10 seconds| Remaining battery capacity has dropped
below 10 days
Red LED flashes three times every 10 seconds| Battery is empty:
Red LED flashes three times when pressing the button| Limiting value was
exceeded/fallen short of
Yellow LED flashes three times| Instrument changes from Wait-mode to Rec-mode.
Yellow LED flashes three times when pressing the button| Instrument is in Rec-
mode
Green and yellow LED flash three times when pressing the button.| Instrument
is in End-mode
Green LED flashes three times when pressing the button| Instrument is in Wait-
mode
Green LED flashes five times when pressing the button| Long pressing of the GO
button has set a time mark.
Green, yellow and red LED flash in succession| The battery has been changed.
5.3. Key functions
A detailed representation of the display readings can be found under Menu
overview.
✓ Instrument in operating status Wait and start criterion Button start
programmed.
Press [GO] for approx. 3 seconds to start the measurement program.
– The measurement program starts and Rec appears in the display.
✓ Instrument is in operating status Wait:
Press [GO] in order to change between displays of upper alarm value, lower alarm value, battery life and last reading.
The displays appear in the specified sequence.
✓ Instrument is in operating status Rec or End:
Press [GO] in order to change between displays of highest saved reading, lowest saved reading, upper alarm value, lower alarm value, battery life and last reading.
The displays appear in the specified sequence.
Using the product
6.1. Connecting a sensor
Observe the following points when connecting sensors to data logger and
measuring points.
Ensure correct polarity of the plugs.
Press the plugs firmly into the ports to ensure leak tightness. However, do not apply force!
Make sure that the plugs are firmly connected to the data logger or that the connections are closed with blanking plugs.
Ensure correct positioning of the sensor to avoid disturbing influences affecting the measurement.
testo 175 T3: Always make sure that you connect the sensor configured (via the software testo Comfort Software) to the individual sockets. The numbers of the connections are printed on the housing.
6.2. Programming data logger
In order to adapt the programming of your data logger to your individual
requirements, you require the testo Comfort Software Basic 5 software. It is
available in the Internet as a free download requiring registration
www.testo.com/download-center.
The instructions for the installation and operation of the software can be found in the testo Comfort Software Basic 5 instruction manual that is downloaded together with the software.
6.3. Menu overview
The menu overview shows exemplary display representations of the data
logger testo 175-T2. The display must be switched on to be able to show the
corresponding indications. This is
accomplished with the software testo Comfort Software.
The indication in the display is updated according to the programmed
measurement rate. Only readings from active channels are displayed.
The channels are also activated via the software testo Comfort Software.
The symbols for upper or lower alarm value light up in operating states Rec
and End, if the programmed alarm value has been exceeded or fallen short off.
After 10 seconds without operating a key the display will return to its
initial state.
Wait mode: Start criterion is programmed, but not yet fulfilled.
① Last reading 10 Start criterion key start / PC start Start criterion
Date/Time| ② Upper alarm value
---|---
|
③ Lower alarm value| ④ Battery capacity in days
|
Last reading 5 (see Fig. ① Wait mode)
10 Measurement value is not saved
Rec mode: Start criterion was fulfilled, data logger saves readings
① Last reading | ② Highest reading |
---|---|
③ Lowest reading | ④ Upper alarm value |
⑤ Lower alarm value | ⑥ Battery capacity in days |
Last reading (see Fig. ① Rec mode)
End mode: Measurement program finished (stop criterion reached – memory full or number of readings) depending on programming
① Last reading | ② Highest reading |
---|---|
③ Lowest reading | ④ Upper alarm value |
⑤ Lower alarm value | ⑥ Battery capacity in days |
Last reading (see Fig. ① End mode)
6.4. Mounting the wall bracket
The scope of delivery does not include mounting materials (e.g. screws,
wall plugs).
✓ The data logger has been removed from the wall bracket.
- Position the wall bracket at the desired place.
- Use a pen or something similar to mark the location for the fastening screws.
- Prepare the fastening location in accordance with the fastening material (e.g. drill hole, insert wall plugs).
- Fasten the wall bracket with suitable screws.
6.5. Securing the data logger
✓ The wall bracket has been mounted.
- Slide the data logger into the wall bracket (1).
- Push the locking pin (2) through the holes in the wall bracket.
- Fasten the lock (3) on the locking pin.
- Pull off the key (4).
6.6. Reading out measurement data
The measurement data remain stored in the data logger after they have been read out and can therefore be read out several times. The measurement data will only be deleted when the data logger is reprogrammed.
Via USB cable
-
Connect the USB cable to a free USB port on the PC.
-
Loosen the screw on the right side of the data logger.
Use a coin for this purpose. -
Open the cover.
-
Plug the USB cable into the Mini USB port (1).
-
Read out the data logger and process the read out data, see separate operating instructions testo Comfort Software.
Via SD card
-
Loosen the screw on the right side of the data logger.
Use a coin for this purpose. -
Open the cover.
-
Push the SD card into the SD card slot (2).
– The display shows Sd (testo 175 T1) or Sd CArd (testo 175 T2, testo 175 T3, testo 175 H1). -
Hold [Go] depressed for longer than 2 seconds.
– The display shows CPY (testo 175 T1) or COPY (testo 175 T2, testo 175 T3, testo 175 H1).
– The yellow LED lights during the copying process.
– The green LED flashes twice and after the copying process the display shows OUT. -
Remove the SD card.
-
Insert the SD card into the SD card slot on the PC.
-
Process the read out data, see separate operating instructions testo Comfort Software.
Maintaining the product
7.1. Changing batteries
The battery change stops the currently running measurement program.
However, stored measurement data are preserved.
-
Read out stored measurement data, see Reading out measurement data.
✓ If it is no longer possible to read out the saved measurement data because the battery capacity is too low:Change the batteries and read out the stored measurement data after.
-
Place the data logger on its front.
-
Loosen the screws on the back of the data logger.
-
Remove the battery compartment cover.
-
Take the empty batteries out of the battery compartment.
-
Insert three new batteries (type AAA). Observe the polarity!
Only use new branded batteries. If a partially exhausted battery is inserted, the calculation of the battery capacity will not be performed correctly.
In order to reach the battery life under application temperatures below -10 °C you should use Energizer L92 AAA-size cells. -
Place the battery compartment cover on the battery compartment.
-
Tighten the screws.
– The display shows rST.
The data logger needs to be reconfigured. For this purpose the software testo Comfort Software must be installed on the computer and a connection to the data logger must be set up. -
Connect the data logger to the PC with a USB cable.
-
Start the software testo Comfort Software and set up a connection to the data logger.
-
Reconfigure the data logger or load the old, saved configuration, see separate operating instructions testo Comfort Software.
– The data logger is once again ready for use.
7.2. Cleaning the instrument
CAUTION
Damage to the sensor!
Ensure that no liquid enters the inside of the housing.
If the housing of the instrument is dirty, clean it with a damp cloth.
Do not use any aggressive cleaning agents or solvents! Weak household cleaning agents or soap suds can be used.
Tips and assistance
8.1. Questions and answers
Question | Possible causes / solution |
---|
FULL appears in the display, the red LED flashes twice, out appears in the display.| Insufficient memory capacity on SD card to save the data.
Remove the SD card, free up more memory space and copy data.
Err appears in the display, the red LED flashes twice, out appears in the display.| An error occurred while saving data to the SD card.
Remove the SD card, free up more memory space and copy data.
nO dAtA appears in the display, the red LED flashes twice.| The logger has not yet recorded any data and is in Wait mode.
Remove the SD card and wait until the logger is in Rec mode.
rST appears in the display.| The battery was changed. No data are recorded.
Reprogram the data logger via the software.
– – – –
appears in the display.| Sensor of data logger defective.
Contact your dealer or the Testo Customer Service.
If you have any questions please contact your local dealer or the Testo Customer Service. You find contact data on the back of this document or in the Internet under www.testo.com/service-contact.
8.2. Accessories and spare parts
Description | Article no. |
---|---|
Wall bracket (black) with lock | 0554 1702 |
Mini USB cable to connect the data logger testo 175 to the PC | 0449 0047 |
SD card to read out the data logger 175 | 0554 8803 |
Batteries (alkaline-manganese AAA- size cells) for applications down to -10
°C| 0515 0009
Batteries (Energizer L92 AAA-size cells) for applications down to -10 °C| 0515
0042
CD testo Comfort Software Professional| 0554 1704
CD testo Comfort Software CFR| 0554 1705
ISO moisture calibration certificate, calibration points 11,3 %rF; 50,0 %rF;
75,3 %rF at +25°C/+77°F; per channel/instrument| 0520 0076
ISO temperature calibration certificate, calibration points -18°C, 0°C, +40°C;
per channel/instrument| 0520 0153
For further accessories and spare parts, please refer to the product catalogues and brochures or look up our website: www.testo.com
Testo SE & Co. KGaA
Testo-Straße 1 79853 Lenzkirch Germany
Tel.: +49 7653 681-0
Fax: +49 7653 681-7699
E-Mail: info@testo.de
www.testo.de
0970 1759
References
- Testo Downloads | Testo, Inc
- testo.com/service-contact
- Testo SE & Co. KGaA | Professionelle Messgeräte und Messlösungen vom Weltmarktführer
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>