stryker Prime Series Stretcher Instruction Manual
- June 12, 2024
- stryker
Table of Contents
Prime Series Stretcher
Product Information
The product is a Stryker medical device, model number 1125. It
is designed to assist with patient care and can be used for various
medical procedures. The product has been manufactured according to
European medical device standards and bears the CE mark. The
manufacturer is Stryker, and the date of manufacture is indicated
in the manual. The product has a safe working load capacity, which
is specified in the manual. It operates on both direct and
alternating current.
Symbols
- Refer to instruction manual/booklet
- Consult instructions for use
- General warning
- Caution
- Warning; crushing of hands
- Warning; non-ionizing radiation
- No pushing
- Do not store the oxygen bottle
- Catalogue number
- Serial number
- Lot (batch) code
- For US Patents see www.stryker.com/patents
- CE mark
- Authorized representative in the European Community
- Manufacturer
- Date of manufacture
- Importer
- Safe working load
- Maximum patient weight
- Mass of equipment with safe working load
- Direct current
- Alternating current
- Dangerous voltage
- Potential equalization
- Protective earth ground
- Protection from water jets
- Type B applied part
- MR Unsafe
- This way up
- Keep dry
- Fragile, handle with care
- Stacking limit by number
Warning/Caution/Note Definition
The words WARNING, CAUTION, and NOTE have special meanings and
should be carefully reviewed:
WARNING
Alerts the reader about a situation which, if not avoided, could
result in death or serious injury. It may also describe potential
serious adverse reactions and safety hazards.
CAUTION
Alerts the reader of a potentially hazardous situation which, if
not avoided, may result in minor or moderate injury to the user or
patient or damage to the product or other property. This includes
special care necessary for the safe and effective use of the device
and the care necessary to avoid damage to a device that may occur
as a result of use or misuse.
Note
Provides special information to make maintenance easier or
important instructions clearer.
Summary of Safety Precautions
Always read and strictly follow the warnings and cautions listed
in the manual. Service should only be performed by qualified
personnel.
Product Usage Instructions
Introduction
This manual provides assistance with the operation and
maintenance of the Stryker product. Before operating or maintaining
the product, it is important to read this manual thoroughly.
Additionally, methods and procedures should be set in place to
educate and train staff on the safe operation and maintenance of
the product.
Pinch Points
The manual contains information about pinch points that should
be taken into consideration during the operation and maintenance of
the product. Care should be taken to avoid injury or damage.
Setup
Refer to the corresponding section in the manual for detailed
instructions on setting up the product for use. Follow the provided
instructions carefully to ensure proper installation and
configuration.
Operation
Refer to the corresponding section in the manual for detailed
instructions on operating the product. Follow the provided
instructions carefully to ensure safe and effective use of the
product.
Option
If the product has additional optional features or functions,
refer to the “Option” section in the manual for instructions on how
to utilize those features. Follow the provided instructions
carefully to maximize the benefits of the product.
Scale System Option
If the product includes a scale system option, refer to the
“Scale System Option” section in the manual for instructions on how
to use this feature. Follow the provided instructions carefully to
ensure accurate measurements.
Please note that these usage instructions are a summary. For
complete and detailed instructions, refer to the specific sections
of the manual mentioned above.
Prime Series® Stretcher
Operations Manual 1125
with Zoom® Motorized Drive
1125-309-005 Rev AB.0
2021-12
CS
HU
RO
DA
IT
RU
DE
KO
SK
EL
LT
SL
ES
LV
SV
ET
NL
TH
FI
PL
TR
EN
FR
PT
ZH
Symbols
Refer to instruction manual/booklet Consult instructions for use General warning Caution Warning; crushing of hands Warning; non-ionizing radiation
No pushing
Do not store the oxygen bottle Catalogue number Serial number Lot (batch) code For US Patents see www.stryker.com/patents CE mark Authorized representative in the European Community European medical device
Manufacturer
Date of manufacture
Importer
Safe working load
1125-309-005 Rev AB.0
EN
~
IPX5
87VL
MR
Maximum patient weight
Mass of equipment with safe working load Direct current
Alternating current
Dangerous voltage
Potential equalization
Protective earth ground Protection from water jets Type B applied part
Medical Equipment Classified by Underwriters Laboratories Inc. With Respect to
Electric Shock, Fire, and Mechanical Hazards Only in Accordance with ANSI/AAMI
ES60601-1: 2005 and CAN/ CSA-C22.2 No. 60601-1:08. *For Prime X option,
Associated Equipment of X-ray Equipment IEC 60601-2-54 In accordance with
European Directive 2012/19/EU on Waste Electrical and Electronic Equipment
(WEEE) as amended, this symbol indicates that the product should be collected
separately for recycling. Do not dispose of as unsorted municipal waste.
Contact local distributor for disposal information. Ensure infected equipment
is decontaminated prior to recycling.
MR Unsafe
This way up
Keep dry
Fragile, handle with care Stacking limit by number
EN
1125-309-005 Rev AB.0
Table of Contents
Warning/Caution/Note Definition ……………………………………………………………………………………………………..3
Summary of safety precautions …………………………………………………………………………………………………3 Pinch
points ……………………………………………………………………………………………………………………6
Introduction ………………………………………………………………………………………………………………………………..7 Product
description ………………………………………………………………………………………………………………..7 Indications for use
………………………………………………………………………………………………………………….7 Clinical
benefits……………………………………………………………………………………………………………………..8 Contraindications
…………………………………………………………………………………………………………………..8 Expected service
life……………………………………………………………………………………………………………….8
Disposal/recycle…………………………………………………………………………………………………………………….8 Specifications
……………………………………………………………………………………………………………………….8 Product illustration – Prime
……………………………………………………………………………………………………..11 Product illustration – Prime X
option ………………………………………………………………………………………….12 Applied parts – electric litter
option ……………………………………………………………………………………………13 Applied parts – Prime X option
…………………………………………………………………………………………………13 Contact information
………………………………………………………………………………………………………………13 Serial number location
…………………………………………………………………………………………………………..14
Setup ………………………………………………………………………………………………………………………………………15 Operation
…………………………………………………………………………………………………………………………………16
Applying and releasing the brakes ……………………………………………………………………………………………16 Prime
base controls………………………………………………………………………………………………………………17 Raising the non-
electric litter …………………………………………………………………………………………………..17 Lowering the non-
electric litter…………………………………………………………………………………………………17 Raising the electric
litter option ………………………………………………………………………………………………..18 Lowering the electric
litter option ……………………………………………………………………………………………..18 Positioning the product
in Trendelenburg …………………………………………………………………………………..18 Positioning the product
in Reverse Trendelenburg ……………………………………………………………………….19 Transporting a patient
without Big Wheel …………………………………………………………………………………..19 Transporting a patient
with Big Wheel ……………………………………………………………………………………….20 Transporting a patient
with Zoom Motorized Drive ……………………………………………………………………….20 Charging Zoom
Motorized Drive battery …………………………………………………………………………………….23 Transporting a
patient manually without Zoom Motorized Drive wheel ………………………………………………24
Raising the siderail ……………………………………………………………………………………………………………….24 Lowering
the siderail……………………………………………………………………………………………………………..25 Positioning the
product with the siderail patient control panel – electric litter option-
Prime only ……………….25 Operator control panel – electric litter option – Prime
only ……………………………………………………………….26 Locking and unlocking the patient control
panel lockout – electric litter option – Prime only……………………..27 Raising or
lowering the Fowler backrest – non-electric litter …………………………………………………………….27
Raising or lowering the Fowler backrest – electric litter
option………………………………………………………….28 Raising or lowering the Gatch – non-electric
litter – Prime only …………………………………………………………29 Raising or lowering the Gatch –
electric litter option – Prime only………………………………………………………29 Positioning the
recovery chair – Prime only …………………………………………………………………………………29 Storing objects
in the base hood………………………………………………………………………………………………30 Hanging devices with
the pump rack option ………………………………………………………………………………..30 Extending or
retracting the power cord with the retractable cord reel option – electric
lift option or electric litter
option ……………………………………………………………………………………………………………………31 Scale system option
……………………………………………………………………………………………………………..31
Scale icon/button identification ………………………………………………………………………………………….32
Weighing a patient………………………………………………………………………………………………………….32 Locking the
scale unit of measure ………………………………………………………………………………………33 Arming or disarming
the Chaperone option …………………………………………………………………………..33 Setting the Chaperone
option alert pattern……………………………………………………………………………33 Setting the Chaperone
option alert volume …………………………………………………………………………..34 Charging the scale system
battery pack option – electric litter option or electric litter option with
Chaperone option – Prime only ………………………………………………………………………………34 Replacing the
batteries for the scale system option – non-electric litter
………………………………………..34 Accessories and parts
…………………………………………………………………………………………………………..35 Attaching the defibrillator tray
……………………………………………………………………………………………37 Converting the defibrillator tray/foot
extender to a defibrillator tray ……………………………………………..38 Converting the
defibrillator tray/foot extender to a foot extender ………………………………………………..38
1125-309-005 Rev AB.0
1 EN
Attaching the footboard/chart holder……………………………………………………………………………………39 Attaching the IV caddy …………………………………………………………………………………………………….39 Positioning or stowing the foot supports – Prime only ………………………………………………………………40 Positioning the two-stage permanently attached IV pole ………………………………………………………….41 Positioning the three-stage permanently attached IV pole ………………………………………………………..42 Attaching and positioning the removable IV pole ……………………………………………………………………43 Attaching the upright oxygen bottle holder ……………………………………………………………………………44 Extending or stowing the serving tray holder/footboard ……………………………………………………………44 Attaching the siderail pads ……………………………………………………………………………………………….45 Locating the patient restraint strap tie-ins …………………………………………………………………………….45 Positioning the upright X-ray cassette holder – Prime X option …………………………………………………..45 Positioning the lateral X-ray cassette holder – Prime X option ……………………………………………………46 Inserting or removing X-ray cassettes – Prime X option ……………………………………………………………47 Cleaning ………………………………………………………………………………………………………………………………….49 Cleaning the product …………………………………………………………………………………………………………….49 Cleaning the mattress……………………………………………………………………………………………………………49 Remove iodine ………………………………………………………………………………………………………………50 Special instructions…………………………………………………………………………………………………………50 Disinfecting ………………………………………………………………………………………………………………………………51 Disinfecting the product …………………………………………………………………………………………………………51 Disinfecting the mattress………………………………………………………………………………………………………..51 Preventive maintenance ………………………………………………………………………………………………………………53 EMC information ………………………………………………………………………………………………………………………..55
EN 2
1125-309-005 Rev AB.0
Warning/Caution/Note Definition
The words WARNING, CAUTION, and NOTE carry special meanings and should be
carefully reviewed.
WARNING
Alerts the reader about a situation which, if not avoided, could result in
death or serious injury. It may also describe potential serious adverse
reactions and safety hazards.
CAUTION
Alerts the reader of a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or damage to the
product or other property. This includes special care necessary for the safe
and effective use of the device and the care necessary to avoid damage to a
device that may occur as a result of use or misuse.
Note – Provides special information to make maintenance easier or important
instructions clearer.
Summary of safety precautions
Always read and strictly follow the warnings and cautions listed on this page.
Service only by qualified personnel.
WARNING · Always allow the product to reach room temperature before you set up
the product or test functional operations.
Permanent product damage may occur. · Always operate the product when all
operators are clear of the mechanisms. · Always use care when you handle the
power cord. Entanglement, damage to the power cord, or potential shock hazards
may occur. If the power cord is damaged, immediately remove the product from
service and contact the appropriate maintenance personnel. · Do not use the
Prime Series stretcher with Zoom Motorized Drive in the MR (Magnetic Resonance
Imaging) environment. The Prime Series stretcher with Zoom Motorized Drive is
MR Unsafe. · Always plug the product directly into a grounded, hospital grade
wall outlet. You can only achieve grounding reliability when you use a
hospital grade outlet. This product is equipped with a hospital grade plug to
protect against electric shock hazard. · Always apply the brakes when a
patient is getting on the product or off the product or when the product is
not moving. Injury could result if the product moves while a patient is
getting on the product or off the product. · Always position the patient in
the center of the product. · Always put the product in the lowest position
with the siderails up and latched when you leave a patient unattended. Do not
leave the product at a higher height. · Always remove any devices that may be
in the way before you raise or lower the litter. · Do not sit on the end of
the product. The product may tip. · Always lock the siderails in the full up
position with the sleep surface horizontal and in the lowest position when you
transport a patient. · Do not use the Prime Series stretcher with Zoom
Motorized Drive in the MR environment. The Prime Series stretcher with Zoom
Motorized Drive is MR (Magnetic Resonance Imaging) Unsafe. · Always unplug the
power cord from the wall outlet before you transport a patient or clean the
product. · Do not attempt to push the product manually with the drive wheel
engaged and the On/Drive-Off/Manual switch in the On position. The product
will be difficult to push and injury may occur. · Always use extra caution
when you use a mattress thicker than 2.5 in. or a mattress overlay with Prime
X option. · Always lock the siderails in the full up position with the sleep
surface flat and in the lowest position when you transport a patient. · Always
keep the patient’s limbs away from the siderail spindles when you raise or
lower the siderail. · Do not allow the siderails to lower on their own. ·
Always keep hands and fingers clear of the Fowler backrest release handles and
the Fowler backrest frame when you lower the Fowler backrest.
1125-309-005 Rev AB.0
3 EN
· Always use caution when you raise a the Fowler backrest while a patient is
on the product. Use proper lifting techniques and get help, if necessary.
· Always keep hands and fingers clear of the foot end push handles when you
lower the Gatch.
· Always keep device lines on the pump rack away from the Gatch handles.
· Always make sure that devices on the pump rack can pass safely through door
openings.
· Do not lift the product by the pump rack.
· Always unplug the power cord from the wall outlet before you transport or
clean the product.
· Always use the retractable cord reel to store the power cord inside the base
when you transport the product.
· Do not use the scale system option or electric lift option or electric
litter option adjacent to or stacked with other devices. If adjacent or
stacked use is necessary, the scale system option or electric lift option or
electric litter option should be observed to make sure that all devices are
operating normally in the configuration in which they will be used.
· Do not place items that weigh more than 30 lb (14 kg) on the defibrillator
tray. Always strap down all devices that you place on the defibrillator tray.
· Always use caution if the defibrillator tray/foot extender, footboard/chart
holder, or upright oxygen bottle holder is attached to avoid pinching your
fingers when you position the foot end push handle option.
· Do not place items that weigh more than 30 lb (14 kg) on the defibrillator
tray/foot extender. Always strap down all devices that you place on the
defibrillator tray.
· Always secure the IV pole to the IV caddy when you transport the product.
· Always store the IV caddy when not in use to avoid product damage.
· Do not sit on the foot supports. This may cause the product to tip.
· Always clear your fingers from the mechanisms when you operate the foot
supports.
· Always stow the foot supports when you transport a patient with the product.
· Always tighten the foot supports before you use the foot supports.
· Do not operate the scale system with the foot supports. Inaccurate readings
may occur.
· Do not operate the Chaperone option with the foot supports. Inaccurate
readings may occur.
· Do not use the IV pole as a push/pull device. Product damage may occur.
· Do not place objects that exceed 40 lb (18 kg) in the upright oxygen bottle
holder.
· Do not place objects that exceed 30 lb (14 kg) on the serving tray.
· Always use caution when you attach restraint straps. Patient or operator
injury may occur. Physical restraints, even if secured, may result in serious
harm to patients and operators, including entanglement, entrapment, physical
injury, or death.
· Always attach restraint straps or devices only at the identified attachment
points of the product. Failure to do so may result in patient or operator
injury. Do not attach restraint straps to the siderail.
· Always refer to the applicable state and federal restrictions and
regulations and the appropriate facility protocols before you use any
restraint strap or device.
· Always take protective measures when you use the upright or lateral X-ray
cassette holder. The X-ray cassette holder does not protect against radiation.
· Always refer to local, state, and federal guidelines in addition to facility
protocols for safety before you use Prime X option with radiation generating
devices. Radiation generating devices may produce residual, stray, or
scattered radiation.
· Always follow the Positioning the upright X-ray cassette holder – Prime X
option instructions to insert the X-ray cassette.
· Always use caution when you take X-rays with the Fowler backrest in the
upright position or when you use a lateral cassette.
· Always follow the Positioning the lateral X-ray cassette holder – Prime X
option instructions to insert the X-ray cassette.
· Always unplug the power cord and turn the On/Drive – Off/Manual switch to
the Off position before service or cleaning.
· Do not steam clean, pressure wash, hose off, or ultrasonically clean the
product.
· Do not clean, service, or perform maintenance while the product is in use.
· Do not immerse the mattress in cleaning or disinfectant solutions. Excess
moisture could cause product malfunction that results in product damage or
patient injury.
· Do not allow fluid to pool on the mattress. Fluids can cause corrosion of
components and may cause the safety and performance of this product to become
unpredictable.
· Always inspect mattress covers for tears, punctures, excessive wear, and
misaligned zippers every time you clean the covers. Remove and replace a
damaged mattress to prevent cross-contamination.
EN 4
1125-309-005 Rev AB.0
· Do not steam clean, pressure wash, hose off, or ultrasonically clean
mattresses. These methods of cleaning may void the product’s warranty.
· Do not steam clean, hose off, or ultrasonically clean the product. Use of
these methods of cleaning is not recommended and may void the product’s
warranty.
· Always disinfect the mattress between patients. Failure to do so could
result in cross-contamination and infection.
CAUTION
· Improper usage of the product can cause injury to the patient or operator.
Operate the product only as described in this manual.
· Do not modify the product or any components of the product. Modifying the
product can cause unpredictable operation resulting in injury to patient or
operator. Modifying the product also voids its warranty.
· Always disconnect the plug from the wall outlet to isolate the Prime Series
stretcher with Zoom Motorized Drive from the supply mains.
· Always position the Prime Series stretcher with Zoom Motorized Drive to
allow for access to the wall outlet.
· Do not use the hydraulics on the base to raise the product with a patient
lift under the product.
· Do not apply the steer pedal when Big Wheel is resting on a threshold or
other raised area. The force required to apply Big Wheel will be higher than
normal which may cause product damage.
· Always immediately unplug the power cord from the wall outlet and turn the
On/Drive-Off/Manual switch to the Off position if large fluid spills occur in
the area of the circuit boards or motors. Fluids can short out controls and
may cause the product to operate erratically or may make some functions
inoperable. Component failure caused by fluids may also cause the product to
operate unpredictably and could cause injury to the patient. Do not put the
product back into service until it is dry and has been tested for safe
operation.
· Remove the batteries if the equipment is placed in storage or will remain
idle for an extended period of time. Each battery weighs 25 lb. Use caution
when removing the batteries to avoid injury.
· Battery posts, terminals, and related accessories contain lead and lead
compounds, chemicals known to the State of California to cause cancer and
birth defects or other reproductive harm. Wash your hands after handling.
Properly dispose of batteries when required.
· The weight capacity of the Gatch is 200 lb (90.7 kg).
· Always make sure that the Gatch prop rod is secure before use.
· Do not sit or stand on the Gatch.
· Always use caution when you attempt to lower the Gatch while the product is
unpowered. Gravity may cause a rapid drop of the Gatch.
· Do not place objects that exceed 60 lb (27 kg) in the base hood.
· Do not sit, step, or stand on the base hood.
· The weight capacity of the pump rack is 40 lb (18 kg).
· Do not use the pump rack as a push/pull device. Product damage may occur.
· Do not use the scale for patients under 50 lb.
· Do not use the scale system reading as a reference for medical treatment.
· Always raise the IV pole before you attach the defibrillator tray/foot
extender to the product. If you do not raise the IV pole, the foot extender
will not operate.
· Do not use the IV pole as a push/pull device. Product damage may occur.
· Do not hang IV bags that exceed 40 lb (18 kg) on the IV pole.
· Always make sure that the IV pole is at a low height to pass through door
openings when you transport a patient.
· Do not use abrasive cleaners to clean the display enclosure for the scale
system option. Do not allow cleaning solutions or other fluids to pool on the
display unit. Wipe dry all surfaces after spills or cleaning.
· The use of accessories, transducers, and cables, other than those specified
or provided by the manufacturer, could result in increased electromagnetic
emissions or decreased electromagnetic immunity and result in improper
operation.
· The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment, for which CISPR 11 class B is normally required, this
equipment might not offer adequate protection to radio frequency communication
services. The user might need to take mitigation measures, such as relocating
or reorienting the equipment.
· Portable RF communications equipment, including peripherals such as antenna
cables and external antennas, should be used no closer than 12 inches (30 cm)
to any part of the Prime stretcher, including cables specified by the
manufacturer.
1125-309-005 Rev AB.0
5 EN
· Avoid stacking or placing other equipment adjacent with other equipment to
prevent improper operation of the products. If such use is necessary,
carefully observe stacked or adjacent equipment to make sure that they are
operating properly.
Pinch points
Figure 1 Prime X option
EN 6
1125-309-005 Rev AB.0
Introduction
This manual assists you with the operation or maintenance of your Stryker
product. Read this manual before operating or maintaining this product. Set
methods and procedures to educate and train your staff on the safe operation
or maintenance of this product.
CAUTION · Improper usage of the product can cause injury to the patient or
operator. Operate the product only as described in this
manual. · Do not modify the product or any components of the product.
Modifying the product can cause unpredictable operation
resulting in injury to patient or operator. Modifying the product also voids
its warranty.
Note · This manual is a permanent part of the product and should remain with
the product even if the product is sold. · Stryker continually seeks
advancements in product design and quality. This manual contains the most
current product
information available at the time of printing. There may be minor
discrepancies between your product and this manual. If you have any questions,
contact Stryker Customer Service or Technical Support at
1-800-327-0770.
Product description
The Stryker Model 1125 Prime Series® stretcher with Zoom® Motorized Drive is a
powered wheeled stretcher that consists of a platform mounted on a wheeled
frame that is designed to support patients in a horizontal position. The
stretcher provides a method of transporting patients within the interior of a
healthcare facility by health professionals or trained representatives of the
facility. The Prime Series stretcher with Zoom Motorized Drive features an
electric drive system which aids the health professional or trained
representative by assisting stretcher movement and maneuverability in various
healthcare facilities. The device can be manually pushed by the user in the
event of power loss to Zoom Motorized Drive. The device has siderails and has
the option available to support the temporary or permanent placement of IV
poles, along with various other options and accessories to assist with the
transport of the patient.
Indications for use
The Stryker Model 1125 Prime Series stretcher with Zoom Motorized Drive is an
electromechanical stretcher that provides a healthcare professional or trained
representative greater maneuverability in steering and moving the stretcher
with significantly less force.
The Prime Series stretcher may be used as a short-term outpatient clinical
evaluation, treatment, minor procedure, and short-term outpatient recovery
platform. The stretcher may include use in, but is not limited to:
· Emergency Department (ED)
· Trauma area
· Post-Anesthesia Care Unit (PACU)
The Prime Series stretcher may be used for minor procedures and short-term
stay (treatment and recovery). See the specifications table for the intended
environmental conditions.
The Stryker Prime Series stretcher has not been evaluated for compliance to
bed standard BS EN 50637. This product is not intended for use for short term
stay with pediatric patients or adult patients with atypical anatomy in
markets that recognize this bed standard for market authorization.
The Prime Series stretcher has a safe working load up to 700 lb (318 kg) and
is intended to be used with all patients, including those mildly to critically
ill. The stretcher may also be used to transport deceased patients within an
enclosed healthcare facility.
The Prime X® option provides an articulating radiographic patient support
surface and a platform below the patient support surface for X-ray cassette
placement. The Prime X option is intended to allow the capture of clinical
X-rays (AP full body, optional full body lateral, and optional upright chest)
when used with a medical X-ray system.
1125-309-005 Rev AB.0
7 EN
Clinical benefits
Patient transport, facilitate treatment, and diagnostic
Contraindications
The Prime Series stretcher with Zoom Motorized Drive is not intended for home
healthcare or domestic use or those directly connected to the public low
voltage power supply network that supplies buildings used for domestic
purposes.
The Prime Series stretcher with Zoom Motorized Drive with Prime X option is
not intended for use with a mattress with a thickness greater than four inches
and is not compatible for use with a C-Arm.
Expected service life
The Prime Series stretcher and the Prime Series stretcher with Prime X option
have a 10 year expected service life under normal use conditions and with
appropriate periodic maintenance.
Disposal/recycle
Always follow the current local recommendations and/or regulations governing
environmental protection and the risks associated with recycling or disposing
of the equipment at the end of its useful life.
Specifications
Prime Series 26 in. width
Prime Series 30 in. width
Prime X option 30 in. width
Note – Safe working load indicates the sum of the patient, mattress, and accessory weight
700 lb
318 kg
700 lb
318 kg
700 lb
318 kg
Maximum patient weight Weight of product standard configuration Weight of
product with all options/ accessories
Overall length
Overall width (siderails up)
Overall width (siderails down)
675 lb
306 kg
675 lb
306 kg
675 lb
306 kg
415 lb
188 kg
435 lb
197 kg
435 lb
197 kg
465 lb
211 kg
485 lb
220 kg
485 lb
220 kg
86 in. (± .5 in.)
34 in. (± 1 in.)
30.25 in. (± .5 in.)
218.4 cm (± 86 in. (± .5 1.27 cm) in.)
86.4 cm
38 in. (± 1 in.)
76.8 cm
30.5 in. (± .5 in.)
218.4 cm (± 86 in. (± .5 218.4 cm (±
1.27 cm)
in.)
1.27 cm)
96.5 cm
38 in. (± 1 in.)
96.5 cm
77.5 cm
30.5 in. (± .5 in.)
77.5 cm
EN 8
1125-309-005 Rev AB.0
Prime Series 26 in. width
Prime Series 30 in. width
Prime X option 30 in. width
Minimum/maximum height
23 in. / 34 in. (± 1 in.)
58 cm / 86 cm (±2.54 cm)
23 in. / 34 in. (± 1 in.)
58 cm / 86 cm (±2.54 cm)
23 in. / 34 in. (± 1 in.)
58 cm / 86 cm (±2.54 cm)
Fowler backrest angle
0° to 90° (± 5°)
Gatch height
5.5 in. (14 cm) minimum
Not applicable
Trendelenburg/Reverse Trendelenburg
+16°/-16°
2.5 in. nominal
6.4 cm
2.5 in. nominal
6.4 cm
2.5 in. nominal
6.4 cm
Minimum clearance
1.75 in. under the hydraulic jacks
4.5 cm
1.75 in. under the hydraulic jacks
4.5 cm
1.75 in. under the hydraulic jacks
4.5 cm
0.75 in. under Zoom 1.9 cm wheels
0.75 in. under Zoom wheels
1.9 cm
0.75 in. under Zoom 1.9 cm wheels
Electrical requirements
100 VAC, 50 Hz, 4A 120 VAC, 60 Hz, 4A 240 VAC, 50/60 Hz, 4A
Battery type
2 x 12 VDC 31Ah battery (2 x 12 VDC) lead acid gel cell battery
Battery voltage
24 VDC
Attenuation equivalent (aluminum equivalence)
Not applicable
Maximum value allowed is 1.7 mm Al
Scale system option weight operating range
50 lb (22.7 kg) to 700 lb (318 kg)
Scale system option accuracy
±3 lb (1.3 kg) for weights less than 100 lb (45 kg) and ±3% for weights greater than or equal to 100 lb (45 kg)*
*To meet the scale system option accuracy claim, the patient surface must be in the flat position (Fowler backrest and Gatch down) and the product cannot exceed 5 degrees of Trendelenburg/Reverse Trendelenburg.
Electric options Electrical requirements Duty cycle
Optional electric litter (available on Prime only)
120 VAC, 60 Hz, 4A 240 VAC, 50 Hz, 4A 240 VAC, 50/60 Hz, 4A
Continuous operation with intermittent loading is 1 min. ON/20 min. OFF
Optional scale system Battery type Battery voltage
Non-electric litter (available on Prime and Prime X option)
4 x AA battery (4 x 1.5 VDC) alkaline type (LR6)
6.0 VDC
1125-309-005 Rev AB.0
9 EN
Optional scale system Battery type Battery voltage
Optional electric litter 1 x rechargeable lithium ion battery pack (0058-135-000) 10.8 VDC, 2.4Ah
Optional scale system with Chaperone® (stretcher exit)
Battery type
Battery voltage
Optional electric litter 1 x rechargeable lithium ion battery pack (0058-134-000) 10.8 VDC, 4.8Ah
Stryker reserves the right to change specifications without notice.
Note – This product is not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
Environmental conditions
Temperature
Operation
50 °F (10 °C)
100 °F (38 °C)
Charging
50 °F (10 °C)
82 °F (28 °C)
Storage and transportation
-4 °F (-20 °C)
140 °F (60 °C)
Relative humidity
75% 30%
75% 30%
95% 10%
Atmospheric pressure
700 hPa
1060 hPa
700 hPa
1060 hPa
500 hPa
1060 hPa
Note – The scale system option (non-electric litter/electric litter option)
operation temperature range is 61 °F (16 °C) to 79 ° F (26 °C).
Specifications listed are approximate and may vary slightly from product to
product or by power supply fluctuations.
In accordance with the European REACH regulation and other environmental
regulatory requirements, the components that contain declarable substances are
listed.
Description Gel cell battery assembly Battery assembly
Number 1040-020-007 1040-010-870
Substance of very high concern (SVHC) chemical name
Lead
Lead
EN 10
1125-309-005 Rev AB.0
Product illustration – Prime
Q A
B
C
P
NO
L
M
K
J D
E
FG
I H
Figure 2 Product illustration – Prime
A
Zoom handle
B
BackSmart® siderail
C
Base hood storage
D
Pump pedal
E
Uni-lower pedal
F
Big Wheel
G
Bumper, integrated
H
Caster with integrated wheel cover
I
Brake/steer pedal
J
Hydraulic release pedal
K
Pump pedal
L
Siderail latch
M
Gatch release handle
N
BackSmart foot end operator control panel, optional
O
BackSmart scale system option
P
Patient control panel lockout, optional
Q
Siderail patient control panel, optional
1125-309-005 Rev AB.0
11 EN
Product illustration – Prime X option A
B
C
D
L
K J
E F
G I
H
Figure 3 Product illustration – Prime X option
A
Zoom handle
B
Patient surface with Clearview Technology® deck
C
Cassette tray
D
BackSmart siderail
E
Pump pedal
F
Uni-lower pedal
G
Bumper, integrated
H
Caster with integrated wheel cover
I
Brake/steer pedal
J
Siderail latch
K
Pump rack, optional
L
BackSmart scale system option
EN 12
1125-309-005 Rev AB.0
Applied parts – electric litter option
Figure 4 Type B applied parts – electric litter option
Applied parts – Prime X option
Contact information
Figure 5 Type B applied parts – Prime X option
Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
1125-309-005 Rev AB.0
13 EN
Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 USA Note – The user
and/or the patient should report any serious product-related incident to both
the manufacturer and the Competent authority of the European Member State
where the user and/or patient is established. To view your operations or
maintenance manual online, see https://techweb.stryker.com/. Have the serial
number (A) of your Stryker product available when calling Stryker Customer
Service or Technical Support. Include the serial number in all written
communication.
Serial number location
A
Figure 6 Serial number location
EN 14
1125-309-005 Rev AB.0
Setup
To unpack your product, see the unpacking instructions that are attached to
the product inside of the shipping crate.
WARNING · Always allow the product to reach room temperature before you set up
the product or test functional operations.
Permanent product damage may occur. · Always operate the product when all
operators are clear of the mechanisms. · Always use care when you handle the
power cord. Entanglement, damage to the power cord, or potential shock hazards
may occur. If the power cord is damaged, immediately remove the product from
service and contact the appropriate maintenance personnel. · Do not use the
Prime Series stretcher with Zoom Motorized Drive in the MR (Magnetic Resonance
Imaging) environment. The Prime Series stretcher with Zoom Motorized Drive is
MR Unsafe. · Always plug the product directly into a grounded, hospital grade
wall outlet. You can only achieve grounding reliability when you use a
hospital grade outlet. This product is equipped with a hospital grade plug to
protect against electric shock hazard.
CAUTION · Always disconnect the plug from the wall outlet to isolate the Prime
Series stretcher with Zoom Motorized Drive from
the supply mains. · Always position the Prime Series stretcher with Zoom
Motorized Drive to allow for access to the wall outlet.
Make sure that the product is working before you put the product into service.
1. Apply the brake. Push on the product to make sure that all four casters
are locked. 2. Release the brake. Push on the product to make sure that all
four casters are unlocked. 3. Raise and lower the litter with the hydraulic
lift system. 4. Raise the product to the highest position and put the product
in the Trendelenburg position. Make sure that the head end
lowers to the full down position. 5. Raise the product to the highest position
and put the product in the Reverse Trendelenburg position. Make sure that the
foot end lowers to the full down position. 6. Apply Big Wheel and make sure
that the product raises and you can guide the product with the front casters
and Big
Wheel. 7. Apply Zoom Motorized Drive. Make sure that you can push the product
forward and backward with the push handles. 8. Make sure that the siderails
raise, lower, and lock in place. 9. Raise and lower the manual Fowler backrest
(head end). 10. Raise and lower the manual Gatch (foot end).
For product with the electric litter option:
1. Complete the setup checklist above. 2. Plug the product into a grounded,
hospital grade wall outlet. Make sure that the LED lights illuminate on the
lockout
keypad. 3. Make sure that each function on the siderail patient control panel
option works. 4. Make sure that each function on the foot end operator control
panel option works. 5. Raise and lower the electric Fowler backrest (head
end). 6. Raise and lower the electric Gatch (foot end).
For product with the scale system option – electric litter option (with or
without Chaperone option):
1. Complete the setup checklist above. 2. Charge the batteries.
1125-309-005 Rev AB.0
15 EN
Operation
Applying and releasing the brakes
WARNING – Always apply the brakes when a patient is getting on the product or
off the product or when the product is not moving. Injury could result if the
product moves while a patient is getting on the product or off the product. To
apply the brakes, push down on the brake (red) side of the brake/steer pedal.
Push on the product to make sure that the brakes work. To release the brakes,
push down on the steer (green) side of the brake/steer pedal.
A
A
Figure 7 Operating the brake/steer pedal Note – Do not push down on the
center of the brake/steer pedal. Always push down on the outer side (A) of the
brake/steer pedal (Figure 7).
EN 16
1125-309-005 Rev AB.0
Prime base controls
CA
B
B
D D
A CA
A C
A BD
Figure 8 Prime base controls and label
Raising the non-electric litter
WARNING · Always position the patient in the center of the product. · Always
put the product in the lowest position with the siderails up and latched when
you leave a patient unattended. Do
not leave the product at a higher height. · Always remove any devices that may
be in the way before you raise or lower the litter. · Do not sit on the end of
the product. The product may tip.
CAUTION – Do not use the hydraulics on the base to raise the product with a
patient lift under the product.
To raise the litter, press down on the pump pedal (A) until you achieve the
desired height (Prime base controls (page 17)).
Lowering the non-electric litter
WARNING · Always position the patient in the center of the product. · Always
put the product in the lowest position with the siderails up and latched when
you leave a patient unattended. Do
not leave the product at a higher height. · Always remove any devices that may
be in the way before you raise or lower the litter. · Do not sit on the end of
the product. The product may tip.
1125-309-005 Rev AB.0
17 EN
CAUTION – Do not use the hydraulics on the base to raise the product with a
patient lift under the product.
To lower the entire litter, press down on the center of the uni-lower pedal
(C) (Prime base controls (page 17)).
To lower the head end of the litter, press down on the pedal (B) or on side of
the uni-lower pedal (C) closest to the head end of the product.
To lower the foot end of the litter, press down on the pedal (D) or press down
on the side of the uni-lower pedal (C) closest to the foot end of the product.
Raising the electric litter option
WARNING · Always position the patient in the center of the product. · Always
put the product in the lowest position with the siderails up and latched when
you leave a patient unattended. Do
not leave the product at a higher height. · Always remove any devices that may
be in the way before you raise or lower the litter. · Do not sit on the end of
the product. The product may tip.
CAUTION – Do not use the hydraulics on the base to raise the product with a
patient lift under the product.
To raise the electric litter, press down on the pump pedal (A) until you
achieve the desired height (Prime base controls (page 17)).
Lowering the electric litter option
WARNING · Always position the patient in the center of the product. · Always
put the product in the lowest position with the siderails up and latched when
you leave a patient unattended. Do
not leave the product at a higher height. · Always remove any devices that may
be in the way before you raise or lower the litter. · Do not sit on the end of
the product. The product may tip.
CAUTION – Do not use the hydraulics on the base to raise the product with a
patient lift under the product.
To lower the entire electric litter, press on the center of the uni-lower
pedal (C) (Prime base controls (page 17)).
To lower the head end of the litter, press down on the pedal (B) or push down
on the side of the uni-lower pedal (C) closest to the head end of the product.
To lower the foot end of the litter, press down on the pedal (D) or push down
on the side of the uni-lower pedal (C) closest to the foot end of the product.
Positioning the product in Trendelenburg
WARNING – Always remove any devices that may be in the way before you raise or
lower the litter.
CAUTION – Do not use the hydraulics on the base to raise the product with a
patient lift under the product. To position the product in the Trendelenburg
position (head down), raise the litter to the highest height (Raising the
nonelectric litter (page 17) or Raising the electric litter option (page 18)).
Note – Raise the litter to the highest height for a greater Trendelenburg
angle.
EN 18
1125-309-005 Rev AB.0
To lower the head end of the product, push down on the head end or foot end
release pedal (D) or push down on the side of the uni-lower pedal (C) closest
to the head end (Prime base controls (page 17)). To lower the product from
Trendelenburg position, push down on the head end and foot end release pedals
(B and D) at the same time or push down on the center of the uni-lower pedal
(C) until the litter is flat.
Positioning the product in Reverse Trendelenburg
WARNING – Always remove any devices that may be in the way before you raise or
lower the litter.
CAUTION – Do not use the hydraulics on the base to raise the product with a
patient lift under the product.
To position the product in the Reverse Trendelenburg position (foot down),
raise the litter to the highest height (Raising the non-electric litter (page
17) or Raising the electric litter option (page 18)). Note – Raise the litter
to the highest height for a greater Trendelenburg angle. To lower the foot end
of the product, push down on the head end or foot end release pedal (B) or
push down on the side of the uni-lower pedal (C) closest to the foot end
(Prime base controls (page 17)). To lower the product from Reverse
Trendelenburg position, push down on the head end and foot end release pedals
(B and D) at the same time or push down on the center of the uni-lower pedal
(C) until the litter is flat.
Transporting a patient without Big Wheel
WARNING · Always position the patient in the center of the product. · Always
lock the siderails in the full up position with the sleep surface horizontal
and in the lowest position when you
transport a patient.
To transport a patient without Big Wheel: 1. Put the brake/steer pedal (A) in
the neutral position (Figure 9).
A
C
B
C
Figure 9 Transporting without Big Wheel
2. Lower the litter to the lowest position. 3. For Model 1125 only, make sure
that the On/Drive-Off/Manual switch is in the Off position. 4. Unplug the
power cord from the hospital grade wall outlet if your product is equipped
with the electric lift option or
electric litter option. 5. Use the push handles to push the product. 6. Apply
the brakes to lock the product in place.
1125-309-005 Rev AB.0
19 EN
Note · Always make sure that the brake is released before you move the product
to avoid operator or patient injury. · With the product in the neutral
position, Big Wheel (B) is elevated and the product rests on all four casters
(C) (Figure
9). In this position, you can move the product in any direction, including
sideways.
Transporting a patient with Big Wheel
WARNING · Always position the patient in the center of the product. · Always
lock the siderails in the full up position with the sleep surface horizontal
and in the lowest position when you
transport a patient.
CAUTION – Do not apply the steer pedal when Big Wheel is resting on a
threshold or other raised area. The force required to apply Big Wheel will be
higher than normal which may cause product damage.
To transport a patient with Big Wheel: 1. Raise the siderails to the highest
position. Pull on each siderail to make sure it is latched. 2. Lower the
litter to the lowest position. 3. For Model 1125 only, make sure that the On
/Drive-Off/Manual switch is in the Off position. 4. Unplug the power cord from
the hospital grade wall outlet if your product is equipped with the electric
lift option or
electric litter option. 5. Put the brake/steer pedal (A) in the steer position
(Figure 10).
A
B
C
Figure 10 Transporting with Big Wheel
6. Use the push handles to push the product. 7. Apply the brakes to lock the
product in place.
Note · Always make sure that the brake is released before you attempt to move
the product. Attempting to move the product
with the brakes applied may result in operator or patient injury. · The Big
Wheel (B) does not pivot. You cannot move the product sideways when you apply
Big Wheel. · When you apply Big Wheel (B), the product is raised and the foot
end casters (C) are elevated. The product rests on the
head end casters and Big Wheel makes the product easier to steer (Figure 10).
Transporting a patient with Zoom Motorized Drive
WARNING · Always position the patient in the center of the product.
EN 20
1125-309-005 Rev AB.0
· Always lock the siderails in the full up position with the sleep surface
horizontal and in the lowest position when you transport a patient.
· Do not use the Prime Series stretcher with Zoom Motorized Drive in the MR
environment. The Prime Series stretcher with Zoom Motorized Drive is MR
(Magnetic Resonance Imaging) Unsafe.
· Always operate the product when all operators are clear of the mechanisms.
CAUTION – Always immediately unplug the power cord from the wall outlet and
turn the On/Drive-Off/Manual switch to the Off position if large fluid spills
occur in the area of the circuit boards or motors. Fluids can short out
controls and may cause the product to operate erratically or may make some
functions inoperable. Component failure caused by fluids may also cause the
product to operate unpredictably and could cause injury to the patient. Do not
put the product back into service until it is dry and has been tested for safe
operation. To transport a patient with Zoom Motorized Drive: 1. Unplug the
power cord from the wall outlet.
Note – The drive function will not operate if the power cord is plugged into
the wall outlet. 2. Put the brake/steer pedal (A) in the steer position
(Figure 11).
A B
Figure 11 On/Drive-Off/Manual switch 3. Lower the litter to the lowest
position. 4. Turn the On/Drive-Off/Manual switch (B) to the ON position. 5.
Check the two LEDs on the drive handle to see if the product is ready for use.
If the green LED is on, the product is
ready for use. If the amber LED is on, the product is not ready for use. Make
sure that the pedal is in the steer position, the power cord is not plugged
in, and the power switch is in the ON position. 6. If the drive handles are in
the stowed position, pull the handles up to the raised position and then push
down on the handles to lock them in the use position (Figure 12).
1125-309-005 Rev AB.0
21 EN
A
Figure 12 Zoom handles in the raised position 7. Grasp the drive handles at
the two raised grip areas. Squeeze one or both of the motion release switches
(A) under the
handles to move the product (Figure 13). Note – You can use either motion
release handle to start movement, but you must release both motion release
handles to stop movement.
Figure 13 Zoom operation 8. Push the handles away from you or pull the
handles toward you to move the product in the desired direction. As you
push or pull, the product will adjust the speed in proportion to the amount of
force that you apply to the drive handles. When the desired speed is reached,
relax the force to maintain the speed. 9. Push or pull the handles again to
increase or decrease the speed of the product. 10. Release both handles to
stop the product. Note – The drive wheel does not pivot. You cannot move the
product directly sideways when the drive wheel is applied. With the drive
wheel in the neutral position and the brakes released, you can move the
product in any direction, including sideways. 11. Apply the brakes to lock the
product in place. 12. Stow the drive handles by lifting up on the handles.
Push the handles into the stowed position (Figure 14).
EN 22
1125-309-005 Rev AB.0
B
Figure 14 Zoom handles in the stowed position Note · Use caution when you
move the product with the drive wheel applied. Make sure that there are no
obstacles near the
product while the drive wheel is applied. Injury to patient, operator, or
bystander or damage to the product or surrounding equipment could occur if the
product collides with an obstacle. · Always make sure that the brake is
released before you attempt to move the product. Attempting to move the
product with the brakes applied may result in operator or patient injury. · If
unanticipated motion occurs, turn the On/Drive-Off/Manual switch to the Off
position. Unplug the power cord from the wall outlet.
Charging Zoom Motorized Drive battery
CAUTION · Remove the batteries if the equipment is placed in storage or will
remain idle for an extended period of time. Each
battery weighs 25 lb. Use caution when removing the batteries to avoid injury.
· Battery posts, terminals, and related accessories contain lead and lead
compounds, chemicals known to the State of
California to cause cancer and birth defects or other reproductive harm. Wash
your hands after handling. Properly dispose of batteries when required.
Two 12 volt batteries are required to power the drive wheel. The drive wheel
will not operate properly if the batteries are not sufficiently charged. When
fully discharged, the batteries require approximately 8 hours of charging time
to recharge (Figure 15). Plug the power cord into a grounded, hospital grade
wall outlet whenever possible to maintain the battery charge levels. The
battery power gauge is located at the head end of the litter (Figure 15). The
six LEDs illuminate individually to indicate the level of battery power that
is available to the product. As the batteries are charging, the LEDs will
flash in succession until all are flashing to indicate that the batteries are
fully charged.
1125-309-005 Rev AB.0
23 EN
Figure 15 Battery power gauge
Transporting a patient manually without Zoom Motorized Drive wheel
WARNING · Always unplug the power cord from the wall outlet before you
transport a patient or clean the product. · Do not attempt to push the product
manually with the drive wheel engaged and the On/Drive-Off/Manual switch in
the On
position. The product will be difficult to push and injury may occur. · Always
position the patient in the center of the product. · Always lock the siderails
in the full up position with the sleep surface horizontal and in the lowest
position when you
transport a patient.
CAUTION – Do not apply the steer pedal when Big Wheel is resting on a
threshold or other raised area. The force required to apply Big Wheel will be
higher than normal which may cause product damage.
To transport a patient manually without Zoom Motorized Drive wheel: 1. Raise
the siderails to the highest position. Pull on each siderail to make sure it
is latched. 2. Unplug the power cord from the wall outlet. 3. Turn the On
/Drive-Off/Manual switch to the Off position. 4. Put the brake/steer pedal in
the steer position.
Note – You can now move the product with Big Wheel for easier manual
transportation. 5. Use the push handles to push the product. 6. Apply the
brakes to lock the product in place.
Note · Always make sure that the brake is released before you move the product
to avoid operator or patient injury. · When you apply Big Wheel, the product
is raised and the foot end casters are elevated. The product rests on the
head end casters and Big Wheel to make the product easier to steer.
This manual transport option allows you to move the product with the
assistance of Big Wheel without the power assistance of Zoom Motorized Drive
wheel.
Raising the siderail
WARNING · Always use extra caution when you use a mattress thicker than 2.5
in. or a mattress overlay with Prime X option. · Always put the product in the
lowest position with the siderails up and latched when you leave a patient
unattended. Do
not leave the product at a higher height. · Always lock the siderails in the
full up position with the sleep surface flat and in the lowest position when
you transport a
patient. · Always keep the patient’s limbs away from the siderail spindles
when you raise or lower the siderail. · Do not allow the siderails to lower on
their own.
To raise the siderail: 1. Use two hands to grasp the siderail. 2. Lift the
siderail toward the foot end of the product until the yellow release latch
clicks into place. Pull on the siderail to
make sure that the siderail is latched.
EN 24
1125-309-005 Rev AB.0
Note · There is a dual siderail release latch option available with latches on
both ends of the product. · Do not use siderails as restraint devices to keep
the patient from exiting the product. The siderails keep the patient from
rolling off the product. The operator must determine how much restraint is
needed to make sure that the patient is safe. · You can use the foot end of
the siderail top rail as a push/pull device. · Siderails only lock in the full
up position.
Lowering the siderail
WARNING · Always use extra caution when you use a mattress thicker than 2.5
in. or a mattress overlay with Prime X option. · Always put the product in the
lowest position with the siderails up and latched when you leave a patient
unattended. Do
not leave the product at a higher height. · Always lock the siderails in the
full up position with the sleep surface flat and in the lowest position when
you transport a
patient. · Always keep the patient’s limbs away from the siderail spindles
when you raise or lower the siderail. · Do not allow the siderails to lower on
their own.
To lower the siderail: 1. Use one hand to grasp the siderail. 2. Use the other
hand to pull up on the yellow release latch. 3. Lift and guide the siderail
toward the head end of the product until the yellow release latch clicks into
place. Pull on the
siderail to make sure that the siderail is latched.
Note · There is a dual siderail release latch option available with latches on
both ends of the product. · Do not use siderails as restraint devices to keep
the patient from exiting the product. The siderails keep the patient from
rolling off the product. The operator must determine how much restraint is
needed to make sure that the patient is safe. · You can use the foot end of
the siderail top rail as a push/pull device. · Siderails only lock in the full
up position.
Positioning the product with the siderail patient control panel – electric
litter option- Prime only
WARNING – Always operate the product when all operators are clear of the
mechanisms.
Use the siderail patient control panel to position the Fowler backrest and
Gatch (Figure 16).
Healthcare professionals must instruct patients how to operate the patient
controls.
Note · Always plug the power cord into a grounded, hospital grade wall outlet.
· The siderail patient control panel is positioned in a staggered location on
each side of the product for easy patient
access. · Each siderail has backlit controls to allow the patient to position
the Fowler backrest and Gatch. When the product is
plugged in and the patient control panel is unlocked, the white buttons are
illuminated.
1125-309-005 Rev AB.0
25 EN
AB
CD DC
BA
A
Fowler down
B
Fowler up
C
Gatch up
D
Gatch down
Figure 16 Siderail patient control panel
Lowers the Fowler (head section) Raises the Fowler (head section) Raises the
Gatch (foot section) Lowers the Gatch (foot section)
Operator control panel – electric litter option – Prime only
WARNING – Always operate the product when all operators are clear of the
mechanisms. Use the operator control panel to adjust the position of the
patient on the product (Figure 17). Note – Always plug the power cord into a
grounded, hospital grade wall outlet.
AB CD
A
Fowler up
B
Fowler down
C
Gatch up
D
Gatch down
Figure 17 Operator control panel – electric litter option
Raises the Fowler backrest Lowers the Fowler backrest Raises the Gatch Lowers
the Gatch
EN 26
1125-309-005 Rev AB.0
Locking and unlocking the patient control panel lockout – electric litter
option Prime only
WARNING – Always operate the product when all operators are clear of the
mechanisms.
To lock the siderail patient control panel, press Unlock/Lock (A) once (Figure
18). Note – The lock icon (C) illuminates amber when the patient controls are
locked. To unlock the siderail patient control panel, press Unlock/Lock (A) a
second time. Note – The unlock icon (B) illuminates green when the patient
controls are unlocked. Note · Always plug the power cord into a grounded,
hospital grade wall outlet. · You can lock the patient control panel to
prevent the patient from moving the backrest and the Gatch. · The siderail
patient control panel is not backlit when you lock the patient control panel.
A BC
Figure 18 Patient control panel lockout
Raising or lowering the Fowler backrest – non-electric litter
WARNING · Always operate the product when all operators are clear of the
mechanisms. · Always keep hands and fingers clear of the Fowler backrest
release handles and the Fowler backrest frame when you
lower the Fowler backrest. · Always use caution when you raise a the Fowler
backrest while a patient is on the product. Use proper lifting techniques
and get help, if necessary.
To raise the Fowler backrest, squeeze one or both of the Fowler backrest
release handles (A) and pull the Fowler backrest up to the desired position
(0° to 90°) (Figure 19).
Note – The Lift Assist® backrest uses patient weight to assist in patient
positioning. The Lift Assist backrest also helps to keep the patient from
sliding toward the foot end of the product when you raise the Fowler backrest.
To lower the Fowler backrest, squeeze one or both of the Fowler backrest
release handles (A) and push the Fowler backrest down to the desired position
(90° to 0°) (Figure 19).
1125-309-005 Rev AB.0
27 EN
A
A
Figure 19 Raising or lowering the Fowler backrest (Prime / Prime X)
Raising or lowering the Fowler backrest – electric litter option
WARNING · Always operate the product when all operators are clear of the
mechanisms. · Always keep hands and fingers clear of the Fowler backrest
release handles and the Fowler backrest frame when you
lower the Fowler backrest. · Always use caution when you raise a the Fowler
backrest while a patient is on the product. Use proper lifting techniques
and get help, if necessary.
To raise the Fowler backrest, press the up button on the Fowler backrest side
of the siderail patient control panel (Positioning the product with the
siderail patient control panel – electric litter option- Prime only (page 25))
or foot end operator control panel (Operator control panel – electric litter
option – Prime only (page 26)) until the Fowler backrest has reached the
desired angle (0° to 70°).
Note – The Lift Assist backrest uses patient weight to assist in patient
positioning. The Lift Assist backrest also helps to keep the patient from
sliding toward the foot end of the product when you raise the Fowler backrest.
To lower the Fowler backrest, press the down button on the Fowler backrest
side of the siderail patient control panel (Positioning the product with the
siderail patient control panel – electric litter option- Prime only (page 25))
or foot end operator control panel (Operator control panel – electric litter
option – Prime only (page 26)) until the Fowler backrest has reached the
desired angle (70° to 0°).
Note – Always plug the power cord into a grounded, hospital grade wall outlet.
Healthcare professionals must instruct patients how to operate the patient
controls.
EN 28
1125-309-005 Rev AB.0
Raising or lowering the Gatch – non-electric litter – Prime only
WARNING – Always keep hands and fingers clear of the foot end push handles
when you lower the Gatch.
CAUTION · The weight capacity of the Gatch is 200 lb (90.7 kg). · Always make
sure that the Gatch prop rod is secure before use. · Do not sit or stand on
the Gatch.
To raise the non-electric Gatch, pump the Gatch pump handle to the left until
you achieve the desired position (see Product illustration – Prime litter).
Lift up on the end of the Gatch and secure the prop rod in the bracket. To
release the prop rod, lift up on the end of the Gatch, swing the prop rod
toward the head end of the product, and lower the foot end. To lower the
Gatch, pull the Gatch release handle until you achieve the desired position
(see Product illustration – Prime litter). Note – You cannot raise the Gatch
manually if the product has the electric litter option.
Raising or lowering the Gatch – electric litter option – Prime only
WARNING · Always keep hands and fingers clear of the foot end push handles
when you lower the Gatch. · Always operate the product when all operators are
clear of the mechanisms.
CAUTION · The weight capacity of the Gatch is 200 lb (90.7 kg). · Do not sit
or stand on the Gatch. · Always use caution when you attempt to lower the
Gatch while the product is unpowered. Gravity may cause a rapid
drop of the Gatch.
To raise the Gatch, press up on the Gatch side of the siderail patient control
panel or the foot end operator control panel. To lower the Gatch, press down
on the Gatch side of the siderail patient control panel or on the foot end
operator control panel. Note – Always plug the power cord into a grounded,
hospital grade wall outlet. Healthcare professionals must instruct patients
how to operate the patient controls.
Positioning the recovery chair – Prime only
WARNING – Always operate the product when all operators are clear of the
mechanisms.
1125-309-005 Rev AB.0
29 EN
To place the product in the recovery chair position (Figure 20):
Figure 20 Recovery chair (Model 1115 shown)
1. Raise the Fowler backrest to a seated position (Raising or lowering the
Fowler backrest – non-electric litter (page 27) or Raising or lowering the
Fowler backrest – electric litter option (page 28)).
Note – To position the recovery chair, the product must have the Lift Assist
backrest option and the Gatch option. 2. Raise the Gatch to the full up
position (Raising or lowering the Gatch – non-electric litter – Prime only
(page 29) or
Raising or lowering the Gatch – electric litter option – Prime only (page
29)). 3. Raise the litter to the highest height (Raising the non-electric
litter (page 17) or Raising the electric litter option (page
18)). 4. Place the product into the Reverse Trendelenburg position
(Positioning the product in Reverse Trendelenburg (page
19)). 5. Reverse steps to lower the product to the flat position.
Storing objects in the base hood
CAUTION · Do not place objects that exceed 60 lb (27 kg) in the base hood. ·
Do not sit, step, or stand on the base hood.
You can store patient belongings in the base hood. Do not use the oxygen
bottle holder cutout to store oxygen bottles or patient belongings.
Hanging devices with the pump rack option
WARNING · Always keep device lines on the pump rack away from the Gatch
handles. · Always remove any devices that may be in the way before you raise
or lower the litter. · Always make sure that devices on the pump rack can pass
safely through door openings. · Do not lift the product by the pump rack.
EN 30
1125-309-005 Rev AB.0
CAUTION · The weight capacity of the pump rack is 40 lb (18 kg). · Do not use
the pump rack as a push/pull device. Product damage may occur.
You can store or hang additional devices on the pump rack that is located on
the foot end of the product. Note – You must select the pump rack option at
the time of purchase.
Extending or retracting the power cord with the retractable cord reel option
electric lift option or electric litter option
WARNING · Always unplug the power cord from the wall outlet before you
transport or clean the product. · Always use the retractable cord reel to
store the power cord inside the base when you transport the product.
To extend the power cord, pull the power cord out from the retractable cord
reel (A) to the desired length (Figure 21).
A
Figure 21 Retractable cord reel
To retract the power cord: 1. Unplug the power cord from the wall outlet. 2.
Tug lightly on the power cord. 3. Guide the power cord into the retractable
cord reel.
Scale system option
WARNING – Do not use the scale system option or electric lift option or
electric litter option adjacent to or stacked with other devices. If adjacent
or stacked use is necessary, the scale system option or electric lift option
or electric litter option should be observed to make sure that all devices are
operating normally in the configuration in which they will be used.
CAUTION – Do not use the scale for patients under 50 lb.
The scale system option is available for products with or without the electric
litter option. The scale system option has a battery backup option. The
standby icon indicates when the product is using the battery backup.
1125-309-005 Rev AB.0
31 EN
The Chaperone option allows you to set zone controls to alert an operator when
a patient may be attempting to exit the product. The Chaperone option is
available for products with the electric litter option and the scale system
option. The scale system option with Chaperone option has a battery backup
option. The standby icon indicates when the product is using the battery
backup.
Raise the foot end push handles before you attempt to operate the scale system
option.
Use the scale system control panel to operate the scale system option (Scale
icon/button identification (page 32)).
Do not touch the product while you weigh or zero the scale system option.
The patient must remain still while the system is weighing. If the patient is
moving, the scale system will try for 20 seconds to get a stable weight or
zero value before displaying the error message ( ).
The error ( ) appears in the display if there is a loose connection or if the
scale is not operating. If you attempt to operate the scale again and the
error appears, contact Stryker Technical Support.
Note – To meet the scale system accuracy claim, the patient surface must be in
the flat position (Fowler backrest and Gatch down) and the product cannot
exceed 5 degrees of Trendelenburg/Reverse Trendelenburg.
Scale icon/button identification
Item A
Icon/button
Name Stand by icon
B
Zone 1 icon
Description
Illuminates green when you plug the power cord into a wall outlet.
Illuminates amber when you unplug the power cord from the wall outlet. Battery
backup functionality will turn on.
The patient can move freely, but the alarm sounds when the patient attempts to
leave the product.
C
Zone 2 icon
The patient can move with limited movement. The alarm sounds when
the patient approaches the siderails or moves toward the foot end.
D
Arm/Disarm button Arms and disarms the Chaperone option (stretcher exit). See Arming
or disarming the Chaperone option (page 33).
E
Weigh button
Weighs the patient. See Weighing a patient (page 32).
F
lb/kg button
Changes the displayed unit of measure. See Locking the scale unit of
measure (page 33).
G
Zero button
Zeros the scale settings. See Locking the scale unit of measure (page
33).
Weighing a patient
CAUTION – Do not use the scale system reading as a reference for medical
treatment. EN 32
1125-309-005 Rev AB.0
1. Raise the foot end push handles before you operate the scale system
option. 2. Press and hold Zero until rel displays on the screen to zero the
scale. 000.0 flashes and then stops flashing.
Note – The scale is not ready for use until the 000.0 stops flashing. 3. Place
the patient on the product. 4. Press lb/kg to select the unit of measure (lb
or kg). 5. Press Weigh to obtain the patient’s weight.
Note – The patient’s weight displays for 40 seconds.
Locking the scale unit of measure
1. Press and hold Weigh, lb/kg, and Zero until diag displays. 2. Press lb/kg
or Zero to toggle to Unit Loc. 3. For scale without Chaperone, press and hold
lb/kg and Zero until rel displays to select Unit Loc. 4. For scale with
Chaperone, press and hold lb/kg and Zero to select Unit Loc. 5. Press lb/kg or
Zero to toggle to lb, kg, or Any. 6. Press and release lb/kg and Zero. Only
and the selected unit of measure displays.
Note – Only will not appear if you select Any. 7. Press and hold Weigh until
quit displays. 8. Press and hold Zero until rel displays. 000.0 flashes and
then stops flashing.
Note – The scale is not ready for use until the 000.0 stops flashing.
Arming or disarming the Chaperone option
When armed, the Chaperone option monitors the patient’s position on the
product. To arm and disarm the Chaperone option: 1. With the patient on the
product, press Arm/Disarm.
Note – This arms Zone 1. Zone 1 illuminates amber when applied. 2. To arm Zone
2, press Arm/Disarm twice within 3 seconds.
Note – This arms Zone 2. Zone 2 illuminates amber when applied. 3. To disarm
the Chaperone option, press and hold Arm/Disarm.
Note – You must zero the scale before you arm the Chaperone option. Note – If
the patient moves outside of the limits of the selected zone, an amber light
flashes, Alrt flashes in the scale system option display, and a sound alarm is
triggered.
Setting the Chaperone option alert pattern
To change the Chaperone option alert pattern: 1. Press and hold both
Arm/Disarm and Weigh together for approximately 6 seconds. Ignore all display
messages until
Ptrn appears on the display (Scale icon/button identification (page 32)). 2.
Release the Arm/Disarm and Weigh buttons.
Note – The pattern setting P (1-10) appears on the display. 3. Press
Arm/Disarm or Weigh to scroll through the pattern options.
Note – A brief sample of the pattern plays as you scroll through the pattern
options.
1125-309-005 Rev AB.0
33 EN
4. Press and hold both Arm/Disarm and Weigh together for 6 seconds until set
appears on the display to save your selected pattern.
5. Release Arm/Disarm and Weigh.
Note – The pattern setting you selected appears on the display. A brief sample
of the pattern confirms your pattern setting.
Setting the Chaperone option alert volume
To change the Chaperone option alert volume: 1. Press and hold both Zero and
lb/kg together until vol appears on the display (Scale icon/button
identification (page
32)). 2. Release Zero and lb/kg.
Note – The volume levels L (1-4) appear on the display. 3. Press Zero or lb/kg
to scroll through the volume levels.
Note – A brief sample of the volume plays as you scroll through the volume
levels. 4. Press and hold both Zero and lb/kg together for 6 seconds until set
appears on the display to save your selected
volume level. 5. Release Zero and lb/kg.
Note – The volume level you selected appears on the display. A brief sample of
the volume level confirms your volume level setting.
Charging the scale system battery pack option – electric litter option or
electric litter option with Chaperone option – Prime only
The scale system with the electric litter option requires one 10.8V Li-ION
battery pack (0058-135-000 without the Chaperone option; 0058-134-000 with the
Chaperone option). Recharge the battery pack when the battery charge indicator
displays one unit remaining. This prevents the scale system option from
shutting down due to drained batteries.
To recharge the scale system Li-ION battery pack, plug the power cord into a
grounded, hospital grade wall outlet. The battery pack fully charges in
approximately three hours.
Note – Always plug the product into a grounded, hospital grade wall outlet
when not in use. This helps the battery to maintain a sufficient charge and
maximizes product performance while on battery power.
Replacing the batteries for the scale system option – non-electric litter
Replace the batteries when the battery charge indicator displays one unit
remaining. This prevents the scale system option from shutting down due to
drained batteries.
When the display flashes Lo batt, the batteries are at their lowest charge and
the scale system option is disabled.
Tools required:
· Phillips screwdriver · Four AA batteries (Alkaline type (LR6))
To replace the batteries: 1. Using a Phillips screwdriver, the screws that
hold the battery compartment cover to the display assembly. 2. Remove and
discard all four AA batteries.
Note – Always dispose of old batteries in accordance with local regulations.
3. Place four new AA batteries as indicated in the battery holder.
Note – Never mix old batteries with new batteries.
EN 34
1125-309-005 Rev AB.0
4. Use the screws removed in step 1 to reinstall the battery compartment cover on the display assembly.
Accessories and parts
These accessories may be available for use with your product. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-0770.
Name Defibrillator tray Defibrillator tray/foot extender Footboard/chart holder Heel pad assembly IV caddy HAVASUTM IV pole, removable IV pole, head end, left IV pole, head end, right IV pole, three-stage, foot end, left, 26″ IV pole, three-stage, foot end, right, 26″ IV pole, three-stage, foot end, left, 30″ IV pole, three-stage, foot end, right, 30″ IV pole, three-stage, foot end, left, 30″ (Prime X option) IV pole, three-stage, foot end, right, 30″ (Prime X option) IV pole, three-stage, head end, left, 26″ IV pole, three- stage, head end, right, 26″ IV pole, two-stage, foot end, left, 26″ IV pole, two-stage, foot end, right, 26″ IV pole, two-stage, foot end, left, 30″ IV pole, two-stage, foot end, right, 30″ IV pole, two-stage, foot end, left, 30″ (Prime X option) IV pole, two-stage, foot end, right, 30″ (Prime X option) IV pole, two-stage, head end, left, 30″ IV pole, two-stage, head end, right, 30″ Mattress, ComfortGel SETM, fire barrier, 26″ Mattress, ComfortGel SE, fire barrier, 30″ Mattress, ComfortGel SE, without fire barrier, 26″ Mattress, ComfortGel SE, without fire barrier, 30″ Mattress, Enhanced Comfort, 3″ x 26″
Part number 1105-045-200 1105-045-400 1105-045-500 1105-045-022 0785-155-000 0390-025-010 1125-060-145 1125-060-140 1105-035-644 1105-035-639 1105-035-344 1105-035-339 1105-035-364 1105-035-361 1125-035-642 1125-035-637 1105-035-640 1105-035-643 1105-035-340 1105-035-343 1105-035-362 1105-035-363 1125-035-338 1125-035-341 1805-034-601 1805-034-301 1805-034-600 1805-034-300 0785-034-613
1125-309-005 Rev AB.0
35 EN
Name
Part number
Mattress, Enhanced Comfort, 3″ x 30″
0785-034-313
Mattress, Enhanced Comfort, 4″ x 26″
0785-034-623
Mattress, Enhanced Comfort, 4″ x 30″
0785-034-323
Mattress, IsoFlex SETM, fire barrier, 30″
1806-034-300
Mattress, IsoFlex SE, without fire barrier, international, 30″ 1806-034-301
Mattress, Ultra Comfort, 4″ x 26″
0785-034-603
Mattress, Ultra Comfort, 4″ x 30″
0785-034-303
Mattress, Ultra Comfort, 5″ x 26″
0785-034-633
Mattress, Ultra Comfort, 5″ x 30″
0785-034-333
Mattress, Ultra Comfort SETM, 4″ x 26″
1704-034-600
Mattress, Ultra Comfort SE, fire barrier, international, 4″ x 1704-034-601 26″
Mattress, Ultra Comfort SE, 3″ x 30″
1703-034-300
Mattress, Ultra Comfort SE, fire barrier, 3″ x 30″
1703-034-301
Mattress, Ultra Comfort SE, 4″ x 30″
1704-034-300
Mattress, Ultra Comfort SE, fire barrier, 4″ x 30″
1704-034-301
Oxygen bottle holder, upright
1115-130-000
Pump bar option, 26″
1105-045-065
Pump bar option, 30″
1105-045-035
Restraint strap package
0785-045-010
Restraint strap, ankle
0785-045-020
Restraint strap, body
0785-045-015
Restraint strap, wrist
0946-044-000
Restraint strap, chest
1010-058-000
Serving tray
1105-045-700
Serving tray holder/footboard
1105-045-800
Siderail pads
1001-052-000
X-ray cassette holder, lateral
1105-045-100
X-ray cassette holder, upright
1105-045-300
Electrical components
Name Actuator, Fowler Actuator, Gatch Battery, Li-ION Smart battery pack
(scale only)
Number 1008-012-020 1008-014-020 0058-135-000
EN 36
1125-309-005 Rev AB.0
Name
Number
Battery, Li-ION Smart battery pack (scale and Chaperone) 0058-134-000
Board, DC control PCB assembly
1008-116-800
Board, non-lift AC assembly
1008-002-800
Board, lift control PCB assembly
1008-002-810
Board, scale control assembly (scale only)
1008-237-850
Board, scale and Chaperone control assembly
1008-237-840
Capacitor, 440 VAC, 35 MFD
0059-087-000
Keypad, scale control (non-electric litter)
1008-037-820
Keypad, siderail control, left
1008-011-017
Keypad, siderail control, right
1008-011-016
Keypad, operator control
1008-015-800
Load cell (scale system option)
1008-037-057
Motor, pump
1008-002-015
Power cord, standard
1028-146-060
Power cord, international, Type, E
1125-060-170
Power cord, international, Type, F
1125-060-180
Power cord, international, Type, G
1125-060-160
Power cord, international, Type, I
1125-060-200
Scale box assembly (non-electric litter)
1070-237-020
Transformer Transformer, international
1008-014-857 1008-014-877
Attaching the defibrillator tray
WARNING · Do not place items that weigh more than 30 lb (14 kg) on the
defibrillator tray. Always strap down all devices that you
place on the defibrillator tray. · Always use caution if the defibrillator
tray/foot extender, footboard/chart holder, or upright oxygen bottle holder is
attached to avoid pinching your fingers when you position the foot end push
handle option.
To attach the defibrillator tray: 1. Insert the defibrillator tray pins into
the sockets at the foot end of the product. 2. Use the strap to secure devices
to the defibrillator tray.
Note · Do not use the defibrillator tray as a push/pull device. Product damage
may occur. · Always raise the foot end push handles when you use accessories
(such as the defibrillator tray/foot extender,
footboard/chart holder, upright oxygen bottle holder) or the accessories will
not function.
1125-309-005 Rev AB.0
37 EN
Converting the defibrillator tray/foot extender to a defibrillator tray
WARNING · Do not place items that weigh more than 30 lb (14 kg) on the
defibrillator tray/foot extender. Always strap down all
devices that you place on the defibrillator tray. · Always use caution if the
defibrillator tray/foot extender, footboard/chart holder, or upright oxygen
bottle holder is
attached to avoid pinching your fingers when you position the foot end push
handle option.
CAUTION – Always raise the IV pole before you attach the defibrillator
tray/foot extender to the product. If you do not raise the IV pole, the foot
extender will not operate.
To convert the defibrillator tray/foot extender to a defibrillator tray: 1.
Pull out the top knob (A) (Figure 22). 2. Pivot the defibrillator tray (B)
until the tray is flat over the foot end of the product. Release the top knob
(A). Make sure
that the defibrillator tray is locked in place. 3. Use the strap to secure
devices to the defibrillator tray.
Note · Do not use the defibrillator tray/foot extender as a push/pull device.
Product damage may occur. · Do not attach items to the foot extender.
B
A
C
D
Figure 22 Defibrillator tray/foot extender
Converting the defibrillator tray/foot extender to a foot extender
WARNING · Do not place items that weigh more than 30 lb (14 kg) on the
defibrillator tray/foot extender. Always strap down all
devices that you place on the defibrillator tray. · Always use caution if the
defibrillator tray/foot extender, footboard/chart holder, or upright oxygen
bottle holder is
attached to avoid pinching your fingers when you position the foot end push
handle option.
CAUTION – Always raise the IV pole before you attach the defibrillator
tray/foot extender to the product. If you do not raise the IV pole, the foot
extender will not operate.
To convert the defibrillator tray/foot extender to a foot extender (Figure
22): 1. Pull out the top knob (A).
EN 38
1125-309-005 Rev AB.0
2. Pivot the defibrillator tray (B) until the tray locks against the foot
extender. 3. Pull out the bottom knob (D) while you hold the defibrillator
tray/foot extender assembly. 4. Lower the foot extender (C) until the foot
extender is flat. 5. Release the bottom knob (D). Push on the foot extender to
make sure that the foot extender is locked in place.
Note · Do not use the defibrillator tray/foot extender as a push/pull device.
Product damage may occur. · Do not attach items to the foot extender.
Attaching the footboard/chart holder
WARNING – Always use caution if the defibrillator tray/foot extender,
footboard/chart holder, or upright oxygen bottle holder is attached to avoid
pinching your fingers when you position the foot end push handle option.
To attach the footboard/chart holder, insert the footboard/chart holder pins
into the sockets at the foot end of the product. Note – Do not use the
footboard/chart holder as a push/pull device. Product damage may occur.
Attaching the IV caddy
WARNING · Always secure the IV pole to the IV caddy when you transport the
product. · Always store the IV caddy when not in use to avoid product damage.
To attach the IV caddy (Figure 23): 1. Lift the IV caddy out of the storage
tray or from the storage clip. 2. Pivot the IV caddy to the desired position.
3. Turn the knob (A) counterclockwise to loosen the pole clamp (C). 4. Pivot
the knob (A) away from the arm connection assembly (B). 5. Open the clamp (C).
6. Place the IV pole into the clamp (C). 7. Close the clamp (C) around the IV
pole and pivot the knob (A) into position. 8. Turn the knob (A) clockwise to
tighten the knob. 9. Reverse steps to disconnect the IV caddy from the
product.
To store the IV caddy, place the IV caddy in the storage tray or secure the
caddy in the storage clip.
1125-309-005 Rev AB.0
39 EN
C
B
A
Figure 23 IV caddy
Positioning or stowing the foot supports – Prime only
WARNING · Do not sit on the foot supports. This may cause the product to tip.
· Always clear your fingers from the mechanisms when you operate the foot
supports. · Always stow the foot supports when you transport a patient with
the product. · Always tighten the foot supports before you use the foot
supports. · Do not operate the scale system with the foot supports. Inaccurate
readings may occur. · Do not operate the Chaperone option with the foot
supports. Inaccurate readings may occur.
To position or stow the foot supports: 1. Loosen the knee knob (A) at the top
of the foot supports to adjust the side-to-side angle of the foot supports
(Figure 24). 2. Tighten the knee knob (A) to lock the foot supports in the
desired position.
CA B
C
A
B
Figure 24 Positioning the foot supports
3. Loosen the leg knob (B) on the side of the foot supports to adjust the
length. 4. Tighten the leg knob (B) to lock the foot supports in the desired
position. 5. Flip the foot supports (C) up before you position the patient. 6.
Reverse steps to stow the foot supports (Figure 25).
EN 40
1125-309-005 Rev AB.0
AC CA
B B
Figure 25 Stowing the foot supports
You cannot purchase the following options and accessories if you select the
foot support: · Foot end push handles · Defibrillator tray · Defibrillator
tray/foot extender · Serving tray holder/footboard · Footboard/chart holder ·
Foot end IV poles Note · Do not use the foot supports to store patient
belongings. · Do not use the foot supports as a push/pull device. Product
damage may occur. · Always apply the brakes when you use the foot supports to
avoid instability. · Do not use the backrest or Gatch with the foot supports.
Positioning the two-stage permanently attached IV pole
WARNING – Do not use the IV pole as a push/pull device. Product damage may
occur.
You can purchase the product with the two-stage IV pole option permanently
attached at the head end, foot end, or both ends of the product. The IV pole
is equipped with a telescopic pole that extends to provide a second height
position. You can fold and store the IV pole when not in use. To position the
two-stage IV pole (Figure 26): 1. Lift and pivot the pole from the storage
position. 2. Push the IV pole down until the IV pole locks in place. 3. To
raise the height of the IV pole, pull up on the telescoping portion (A) until
the pole locks in place at the fully raised
position. 4. Rotate the IV hangers (B) to the desired position and hang the IV
bags. 5. To lower the IV pole, hold the telescoping portion of the IV pole,
turn the latch (C), and lower the telescoping portion.
Note · Do not hang IV bags that exceed 40 lb (18 kg) on the IV pole. · Always
make sure that the IV pole is at a low height to pass safely through door
openings when you transport a
patient.
1125-309-005 Rev AB.0
41 EN
B A
A
C
Figure 26 Positioning the 2 stage permanently attached IV pole
Positioning the three-stage permanently attached IV pole
WARNING – Do not use the IV pole as a push/pull device. Product damage may
occur. You can purchase the product with the three-stage IV pole option
permanently attached at the head end, foot end, or both ends of the product.
The IV pole is equipped with a telescopic pole that extends to provide a
second and third height position. You can also fold and store the IV pole when
not in use. To position the three-stage IV pole (Figure 27): 1. Lift and pivot
the pole from the storage position. 2. Push the IV pole down until the pole
locks in place. 3. To raise the height of the IV pole, pull up on the
telescoping portion (A) until the pole locks into place at the fully raised
position. 4. For a higher IV pole, pull up on section (B). Release section (B)
at any desired height to lock the pole in place. 5. Rotate the IV hangers (C)
to the desired position and hang the IV bags. 6. To lower the IV pole, push up
on the yellow portion of the grip (D) while holding on to section (B) until
the pole lowers. 7. Turn the latch (E) and lower the IV pole telescoping
portion.
Note · Do not hang IV bags that exceed 12 lb (5 kg) total for all bags on the
IV pole. · Do not hang IV bags that exceed 9.3 lb (4.2 kg) on a single IV
hanger. · Always make sure that the IV pole is at a low height to allow the
pole to pass safely through door openings when you
transport a patient.
EN 42
1125-309-005 Rev AB.0
C
B
D
A E
Figure 27 Positioning the three-stage permanently attached IV pole
Attaching and positioning the removable IV pole
CAUTION · Do not use the IV pole as a push/pull device. Product damage may
occur. · Do not hang IV bags that exceed 40 lb (18 kg) on the IV pole. ·
Always make sure that the IV pole is at a low height to pass through door
openings when you transport a patient. To attach and position the removable IV
pole (Figure 28): 1. Insert the IV pole into a socket at the head end or foot
end of the product. 2. Turn the knob (A) counterclockwise and pull up on the
telescoping portion (B) until you reach the desired height. 3. Turn the knob
(A) clockwise to lock the telescoping portion in place.
1125-309-005 Rev AB.0
43 EN
B A
Figure 28 Removable IV pole
Attaching the upright oxygen bottle holder
WARNING · Do not place objects that exceed 40 lb (18 kg) in the upright oxygen
bottle holder. · Always use caution if the defibrillator tray/foot extender,
footboard/chart holder, or upright oxygen bottle holder is
attached to avoid pinching your fingers when you position the foot end push
handle option.
The upright oxygen bottle holder supports an oxygen bottle in a vertical
position. To attach the upright oxygen bottle holder: 1. Insert the support
bar into any of the IV sockets. 2. Insert the cotter pin through the hole in
the support bar to secure the bottle holder to the product.
Note – Do not use the upright oxygen bottle holder as a push/pull device.
Product damage may occur.
Extending or stowing the serving tray holder/footboard
WARNING – Do not place objects that exceed 30 lb (14 kg) on the serving tray.
To fit the serving tray on the siderail, pull out on both sides of the serving
tray and position the tray over the siderails. To stow the serving tray: 1.
Remove the serving tray from the siderails. 2. Push in the sides of the
serving tray. 3. Store the serving tray in the footboard.
Note – Do not use the serving tray/footboard as a push/pull device. Product
damage may occur.
EN 44
1125-309-005 Rev AB.0
Attaching the siderail pads
To attach the siderail pads: 1. Tuck the siderail pad between the mattress and
the siderail. 2. Fasten the Velcro® straps around the top of the siderail to
secure the siderail pad.
Locating the patient restraint strap tie-ins
WARNING · Always use caution when you attach restraint straps. Patient or
operator injury may occur. Physical restraints, even if
secured, may result in serious harm to patients and operators, including
entanglement, entrapment, physical injury, or death. · Always attach restraint
straps or devices only at the identified attachment points of the product.
Failure to do so may result in patient or operator injury. Do not attach
restraint straps to the siderail. · Always refer to the applicable state and
federal restrictions and regulations and the appropriate facility protocols
before you use any restraint strap or device.
There are eight patient restraint strap tie-in locations on the litter
assembly for attaching patient restraint straps (Figure 29 or Figure 30).
Figure 30 Prime X option restraint strap locations
Figure 29 Prime restraint strap locations
Note – Restraint straps are Type B applied parts.
Positioning the upright X-ray cassette holder – Prime X option
WARNING · Always take protective measures when you use the upright or lateral
X-ray cassette holder. The X-ray cassette holder
does not protect against radiation. · Always refer to local, state, and
federal guidelines in addition to facility protocols for safety before you use
Prime X
option with radiation generating devices. Radiation generating devices may
produce residual, stray, or scattered radiation. · Always use extra caution
when you use a mattress thicker than 2.5 in. or a mattress overlay with Prime
X option. · Always follow the Positioning the upright X-ray cassette holder –
Prime X option instructions to insert the X-ray cassette.
1125-309-005 Rev AB.0
45 EN
The X-ray cassette holder attaches to the Fowler backrest weldment to support
X-ray cassettes. You can take X-rays while a patient is on the product. You
can also adjust the cassette’s position before you take an X-ray. To position
the X-ray cassette holder: 1. Apply the brakes. Push on the product to make
sure that the brakes are working. 2. Raise the Fowler backrest section to full
height. 3. Insert the lower retainer guides (A) under the Fowler backrest
weldment bar (Figure 31). 4. Raise the cassette holder until the retainer
guides latch onto the Fowler backrest weldment. Make sure that the cassette
holder is secure.
C
C
B
A
D
E
Figure 31 X-ray cassette holder
5. Insert an X-ray cassette from either side of the cassette holder or pull
the slider release arrow (B) to release the cassette holder from the Fowler,
position the cassette, and secure the cassette holder in the Fowler backrest
weldment.
6. To adjust the height of the cassette, loosen the knob (E) and move the
cassette support rail (D) up or down until you reach the desired height.
7. Tighten the knob (E) to secure the cassette support rail in place. 8. When
the X-ray process is complete, pull up on the slider release arrow (B) to
release the cassette holder from the
Fowler backrest weldment. 9. Remove the X-ray cassette from the cassette
holder. 10. Close and stow the cassette holder.
Positioning the lateral X-ray cassette holder – Prime X option
WARNING · Always take protective measures when you use the upright or lateral
X-ray cassette holder. The X-ray cassette holder
does not protect against radiation. · Always refer to local, state, and
federal guidelines in addition to facility protocols for safety before you use
Prime X
option with radiation generating devices. Radiation generating devices may
produce residual, stray, or scattered radiation.
EN 46
1125-309-005 Rev AB.0
· Always use caution when you take X-rays with the Fowler backrest in the
upright position or when you use a lateral cassette.
· Always use extra caution when you use a mattress thicker than 2.5 in. or a
mattress overlay with Prime X option. · Always follow the Positioning the
lateral X-ray cassette holder – Prime X option instructions to insert the
X-ray cassette.
To position the lateral X-ray cassette holder: 1. Apply the brakes. Push on
the product to make sure that the brakes are working.
D
B
C A
B C
Figure 32 Lateral X-ray cassette holder
2. Press down on the yellow release button (A) to open the lateral cassette
holder (Figure 32). 3. Slide the flat base (B) between the mattress and the
patient platform. 4. Position the lateral cassette holder in the desired
location. 5. Place the X-ray cassette in the cassette canal (C). 6. Pull up on
the cassette hook (D) to adjust the height of the arm to fit over the X-ray
cassette to hold the cassette in
place. 7. Reverse steps to remove the cassette and stow the lateral cassette
holder.
Inserting or removing X-ray cassettes – Prime X option
WARNING · Always refer to local, state, and federal guidelines in addition to
facility protocols for safety before you use Prime X
option with radiation generating devices. Radiation generating devices may
produce residual, stray, or scattered radiation. · Always use caution when you
take X-rays with the Fowler backrest in the upright position or when you use a
lateral cassette.
Prime X option provides both an articulating radiographic support surface and
a platform below the patient support surface for X-ray cassette placement.
Working with medical X-ray systems, the radiographic support surface allows
the capture of clinical X-rays (AP full body, optional full body lateral, and
optional upright chest) while the patient is on the product. You can insert
cassettes from the head end, foot end, and either side of the product.
To insert an X-ray cassette:
1. Center the patient on the product with the position indicator labels
located on all sides of the product (Figure 33).
2. Insert an X-ray cassette below the patient surface. Use the cassette
guides to assist in positioning the X-ray cassette.
1125-309-005 Rev AB.0
47 EN
Figure 33 Inserting or removing X-ray cassettes on Prime X option
Note · Do not use a mattress with a thickness greater than four inches with
the Prime X option. · Do not use a C-Arm with Prime X option. Prime X option
is not compatible with a C-Arm.
EN 48
1125-309-005 Rev AB.0
Cleaning
Cleaning the product
WARNING · Always unplug the power cord and turn the On/Drive – Off/Manual
switch to the Off position before service or cleaning. · Do not steam clean,
pressure wash, hose off, or ultrasonically clean the product.
CAUTION – Do not use abrasive cleaners to clean the display enclosure for the
scale system option. Do not allow cleaning solutions or other fluids to pool
on the display unit. Wipe dry all surfaces after spills or cleaning.
These instructions provide recommended cleaning methods for Model 1125 Prime
Series Stretcher.
Recommended cleaning method: 1. If the product is equipped with the electric
lift option or electric litter option, unplug the power cord from the wall
outlet
before you transport or clean the product. 2. Remove the mattress from the
product. 3. Follow the cleaning solution manufacturer’s dilution
recommendations. 4. Dry the product. Do not place the mattress on the product
until the product is dry.
Note · Direct skin contact with visibly soiled, permeable material may
increase the risk of infection. · Clean the base hood storage area. · Clean
the bottom of the brake pads to prevent wax or floor remnant buildup. · Some
cleaning agents are corrosive in nature and may cause damage to the product if
you use them improperly. If
you do not rinse and dry the product well, a corrosive residue may be left on
the surface of the product that could cause premature corrosion of critical
components. Failure to follow these cleaning instructions may void your
warranty.
Cleaning the mattress
WARNING · Do not clean, service, or perform maintenance while the product is
in use. · Do not immerse the mattress in cleaning or disinfectant solutions.
Excess moisture could cause product malfunction that
results in product damage or patient injury. · Do not allow fluid to pool on
the mattress. Fluids can cause corrosion of components and may cause the
safety and
performance of this product to become unpredictable. · Always inspect mattress
covers for tears, punctures, excessive wear, and misaligned zippers every time
you clean the
covers. Remove and replace a damaged mattress to prevent cross-contamination.
· Do not steam clean, pressure wash, hose off, or ultrasonically clean
mattresses. These methods of cleaning may void
the product’s warranty.
The life of the mattress can be affected by an increase in frequency of usage,
which might include more frequent cleaning and disinfection.
Recommended cleaning method: 1. Use a clean, soft cloth to wipe down the
entire mattress with a mild soap and water solution to remove foreign
material. 2. Wipe down the mattress with a clean, dry cloth to remove any
excess liquid or cleaning agents. 3. Rinse and dry covers after cleaning. 4.
Disinfect as needed with a hospital grade disinfectant after cleaning has been
completed (see Disinfecting the
mattress).
1125-309-005 Rev AB.0
49 EN
Note
· Do not iron, dry-clean, or tumble dry the mattress, as this will cause
malfunction and damage the product.
· The mattress cover must be completely dry before you store, add linens, or
place a patient on the mattress to prevent impairment of the product
performance.
· Avoid over-exposure to alcohol or hydrogen peroxide. The cover material will
swell.
· Do not allow liquid to seep into the zipper area and watershed cover
barrier. Fluids allowed to come in contact with the zipper may leak into the
mattress which could impair the product performance.
· Some cleaning agents are corrosive in nature and may cause damage to the
product if you use them improperly. If you do not rinse and dry the product
well, a corrosive residue may be left on the surface of the product that could
cause premature corrosion of critical components. Failure to follow these
cleaning instructions may void your warranty.
Remove iodine
1. Make a solution of 1 to 2 tablespoons of sodium thiosulfate in a pint of
warm water. Use the solution to wipe down the stained area.
2. Clean the stain as soon as possible after the stain occurs. 3. If stains
are not immediately removed, allow solution to soak or stand on the mattress
before you wipe the mattress. 4. Rinse the mattresses which have been exposed
to the solution with clear water before you return the mattresses to
service.
Note – Failure to follow these directions when you use these types of cleaners
may void this product’s warranty.
Special instructions
Velcro® Solids or stains Hard-to-clean spots
Laundering
Saturate with disinfectant, rinse with water, and allow the solution to
evaporate.
Use neutral soaps and warm water. Do not use harsh cleansers, solvents, or
abrasive cleaners.
Use standard household cleansers or vinyl cleansers and a soft bristle brush
on troublesome spots or stains. Pre-soak dried-on soil.
Laundering is not recommended. Laundering may substantially decrease the
useful life of the mattress.
EN 50
1125-309-005 Rev AB.0
Disinfecting
Disinfecting the product
WARNING · Do not clean, service, or perform maintenance while the product is
in use. · Do not steam clean, hose off, or ultrasonically clean the product.
Use of these methods of cleaning is not recommended
and may void the product’s warranty.
Recommended disinfectants:
· Quaternaries (active ingredient – ammonium chloride) that contain less than
3% glycol ether · Phenolic disinfectant (active ingredient – o-phenylphenol) ·
Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts
water which equals 520 ppm available
chlorine (40 mL of a 5.25% bleach solution per 4000 mL water)) · 70% isopropyl
alcohol
Recommended disinfection method: 1. Follow the disinfectant solution
manufacturer’s dilution recommendations. 2. Hand wash all surfaces of the
product with a disinfectant solution. 3. Avoid over-saturation and make sure
that the product does not stay wet longer than the chemical manufacturer’s
guidelines for proper disinfecting. 4. Dry the product. Do not place the
mattress on the product until the product is dry. 5. Disinfect the Velcro®
after every use. Saturate the Velcro® with disinfectant, rinse with water, and
allow the
disinfectant to evaporate (appropriate disinfectant is determined by the
facility). 6. Check functionality before you return the product to service.
· Raise and lower the product · Lock and unlock the brake/steer pedal in both
positions · Latch and unlatch the siderails · Raise and lower the Fowler
backrest · Raise and lower the Gatch · Make sure all components have proper
lubrication · Make sure all labels are intact
Note · Direct skin contact with visibly soiled, permeable material may
increase the risk of infection. · Some cleaning agents are corrosive in nature
and may cause damage to the product if you use them improperly. If
you do not rinse and dry the product well, a corrosive residue may be left on
the surface of the product that could cause premature corrosion of critical
components. Failure to follow these cleaning instructions may void your
warranty.
Disinfecting the mattress
WARNING – Always disinfect the mattress between patients. Failure to do so
could result in cross-contamination and infection.
Recommended disinfectants:
· Quaternaries (active ingredient – ammonium chloride) that contain less than
3% glycol ether
1125-309-005 Rev AB.0
51 EN
· Phenolic disinfectant (active ingredient – o-phenylphenol)
· Chlorinated bleach solution (5.25% bleach diluted 1 part bleach to 100 parts
water which equals 520 ppm available chlorine (40 mL of a 5.25% bleach
solution per 4000 mL water))
· 70% isopropyl alcohol
Recommended disinfection method:
1. Make sure that the mattress is clean and dry before you apply
disinfectants.
2. Wipe down the mattress with a clean, dry cloth to remove any excess liquid
or disinfectant.
3. Rinse and dry covers after disinfection.
Note · The mattress cover must be dry before you store or add linens. Failure
to remove excess disinfectant could cause
degradation of the cover material. · Some cleaning agents are corrosive in
nature and may cause damage to the product if you use them improperly. If
you do not rinse and dry the product well, a corrosive residue may be left on
the surface of the product that could cause premature corrosion of critical
components. Failure to follow these cleaning instructions may void your
warranty. · Frequent or prolonged exposure to higher concentrations of
disinfectant solutions may prematurely age the cover fabric. · The use of
accelerated hydrogen peroxides or quaternaries that contain glycol ethers may
damage the cover.
EN 52
1125-309-005 Rev AB.0
Preventive maintenance
Remove product from service before you perform the preventive maintenance
inspection. Check all items listed during annual preventive maintenance for
all Stryker Medical products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by
qualified personnel.
Note – Clean and disinfect the exterior of the mattress before inspection, if
applicable.
Inspect the following items:
All fasteners are secure Siderails move and latch Casters lock when you apply
the brake pedal Casters are secure and swivel Casters are free of wax and
debris Brake mechanism works Steer function works Check for cracks Fowler
backrest raises, lowers, and latches Gatch raises, lowers, and latches
(option) Trendelenburg/Reverse Trendelenburg operates from all locations
Ground chain intact No leaks at hydraulic connections Hydraulic jacks hold
Lubricate where required Body restraints work (option) IV pole is intact and
adjusts and latches in all positions (option) Oxygen bottle holder is intact
and operates (option) No rips or cracks in the mattress cover Accessories and
mounting hardware are in good condition Battery powered functionality (option)
Cables are not worn or pinched (option) Power cord and plug are free of damage
(option) All electrical connections are tight (option) All grounds secure to
the frame (option) Ground impedance not more than 200 m (milliohms) (option)
(Max: test points include electronics enclosure and motor chassis mounted to
the base of the product) Current leakage not more than 300 µA (microamps) (per
UL 60606-1-1) (option) Batteries charged (scale system option) Display housing
is intact and not damaged (scale system option) Load cells are intact and not
damaged (scale system option) Foot support knee knob mechanism functions
(option) Foot support leg knob mechanism functions (option) Foot support
extends to the full extended position and stops in the correct position
(option) Foot support self-tapping screws (6) are secured and are not stripped
(option)
1125-309-005 Rev AB.0
53 EN
Scale calibrated. Recalibrate, if necessary (scale system option) Upright
X-ray cassette holder is working (Prime X option) No damage to the Fowler
backrest skin and foot skin (Prime X option) No damage to the head and foot
trays (Prime X option) Bolt and nut through each Fowler backrest pivot are not
loose (Prime X option) No excessive play in the drive handles Press the handle
switches and make sure that the product does not move unless the handles are
pushed forward or pulled back Press the handle switches and make sure that the
product moves forward and backward
Product serial number: Completed by:
Date:
EN 54
1125-309-005 Rev AB.0
EMC information
CAUTION
· The use of accessories, transducers, and cables, other than those specified
or provided by the manufacturer, could result in increased electromagnetic
emissions or decreased electromagnetic immunity and result in improper
operation.
· The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment, for which CISPR 11 class B is normally required, this
equipment might not offer adequate protection to radio frequency communication
services. The user might need to take mitigation measures, such as relocating
or reorienting the equipment.
Guidance and manufacturer’s declaration – electromagnetic emissions
The electric lift option, electric litter option, or scale system option is intended for use in an electromagnetic environment specified below. The operator of the electric lift option, electric litter option, or scale system option should make sure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
RF Emissions CISPR 11
Group 1
The electric lift option, electric litter option, or scale system option uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11
Harmonic Emissions IEC 61000-3-2
Voltage Fluctuations Flicker Emissions IEC 61000-3-3
Class A Class A Complies
The electric lift option, electric litter option, or scale system option is suitable for use in all establishments other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes.
- Applies to electric lift option or electric litter option only.
CAUTION · Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should
be used no closer than 12 inches (30 cm) to any part of the Prime stretcher, including cables specified by the manufacturer. · Avoid stacking or placing other equipment adjacent with other equipment to prevent improper operation of the products. If such use is necessary, carefully observe stacked or adjacent equipment to make sure that they are operating properly.
Recommended separation distances between portable and mobile RF communications
equipment and the electric lift option, electric litter option, or scale
system option.
The electric lift option, electric litter option, or scale system option is
intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the electric lift
option, electric litter option, or scale system option can help prevent
electromagnetic interferences by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
electric lift option, electric litter option, or scale system option as
recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power of transmitter W
0.01
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz D=(1.2) (P) 0.12
80 MHz to 800 MHz D=(1.2) (P) 0.12
800 MHz to 2.7 GHz D=(2.3) (P) 0.23
1125-309-005 Rev AB.0
55 EN
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance (D) in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 – At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2 – These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Guidance and manufacturer’s declaration – electromagnetic immunity
The electric lift option, electric litter option, or scale system option is intended for use in the electromagnetic environment specified below. The customer or the user of the electric lift option, electric litter option, or scale system option should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment guidance
Electrostatic Discharge (ESD) IEC 61000-4-2
±8 kV contact ±15 kV air
±8 kV contact ±15 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrostatic fast transient/ burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, voltage variations and short interruptions on power supply input
lines IEC 61000-4-11*
±2 kV for power supply lines ±1 kV for input/output lines
±2 kV for power supply lines ±1 kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
±1 kV line(s) to line(s) ±2 kV line(s) to earth
±1 kV line(s) to line(s) ±2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital environment.
0% UT for 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT for 1 cycle
70% UT (30% dip in UT) for 25 cycles
0% UT for 250 cycles
0% UT for 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270°, and 315°
0% UT for 1 cycle
70% UT (30% dip in UT) for 25 cycles
0% UT for 250 cycles
Mains power quality should be that of a typical commercial or hospital environment. If the user of the electric lift option or electric litter option requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterrupted power supply or a battery.
Power frequency (50/60 Hz) 30 A/m magnetic field IEC 61000-4-8
30 A/m
Note: Ut is the AC mains voltage before application of the test level.
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
*Applies to the electric lift option or electric litter option only.
EN 56
1125-309-005 Rev AB.0
Guidance and manufacturer’s declaration – electromagnetic immunity
The electric lift option, electric litter option, or scale system option is intended for use in the electromagnetic environment specified below. The customer or user of the electric lift option, electric litter option, or scale system option should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment guidance
Portable and mobile RF communications equipment should be used no closer to any part of the electric lift option, electric litter option, or scale system option, including cables, than the recommended separation distance calculated from the equation appropriate for the frequency of the transmitter.
Recommended separation distance
D=(1.2) (P)
Conducted RF IEC 61000- 4-6*
Radiated RF IEC 61000-4-3
3 Vrms 6 Vrms in ISM and amateur radio bands 150kHz to 80MHz
3 V/m 80 MHz to 2.7 GHz
3 Vrms
6 Vrms in ISM and amateur radio bands
3 V/m
D=(1.2) (P)
80 MHz to 800 MHz
D=(2.3) (P)
800 MHz to 2.7 GHz
where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and D is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey a, should be less than the compliance level in each frequency range b.
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
Note 3: The ISM (industrial, scientific and medical) bands between 0.15 MHz
and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
27.283 MHz; and 40.66 MHz to 40.70 MHz.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the electric lift option,
electric litter option, or scale system option is used exceeds the applicable
RF compliance level above, the electric lift option, electric litter option,
or scale system option should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the electric lift option, electric litter option,
or scale system option.
b Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3
V/m.
*Applies to electric litter option only.
1125-309-005 Rev AB.0
57 EN
Lehátko Prime Series®
Pírucka pro obsluhu 1125
s motorizovaným pohonem Zoom®
CS
1125-309-005 Rev AB.0
2021-12
Symboly
Prostudujte si pírucku / brozuru s pokyny Pectte si návod k pouzití Obecné varování Upozornní Varování; rozdrcení rukou Varování; neionizující záení
Netlacte
Neskladujte kyslíkovou láhev Katalogové císlo Sériové císlo Císlo sarze Patenty USA viz www.stryker.com/patents Znacka CE Zplnomocnný zástupce pro Evropské spolecenství Evropský zdravotnický prostedek
Výrobce
Datum výroby
Dovozce
Bezpecná pracovní zátz
1125-309-005 Rev AB.0
CS
~
IPX5
87VL
MR
Maximální hmotnost pacienta
Hmotnost výrobku vcetn bezpecné pracovní zátze Stejnosmrný proud
Stídavý proud
Nebezpecné naptí
Ekvalizace potenciálu
Ochranné uzemnní Ochrana proti silnému proudu vody Pílozná cást typu B
Zdravotnické vybavení klasifikované spolecností Underwriters Laboratories
Inc., pouze pokud jde o nebezpecí zásahu elektrickým proudem, nebezpecí pozáru
nebo mechanická nebezpecí, podle norem ANSI/AAMI ES60601-1:2005 a
CAN/CSA-C22.2 c. 60601-1:08. *Pro volitelný doplnk Prime X viz Píslusenství
RTG zaízení, IEC 60601-2-54 Podle smrnice Evropského parlamentu a Rady
2012/19/EU o odpadních elektrických a elektronických zaízeních (OEEZ) ve znní
pozdjsích pedpis tento symbol oznacuje, ze výrobek se musí sbírat k recyklaci
oddlen. Nelikvidujte s bzným netídným odpadem. Informace o likvidaci si
vyzádejte od místního distributora. Ped recyklací zajistte dekontaminaci
infikovaných prostedk.
Není bezpecné pro vysetení MR
Touto stranou nahoru
Uchovávejte v suchu
Kehké, manipulujte opatrn Limit stohování podle poctu
CS
1125-309-005 Rev AB.0
Obsah
Definice varování/upozornní/poznámky …………………………………………………………………………………………..3
Pehled bezpecnostních opatení ………………………………………………………………………………………………3 Místa, kde
hrozí skípnutí …………………………………………………………………………………………………..6
Úvod ………………………………………………………………………………………………………………………………………… 7 Popis výrobku
……………………………………………………………………………………………………………………….7 Indikace k pouzití
…………………………………………………………………………………………………………………..7 Klinické
pínosy……………………………………………………………………………………………………………………..8 Kontraindikace
………………………………………………………………………………………………………………………8 Ocekávaná zivotnost
………………………………………………………………………………………………………………8 Likvidace / recyklace
………………………………………………………………………………………………………………8 Technické parametry
………………………………………………………………………………………………………………8 Obrázek výrobku
Prime……………………………………………………………………………………………………….11 Ilustrace produktu moznost
Prime X ……………………………………………………………………………………….12 Pílozné cásti elektricky
ovládaná lozná plocha nosítek (volitelný doplnk)………………………………………13 Pílozné
cásti volitelný doplnk Prime X…………………………………………………………………………………..13 Kontaktní
informace ……………………………………………………………………………………………………………..13 Umístní sériového
císla………………………………………………………………………………………………………..14
Píprava …………………………………………………………………………………………………………………………………..15 Provoz
…………………………………………………………………………………………………………………………………….16
Zabrzdní a odbrzdní brzd ……………………………………………………………………………………………………16 Ovládání
podvozku Prime ………………………………………………………………………………………………………17 Zvýsení nosítek bez
elektrického mechanismu ……………………………………………………………………………17 Snízení nosítek bez
elektrického mechanismu ……………………………………………………………………………17 Zvýsení elektricky
ovládané lozné plochy nosítek (volitelný doplnk)………………………………………………..18 Snízení
volitelné elektricky ovládané lozné plochy nosítek……………………………………………………………..18
Umístní výrobku do Trendelenburgovy polohy …………………………………………………………………………..19
Umístní výrobku do obrácené Trendelenburgovy polohy
……………………………………………………………..19 Peprava pacienta bez pouzití kola Big Wheel
…………………………………………………………………………….19 Peprava pacienta za pouzití kola Big Wheel
………………………………………………………………………………20 Peprava pacienta za pouzití motorizovaného
pohonu Zoom ………………………………………………………….21 Dobíjení baterie motorizovaného pohonu
Zoom…………………………………………………………………………..23 Peprava pacienta rucn bez pouzití kola
motorizovaného pohonu Zoom ………………………………………….24 Zvednutí
postranice………………………………………………………………………………………………………………24 Spustní
postranice………………………………………………………………………………………………………………25 Nastavení polohy
výrobku s ovládacím panelem pro pacienta na postranici varianta s elektricky
ovládanou
loznou plochou nosítek pouze Prime…………………………………………………………………………..25 Ovládací
panel pro obsluhu volitelná elektricky ovládaná lozná plocha nosítek pouze
Prime ……………..26 Zablokování a odblokování funkcí ovládacího panelu pro
pacienta volitelná elektricky ovládaná lozná
plocha nosítek pouze Prime ……………………………………………………………………………………..27 Zvednutí
nebo snízení zádové cásti nosítka bez elektrického mechanismu
……………………………………..27 Zvednutí nebo snízení zádové cásti varianta s elektricky
ovládanou loznou plochou nosítek ……………….28 Zvedání a sklápní podkolenní cásti
nosítka bez elektrického mechanismu pouze Prime………………….29 Zvedání a
sklápní podkolenní cásti nosítka s elektrickým mechanismem pouze
Prime……………………29 Poloha zotavovacího kesla pouze Prime
………………………………………………………………………………..30 Ukládání pedmt do krytu podvozku
………………………………………………………………………………………30 Zavsování zdravotnických prostedk pi
pouzití volitelné konzoly pro pumpu …………………………………..31 Vytazení a zatazení
napájecího kabelu pomocí volitelného navijáku na kabel pouze u volitelného
elektrického zvedacího mechanismu nebo volitelné elektricky ovládané lozné
plochy nosítek ………………………………………………………………………………………………………………….. 31 Systém váhy
(volitelný doplnk) ………………………………………………………………………………………………32 Identifikace
ikon/tlacítek váhy ……………………………………………………………………………………………32 Vázení pacienta
…………………………………………………………………………………………………………….33 Uzamknutí mrné jednotky na váze
……………………………………………………………………………………33 Aktivace a deaktivace funkce Chaperone
…………………………………………………………………………….33 Nastavení alarm pro funkci
Chaperone………………………………………………………………………………34 Nastavení hlasitosti alarmu pro
funkci Chaperone ………………………………………………………………….34 Nabíjení sady baterií systému
váhy (volitelný doplnk) elektricky ovládaná lozná plocha nosítek
(volitelný doplnk) nebo elektricky ovládaná lozná plocha nosítek s funkcí
Chaperone (volitelný doplnk) pouze Prime
……………………………………………………………………………………….34 Výmna baterií volitelného systému váhy
nosítka bez elektrického mechanismu ………………………..35
1125-309-005 Rev AB.0
1 CS
Doplky a soucásti ……………………………………………………………………………………………………………….35 Pipevnní desky pro defibrilátor………………………………………………………………………………………..38 Pemna desky pro defibrilátor / nozní nástavec v desku pro defibrilátor ……………………………………..38 Pemna desky pro defibrilátor / nozní nástavec v nozní nástavec……………………………………………..39 Pipevnní dolní pelesti / drzáku dokument …………………………………………………………………………40 Pipevnní IV podstavce ………………………………………………………………………………………………….40 Nastavení nebo ulození oprek pro nohy pouze Prime………………………………………………………….40 Umístní dvoudílného trvale pipevnného IV stojanu …………………………………………………………….42 Umístní trojdílného trvale pipevnného IV stojanu ……………………………………………………………….42 Pipevnní a úprava polohy snímatelného IV stojanu………………………………………………………………43 Pipojení svislého drzáku kyslíkové láhve …………………………………………………………………………….44 Vytazení nebo odstranní drzáku servírovacího podnosu / dolní pelesti ………………………………………44 Pipevnní podlozek postranice …………………………………………………………………………………………45 Umístní vázání pro pacientské upínací pásy ……………………………………………………………………….45 Umístní svislého drzáku RTG kazety volitelný doplnk Prime X …………………………………………….45 Umístní bocního drzáku RTG kazety volitelný doplnk Prime X…………………..