ottobock 501D1 Ball Catch Instruction Manual

June 12, 2024
ottobock

ottobock 501D1 Ball Catch - Cover

501D1
Assembly instructions

501D1 Ball Catch

ottobock 501D1 Ball Catch - Product Overview 1

Drill whole measurement

Size a [mm] b [mm]
17K29=4 16 10
17K29=5, *=6 21 7

Foreword

INFORMATION
Date of last update: 2021-12-08

► Please read this document carefully before using the product and observe the safety notices.
► Instruct the user in the safe use of the product.
► Please contact the manufacturer if you have questions about the product or in case of problems.
► Report each serious incident related to the product to the manufacturer and to the relevant authority in your country. This is particularly import ant when there is a decline in the health state.
► Please keep this document for your records.
These instructions for use provide important information on the processing of the 501D1 spring-loaded thrust piece.

Product description

Components (see fig. 1)

Item Designation Article number
1 Clamping sleeve 506S1=3×14
2 Spring-loaded thrust piece 501D1
3 Orthotic knee joint 17K29*

2.1 Available sizes
The 501D1 spring-loaded thrust piece made of stainless steel is available in one size.

2.2 Components
Scope of delivery

Item Piece(s) Designation Reference number
not illus­trated 1 Spring-loaded thrust piece 501D1

Intended use

3.1 Indications for use
The product is intended exclusively for orthotic fittings of the lower limbs.

3.2 Combination possibilities
The 501D1 spring-loaded thrust piece can be combined with the 17K29=4, =5 and =6 orthotic knee joints.

3.3 Lifetime
The product is designed for a lifetime of 3 years when used as intended and assembled professionally.

3.4 Qualification
Patients may be fitted with the product only by trained qualified personnel. The qualified personnel must be familiar with the handling of the various techniques, materials, machines and tools.

Safety

4.1 Explanation of warning symbols

Warning regarding possible risks of accident or injury.

4.2 Safety instructions

Incorrect alignment, assembly or adjustment Risk of injury due to change in or loss of functionality
► Assembly, adjustment and maintenance operations may only be com pleted by qualified personnel.
► Observe the alignment, assembly and adjustment instructions.

Preparing the product for use

Preparing the threaded bore
> Required tools: 724S6=3.2 drill, 726W8 counterbore, 731B21=M4 screw tap.

  1. Remove the clamping sleeve from the upper joint section and take off the ring lock.
  2. Determine the position of the bore according to the table and mark it on the bar (see page 2).
  3. Drill the Ø 3.2 mm core hole.
  4. Deburr the bore with the 726W8 counterbore.
  5. Cut the M4 thread.

Installing the product

  1. Screw the spring-loaded thrust piece into the thread and secure it with Loctite 241. The thrust piece has to be flush with the surface of the bar.
  2. Slide on the ring lock.
  3. Check the function of the thrust piece and readjust if needed.
  4. Insert the clamping sleeve.

Cleaning

Promptly clean the product after contact with water containing salt, chlorine or soap, or if it gets dirty.

  1. Rinse the product with clear fresh water.
  2. Dry the product with a cloth or allow it to air dry. Avoid exposure to direct heat (e. g. from an oven or radiator).

Maintenance

INFORMATION
The product may be exposed to increased loads by the patient.
► Shorten the maintenance intervals according to the expected loads.

Disposal

Dispose of the product in accordance with national regulations.

Legal information

All legal conditions are subject to the respective national laws of the country of use and may vary accordingly.

9.1 Liability
The manufacturer will only assume liability if the product is used in accord acne with the descriptions and instructions provided in this document. The manufacturer will not assume liability for damage caused by disregarding the information in this document, particularly due to improper use or unauthored iced modification of the product.

9.2 CE conformity
The product meets the requirements of Regulation (EU) 2017/745 on medic al devices. The CE declaration of conformity can be downloaded from the manufacturer’s website.

Ottobock SE & Co. KGaA
Max-Näder-Straße 15 · 37115 Duderstadt · Germany
T +49 5527 848-0 · F +49 5527 848-3360
healthcare@ottobock.de · www.ottobock.com

© Ottobock · 647G95=all_INT-17-2203