ATEC CM-3CDSRG-32 3-Phase Coupler or Decoupler User Manual

June 12, 2024
ATEC

CM-3CDSRG-32 3-Phase Coupler or Decoupler

Product Information

The Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler is
a device designed for immunity testing. It is manufactured by
Thermo Fisher Scientific Corporation, a leading provider of
scientific and laboratory equipment. The CM-3CDSRG-32 is designed
to comply with safety standard regulations and comes with various
safety features to ensure safe operation.

The device is equipped with ventilation openings to ensure
proper airflow and prevent overheating. It should be situated away
from heat sources and not installed in cabinets or other enclosed
spaces that may impede ventilation.

The CM-3CDSRG-32 is powered by a specific type of power source,
as described in the operating instructions or marked on the
instrument. It is important to connect the device only to the
recommended power source to avoid any electrical issues.

For proper grounding, the device comes with a grounded power
cord. It is essential to ensure that the grounding is not defeated
during operation to maintain safety.

The instrument should be cleaned as recommended by the
manufacturer to maintain its performance and longevity.

Product Usage Instructions

Before operating the Thermo Scientific CM-3CDSRG-32 3-Phase
Coupler/Decoupler, it is crucial to carefully read and understand
all safety and operating instructions provided in the user
manual.

  1. Ensure that the instrument is placed on a recommended cart or
    stand provided by the manufacturer or included as part of the CE
    Master system.

  2. Avoid mounting the instrument on walls or ceilings.

  3. Situate the instrument in a location that allows proper
    ventilation and does not obstruct airflow through the ventilation
    openings.

  4. Avoid placing the instrument near water or any other sources of
    moisture.

  5. Connect the instrument to the appropriate power source as
    described in the operating instructions or marked on the
    device.

  6. Make sure the instrument’s power cord is properly grounded and
    not likely to be pinched or walked on.

  7. Clean the instrument using the recommended cleaning methods
    provided by the manufacturer.

Following these usage instructions will help ensure safe and
optimal operation of the Thermo Scientific CM-3CDSRG-32 3-Phase
Coupler/Decoupler during immunity testing.

Advanced Test Equipment Corp.
www.atecorp.com 800-404-ATEC (2832)
CM-3CDSRG-32
Operations Manual
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

CM-3CDSRG-32
Operations Manual
MA-95-532-005-00 REV A
Copyright 2011 Thermo Fisher Scientific Corporation All Rights Reserved
Thermo Fisher Scientific Corporation 200 Research Drive
Wilmington, MA 01887-4442 USA Tel: 978-275-0800 Fax: 978-275-0850
http://www.thermo.com/esd
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

The copyright protection afforded to Thermo Fisher Scientific Corporation includes all forms of media material and information including, without limitation, the software, generated icons and logos etc. and all associated documentation. This operating manual and associated software contained within may not be duplicated in any way without the express permission of Thermo Fisher Scientific Corporation. Printable products of the software such as test reports and graphs, and associated files of the same, are the property of the purchaser and may be subject to their own copyright restrictions. The software described in this manual may be copied by, strictly for the use of, the purchaser in accordance with the furnished license agreement. TM Trademark Credits IBM is a registered trademark of International Business Machines Corporation. Microsoft Windows and MS-DOS are registered trademarks of Microsoft Corporation. Notice In no event will Thermo Fisher Scientific Corporation be liable for any damages, consequential, incidental or indirect (including loss of business profits, business interruption, loss of business information, and the like) arising out of the use or inability to use the hardware or software supplied. The author and publisher of the operating manual make no warranty of any kind, expressed or implied, with regard to the instructions and suggestions contained within. The author and publisher of this operating manual shall not be liable in any event of incidental or consequential damages in connection with, or arising out of, the furnishing, undertaking or use while following verbal or written instructions and suggestions in any interpretation of its contents. Thermo Fisher Scientific Corporation’s products undergo continuous development and the company reserves the right to change their specifications without obligation or prior notification.

Amendment History

Issue A

Date

Author

10/26/2011 RRA

Amendment Initial Release

Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

TABLE OF CONTENTS

PRODUCT SAFETY

1

IMMUNITY TESTING GUIDELINES

3

SAFETY CONCERNS DURING IMMUNITY TESTING

4

INAPPROPRIATE USES

5

DESCRIPTION

6

Assembly Instructions

7

Cabling and Interconnections

9

Electrical Supply Requirements

11

Use of the CM-3CDSRG-32 with Ground-Fault Interrupters (GFI)

12

Physical Environment

13

Operating limits

13

Storage limits

13

Test Area Considerations and Site Preparation

14

Special Considerations for Immunity Testing

14

Handling, Transportation, Storage

15

CM-3CDSRG-32 BASICS

16

PROGRAMMING CM-3CDSRG-32

19

MAINTENANCE

22

Service and Maintenance Issues

22

Factory Inspection and Refurbishment

22

Calibration

22

Decommission Issues

22

TROUBLESHOOTING

23

Troubleshooting Guide

23

Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Product Safety
The following safety instructions have been included in compliance with safety standard regulations. Please read them carefully.
Read Instructions — Read all safety and operating instructions before operating the instrument.
Retain Instructions — Retain all safety and operating instructions for future reference.
Heed Warnings – Adhere to all warnings on the instrument and in the operating instructions.
Follow Instructions — Follow all operating and use instructions.
Water and Moisture — Do not use the instrument near water.
Carts and Stands — Use the instrument only with a cart or stand that is recommended by the manufacturer or included as part of the CE Master system.
Wall or Ceiling Mounting — Do not mount the instrument on a wall or ceiling.
Ventilation — The instrument should be situated so that its location or position does not interfere with its proper ventilation. Do not install in a cabinet or other situation that may impede the flow of air through the ventilation openings.
Heat — The instrument should be situated away from heat sources such as heat registers or other instruments which produce heat.
Power Sources — Connect the instrument only to the type of power source described in the operating instructions or as marked on the instrument.
Grounding or Polarization — Take precautions to insure that the grounding of the instrument is not defeated. Operate only with grounded power cord.
Power Cord Protection — Place power supply cords so that they are not likely to be walked on or pinched by items placed on them or against them. Pay particular attention to cords at plugs, convenience receptacles, and the point where they enter and exit the instrument.
Cleaning — Clean the instrument only as recommended by the manufacturer.
Non-Use Periods — Unplug the power cord of the instrument when it will be unused for a long period of time.
Lifting and Carrying — When moving or installing a CM-3CDSRG-32 follow the instructions given in the Installation section of this manual; The CM- 3CDSRG-32 requires two people for moving.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler 1

Object and Liquid Entry — Take care that objects do not fall and that liquids are not spilled into the enclosure through openings.
Defects and Abnormal Stress — Whenever it is likely that the normal operation has been impaired, make the equipment inoperable and secure it against further operation. Normal operation is likely to be impaired if, for example, the instrument:
Shows visible damage. Fails to perform the intended function. Has been subject to prolonged storage under unfavorable conditions. Has been subjected to severe transport stresses.
Damage Requiring Service — The instrument should be serviced by qualified service personnel when:
The power supply cord or the plug has been damaged. Objects have fallen or liquid has been spilled into the instrument. The instrument has been exposed to rain. The instrument does not appear to operate normally or exhibits a marked change in performance. The instrument has been dropped, or the enclosure has been damaged.
Sitting or Climbing — Do not sit or climb upon the instrument or use it as a step or ladder.
Adapters, cables, and accessories — Use only adapters, cables, and accessories which have been approved for use with this unit.
Use only MultiContact Safety Plugs for connections to the front panel of the CM-3CDSRG-32 ­ use of any other connections may result in risk of electric shock.
EUT load may restart automatically The thermal protection within the CM- 3CDSRG-32 is self-resetting. In the event of an Overload, as indicated by the chassis-mount Overload indicator, switch the EUT power switch off or remove EUT Mains input and wait for Overload to clear.
The CM-3CDSRG-32 3-Phase Coupler/Decoupler is designed to operate with the Thermo Scientific EMCPRO series. Use of the CM-3CDSRG-32 with any other equipment may result in risk of electric shock, damage to equipment, and risk of fire.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Immunity Testing Guidelines
DO NOT WORK ALONE.
Do not use the equipment in conditions other than reasonable laboratory conditions. There should be no condensing humidity or water standing on the floor or work surfaces; there shouldn’t be significant dust or other contamination.
Ensure that NO ONE is touching the equipment under test (EUT) during the test or immediately after the test until AC power to the EUT has been turned off.
Ensure that there is a barrier to act as protection in case the equipment under test explodes. This may happen due to power-follow1 after a failure. The barrier should be interlocked to prevent surging and to disconnect all AC if the barrier is removed.
The equipment under test must be surrounded by sufficient insulating material to withstand twice the surge voltage. Consider distance to the floor or table and walls if air is the insulating material.
Ensure that the proper supply mains voltages are applied to both Thermo Scientific equipment and to the equipment under test, and that the AC branch circuit is capable of supplying the current.
The ground (protective earth), neutral and phase lines of the AC supply to the equipment under test supply must be connected properly. Do not defeat the protective earth connection.
The ground (protective earth), neutral and phase lines of the AC supply to the Thermo Scientific equipment must be connected properly. Do not defeat the protective earth connection.
When surging a powered EUT, the mains supply to the equipment under test must be capable of handling the potential AC fault current (e.g. do not use a UPS to power the EUT).
Use only equipment which is designed to be safe for the test being performed.
Do not test in a potentially explosive atmosphere (e.g. gas fumes).
Never use equipment that is operating in a strange manner, or that shows clear indication of abuse.
If probes are in use, be sure they are differential probes which have no ground connection to the surged ground or to the equipment under test.

1 Power-follow is a condition where the Surge event causes a low-impedance path which full mains current may then flow through. One example of this is a gas-tube arrestor, which maintains a high impedance until a surge event switches it to a low-impedance state; the low-impedance state is maintained until the mains voltage drops below a critical threshold. Similar effects can be seen due to electrical arcs or exploded components.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

3

Safety Concerns During Immunity Testing
Immunity testing is hazardous. The equipment under test (EUT) can ignite, possibly explosively. Noxious, toxic and sometimes fatal fumes can be generated by the burning equipment. Accumulated gases may ignite explosively (i.e., flashover). In an environment where Immunity testing takes place, it is absolutely crucial that these minimum safety precautions be taken:
Immunity testing should be performed only by properly trained test personnel, who are experienced in conducting such test, or be observed and supervised by such experienced personnel. No person subject to heart or neurological conditions should be allowed to conduct surge tests. Persons with pacemakers should not be allowed in or near the area where testing is conducted.
Never leave a procedure or a test setup unattended. All personnel working in the area must be shielded with appropriate eye protection, body
protection and electrical protection. They should not be allowed to work in a direct line of a possible explosion of the equipment under test. The test area should be a clear and unobstructed environment dedicated to such tests. The test area should be equipped with ventilating hoods and blowers to remove gases that may be caused by exploding or burning components. The test area should have nonflammable walls and floors plus shielding to contain exploding parts and flames. There must be fire extinguishers certified for use in electrical and chemical fires readily available at the test site. DO NOT USE WATER TO EXTINGUISH AN ELECTRICAL FIRE. All flammable materials and debris must be outside the test area, and the area must be well marked, preferably by physical barriers, to prevent accidental intervention by nontest personnel while a test is in progress.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Inappropriate Uses
The unit must be kept within the environmental limits of the operating requirements. This includes not using the unit in condensing humidity. There are no User Serviceable parts inside; do not remove covers. Use only adapters and cables and accessories which have been approved for use with this unit. Connect only to Thermo Scientific equipment which has been approved for use with this unit. DO NOT connect to any other test equipment. DO NOT modify front panel jumpers for multiple line Surge coupling. Read and follow directions for connecting equipment.
Operating the CM-3CDSRG-32 with any equipment other than those recommended may damage equipment, create a shock hazard, and create a fire risk!

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

5

Description
The CM-3CDSRG-32 3-Phase Coupler/Decoupler is a single-enclosure, floor- standing or desktop test system designed to couple SURGE events to a 3-phase AC Mains. Single line coupling for Surge per IEC standards with EMCPRO PLUS EMC Test System. A Coupler/Decoupler is required to apply surge to the AC Mains of the Equipment Under Test. The Coupler/Decoupler connects the output of a surge generator to the EUT AC Mains [Coupler] and minimize surge or surge energy which may couple to the AC Mains of the building [Decoupler]. The CM- 3CDSRG-32 is a manual tester, in which the SURGE operation is programmed from the front panel via programming jumpers. Programming is fast and simple. Operation is semiautomatic when used with the EMCPro series test equipment, a Personal Computer, and the CEWare software, where the CEWare software can provide pauses and prompt the operator to move jumpers for changing coupling modes.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Installation
Assembly Instructions
The CM-3CDSRG-32 requires two people for unpacking. Check the packaging materials and follow any special unpacking instructions shipped with your unit. The CM-3CDSRG-32 comes fully assembled. Save all shipping materials (shipping boxes, foam pieces, bags, and special instructions) for possible future shipments. The original shipping materials must be used whenever the CM-3CDSRG-32 is returned for service.
Voltage Selection
The CM-3CDSRG-32 must be set for the mains voltage at your site. This is usually done before shipment, but must be checked before installation and use. On the rear of the, CM-3CDSRG-32 check that the voltage select block is set to the proper mains voltage. If not correct, remove line cord, pry the voltage select block from the inlet assembly using a small screw driver; rotate the block to the proper voltage, as shown to the right, and fully insert in holder.
Use AC Mains Line Cord part number 02-701-190-00 CONN,AC CORD,10A 125V NEMA 18G BLK 7’6″ for CM-3CDSRG-32 AC input connection (provided):

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

7

Carefully check all shipping materials for accessories which will be packaged with the CM-3CDSRG-32 including a standard AC input cord:
Part#: 01-301-995-00 (SURGE CABLE) QTY=1 Description: CABLE, 1 Meter, 13AWG, 2 conductors
PART#: 02-701-851-00 (SYNCH CABLE) QTY=1 Description: CONN, AC CORD, IEC-320-C19 EA END, 250V VDE
Part#: 01-301-795-00 (COUPLING MODE JUMPER) QTY=2 Description: BLACK JUMPER PLUG ELBOW, 4mm 32A 6kV
Part#: 95-332-904-00 (3-Phase EUT MAINS INPUT CABLE) QTY=1 Description: ASSY, AC EUT CORD, 32A, IEC-309
Part# 02-709-166-00 (3-Phase INPUT CABLE RECEPTICALE) QTY=1 Description: CONN P&S, RECPT, 32A 288/500V, 5W, BLACK, VDE

Part # 01-301-915-00 QTY=10 (EUT output pins) SAFETY PLUG,4mm,32A

Part: 01-301-880-00 QTY=8 (EUT output pin cover) BLUE, OUTER COVER

Part:01-301-905-00 QTY=2 (EUT output pin cover) YELLOW, OUTER COVER

Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Cabling and Interconnections
The CM-3CDSRG-32 can be placed alongside a Surge generator, placed on top of a Surge generator, or placed under a Surge generator.

The CM-3CDSRG-32 has an AC Mains line cord which plugs into the rear of the unit, via a standard IEC-320 connector. This provides fan and control power to the CM-3CDSRG-32 , and is required for operation. The EUT Power input is a pin and sleeve connector, rated for the current and voltage of the Equipment Under Test. A power cord is provided to connect the CM-3CDSRG-32 to the building AC mains.

A matching pin and sleeve connector (CONN P&S, RECPT, 32A 288/500V, 5W, BLACK, VDE) for connection to the building mains is provided, and installation of this connector may require the services of an electrician.

Connection to the Equipment Under Test — the EUT Load Connection to the Equipment Under Test (EUT) is via five safety sockets on the lower right of the CM-3CDSRG-32 front panel. Five, unassembled safety socket connectors are provided with the CM-3CDSRG-32. Additional connectors are available from MultiContact.

The cable to the Equipment Under Test must be built to the customer requirements. This wiring should be done to meet all applicable, local, safety requirements. In general — but not exclusively — these requirements include the following:

The wiring to the EUT must be rated to withstand the applied mains voltage.
The wiring to the EUT should be rated to withstand twice the applied pulse voltage, or the wiring should be isolated from the operator and ground so that insulation breakdown does not create a hazard.
The wiring to the EUT should be rated for the expected fault current of the CM-3CDSRG-32 (32 amperes) or additional current limiting and protection should be added.
WHERE CONNECTION TO PE IS REQUIRED FOR SAFETY, A PE CONNECTION MUST BE PROVIDED.
It is considered good practice for the PE wired connection to be made longer than any of the other connections; in the event of an accidental disconnection, the PE terminal is disconnected last.
In addition, some loads may require proper phasing for proper operation. The user must determine the phase relationships of the building AC mains and check this relationship at the CM-3CDSRG-32 outputs.
Caution
Construction and assembly of cord, and connection of EUT loads should be done only by trained personnel with knowledge of electrical safety practices, and practical knowledge of the electrical and safety requirements of the load.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

9

Connecting the CM-3CDSRG-32 to EMCPro
The CM-3CDSRG-32 accessories kit includes a line synchronization cable PART# 02-701-85100 This cable connects the synch output of the CM-3CDSRG-32 to the EUT Mains Input connector on the rear of the EMCPro generator.
The synch output carries the AC line voltage. When this connection to an EMCPro is in place, the tester’s 250V mains limitation applies to the CM- 3CDSRG-32 .
The synch output of the CM-3CDSRG-32 is not intended for general use as a utility outlet — do not use this outlet for other devices.
Alternatively, this cable can be replaced with a standard IEC cord, connecting the EMCPro to a single-phase AC Mains. In situations where a neutral is not available, this can be a simple way to establish line synchronization.
The surge cable Part# 01-301-995-00 which connects the CM-3CDSRG-32 surge input to the EMCPro H Coil I/O Line Surge Output connections. DO NOT CONNECT to the EMCPro Surge Output To Telecom Lines connectors — this output is not rated for AC mains.
These connections are color-coded; match the white connector on the CM- 3CDSRG-32 to the white connector on the surge generator.
The surge cable set should be disconnected from both the CM-3CDSRG-32 and the surge generator when not performing surge tests.
Direct Current Power to EUT The CM-3CDSRG-32 can accept a single-phase AC, or Direct Current (DC) power source at the EUT input and provide it to the EUT output. When connecting a single-phase or DC source it is important to note the assignment of EUT inlet and outlet connectors.
The CM-3CDSRG-32 is internally protected at 32 amperes. When used with a single-phase AC, or Direct Current (DC) power source, if the EUT load or cabling requires current limiting protection of less than the CM-3CDSRG-32 rating, it is the responsibility of the user to provide that protection.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Electrical Supply Requirements
The CM-3CDSRG-32 system power is configured for one of two voltage ranges:
100 – 120 Vac, 50/60 Hz 220 – 240 Vac, 50/60 Hz
The voltage range configured is identified by label on the rear panel of the CM-3CDSRG32 . The voltage range may be easily changed with a screwdriver, following the directions of the Installation, Voltage Selection section of this manual.
The CM-3CDSRG-32 is rated for a EUT, AC power source of:
CM-3CDSRG-32 250/433 Vac, 32 ampere, 3-phase, 5-wire, 50/60 Hz (U.S.) DC: 30Adc up to 24Vdc, 25A 24 to 48V, 8A up to 100V
CM-3CDSRG-32 is restricted to 250/433 VAC when the sync cable is connected to an EMCPro.
The EUT, AC power source should be fused or protected for a maximum of 100 amperes.
The overcurrent protection of the CM-3CDSRG-32 includes detection of imbalanced phasecurrents. When using the CM-3CDSRG-32 with a one-phase load, a two phase load, or a load with extreme imbalance between phase currents, the overcurrent protection will trip at a load current of 85% of the CM-3CDSRG-32 rating.
If testing with mains synchronization is required — the surge pulses synchronized with the AC Mains — then a neutral reference connection must be provided at the EUT inlet connector. If neutral is not available, then a neutral reference must be created via a delta-towye transformer or load, and the neutral reference should be externally fused or protected for the rating of transformer and wiring.
The synch cable between the CM-3CDSRG-32 and surge generator can be replaced with a standard IEC cordset, connecting the surge generator to a single-phase AC Mains. In situations where a neutral is not available, this can be a simple way to establish a line synchronization reference. The user must then verify by experiment the synchronization relationship between the single phase and the three phase lines.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

11

The EUT power source must have adequate capacity for full power-follow2 in the event that testing causes an equipment failure. Using a supply without sufficient capacity may result in a failure being masked.
Acceptable power sources include: A dedicated branch circuit The dedicated output of a motor-generator (MG) set. The output of a properly rated and installed isolation transformer.
Use of the CM-3CDSRG-32 with Ground-Fault Interrupters (GFI)
Problems may result when using the CM-3CDSRG-32 on AC mains protected by GroundFault Interrupters (GFIs).
The problem is due to a clamp circuit, protecting the building Mains from excessive voltage during surge, and from the coupling network mandated by IEC 61000-4-5 which connects the surge generator to the AC mains during the surge pulse, and creates a short pulse of mains current in PE, tripping the GFI.
There are two solutions to this problem: The first solution is to power the EUT from a source that does not have a GFI device
installed. This solution may not be practical in some locations due to building codes or safety regulations.
The second solution is to install an isolation transformer between the AC mains and CM3CDSRG-32 . The isolation transformer should be rated for the current of the EUT load.
Note A solid-state regulated line source such as an uninterruptable power supply (UPS) or a power frequency amplifier will generally not have the instantaneous capacity or full power-follow required for Surge and PQF testing, and therefore should not be used as the AC power source for any
equipment being tested.
2 Power-follow is a condition where the Surge event causes a low-impedance path through which fault level mains current may then flow: possibly hundreds of amperes or greater. One example of this is a gas-tube arrestor, which maintains a high impedance until a surge event switches it to a low-impedance state; the low-impedance state is maintained until the mains voltage drops below a critical threshold. Similar effects can be seen due to electrical arcs or exploded components.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Physical Environment

The CM-3CDSRG-32 systems are intended for operation in a laboratory environment, protected from excess dust, humidity, and temperature.
No condensing humidity or standing water on the floor or work surfaces. No significant dust or other contamination.

Operating limits
Temperature: Humidity: Altitude:

15 – 40° C 10 – 85%, non-condensing 10,000 feet max.

Storage limits
Temperature: Humidity: Altitude:

0 – 60° C 10 – 90%, non-condensing 10,000 feet max.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

13

Test Area Considerations and Site Preparation
Conduct Immunity testing under sound laboratory conditions. Verify the following:
· No condensing humidity or standing water on the floor or work surfaces · No significant dust or other contaminants · Clear and unobstructed vision · Adequate ventilation, including ventilating hoods and blowers to remove gases · Nonflammable walls and floors · Barriers surrounding the EUT to contain exploding parts and flames · Appropriate fire extinguishers for electrical and chemical fires. DO NOT USE WATER
TO EXTINGUISH AN ELECTRICAL FIRE. · No flammable materials or debris inside the test area. The area must be well marked,
preferably by physical barriers · Keep unauthorized personnel out of the area during testing
Allow at least six inches clearance from the wall or other equipment on both sides of the instrument for proper air flow. The system must be mounted on a level surface. EMCPRO and CM-3CDSRG-32 may be placed on a table that can support the weight of the units. They may also be placed on a sturdy cart. However, since the unit is heavy, make certain that the instrument and cart combination is stable, and cannot be easily tipped particularly if the wheels strike a bump on the floor such as the edge of a rug, a pebble or a small piece of hardware.
Special Considerations for Immunity Testing
Immunity testing is best done in an area cleared of obstructions. Clearly mark the boundaries of the area. There is the possibility that the surge may flash over to circuits or metal objects not directly under test, and that components in the equipment being tested may explode or ignite under the stress of the surge test. Whenever possible, enclose the equipment under test within a fireproof and explosion-proof barrier having insulation capable of withstanding at least twice the maximum surge voltage. Never allow direct line-of-site view of components that may explode or ignite. If visual observation is required, use a robust transparent barrier of suitable thickness for protection. Immunity testing should only be carried out by fully trained personnel who are informed of the hazards of such testing, and who have full control over all of the test equipment in the area.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Handling, Transportation, Storage
Save all shipping materials (shipping boxes, foam pieces, bags, and special instructions) for possible future shipments.
The CM-3CDSRG-32 requires two people for unpacking. Follow the instructions in the INSTALLATION section of this manual and any special instructions shipped with the unit.
Issues related to transportation of CM-3CDSRG-32 models are identical to those for installation; the weight of the unit requires that two people help in packing and movement. When packing the unit for transportation the original shipping materials should be used, or new materials of equivalent strength and durability.
For long storage intervals, unplug the power cords of the instrument and cover the instrument to protect from dust and liquid spills.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

15

CM-3CDSRG-32 Basics
A Coupler/Decoupler is required to apply surge to the AC Mains of the Equipment Under Test. The Coupler/Decoupler connects the output of a surge generator to the EUT AC Mains [Coupler] and minimize surge energy which may couple to the AC Mains of the building [Decoupler]. The circuitry to perform this is very simple. The generator is connected to the EUT output. Filter inductors and capacitors in each line attenuate surge energy and reduce the amount of energy escaping into the building AC Mains. The complications in this circuitry are in the details: constructing filter inductors which can block the volt-second product of a 6000 volt, 1.2x50uS pulse, yet which have very low coupling capacitance to avoid passing SURGE energy. They must also safely couple the surge generator to a high-voltage AC Mains.
Basic De-Coupler Network
For surges, the CM-3CDSRG-32 is designed to operate with a Thermo Scientific EMCPro series surge generator. The coupling network required to connect the surge to the AC Mains is built into the EMCPro.
Operating the CM-3CDSRG-32 with any equipment other than those recommended may damage equipment, create a shock hazard, and create a fire risk!
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

The CM-3CDSRG-32 is programmed from the front panel, and all connections to the Equipment Under Test (EUT) are made from the front panel. The front panel controls and features are broken into three major groups: the Surge generator to CM-3CDSRG-32 Interconnections, the Equipment Under Test Connections, and the User Interface (Programming).

EMCPro to CM-3CDSRG-32 Interconnections
The Surge cable Part# 01-301-995-00 connects the CM-3CDSRG-32 surge input to the EMCPro : H Coil / Surge Output connections.
DO NOT CONNECT to the EMCPro Surge Output To Telecom Lines connectors — this output is not rated for AC mains.
These connections are color-coded; match the white connector on the CM- 3CDSRG-32 to the white connector on the surge generator.
AC line sync surges, you must install line synchronization cable PART# 02-701-851-00 This cable connects the synch output of the CM-3CDSRG-32 to the EUT Mains Input connector on the rear of the EMCPro generator.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

17

Equipment Under Test Connections
Connection to the Equipment Under Test (EUT) is via five safety sockets on the lower right of the CM-3CDSRG-32 front panel. Five, unassembled safety socket connectors are provided. Additional connectors are available from MultiContact.
The cable to the Equipment Under Test must be built to the customer requirements. This wiring should be done to meet all applicable, local, safety requirements. In general — but not exclusively — these requirements include the following:
The wiring to the EUT must be rated to withstand the applied mains voltage. The wiring to the EUT should be rated to withstand twice the applied pulse
voltage, or the wiring should be isolated from the operator and ground so that insulation breakdown is not a problem. The wiring to the EUT should be rated for the expected fault current of the CM3CDSRG-32 (32 amperes) or additional current limiting and protection should be added. WHERE CONNECTION TO PE IS REQUIRED FOR SAFETY, A PE CONNECTION MUST BE PROVIDED.
User Interface (Programming)
Feedback to the operator is provided by three indicators on the front of the unit:
SYSTEM POWER is lit when the unit is powered — the rear power switch is set to ON and the system power cord connected. This indicator must be lit for the unit to operate.
EUT POWER is lit when the EUT power switch on the front panel is set to ON — the EUT is energized.
OVERLOAD is lit when there is an over temperature or over current fault and the EUT power switch on the front panel is set to ON. The unit will not operate while this indicator is lit.
Warning! EUT load may restart automatically!
The thermal protection within the CM-3CDSRG-32 is self-resetting. In the event of an Overload, as indicated by the chassis-mount Overload indicator, switch the EUT power switch off or remove EUT Mains input and wait for Overload to clear.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Programming CM-3CDSRG-32
To program surges, first connect the CM-3CDSRG-32 and Surge generator using the surge cable set provided with the CM-3CDSRG-32 .
Connect SURGE HI to SURGE HI, and SURGE LO to SURGE LO. Use Surge cable Part# 01-301-995-00 supplied with the CM-3CDSRG-32 — do not substitute. The surge cable set should be disconnected from both the CM-3CDSRG-32 and the Surge generator when not performing tests.
The CM-3CDSRG-32 is manually programmed for surge mode by moving the SURGE HI and SURGE LO jumpers on the front panel. There are two jumpers required for any surge event; one jumper is placed in the SURGE HI central socket and bridges to one of the outer sockets; one jumper is placed in the SURGE LO central socket and bridges to one of the outer sockets.

As an example, to program SURGE HI to L1 and SURGE LO to PE, place the jumpers as shown below.

SURGE HI should be set to one line to be surged; SURGE LO should be set to the other (SURGE HI and SURGE LO should always be set to different lines; if set to the same line the surge generator will be shorted, and no surge voltage will be delivered. This is a safe condition -the surge generator is designed to withstand shorting — but it does not provide a valid test.).

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

19

When surging a line, the polarity of SURGE HI with respect to SURGE LO is always as set at the Surge generator – if the Surge generator is set for a positive surge voltage, SURGE HI will be positive with respect to SURGE LO. Other than polarity, SURGE HI and SURGE LO may be interchanged freely.

There is minimal attenuation of surge waveforms through the CM-3CDSRG-32 ; the voltage programmed at the Surge generator should be delivered to the EUT outlets without attenuation.

The CM-3CDSRG-32 supports single-line to single-line surges.

In all there are 20 coupling modes, of which some are redundant. Eliminating the redundant coupling modes leaves 10 different coupling modes. These are:

Surge from

To

L1

L2

L1

L3

L1

N

L1

PE

L2

L3

L2

N

L2

PE

L3

N

L3

PE

N

PE

For surging to the CM-3CDSRG-32 the EMCPRO must be programmed with Output: Coupling . . . DIRECT. This sets the surge output to the safety sockets. To synchronize to the AC Mains, set the Phase Synch field to L1, and turn the EMCPRO EUT switch ON (If synchronization is not required, set the Phase Synch field to RND, and the EMCPRO EUT switch may be ON or OFF). See the EMCPRO manual for more information on programming options.

¦ SURGE: USER DEF ¦¦¦¦¦¦¦¦¦¦¦¦¦¦EUT OFF¦

Waveform ………. 12 Ohm

Voltage ……….. +2000 V

Output:Coupling … DIRECT

Phase Sync:Angle .. L1:180 deg

Number of Tests … 500

Time Between Tests

60 s

RUN ¦¦ ¦¦ ¦¦ ¦¦ ¦¦ ¦¦EXIT

Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

For performing AC line sync surges, you must install line synchronization cable PART# 02-701-851-00 This cable connects the synch output of the CM- 3CDSRG-32 to the EUT Mains Input connector on the rear of the EMCPro generator.
Most of the EMCPRO pre-programmed tests can also be used with the CM-3CDSRG-32 . To use a pre-programmed test the test must offer the option under coupling of Output: Coupling: DIRECT. See the EMCPRO manual for more information on pre-programmed tests.
Pre-programmed tests may use line synchronization; set the Phase Synch field to L1, and turn the EMCPRO EUT switch ON. When testing is performed with mains synchronization, a neutral reference must be provided at the EUT inlet connector. If neutral is not available, then:
a neutral reference must be created via a delta-to-wye transformer or load, and the neutral reference must be externally fused or protected for the rating of transformer and wiring, or;
The synch cable between the CM-3CDSRG-32 and EMCPRO can be replaced with a standard IEC cordset, and connected to a single-phase AC Mains. The user must then verify by experiment the synchronization relationship between the single phase and the three phase lines.
When using line synchronization, the EMCPRO will always synchronize to the L1 connection of the CM-3CDSRG-32. To synchronize to another phase, add 120° or 240° to the desired value and enter Phase Sync:Angle: L1: field of the surge generator. The phase synchronization should be checked with an oscilloscope by programming a low-voltage surge — within the voltage rating of the oscilloscope probes used — and measuring the phase reference.
For all pre-programmed tests, remember that you are only testing the single coupling mode that you have programmed at the CM-3CDSRG-32 front panel. In many cases, for a complete test, you will have to change coupling mode and continue until all modes have been tested.
The CM-3CDSRG-32 is a manual unit; it is programmed by moving jumpers and on the front panel. Since the EMCPro is an automatic unit some care must be taken on interfacing the two units and in programming. The CEWare software provides pauses and prompt features which can prompt the operator with pre-programmed messages and instructions for setting the CM3CDSRG-32 jumpers — allowing semi- automatic testing.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

21

Maintenance
Service and Maintenance Issues
There are no user serviceable parts within the CM-3CDSRG-32 . Service is solely by Thermo Fisher Scientific trained, authorized technicians. The system power connector contains a fuse block, holding two, 5mm, 0.5A 250V timedelay fuses. These may be replaced by removing the mains cordset from the power connector and prying the cover off the fuse holder with a small screw driver.
Factory Inspection and Refurbishment
Factory Inspection and Refurbishment is available. Consult factory for terms and conditions.
Calibration
Calibration of the CM-3CDSRG-32 is not required.
Decommission Issues
There are no hazardous materials within the CM-3CDSRG-32 and no special issues in decommissioning.
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Troubleshooting
This section aids in finding and solving any problems you may experience.
Troubleshooting Guide
In the event of a problem with the CM-3CDSRG-32 please follow the following procedure:
Verify that you have properly selected all test parameters on the EMCPRO. Check that the CM-3CDSRG-32 and EMCPRO are properly connected to the AC
Mains. Check that the EUT is properly connected to the CM-3CDSRG-32 . Check all interconnections for loose connections. Try cycling power to the EUT on the EMCPRO, and CM-3CDSRG-32. Check the trouble-shooting guide( next page) for your symptoms. Contact Customer Service for guidance.

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

23

CM-3CDSRG-32 Troubleshooting Guide

The EUT cannot be powered The OVERLOAD indicator is lit

Check that: The SYSTEM POWER indicator is lit. The SYSTEM POWER switch is ON (rear) The System power line cord is inserted and
connected to the building AC mains The OVERLOAD indicator is not lit The EUT POWER switch is ON The EUT POWER indicator is lit Either there was an over- current condition on the EUT load, or an internal overtemperature condition within the unit. Both conditions are self-resetting.

Turn the EUT POWER switch to off. Check the EUT load for possible shorts. Check that the units fans are operating. Wait 2 minutes and retry the EUT POWER switch.

The SYSTEM POWER indicator is NOT lit.

WARNING! If the EUT POWER switch is ON, the EUT may start unexpectedly when the overtemperature condition resets. Check that: The SYSTEM POWER switch is ON (rear) The System power line cord is inserted and
connected to the building AC mains The voltage configuration block in the
System power inlet receptacle is set to the proper Mains voltage. The fuses in the System power inlet receptacle are good.

Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

Filing a Problem Report When reporting a problem to Thermo Fisher Scientific Customer Service, please include the serial number and model number from the rear panel of the CM-3CDSRG-32.
Mail, Fax, or phone your problem report to:
Customer Service Thermo Fisher Scientific Corporation,
200 Research Drive Wilmington, MA 01887-4442 USA
Tel: 978-275-0800 Fax: 978-275-0850 http://www.thermo.com/esd

Thermo Fisher Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

25

WEEE Compliance:
This product is required to comply with the European Union’s Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following symbol:
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State, and this product should be disposed of or recycled through them. Further information on Thermo Fisher Scientific’s compliance with these Directives, the recyclers in your country, and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at www.thermo.com/WEEERoHS
Thermo Scientific CM-3CDSRG-32 3-Phase Coupler/Decoupler

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals