KaWe Eurolight C30 OP Otoscopes User Manual

June 12, 2024
KaWe

KaWe Eurolight C30 OP Otoscopes

Product Information

The KaWe EUROLIGHT, KaWe COMBILIGHT, and KaWe PICCOLIGHT are medical devices used for examination purposes. These devices are designed for professional use and come with various features and accessories.

Features

  • EUROLIGHT and COMBILIGHT models have interchangeable heads
  • PICCOLIGHT models require 2x Mignon Typ AA 1.5V batteries
  • The devices have a temperature limitation
  • Reusable ear funnels are sterilizable
  • Various accessories and spare parts are available

Intended Use

The devices are intended to be used by healthcare professionals for diagnostic examinations of the ear, nose, and throat. They provide illumination and magnification for accurate examination and diagnosis.

Product Usage Instructions

  1. Battery Installation:
    For EUROLIGHT and COMBILIGHT models:

    • Insert batteries or rechargeable battery into the handle.

    • Refer to Chapter 8 for battery replacement or recharge instructions.
      For PICCOLIGHT models:

    • Insert 2x Mignon Typ AA 1.5V batteries into the handle.

  2. Head Assembly:

    • For C30 / F.O.30 variants:
    • Attach and lock the head, ensuring it aligns with the notches.
  3. Lamp Replacement:

    • To replace the lamp:
      For EUROLIGHT / COMBILIGHT models:

    • a) Remove the lamp from the otoscope head.

    • For PICCOLIGHT models:

    • a) Swing the magnifying glass upwards and remove the lamp from the adapter.

    • b) Unscrew the lamp from the adapter.

    • c) Screw the new lamp onto the adapter.

    • d) Insert the lamp with adapter back into the holder until it clicks into place.

  4. Hygienic Preparation:

    • For cleaning and sterilization:
    • Cleaning: Manual or automatic cleaning methods can be used.
    • Sterilization: Only reusable ear funnels can be sterilized.
    • Relative Humidity: 30% – 75%
  5. Disposal:

    • Follow the local regulations for the disposal of the product and its components.
  6. Additional Information:

    • Refer to the User’s Manual for detailed information on symbols, safety guidelines, contraindications, preparation for use, and available accessories and spare parts.

Symbols used in this User’s Manual and on the product package

KaWe-Eurolight-C30-OP-Otoscopes-FIG-1 \(1\)Purpose and intended use

The otoscope with an attached ear funnel is designed for medical examination of the external auditory canal up to the ear drum. Following all instructions detailed herein, as well as all important guidelines and requirements also qualifi es as proper use. Any other use of the device is not its intended use, and the manufacturer disclaims any liability that may arise as a result of such unintended use. Otoscopes are used to diagnose and monitor diff erent diseases and functional disorders in the areas of the outer ear, auditory canal, ear drum and inner ear.

  • This device is recommended for use in adult patients, children and infants.
  • Only authorized and trained professionals may use the device!

Safety guidelines/contraindications

  • Check the device for proper functioning before each use. Do not use the device if the device is damaged.
  • Do not use the device without an ear funnel!
  • Using the device on damaged tissue is not recommended!
  • In case of malformations of outer or inner ear which may cause the examination to end in an injury, using the device is not allowed.
  • Choose the ear funnel size for each patient individually to avoid damaging the ear or the auditory canal.
  • Keep the device out of strong magnetic fields (MRI).
  • Do not shine the light from the device directly into eyes!
  • Like all electrical diagnostic devices, KaWe otoscopes are subject to special electromagnetic compatibility precautions.
  • Mobile high frequency telecommunications equipment may interfere with KaWe otoscopes’ function.
  • Reusable ear funnels shall be sterilized before use!
  • Use only original KaWe spare parts and components.

In the event of an error, increased heat build-up can occur, which can lead to the risk of burns. In case of intense heating, turn the handle off immediately. Disposable ear funnels are single-use only and to be disposed of immediately after use.

Assembly and service

EUROLIGHT / COMBILIGHT

  1. Insert the battery or the rechargeable battery into the handle. Check for correct polarity and for correct lamps (2.5 V or 3.5 V)! Also follow the instructions of Section 8: Maintenance and Service.

  2. C10 options: Push the head all the way down, hold it in place and tighten the screw. Check the screw groove on the head.
    C30 / F.O.30 options: Push the head all the way down, hold it in place and tighten the screw. Check the grooves!

  3. Turning the lamp on: Press on the latching switch and turn the illumination regulator to the left. To turn the lamp off, turn the regulator back to its starting position (0). The illumination intensity changes gradually.

  4. Battery/rechargeable battery change: see Section 8: Maintenance and Service.

  5. Note: You can charge rechargeable batteries (NiMH, Li-Ion) inside the handle using MedCharge 4000 charger. Please follow a separate User’s Manual for MedCharge 4000 Сharger.

PICCOLIGHT

  1. Insert batteries (2x Mignon or AA 1.5 V batteries) into the handle. Check the polarity! Also follow instructions in Section 8: Maintenance and Service.
  2. Screw the head onto the handle.
  3. Use the sliding switch to turn the lamp on or off.
  4. You can change batteries/rechargeable batteries by unscrewing the head from the handle. Also follow instructions in Section 8: Maintenance and Service.

Duration of operation
The devices are not intended for continuous operation, the maximum duration of one use is 1 minute with a 10 minutes pause between uses.

Preparation for use

  1. Check the assembled otoscope for damage, impurities and compatibility. Check that the moving parts (i.e. magnifying glass frame, plug junctions) move smoothly.
  2. Check that the lamp is well secured and functions properly.
  3. Check the battery charge. If all those checks have been successful, you can now use the otoscope to examine a patient after sanitizing it and attaching a new ear funnel.

Sanitary preparation

Disposable ear funnels are single-use only and therefore are not to be cleaned or re-used! Clean reusable ear funnels immediately after use to prevent impurities from drying up.

Manual/automated cleaning
Protect the device from moisture! Never allow the otoscope to come into contact with fluids and ensure that liquids do not penetrate the housing. The exterior surface of the device can be cleaned with a damp, soft and lint-free cloth. Simple cleaning and or disinfection with an alcohol-based surface disinfecting agent is sufficient.

  • Do not sterilize the device!
  • Do not use abrasive cleaning agents!
  • The magnifying glass is only to be cleaned with glass cleaner!

No automated cleaning possible/required!

Disinfection
The otoscope can be disinfected by wiping it with a cloth dampened with alcohol (e.g. isopropyl). Avoid moisture from getting on the device! Disinfect reusable ear funnels with a suitable disinfectant (i.e. Helipur®). Follow the concentrations and reaction times indicated by disinfectant manufacturer.

Sterilization
Only reusable ear funnels are to be sterilized! Proper steam sterilization can only be done after sufficient pre-cleaning!

  • Maximum hot steam sterilization temperature :…………….134 °C
  • Minimum hot steam sterilization time : …………………………3 Min.

Storage and transport

Store and transport the device in clean conditions free of dustand moisture.

Ambient temperature

  • Storage: -10 °C to +45 °C
  • Transport: -20 °C to +50 °C
  • Operation: +5 °C to +35°C
  • Relative air humidity: 30 % – 75 %

Maintenance and service

Any damaged parts (i.e. reusable ear funnels) must be discarded and replaced! In case the illumination intensity is low, check for the following causes:

  1. Check the battery/chargeable battery charge. If necessary, replace or recharge the battery.
  2. Clean or replace the lamp. If reasons 1 and 2 above are ruled out, it is possible that the device has to be repaired or replaced. Contact your supplier.

Battery change
Check the battery charge regularly. Reduced battery voltage (caused by wear) leads to diminishing illumination and in some cases to blinking LED lamps. In both cases replacing the batteries is recommended. Use high quality alkaline batteries or fully charged rechargeable batteries to replace the old ones. Please follow the User’s Manual for the rechargeable batteries or charger devices!

EUROLIGHT / COMBILIGHT

  • a) battery (2x Baby / C type batteries)
  • b) rechargeable battery

PICCOLIGHT

Lamp change
Satisfactory operation of the device with the best illumination power can only be guaranteed with the use of original KaWe replacement lamps.
EUROLIGHT F.O. / COMBILIGHT F.O. / PICCOLIGHT F.O

  1. Remove the head from the handle. To do that, press the corrugated bushing down.
  2. Pull the lamp from the bottom of the head.
  3. Clean the bulb of the replacement lamp with alcohol. The glass bulb should be clean and free of fingerprints (grease spots).
  4. Push the lamp all the way into the head.

PICCOLIGHT C

  1. Swing the magnifying glass up and grasping the adapter, pull out the lamp.
  2. Detach the lamp by unscrewing it from the adapter.
  3. Clean the bulb of the replacement lamp with alcohol. The glass bulb should be clean and free of fingerprints (grease spots).
  4. Screw the new lamp onto the adapter.
  5. Push the lamp together with the adapter fully back into the slot.

EUROLIGHT C / COMBILIGHT C

  1. Unscrew the lamp from the otoscope head.
  2. Remove the black plastic bushing from the lamp carefully.
  3. Clean the bulb of the replacement lamp with alcohol. The glass bulb should be clean and free of fingerprints (grease spots).
  4. Screw a replacement lamp into the head and cover it with plastic bushing.

Accessories and spare parts

  • Disposable ear funnels: anthracite, Ø 2.5 mm / 4.0 mm black, Ø 2.5 mm / 4.0 mm

  • Reusable ear funnels: black, Ø 2.0 mm / 2.5 mm / 3.0 mm / 3.5 mm / 4.0 mm / 4.5 mm / 5.0 mm / 9.0 mm*

  • Only for EUROLIGHT C

  • Lamp support 2.5 V
    Replace the lamp the same way it is done with EURO-/ COMBILIGHT C.

  • Set of laryngeal mirrors

  • Tongue depressor holder

  • Nose spreader speculum
    Turn the regulator screw to open or close.

  • Ball with adapter for pneumatic tests
    Adapter is required only for EURO-/COMBILIGHT F.O. and PICCOLIGHT models.

  • Nose light Replace the lamp the same way it is done with EURO-/ COMBILIGHT F.O.

  • Finoff transilluminator, XL
    Replace the lamp the same way it is done with EURO-/ COMBILIGHT F.O.

  • Laryngeal mirror, straight Replace the lamp by removing the mirror head and unscrewing the lamp.

You can find other accessories and spare parts in our product catalogue at https://www.kawemed.com. or ask your supplier
Disposal: Dispose of the device according to local regulations. Icons on the device and the package contain information on disposal requirements. Defective and/or recyclable electrical / electronic devices and rechargeable batteries must be transferred to a designated waste collection point for recycling.

User and patient instructions

Please inform the manufacturer and the competent authorities of the relevant member state where the user and/or patient is located immediately about any serious product-related incidents.

Annex

Warranty: The warranty for LED lamps is fi ve years after the sale, provided that the device was operated as intended, according to this User’s Manual and using original KaWe spare parts. In other cases, the warranty is two years as provided for by law (except for illumination sources, batteries and rechargeable batteries). If you have any questions, or your device requires repair, please contact your supplier.
Contact details: Please contact your supplier or the manufacturer (the address is provided on the back of the page).
Basic UDI-DI: 4030155KaWe0101XR

CONTACT

  • KIRCHNER & WILHELM GmbH + Co. KG
  • Eberhardstraße 56
  • 71679 Asperg,
  • Germany
  • Phone: +49 7141 68188-0
  • Fax: +49 7141 68188-11
  • e-mail: info@kawemed.de.
  • www.kawemed.com.

All information is without guarantee and subject to change.