My Weigh iBalance i401 Digital Precision Scale User Manual

June 12, 2024
My Weigh

IBALANCE i401
USER MANUAL

My Weigh iBalance i401 Digital Precision Scale -

Capacity
4009 x0.0059
141020000202

Thank you for purchasing the My Weigh® iBalance® 401′ digital scale. Please read all operating instructions carefully before use. This electronic scale is a precision instrument. With normal care and proper treatment, it will provide years of reliable service. For more information please visit www.myweigh.com
Never load the scale with more than the maximum capacity. Although the iBalance® 401′ is designed to be extremely durable with extra overload protection built into the case, overloading will permanently damage it! Avoid any exposure to extreme heat or cold, your scale works better when operated at normal room temperature. Keep your scale in a clean environment. Dust, dirt, moisture, vibration, air currents and/or a close proximity to other electronic equipment can all cause an adverse effect on the reliability and accuracy of your scale. Handle with care. Gently apply all items to be weighed onto tray top. Avoid shaking, dropping or otherwise shocking the scale. Scales are delicate instruments and unlike cellular phones, scales have delicate sensors that determine how much an item weighs. If you drop or shock your scale, these sensors leerthe shock and are sometimes destroyed. This happens with all digital scales. We design our scales to be as resistant to shock or drops as possible, however there is no way for us to protect 100% against load cell or sensor damage.
Failure to follow these instructions will void your warranty.
Always allow the scale to acclimate to normal room temperature for at least one hour before use. Give your scale sufficient warm up time. Usually 30-60 seconds before calibration to give the internal components a chance to stabilize.

PRECAUTIONS BEFORE USING THE BALANCE

  1. Matter charged with static electricity can affect accuracy. Discharge all static electricity. For example, one method is to use Static-Guard spray, and spray it on both sides of the weighing platform.
  2. Before the iniialuse, please remove the protection screw. R-install thisscrew before transport to help avoid possible damage. The balance must be in an exactly horizontal position in order to achieve accurate measurement results. In order to bring the balance nto a horizonta position, the adjustable fet are tumed either lockwise or counter-clockwise until the air bubble on the front panel isin the center of the marked dirce.
  3. Please use an independent power outlt to avoid nterference from other electrical applances.
  4. Don’t put any object on the platform before powering on.
  5. When possible please allow the sale to warm up forseveral minutes before operation.
  6. Items should always be placed on the center f the platform when being weighed.
  7. For optimum accuracy, recalibrate before each use.

SCALE FEATURES

My Weigh iBalance i401 Digital Precision Scale - fig1

PROTECTION SCREW
on the left side of the scale

My Weigh iBalance i401 Digital Precision Scale - fig2

ADJUSTABLE FEET
on bottom of each comer of the scale

My Weigh iBalance i401 Digital Precision Scale - fig3

DATA TRANSMISSION PORT
onright side of the scale

My Weigh iBalance i401 Digital Precision Scale - fig4

KEYPAD FUNCTIONS

| ON/OFE pover switch.
---|---
| SAMPLNG & COUNTING,
| UNIT selecton
| TARE is used to deduct the weight of an item or container.The symbol will appear and reading will go to zero,Press it again to exit the tare mode (when empty), the tare indication willdisappear.
|  ZEROisused toretu thedispay o zeroit asmallweightreading s leftwhile
DISPLAY SYMBOLS
| Scaleis in ZERO mode.
| SaleisinTAREmode.
|  BATIERY needs echarging.
|  Thedisplayreadingis STABLE.
• In charge|  Scaleisin the proces ofRECHARGE
• pcs| Scaleisin COUNT made.

POWER SUPPLY

Thei401™ s powered by DC6V rechargeable battery or directly by AC power adapter.

‘OPERATION INSTRUCTIONS

Weighing Procedures

  1. Press to turn on the scale.
    When the power is turned on, the scale will countdown for a few seconds and “0” will appear on the display.

  2. Select the weighing unit with key.
    Press to select a weighing unit g, ct, oz , ort, lb, dr, gn,dwt, MM, tl.1, tl.T, tl.H, t, 1/8,1/4 . Once the unit has been selected, the selected unit will be displayed next to the weight value.

  3. Start weighing Verify the reading is “0”. Place objects on the weighing platform to weigh. When the reading becomes stable, the stable indicator is displayed.

Tare
Tare can be used for eliminating the weight value of an empty container. Place an empty container on the scale and press . Then place the items to be weighed in the container.
NOTE: When all weight is removed from the weighing tray, the tared value of a container will be displayed as a negative number. Press again to return the scale to zero.

ERROR MESSAGES
When an applied load exceeds the capacity, “—–” will appear on the display. Remove excessive load immediately. The unit may return to normal operation. Remember: You can permanently damage the scale and void your warranty by overloading it! The LCD will display”ERR “if the weight placed on the platform is incorrect during calibration.
CALIBRATION
When to calibrate – calibration is RARELY required. Calibration may be required when the scale is first set up for use, or if the scale is moved to a different altitude or new location. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with time and use, mechanical deviations can occur.
How to calibrate: you must have an accurate 300g weight or combination of weights in order to calibrate

  1. Press and hold (do not release) and power the scale on with , release when the LCD shows”CAL”.

  2. Press again, it shows”000000″. This is where you input the calibration weight you are going to use to calibrate (300 grams is recommended).
    Use to cycle through the 6 zeros shown on the screen and use on the selected digit to adjust selected digit. To calibrate using the recommended 300 grams, the screen must read “000300”. Once this is set, place the 300 gram calibration weight on the tray.

  3. Wait for 3 seconds, then press   to finish the calibration process.

COUNTING FUNCTION

  1. Press , the display il show” 10 Cr pes” (means samplesize s 10 pcs)
  2. Press to toggle sample size, 10, ” “20. ” “50, ” “100″pes will appearin. Sop atthe one you want o use.
  3. Putthe exactquantity of samples desired o the platform and press , the set sample size il ppear.
  4. Keep adding objects o be counted on the pan, thetotal number of the abjects will b displayed. I the unit weight s too small for the counting resolution, the display willshowErrPcs”

WEIGHT RESPONSE SPEED AND DIVISION

The 401 allows you to adjust the scale reaction time and division selction.

  1. Press and hold (do not release) and power the scale on with . Wait until the display shows”nb0 – nb7″ and release
  2. Press again to select your response speed. (nb0: fastest, nb7 slowest) Press   to confirm.
  3. You will now enter division selection. The display will show e.g “d 0.01′: You can press to select, and press to confirm. The scale will return to normal weighing mode.

SELECTING DESIRED WEIGHING MODES

There are 15 weighing modes which can be easily enabled or disabled for ease of use, to do this:

  1.  Press and hold (do not release) and press and release the ON/OFF key. Release the key a the end of the self test
  2. The display will show “Sln”, press to and choose “Sln” and press to confirm(Be sure to select the”Sln” section.).
  3. Then the display will show”On x”or”OFF x”(“x”is unit);”On” means enable,”OFF” means disable .There are fifteen units preset, press to toggle through the weighing modes and choose the desired unit. Choose “On” or”OFF” by pressing and press to confirm.
  4. Then the scale will ask you to choose the default initial weighing mode. The display will show “init X” (“X” is unit), press gi to -0- toggle and choose initial unit; press to confirm.

RANGE OF ZERO TRACKING AND ZERO DISPLAY SELECTION

Zerotracking enables high precision scales o compensate during wind fluctuations and vibaions. t s ossile o adjust the level ofassistance offered by zerotracking.Todo thisuse the following seps:

  1. Press and hold   (do not release) and power the scale on with . Wait until the display shows”0.Od, 0.5d. 1.0d. 1.5d. 2.0d 3.0d”. Then press a to select the range of zero tracking and press o- to confirm.
  2. The display will then show “ZEr-S”or”ZEr-L” press to select the zero display range. (ZEr-S means Od and ZEN!. means ± 3.0d) Press to confirm.
  3. To select the baud rate press   , you can choose between 1200, 2400,4800 and 9600. Press to confirm.
  4. You can choose the communication method by pressing   to toggle (Co: send in succession, st: send steadily) , press to confirm. After that the scale will return to normal weighing mode.

DATA TRANSMISSION – SERIES RS-232 INTERFACE (only for communication)

  1. iBalance 02325 -RS232 C’s UART signal
  2.  Format
(1)Baudrate: 1200 bps 2400 bps 4800bps 9600 bps
(2) Databits: 8bits
(3)Paritybit:  none
(4) 5Stophit: 1 bit
(5)Code ASCl

DATA FORMAT :

HEADT HEAD2 DATA UNIT
CR
123 456 7891011121314 1516171819 2021
HEADT (2BYTES)
0L-overload
ST-stable
HEAD? (2BYTES)
NT – net weight mode
US-unstable
DATA(8BYTE)
20 (HEX) =”-“(negative sign)
20 (HEX) =” *(blank)
2E (HEX) =”.”(decmal point)
UNIT (4 byte)

g=20(HEX; 20(HEX;
kg=20(HEX); 20(HEX;
A=20(HX; 20(HX;
ot=20(HE0; 6F(HEX;
R=0D (HEX); 0A (HEX)
20(HEX; 67 (HEX)
67 (HEX) 6B (HEX)
63(HEN); 74 (HEX)
TA(HEN; 74 (HEX)

TRANSMISSION EXAMPLE
stablenet +0.168

HEAD, HEAD, DATA UNIT CR
ST, NT +0.168 g OA, OD

DISPLAY UNITS

  • g GRAM
  • GN GRAIN (UK)
  • tl.H TAEL (HONG KONG)
  • ct CARAT
  • dwt PENNYWEIGHT
  • t TOLA (INDIA)
  • oz OUNCE
  • MM MOMME (1PN)
  • 1/8 1/8 OUNCE
  • ort TROY OUNCE
  • tIJ JEWELRY TAEL (HONG KONG)
  • 1/4 1/4 OUNCE
  • dr DRAM
  • tI.T TAEL (TWN)

SPECIFICATIONS

Capacity      400g x   0.005g| Units g, ct, oz , ort, lb, dr, gn,dwt,
MM, t1.1, tl.T, tl.H, t, 1/8,1/4
Scale dimension| 270mm x 200mm x 70mm
Tray dimension| 116mm diameter
Scale Weight| 920g
Operating temperature| Optimum 10-40°C (50-104°f)
Power Source| DC 6V rechargeable battery / AC power adaptor
Tare range| Up to scale’s maximum capacity

My Weigh iBalance i401 Digital Precision Scale -
icon12

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals