SALUKI TECHNOLOGY SCKCL03-N Mechanical Calibration Kit Instructions
- June 12, 2024
- SALUKI TECHNOLOGY
Table of Contents
SALUKI SCKCL03-N Mechanical Calibration Kit
Product Information
- The SCKCL03-N High Precision Calibration Kit is designed for calibrating network analyzers with an N-type interface. It is capable of operating in the frequency range of DC to 3GHz, providing accurate calibration and measurement results.
- This calibration kit includes precision standard devices that ensure reliable and repeatable connections. It consists of various components such as open, short, and load terminations, as well as adapters to change the sex of the test port.
- The key features of the SCKCL03-N High Precision Calibration Kit are:
- Frequency Range: DC to 3GHz
- VSWR: 500 times cycles
Product Usage Instructions
To use the SCKCL03-N High Precision Calibration Kit, follow these steps:
- Ensure that your network analyzer is powered off and disconnected from any power source.
- Identify the required calibration component based on your measurement needs:
- Open Termination: Connect the “Open: N (M)” or “Open: N (F)” termination to the test port of your network analyzer. This termination provides an open circuit for calibration.
- Short Termination: Connect the “Short: N (M)” or “Short: N (F)” termination to the test port of your network analyzer. This termination provides a short circuit for calibration.
- Load Termination: Connect the “Load: N (M)” termination to the test port of your network analyzer. This termination provides a calibrated load for calibration.
- If needed, use the included torque wrench to ensure a secure and reliable connection between the calibration component and the test port.
- Once the calibration component is connected, power on your network analyzer and follow its calibration procedure. Refer to your network analyzer’s user manual for specific instructions on calibration.
- After calibration is complete, disconnect the calibration component from the test port and store it in a safe place for future use.
By following these instructions, you can effectively calibrate your network analyzer using the SCKCL03-N High Precision Calibration Kit.
Key Features
- Frequency Range: DC to 3GHz
- VSWR : <1.014
- Calibration standards to perform full two-port calibration
- Adapters and torque wrench included
-
500 times Cycles
Overview
Saluki’s SCKCL03-N mechanical calibration kit contains precision standard devices to calibrate network analyzers in the N-type interface. This kit also contains adapters to change the sex of the test port, and a torque wrench to ensure a repeatable reliable connection.
Model
|
Frequency
|
Spec.
|
List
|
qty.
---|---|---|---|---
| | | Open: N (M)| 1
| | | Open: N (F)| 1
| | | Short: N (M)| 1
SCKCL03-N
(Calibration Level)
|
DC – 3GHz
| Open: ±0.85°
Short: ±0.65°
Load: VSWR<1.014
Adaptor: VSWR<1.029
| Short: N (F)| 1
Load: N (M)| 1
Load: N (F)| 1
N(f)-N(f) adaptor| 1
| | | N(m)-N(m) adaptor| 1
| | | N(m)-N(f) adaptor| 1
| | | N-type Torque wrench| 1
Material Information
Basic Data
Interface| IEEE Std 287
Mating Cycles| >500 cycles
Coupling Torque| 0.9Nm
Open-End Wrench Size| 19mm
Electrical Data
Frequency| DC – 3GHz
Impedance| 50Ω
Average Power| ≤0.5W
Materials
Center Conductor| Au-plated Beryllim Copper
Body| Copper Plating / Staineless Stell, Passivated
Connector Nut| Stainless Steel, Passivated
Note: Information will conduct the necessary updates,the contents of this document are subject to change without notice