INVACARE RPL450-2 Battery Powered Patient Lift
Product Information
- Product Name: Battery-Powered
Patient Lift Model Numbers: RPL450-2, RPL600-2
Product Usage Instructions
Assembling the Patient Lift:
- Unpack the patient lift.
Assemble the mast to the base.
Assemble the boom actuator.
Install the leg actuator to the base.
Mount the battery charger.
Install the shifter handle.
Operating the Patient Lift:
- Use the hand control buttons to operate the patient lift.
Activate a mechanical emergency release if needed.
Perform an emergency stop if necessary.
Mount the battery charger and charge the battery.
Patient Lifting:
- Prepare the lift for use by positioning it properly.
Attach slings to the lift.
Lift or move the patient using the patient lift.
Patient Transferring:
- Transfer the patient to a commode chair or standard commode using the patient lift.
Transferring to a commode chair:
Follow the instructions for transferring to a commode chair.
Transferring to a standard commode:
Follow the instructions for transferring to a standard commode.
Troubleshooting:
Refer to the troubleshooting section in the manual for assistance in resolving
any issues that may arise.
Maintenance:
- Perform regular maintenance checks using the maintenance safety inspection checklist.
- Lubricate the lift as instructed.
- Detect wear and damage and take appropriate action.
- Clean the sling and lift regularly.
- Replace components such as the boom actuator, hanger bar, casters, and forks as needed.
Accessories:
- The Reliant Scale RLS6 is an optional accessory.
- Follow the instructions for removing the hanger bar and installing the Reliant Scale.
- Operate the scale using the keypad functions.
- Weigh the patient as needed.
- Replace the battery and calibrate the Reliant Scale if necessary.
- Refer to the troubleshooting section for assistance with any issues related to the Reliant Scale. Please note that this is a summary of the product information and usage instructions. For complete and detailed instructions, please refer to the user manual provided with the product.
Invacare® ReliantTM 450/600
RPL450-2, RPL600-2
en-us
Battery-Powered Patient Lift
User Manual
This manual MUST be given to the user of the product. BEFORE using this product, this manual MUST be read and saved for future reference. PROVIDER: Keep this manual. The procedures in this manual MUST be performed by a qualified technician.
© 2022 Invacare Corporation All rights reserved. Republication, duplication or
modification in whole or in part is prohibited without prior written
permission from Invacare. Trademarks are identified by TM and ®. All
trademarks are owned by or licensed to Invacare Corporation or its
subsidiaries unless otherwise noted. Invacare reserves the right to alter
product specifications without further notice. Making Life’s Experiences
Possible is a registered trademark in the U.S.A. Tous droits réservés. La
réédition, la reproduction ou la modification du présent document en tout ou
en partie est interdite sans l’autorisation écrite d’Invacare. Les marques de
commerce sont indiquées par les symboles TM et ®. Toutes les marques de
commerce sont la propriété d’Invacare Corporation ou de ses filiales, ou
leur ont été accordées sous licence, à moins d’indication contraire. Invacare
se réserve le droit de modifier les spécifications du produit sans préavis.
Making Life’s Experiences Possible est une marque déposée américaine. Todos
los derechos reservados. Queda prohibido volver a publicar, copiar o modificar
el presente documento, en parte o por completo, sin el previo consentimiento
por escrito de Invacare. Las marcas comerciales se identifican con TM y ®.
Todas las marcas comerciales son propiedad de Invacare Corporation o de sus
filiales, o bien Invacare Corporation o
sus filiales cuentan con las correspondientes licencias, a menos que se
estipule lo contrario. Invacare se reserva el derecho de modificar las
especificaciones del producto sin previo aviso. Making Life’s Experiences
Possible es una marca comercial registrada en EE. UU.
Contents
1 General 1.1 Symbols 1.1.1 Warnings 1.1.2 Other Symbols 1.2 Service Life 1.3
Limited Warranty
2 Safety 2.1 General Guidelines 2.2 Operating Information 2.2.1 General 2.2.2
Pinch Points and Positioning 2.2.3 Lift Assembly 2.2.4 Lift Operation 2.2.5
Using the Sling 2.2.6 Lifting the Patient 2.2.7 Transferring the Patient 2.2.8
Maintenance 2.2.9 Electrical and Grounding 2.2.10 Disposal 2.3 Radio Frequency
Interference
3 Product Labeling
4 Technical Data 4.1 Patient Lift 4.2 Full-Body, Divided-Leg, and Toileting
Slings 4.3 Reliant Scale RLS6
5 Assembly
20
5.1 Assembling the Patient Lift
20
5.1.1 Unpacking the Patient Lift
20
5 5 5 5 6
5.1.2 Assembling the Mast to the Base 5.1.3 Assembling the Boom Actuator 5.1.4 Installing the Leg Actuator to the Base 5.1.5 Mounting the Battery Charger 5.2 Installing the Shifter Handle
21 22 24 25 27
6 6 Operation
28
8 8 9 9 10
6.1 Operating the Patient Lift 6.1.1 Using the Hand Control Buttons 6.1.2
Activating a Mechanical Emergency Release 6.1.3 Performing an Emergency Stop
6.2 Mounting the Battery Charger Charging the Battery
28 28 30 31 32
10 7 Patient Lifting
34
10
7.1 Preparing the Lift for Use
34
11
7.1.1 Positioning the Lift for Use
34
12
7.1.2 Attaching Slings to the Lift
35
13
7.2 Lifting/Moving the Patient
36
13 8 Patient Transferring
39
14
8.1 Transferring the Patient for Using a Commode Chair or
14
Standard Commode
39
15
8.1.1 Transferring to a Commode Chair
41
16
8.1.2 Transferring to a Standard Commode
42
17 17
8.2 Transferring the Patient to a Bathing Unit 8.3 Transferring the Patient to a Wheelchair
43 43
18 9 Troubleshooting
46
19
9.1 Troubleshooting
46
10 Maintenance
49
10.1 Maintenance Safety Inspection Checklist
49
10.2 Lubricating the Lift
52
10.3 Detecting Wear and Damage
52
10.4 Cleaning the Sling and Lift
52
10.5 Replacing the Boom Actuator
53
10.5.1 Checking and Tightening the Mast Pivot Bolt
54
10.6 Replacing the Hanger Bar
55
10.7 Maintaining the Base Adjustment
57
10.8 Replacing the Casters and Forks
58
10.8.1 Replacing the Rear Casters
58
10.8.2 Replacing the Front Casters
59
10.8.3 Replacing the Forks
59
11 Accessories
60
11.1 Reliant Scale RLS6
60
11.1.1 Removing the Hanger Bar
60
11.1.2 Installing the Reliant Scale
62
11.2 Operating the Scale
64
11.2.1 Keypad Functions
64
11.3 Weighing the Patient
65
11.4 Replacing the Battery
68
11.5 Calibrating the Reliant Scale
68
11.6 Troubleshooting
70
11.6.1 Display Codes
70
General
1 General
1.1 Symbols
1.1.1 Warnings Signal words are used in this manual and apply to hazards or
unsafe practices which could result in personal injury or property damage. See
the information below for definitions of the signal words.
DANGER! -Danger indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING! -Warning indicates a potentially hazardous situation which, if not
avoided, could result in death or serious injury.
ATTENTION! -Caution indicates a potentially hazardous situation which, if not
avoided, may result in property damage or minor injury or both.
-Indicates a hazardous situation that could result in damage to property if it is not avoided.
Gives useful tips, recommendations and information for efficient, trouble-free use.
Other Symbols
Date of fabrication
1145810-E
5
Invacare® ReliantTM 450/600
1.2 Service Life
Liftthe expected service life is eight years, presuming the product is used
daily and in accordance with safety instructions and maintenance instructions
stated in this manual. Slingrefer to the Sling User Manual for expected
service life information.
1.3 Limited Warranty
THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO
PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser/user of our products.
This warranty gives you specific legal rights and you may also have other
legal rights which vary from state to state. Invacare warrants the products
manufactured to be free from defects in materials and workmanship for a period
of three years on the lift and one year on the slings, hydraulic pump /
electric components from the date of purchase. If within such warranty period
any such product shall be proven to be defective, such product shall be
repaired or replaced, at Invacare’s option. This warranty does not include any
labor or shipping charges incurred in replacement part installation or repair
of any such product. Invacare’s sole obligation and your exclusive remedy
under this warranty shall be limited to such repair and/or replacement. For
warranty service, please contact the dealer from whom you purchased your
Invacare product. In the event you do not receive satisfactory warranty
service, please write directly to Invacare at the address on the back cover,
provide dealer’s name, address, date of purchase, indicate nature of the
defect. Invacare Corporation will issue a serialized return authorization. The
defective unit or parts MUST be returned for warranty inspection using the
serial number, when applicable as identification within 30 days of return
authorization date. Do not return products to our factory without our prior
consent. C.O.D. shipments will be refused; please prepay shipping charges.
6
1145810-E
1 General
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL
NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED, PRODUCTS
SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE,
PRODUCTS MODIFIED WITHOUT INVACARE’S EXPRESS WRITTEN CONSENT (INCLUDING, BUT
NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR
ATTACHMENTS; PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT
WITHOUT THE SPECIFIC CONSENT OF INVACARE, OR TO A PRODUCT DAMAGED BY
CIRCUMSTANCES BEYOND INVACARE’S CONTROL, AND SUCH EVALUATION WILL BE SOLELY
DETERMINED BY INVACARE. THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM
NORMAL WEAR OR FAILURE TO ADHERE TO THE INSTRUCTIONS IN THIS MANUAL. THE
FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER EXPRESS WARRANTIES.
IMPLIED WARRANTIES, IF ANY, INCLUDING THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, SHALL NOT EXTEND BEYOND
THE DURATION OF THE EXPRESSED WARRANTY PROVIDED HEREIN AND THE REMEDY FOR
VIOLATIONS OF ANY IMPLIED WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT
OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN. INVACARE
SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW EXCLUSION OR LIMITATION OF INCIDENTAL OR
CONSEQUENTIAL DAMAGE, OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY LASTS, SO
THE ABOVE EXCLUSIONS AND LIMITATIONS MAY NOT APPLY TO YOU. THIS WARRANTY SHALL
BE EXTENDED TO COMPLY WITH STATE OR PROVINCIAL LAWS AND REQUIREMENTS.
1145810-E
7
2 Safety
2 Safety
The Safety section contains important information for the safe operation and
use of this product.
2.1 General Guidelines
WARNING! DO NOT use this product or any available optional equipment without
first completely reading and understanding these instructions and any
additional instructional material such as user manuals, service manuals or
instruction sheets supplied with this product or optional equipment. If you
are unable to understand the warnings, cautions or instructions, contact a
healthcare professional, provider or technical personnel before attempting to
use this equipment; otherwise, injury or damage may occur.
ACCESSORIES WARNING Invacare products are specifically designed and
manufactured for use in conjunction with Invacare accessories. Accessories
designed by other manufacturers have not been tested by Invacare and are not
recommended for use with Invacare products.
THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT
NOTICE.
Check all parts for shipping damage before using. In case of damage, DO NOT
use the equipment. Contact the Dealer for further instructions.
1145810-E
8
Invacare® ReliantTM 450/600
2.2 Operating Information
2.2.1 General
WARNING! -The Invacare patient lift is NOT a transport device. It is intended
to transfer an individual from one resting surface to another (such as a bed
to a wheelchair). -DO NOT attempt any transfer without approval of the
patient’s physician, nurse, or medical assistant. Thoroughly read the
instructions in this user manual, observe a trained team of experts perform
the lifting procedures, and then perform the entire lift procedure several
times with proper supervision and a capable individual acting as a patient.
-Use common sense in all lifts. Special care MUST BE taken with people with
disabilities who cannot cooperate while being lifted. -Invacare slings and
patient lift accessories are specifically designed to be used in conjunction
with Invacare patient lifts. Slings and accessories designed by other
manufacturers are not to be utilized as a component of the Invacare patient
lift system. -If the patient lift is exposed to extreme temperature (above
100°F or below 32°F), high humidity, and/or becomes wet prior to use, ensure
the handgrips do not twist on the patient lift handle; otherwise, damage or
injury may occur.
9
1145810-E
2.2.2 Pinch Points and Positioning
WARNING! -Be aware of pinch points, such as between the boom and hanger bar
and at the base of the lift, to prevent pinching fingers. -When positioning
the lift, be aware of the position of the hanger bar and the patient. Injury
could occur.
2 Safety
Fig. 2-1
2.2.3 Lift Assembly
WARNING! -DO NOT overtighten the mounting hardware. This will damage the
mounting brackets.
2.2.4 Lift Operation
WARNING! -Although Invacare recommends that two assistants be used for all
lifting preparation and transferring-from and transferring-to procedures, our
equipment will permit proper operation by one assistant. The use of one
assistant is based on the evaluation of the healthcare professional for each
individual case. -Make sure there is an audible click when mounting the
battery on the battery charger to confirm proper mounting; otherwise, injury
or damage may occur. -DO NOT exceed the maximum weight limitation of the
patient lift. The weight limitation for the RPL450-2 is 450 lb; the weight
limitation for the RPL600-2 is 600 lb. -ALWAYS keep hands and fingers clear of
moving parts to avoid injury.
1145810-E
10
Invacare® ReliantTM 450/600
2.2.5 Using the Sling
WARNING! -Use an Invacare-approved sling recommended by the individual’s
doctor, nurse or medical assistant for the comfort and safety of the
individual being lifted. -DO NOT use any kind of plastic-back incontinence pad
or seating cushion between the patient and the sling material that may cause
the patient to slide out of the sling during transfer. -After each laundering
(in accordance with instructions on the sling), inspect the sling(s) for wear,
tears, and loose stitching. -Bleached, torn, cut, frayed, or broken slings are
unsafe and could result in injury. Discard them immediately. -DO NOT alter the
slings. -Be sure to check the sling attachments each time the sling is removed
and replaced to ensure it is properly attached before the patient is removed
from a stationary object (bed, chair, or commode). -If the patient is in a
wheelchair, secure the wheel locks in place to prevent the chair from moving
forward or backward. -When connecting slings equipped with color-coded straps
to the patient lift, the shortest of the straps MUST be at the back of the
patient for support. Using the long section will leave little or no support
for the patient’s back. The loops of the sling are color coded and can be used
to place the patient in various positions. The colors make it easy to connect
both sides of the sling equally. Make sure there is sufficient head support
when lifting a patient.
11
1145810-E
2 Safety
2.2.6 Lifting the Patient
WARNING! -When using an adjustable-base lift, the legs MUST be in the maximum
Opened/Locked position before lifting the patient. -When elevating a few
inches off the surface of the stationary object (wheelchair, commode, or bed)
and before moving the patient, check again to make sure the sling is properly
connected to the hooks of the hanger bar. If any attachments are not properly
in place, lower the patient back onto the stationary object (wheelchair,
commode, or bed) and correct the attachments. -Adjustments for safety and
comfort should be made before moving the patient. The patient’s arms should be
inside of the straps. -Invacare slings are made specifically for use with
Invacare patient lifts. For the safety of the patient, DO NOT intermix slings
and patient lifts of different manufacturers. The warranty will be voided.
-Check the sling attachments each time the sling is removed and replaced to
ensure it is properly attached before the patient is removed from a stationary
object (bed, chair or commode). -During transfer, with the patient suspended
in a sling attached to the lift, DO NOT roll the caster base over uneven
surfaces that could cause the patient lift to tip over. Use the steering
handle on the mast at all times to push or pull the patient lift. -Invacare
does not recommend locking the rear casters of the patient lift when lifting
an individual. Doing so could cause the lift to tip and endanger the patient
and assistants. Invacare does recommend leaving the rear casters unlocked
during lifting procedures to allow the patient lift to stabilize itself when
the patient is initially lifted from a chair, bed, or any stationary object.
1145810-E
12
Invacare® ReliantTM 450/600
2.2.7 Transferring the Patient
WARNING! To reduce the risk of injury or damage: -When elevated a few inches
off the surface of the stationary object (wheelchair, commode, or bed) and
before moving the patient, check again to make sure the sling is properly
connected to the hooks of the hanger bar. If any attachments are not properly
in place, lower the patient back onto the stationary object (wheelchair,
commode, or bed) and correct the attachments. -Wheelchair wheel locks MUST be
in a locked position before lowering the patient into the wheelchair for
transport. -Before transferring, check that the wheelchair weight capacity can
withstand the patient’s weight. -Be sure to check the sling attachments each
time the sling is removed and replaced to ensure it is properly attached
before the patient is removed from the bed or chair. -The mast pivot under the
rubber boot must be tight to ensure safe use of the patient lift. The bolt
must be checked at least every six months in conjunction with periodic
maintenance.
2.2.8 Maintenance
WARNING! -Maintenance MUST be performed ONLY by qualified personnel. -The pump
is sealed at the factory. DO NOT attempt to open the pump or obtain local
service as this will VOID the warranty and might result in damage. Consult
your dealer or contact Invacare for more information. -After the first year of
use, the hooks of the hanger bar and the mounting brackets of the boom should
be inspected every three months to determine the extent of wear. If these
parts become worn, they must be replaced. -Casters and axle bolts must be
inspected every six months to check for tightness and wear. After the first 12
months of operation, inspect the hanger bar and the eye of the boom to which
it attaches for wear. If the metal is worn, the parts MUST be replaced. Repeat
this inspection every six months thereafter. -Regular maintenance of patient
lifts and accessories is necessary to assure proper operation. -DO NOT
overtighten the mounting hardware. This will damage the mounting brackets.
-After the first six months of operation, inspect all pivot points and
fasteners for wear. If the metal is worn, the parts MUST be replaced. Repeat
this inspection every six months.
13
1145810-E
2 Safety
2.2.9 Electrical and Grounding
WARNING! GROUNDING INSTRUCTIONS -DO NOT, under any circumstances, cut or
remove the round grounding prong from any plug used with or for Invacare
products. Some devices are equipped with three-prong (grounding) plugs for
protection against possible shock hazards. When a two-prong wall receptacle is
encountered, it is the personal responsibility and obligation of the customer
to contact a qualified electrician and have the two-prong receptacle replaced
with a properly grounded three-prong wall receptacle in accordance with the
National Electrical Code. If you must use an extension cord, use ONLY a three-
wire extension cord having the same or higher electrical rating as the device
being connected. In addition, Invacare has placed RED/ORANGE warning tags on
some equipment. DO NOT remove these tags. -Carefully read the battery/battery
charger information prior to installing, servicing, or operating the patient
lift.
2.2.10 Disposal
WARNING! -This product has been supplied from an environmentally aware
manufacturer that complies with the Waste Electrical and Electronic Equipment
(WEEE) Directive 2002/96/CE. -The device contains lead-acid batteries. DO NOT
dispose of the batteries in normal household waste; they MUST be taken to a
proper disposal site. Contact your local waste management company for
information. -This product may contain substances that could be harmful to the
environment if disposed of in places (such as landfills) that are not
appropriate according to legislation. -Please be environmentally responsible
and recycle this product through your recycling facility at its end of life.
1145810-E
14
Invacare® ReliantTM 450/600
2.3 Radio Frequency Interference
WARNING! Risk of Injury or Damage -Most electronic equipment is influenced by
Radio Frequency Interference (RFI). CAUTION should be exercised with regard to
the use of portable communication equipment in the area around such equipment.
If RFI causes erratic behavior, PUSH the Red Power Switch OFF IMMEDIATELY. DO
NOT turn the Power Switch ON while transmission is in progress.
15
1145810-E
3 Product Labeling
3 Product Labeling
Fig. 3-1
1145810-E
16
Technical Data
4.1 Patient Lift
Height at Sling HookupMaximum Height at Sling HookupMinimum Base WidthOpen
Base WidthClosed Base Height (Clearance) Base Length Caster Size (Front/Rear)
Sling Options Weight Capacity Weight IN Carton Weight OUT of Carton Battery
Charger Input Charger Output/Charging Time Accessories
1145810-E
4 Technical Data
450 LB
600 LB
Low Profile RPL450-2
Low Profile RPL600-2
74 inches
68 inches
24 inches
28 inches
41.0 inches
26.5 inches
4.5 inches
48.0 inches
3.0/5.0 inches
3 styles
1 style
450 lb
600 lb
136 lb
141 lb
109 lb
115 lb
24V DC (RCHBL)
100-240V AC
29.5V DC maximum 6 hours
Digital Scale
17
Invacare® ReliantTM 450/600
Audio Low Battery Alarm Motor Safety Devices Approximate Lifts per Charge (varies depending upon load and stroke) Warranty Pump/Electronics
Yes Anti-Entrapment 100-200 cycles per charge
1 year
4.2 Full-Body, Divided-Leg, and Toileting Slings
Specification Size
Full Body
R110 R112*
M
R111 R113*
L
R114 XL
Full Body w/ Commode
Divided Leg
R115 M
R116 R117* R100P R100
L
XL
P
M
R101 L
Toileting
Heavy Duty w/o Heavy Duty w/
Commode
Commode
R102 R121
R140*
R141*
XL
L
N/A
N/A
Width
41.5 45.5 45.5 41.5 45.4 45.5
33
37.5
41
44.5
36
45
45
Commode Opening
N/A N/A N/A
8
8
9
N/A N/A N/A N/A
N/A
N/A
13
Length
54.7 60.5 65.3 54.7 60.5 65.3 59.8 62.8 67.8 72.3
37
55
55
Commode Opening
N/A N/A N/A
11
11
13
N/A N/A N/A N/A
N/A
N/A
13
Back
N/A N/A N/A N/A N/A N/A 29.2 35.7 41.7 43.2
N/A
N/A
N/A
Weight Capacity (lb)
450
450
450
450
450
450
450
450
450
450
450
600
600
*Slings are made of a mesh material. All dimensions are in inches except where noted.
18
1145810-E
4.3 Reliant Scale RLS6
Weight Range Resolution Display Automatic Power Down Size Weight Power Battery
Life Temperature Range
1145810-E
4 Technical Data
Up to 600 lb (272.7 kg) +/- 0.2 lb (0.1 kg) Liquid Crystal Digital Two (2)
minutes 5 3/4 in L X 3 5/8 in W X 4 in H (14.6 X 9.2 X 10.2 cm) .55 lb (.25
kg) Nine (9) volt alkaline battery (included) Approximately 1500 readings 50°
to 104°F (10° to 40°C)
19
Assembly
5 Assembly
5.1 Assembling the Patient Lift
WARNING! -Use only Invacare parts in the assembly of the patient lift. The
base legs, the mast, boom, pump assembly, and the hanger bar are manufactured
to specifications that assure correct alignment of all parts for safe,
functional operation.
5.1.1 Unpacking the Patient Lift
1. Unpack the components from the shipping carton.
DO NOT remove the plastic wrap that secures the boom to the mast in place. The
plastic wrap will be removed in 5.1.3 Assembling the Boom Actuator, page 22.
Fig. 5-1
1145810-E
20
Invacare® ReliantTM 450/600
5.1.2 Assembling the Mast to the Base
1. If locking-type casters are on the patient lift, lock them. 2. Remove the
shoulder bolt, nut, and washer that secure the mast in the U-shape
cutout of the base. 3. Position the mast in an upright position, and place the
mast into the U-shaped
cutout of the base. 4. Insert the shoulder bolt with washers through the base
and mast. 5. Secure with a nut.
Fig. 5-2
21
1145810-E
5 Assembly
Fig. 5-3
WARNING! The mast may be removed from the base for storage or transport. Each
time the mast is removed and returned to the base, the mast MUST be properly
secured to the base assembly.
5.1.3 Assembling the Boom Actuator
CAUTION! DO NOT overtighten the nut and bolt. This damages the mounting
bracket.
1145810-E
22
Invacare® ReliantTM 450/600
1. Remove the shoulder bolt, washer, and nut from the mounting bracket on the
boom assembly. Ensure the pre-installed bushings are in the actuator rod eye.
2. Unpack the pinch guard from the patient lift carton. The bottom of the
boom actuator assembly will already be assembled to the mast mounting bracket.
3. Cut the plastic wrap that secures the boom and mast together.
4. Lift up on the boom and place it on your left shoulder. 5. Let the boom
actuator rest on your shoulder and rotate
the shaft extension of the actuator assembly until it lines up with the
mounting holes in the boom assembly.
Fig. 5-4
6. Place the pinch guard over the shaft extension of the actuator.
7. Align the holes of the boom assembly mounting bracket with the actuator
holes, and insert the bolt. Secure with a nut. Ensure the bolt is completely
through the holes of the boom assembly mounting bracket and the actuator
assembly. The boom assembly will pivot easily if the mounting hardware is
aligned properly when the boom assembly is secured to the mast.
23
1145810-E
5.1.4 Installing the Leg Actuator to the Base 1. Slide the leg actuator into the slot in the base of the patient lift (Detail B).
5 Assembly
Fig. 5-5
2. To secure the leg actuator to the pivot bracket (Detail A):
WARNING! Ensure there is sufficient room to turn the patient lift on its side
and the floor area is clear of debris; otherwise, injury to personnel or
damage to the patient lift may occur.
1145810-E
24
Invacare® ReliantTM 450/600
Fig. 5-6 A. Turn the patient lift on its side. B. Position the slot in the leg
actuator over the pivot bracket. C. Install the pin through the leg actuator
and pivot bracket and secure with a hitch pin. D. Return the patient lift to
the upright position. 3. To secure the leg actuator to the mast bracket: A.
Position the leg actuator between the mast brackets. B. Move the legs to align
the holes in the leg actuator with the holes in the mast bracket. C. Install
the pin through the holes of the leg actuator and mast bracket and secure with
a hitch pin. 4. Plug the hand control (not shown) into the bottom of the
control box. 5.1.5 Mounting the Battery Charger
Refer to your local regulations concerning proper mounting procedures.
25
1145810-E
1. Place the battery charger with the mounting bracket on the wall at the desired position. 2. With a pencil, mark the middle hole position. 3. Measure 6 1/2 inches below the pencil mark, and drill one mounting hole. 4. Install the bottom mounting screw until there is an approximate 1/8-inch gap between the screw head and the wall. 5. Install the battery charger with mounting bracket onto the bottom mounting screw. 6. Drill the remaining two mounting holes. 7. Install the two remaining mounting screws through the mounting bracket and into the wall. Tighten securely. 8. Plug the battery charger into the wall electrical outlet. 9. Verify ON is illuminated.
5 Assembly
Fig. 5-7
1145810-E
26
Invacare® ReliantTM 450/600
5.2 Installing the Shifter Handle
1. Remove the shifter handle from the packaging carton. 2. Align the shifter
handle threads with the threaded opening in the base. 3. Turn the shifter
handle clockwise and securely tighten into the base.
Fig. 5-8
27
1145810-E
6 Operation
6 Operation
6.1 Operating the Patient Lift
WARNING! -DO NOT attempt any transfer without approval of the patient’s
physician, nurse, or medical assistant. Thoroughly read the instructions in
this Owner’s Manual, observe a trained team of experts performing the lifting
procedures, and then perform the entire lift procedure several times with
proper supervision and a capable individual acting as a patient. -The legs of
the lift must be in the maximum open position and the shifter handle locked in
place for optimum stability and safety. If it is necessary to close the legs
of the lift to maneuver the lift under a bed, close the legs of the lift only
as long as it takes to position the lift over the patient and lift the patient
off the surface of the bed. When the legs of the lift are no longer under the
bed, return the legs of the lift to the maximum open position and immediately
lock the shifter handle.
Although Invacare recommends that two assistants be used for all lifting
preparation and transferring from and transferring to procedures, the
equipment will permit proper operation by one assistant. The use of one
assistant is based on the evaluation of the healthcare professional for each
individual case.
6.1.1 Using the Hand Control Buttons
WARNING! DO NOT lock the rear casters of the patient lift when lifting an
individual. Locking the rear casters could cause the patient lift to tip and
endanger the patient and assistants.
The hand control is used to raise/lower the boom or to open/close the legs of
the base for stability when lifting a patient.
1145810-E
28
Invacare® ReliantTM 450/600
Raising/Lowering the Boom
To raise the patient lift, press the Boom Up button (Up arrow) to raise the
boom and the patient. To lower the patient lift, press the Boom Down button
(Down arrow) to lower the boom and the patient.
Opening/Closing the Legs
To open the legs, press the Legs Open button. To close the legs, press the
Legs Closed button.
Fig. 6-1
29
1145810-E
6.1.2 Activating a Mechanical Emergency Release
Primary Emergency Release To activate the primary emergency release, insert a
pen into the hole labeled “emergency” on the control box of the lift and push
down on the boom at the same time. All lift actuators are equipped with a
mechanical emergency release. The mechanical release enables the actuator to
retract without power. The actuator will retract only while under load and the
mechanical emergency release is pulled. The release is colored reddish orange
with the word “Emergency” spelled out in white.
Secondary Emergency Release It is recommended that the primary emergency
release be used first. The secondary emergency release is only a backup to the
primary emergency release.
In cases where the primary release is either not functioning or is
unreachable, a secondary emergency release may be used. To activate the
secondary release, pull up on the EMERGENCY grip and push down on the boom at
the same time.
1145810-E
Operation
Fig. 6-2 Fig. 6-3
30
Invacare® ReliantTM 450/600
6.1.3 Performing an Emergency Stop Press the RED/ORANGE emergency button on
the control box to stop the lift arms and the patient from raising or
lowering. Rotate the RED/ORANGE emergency stop button clockwise to disengage
the emergency stop.
Fig. 6-4
31
1145810-E
6.2 Mounting the Battery Charger Charging the Battery
Invacare recommends the battery be recharged daily to prolong battery life. An
audible alarm will sound (horn will beep) when the battery is low.
1. Lift up on the handle on the back of the battery. 2. Lift the battery up
and out, away from the control
box.
CAUTION! -Make sure there is an audible click when mounting the battery on the
battery charger to confirm proper mounting; otherwise, injury or damage may
occur.
3. Place the battery on the battery charger as shown. Make sure there is an
audible click. The charge LED will illuminate. When charging is complete, the
charge LED will stop illuminating. A full recharge will take approximately
four hours.
6 Operation
Fig. 6-5
1145810-E
32
Invacare® ReliantTM 450/600
4. Lift up on the handle on the back of the battery. 5. Lift the battery up
and out, away from the battery charger.
CAUTION! -Make sure there is an audible click when mounting the battery on the
battery charger to confirm proper mounting; otherwise, injury or damage may
occur. 6. Reinstall the battery onto the control box as shown. Make sure there
is an audible click.
The battery mounts to the control box and battery charger as shown.
33
1145810-E
Patient Lifting
7.1 Preparing the Lift for Use
Although Invacare recommends that two assistants be used for all lifting
preparation, transferring from and transferring to procedures, our equipment
will permit proper operation by one assistant. The use of one assistant is
based on the evaluation of the healthcare professional for each individual
case.
7.1.1 Positioning the Lift for Use Refer to 2 Safety, page 8 before proceeding
and observe all warnings indicated. Before positioning the legs of the patient
lift under a bed, make sure the area is clear of any obstructions.
WARNING! -The legs of the lift must be in the maximum open position for
optimum stability and safety. If it is necessary to close the legs of the lift
to maneuver the lift under a bed, close the legs of the lift only as long as
it takes to position the lift over the patient and lift the patient off the
surface of the bed. When the legs of the lift are no longer under the bed,
return the legs of the lift to the maximum open position and immediately lock
the shifter handle.
1. Press the Legs Open button on the hand control to open the legs of the
patient lift to maximum.
2. Position the patient lift using the steering handle.
3. Press the Boom Down button on the hand control to lower the boom for easy
attachment of the sling.
1145810-E
34
Invacare® ReliantTM 450/600
Fig. 7-1 7.1.2 Attaching Slings to the Lift 1. Place the straps of the sling
over the hooks on the hanger bar. 2. Match the corresponding colors on each
side of the sling for an even lift of the patient.
Model Nos. R110R117 Full Body Slings and Model No. R121 Toileting Sling have
four sling straps. Model Nos. R100R102 have six sling straps. Invacare Lift
Hanger Bars have three hookup points per side. The middle hookup is used ONLY
for slings that have three sets of straps per side.
35
1145810-E
3. Use the lift. Refer to 7.2 Lifting/Moving the Patient, page 36.
7 Patient Lifting
Fig. 7-2
7.2 Lifting/Moving the Patient
WARNING! -DO NOT lock the rear casters of the patient lift when lifting an
individual. Locking the rear casters could cause the patient lift to tip and
endanger the patient and assistants. -DO NOT move the patient if the sling is
not properly connected to the hooks of the hanger bar. When the sling is
elevated a few inches off of the stationary surface and before moving the
patient, check again to make sure the sling is properly connected to the hooks
of the hanger bar. If any attachments are NOT properly in place, lower the
patient back onto the stationary surface and correct this problem; otherwise,
injury or damage may occur. -Adjust for safety and comfort before moving the
patient. -DO NOT use slings and patient lifts from different manufacturers.
Invacare slings are made specifically for use with Invacare patient lifts.
Injury or damage may occur.
1145810-E
36
Invacare® ReliantTM 450/600
1. Press the boom UP ( ) button to raise the patient high enough to clear the
bed. The patient’s weight will be fully supported by the patient lift.
The boom will stay in position until the boom DOWN ( ) button is pressed.
When the patient is lifted from the bed (with the patient’s head supported by
the sling and/or an assistant), he/she will be raised to a sitting position
(Detail A).
2. Place the patient’s arms inside the sling. WARNING! -When the sling is
elevated a few inches off the surface of the bed and before moving the
patient, check again to ensure the sling is properly connected to the hooks of
the hanger bar. If any attachments are NOT properly in place, lower the
patient back onto the stationary surface and correct this problem; otherwise,
injury or damage may occur. -Adjustments for safety and comfort should be made
before moving the patient. The patient’s arms should be inside of the straps.
-Invacare slings are made specifically for use with Invacare Patient Lifts.
For the safety of the patient, DO NOT use slings and patient lifts from
different manufacturers.
3. When the patient is clear of the bed surface, swing his/her feet off the
bed (Detail B). 4. Using the steering handle, move the lift away from the bed.
5. When moving the patient lift away from the bed, turn the patient so he/she
faces the assistant operating
the patient lift (Detail C).
6. Press the DOWN ( ) button to lower the patient until his/her feet rest on
the base of the lift. The lower center of gravity provides stability, which
makes the patient feel more secure and the lift easier to move.
7. Pull the patient lift away from the bed, and then push it from behind with
both hands firmly on the steering handle.
37
1145810-E
7 Patient Lifting
Fig. 7-3
1145810-E
38
Invacare® ReliantTM 450/600
Patient Transferring
Transferring the Patient for Using a Commode Chair or Standard Commode
Slings with commode openings are designed to be used with either a commode
chair or standard commode. Invacare recommends that the sling remain connected
to the hanger bar hooks during the patient’s use of either the commode chair
or a standard commode. Although Invacare recommends that two assistants be
used for all lifting preparation, and transferring-from and transferring-to
procedures, our equipment will permit proper operation by one assistant. The
use of one assistant is based on the evaluation of the healthcare professional
for each individual case.
39
1145810-E
8 Patient Transferring
WARNING! -DO NOT attempt any transfer of a patient without approval of the
patient’s physician, nurse, or medical assistant. -DO NOT move the patient if
the sling is not properly connected to the hooks of the hanger bar. When the
sling is a few inches off the stationary surface and before moving the
patient, check to make sure the sling is properly connected to the hooks of
the hanger bar. If any attachments are NOT properly in place, lower the
patient back onto the stationary surface and correct this problem; otherwise,
injury or damage may occur. -Adjust for safety and comfort before moving the
patient. The patient’s arms should be inside the straps. -DO NOT use slings
and patient lifts from different manufacturers. Invacare slings are made
specifically for use with Invacare patient lifts. Otherwise, injury or damage
may occur. -Be sure to check the sling attachments each time the sling is
removed and replaced to ensure it is properly attached before the patient is
removed from a bed or chair. -DO NOT lock the rear casters of the patient lift
when lifting an individual. Doing so could cause the lift to tip and endanger
the patient and assistants. Invacare DOES recommend that the rear casters be
left unlocked during lifting procedures to allow the patient lift to stabilize
itself when the patient is initially lifted from a chair, bed, or any
stationary object. -Note the legs of the lift must be in the maximum open
position for optimum stability and safety. If it is necessary to close the
legs of the lift to maneuver the lift under a bed, close the legs of the lift
only as long as it takes to position the lift over the patient and lift the
patient off the surface of the bed. When the legs of the lift are no longer
under the bed, return the legs of the lift to the maximum open position.
1145810-E
40
Invacare® ReliantTM 450/600
8.1.1 Transferring to a Commode Chair 1. Before lifting the patient from the
bed, refer to 7.1.2 Attaching Slings to the Lift,
page 35. For operation of the patient lift, refer to 6 Operation, page 28. 2.
Lift the patient from the bed. Refer to 7.2 Lifting/Moving the Patient, page
36. 3. Press the Boom Up button to elevate the patient high enough to clear
the
commode chair arms and have his/her weight supported by the patient lift. 4.
With the help of both assistants, guide the patient onto the commode chair. 5.
Press the Boom Down button to lower the patient onto the commode chair, and
leave the sling attached to the hanger bar hooks. 6. When complete, recheck
for correct attachments. 7. Press the Boom Up button to raise the patient off
the commode chair. 8. When the patient is clear of the commode surface, use
the steering handles to
move the lift away from the commode chair. 9. To return the patient to bed,
reverse the procedures for lifting the patient,
operation, and sling attachment. 10. To return or place the patient in a
wheelchair, refer to 8.3 Transferring the Patient
to a Wheelchair, page 43.
41
Fig. 8-1
1145810-E
8 Patient Transferring
8.1.2 Transferring to a Standard Commode Although Invacare recommends two
assistants be used for all lifting preparation and transferring from and
transferring to procedures, the equipment will permit proper operation by one
assistant. The use of one assistant is based on the evaluation of the
healthcare professional for each individual case. The Invacare patient lift is
NOT intended as a transport device. If the bathroom facilities are NOT near
the bed or if the patient lift cannot be easily maneuvered toward the commode,
the patient MUST be transferred to a wheelchair and transported to the
bathroom facilities before using the patient lift again to position the
patient on a standard commode. Refer to 8.3 Transferring the Patient to a
Wheelchair, page 43. Before transferring the patient, the patient lift should
be guided to the bathroom facilities to ensure it can be easily maneuvered
toward the commode.
1. Before lifting the patient from the bed, refer to 7.1.2 Attaching Slings
to the Lift, page 35. For operation of the patient lift, refer to 6 Operation,
page 28.
2. Transport the patient to the bathroom facility. 3. Press the Boom Up/Down
buttons to elevate the patient high enough to clear the standard commode and
have his/her weight
supported by the patient lift. 4. With the help of both assistants, guide the
patient onto the standard commode. 5. Press the Boom Down button to lower the
patient onto the standard commode, and leave the sling attached
to the hanger bar hooks. 6. When complete, recheck for correct attachments.
7. Press the Boom Up button to raise the patient off the commode.
8. When the patient is clear of the standard commode surface, use the
steering handle to move the lift away from the standard commode.
1145810-E
42
Invacare® ReliantTM 450/600
9. To return or place the patient in a wheelchair, refer to 8.3 Transferring
the Patient to a Wheelchair, page 43. 10. To return the patient to bed,
reverse the procedures for patient lift, operation, and sling attachment.
8.2 Transferring the Patient to a Bathing Unit
There are many portable bathing apparatuses; this is an example of one. Refer
to your particular portable bath instructions and use them in conjunction with
this Owner’s Manual.
1. Lift the patient from the bed. Refer to 7.2 Lifting/Moving the Patient,
page 36.
2. Press the Boom Up/Down buttons to elevate the patient high enough to clear
the bed and be able to slide the portable bathtub under the patient.
3. Press the Boom Down button to lowerthe patient into the portable bathtub.
4. Detach the sling from the hanger bar hooks, and attach the portable bathtub
straps to the patient lift. 5. Press the Boom Up button to raise the sides of
the portable bath tub. 6. Bathe the patient. 7. Reverse the steps in this
procedure to return the patient to bed.
8.3 Transferring the Patient to a Wheelchair
1. Lift the patient from the bed. Refer to 7.2 Lifting/Moving the Patient,
page 36 2. Ensure the legs of the lift (with the patient) are in the open
position. Press the Legs Open button until the legs are in the
maximum open position. 3. Move the wheelchair into position.
43
1145810-E
8 Patient Transferring
4. Engage the rear wheel locks of the wheelchair to prevent movement of the
chair.
WARNING! -The wheelchair wheel locks MUST be in a locked position before
lowering the patient into the wheelchair for transport; otherwise, injury may
occur.
Use the straps or handles on the side and the back of the sling to guide the
patient’s hips into the seat as far back as possible for proper positioning.
5. Position the patient over the seat with his/her back against the back of
the chair.
6. Begin to lower the patient by opening the control valve or by pressing the
DOWN ( ) button. 7. With one assistant standing behind the chair and the other
operating the patient lift, the assistant behind the chair must pull
back on the grab handle (on select models) or sides of the sling to seat the
patient well into the back of the chair. This seating position helps to
maintain a good center of balance and prevent the chair from tipping forward.
8. Leave the sling in place.
Remove the sling only if using a divided-leg sling.
9. To return the patient to the seating surface, reverse the steps in this
procedure for patient lifting, operation and sling attachments.
WARNING! -Be sure to check the sling attachments each time the sling is
removed and replaced to ensure it is properly attached before the patient is
removed from the bed or chair.
1145810-E
44
Invacare® ReliantTM 450/600
Fig. 8-2
45
1145810-E
Troubleshooting
Troubleshooting
SYMPTOMS
Patient lift feels loose.
FAULTS
Mast/Base joint is loose. Tie-rods are loose.
Casters/Brakes noisy or stiff. Fluff or debris in bearings.
Noisy or dry sound from pivots.
Needs lubrication.
9 Troubleshooting
SOLUTION Refer to 5 Assembly, page 20. Refer to 10.7 Maintaining the Base
Adjustment, page 57. Refer to 10.8 Replacing the Casters and Forks, page 58.
Refer to 10.2 Lubricating the Lift, page 52.
1145810-E
46
Invacare® ReliantTM 450/600
SYMPTOMS
FAULTS Hand control or actuator connector is loose.
Battery low.
SOLUTION
Charge the batteries. Refer to 6.2 Mounting the Battery Charger Charging the Battery, page 32.
RED emergency stop button pressed IN.
Rotate the RED emergency stop button CLOCKWISE until it pops out.
Actuator fails to lift or legs fail to open when button is pressed.
Battery not connected properly to control box.
Reconnect the battery to the control box. Refer to 6.2 Mounting the Battery
Charger Charging the Battery, page 32.
Replace the battery pack. Refer to 6.2 Mounting the Battery Charger Charging
the Battery, page 32.
The connecting terminals are damaged.
Boom or leg actuator requires service or load is too high.
Refer to 10.5 Replacing the Boom Actuator, page 535.1.4 Installing the Leg Actuator to the Base, page 24 Contact your dealer.
Unusual noise from actuator.
Actuator is worn or damaged or spindle is bent.
Refer to 10.5 Replacing the Boom Actuator, page 535.1.4 Installing the Leg Actuator to the Base, page 24 Contact your dealer.
47
1145810-E
SYMPTOMS
Lift arms will not lower in uppermost position.
FAULTS
SOLUTION
Lift arms require a minimum weight load to lower from the uppermost position.
Pull down slightly on the lift arms.
9 Troubleshooting
Lift arms will not lower during a power retraction.
Shoulder bolt at the junction of the boom and mast may not be properly installed.
Refer to 10.5.1 Checking and Tightening the Mast Pivot Bolt, page 54.
If symptoms are not remedied by the suggested solutions, contact your dealer or Invacare.
1145810-E
48
Invacare® ReliantTM 450/600
10 Maintenance
10.1 Maintenance Safety Inspection Checklist
PART
TASK
Inspect for missing hardware.
Ensure the base opens/closes with ease.
Inspect the casters and axle bolts for tightness. CASTER BASE
Inspect the casters for smooth swivel and roll.
Inspect and clear the wheels of debris.
Inspect the pivot joints for wear.
MAST
Inspect the mastit MUST be securely assembled to the boom.
Inspect for bends or deflections.
Inspect the pivot joints for wear.
INSTITUTIONAL
INITIAL
MONTHLY
INSPECTION INSPECTION/
ADJUSTMENT
IN-HOME SIX-MONTH INSPECTION
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
49
1145810-E
PART
TASK
Check all hardware and the hanger bar supports.
10 Maintenance
INSTITUTIONAL
INITIAL
MONTHLY
INSPECTION INSPECTION/
ADJUSTMENT
IN-HOME SIX-MONTH INSPECTION
X
X
X
Inspect for bends or deflections.
X
X
X
Inspect bolted joints of boom for wear.
X
X
X
BOOM
Inspect to ensure the boom is centered between the base legs.
X
X
X
Inspect the mast pivot bolt under the rubber boot to ensure the bolt is tightly secured.
X
X
X
Inspect the pivot joints for wear.
X
X
X
Inspect the bolt/hooks for wear or damage.
X
X
X
HANGER BAR Inspect the sling hooks for wear or deflection.
X
X
X
Inspect the pivot joints for wear.
X
X
X
CLEANING
Clean as necessary.
X
X
X
1145810-E
50
Invacare® ReliantTM 450/600
PART
TASK
INSTITUTIONAL
INITIAL
MONTHLY
INSPECTION INSPECTION/
ADJUSTMENT
IN-HOME SIX-MONTH INSPECTION
CHECK ALL SLING ATTACHMENTS each time the lift is used to ensure proper connection and patient safety.
X
X
X
SLINGS AND
HARDWARE Inspect sling material for wear.
X
X
X
Inspect straps for wear.
X
X
X
The Invacare patient lift is designed to provide a maximum of safe, efficient, and satisfactory service with minimum care and maintenance. All parts of the Invacare lift are made of the best grades of steel, but metal-to-metal contact will cause wear after considerable use. There is no adjustment or maintenance of either the casters or brakes, other than cleaning, lubrication, and checking the axle and swivel bolts for tightness. Remove all debris and other substances from the wheel and swivel bearings. If any parts are worn, immediately replace these parts. If you question the safety of any part of the lift, immediately contact your dealer and advise him/her of the issue.
51
1145810-E
10.2 Lubricating the Lift
10 Maintenance
The Invacare lift is designed for minimum maintenance. However, a six-month
check and lubrication should ensure continued safety and reliability. Keep the
lift and slings clean and in good working order. Any defect should be noted
and reported to your dealer as soon as possible. The casters MUST swivel and
roll smoothly. A light grease (waterproof auto lubricant) may be applied to
the ball bearing swivel of the casters once a year. Apply more frequently if
the casters are exposed to extreme moist conditions. Refer to the FIGURE for
lubrication points. Lubricate all pivot points. Wipe all excess lubricant from
lift surface.
1. Hanger Bar
2. Boom Mounting Bracket
3. Boom/Mast Mount
4. Mast Mounting Bracket
Fig. 10-1
10.3 Detecting Wear and Damage
It is important to inspect all stressed parts, such as slings, the hanger bar,
and any pivot for slings for signs of cracking, fraying, deformation, or
deterioration. Replace any defective parts immediately and ensure the lift is
not used until repairs are made.
10.4 Cleaning the Sling and Lift
Laundering should always be done with dark colors. The sling should be washed
regularly in a water temperature not to exceed 180°F (82°C). Refer to the
sling user manual for complete cleaning instructions.
1145810-E
52
Invacare® ReliantTM 450/600
A soft cloth, dampened with water and a small amount of mild detergent, is all
that is needed to clean the patient lift. The lift can be cleaned with non-
abrasive cleaners.
10.5 Replacing the Boom Actuator
1. Remove the nut, washer, and shoulder bolt that secure the actuator to the
mast mounting bracket. 2. Rest the boom on your shoulder and remove the nut,
bolt, and pinch guard from the boom mounting bracket. 3. Remove the boom
actuator assembly. 4. Reverse the steps for installation.
CAUTION! DO NOT overtighten the nut and bolt. This damages the mounting
bracket.
53
1145810-E
10 Maintenance
Fig. 10-2
10.5.1 Checking and Tightening the Mast Pivot Bolt 1. Lift up the back of the
rubber boot and slide it off the mast along the boom. 2. Ensure the bolt is
through the bracket and the locknut is tight and secure.
1145810-E
54
Invacare® ReliantTM 450/600
3. If needed, do one or both of the following: l Tighten the locknut, and
back off the locknut 1/8 of a turn. l Replace the locknut.
4. Reposition the rubber boot.
Fig. 10-3
10.6 Replacing the Hanger Bar
WARNING! -After the first year of use, the hooks of the hanger bar and
mounting brackets of the boom should be inspected every six months to
determine the extent of wear. If these parts become worn, they must be
replaced.
55
1145810-E
1. Note the hanger bar comes attached to the boom with a bolt, nut, washers, and pinch guard. Refer to the images in this section for bolt and washer placement. Washers should be placed against both sides of the hanger bar bracket to eliminate damage. The hanger bar has three hooks on both ends to attach the sling that supports the patient during lifting.
10 Maintenance
2. Remove the existing hardware and replace the hanger bar. 3. Tighten the
locknut onto the bolt until the locknut bottoms out.
CAUTION! -DO NOT overtighten the nut and bolt. This damages the mounting
bracket.
Fig. 10-4
1145810-E
56
Invacare® ReliantTM 450/600
10.7 Maintaining the Base Adjustment
The base adjustment should not require any attention other than:
1. Check the squareness of the legs when in the closed position. 2. Place a
square on the inside of the legs and base to determine 90° alignment. 3.
Adjust the linkage rods until 90° alignment is achieved.
Fig. 10-5
57
1145810-E
10.8 Replacing the Casters and Forks
10.8.1 Replacing the Rear Casters
1. Place the lift on its side. 2. Remove the bolt and locknut that secure the
existing rear caster to the fork.
The bushing will be loose and may fall out of the caster. The existing bushing
will be reused. Examine and replace the bushing if it is worn.
3. Install the new/existing bushing into the new rear caster. 4. Align the
mounting holes in the new rear caster and the fork. 5. Install the bolt
through the fork and new rear caster and tighten securely with
the locknut.
10 Maintenance
Fig. 10-6
1145810-E
58
Invacare® ReliantTM 450/600
10.8.2 Replacing the Front Casters
1. Place the lift on its side. 2. Remove the bolt and locknut that secure the
existing front caster assembly to
the fork.
The front caster assembly consists of two casters and a washer in between.
The washer will fall out from between the two casters. The washer can be
reused. Examine and replace the washer if it is worn.
3. Position the new or existing washer between the two new casters. 4. Align
the mounting holes in the new front caster assembly and the fork. 5. Install
the bolt through the fork and the new front caster assembly and tighten
securely with the locknut.
Fig. 10-7
10.8.3 Replacing the Forks
1. Place the lift on its side. 2. Remove the front or rear caster from the
lift. Refer to 10.8 Replacing the Casters and Forks, page 58. 3. Unscrew the
existing fork from the base. 4. Install the new fork onto the base. 5. Install
the front or rear caster onto the lift. Refer to 10.8 Replacing the Casters
and Forks, page 58.
59
1145810-E
11 Accessories
11 Accessories
11.1 Reliant Scale RLS6
The Reliant Scale is a compact precision scale system designed specifically
for the Invacare patient lift system.
DANGER! Risk of Death, Injury or Damage -Improper use of this product may
cause death, injury or damage. -If you are unable to understand the warnings,
cautions, or instructions, contact a healthcare professional, dealer, or
technical personnel before attempting to use this equipment. -DO NOT use this
product or any available optional equipment without first completely reading
and understanding these instructions and any additional instructional material
such as user manuals, service manuals or instruction sheets supplied with this
product or optional equipment.
11.1.1 Removing the Hanger Bar
1. Remove the shoulder bolt, locknut, pinch guard, and washer that secure the
hanger bar to the hanger bar bracket.
Note the hanger bar hardware varies. For more information, see 10.6 Replacing
the Hanger Bar, page 55
Save the shoulder bolt to secure the Reliant Scale to the boom. The pinch
guard will not be used when the scale is installed. Save the pinch guard for
future use of the hanger bar without the scale.
Removing the hanger bar hardware will release the hanger bar, two nylon
washers and the hanger bar pin. Save the hanger bar pin and the two nylon
washers for future use of the hanger bar without the scale.
1145810-E
60
Invacare® ReliantTM 450/600
Fig. 11-1
61
1145810-E
11.1.2 Installing the Reliant Scale
WARNING! -The patient and sling MUST be removed from the lift during ALL
installation procedures.
1. Position the load cell assembly of the Reliant Scale into the hanger bar
bracket. Refer to Detail A. Use a 1/4-inch nylon washer.
11 Accessories
2. Secure the Reliant Scale to the hanger bar bracket with a new locknut and the hardware removed in 11.1.1 Removing the Hanger Bar, page 60. Refer to Detail A for washer orientation. Ensure the shaft of the shoulder bolt passes through both sides of the hanger bar bracket so the detent balls are visible on the other side of the bracket.
3. Tighten the locknut onto the shoulder bolt until the locknut bottoms out.
4. Insert the hanger bar pin with two nylon washers through the hanger bar.
Refer to Detail B. 5. Align the mounting holes in the hanger bar pin with the
mounting holes in the load cell assembly. Refer to Detail B. 6. Secure the
hanger bar pin to the load cell assembly with the provided mounting screw and
locknut. Securely tighten. Refer to
Detail B.
WARNING! -After ANY adjustments, repair or service and BEFORE use, make sure
all attaching hardware is securely tightened; otherwise, injury or damage may
occur.
1145810-E
62
Invacare® ReliantTM 450/600
7. Weigh the patient. Refer to 11.2 Operating the Scale, page 64 and 11.3
Weighing the Patient, page 65.
Fig. 11-2
63
1145810-E
11.2 Operating the Scale
11.2.1 Keypad Functions
Key
Indicator Indicator Location
ON OFF
OFF
Center of the Display Window
11 Accessories
Definition
Press this key to apply power to the scale and turn on the unit. When the
scale is on, press the button to turn it off.
ZERO
ZERO
Lower Left Corner of the Display Window
Press this key when the scale is on to reset the weight in the display window to zero.
UNITS
lb or kg
Upper Right Corner of the Press this key to change the unit of measurement from pounds (lb) to
Display Window
kilograms (kg).
LOCK UNLOCK
LOCK
Lower Right Corner of the Display Window
Press this key to lock or unlock a weight value in the display window. The weight displayed is a STORED weight and not the current weight when the key is used in the lock position.
N/A
LO BAT
Center of the Display Window
Note this indicator appears in the display window when the battery is low.
1145810-E
64
Invacare® ReliantTM 450/600
Fig. 11-3
11.3 Weighing the Patient
WARNING! -The weight capacity is limited to the lowest-rated capacity of any
one of the components in use (for example, the Patient Lift, Sling or Scale).
The patient’s weight MUST not exceed the lowest-rated capacity of any
component. -DO NOT operate key strokes with pointed objects (for example,
pencils, pens, or fingernails); otherwise, damage to the keypad will result.
1. Attach the sling straps to the hanger bar. For proper attachment
instructions, refer to 7.1.2 Attaching Slings to the Lift, page 35. For
improved accuracy on the 450 and 600 Reliant Lifts, attach all sling straps to
the center hooks on both sides of the hanger bar. However, patient comfort may
require a different configuration of the sling straps.
65
1145810-E
11 Accessories
2. Press the ON/OFF key.
The display will indicate the last weight that was measured. The word “LOCK”
displays in the display box.
Press the ZERO key to avoid capturing the weight of the sling and the
hardware. If the ZERO key is not pressed, the weight of the sling and the
weight of the hardware are included in the weight that displays. FOR AN
ACCURATE WEIGHT READING, PRESS THE ZERO KEY.
3. Press the ZERO key. The following occur: l The scale resets to zero and
the word “ZERO” displays. l The lock function turns off and the word “LOCK”
disappears from the display. The scale is now active and continually updates
the weight display.
4. Place the patient in the sling. For patient placement instructions, refer
to 6 Operation, page 28. 5. Activate the lift mechanism to raise the patient
until he/she is completely supported by the lift. For patient lifting
instructions,
refer to 7 Patient Lifting, page 34. 6. Note the weight display. 7. When the
weight display becomes stable, press the LOCK button to lock the weight
display. The word “LOCK” displays in the
window.
If it is necessary to unlock the weight while the patient is still supported
by the lift, the UNLOCK button may be pressed. The weight unlocks, and the
word “LOCK” disappears from the display. The weight value will then be
updated. Once the weight becomes stable, the weight can be locked again by
pressing the LOCK button.
Stable is defined as the weight fluctuating two tenths of a pound. For
example, for a patient weighing 100 pounds, the scale will fluctuate between
99.8 and 100.2 until the LOCK key is pressed. Weight display fluctuation is
normal, as
Press the UNITS button to toggle between pounds and kilograms. The selected
unit displays in the window.
1145810-E
66
Invacare® ReliantTM 450/600
8. Lower the lift and remove the sling from the patient.
The patient’s weight continues to display in the window. The display will turn
off after a two-minute period of non-use (no changes in weight exceeding five
pounds [two kilograms)]. The time delay for automatic shutoff CANNOT be
adjusted.
After the display turns off, press the ON/OFF button to recall the weight.
Press the ON/OFF button a second time to turn off the unit.
67
1145810-E
11 Accessories
11.4 Replacing the Battery
The scale is powered by a nine-volt alkaline battery that should provide
approximately 1500 readings before needing replacement.
When battery replacement is needed, LO BAT will appear on the display. To
replace the battery:
1. Slide the battery door open in the direction of the arrow. 2. Remove the
existing battery. 3. Install the new battery. 4. Reinstall the battery door.
Fig. 11-4
11.5 Calibrating the Reliant Scale
The Reliant Scale will be calibrated at the factory with the load cell. Should
it be necessary to recalibrate the scale, complete the following instructions.
1145810-E
68
Invacare® ReliantTM 450/600
1. Remove the patient and the sling from the scale to properly calibrate the
Reliant Scale. For patient removal instructions, refer to 6 Operation, page
28. For sling detachment instructions, refer to 7 Patient Lifting, page 34.
2. With the Reliant Scale on, remove the four screws on the back of the
enclosure and remove the front cover to expose the PC board.
3. Press the CAL button on the PC board. The CAL switch is not labeled but is
the only button on the PC board and is located in the lower right corner. Once
the CAL button is pressed, CAL1 displays in the window.
CAL1 selects the calibration mode using 50 pounds of calibrated weight.
CAL2 selects the calibration mode using 200 pounds of calibrated weight.
Press the UNITS key to toggle between CAL1 and CAL2.
4. When the desired calibration mode displays, press the LOCK/UNLOCK key.
UnLd displays in the window.
5. Ensure there is no load on the scale, and press the ZERO key. 6. Note a
dashed line (——) will scroll across the window and then the word LOAD will
display. 7. Perform one of the following:
l FOR CAL1 OPTIONsupport 50 pounds of calibrated weight from the scale and
press the ZERO key. l FOR CAL2 OPTIONsupport 200 pounds of calibrated weight
from the scale and press the ZERO key. 8. Note a dashed line (——) will scroll
across the window, and then the word DONE will display. 9. Remove the weight
from the scale and press the LOCK/UNLOCK key.
The unit enters a temporary test mode and does not lock the weight display.
This allows weight to be loaded and unloaded to check the calibration.
69
1145810-E
10. Turn off the unit by pressing the ON/OFF key.
11. Turn on the unit by pressing the ON/OFF key. The scale is now in normal
operation.
11 Accessories
11.6 Troubleshooting
SYMPTOM
Unit does NOT work properly.
Battery has been replaced and the unit still does NOT work properly.
11.6.1 Display Codes
CODE
DESCRIPTION
PROBABLE CAUSE
Battery failure
SOLUTION Check the battery. Replace if necessary.
Contact Invacare for service at 1-800-3336900.
Calibration Requiredindicates improper stored calibration data. Calibration is necessary.
Over Capacityindicates a weight exceeding the capacity has been loaded on the scale.
1145810-E
70
Notes
1145810-E
71
Canada United States:
Invacare Canada L. P. · 570 Matheson Blvd E Unit 8, CDN Mississauga, Ontario, L4Z 4G4 Canada · 800-668-5324 · www.invacare.ca Invacare Corporation · One Invacare Way, Elyria, Ohio, 44035 · Tel: 800 333 6900 · www.invacare.com
Invacare Corporation · One Invacare Way · Elyria, OH 44035, USA
1145810-E 2022-07-12