Digi-Pas DWL9000XY Digias 2 Axis Ultra Precision Inclinometer
Product Information
The DWL9000XY is a 2-Axis Ultra Precision Inclinometer manufactured by Digipas Technologies Inc. It is designed to measure the tilt or inclination of an object with high accuracy and precision. The device features a colour TFT LCD display, a measuring range of 0.2 arcsec. in single axis mode and 25 sec. in 2-axis mode, and a resolution of 2.0g. It is powered by 4 AAA 1.5V batteries or can be connected to a USB power source. The inclinometer is made of PC ABS and alloy steel materials and has connectivity options such as USB and Bluetooth. The package includes a 5-meter USB 2.0 cable and comes with PC SYNC software for data synchronization. The device operates within a temperature range specified by the manufacturer and has a dimension of 188 x 62 x 37 mm with a nett weight of approximately 1150 grams.
Product Usage Instructions
- Setting Up: To set up the DWL9000XY inclinometer, follow the operation procedure provided in the instruction manual. Mount the device securely on a fixture or workpiece using the mounting instructions provided.
- Storage & Cleaning: When not in use, store the inclinometer in a safe and dry place. Avoid exposing it to extreme temperatures or humidity. To clean the device, use a soft cloth or brush to remove any dirt or debris. Do not use solvents or harsh chemicals that may damage the device.
- Warranty: Refer to the warranty section in the instruction manual for information regarding product warranty and support.
- User Calibration: For user calibration instructions, refer to Appendix 1 in the instruction manual.
Note: Always refer to the latest version of the instruction manual available on the Digi-Pas website for the most up-to-date information and instructions.
INTELLECTUAL PROPERTY
This manual contains propriety information, which is protected by copyrights.
All rights are re-served. No part of this manual may be photocopied,
reproduced, redistributed or translated to another language without the prior
written consent of Digipas Technologies Inc.
The information in this manual was correct at the time of printing, stored in CD or uploaded in the Company website. However, Digipas Technologies Inc. will continue to improve products and reserves the rights to change specification and maintenance procedures at any time without prior notice.
Digi-Pas® Products were manufactured under ISO9001 & ISO14001 standards, tested to comply by the followings certification bodies:
DEVICE OVERVIEW
Technical Specification
Measuring Range (Single Axis Mode)| 0 to 7200 arcsec.
(0.0000° to ± 2.0000°)
---|---
Measuring Range (2-Axis Mode)| 0 to 3600 arcsec.
(0.0000° to ± 1.0000°)
Resolution| 0.2 arcsec. (≤ 1μm/M)
Accuracy| ± 1 arcsec at 0 to ±100 arcsec.
± 1.2 arcsec at other angles
Repeatability| 0.2 arcsec. (≤ 1μm/M)
Measurement Speed| ≤ 25 Sec.
Vibrometer| 2.0g
Display| Colour TFT LCD
Power Supply| 4 x AAA 1.5V Batteries / USB Power source
Material| PC ABS / Alloy Steel
Connectivity| USB 2.0 Cable (≤ 5 metre) Bluetooth (≤30 metre)
PC SYNC Software| Professional Edition (Included)
Operating Temperature| +10°C to +40°C
(Calibrated for the entire temperature range)
Storage Temperature| -20°C to +60°C
Dimension (mm)| 188 x 62 x 37
Net Weight (Approx.)| 1150 gram
Table 1. Technical specification of DWL-9000XY Ultra Precision Inclinometer
Notes:
- Product performance to specification complies with accredited Calibration & Test Providers in USA, Japan, UK, and Germany to conform with NIST, JIS, UKAS & DIN under the International Laboratory Accreditation Cooperation (ILAC) and American Association for Laboratory Accreditation (A2LA). For more information, please visit Z “www.digipas.com”.
Dimension of the Tilt Sensor Module
Overview
List of items included in DWL-9000XY product package
Item No. | Description | Quantity |
---|---|---|
1 | DWL 9000XY Ultra Precision Inclinometer | 1 unit |
2 | Certificate of Calibration | 1 set |
3 | AAA batteries | 4 pcs |
4 | USB 2.0 cable (3 Meters) | 1 pc |
5 | Security dongle | 1 pc |
6 | M5 Button head screws | 3 pcs |
Table 1. Packaging content
SETTING UP
Operation Procedure
- Insert 4 pieces of “AAA” batteries into the battery compartment and press ON/OFF but- ton. Alternatively, insert USB power source to the USB Port to power up the device. Take note that the device performance might be affected when poorly regulated USB power source is used.
- Initial Setup screen pops up. Press ON/OFF button to scroll up or MODE button to scroll down for selecting “Default Units” or “Current Device Location”. Press “UNITS” or “HOLD” button to input the selected option. Select the nearest option when your location is not available in the default list. Scroll to select “Done” to continue.
- Allow sufficient time for the device to stabilize during initial power-up. When the thermometer on the screen turns green, the device is ready to use.
Note: For maximum accuracy, perform calibration (refer to APPENDIX: User Calibration) or Absolute Level (refer to Chapter 4: Absolute Level for more details) before measurement.
Mounting Device On Fixture or Work Piece
The two threaded holes are provided for mounting the device onto user-defined
fixtures/machinery.
To mount the device onto user’s defined fixture or workpiece. Remove both the set screw (M5), then replace with the appropriate type of screws specified by the user.
USER INTERFACE
Single Axis Mode Interface
Single Axis Mode Display Screen and Button Function
Single Axis Mode Operation
Place the digital level on the surface to be measured.
Please ensure the contact surfaces of the device and measuring plane are clean
and free from dust particles.
The green arrow sign on display indicates the higher side.
User may “freeze” the screen by pressing the HOLD button.
The icon pops up to indicate the screen is paused. To resume operation, press
the hold button once.
Dual-Axis Mode Interface
Dual-Axis Mode Display Screen and Button Function
Place the digital level on the surface to be measured.
Please ensure the contact surfaces of the device and measuring plane are clean
and free from dust particles.
The “Target Ball” or “Bull Eye” move towards the measured position similar to traditional “Bubble” vials. The green arrow sign on display indicates the higher side of plane.
Once the measurement stabilizes, the “Target Ball” or “Bull Eye” blinks.
Main Menu Icon Screen Display and Button Function
Main Menu Icon Features
ABSOLUTE LEVEL SETTING
Enable user to ensure each measurement reading is in accordance to maximum
device accuracy specified.
ALTERNATE ZERO SETTING
Enable user to measure relative angles at a common plane with respect to a
reference angle. Set any angle to 0” as a reference.
SYNC MENU
Enables user to select the connectivity (USB/Bluetooth) with a computer
installed with PC Sync Software.
ANGLE METER
Enables user to obtain real-time continuous angle measurements displayed in
line graph.
VIBRO METER
Enables user to obtain real-time continuous vibration measurements displayed
in line graph.
SETTING MENU
Enable user to modify various parameters of the device.
FEATURES AND SETTING
Absolute Level Screen Display and Button Function
Place the device on the surface to be measured. Press MODE button to start the
measurement and wait until the loading bar is full.
Please ensure the contact surfaces of the device and measuring plane are clean
and free from dust particles.
Note: The device is able to auto detect its position is single axis
position or dual-axis.
Turn the device 180° and press MODE button again to start the measurement.
When completed the above settings, the ‘ABS’ logo is shown to indicate that the device is in the Absolute Level mode.
Alternate Zero Screen Display and Button Function
Place the digital level on the surface to be measured. Press MODE button to
set the angle to 0” as a reference.
Please ensure the contact surfaces of the device and measuring plane are clean
and free from dust particles.
Note: The device is able to auto detect its position is single axis
position or 2-axis.
The ‘Alt. Zero’ logo is shown to indicate that the device is in Alternate Zero mode.
User may “freeze” the screen by pressing the HOLD button. The icon pops up to indicate the screen is paused. To resume operation, press the hold button once.
SYNC Menu Screen Display and Button Function
The status bar indicates current selected mode as USB. To change to Blue-tooth mode, scroll the ‘right’ button and then press MODE button.
Reboot the device by turning off the device and then turn on the device again to activate the new setting.
Once Bluetooth mode is activated, the Bluetooth Icon will appear on the top right display.
Angle Meter Screen Display and Button Function
Place the digital level on the surface to be measured.
Please ensure the contact surfaces of the device and measuring plane are clean
and free from dust particles.
The Angle Meter displays real-time continuous angle measurements in line graph.
User may “freeze” the screen by pressing the HOLD button.
The icon pops up to indicate the screen is paused. To resume operation, press
the hold button once.
Vibro Meter Screen Display and Button Function
Place the device on the surface to be measured.
Press Auto Set button to set the line graph to the centre of display.
Please ensure the contact surfaces of the device and measuring plane are clean
and free from dust particles.
When the device detects the vibration, the relative magnitude (g) is shown on the graph.
To change the scale and speed of the line graph, press the scale selection button or the speed selection button accordingly.
Setting Menu Display Screen and Button Function
-
Units
To change the measuring unit (arcsec., mm/M, mil(NATO) & mrad). -
Current Device Location
When the device detects the vibration, the relative magnitude (g) is shown on the graph. -
AutoOff
-
To set automatic power off according to user defined time period.
-
AutoDim
To set automatic dim according to user defined time period. -
Buzzer
To turn on/off the device buzzer. -
Display Brightness
To set LCD brightness according to user defined level.
STORAGE AND CLEANING
Storage
Keep the device in the equipment box and maintain the storage temperature
within -20°C to 60°C or -4°F to 140°F.
When the device is not in used, the batteries are to be removed from the
device.
Cleaning
- Keep the device dry and clean. Remove any moisture or dirt with a soft dry cloth before measurement to obtain the maximum accuracy. Do not use harsh chemicals, strong de-tergents or cleaning solvent to clean the device.
- Do not submerge device in liquid while cleaning.
WARRANTY
Digi-Pas® 2-Axis Ultra Precision Inclinometer is warranted to the original purchaser to be free from defects in workmanship and material. Digipas Technologies Inc. will, at its option, repair or replace any defective part which may malfunction under normal and proper use period of 2 (two) of pur- chase. The foregoing warranty shall not apply to defects resulting from misuse, abuse, assignment, or transfer by the Buyer. or interfacing, outside of environment the product. Digipas Technologies Inc. that the operation of instrument software, or firmware, will be uninterrupted or error-free. The exclusive remedy under any and all warrants and guarantees expressed herein, and we shall not be liable for damages from loss or delay of equipment uses, consequential, or incidental dam-age. No other Warranty is expressed or implied. Digipas Technologies Inc. specifically disclaims the implied warranties of merchantability and fitness for a particular purpose.
Limitation of Liability:
In no event shall DIGIPAS TECHNOLOGIES INC., (hereinafter, “the Company”) be
liable to you or any third party for any indirect, special, incidental,
exemplary or consequential damages of any kind resulting from any breach of
warranty or from the performance or use of the product. This includes without
limitation: property damage, loss of value of the product or any third party
prod-ucts that are used with the product, or loss of use of the product or any
third party products that are used with the product, even if the Company has
been advised of the possibility of such loss or damages. The Company’s total
cumulative liability arising from or related to the product, whether in
contract, tort (including negligence) or otherwise, shall not exceed the
amount actually paid by you for the product. Some states and/or jurisdictions
do not allow the exclusion or limitation of incidental or consequential
damages, so the above limitations or exclusions may not apply to you. If any
limitation of liability shall be deemed invalid by any applicable law, then
the limitations of liability set forth above shall apply to the maximum extent
permitted under applicable law.
Manufactured by: JSB TECH PTE LTD
Email: info@digipas.com
www.digipas.com
APPENDIX 1: USER CALIBRATION
Calibration Instruments:
- Granite Table:
Grade AA (Levelled to ≤0.5 arcsec.)
Calibration Procedure:
Ensure the DWL9000XY device is in power OFF condition. Press and hold the CAL
button by using a small pin and simultaneously press the ON/OFF button.
The LCD screen displays “Calibration 1”.
Place the device to Position 1. Press the MODE button once and wait until the countdown reach “0”, the LCD screen displays “Calibration 2”.
Turn the device 180° to Position 2. Press the MODE button again to start the calibration at “Position 2” and wait until the countdown reach “0”, the LCD screen displays “Cali-bration 3”.
Place the device to Position 3. Press the MODE button once and wait until the countdown reach “0”, the LCD screen display “Calibration 4”.
Turn the device 180° to Position 4. Press the MODE button again and wait until the countdown reach “0”, the LCD screen will switch to measuring mode once calibration is completed.
Precaution:
To achieve maximum accuracy, the device must be held firmly on to jig during
calibration process. Any movement on countdown during each calibration
procedure would affect device accuracy.
© 2021 Digipas Technologies Inc. All Rights Reserved.