PHILIPS UltraFill Home Oxygen System User Guide
- June 12, 2024
- Philips
Table of Contents
PHILIPS UltraFill Home Oxygen System
Startup checklist
Before filling UltraFill oxygen cylinders, check to make sure that the system is setup correctly. Follow the Startup checklist below to ensure proper filling.
| | Action| How to verify
---|---|---|---
1| | Tubing is properly connected between UltraFill and the oxygen
concentrator. Check for kinks.| Inspect visually.
2| | Oxygen concentrator is powered on.| Green light on top panel should be
lit. If not, check that the power cord is plugged in and the power button is
depressed. If the problem persists, contact your homecare provider.
3| | Oxygen concentration level is properly set. EverFlo should be at or below
3 LPM. Millennium M10 should be at or below 7 LPM.| Check the concentrator’s
flow meter.
4| | UltraFill is powered on.| Green light on top panel should be lit. If not,
check that the power cord is plugged in and the power button is depressed. If
the problem persists, contact your homecare provider.
5| | Conserver or regulator is in the “Off” position.| Inspect visually
6| | If the conserving
device or regulator is not connected during filling, check that the cylinder
valve is closed.| Using the cylinder valve wrench provided, turn the valve
head clockwise until tight.
7| | Cylinder is properly
connected to the UltraFill device.| One or more lights on the fill-level
indicator is solid. No lights should be flashing
8| | Fill Start/Stop
button is depressed and filling has started.| If cylinder is not full, one
light on the fill-level indicator will flash.
9| | Filling complete.
Release cylinder from UltraFill.| All lights on the fill-level indicator are
solidly lit.
Safe transport
Reasonable care should be taken when moving the UltraFill device across an uneven surface. Follow the general guidelines below when attempting to move UltraFill across the floor, through doorways, and into an elevator.
Correct transport method over uneven surface
The proper method for moving the UltraFill device is to first remove the cylinder, then either lift the unit using its two handles or pull the device along its long orientation using one handle.
Incorrect transport method over uneven surface
Never attempt to move the UltraFill device with the cylinder attached or by pushing the top of the unit along its horizontal axis.
Customers Support
Philips Healthcare is part of Royal Philips Electronics
How to reach us
www.philips.com/healthcare
healthcare@philips.com
Asia
+49 7031 463 2254
Europe, Middle East, Africa
+49 7031 463 2254
Latin America
+55 11 2125 0744
North America
+1 425 487 7000
800 285 5585
(toll free, US only)
Philips Respironics
1010 Murry Ridge Lane
Murrysville, PA 15668
Customer Service
+1 724 387 4000
800 345 6443
(toll free, US only)
Philips Respironics
International Headquarters
+33 1 47 28 30 82
Scan with your cell phone to learn more
Respironics, EverFlo, Millennium, and UltraFill are trademarks of Respironics, Inc. and its affiliates. All rights reserved.
Please visit http://ultrafill.respironics.com
© 2010 Koninklijke Philips Electronics N.V.
All rights are reserved.
Philips Healthcare reserves the right to make changes in specifications and/or
to discontinue
any product at any time without notice or obligation and will not be liable
for any consequences
resulting from the use of this publication.
CAUTION: US federal law restricts these devices to sale by or on the order of
a physician.
Geyer WMB 12/20/10 MCI 4103767 PN 1080388
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>