DENTALEZ Lubricated Prophy Angle
Instruction Manual
NON-STERILE
INDICATIONS FOR USE
Used intraorally by trained dental professionals for drilling and preparation
of dental cavities for restoration, such as fillings, disking, cavity and
crown preps, polishing, post and pin drilling and pin setting.
Caution: U.S. Federal law restricts this device to sale by or on the order of
a licensed physician.
DO NOT USE FOR ORAL SURGERY
Before operating the handpiece, carefully read and follow all of these
instructions. Save these instructions for future reference and observe all
warnings and cautions.
WARNING Under normal operating conditions, air is exhausted from the back
of the head of this handpiece. If air is directed into an open soft tissue
wound or beneath the mucosa or dermis, injury to the patient could result from
air emphysema of air embolism.
WARNING This angle has no internal air or water cooling capabilities.
External water for cooling maybe required.
FEATURES
Rated for use with Star Motors and Attachments Titan Five Star Prophy Angle:
Star Titan 3
Part #265803
Controlled Speed Reduces Prophy Paste ‘Spin-off’ Field Replaceable Cartridge
Quick, Easy Cleaning Compatible with Titan 3 and Titan T Systems
Steam Sterilization: 135° C
Serializable up to 135°C
Motor | Accessories | Ratio | Max Allowable RPM |
---|---|---|---|
5K | M TA | 1:01 | 5,000 |
20K | Torque Multiplier & MTA | 4:01 | 5,000 |
INSTALLATION AND OPERATION
PROPHY ANGLE INSTALLATION FOR TITAN 3 OR TITAN T SYSTEMS
- For the Titan 3 or Titan T systems thread angle onto motor-tangle adapter (FIG. 1), Star Titan 3 Part #264073 or Titan T Part # 263970, by hand. Seat completely. CAUTION When angle does not seat on adapter, gears are not meshed correctly. Remove angle; spin motor-to-angle adapter drive gear and repeat step one.
- Fit angle disconnect wrench (FIG. 2), Star Part #203048, furnished with every angle, to mating indentations on angle neck; tighten securely.
PROPHY CUP INSTALLATION
The prophy angle uses either a screw-on or a snap-on type polishing cup. When
using a snap-on type cup, use the screw-to-button adapter, Star Part #258950.
Thread the screw-to-button adapter (FIG. 3), or polishing cup, with a
clockwise rotation until snug into the hollow cartridge shaft of the prophy
angle.
Operate the Titan motor in the forward direction only. If the Titan motor is
operated in reverse direction, the prophy cup will separate from the motor.
PROPHY CUP REMOVAL
Unthread the adapter, or polishing cup, counterclockwise. If the screw-to-
button adapter is difficult to disengage, remove the prophy cup, and use a
small screwdriver to loosen the screw-to-button adapter.
SAFETY INSTRUCTIONS
PLEASE OBSERVE SAFETY INSTRUCTIONS
Before use, verify concentric rotation by operating the handpiece outside the
oral cavity in a safe place.
LUBRICATING
- Lubricate with Dental Lube II, Star Part #262539 once a day or before each sterilization procedure.
- Remove prophy cup.
- Place a drop of Dental Lube II down the center of the hollow cartridge shaft (allow a few seconds for the lubricant to drain down the shaft). (See FIG. 4)
- Place a drop of Dental Lube II at the base of the angle. Allow a few seconds for the lubricant to drain and reach the bronze bushing. (See FIG. 5)
CLEANING AND STERILIZATION
CAUTION| It is the responsibility of the user to use only FDA
compliant sterilizing equipment and accessories that are capable of meeting
the manufacturers recommendations below.
---|---
CAUTION| Do not exceed 135°C
Do not use :
• dry heat sterilizer
• cold sterilizers
• ultrasonic cleaners
•immersion techniques
• sterilizer bags containing disinfectants
• detergents
• surface wipes
• surface sprays
• chemical disinfectors
• hand soaps
• surgical milk
LIMITS ON REPROCESSING| Repeated processing has minimal effect on these
Adapters. End of life is normally determined by wear and damage due to use.
INSTRUCTIONS
WARNING| The angle must be cleaned and sterilized before first use and
after each patient use.
CONTAINMENT AND TRANSPORTATION:| It is recommended that angles are
reprocessed as soon as is reasonably practical following use.
PREPARATION FOR CLEANING:| Wear puncture resistant gloves to prepare the
prophy angle for sterilization. Remove prophy cup or screw-to-button adapter
from prophy angle. Remove angel from motor-to-angle adaptor.
CLEANING AUTOMATED:| Washer-disinfectors are not recommended.
CLEANING MANUAL:| 1. Clean and inspect once a day.| Old Style:
-No Markings on End Cap
– Unthread End Cap in
Counter Clockwise Direction| New Style:
-c◀ ▶34o and L
Marked On End Cap
– Unthread End Cap in Clockwise Direction
---|---
2. Clean all external surfaces (FIG.6) with isopropyl alcohol and a lint-
free cloth or brush.
3. Use the plastic wrench to unthread the end cap by turning clockwise. (See
FIG.7):
4. Remove the cartridge and inspect as follows:5. If prophy paste is not found on thecartridge, internal cleaning
is not necessary. Simply wipe off with lint-free cloth.
6. If prophy paste is found on the cartridge, clean off the paste manually.
The gear assembly and end cap may be ultrasonically cleaned (FIG.8).
NOTICE
Do not ultrasonically clean any other components.
CLEANING MANUAL CONTINUED:| 7. When prophy paste is consistently found in the cartridge, install a new bronze bushing, or a new cartridge, Star Part
258948, which includes the bronze bushing. Replacing the bronze bushing
renews the seal and protects the gear and bearing from abrasive paste.
8. Wipe the inside of the housing with a lint-free cloth or cotton swab.
Note: Do not immerse head in solvent.
9. After cleaning, apply one drop of Denta Lube II, Star Part #262539, to the
inside diameter of the O-ring and to the gear as shown. (See FIG. 9) 10. To
reassemble, insert gear assembly into bronze bushing and guide shaft through
the hole in O-ring. Install cartridge into prophy head angle.
11. Turn end cap counterclockwise and hand tighten to secure. Use wrench to
fully seat – 30° rotation maximum.
NOTICE Do not overtighten as poor performance and damage to angle may
result
---|---
LUBRICATION:| Lubricate – See ‘Lubrication’ instruction section.
DISINFECTION:| Use steam sterilization exclusively for complete sterility as
outlined below.
MAINTENANCE:| Angle should be routinely maintained, checked for corrosion,
worn parts, etc. and should be replaced on an as needed basis.
INSPECTION & FUNCTION TESTING:| See Installation, Operating Recommendations,
and/or Troubleshooting sections of this manual.
PACKAGING:| Single: A standard sterilization packaging material must be used.
Ensure that the pack is large enough to contain the item without stressing the
seals.
WRAPPED INSTRUMENT STERILIZATION:| Gravity Type Autoclave:
Autoclave:
• Minimum of 15 minutes at 132° C
• Minimum of 30 minutes at 121° C
Pre-Vacuum
• Minimum of 4
Do not exceed 135° C
DRYING:| Dry the prophy angle for 30 minutes minimum using the sterilization
dry cycle and allow the prophy angle to cool to room temperature in bag prior
to use. If moisture is present, increase your sterilizer dry cycle
appropriately.
STORAGE:| Store in bag after sterilization until ready to use.
ADDITIONAL INFORMATION:| Do not exceed the sterilizer manufacturer’s
guidelines for load capacity. Use only paper or paper/plastic bags.
MANUFACTURER CONTACT:| See first or last panels of this manual.
TROUBLESHOOTING
PROBLEM: LOW POWER
Cause: Air leak at motor
Correction: Replace gasket; refer to Titan Motor Instruction Manual.
Cause: Debris build-up
Correction: Refer to Cleaning and Lubricating section.
Cause: Faulty angle
Correction: Replace cartridge.
Note: If a problem persists, contact dealer for further assistance; or,
contact Star technical service at: 1-866-DTE-INFO
(1-866-383-4636)
CARTRIDGE REPLACEMENT
-
Remove the end cap using the plastic wrench to turn it clockwise. (See illustration under Cleaning and Lubricating.)
Note: Do not use pliers on the bronze bushing. -
Pull bronze bushing to remove cartridge from prophy head as shown below in (FIG.10). If cartridge is not removable by hand, thread a prophy cup or a screw-to-button adapter in the cartridge and tug on it.
-
Separate gear assembly from the bronze bushing by pulling apart as shown above.
-
If gears are worn, replace cartridge, Star Part #258948. If not, simply wipe clean with a lint free cloth.
-
Wipe the inside of the housing with a lint free cloth or cotton swab, barely dampened with solvent.
Note: Do not immerse head in solvent. -
Guide the shaft of the gear assembly through the center of the bronze bushing.
Install the assembled cartridge into the angle housing. -
Position end cap and turn counterclockwise. Hand tighten to secure. Use wrench to fully seat – 30° maximum.
NOTICE
Do not overtighten as poor performance and damage to angle may result.
REPLACEMENT PART NUMBERS
ALWAYS INSIST ON STAR PARTS FOR BEST PERFORMANCE.
DESCRIPTION | ITEM NUMBER |
---|
Titan 3 Five Star Prophy Angle Cartridge Lubricated (Includes cartridge and
bronze bushing with O-ring)| 258948
Transmission Gear Assembly Lubricated| 259008
Prophy Angle End Caps (Package of 2) Lubricated| 265805
Screw-to-Button Adapter| 258950
Angle Disconnect Wrench| 203048
O-Rings (Package of 3)| 258951
Denta Lube II| 262539
Titan 3 Motor-to-Angle Adapter| 264073
Titan T Motor-to-Angle Adapter| 263970
IMPORTANT NOTICE ABOUT RETURNING PRODUCT TO DENTALEZ INC.
Anyone wishing to return any DENTALEZ Inc. product to a DENTALEZ facility MUST
obtain the proper RETURN AUTHORIZATION NUMBER from the Customer Service
Department. The RA Number must appear clearly marked on both the outside and
inside of the returned product carton. When calling for a return authorization
number you will be asked to provide the following information:
To Obtain A Return Authorization Number Call 1-866-DTE-INFO| 1. Product Name,
Model Number or Part Number
2. Serial Number of Product
3. Dealer Return Purchase Order Number
4. Reason For Return
5. Copy of Invoice Showing Purchase
---|---
Items will be evaluated. Return Authorization Number is not a guarantee for
credit. Your cooperation with this procedure will help us expedite processing
of any returns.
FOR NEW PRODUCTS, MANUALS AND TECHNICAL INFORMATION, CALL 1-866-DTE-INFO OR
VISIT OUR WEB SITE @ WWW.DENTALEZ.COM
LIMITED WARRANTY
O-Rings and Seals are specifically excluded from this warranty.
DENTALEZ warrants the Titan Lubricated Five Star Prophy Angle to be free of
defects in material and workmanship, under normal usage, under the following
terms:
Star Products: | *Warranty Period:** |
---|---|
Lubricated Five Star Prophy Angle | 3 Months from date of purchase |
Please note the following additional terms of our warranty and return policy:
-
Warranty covers manufacturing defects only and does not cover defects resulting from abuse, accidents, misuse, improper handling, cleaning, care or maintenance or the failure to follow or observe our operating/use, maintenance and/or installation instructions. Damage resulting from the use of chemicals, cleaners, disinfectants or sterilization is not covered under this warranty nor is normal wear and tear. Failure to use authorized parts or an authorized repair facility voids this warranty.
-
Liability is limited to repair or replacement of the defective part(s) or products at our sole discretion.
All other liabilities, in particular liability for damages, including, without limitation, special, exemplary, consequential or incidental damages are excluded -
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NO EMPLOYEE, REPRESENTATIVE OR DEALER IS AUTHORIZED TO CHANGE THIS WARRANTY IN ANY WAY OR TO GRANT ANY OTHER WARRANTY.
*Provided conditions in warranty are met.
WARRANTY REGISTRATION
Please visit the website below to register your Star product. This will ensure
proper registration and prompt and accurate service should any of your
purchases need evaluation or repair during the warranty period. Star Products:
WWW.DENTALEZ.com/warranty
**** DENTALEZ, Inc., Star Dental® Division
1816 Colonial Village Lane
Lancaster, PA 17601 USA
058952 Rev U Date of Issue: 2022-11-01 ©2022 DENTALEZ, Inc.