NATIONAL INSTRUMENTS PXIe-5163 PXIe 14-bit PXI Oscilloscope
Product Information
The PXIe-5163 is a 200 MHz, 1 GS/s oscilloscope. It is a PXIe module that is designed to be used in a PXI chassis with a compatible controller. The oscilloscope provides high-speed data acquisition and signal analysis capabilities.
Contents of the PXIe-5163 Kit
- PXIe-5163 oscilloscope module
- PXIe-5163 Safety, Environmental, and Regulatory Information
- PXIe-5163 Getting Started Guide (this document)
- Maintain Forced-Air Cooling Note to Users
- Export Requirements Note to Users Information
Probe Options for the PXIe-5163
NI Part Number | Description |
---|---|
784254-01 | 400 MHz, 2 m, 10X cable divider probe |
784253-01 | 500 MHz, 1.2 m, 10X cable divider probe |
781785-01 | 800 MHz differential active probe |
784255-01 | 1 GHz single-ended active probe |
784256-01 | 1.5 GHz single-ended active probe |
784257-01 | 2.5 GHz single-ended active probe |
783630-01 | 100x passive probe, 500 MHz |
783629-01 | 10x passive probe, 500 MHz |
786846-01 | 5 A RMS, 50 MHz current probe |
786847-01 | 5 A RMS, 120 MHz current probe |
786848-01 | 500 A RMS, 2 MHz current probe |
786849-01 | 150 A RMS, 10 MHz current probe |
785561-01 | 30 A RMS, 50 MHz current probe |
785562-01 | 30 A RMS, 100 MHz current probe |
Product Usage Instructions
Verifying the System Requirements
Before using the NI-SCOPE instrument driver, ensure that your system meets
the minimum system requirements. Refer to the product readme for more
information about the requirements and supported application development
environments (ADEs).
Unpacking the Kit
Follow these steps to unpack the PXIe-5163 kit:
- Refer to Figure 1 to identify the contents of the kit.
- Ground yourself using a grounding strap or by holding a grounded object.
- Touch the antistatic package to a metal part of the computer chassis.
- Remove the module from the package and inspect it for loose components or damage.
- Unpack any other items and documentation from the kit.
- Store the module in the antistatic package when not in use.
Preparing the Environment
Ensure that the environment in which you are using the PXIe-5163 meets the
specified specifications.
GETTING STARTED GUIDE
PXIe-5163
PXIe, 200 MHz, 1 GS/s, 14-bit PXI Oscilloscope
Note: Before you begin, install and configure your chassis and
controller.
This document explains how to install, configure, and test the PXIe-5163. The
PXIe-5163 is a 200 MHz, 1 GS/s oscilloscope.
Documentation for the PXIe-5163 is available from the Start menu by navigating
to NI-SCOPE Documentation in the National Instruments folder.
Verifying the System Requirements
To use the NI-SCOPE instrument driver, your system must meet certain
requirements.
Refer to the product readme, which is available on the driver software media
or online at ni.com/manuals, for more information
about minimum system requirements, recommended system, and supported
application development environments (ADEs).
Unpacking the Kit
Refer to the following figure to identify the contents of the PXIe-5163
kit.
Figure 1. PXIe-5163 Kit Contents
- PXIe-5163
- PXIe-5163 Safety, Environmental, and Regulatory Information
- PXIe-5163 Getting Started Guide (this document)
- Maintain Forced-Air Cooling Note to Users
- Export Requirements Note to Users
Notice: To prevent electrostatic discharge (ESD) from damaging the module, ground yourself using a grounding strap or by holding a grounded object, such as your computer chassis.
- Touch the antistatic package to a metal part of the computer chassis.
- Remove the module from the package and inspect it for loose components or other signs of damage.
Notice: Never touch the exposed pins of connectors.
Note: Do not install a module if it appears damaged in any way. 3. Unpack any other items and documentation from the kit.
Store the module in the antistatic package when the module is not in use.
Other Equipment
There are several required items not included in your device kit that you need
to operate the PXIe-5163. Your application may require additional items not
included in your kit to install or operate your device.
Required Items
- A PXI Express chassis and chassis documentation
- A PXI Controller or PC with a Device for PXI Remote Control system that meets the system requirements specified in this guide and chassis documentation
Optional Items
- PXI Slot Blocker Kit, 5 slot blockers (NI part number 199198-01)
- PXI Filler Panel Kit, 17 filler panels (NI part number 778646-01)
- BNC-to-BNC cable, 50 Ω, 1 m (NI part number 781887-01)
- AUX 0 connectivity:
- SCB-19 noise rejecting, shielded AUX 0 connector block (NI part number 783959-01)
- SHH19-MH19-AUX shielded single-ended cable, 1 m (NI part number 784091-01)
- Mini-HDMI Breakout Cable to 6 BNC, 6 in. (NI part number 786979-01)
- Oscilloscope probe compensator, SMB (NI part number 786983-01)
- Probes—refer to the following table for several options compatible with the PXIe-5163.
Table 1. Probe Options for the PXIe-5163
Model | NI Part Number | Description |
---|---|---|
CP400X | 784254-01 | 400 MHz, 2 m, 10X cable divider probe |
CP500X | 784253-01 | 500 MHz, 1.2 m, 10X cable divider probe |
NI 5191 | 781785-01 | 800 MHz differential active probe |
SA1000X | 784255-01 | 1 GHz single-ended active probe |
SA1500X | 784256-01 | 1.5 GHz single-ended active probe |
SA2500X | 784257-01 | 2.5 GHz single-ended active probe |
SP500C | 783630-01 | 100x passive probe, 500 MHz |
SP500X | 783629-01 | 10x passive probe, 500 MHz |
CC0550X | 786846-01 | 5 A RMS, 50 MHz current probe |
CC05120X | 786847-01 | 5 A RMS, 120 MHz current probe |
Model | NI Part Number | Description |
--- | --- | --- |
CC5002X | 786848-01 | 500 A RMS, 2 MHz current probe |
CC15010X | 786849-01 | 150 A RMS, 10 MHz current probe |
CC3050X | 785561-01 | 30 A RMS, 50 MHz current probe |
CC30100X | 785562-01 | 30 A RMS, 100 MHz current probe |
Visit ni.com for more information about these additional items.
Prep aring the Environment
Ensure that the environment in which you are using the PXIe-5163 meets the
following specifications.
Temperature
- Storage -40 °C to 71 °C
Humidity
- Operating 10% to 90%, noncondensing
- Storage 5% to 95%, noncondensing
- Pollution Degree 2
- Maximum altitude 4,600 m (570 mbar) (at 25 °C ambient temperature)
Indoor use only.
Installing the Software
You must be an Administrator to install NI software on your computer.
-
Install an ADE, such as LabVIEW or LabWindows™/CVI™.
-
Visit ni.com/downloads/drivers and search for NI-SCOPE.
-
Download the latest version of NI-SCOPE, extract the downloaded files, and run the executable (.exe) file.
Driver support for the PXIe-5163 was first available in NI-SCOPE 18.7. -
Follow the instructions in the installation prompts.
Note: Windows users may see access and security messages during installation. Accept the prompts to complete the installation. -
When the installer completes, select Restart in the dialog box that prompts you to restart, shut down, or restart later.
Installing the PXIe-5163
Notice To prevent damage to the PXIe-5163 caused by ESD or contamination,
handle the module using the edges or the metal bracket.
- Ensure the AC power source is connected to the chassis before installing the module. The AC power cord grounds the chassis and protects it from electrical damage while you install the module.
- Power off the chassis.
- Inspect the slot pins on the chassis backplane for any bends or damage prior to installation. Do not install a module if the backplane is damaged.
- Remove the black plastic covers from all the captive screws on the module front panel.
- Identify a supported slot in the chassis. The following figure shows the symbols that indicate the slot types.
Figure 2. Chassis Compatibility Symbols
-
PXI Express System Controller Slot
-
PXI Peripheral Slot
-
PXI Express Hybrid Peripheral Slot
-
PXI Express System Timing Slot
-
PXI Express Peripheral Slot
PXIe-5163 modules can be placed in PXI Express peripheral slots, PXI Express hybrid peripheral slots, or PXI Express system timing slots. -
Touch any metal part of the chassis to discharge static electricity.
-
Ensure that the ejector handle is in the downward (unlatched) position.
-
Place the module edges into the module guides at the top and bottom of the chassis. Slide the module into the slot until it is fully inserted.
Figure 3. Module Installation -
Chassis
-
Hardware Module
-
Ejector Handle in Downward (Unlatched) Position
-
Latch the module in place by pulling up on the ejector handle.
-
Secure the module front panel to the chassis using the front-panel mounting screws.
Note: Tightening the top and bottom mounting screws increases mechanical stability and also electrically connects the front panel to the chassis, which can improve the signal quality and electromagnetic performance. -
Cover all empty slots using EMC filler panels or fill using slot blockers to maximize cooling air flow, depending on your application.
-
Power on the chassis.
PXIe-5163 Front Panel and Pinout
Front Panel
Table 2. Connectors
Signal | Connector Type | Description |
---|---|---|
CH 0 and CH 1 | BNC female | Analog input connection; digitizes data and |
triggers acquisitions.
CLK IN| SMB| Imports an external reference clock or sample clock to the
oscilloscope.
PFI 0| SMB| PFI line for digital trigger input/output, probe compensation.
AUX 0| MHDMR| Reference clock input, reference clock output, bidirectional
digital PFI, and 3.3 V power output.
Table 3. Access LED
Color | Indications |
---|---|
Off | The module is not yet functional or the module has detected a problem |
with a power rail.
Amber| The module is being accessed.
Green| The module is ready to be programmed.
Table 4. Active LED
Color | Indications |
---|---|
Off | Module is not armed, triggered, or acquiring a waveform. |
Amber | The module is armed and waiting for a Reference (Stop) Trigger. |
Green | The module has received a Reference (Stop) Trigger. Also indicates that |
the module is acquiring a waveform.
Red| The module has detected an error. The LED remains red until the error
condition is removed. The following scenarios are examples of potential
errors:
• Detected an unlocked condition on a previously locked PLL. A PLL that is unlocked while in reset does not show an error.
• Unable to detect the external sample clock.
• Detected an overheating or overpowering error.
Note Certain driver interactions may cause the Active LED to flash red. An error condition does not exist unless the Active LED remains red.
AUX 0 Connector Pinout
Table 5. AUX 0 Connector Pin Assignments
Pin | Signal | Signal Description |
---|---|---|
1 | GND | Ground reference for signals |
2 | CLK IN | Used to import an external reference clock |
3 | GND | Ground reference for signals |
4 | GND | Ground reference for signals |
5 | CLK OUT | Used to export the reference clock |
6 | GND | Ground reference for signals |
7 | GND | Ground reference for signals |
8 | AUX 0/PFI 0 | Bidirectional PFI line |
9 | AUX 0/PFI 1 | Bidirectional PFI line |
10 | GND | Ground reference for signals |
11 | AUX 0/PFI 2 | Bidirectional PFI line |
12 | AUX 0/PFI 3 | Bidirectional PFI line |
13 | GND | Ground reference for signals |
14 | AUX 0/PFI 4 | Bidirectional PFI line |
15 | AUX 0/PFI 5 | Bidirectional PFI line |
16 | AUX 0/PFI 6 | Bidirectional PFI line |
17 | AUX 0/PFI 7 | Bidirectional PFI line |
18 | +3.3 V | +3.3 V power (200 mA maximum) |
19 | GND | Ground reference for signals |
PXIe-5163 SCB-19 Pinout
You can use the SCB-19 connector block to connect digital signals to the AUX 0
connector on the PXIe-5163 front panel. Refer to the following figure and
table for information about the SCB-19 signals when connected to the AUX 0
front panel connector.
Table 6. SCB-19 Signal Descriptions
Pin | Signal | Signal Description |
---|---|---|
1 | PFI 0 | Bidirectional PFI line |
2 | PFI 1 | Bidirectional PFI line |
3 | PFI 2 | Bidirectional PFI line |
4 | PFI 3 | Bidirectional PFI line |
5 | NC | No connection |
6 | CLK IN | Used to import an external reference clock or sample clock |
7 | NC | No connection |
--- | --- | --- |
8 | CLK OUT | Used to export the reference clock |
9 | PFI 4 | Bidirectional PFI line |
10 | PFI 5 | Bidirectional PFI line |
11 | PFI 6 | Bidirectional PFI line |
12 | PFI 7 | Bidirectional PFI line |
13 | +3.3 V | +3.3 V power (200 mA maximum) |
14 to 26 | GND | Ground reference for signals |
PXIe-5163 Mini-HDMI Breakout Cable to 6 BNC Pinout
You can use the Mini-HDMI Breakout Cable to 6 BNC to connect digital signals
to the AUX 0 connector on the PXIe-5163 front panel. Refer to the following
figure and table for information about the Mini-HDMI Breakout Cable to 6 BNC
signals when connected to the AUX 0 front panel connector.
PFI 1You can use the Mini-HDMI Breakout Cable to 6 BNC to connect digital
signals to the AUX 0 connector on the PXIe-5163 front panel. Refer to the
following figure and table for information about the Mini-HDMI Breakout Cable
to 6 BNC signals when connected to the AUX 0 front panel connector.
Table 7. Mini-HDMI Breakout Cable to 6 BNC Signal Descriptions
Signal | Connector Type | Description |
---|---|---|
CLK IN |
BNC female
| Used to import an external reference clock
CLK OUT| Used to export the reference clock
PFI 0| Bidirectional PFI line
PFI 1| Bidirectional PFI line
PFI 2| Bidirectional PFI line
PFI 3| Bidirectional PFI line
Configuring the PXIe-5163 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which NI hardware products are in the system and
how they are configured. MAX is automatically installed with NI-SCOPE.
-
Launch MAX.
-
In the configuration tree, expand Devices and Interfaces to see the list of installed NI hardware.
Installed modules appear under the name of their associated chassis. -
Expand your Chassis tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note If you do not see your module listed, pressto refresh the list of installed modules. If the module is still not listed, power off the system, ensure the module is correctly installed, and restart. -
Record the identifier MAX assigns to the hardware. Use this identifier when programming the PXIe-5163.
-
Self-test the hardware by selecting the item in the configuration tree and clicking Self-Test in the MAX toolbar.
The MAX self-test performs a basic verification of hardware resources.
PXIe-5163 Self-Calibration
For the PXIe-5163, self-calibration yields the following benefits:
- Characterizes onboard sample clock phase adjustment circuitry
- Characterizes onboard intermodular synchronization measurement circuitry
- Corrects for DC gain and offset errors within the oscilloscope for all ranges and both input impedance paths based on comparison to a precision, high-stability internal voltage reference
- Calibrates offset error of interleaving ADCs
- Takes approximately 1 minute to complete
Running Self-Calibration
Running self-calibration enables the PXIe-5163 to yield full performance over
its operating temperature range and two-year external calibration cycle.
Before running self-calibration, warm up the PXIe-5163 for at least 15
minutes.
Note: Warm-up begins after the chassis and controller or PC is powered and NI-SCOPE is loaded and recognizes the PXIe-5163.
PXIe-5163 modules are externally calibrated at the factory, but for optimal performance, use self-calibration in any of the following situations:
- After first installing the PXIe-5163 in your chassis
- After any module in the chassis is installed, uninstalled, or moved
- When the system is in an environment where the ambient temperature varies or the module temperature has drifted more than ±5 °C from the previous self-calibration
- To periodically adjust for small performance drifts that occur with product aging
- When 90 days have elapsed since the previous self-calibration
Tip You can use the nodes on the NI-SCOPE External Calibration palette to programmatically return the date and module temperature of the previous self- calibration. You can also view this information by selecting the device in MAX.
Self-calibrate the PXIe-5163 programmatically in NI-SCOPE or on demand in MAX.
Option Description
NI-SCOPE: Use the Self Calibrate node to self-calibrate the module.
MAX: Select the device in the Devices and Interfaces menu and click the
Self-Calibrate button.
When the two-year external calibration interval expires, an external calibration is required to ensure performance that is within specification over the subsequent two years.
First Measurements
Making a Measurement with Instrument Studio
- Connect CH 0 to an input signal.
- Launch Instrument Studio at Start» National Instruments.
- In the instrument header menu in the upper-right corner of the panel ( ), add the PXIe-5163 to the large panel with Add/Remove Devices.
- Click Auto to automatically configure device settings for the detected signal.
- If the soft front panel is not already running, click Run/Stop.
- Add oscilloscope measurements to the channel by selecting Add/Remove in the measurement table of the oscilloscope.
For more information on the measurements available for oscilloscopes in InstrumentStudio, refer to the InstrumentStudio Manual at ni.com/manuals.
Making a Measurement with LabVIEW
-
Launch LabVIEW.
-
Select Help»Find Examples.
-
Open the example VI that you want to use by selecting Hardware Input and Output» Modular Instruments»NI-SCOPE (High-Speed Digitizers).
Tip If you are not sure which example to run, use the Quick Start VI, which is found under Hardware Input and Output»Modular Instruments»NI-SCOPE (High-Speed Digitizers)»Demos»niScope EX Quick Start.vi. -
Follow any setup instructions in the VI and specify any desired settings.
-
Click Run to run the example program.
PXIe-5163 Compensating Passive Probes
Compensating passive probes increases the accuracy of your measurements by
matching the capacitance of your probe to the capacitance of a particular
oscilloscope input channel.
Before beginning this procedure, complete the following:
- Set up your oscilloscope in a compatible chassis;
- Install a version of NI-SCOPE compatible with your oscilloscope on your system; and
- Configure a panel and layout in InstrumentStudio to include your oscilloscope.
Note: For more information on using InstrumentStudio, refer to the InstrumentStudio Manual at ni.com/manuals.
The following equipment is required for this procedure:
If using the Oscilloscope Probe Compensator|
- Oscilloscope Probe Compensator (NI part number 786983-01)
---|---
If connecting the probe directly to the oscilloscope PFI|
- A 50 Ω SMB (f)-to-BNC (f) cable, such as the SMB-100 (NI part number 781449-01)
- A probe tip–to-BNC (m) adapter
The PXIe-5163 can output a 1 kHz square wave that you can use to compensate
passive probes.
To compensate a passive probe, complete the following steps:
-
Connect the BNC end of the probe to an input channel of your oscilloscope.
-
Depending on your connection method, connect the probe tip to the oscilloscope:
- Oscilloscope Probe Compensator
- Connect the SMB end of the Oscilloscope Probe Compensator to the PFI connector of the oscilloscope that generates the probe compensation signal.
- Connect the ground clip of the probe to the ground terminal.
- Contact the tip of the probe to the square wave terminal.
- Direct connection to PFI
- Connect the SMB end of the SMB (f)-to-BNC (f) cable to the PFI connector of the oscilloscope that generates the probe compensation signal.
- Attach the BNC adapter to the tip of the probe.
- Connect the probe tip via the BNC adapter to the BNC end of the cable.
-
Enable the probe compensation signal:
Option Description
InstrumentStudio: In the instrument header menu ( ) in the upper-right corner of the panel, select Probe Compensation » Enabled.
NI-SCOPE: Call Probe Compensation Signal Start. -
In InstrumentStudio, configure the settings of the input channel to which you connected the probe.
- Set the channel to On to display the signal.
- Set the Input impedance of the channel to 1 MΩ.
- Match the Probe attenuation setting to the setting on your probe.
-
Adjust the vertical range on the input channel until the signal starts to clip and then increase the vertical range by one step so that it no longer clips.
This process ensures you are using the maximum dynamic range of the ADC. -
Examine the digitized signal and adjust the tunable capacitor on the probe.
The probe is correctly compensated when the waveform appears as square as possible:
Once compensated, the probe conveys signals to the input channel accurately, without artificially attenuating or amplifying frequency components of the signal.
Tip: For the most accurate measurements:
- Compensate probes for each channel of the oscilloscope;
- Use a compensated probe only with the channel you used to compensate it; and
- Compensate your probes frequently.
Programming the PXIe-5163
You can acquire data interactively using InstrumentStudio, or you can use
the NI-SCOPE instrument driver to program your device in the supported ADE of
your choice.
Table 8. PXIe-5163 Programming Options
Application | Location | Description |
---|---|---|
InstrumentStudio | Note InstrumentStudio is supported only on 64-bit |
systems. If you are using a 32-bit system, use your driver’s specific soft front panel instead of InstrumentStudio.
InstrumentStudio is automatically installed when you install the
NI-SCOPE driver. You can access InstrumentStudio in one of the following ways:
• From the Windows start menu, select National Instruments » [Driver] Soft Front Panel. This launches InstrumentStudio and runs a soft front panel populated with NI-SCOPE devices.
• From the Windows Start menu, select National Instruments » InstrumentStudio [year]. This launches InstrumentStudio and runs a soft front panel populated with devices detected on your system.
• From Measurement & Automation Explorer (MAX), select a device and then click Test Panels…. This
launches InstrumentStudio and runs a soft front panel for the device you selected.
| When you install NI-SCOPE on a 64-bit system, you can monitor, control, and record measurements from supported devices using InstrumentStudio.
InstrumentStudio is a software-based front panel application that allows you to perform interactive measurements on several different device types in a single program.
Application | Location | Description |
---|
NI-SCOPE
Instrument Driver
| LabVIEW—Available on the LabVIEW Functions palette at Measurement I/O » NI-SCOPE.| The NI-SCOPE API
configures and operates the device hardware and provides customizable acquisition, control, analysis, and measurement options using LabVIEW VIs or LabWindows/CVI functions.
C or LabWindows/CVI—Available at Program Files » IVI Foundation »
IVI » Drivers » niScope.
Microsoft Visual C/C++, .NET—Use examples located in the file directory
Users\Public\Public Documents\National Instruments\NI–SCOPE \examples.| You
can modify an NI-SCOPE C example to create an application with Microsoft
Visual C/C++, or modify
a .NET example to create an application with Visual Basic .NET or Visual C#.
Copy an NI-SCOPE example to copy required project settings for include paths and library files. Alternatively, refer to the Creating an Application section of the NI High-Speed Digitizers Help to manually add all required include and library files to your project.
NI-SCOPE Examples
Examples demonstrate the functionality of the device and serve as programming
models and building blocks for your own applications.
The NI Example Finder is a utility available for some ADEs that organizes
examples into categories and allows you to easily browse and search installed
examples. You can see descriptions and compatible hardware models for each
example, or see all the examples compatible with one particular hardware
model.
To locate examples, refer to the following table:
Application Development Environment (ADE) | Locating Examples |
---|---|
LabVIEW or LabWindows/CVI | Locate examples with the NI Example Finder. Within |
LabVIEW or LabWindows/CVI, select Help » Find Examples , and navigate
to Hardware Input and Output » Modular Instruments.
Visual C/C++ or Visual Basic| Locate examples in the file directory
Users\Public
\Public Documents\National Instruments\NI– SCOPE\examples.
Visual C# or Visual Basic .NET
Troubleshooting
If an issue persists after you complete a troubleshooting procedure, contact NI technical support or visit ni.com/support.
Why Is the ACCESS LED Off When the Chassis Is On?
The LEDs may not light until the module has been configured in MAX. Before
proceeding, verify that the PXIe-5163 appears in MAX.
If the ACCESS LED fails to light after you power on the chassis, a problem may
exist with the chassis power rails, a hardware module, or the LED.
Notice Apply external signals only while the PXIe-5163 is powered on. Applying external signals while the module is powered off may cause damage.
-
Disconnect any signals from the module front panels.
-
Power off the chassis.
-
Remove the module from the chassis and inspect it for damage. Do not reinstall a damaged module.
-
Install the module in a different chassis slot from which you removed it.
-
Power on the chassis.
Note: If you are using a PC with a device for PXI remote control system, power on the chassis before powering on the computer. -
Verify that the module appears in MAX.
-
Reset the module in MAX and perform a self-test.
What Should I Do if the PXIe-5163 Doesn’t Appear in MAX?
-
In the MAX configuration tree, expand Devices and Interfaces.
-
Expand the Chassis tree to see the list of installed hardware, and press
to refresh the list. -
If the module is still not listed, power off the system, ensure that all hardware is correctly installed, and restart the system.
-
Navigate to the Device Manager.
Operating System Description
Windows 10/8.1: Right-click the Start button, and select Device Manager.
Windows 7: Select Start »Control Panel » Device Manager. -
Verify the PXIe-5163 appears in the Device Manager.
-
Under an NI entry, confirm that a PXIe-5163 entry appears.
Note If you are using a PC with a device for PXI remote control system, under System Devices, also confirm that no error conditions appear for the PCI-to- PCI Bridge. -
If error conditions appear, reinstall NI-SCOPE and the PXIe-5163.
-
What Should I Do if the PXIe-5163 Fails the Self-Test?
- Restart the system.
- Launch MAX, and perform the self-test again.
- Power off the chassis.
- Reinstall the failed module in a different slot.
- Power on the chassis.
- Perform the self-test again.
Where To Go Next
Located in the hardware kit
EXPLORE the application development environment (ADE) for your application.
Learn LabVIEW Basics
Getting Started with LabWindows/CVI
Located online atni.com/manuals
LEARN about hardware features or review device specifications.
NI PXIe-4136 Specifications OR
NI PXIe-4137 Specifications
NI DC Power Supplies and SMUs Help*
Located using the NI Example Finder
CREATE custom applications within an application programming interface (API).
NI-DCPower Soft Front Panel
NI-DCPower Instrument Driver
NI DCPower Examples
NI DC Power Supplies and SMUs Help
DISCOVER
more about your products through ni.com.
Support
ni.com/support
Power Supplies
Solutions
ni.com/powersupplies
Services
ni.com/services
NI Community
ni.com/community
Worldwide Support and Services
The National Instruments website is your complete resource for technical
support. At ni.com/support, you have access to
everything from troubleshooting and application development self-help
resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation
Services, repairs, extended warranty, and other services.
Visit ni.com/register to register your National
Instruments product. Product registration facilitates technical support and
ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council
of the European Communities using the manufacturer’s declaration of
conformity. This system affords the user protection for electromagnetic
compatibility (EMC) and product safety. You can obtain the DoC for your
product by visiting ni.com/certification. If
your product supports calibration, you can obtain the calibration certificate
for your product at ni.com/calibration.
National Instruments corporate headquarters is located at 11500 North Mopac
Expressway, Austin, Texas, 78759-3504. National Instruments also has offices
located around the world. For telephone support in the United States, create
your service request at ni.com/support or dial 1 866
ASK MYNI (275 6964). For telephone support outside the United States, visit
the Worldwide Offices section of ni.com/niglobal to
access the branch office websites, which provide up-to-date contact
information, support phone numbers, email addresses, and current events.
Refer to the NI Trademarks and Logo Guidelines at
ni.com/trademarks for information on National
Instruments trademarks. Other product and company names mentioned herein are
trademarks or trade names of their respective companies. For patents covering
National Instruments products/technology, refer to the appropriate location:
Help» Patents in your software, the patents.txt file on your media, or the
National Instruments Patent Notice at ni.com/patents.
You can find information about end-user license agreements (EULAs) and third-
party legal notices in the readme file for your NI product. Refer to the
Export Compliance Information at ni.com/legal/export-
compliance for the National
Instruments global trade compliance policy and how to obtain relevant HTS
codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED
WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL
NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in
this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR
252.227-7014, and DFAR 252.227-7015.
© 2015 National Instruments. All rights reserved.
374874C-01 Sep15
References
- Test and Measurement Systems, a part of Emerson - NI
- NI Community - National Instruments
- NI Driver Downloads - NI
- Product Documentation - NI
- Contact Us - NI
- Oscilloscopes and Digitizers - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- NI Trademarks and Logo Guidelines - NI
- Test and Measurement Systems, a part of Emerson - NI
- NI Driver Downloads - NI
- Product Documentation - NI
- Contact Us - NI
- Log In - National Instruments
- NI Services - NI
- Support - NI
- CC3050X NI Oscilloscope Probe | Apex Waves
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>